FCC_SAR_Appendix_A

FCC ID: YHLBLUJENNYSLD

RF Exposure Info

Download: PDF
FCCID_4079688

                                                                             Date: 2018-10-31

Test Laboratory: UnionTrust

System Check_H835

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium parameters used: f = 835 MHz; σ = 0.918 mho/m; εr = 42.9; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY4 (High Precision Assessment)

DASY4 Configuration:

      Probe: ES3DV3 - SN3089; ConvF(6.14, 6.14, 6.14); Calibrated: 2018-5-18
      Sensor-Surface: 3mm (Mechanical Surface Detection)
      Electronics: DAE4 Sn662; Calibrated: 2018-5-11
      Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
      Measurement SW: DASY4, V4.7 Build 55; Postprocessing SW: SEMCAD, V1.8 Build 176


system check/Area Scan (51x101x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (interpolated) = 1.17 mW/g

system check/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 33.4 V/m; Power Drift = -0.09 dB
Peak SAR (extrapolated) = 1.54 W/kg
SAR(1 g) = 0.990 mW/g; SAR(10 g) = 0.636 mW/g
Maximum value of SAR (measured) = 1.16 mW/g


                                                                             Date: 2018-10-31

Test Laboratory: UnionTrust

System Check_H1900

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium parameters used: f = 1900 MHz; σ = 1.41 mho/m; εr = 41.3; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY4 (High Precision Assessment)

DASY4 Configuration:

      Probe: ES3DV3 - SN3089; ConvF(4.81, 4.81, 4.81); Calibrated: 2018-5-18
      Sensor-Surface: 3mm (Mechanical Surface Detection)
      Electronics: DAE4 Sn662; Calibrated: 2018-5-11
      Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP-1376
      Measurement SW: DASY4, V4.7 Build 55; Postprocessing SW: SEMCAD, V1.8 Build 176


system check/Area Scan (51x71x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (interpolated) = 6.03 mW/g

system check/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 62.6 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 7.92 W/kg
SAR(1 g) = 4.2 mW/g; SAR(10 g) = 2.15 mW/g
Maximum value of SAR (measured) = 5.33 mW/g


                                                                             Date: 2018-11-02

Test Laboratory: UnionTrust

System Check_B835

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium parameters used: f = 835 MHz; σ = 0.98 mho/m; εr = 55.8; ρ = 1000 kg/m3
Phantom section: Center Section
Measurement Standard: DASY4 (High Precision Assessment)

DASY4 Configuration:

      Probe: ES3DV3 - SN3089; ConvF(6.21, 6.21, 6.21); Calibrated: 2018-5-18
      Sensor-Surface: 3mm (Mechanical Surface Detection)
      Electronics: DAE4 Sn662; Calibrated: 2018-5-11
      Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1125
      Measurement SW: DASY4, V4.7 Build 55; Postprocessing SW: SEMCAD, V1.8 Build 176


system check/Area Scan (51x101x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (interpolated) = 1.10 mW/g

system check/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 34.4 V/m; Power Drift = -0.04 dB
Peak SAR (extrapolated) = 1.44 W/kg
SAR(1 g) = 0.944 mW/g; SAR(10 g) = 0.612 mW/g
Maximum value of SAR (measured) = 1.10 mW/g


                                                                                Date: 2018-11-02

Test Laboratory: UnionTrust

System Check_B1900

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium parameters used: f = 1900 MHz; σ = 1.55 mho/m; εr = 54; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY4 (High Precision Assessment)

DASY4 Configuration:

      Probe: ES3DV3 - SN3089; ConvF(4.52, 4.52, 4.52); Calibrated: 2018-5-18
      Sensor-Surface: 3mm (Mechanical Surface Detection)
      Electronics: DAE4 Sn662; Calibrated: 2018-5-11
      Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP-1376
      Measurement SW: DASY4, V4.7 Build 55; Postprocessing SW: SEMCAD, V1.8 Build 176


system check/Area Scan (51x71x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (interpolated) = 6.40 mW/g

system check/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 60.0 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 7.75 W/kg
SAR(1 g) = 4.2 mW/g; SAR(10 g) = 2.16 mW/g
Maximum value of SAR (measured) = 5.34 mW/g



Document Created: 2018-11-07 17:58:17
Document Modified: 2018-11-07 17:58:17