SAR report part 1

FCC ID: Q5EDP58001

RF Exposure Info

Download: PDF
FCCID_4231911

                                                   TEST REPORT
Report No........................................ :      CHTEW19030220                        Report verificaiton:

Project No. ...................................... :     SHT1901068505EW
FCC ID ............................................. :   Q5EDP58001

Applicant’s name ........................... :           Kirisun Communication Co.,Ltd.

Address ............................................ :   3rd Floor, Building A, Tongfang Information Habour, No.11
                                                         Langshan Road, Nanshan District, Shenzhen 518057, P.R.China
Manufacturer.....................................:       Kirisun Communication Co.,Ltd.

Address.............................................:    3rd Floor, Building A, Tongfang Information Habour, No.11
                                                         Langshan Road, Nanshan District, Shenzhen 518057, P.R.China
Test item description ..................... :            DMR Two Way Radio

Trade Mark....................................... :      KIRISUN
Model/Type reference ...................... :            DP580-01

Listed Model(s) ................................ :       DP585-01
                                                          FCC 47 CFR Part2.1093
Standard ......................................... :      IEEE Std C95.1, 1999 Edition
                                                          IEEE 1528: 2013
Date of receipt of test sample………:                       Jan. 31, 2019

Date of testing………………………..:                              Feb. 01, 2019- Feb. 25, 2019

Date of issue………………………….:                                Mar. 25, 2019

Result .................. …………………...:                    PASS

Compiled by
( position+printed name+signature) . :                   File administrators:   Xiaodong Zhao

Supervised by
( position+printed name+signature) . :                   Test Engineer:         Xiaodong Zhao

Approved by
( position+printed name+signature) . :                   Manager:                   Hans Hu

Testing Laboratory Name ............. :                  Shenzhen Huatongwei International Inspection Co., Ltd

Address ............................................ :   1/F, Bldg 3, Hongfa Hi-tech Industrial Park, Genyu Road, Tianliao,
                                                         Gongming, Shenzhen, China

Shenzhen Huatongwei International Inspection Co., Ltd. All rights reserved.
This publication may be reproduced in whole or in part for non-commercial purposes as long as the
Shenzhen Huatongwei International Inspection Co., Ltd is acknowledged as copyright owner and source of
the material. Shenzhen Huatongwei International Inspection Co., Ltd takes no responsibility for and will not
assume liability for damages resulting from the reader's interpretation of the reproduced material due to its
placement and context.
The test report merely correspond to the test sample.



                                                                    Page: 1 of 30


Report No:    CHTEW19030220                     Page: 2 of 30                      Issued: 2019-03-25



                                                  Contents
1.      Test Standards and Report version                                                                    3
 1.1.   Test Standards                                                                                  3
 1.2.   Report version                                                                                  3
2.      Summary                                                                                              4
 2.1.   Client Information                                                                              4
 2.2.   Product Description                                                                             4
 2.3.   Test frequency list                                                                             6
3.      Test Environment                                                                                     7
 3.1.   Test laboratory                                                                                 7
 3.2.   Test Facility                                                                                   7
 3.3.   Environmental conditions                                                                        7
4.      Equipments Used during the Test                                                                      8
5.      Measurement Uncertainty                                                                              9
6.      SAR Measurements System Configuration                                                               10
 6.1.   SAR Measurement Set-up                                                                          10
 6.2.   DASY5 E-field Probe System                                                                      11
 6.3.   Phantoms                                                                                        12
 6.4.   Device Holder                                                                                   12
7.      SAR Test Procedure                                                                                  13
 7.1.   Scanning Procedure                                                                              13
 7.2.   Data Storage and Evaluation                                                                     15
8.      Position of the wireless device in relation to the phantom                                          17
 8.1.   Front-of-face                                                                                   17
 8.2.   Body Position                                                                                   17
9.      System Check                                                                                        18
 9.1.   Tissue Dielectric Parameters                                                                    18
 9.2.   SAR System Check                                                                                19
10.     SAR Exposure Limits                                                                                 23
11.     Conducted Power Measurement Results                                                                 24
12.     Maximum Tune-up Limit                                                                               25
13.     SAR Measurement Results                                                                             26
14.     Simultaneous Transmission analysis                                                                  28
15.     Test Setup Photos                                                                                   29
16.     External and Internal Photos of the EUT                                                             30




Shenzhen Huatongwei International Inspection Co., Ltd.               Report Template Version: V01 (2018-08)


Report No:     CHTEW19030220                    Page: 3 of 30                 Issued: 2019-03-25

1 . Test Standards and Report version
1.1. Test Standards
The tests were performed according to following standards:
FCC 47 Part 2.1093: Radiofrequency Radiation Exposure Evaluation:Portable Devices
IEEE Std C95.1, 1999 Edition: IEEE Standard for Safety Levels with Respect to Human Exposure to Radio
Frequency Electromagnetic Fields, 3 kHz to 300 GHz
IEEE Std 1528™-2013: IEEE Recommended Practice for Determining the Peak Spatial-Average Specific
Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement
Techniques.
KDB 865664 D01 SAR Measurement 100 MHz to 6 GHz v01r04: SAR Measurement Requirements for 100
MHz to 6 GHz
KDB 865664 D02 RF Exposure Reporting v01r02: RF Exposure Compliance Reporting and Documentation
Considerations
KDB 447498 D01 General RF Exposure Guidance v06: Mobile and Portable Device RF Exposure Procedures
and Equipment Authorization Policies
KDB 643646 D01:SAR Test for PTT Radios v01r03: SAR Test Reduction Considerations for Occupational
PTT Radios
1.2. Report version
      Revision No.                Date of issue                         Description
             N/A                   2019-03-25                             Original




Shenzhen Huatongwei International Inspection Co., Ltd.          Report Template Version: V01 (2018-08)


Report No:      CHTEW19030220                   Page: 4 of 30                   Issued: 2019-03-25

2. Summary
2.1. Client Information
Applicant:                    Kirisun Communication Co.,Ltd.
                              3rd Floor, Building A, Tongfang Information Habour, No.11 Langshan Road,
Address:
                              Nanshan District, Shenzhen 518057, P.R.China
Manufacturer:                 Kirisun Communication Co.,Ltd.
                              3rd Floor, Building A, Tongfang Information Habour, No.11 Langshan Road,
Address:
                              Nanshan District, Shenzhen 518057, P.R.China

2.2. Product Description
Name of EUT:                  DMR Two Way Radio
Trade mark:                   KIRISUN
Model/Type reference:         DP580-01

Listed model(s):              DP585-01
Accessories                   Belt Clip
Device Category:              Portable
Product stage:                Production unit
RF Exposure Environment:      Occupational/Controlled
Power supply:                 DC 7.4V
Hardware version:             V1.0

Software version:             V1.2.0.9
                              Overall (Length x Width x Thickness):115 x 60 x 35mm
Device Dimension:             Antenna(Length):158mm(136~151MHz),143mm(151~162MHz),
                              130mm(162~174MHz)
Maximum SAR Value
                              Front-of-face:      25mm
Separation Distance:
                              Body:               0mm
                              Front-of-face:      1.215 W/kg
Maximun SAR Value (1g):
                              Body:               4.468 W/kg
PMR
Operation Frequency Range:    136~174MHz
Rated Output Power:              High Power:      5W(36.99dBm)        Low Power:     1W(30.00dBm)
Modulation Type:              Analog:             FM
                              Digital:            4FSK
Channel Separation:           Analog:                 12.5kHz         20kHz             25kHz
                              Digital:                6.25kHz         12.5kHz
Digital Type:                 DMR
Antenna type:                 External




Shenzhen Huatongwei International Inspection Co., Ltd.           Report Template Version: V01 (2018-08)


Report No:      CHTEW19030220                  Page: 5 of 30                       Issued: 2019-03-25

Bluetooth
Version:                       Supported BT4.2+EDR
Modulation:                    GFSK,π/4QPSK,8DPSK
Operation frequency:           2402MHz~2480MHz
Channel number:                79
Channel separation:            1MHz
Antenna type:                  FPC
Bluetooth
Version:                       Supported BT4.2+BLE
Modulation:                    GFSK
Operation frequency:           2402MHz~2480MHz
Channel number:                40
Channel separation:            2MHz
Antenna type:                  FPC
Remark:
The EUT battery must be fully charged and checked periodically during the test to ascertain uniform power.




Shenzhen Huatongwei International Inspection Co., Ltd.             Report Template Version: V01 (2018-08)


Report No:     CHTEW19030220                    Page: 6 of 30                         Issued: 2019-03-25

2.3. Test frequency list
When the frequency channels required for SAR testing are not specified, the following should be applied to
determine the number of required test channels.The test channels should be evenly spread across the
transmission frequency band of each wireless mode.



Nc is the number of test channels, rounded to the nearest integer,
fhigh and flow are the highest and lowest channel frequencies within the transmission band,
fc is the mid-band channel frequency,
all frequencies are in MHz.

             Operation Frequency                Test Frequency
  Start Frequency        Stop Frequency            number

         136                   174                     5



                             Channel                                          Test Frequency (MHz)
  ModulationType                                 Test Channel
                            Separation                                       TX                  RX
                                                      CH1                136.0250             136.0250
                                                      CH2                145.5000             145.5000
       Analog                12.5kHz                  CH3                155.0000             155.0000
                                                      CH4                164.5000             164.5000
                                                      CH5                173.9750             173.9750
                                                      CH1                136.0250             136.0250
                                                      CH2                145.5000             145.5000
       Digital               12.5kHz                  CH3                155.0000             155.0000
                                                      CH4                164.5000             164.5000
                                                      CH5                173.9750             173.9750




Shenzhen Huatongwei International Inspection Co., Ltd.                Report Template Version: V01 (2018-08)


Report No:     CHTEW19030220                   Page: 7 of 30                         Issued: 2019-03-25

3. Test Environment
3.1. Test laboratory
Laboratory:Shenzhen Huatongwei International Inspection Co., Ltd.
Address: 1/F, Bldg 3, Hongfa Hi-tech Industrial Park, Genyu Road, Tianliao, Gongming, Shenzhen, China
3.2. Test Facility
CNAS-Lab Code: L1225

Shenzhen Huatongwei International Inspection Co., Ltd. has been assessed and proved to be in compliance
with CNAS-CL01 Accreditation Criteria for Testing and Calibration Laboratories (identical to ISO/IEC17025:
2005 General Requirements) for the Competence of Testing and Calibration Laboratories.

A2LA-Lab Cert. No.: 3902.01

Shenzhen Huatongwei International Inspection Co., Ltd. EMC Laboratory has been accredited by A2LA for
technical competence in the field of electrical testing, and proved to be in compliance with ISO/IEC 17025:
2005 General Requirements for the Competence of Testing and Calibration Laboratories and any additional
program requirements in the identified field of testing.

FCC-Registration No.: 762235

Shenzhen Huatongwei International Inspection Co., Ltd. EMC Laboratory has been registered and fully
described in a report filed with the FCC (Federal Communications Commission). The acceptance letter from
the FCC is maintained in our files. Registration 762235.

IC-Registration No.: 5377B-1

Two 3m Alternate Test Site of Shenzhen Huatongwei International Inspection Co., Ltd. has been registered by
Certification and Engineering Bureau of Industry Canada for the performance of radiated measurements with
Registration No. 5377B-1.

ACA

Shenzhen Huatongwei International Inspection Co., Ltd. EMC Laboratory can also perform testing for the
Australian C-Tick mark as a result of our A2LA accreditation.


3.3. Environmental conditions
During the measurement the environmental conditions were within the listed ranges:

Ambient temperature                                       18 °C to 25 °C

Ambient humidity                                          30%RH to 70%RH

Air Pressure                                              950-1050mbar




Shenzhen Huatongwei International Inspection Co., Ltd.              Report Template Version: V01 (2018-08)


Report No:      CHTEW19030220                         Page: 8 of 30                     Issued: 2019-03-25

4. Equipments Used during the Test
                                                                                        Cal. date     Due date
Used         Test Equipment           Manufacturer         Model No.     Serial No.
                                                                                      (YY-MM-DD)    (YY-MM-DD)
       Data Acquisition Electronics
 ●     DAEx
                                        SPEAG               DAE4           1549       2018/04/25    2019/04/24

 ●     E-field Probe                    SPEAG              EX3DV4          7494       2018/02/26    2019/02/25

       Universal Radio
 ○     Communication Tester
                                          R&S              CMW500         137681      2018/07/11    2019/07/10

 ● Tissue-equivalent liquids Validation
 ○     Dielectric Assessment Kit        SPEAG              DAK-3.5         1267          N/A           N/A

 ●     Dielectric Assessment Kit        SPEAG               DAK-12         1130          N/A           N/A

 ●     Network analyzer                 Keysight           E5071C       MY46733048    2018/09/19    2019/09/18

 ○ System Validation
 ●     System Validation Antenna        SPEAG              CLA-150         4024       2018/02/21    2021/02/20

 ○     System Validation Dipole         SPEAG              D450V3          1102       2018/02/23    2021/02/22

 ○     System Validation Dipole         SPEAG              D750V3          1180       2018/02/07    2021/02/06

 ○     System Validation Dipole         SPEAG              D835V2         4d238       2018/02/19    2021/02/18

 ○     System Validation Dipole         SPEAG              D1750V2         1164       2018/02/06    2021/02/05

 ○     System Validation Dipole         SPEAG              D1900V2        5d226       2018/02/22    2021/02/21

 ○     System Validation Dipole         SPEAG              D2450V2         1009       2018/02/05    2021/02/04

 ○     System Validation Dipole         SPEAG              D2600V2         1150       2018/02/05    2021/02/04

 ○     System Validation Dipole         SPEAG              D5GHzV2         1273       2018/02/21    2021/02/20

 ●     Signal Generator                   R&S              SMB100A        114360      2018/08/21    2019/08/20

 ●     Power Viewer for Windows           R&S                N/A            N/A          N/A           N/A

 ●     Power sensor                       R&S              NRP18A         101010      2018/08/21    2019/08/20

 ●     Power sensor                       R&S              NRP18A         101011      2018/08/21    2019/08/20

 ●     Power Amplifier                   BONN           BLWA 0160-2M     1811887      2018/11/15    2019/11/14

 ●     Dual Directional Coupler       Mini-Circuits     ZHDC-10-62-S+   F975001814    2018/11/15    2019/11/14

 ●     Attenuator                     Mini-Circuits       VAT-3W2+         1819       2018/11/15    2019/11/14

 ●     Attenuator                     Mini-Circuits       VAT-10W2+        1741       2018/11/15    2019/11/14

Note:
1. The Probe,Dipole and DAE calibration reference to the Appendix B and C.
2. Referring to KDB865664 D01, the dipole calibration interval can be extended to 3 years with justifcatio.
    The dipole are also not physically damaged or repaired during the interval.




Shenzhen Huatongwei International Inspection Co., Ltd.                   Report Template Version: V01 (2018-08)


Report No:   CHTEW19030220                    Page: 9 of 30                       Issued: 2019-03-25

5. Measurement Uncertainty
Per KDB 865664 D01 SAR Measurement 100 MHz to 6 GHz, when the highest measured 1-g SAR within a
frequency band is < 1.5 W/kg. The equivalent ratio (1.5/1.6) should be applied to extremity and occupational
exposure conditions.The expanded SAR measurement uncertainty must be ≤ 30%, for a confidence interval of
k = 2. If these conditions are met,extensive SAR measurement uncertainty analysis described in IEEE Std
1528-2013 is not required in SAR reports submitted for equipment approval.




Shenzhen Huatongwei International Inspection Co., Ltd.             Report Template Version: V01 (2018-08)


Report No:    CHTEW19030220                     Page: 10 of 30                        Issued: 2019-03-25

6. SAR Measurements System Configuration
6.1. SAR Measurement Set-up

  The DASY5 system for performing compliance tests consists of the following items:

  A standard high precision 6-axis robot (Stäubli RX family) with controller and software. An arm extension for
  accommodating the data acquisition electronics (DAE).

  A dosimetric probe, i.e. an isotropic E-field probe optimized and calibrated for usage in tissue simulating
  liquid. The probe is equipped with an optical surface detector system.

  A data acquisition electronic (DAE) which performs the signal amplification, signal multiplexing, AD-
  conversion, offset measurements, mechanical surface detection, collision detection, etc. The unit is battery
  powered with standard or rechargeable batteries. The signal is optically transmitted to the EOC.

  A unit to operate the optical surface detector which is connected to the EOC.

  The Electro-Optical Coupler (EOC) performs the conversion from the optical into a digital electric signal of
  the DAE. The EOC is connected to the DASY5 measurement server.

  The DASY5 measurement server, which performs all real-time data evaluation for field measurements and
  surface detection, controls robot movements and handles safety operation. A computer operating Windows
  2003.

  DASY5 software and SEMCAD data evaluation software.

  Remote control with teach panel and additional circuitry for robot safety such as warning lamps, etc.

  The generic twin phantom enabling the testing of left-hand and right-hand usage.

  The device holder for handheld Mobile Phones.

  Tissue simulating liquid mixed according to the given recipes.

  System validation dipoles allowing to validate the proper functioning of the system.




Shenzhen Huatongwei International Inspection Co., Ltd.                Report Template Version: V01 (2018-08)


Report No:      CHTEW19030220                     Page: 11 of 30                     Issued: 2019—03—25

6.2. DASY5 E—field Probe System
The SAR measurements were conducted with the dosimetric probe EX3DV4 (manufactured by SPEAG),
designed in the classical triangular configuration and optimized for dosimetric evaluation.

®    Probe Specification

ConstructionSymmetrical design with triangular core
                        Interleaved sensors
                        Built—in shielding against static charges
                        PEEK enclosure material (resistant to organic solvents, e.g., DGBE)

Calibration|SO/IEC 17025 calibration service available.

Frequency                  10 MHz to 10 GHz;
                           Linearity: + 0.2 dB (30 MHz to 10 GHz)

Directivity                +0.1 dB in TSL (rotation around probe axis)
                           +0.3 dB in TSL (rotation normal to probe axis)

Dynamic Range              10 uW/g to > 100 mW/g;
                           Linearity: + 0.2 dB

Dimensions                 Overall length: 337 mm (Tip: 20 mm)
                           Tip diameter: 2.5 mm (Body: 12 mm)
                           Distance from probe tip to dipole centers: 2.0 mm

Application                General dosimetry up to 10 GHz
                           Dosimetry in strong gradient fields
                           Compliance tests of Mobile Phones

Compatibility              DASY3, DASY4, DASY52 SAR and higher, EASY4/MRI

®    Isotropic E—Field Probe

The isotropic E—Field probe has been fully calibrated and assessed for isotropicity, and boundary effect within
a controlled environment. Depending on the frequency for which the probe is calibrated the method utilized for
calibration will change.

The E—Field probe utilizes a triangular sensor arrangement as detailed in the diagram below:
              DIPOLE SENSOR
       Hlnfl-nEsl
                     flmch% cror
                                        / ‘Ix'   //
              —]‘} DiobE




Shenzhen Huatongwei International Inspection Co., Ltd.                Report Template Version: VO1 (2018—08)


Report No:    CHTEW19030220                     Page: 12 of 30                       Issued: 2019-03-25

6.3. Phantoms
Phantom for compliance testing of handheld and body-mounted wireless devices in the frequency range of 30
MHz to 6 GHz. ELI isfully compatible with standard and all known tissuesimulating liquids. ELI has been
optimized regarding its performance and can beintegrated into our standard phantom tables. A cover prevents
evaporation ofthe liquid. Reference markings on the phantom allow installation of thecomplete setup, including
all predefined phantom positions and measurementgrids, by teaching three points. The phantom is compatible
with all SPEAGdosimetric probes and dipoles.




                                                 ELI Phantom


6.4. Device Holder
The device was placed in the device holder (illustrated below) that is supplied by SPEAG as an integral part of
the DASY system.
The DASY device holder is designed to cope with the different positions given in the standard. It has two
scales for device rotation (with respect to the body axis) and device inclination (with respect to the line
between the ear reference points). The rotation centers for both scales is the ear reference point (ERP). Thus
the device needs no repositioning when changing the angles.




                                      Device holder supplied by SPEAG




Shenzhen Huatongwei International Inspection Co., Ltd.               Report Template Version: V01 (2018-08)


Report No:    CHTEW19030220                     Page: 13 of 30                        Issued: 2019-03-25

7. SAR Test Procedure
7.1. Scanning Procedure
  The DASY5 installation includes predefined files with recommended procedures for measurements and
  validation. They are read-only document files and destined as fully defined but unmeasured masks. All test
  positions (head or body-worn) are tested with the same configuration of test steps differing only in the grid
  definition for the different test positions.

  The “reference” and “drift” measurements are located at the beginning and end of the batch process. They
  measure the field drift at one single point in the liquid over the complete procedure. The indicated drift is
  mainly the variation of the DUT’s output power and should vary max. ± 5 %.

  The “surface check” measurement tests the optical surface detection system of the DASY5 system by
  repeatedly detecting the surface with the optical and mechanical surface detector and comparing the results.
  The output gives the detecting heights of both systems, the difference between the two systems and the
  standard deviation of the detection repeatability. Air bubbles or refraction in the liquid due to separation of
  the sugar-water mixture gives poor repeatability (above ± 0.1mm). To prevent wrong results tests are only
  executed when the liquid is free of air bubbles. The difference between the optical surface detection and the
  actual surface depends on the probe and is specified with each probe (It does not depend on the surface
  reflectivity or the probe angle to the surface within ± 30°.)

  Area Scan
  The Area Scan is used as a fast scan in two dimensions to find the area of high field values before running
  a detailed measurement around the hot spot.Before starting the area scan a grid spacing of 15 mm x 15
  mm is set. During the scan the distance of the probe to the phantom remains unchanged. After finishing
  area scan, the field maxima within a range of 2 dB will be ascertained.

  Zoom Scan
  After the maximum interpolated values were calculated between the points in the cube, the SAR was
  averaged over the spatial volume (1g or 10g) using a 3D-Spline interpolation algorithm. The 3D-spline is
  composed of three one-dimensional splines with the “Not a knot” condition (in x, y, and z directions). The
  volume was then integrated with the trapezoidal algorithm.

  Spatial Peak Detection
  The procedure for spatial peak SAR evaluation has been implemented and can determine values of masses
  of 1g and 10g, as well as for user-specific masses.The DASY5 system allows evaluations that combine
  measured data and robot positions, such as:
  • maximum search
  • extrapolation
  • boundary correction
  • peak search for averaged SAR
  During a maximum search, global and local maxima searches are automatically performed in 2-D after each
  Area Scan measurement with at least 6 measurement points. It is based on the evaluation of the local SAR
  gradient calculated by the Quadratic Shepard’s method. The algorithm will find the global maximum and all
  local maxima within -2 dB of the global maxima for all SAR distributions.
  Extrapolation routines are used to obtain SAR values between the lowest measurement points and the
  inner phantom surface. The extrapolation distance is determined by the surface detection distance and the
  probe sensor offset. Several measurements at different distances are necessary for the extrapolation.
  Extrapolation routines require at least 10 measurement points in 3-D space.
  They are used in the Zoom Scan to obtain SAR values between the lowest measurement points and the
  inner phantom surface. The routine uses the modified Quadratic Shepard’s method for extrapolation.

  A Z-axis scan measures the total SAR value at the x-and y-position of the maximum SAR value found
  during the cube scan. The probe is moved away in z-direction from the bottom of the SAM phantom in 5mm
  steps.




Shenzhen Huatongwei International Inspection Co., Ltd.                Report Template Version: V01 (2018-08)


Report No:   CHTEW19030220                    Page: 14 of 30                 Issued: 2019-03-25

  Table 1: Area and Zoom Scan Resolutions per FCC KDB Publication 865664 D01v04




Shenzhen Huatongwei International Inspection Co., Ltd.         Report Template Version: V01 (2018-08)


Report No:    CHTEW19030220                       Page: 15 of 30                       Issued: 2019-03-25

7.2. Data Storage and Evaluation
  Data Storage
  The DASY5 software stores the acquired data from the data acquisition electronics as raw data (in microvolt
  readings from the probe sensors), together with all necessary software parameters for the data evaluation
  (probe calibration data, liquid parameters and device frequency and modulation data) in measurement files
  with the extension “.DA4”. The software evaluates the desired unit and format for output each time the data
  is visualized or exported. This allows verification of the complete software setup even after the
  measurement and allows correction of incorrect parameter settings. For example, if a measurement has
  been performed with a wrong crest factor parameter in the device setup, the parameter can be corrected
  afterwards and the data can be re-evaluated.

  The measured data can be visualized or exported in different units or formats, depending on the selected
  probe type ([V/m], [A/m], [°C], [mW/g], [mW/cm²], [dBrel], etc.). Some of these units are not available in
  certain situations or show meaningless results, e.g., a SAR output in a lossless media will always be zero.
  Raw data can also be exported to perform the evaluation with other software packages.

  Data Evaluation
  The SEMCAD software automatically executes the following procedures to calculate the field units from the
  microvolt readings at the probe connector. The parameters used in the evaluation are stored in the
  configuration modules of the software:

  Probe parameters:       Sensitivity:                   Normi, ai0, ai1, ai2
                          Conversion factor:             ConvFi
                          Diode compression point:       Dcpi
  Device parameters:      Frequency:                     f
                          Crest factor:                  cf
  Media parameters:       Conductivity:                  σ
                          Density:                       ρ

  These parameters must be set correctly in the software. They can be found in the component documents or
  they can be imported into the software from the configuration files issued for the DASY5 components. In the
  direct measuring mode of the multimeter option, the parameters of the actual system setup are used. In the
  scan visualization and export modes, the parameters stored in the corresponding document files are used.

  The first step of the evaluation is a linearization of the filtered input signal to account for the compression
  characteristics of the detector diode. The compensation depends on the input signal, the diode type and the
  DC-transmission factor from the diode to the evaluation electronics. If the exciting field is pulsed, the crest
  factor of the signal must be known to correctly compensate for peak power. The formula for each channel
  can be given as:


  Vi:      compensated signal of channel ( i = x, y, z )
  Ui:      input signal of channel ( i = x, y, z )
  cf:      crest factor of exciting field (DASY parameter)
  dcpi:    diode compression point (DASY parameter)

  From the compensated input signals the primary field data for each channel can be evaluated:




  Vi:        compensated signal of channel ( i = x, y, z )
  Normi:     sensor sensitivity of channel ( i = x, y, z ),
             [mV/(V/m)2] for E-field Probes
  ConvF:     sensitivity enhancement in solution
  aij:       sensor sensitivity factors for H-field probes
  f:         carrier frequency [GHz]
  Ei:        electric field strength of channel i in V/m
  Hi:        magnetic field strength of channel i in A/m




Shenzhen Huatongwei International Inspection Co., Ltd.                  Report Template Version: V01 (2018-08)


Report No:    CHTEW19030220                      Page: 16 of 30                         Issued: 2019-03-25

The RSS value of the field components gives the total field strength (Hermitian magnitude):



  The primary field data are used to calculate the derived field units.



  SAR:       local specific absorption rate in mW/g
  Etot:      total field strength in V/m
  σ:         conductivity in [mho/m] or [Siemens/m]
  ρ:         equivalent tissue density in g/cm3

  Note that the density is normally set to 1 (or 1.06), to account for actual brain density rather than the density
  of the simulation liquid.




Shenzhen Huatongwei International Inspection Co., Ltd.                    Report Template Version: V01 (2018-08)


Report No:    CHTEW19030220                      Page: 17 of 30                        Issued: 2019-03-25

8. Position of the wireless device in relation to the phantom
8.1. Front-of-face
  A typical example of a front-of-face device is a two-way radio that is held at a distance from the face of the
  user when transmitting. In these cases the device under test shall be positioned at the distance to the
  phantom surface that corresponds to the intended use as specified by the manufacturer in the user
  instructions. If the intended use is not specified, a separation distance of 25 mm between the phantom
  surface and the device shall be used.




8.2. Body Position
  A typical example of a body-worn device is a mobile phone, wireless enabled PDA or other battery operated
  wireless device with the ability to transmit while mounted on a person’s body using a carry accessory
  approved by the wireless device manufacturer.




Shenzhen Huatongwei International Inspection Co., Ltd.                Report Template Version: V01 (2018-08)


Report No:   CHTEW19030220                          Page: 18 of 30                        Issued: 2019-03-25

9. System Check
9.1. Tissue Dielectric Parameters
It’s satisfying the latest tissue dielectric parameters requirements proposed by the KDB865664 D01.

                                         Targets for tissue simulating liquid
                                   Tissue dielectric parameters for head and body
     Target Frequency                              Head                                      Body
          (MHz)                          εr                 σ(s/m)                  εr               σ(s/m)
            150                         52.3                 0.76                  61.9               0.80

CheckResult:
                           Dielectric performance of Head tissue simulating liquid

                      εr                        σ(s/m)
Frequency                                                       Delta      Delta             Temp
                                                                                     Limit               Date
  (MHz)                                                          (εr)       (σ)              (℃)
             Target     Measured        Target     Measured

   150        52.30        53.22         0.76        0.75       1.76%     -0.92%     ±5%     22.5    2019-02-19


                           Dielectric performance of Body tissue simulating liquid

                      εr                        σ(s/m)
Frequency                                                       Delta      Delta             Temp
                                                                                     Limit               Date
  (MHz)                                                          (εr)       (σ)              (℃)
             Target     Measured        Target     Measured

   150        61.90        61.45         0.80        0.81      -0.72%     1.13%      ±5%     22.5    2019-02-20




Shenzhen Huatongwei International Inspection Co., Ltd.                   Report Template Version: V01 (2018-08)


Report No:   CHTEW19030220                     Page: 19 of 30                       Issued: 2019-03-25

9.2. SAR System Check
The purpose of the system check is to verify that the system operates within its specifications at the decice
test frequency.The system check is simple check of repeatability to make sure that the system works correctly
at the time of the compliance test;

System check results have to be equal or near the values determined during dipole calibration with the
relevant liquids and test system (±10%).
System check is performed regularly on all frequency bands where tests are performed with the DASY5
system.




                                      System Performance Check Setup




                                           Photo of Dipole Setup




Shenzhen Huatongwei International Inspection Co., Ltd.              Report Template Version: V01 (2018-08)


Report No:    CHTEW19030220                           Page: 20 of 30                           Issued: 2019-03-25

Check Result:


                                                           Head

                        1g SAR                         10g SAR
Frequency
                                                                             Delta   Delta           Temp
  (MHz)                                                                                      Limit             Date
             Target   Normalize   Measured   Target   Normalize   Measured   (1g)    (10g)            (℃)
              1W       to 1W       1W         1W       to 1W        1W

   150       3.68       3.86        3.86     2.45       2.56        2.56     4.89%   4.49%   ±10%    22.5    2019-02-19




                                                           Body

                        1g SAR                         10g SAR
Frequency
                                                                             Delta   Delta           Temp
  (MHz)                                                                                      Limit             Date
                                                                             (1g)    (10g)            (℃)
             Target   Normalize   Measured   Target   Normalize   Measured
              1W       to 1W       1W         1W       to 1W        1W

   150       3.75       3.98        3.98     2.50       2.59        2.59     6.13%   3.60%   ±10%    22.5    2019-02-20

Note:
1. the graph results see follow.




Shenzhen Huatongwei International Inspection Co., Ltd.                        Report Template Version: V01 (2018-08)


Report No:    CHTEW19030220                   Page: 21 of 30                     Issued: 2019—03—25

Plots of System Performance Check

SystemPerformanceCheck—Head 150MHz

DUT : Antenna 150 MHz; Type: CLA150; Serial: 4024
Date: 2019—02—19
Communication System: UID 0, A—CW (0); Frequency: 150 MHz;Duty Cycle: 1:1
Medium parameters used: f= 150 MHz; a = 0.753 S/m; s, = 53.218; p = 1000 kg/m3
Phantom section: Flat Section

DASY Configuration:
     e   Probe: EX3DV4 — SN7494; ConvF(13.63, 13.63, 13.63) @ 150 MHz; Calibrated: 2/26/2018
         Sensor—Surface: 1.4mm (Mechanical Surface Detection)
         Electronics: DAE4 Sn1549; Calibrated: 4/25/2018
         Phantom: ELI V8.0 ; Type: QD OVA 004 AA ; Serial: 2078
         DASY52 52.10.1(1476); SEMCAD X 14.6.12(7450)

Head/d=0mm, Pin=1W/Area Scan (81x81x1): interpolated grid: dx=1.500 mm, dy=1.500
mm
Maximum value of SAR (interpolated) = 5.50 W/kg

Head/d=0mm, Pin=1W/Zoom Scan (5x5x7) (5x5x7)/Cube 0: Measurement grid:
dx=8mm, dy=8mm, dz=5mm
Reference Value = 81.34 V/m; Power Drift = —0.11 dB
Peak SAR (extrapolated) = 6.87 W/kg
SAR(1 g) = 3.86 W/kg; SAR(10 g) = 2.56 W/kg
Maximum value of SAR (measured) = 5.52 W/kg
   dB




     —1.96




     —3.92




     —5.87




     —1.83




     —9.79
               0 dB = 5.52 W/kg = 7. 42 dBW/kg




Shenzhen Huatongwei International Inspection Co., Ltd.            Report Template Version: VO1 (2018—08)


Report No:    CHTEW19030220                    Page: 22 of 30                    Issued: 2019—03—25

SystemPerformanceCheck—Body 150MHz

DUT : Antenna 150 MHz; Type: CLA150; Serial: 4024
Date: 2019—02—20
Communication System: UID 0, A—CW (0); Frequency: 150 MHz;Duty Cycle: 1:1
Medium parameters used: f= 150 MHz; a = 0.809 S/m; s, = 61.453; p = 1000 kg/m3
Phantom section: Flat Section

DASY Configuration:
         Probe: EX3DV4 — SN7494; ConvF(12.81, 12.81, 12.81) @ 150 MHz; Calibrated: 2/26/2018
         Sensor—Surface: 1.4mm (Mechanical Surface Detection)
         Electronics: DAE4 Sn1549; Calibrated: 4/25/2018
         Phantom: ELI V8.0 ; Type: QD OVA 004 AA ; Serial: 2078
         DASY52 52.10.1(1476); SEMCAD X 14.6.12(7450)

Body/d=Omm, Pin=1W/Area Scan (81x81x1): interpolated grid: dx=1.500 mm, dy=1.500
mm
Maximum value of SAR (interpolated) = 5.89 W/kg

Body/d=O0mm, Pin=1W/Zoom Scan (5x5x7) (5x5x7)/Cube 0: Measurement grid:
dx=8mm, dy=8mm, dz=5mm
Reference Value = 80.36 V/m; Power Drift = —0.05 dB
Peak SAR (extrapolated) = 7.76 W/kg
SAR(1 g) = 3.98 W/kg:; SAR(10 g) = 2.59 W/kg
Maximum value of SAR (measured) = 5.92 W/kg
   dB




     —2.07




     —4.14




     —6.22




     —8.29




     —10.36
               0 dB = 5.92 W/kg = 7. 72 dBW/kg




Shenzhen Huatongwei International Inspection Co., Ltd.            Report Template Version: VO1 (2018—08)


Report No:     CHTEW19030220                   Page: 23 of 30                       Issued: 2019-03-25

10. SAR Exposure Limits
SAR assessments have been made in line with the requirements of FCC 47 CFR § 2.1093.

                                                                   Limit (W/kg)
             Type Exposure                    General Population /                   Occupational /
                                       Uncontrolled Exposure Environment   Controlled Exposure Environment
 Spatial Average SAR
                                                     0.08                                0.4
 (whole body)
 Spatial Peak SAR
                                                      1.6                                8.0
 (1g cube tissue for head and trunk)
 Spatial Peak SAR
                                                      4.0                                20.0
 (10g for limb)

Population/Uncontrolled Environments: are defined as locations where there is the exposure of individual who
have no knowledge or control of their exposure.

Occupational/Controlled Environments: are defined as locations where there is exposure that may be incurred
by people who are aware of the potential for exposure (i.e. as a result of employment or occupation).




Shenzhen Huatongwei International Inspection Co., Ltd.              Report Template Version: V01 (2018-08)


Report No:      CHTEW19030220                 Page: 24 of 30                         Issued: 2019-03-25

11. Conducted Power Measurement Results
                                                   PMR
                           Channel                          Frequency                     Conducted Power
       Mode
                          Separation              Channel                 MHz                 (dBm)
                                                   CH1                  136.0250               36.90
                                                   CH2                  145.5000               36.70
      Analog               12.5KHz                 CH3                  155.0000               36.50
                                                   CH4                  164.5000               36.70
                                                   CH5                  173.9750               36.80
                                                   CH1                  136.0250               36.50
                                                   CH2                  145.5000               36.30
       Digtal              12.5KHz                 CH3                  155.0000               36.30
                                                   CH4                  164.5000               36.40
                                                   CH5                  173.9750               36.50

Duty Factor Measured Results
                                          T on               Period                          Crest Factor
     Mode              Type                                                  Duty Cycle
                                          (ms)                (ms)                          (1/duty cycle)
     Digtal            4FSK               27.61              60.502           45.64%              2.1913
Duty Cycle plot




Bluetooth Conducted Power
                                              Bluetooth
        Mode                    Channel                  Frequency (MHz)           Conducted power (dBm)
                                   0                          2402                         7.97
        GFSK                      39                          2441                         8.72
                                  78                          2480                         8.76
                                   0                          2402                         6.51
      π/4QPSK                     39                          2441                         7.47
                                  78                          2480                         7.52
                                   0                          2402                         6.86
       8DPSK                      39                          2441                         7.78
                                  78                          2480                         7.84
                                   0                          2402                         -1.34
         BLE                      19                          2440                         0.15
                                  39                          2480                         -0.02




Shenzhen Huatongwei International Inspection Co., Ltd.                Report Template Version: V01 (2018-08)


Report No:    CHTEW19030220                        Page: 25 of 30                          Issued: 2019-03-25



12. Maximum Tune-up Limit
This device operates using the following maximum output power specifications. SAR values were scaled to
the maximum allowed power to determine compliance per KDB publication 447498 D01.

                                                       PMR
                              Channel Separation         Operation Frequency              Maximum tune-up power
        Mode
                                   (KHz)                    Range (MHz)                          (dBm)
       Analog                                                                                      36.99
                                     12.5                     136~174
        Digtial                                                                                    36.99


                                                   Bluetooth
             Mode                                            Maximum Tune-up (dBm)
             GFSK                                                          9.00

          π/4QPSK                                                          8.00

             8DPSK                                                         8.00

         GFSK(BLE)                                                         0.50


Per KDB 447498 D01, the 1-g and 10-g SAR test exclusion thresholds for 100MHz to 6GHz at test
separation distances ≦50mm are determined by:
[(max. Power of channel, including tune-up tolerance, mW) / (min. test separation distance, mm)] *
[√f(GHz)] ≦3.0 for 1-g SAR
                                                      Tune-up Power          Separation Distance
                  Frequency                                                                           Exclusion
Band/Mode                           Position
                    (GHz)                            dBm            mW             (mm)              Thresholds

                                 Front-of-face                                       25                    0.5
 Bluetooth          2.45                             9.00           7.94
                                  Body-worn                                          0                     2.5
Per KDB 447498 D01, when the minimum test separation distance is <5mm, a distance of 5mm is
applied to determine SAR test exclusion.
The test exclusion thereshold is ≦3, SAR testing is not required.




Shenzhen Huatongwei International Inspection Co., Ltd.                     Report Template Version: V01 (2018-08)


Report No:      CHTEW19030220                            Page: 26 of 30                               Issued: 2019-03-25

13. SAR Measurement Results

                                                         Front-of-face
                                                                     Tune-                 Measured    Report    50% Duty
                             Frequency       Conducted     Tune-
            Channel                                                     up      Power      SAR(1g)    SAR(1g)   factor SAR   Test
  Mode                                         Power      up limit
           Separation                                                scaling   Drift(dB)                                     Plot
                         CH         MHz        (dBm)      (dBm)                             (W/kg)     (W/kg)    (W/kg)
                                                                      factor

                        CH1       136.0250     36.90       36.99      1.02       0.11       0.500      0.510      0.255       -

                        CH2       145.5000     36.70       36.99      1.07         -          -          -          -         -

 Analog     12.5KHz     CH3       155.0000     36.50       36.99      1.12      -0.15       2.170      2.430      1.215       1

                        CH4       164.5000     36.70       36.99      1.07         -          -          -          -         -

                        CH5       173.9750     36.80       36.99      1.04      -0.15       2.190      2.278      1.139       -

                        CH1       136.0250     36.50       36.99      1.12       0.18       0.181      0.203      0.101       -

                        CH2       145.5000     36.30       36.99      1.17         -          -          -          -         -

 Digtial    12.5KHz     CH3       155.0000     36.30       36.99      1.17       0.05       1.140      1.334      0.667       3

                        CH4       164.5000     36.40       36.99      1.15         -          -          -          -         -

                        CH5       173.9750     36.50       36.99      1.12       0.03       1.087      1.217      0.609       -

 Worse case mode(DP585-01)

 Analog     12.5KHz     CH3       155.0000     36.50       36.99      1.12       0.14       2.090      2.341      1.170

Note:
1. The value with blue color is the maximum SAR Value of each test band.
2. Batteries are fully charged at the beginning of the SAR measurements.
3. When the SAR for all antennas tested using the default battery is ≤ 3.5 W/kg (50% PTT duty factor),
   testing of all other required channels is not necessary.
4. When the SAR of an antenna tested on the highest output power using the default battery is > 3.5 W/Kg
   and ≤ 4.0 W/Kg (50% PTT duty factor), testing of the immediately adjacent channel(s) is not necessary,
   but testing of other required channels may still be required.




Shenzhen Huatongwei International Inspection Co., Ltd.                            Report Template Version: V01 (2018-08)


Report No:      CHTEW19030220                         Page: 27 of 30                               Issued: 2019-03-25

                                                    Body-worn (Rear)
                                                                  Tune-                 Measured    Report    50% Duty
                          Frequency       Conducted     Tune-
            Channel                                                  up      Power      SAR(1g)    SAR(1g)   factor SAR   Test
 Mode                                       Power      up limit
           Separation                                             scaling   Drift(dB)                                     Plot
                        CH       MHz        (dBm)      (dBm)                             (W/kg)     (W/kg)    (W/kg)
                                                                   factor

                        CH1    136.0250     36.90       36.99      1.02      -0.12       8.760      8.935      4.468       2

                        CH2    145.5000     36.70       36.99      1.07      -0.03       2.327      2.490      1.245       -

 Analog    12.5KHz      CH3    155.0000     36.50       36.99      1.12       0.01       2.410      2.699      1.350       -

                        CH4    164.5000     36.70       36.99      1.07       0.05       1.972      2.110      1.055       -

                        CH5    173.9750     36.80       36.99      1.04       0.14       1.890      1.966      0.983       -

                        CH1    136.0250     36.50       36.99      1.12      -0.12       4.490      5.029      2.514       4

                        CH2    145.5000     36.30       36.99      1.17         -          -          -          -         -

 Digtial   12.5KHz      CH3    155.0000     36.30       36.99      1.17       0.14       1.100      1.287      0.644       -

                        CH4    164.5000     36.40       36.99      1.15         -          -          -          -         -

                        CH5    173.9750     36.50       36.99      1.12       0.11       1.123      1.258      0.629       -

Note:
1. The value with blue color is the maximum SAR Value of each test band.
2. Batteries are fully charged at the beginning of the SAR measurements.
3. The Body-worn SAR evaluation was performed with the Leather Case body-worn accessory attached to
   the DUT and touching the outer surface of the planar phantom.
4. When the SAR for all antennas tested using the default battery is ≤ 3.5 W/kg (50% PTT duty factor),
   testing of all other required channels is not necessary.
5. When the SAR of an antenna tested on the highest output power using the default battery is > 3.5 W/Kg
   and ≤ 4.0 W/Kg (50% PTT duty factor), testing of the immediately adjacent channel(s) is not necessary,
   but testing of other required channels may still be required.




SAR Test Data Plots to the Appendix A.




Shenzhen Huatongwei International Inspection Co., Ltd.                         Report Template Version: V01 (2018-08)


Report No:       CHTEW19030220                Page: 28 of 30                         Issued: 2019-03-25

14. Simultaneous Transmission analysis

No.      Simultaneous Transmission Configurations        Front-of-face   Body-worn       Note
 1               PMR + Bluetooth(data)                    Yes            Yes
General note:
1. The reported SAR summation is calculated based on the same configuration and test position.
2. For simultaneous transmission analysis, Bluetooth SAR is estimated per KDB 447498 D01 based on the
   formula below
    a) [(max. Power of channel, including tune-up tolerance, mW) / (min. test separation distance, mm)] *
         [√f(GHz)/x]W/kg for test separation distances ≦50mm; whetn x=7.5 for 1-g SAR, and x=18.75 for
         10-g SAR.
    b) When the minimum separation distance is <5mm, the distance is used 5mm to determine SAR test
         exclusion
    c) 0.4 W/kg for 1-g SAR and 1.0W/kg for 10-g SAR, when the test separation distances is >50mm.

   Bluetooth           Exposure position        Front-of-face            Body-worn
   Max power            Test separation               25mm                 0mm
      9.00 dBm       Estimated SAR (W/kg)             0.066                0.332

According to KDB 447498 Section 7.2,
The [Σ of (the highest measured or estimated SAR for each standalone antenna configuration, adjusted for
maximum tune-up tolerance) / 1.6 W/kg] + [Σ of MPE ratios] is ≤ 1.0.
                                   Max SAR (W/kg)
      Exposure
                                                                          Ratios
      Position
                           PMR                 Bleutooth

  Front of Face            1.215                    0.066                   0.2

      Body-worn            4.468                    0.332                   0.8




Shenzhen Huatongwei International Inspection Co., Ltd.               Report Template Version: V01 (2018-08)


Report No:     CHTEW19030220                   Page: 29 of 30                    Issued: 2019—03—25

15. Test Setup Photos




             Liquid depth in the ELI phantom               Front—of—face(25mm) DP580—01_136~151MHz




     Body—worn(Omm) DP580—01_151~162MHz                    Front—of—face(25mm) DP580—01_162~174MHz




Shenzhen Huatongwei International Inspection Co., Ltd.            Report Template Version: VO1 (2018—08)


Report No:   CHTEW19030220                    Page: 30 of 30                  Issued: 2019-03-25




     Body-worn(0mm) DP580-01_162~174MHz                  Front-of-face(25mm) DP585-01_151~162MHz


16. External and Internal Photos of the EUT
Please refer to the test report No.: CHTEW19030201


                                    ---------End of Report---------




Shenzhen Huatongwei International Inspection Co., Ltd.          Report Template Version: V01 (2018-08)



Document Created: 2019-03-28 09:12:09
Document Modified: 2019-03-28 09:12:09

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC