I19Z60700-SEM01_SAR_Rev1_Part5

FCC ID: YAMPNC550B9

RF Exposure Info

Download: PDF
FCCID_4368942

Calibratio
      ;    n Laboratory of                                   N/z
                                                              ty
                                                            St‘s              S   Schweizerischer Kalibrierdienst
Schmid & Partner                                           ila‘
                                                             \—/E“E—!'_       c   Service suisse d‘étalonnage
  Englneenng AG                                            T nds                  Servizio svizzero di taratura
                              a       m                     zz\wy
Zeughausstrasse 43, 8004 Zurich, Switzerland                *4/fi\\y‘          S   Swiss Calibration Service
                                                               Arolale


Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates
Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) EC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   * Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
     e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2450V2—853_Jul17                               Page 2 of 8


()
 S
 Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version
      Extrapolation
      Phantom
      Distance Dipole Center — TSL
                                                                DASY5
                                                      Advanced Extrapolation
                                                       Modular Flat Phantom
                                                                10 mm
                                                                                                    V52.10.0




                                                                                                   with Spacer
      Zoom Scan Resolution                              dx, dy, dz =5 mm
      Frequency                                         2450 MHz # 1 MHz


 Head TSL barameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
      Nominal Head TSL parameters                                  22.0 °C            39.2               1.80 mho/m
      Measured Head TSL parameters                              (22.0 + 0.2) °C    37.8 x 6 %         1.87 mho/m # 6 %
      Head TSL temperature change during test                      <0.5 °C

 SAR result with Head TSL

      SAR averaged over 1 cm‘ (1 g) of Head TSL                  Condition
      SAR measured                                        250 mW input power                      13.4 Wikg
      SAR for nominal Head TSL parameters                     normalized to 1W         52.2 Wikg a 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                 condition
      SAR measured                                        250 mW input power                      6.26 Wikg
      SAR for nominal Head TSL parameters                     normalized to 1W         24.7 Wikg 2 16.5 % (k=2)


 Body TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature        Permittivity         Conductivity
       Nominal Body TSL parameters                                 22.0 °C            §2.7               1.95 mho/m
       Measured Body TSL parameters                             (22.0 £0.2) °C     52.1 #6%           2.04 mho/m £ 6 %
       Body TSL temperature change during test                     <0.5 °C             a~—                     ————


 SAR result with Body TSL

       SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
       SAR measured                                       250 mW input power                      12.9 Wikg
       SAR for nominal Body TSL parameters                    normalized to 1W          50.4 Wikg 17.0 % (k=2)

       SAR averaged over 10 cm* (10 g) of Body TSL                condition

       SAR measured                                       250 mW input power                   6.03 W/kg
       SAR for nominal Body TSL parameters                    normalized to 1W          23.8 Wikg x 16.5 % (k=2)



 Certificate No: D2450V2—853_Jul1 7                    Page 3 of 8




                                                                                  ~ o crucce


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                          52.0 Q + 5.0 jQ
      Return Loss                                                                      —25.6 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                          49.6 Q + 6.3 jQ
      Return Loss                                                                      —24.0 dB




General Antenna Parameters and Design

    IEectrical Delay (one direction)                        [                           1.161 ns                        41

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductorof the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                   SPEAG

      Manufactured on                                                             November 10, 2009




Certificate No: D2450V2—853_Jul17                        Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                       Date: 20.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.87 $/m; s = 37.8; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e    Probe: EX3DV4 — SN7349; ConvF(8.12, 8.12, 8.12); Calibrated: 31.05.2017;

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 112.7 V/m; Power Drift = —0.00 dB
Peak SAR (extrapolated) = 27.0 W/kg
SAR(1 g) = 13.4 W/kg; SAR(10 g) = 6.26 W/kg
Maximum value of SAR (measured) = 21.5 W/kg

        dB
         0


        ~4.00


        —8.00

        —12.00

        —16.00


        —20.00
                   0 dB = 21.5 W/kg = 13.32 dBW/kg




Certificate No: D2450V2—853_Jul17               Page 5 of 8




                                                                      ~ o nu ce    —


©
Impedance Measurement Plot for Head TSL




                                                                  20 Jul 2017 11:54:15
         CHZ sii      1 U FS          4151965 a       4.9570 a 322.01 pH      2 450.000 000 MHz
                                                  aer     T    .




         Av
         188


         H1d

          CH2




         Ca




         Av
         16°

          H1d

                    START 2 250.000 @00 MHz                            STOP 2 650.000 000 MHz




Certificate No: D2450V2—853_Jul17                     Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                               Date: 21.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 2.04 $/m; & = 52.1; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e   Probe: EX3DV4 — SN7349; ConvF(8.1, 8.1, 8.1); Calibrated: 31.05.2017;

   *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e   Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *   Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   &   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 104.1 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 25.5 W/kg
SAR(I g) = 12.9 W/kg; SAR(10 g) = 6.03 W/kg
Maximum value of SAR (measured) = 20.0 W/kg

       dB
       0
        —5.00
        —10.00
        —15.00
        —20.00
        —25.00
                   0 dB = 20.0 W/kg = 13.01 dBW/kg




Certificate No: D2450V2—853_Jul17              Page 7 of 8


©
Impedance Measurement Plot for Body TSL




                                                                       21 Jul 2017 14142134
         CHZ sii      i uo rs         1149.619 a       6.2656 2 407.02 pH         2450.000 000 MHz
                                                   —            _




         fAvg
         16


         H1d

          CH2




          Ca




          Av
          16"

          H1d

                    START 2 250.000 000 MHz                                 STOP 2 650.000 000 MHz




Certificate No: D2450V2—853_Jul17                        Page 8 of 8


&
duveve y   arik ame   m k pecme ewe   n nrneme n ne n n n n    o on n en n en




              Calibration Laboratory of                                                 wl/7                                   S   Schweizerischer Kalibrierdienst




                                                                                )§                &
              Schmid & Partner




                                                                                   9.
                                                                                                                                   Service suisse d‘étalonnage
                 Engineering AG                                                                                                    Servizio svizzero di taratura
              Zeughausstrasse 43, 8004 Zurich, Switzerland                              LN3                                        Swiss Calibration Service
                                                                                          Zulitabs®

              Accredited by the Swiss Accreditation Service (SAS)                                                               Accreditation No.: SCS 0108
              The Swiss Accreditation Service is one of the signatories to the EA
              Multilateral Agreementfor the recognition of calibration certificates

              Client         CTTL—BJ (Auden)                                                                        Certificate No: D2600V2—1012_Jul17

              CALIBRATION CERTIFICATE
               Object                                     D2600V2 — SN:1012


               Calibration procedure(s)                   QA CAL—05.v9
                                                          Calibration procedure for dipole validation kits above 700 MHz



               Calibration date:                          July 21, 2017


               This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
               The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


               All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


               Calibration Equipment used (M&TE critical for calibration)


               Primary Standards                              ID #                  Cal Date (Certificate No.)                            Scheduled Calibration
               Power meter NRP                                SN: 104778            04—Apr—17 (No. 217—02521/02522)                       Apr—18
                Power sensor NRP—Z291                         SN: 103244            04—Apr—17 (No. 217—02521)                             Apr—18
                Power sensor NRP—Z291                         SN: 108245            04—Apr—17 (No. 217—02522)                             Apr—18
                Reference 20 dB Attenuator                    SN: 5058 (20k)        07—Apr—17 (No. 217—02528)                             Apr—18
               Type—N mismatch combination                    SN: 5047.2 / 06827    O7—Apr—17 (No. 217—02529)                             Apr—18
                Reference Probe EX3DV4                        SN: 7349              31—May—17 (No. EX3—7349_May17)                        May—18
                DAE4                                          SN: 601               28—Mar—17 (No. DAE4—601_Mar17)                        Mar—18


                Secondary Standards                           ID #                  Check Date (in house)                                 Scheduled Check
                Power meter EPM—442A                          SN: GB37480704        07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
                Power sensor HP 8481A                         SN: US37202783        07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
                Power sensor HP 8481A                         SN: MY41092317        07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
                RF generator R&S SMT—06                       SN: 100972                15—Jun—15 (in house check Oct—16)                 In house check: Oct—18
                Network Analyzer HP 8753E                     SN: US37390585            18—Oct—01 (in house check Oct—16)                 In house check: Oct—17


                                                              Name                                    Function                             Si       ture
                Calibrated by:                                Michael Weber                           Laboratory Technician
                                                                                                                                                s

                Approved by:                                  Katja Pokovic                           Technical Manager            //%%



                                                                                                                                          Issued: July 24, 2017
                This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


               Certificate No: D2600V2—1012_Jul17                                          Page 1 of 8




                                                                                                                              ~ onz   s                —


Calibn:ation Laboratory of                                  s\o\‘;i‘\ijj'm,,’
Schmid & Partner                                           ilfi%./{@                 S      Schweizerischer Kalibrierdienst
                                                                                           Service suisse d‘étalonnage
  Engineering AG                                           yyp: l                   C      Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                €4,/fi\\x}               S      Swiss Calibration Service
                                                                "thilale


Accredited by the Swiss Accreditation Service (SAS)                                  Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates
Glossary:
TSL                         tissue simulating liquid
ConvrF                      sensitivity in TSL / NORM x,y,z
N/A                         not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "[EEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   * Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2600V2—1012_Jul17                               Page 2 of 8




                                                                                ,,,,,,,,


(iléfll,
          Measurement Conditions
             DASY system configuration, as far as not given on page 1.
              DASY Version                                             DASY5                                 V52.10.0
               Extrapolation                                   Advanced Extrapolation
               Phantom                                          Modular Flat Phantom
               Distance Dipole Center — TSL                               10 mm                             with Spacer
               Zoom Scan Resolution                               dx, dy, dz =5 mm
               Frequency                                          2600 MHz + 1 MHz


          Head TSL parameters
             The following parameters and calculations were applied.
                                                                          Temperature       Permittivity          Conductivity
                Nominal Head TSL parameters                                  22.0 °C           39.0                1.96 mho/m
                Measured Head TSL parameters                              (22.0 + 0.2) °C   37.2 46 %           2.04 mho/m + 6 %
                Head TSL temperature change during test                      <0.5 °C                                    ——

          SAR result with Head TSL

                SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
                SAR measured                                       250 mW input power                      14.9 W/kg
                SAR for nominal Head TSL parameters                     normalized to 1W        57.9 Wikg + 17.0 % (k=2)

                SAR averaged over 10 cm* (10 g) of Head TSL                 condition
                SAR measured                                       250 mW input power                      6.57 Wi/kg
                SAR for nominal Head TSL parameters                     normalized to 1W         25.8 Wikg a 16.5 % (k=2)


          Body TSL parameters
              The following parameters and calculations were applied.
                                                                          Temperature       Permittivity          Conductivity
                Nominal Body TSL parameters                                  22.0 °C            52.5               216 mho/m
                Measured Body TSL parameters                              (22.0 £ 0.2) °C    51.6 4 6 %         2.22 mho/m + 6 %
                Body TSL temperature change during test                      <0.5 °C            ~~~                     _


          SAR result with Body TSL

                SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
                SAR measured                                        250 mW input power                     14.1 W/kg
                SAR for nominal Body TSL parameters                     normalized to 1W         55.5 Wikg + 17.0 % (k=2)

                SAR averaged over 10 cm‘ (10 g) of Body TSL                 condition

                SAR measured                                        250 mW input power                  6.25 W/kg
                SAR for nominal Body TSL parameters                     normalized to 1W         24.8 Wikg # 16.5 % (k=2)



          Certificate No: D2600V2—1012_Jul17                     Page 3 of 8


(uém»
        Appendix (Additional assessments outside the scope of SCS 0108)

        Antenna Parameters with Head TSL

              Impedance, transformed to feed point                                           46.8 Q — 5.0 jQ
              Return Loss                                                                       — 24.2 dB



        Antenna Parameters with Body TSL

              Impedance, transformed to feed point                                           43.5 0 — 5.3 jQ
              Return Loss                                                                      —21.0 dB



        General Antenna Parameters and Design

            [ Electrical Delay (one direction)                                                  1.151 ns

        After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

        The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
        second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
        are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
        "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
        according to the Standard.
        No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
        feedpoint may be damaged.



        Additional EUT Data

              Manufactured by                                                                    SPEAG
              Manufactured on                                                               October 30, 2007




        Certificate No: D2600V2—1012_Jul17                        Page 4 of 8


_:                                                                       |
@77T,L
y
                                                                                 1O _


  DASY5 Validation Report for Head TSL

                                                                                        Date: 20.07.2017
  Test Laboratory: SPEAG, Zurich, Switzerland

  DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN: 1012

  Communication System: UID 0 — CW; Frequency: 2600 MHz
  Medium parameters used: f = 2600 MHz; 0 = 2.04 S/m; & = 37.2; p = 1000 kg/m*
  Phantom section: Flat Section
  Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

  DASY52 Configuration:

     *    Probe: EX3DV4 — SN7349; ConvF(7.96, 7.96, 7.96); Calibrated: 31.05.2017;

     *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

     *    Electronics: DAE4 $n601; Calibrated: 28.03.2017

     *    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

     «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


  Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
  Measurement grid: dx=5mm, dy=5mm, dz=5mm
  Reference Value = 113.6 V/m; Power Drift = 0.05 dB
  Peak SAR (extrapolated) = 32.3 W/kg
  SAR(1 g) = 14.9 W/kg; SAR(10 g) = 6.57 W/kg
  Maximum value of SAR (measured) = 25.0 W/kg


          dB
          0

          —5.00

          —10.00

          —15.00

          —20.00

          —25.00
                     0 dB = 25.0 W/kg = 13.98 dBW/kg




  Certificate No: D2600V2—1012_Jul17              Page 5 of 8


®
Impedance Measurement Plot for Head TSL




                                                                  20 Jul 2017 12:14:05
                S11     1 U FS          5: 46.836 0   —5.0410 0 12143 pF      2 600.000 000 MHz
                                                      ~~1



         Del

         Ca




         Av
         16 *


         H1d

          CH2




         Ca




         Av
         16s


          H1d

                      START 2 400.000 000 MHz                           STOP 2 800.000 000 MHz




Certificate No: D2600V2—1012_Jul17                      Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                      Date: 21.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN: 1012

Communication System: UID 0 — CW; Frequency: 2600 MHz
Medium parameters used: f = 2600 MHz; 0 = 2.22 $/m; & = 51.6; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e   Probe: EX3DV4 — SN7349; ConvF(7.94, 7.94, 7.94); Calibrated: 31.05.2017;

   e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e   Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *   Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   e   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 106.6 V/m; Power Drift = —0.07 dB
Peak SAR (extrapolated) = 30.1 W/kg
SAR(I g) = 14.1 W/kg; SAR(10 g) = 6.25 W/ikg
Maximum value of SAR (measured) = 23.4 W/kg


       dB
        0

       —5.20
       ~10.40
        —15.60
        —20.80
        —26.00
                   0 dB = 23.4 W/kg = 13.69 dBW/kg




Certificate No: D2600V2—1012_Jul17             Page 7 of 8




                                                                    ~ o us        ~


mm
  TL
y




Impedance Measurement Plot for Body TSL




                                                                       21 Jul 2017 14:S50:41
                 §11     1 U FS          1: 43.541. a   —5.3223 a   11.501 pF      2 600.000 000 MHz




         Del

         Ca




         Av
         16 *


         H1d

          CH2




         Ca




          fivg
          16


          H1d

                       START 2 400.000 000 MHz                               STOP 2 800.000 000 MHz




Certificate No: D2600V2—1012_Jul17                         Page 8 of 8


                                                                    No. I19Z60700-SEM01
                                                                      Page 199 of 202


ANNEX I        Extended Calibration SAR Dipole
Referring to KDB865664 D01, if dipoles are verified in return loss ( <-20dBm, within 20% of prior
calibration), and in impedance ( within 5 ohm of prior calibration), the annual calibration is not
necessary and the calibration interval can be extended.

Justification of Extended Calibration SAR Dipole D750V3– serial no.1017
 Head
                                              Real                         Imaginary
 Date       of    Return-Loss                                 Delta                         Delta
                                  Delta (%)   Impedance                    Impedance
 Measurement      (dB)                                        (ohm)                         (johm)
                                              (ohm)                        (johm)
 2017-7-19        -27.5                       54.4                         0.5
 2018-7-17        -26.0           5.5         55.1            0.7          -1.0             -1.5


 Body
                                              Real                         Imaginary
 Date       of    Return-Loss                                 Delta                         Delta
                                  Delta (%)   Impedance                    Impedance
 Measurement      (dB)                                        (ohm)                         (johm)
                                              (ohm)                        (johm)
 2017-7-19        -29.1                       49.3                         -3.4
 2018-7-17        -29.3           -0.69       51.5            2.2          -3.1             0.3




Justification of Extended Calibration SAR Dipole D835V2– serial no.4d069
 Head
                                           Real                           Imaginary
 Date     of     Return-Loss    Delta                                                      Delta
                                           Impedance    Delta (ohm)       Impedance
 Measurement     (dB)           (%)                                                        (johm)
                                           (ohm)                          (johm)
 2017-7-19       -32.4                     52.1                           -1.2
 2018-7-17       -30.3          6.5        53.0         1.1               -1.0             0.2


 Body
                                              Real                         Imaginary
 Date       of    Return-Loss                                 Delta                         Delta
                                  Delta (%)   Impedance                    Impedance
 Measurement      (dB)                                        (ohm)                         (johm)
                                              (ohm)                        (johm)
 2017-7-19        -26.9                       47.9                         -3.9
 2018-7-17        -25.5           5.2         48.5            0.6          -5.0             -1.1




                                                               ©Copyright. All rights reserved by CTTL.


                                                                   No. I19Z60700-SEM01
                                                                     Page 200 of 202


Justification of Extended Calibration SAR Dipole D1750V2– serial no.1003
 Head
                                               Real                        Imaginary
 Date      of     Return-Loss                               Delta                          Delta
                                  Delta (%)    Impedance                   Impedance
 Measurement      (dB)                                      (ohm)                          (johm)
                                               (ohm)                       (johm)
 2017-7-21        -37.1                        50.9                        1.1
 2018-7-19        -35.6           4.0          49.5         -1.4           -1.6            -2.7


 Body
                                               Real                        Imaginary
 Date       of    Return-Loss                               Delta                          Delta
                                  Delta (%)    Impedance                   Impedance
 Measurement      (dB)                                      (ohm)                          (johm)
                                               (ohm)                       (johm)
 2017-7-21        -30.2                        47.0                        0.1
 2018-7-19        -27.6           8.6          46.0         -1.0           -0.3            -0.4




Justification of Extended Calibration SAR Dipole D1900V2– serial no.5d101
 Head
                                              Real                       Imaginary
 Date     of     Return-Loss                               Delta                          Delta
                                Delta (%)     Impedance                  Impedance
 Measurement     (dB)                                      (ohm)                          (johm)
                                              (ohm)                      (johm)
 2017-7-26       -24.5                        51.7                       5.8
 2018-7-24       -22.9          6.5           50.6         -1.1          7.2              1.4


 Body
                                               Real                        Imaginary
 Date      of     Return-Loss                               Delta                          Delta
                                  Delta (%)    Impedance                   Impedance
 Measurement      (dB)                                      (ohm)                          (johm)
                                               (ohm)                       (johm)
 2017-7-26        -22.0                        46.2                        6.6
 2018-7-24        -21.4           2.7          46.4         0.2            7.4             0.8




                                                              ©Copyright. All rights reserved by CTTL.


                                                                    No. I19Z60700-SEM01
                                                                      Page 201 of 202


Justification of Extended Calibration SAR Dipole D2450V2– serial no.853
 Head
                                               Real                         Imaginary
 Date       of    Return-Loss     Delta                     Delta                             Delta
                                               Impedance                    Impedance
 Measurement      (dB)            (%)                       (ohm)                             (johm)
                                               (ohm)                        (johm)
 2017-7-21        -25.6                        52.0                         5.0
 2018-7-19        -23.1           9.8          53.6         1.6             6.3               1.3


 Body
                                               Real                         Imaginary
 Date       of    Return-Loss                               Delta                             Delta
                                  Delta (%)    Impedance                    Impedance
 Measurement      (dB)                                      (ohm)                             (johm)
                                               (ohm)                        (johm)
 2017-7-21        -24.0                        49.6                         6.3
 2018-7-19        -22.0           8.3          50.4         0.8             8.0               1.7




Justification of Extended Calibration SAR Dipole D2600V2– serial no.1012
 Head
                                              Real                          Imaginary
 Date     of     Return-Loss                               Delta                             Delta
                                Delta (%)     Impedance                     Impedance
 Measurement     (dB)                                      (ohm)                             (johm)
                                              (ohm)                         (johm)
 2017-7-21       -24.2                        46.8                          -5.0
 2018-7-19       -23.7          2.1           47.7         0.9              -5.9             -0.9


 Body
                                               Real                         Imaginary
 Date      of     Return-Loss                               Delta                             Delta
                                  Delta (%)    Impedance                    Impedance
 Measurement      (dB)                                      (ohm)                             (johm)
                                               (ohm)                        (johm)
 2017-7-21        -21.0                        43.5                         -5.3
 2018-7-19        -22.4           -6.7         44.4         0.9             -4.5              0.8




                                                                 ©Copyright. All rights reserved by CTTL.


                                       No. I19Z60700-SEM01
                                         Page 202 of 202


ANNEX J Accreditation Certificate




                                    ©Copyright. All rights reserved by CTTL.



Document Created: 2019-07-16 10:48:55
Document Modified: 2019-07-16 10:48:55

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC