SAR test report 1 of 3

FCC ID: PV7-WIBEAR11N-DF2

RF Exposure Info

Download: PDF
FCCID_4378240

                                                                                              SAR Evaluation Report
                                                                                                         EUT Information
                                                        Manufacturer               Remote Diagnostic Technologies Ltd
                                                        Model Name                 Tempus Pro
                                                        FCC ID                     contains NCMOMO6012; PV7-WIBEAR11N-DF2
                                                        IC Number                  contains 2734A-MO6012; 7738A-WB11NDF2
                                                        EUT Type                   medical vital signs monitor

                                                                                                           Prepared by
                                                                                   IMST GmbH, Test Center
                                                                                   Carl-Friedrich-Gauß-Str. 2 – 4
                                                        Testing Laboratory
                                                                                   47475 Kamp-Lintfort
                                                                                   Germany

                                                                                                    The Test Center facility ‘Dosimetric Test Lab’ within IMST GmbH is accredited by
                                                                                                    the German National ‘Deutsche Akkreditierungsstelle GmbH (DAkkS)’ for testing
                                                                                                    according to the scope as listed in the accreditation certificate: D-PL-12139-01-00.

                                                                                                    The German Bundesnetzagentur (BNetzA) recognizes IMST GmbH as CAB-EMC
                                                                                                    on the basis of the Council Decision of 22. June 1998 concerning the conclusion
                                                        Laboratory Accreditation                    of the MRA between the European Community and the United States of America
                                                                                                    (1999/178/EC) in accordance with § 4 of the Recognition Ordinance of 11.
                                                                                                    January 2016. The recognition is valid until 20. July 2021 under the registration
                                                                                                    number: BNetzA-CAB-16/21-14.
                                                                                                    IMST GmbH is recognized as a wireless testing laboratory to perform equipment
                                                                                                    testing in Canada under the terms of Phase I of the Canada-EU Comprehensive
                                                                                                    Economic and Trade Agreement (CETA) under CAB identifier: DE0010.

                                                                                                           Prepared for
                                                                                   Remote Diagnostic Technologies Ltd
                                                                                   Pavilion C2 Ashwood Park, Ashwood Way,
                                                        Applicant / Manufacturer
                                                                                   Basingstoke, Hampshire RG23 8BG
                                                                                   United Kingdom

                                                                                                         Test Specification
                                                        Applied Rules/Standards    IEEE 1528-2013, FCC CFR 47 § 2.1093, RSS-102 Issue 5
                                                        Usage Configuration        body-supported conditions (lap-held)
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                        Exposure Category             general public / uncontrolled exposure              occupational / controlled exposure
                                                        Test Result                   PASS                                                FAIL

                                                                                                       Report Information
SAR_Report_FCC_ISED_Extremity_v2.0




                                                        Data Stored                60320_6190299
SAR_Report_FCC_ISED_Body_v2.1




                                                        Issue Date                 April 15, 2019
                                                        Revision Date              July 09, 2019
                                                        Revision Number            1 (A new revision replaces all previous revisions and thus, become invalid herewith)

                                                                                   This report relates only to the item(s) evaluated. This report shall not be reproduced, except in its
                                                                                   entirety, without the prior written approval of IMST GmbH. The results and statements contained in
                                                        Remarks                    this report reflect the evaluation for the certain model described above. The manufacturer is
                                                                                   responsible for ensuring that all production devices meet the intent of the requirements described
                                              444/AR




                                                                                   in this report.



                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                    Page 1 of 109


                                                       Issue Date: April 15, 2019                                             Revision Date: July 09, 2019                                        Revision No.: 1


                                                       Table of Contents

                                                       1        Subject of Investigation .......................................................................................................................... 3
                                                            1.1          Technical Data of EUT ..................................................................................................................................... 3
                                                            1.2          Picture of the DUT and Antenna Location ........................................................................................................ 3
                                                            1.3          Test Specification / Normative References ...................................................................................................... 5
                                                            1.4          Attestation of Test Results ............................................................................................................................... 5
                                                       2        Exposure Criteria and Limits .................................................................................................................. 6
                                                            2.1          SAR Limits ....................................................................................................................................................... 6
                                                            2.2          Exposure Categories ........................................................................................................................................ 6
                                                            2.3          Distinction between Maximum Permissible Exposure and SAR Limits ............................................................ 6
                                                       3        The Measurement System....................................................................................................................... 7
                                                            3.1          Phantoms ......................................................................................................................................................... 8
                                                            3.2          E-Field-Probes ................................................................................................................................................. 9
                                                       4        Measurement Procedure .......................................................................................................................10
                                                            4.1          General Requirement ..................................................................................................................................... 10
                                                            4.2          Measurement Procedure ................................................................................................................................ 10
                                                            4.3          Additional Information for IEEE 802.11 (WiFi) Transmitters ........................................................................... 11
                                                            4.4          Measurement Variability ................................................................................................................................. 12
                                                       5        System Verification and Test Conditions ............................................................................................13
                                                            5.1          Date of Testing ............................................................................................................................................... 13
                                                            5.2          Environment Conditions ................................................................................................................................. 13
                                                            5.3          Tissue Simulating Liquid Recipes .................................................................................................................. 13
                                                            5.4          Tissue Simulating Liquid Parameters ............................................................................................................. 14
                                                            5.5          Simplified Performance Checking .................................................................................................................. 15
                                                       6        Measurement Conditions ......................................................................................................................15
                                                            6.1          SAR Test Conditions ...................................................................................................................................... 15
                                                            6.2          Tune-Up Information ...................................................................................................................................... 16
                                                            6.3          Measured Output Power ................................................................................................................................ 17
                                                            6.4          Standalone SAR Test Exclusion .................................................................................................................... 20
                                                       7        SAR Test Results ...................................................................................................................................21
                                                       8        Simultaneous Transmission Consideration ........................................................................................22
                                                       9        Administrative Measurement Data .......................................................................................................23
                                                            9.1          Calibration of Test Equipment ........................................................................................................................ 23
                                                            9.2          Uncertainty Assessment ................................................................................................................................ 24
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                       10       Report History ........................................................................................................................................26
                                                            Appendix A - Pictures ................................................................................................................................................... 28
                                                            Appendix B - SAR Distribution Plots ............................................................................................................................. 31
                                                            Appendix C - System Verification Plots ........................................................................................................................ 36
SAR_Report_FCC_ISED_Body_v2.1




                                                            Appendix D – Certificates of Conformity ....................................................................................................................... 39
                                                            Appendix E – Calibration Certificates for DAEs ............................................................................................................ 42
                                                            Appendix F – Calibration Certificates for E-Field Probes.............................................................................................. 52
                                                            Appendix G – Calibration Certificates for Dipoles ......................................................................................................... 73
                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                                                                   Page 2 of 109


                                                       Issue Date: April 15, 2019                     Revision Date: July 09, 2019                    Revision No.: 1


                                                       1 Subject of Investigation
                                                       The Tempus Pro is medical vital signs monitor. It is operating in 850 MHz, 1900 MHz and 2.4 GHz frequency
                                                       range and has three integrated antennas.

                                                       1.1 Technical Data of EUT
                                                                                                     Product Specifications
                                                        Manufacturer                 Remote Diagnostic Technologies     Ltd
                                                        Model Name / Number          00-1026-R
                                                        Brand Name                   Tempus Pro
                                                        Serial Number of DUT         602426
                                                        IMST DUT Number              01
                                                        Hardware Version             Trizeps VII
                                                        Software Version             V7.01
                                                        Integrated Transmitter       Option GTM601W                                     U-Blox ELLA W-163
                                                        Operation Mode               GSM/GPRS 850/1900, WCDMA B2/B5                     IEEE 802.11b/g/n; Bluetooth V2.3/V3.0
                                                        Operation Frequency Range    850 MHz, 1900 MHz                                  2.4 GHz
                                                        TX Antenna Type              integrated (1x WWAN, 1x WLAN, 1x BT)
                                                        Usage Configuration          body-supported conditions (lap-held)
                                                        Max. Output Power            refer to chapter 6.2
                                                        Power Supply                 internal Li-Ion 7.4V 10.2Ah 75.5 Wh
                                                        DUT Stage                         production unit         identical prototype
                                                        Notes: -


                                                       1.2 Picture of the DUT and Antenna Location
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                                                                   Fig. 1:    Front view of the DUT.



                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                 Page 3 of 109


                                                       Issue Date: April 15, 2019                        Revision Date: July 09, 2019            Revision No.: 1




                                                                                                   Fig. 2:    Antenna location of the DUT.




                                                                                                   Antenna Type and Location
                                                       Antenna                             BLUETOOTH                           WWAN                      WLAN
                                                       Reference                              A1                                 A2                        A3

                                                                                                         Configuration Edge
                                                       Side of DUT                  Left               Top             Right            Bottom   Back              Front
                                                       Reference                    (1)                (2)              (3)              (4)      (5)               (6)
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                            Page 4 of 109


                                                       Issue Date: April 15, 2019                                       Revision Date: July 09, 2019                    Revision No.: 1


                                                       1.3 Test Specification / Normative References
                                                       The tests documented in this report were performed according to the standards and rules described below.

                                                                                                                         Test Specifications
                                                               Test Standard / Rule                                                  Description                                         Issue Date
                                                                                                   IEEE Recommended Practice for Determining the Peak Spatial-
                                                              IEEE 1528-2013                       Average Specific Absorption Rate (SAR) in the Human Head from                  June 14, 2013
                                                                                                   Wireless Communications Devices: Measurement Techniques.
                                                                                                   Code of Federal Regulations; Title 47.
                                                              FCC CFR 47 § 2.1091                                                                                                 October 01, 2010
                                                                                                   Radiofrequency radiation exposure evaluation: Mobile Devices.
                                                                                                   Code of Federal Regulations; Title 47.
                                                              FCC CFR 47 § 2.1093                                                                                                 October 01, 2010
                                                                                                   Radiofrequency radiation exposure evaluation: Portable Devices.
                                                                                                   Radio Frequency (RF) Exposure Compliance of
                                                              RSS-102, Issue 5                                                                                                    March, 2015
                                                                                                   Radiocommunication Apparatus (All Frequency Bands)
                                                                                                                 Measurement Methodology KDB
                                                              KDB 865664 D01 v01r04                SAR measurement 100 MHz to 6 GHz                                               August 07, 2015
                                                              KDB 865664 D02 v01r01                Exposure Reporting                                                             October 23, 2015
                                                                                                                               Product KDB
                                                              KDB 447498 D01 v06                   General RF Exposure Guidance                                                   October 23, 2015
                                                              KDB 616217 D04 v01r02                SAR for Laptop and Tablets                                                     October 23, 2015
                                                                                                                             Technology KDB
                                                              KDB 248227 D01 v02r02                802.11 Wi-Fi SAR                                                               October 23, 2015

                                                       Table 1:     Normative references.

                                                       1.4 Attestation of Test Results
                                                                                                            Highest Reported SAR1g [W/kg]
                                                          Equipment                                                                                                                      SAR1g Limit
                                                                                 Band                  Body / Back Side of DUT                Simultaneous Transmission
                                                            Class                                                                                                                          [W/kg]

                                                              PCE            WCDMA B5                               0.027

                                                              DTS           WLAN 2.4 GHz                            0.006                               0.427                         1.6              PASS

                                                              DSS             Bluetooth                            0.400*

                                                        Notes: To establish a connection at a specific channel and with maximum output power, base station simulator
                                                               has been used for 2G and 3G testing. Engineering test software has been used for WLAN measurements.
                                                               All measured SAR results and considered configurations are shown in chapter 7

                                                       Table 2:     Highest reported SAR results.
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1




                                                       Prepared by:          ........................................                    Reviewed by:       ........................................

                                                                             Alexander Rahn                                                                 Dessislava Patrishkova
                                                                             Test Engineer                                                                  Quality Assurance
                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                                        Page 5 of 109


                                                       Issue Date: April 15, 2019                           Revision Date: July 09, 2019                      Revision No.: 1


                                                       2 Exposure Criteria and Limits

                                                       2.1 SAR Limits
                                                                                                          Human Exposure Limits
                                                                                                                         Uncontrolled Environment             Controlled Environment
                                                                                                                            (General Population)                  (Occupational)
                                                                                Condition
                                                                                                                        SAR Limit                            SAR Limit
                                                                                                                                            Mass Avg.                            Mass Avg.
                                                                                                                         [W/kg]                               [W/kg]

                                                        SAR averaged over the whole body mass                               0.08            whole body          0.4              whole body

                                                        Peak spatially-averaged SAR for the head, neck & trunk                  1.6        1g of tissue*        8.0             1g of tissue*

                                                        Peak spatially-averaged SAR in the limbs                                4.0        10g of tissue*       20.0            10g of tissue*

                                                        Note: *Defined as a tissue volume in the shape of a cube

                                                       Table 3:     SAR limits specified in IEEE Standard C95.1-2005 and Health Canada’s Safety Code 6.

                                                       In this report the comparison between the exposure limits and the measured data is made using the spatial
                                                       peak SAR; the power level of the device under test guarantees that the whole body averaged SAR is not
                                                       exceeded.

                                                       2.2 Exposure Categories
                                                                                                General Public / Uncontrolled Exposure
                                                        General population comprises individuals of all ages and of varying health status, and may include particularly susceptible groups or
                                                        individuals. In many cases, members of the public are unaware of their exposure to electromagnetic fields. Moreover, individual
                                                        members of the public cannot reasonably be expected to take precautions to minimize or avoid exposure.

                                                                                                   Occupational / Controlled Exposure
                                                        The occupationally exposed population consists of adults who are generally exposed under known conditions and are trained to be
                                                        aware of potential risk and to take appropriate precautions.

                                                       Table 4:      RF exposure categories.

                                                       2.3 Distinction between Maximum Permissible Exposure and SAR Limits
                                                       The biological relevant parameter describing the effects of electromagnetic fields in the frequency range of
                                                       interest is the specific absorption rate SAR (dimension: power/mass). It is a measure of the power absorbed
                                                       per unit mass. The SAR may be spatially averaged over the total mass of an exposed body or its parts. The
                                                       SAR is calculated from the r.m.s. electric field strength E inside the human body, the conductivity  and the
                                                       mass density  of the biological tissue:
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                                                                    E2    T
                                                                                            SAR      c
                                                                                                         t           t 0 
                                                                                                                                      .                                                 (1)


                                                                                                                                                    T /  t as a function of the specific
SAR_Report_FCC_ISED_Body_v2.1




                                                       The specific absorption rate describes the initial rate of temperature rise
                                                       heat capacity c of the tissue. A limitation of the specific absorption rate prevents an excessive heating of the
                                                       human body by electromagnetic energy.

                                                       As it is sometimes difficult to determine the SAR directly by measurement (e.g. whole body averaged SAR),
                                                       the standard specifies more readily measurable maximum permissible exposures in terms of external electric
                                                       E and magnetic field strength H and power density S, derived from the SAR limits. The limits for E, H and S
                                              444/AR




                                                       have been fixed so that even under worst case conditions, the limits for the specific absorption rate SAR are
                                                       not exceeded.

                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                           Page 6 of 109


                                                       Issue Date: April 15, 2019                     Revision Date: July 09, 2019          Revision No.: 1


                                                       3 The Measurement System

                                                       DASY is an abbreviation of „Dosimetric Assessment System“ and describes a system that is able to
                                                       determine the SAR distribution inside a phantom of a human being according to different standards. The
                                                       DASY4 system consists of the following items as shown in Fig: 3. Additionally, Fig: 4 shows the equipment,
                                                       similar to the installations in other laboratories.

                                                              Fully compliant with all current measurement standards as stated in Fig. 5
                                                              High precision robot with controller
                                                              Measurement server (for surveillance of the robot operation and signal filtering)
                                                              Data acquisition electronics DAE (for signal amplification and filtering)
                                                              Field probes calibrated for use in liquids
                                                              Electro-optical converter EOC (conversion from the optical into a digital signal)
                                                              Light beam (improving of the absolute probe positioning accuracy)
                                                              Two SAM phantoms filled with tissue simulating liquid
                                                              DASY4 software
                                                              SEMCAD




                                                       Fig. 3:   The DASY4 measurement system.
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                       Page 7 of 109


                                                       Issue Date: April 15, 2019                            Revision Date: July 09, 2019                  Revision No.: 1




                                                       Fig. 4:   The measurement set-up with a DASY system and phantoms containing tissue simulating liquid.

                                                       The EUT operating at the maximum power level is placed by a non-metallic device holder (delivered from
                                                       Schmid & Partner) in the above described positions at a shell phantom of a human being. The distribution of
                                                       the electric field strength E is measured in the tissue simulating liquid within the shell phantom. For this
                                                       miniaturised field probes with high sensitivity and low field disturbance are used. Afterwards the corre-
                                                       sponding SAR values are calculated with the known electrical conductivity  and the mass density  of the
                                                       tissue in the SEMCAD FDTD software. The software is able to determine the averaged SAR values
                                                       (averaging region 1 g or 10 g) for compliance testing.

                                                       The measurements are done by two scans: first a coarse scan determines the region of the maximum SAR
                                                       afterwards the averaged SAR is measured in a second scan within the shape of a cube.

                                                       3.1 Phantoms
                                                                                                        TWIN SAM PHANTOM V4.0
                                                                                         Specific Anthropomorphic Mannequin defined in IEEE 1528 and IEC 62209-1 and delivered by
                                                                                         Schmid & Partner Engineering AG. It enables the dosimetric evaluation of left and right hand
                                                                                         phone usage as well as body mounted usage at the flat phantom region.
                                                                                         The details and the Certificate of conformity can be found in Fig. 6.
                                                        Shell Thickness                  2 ± 0.2 mm (6 ± 0.2 mm at ear point)
                                                                                         Length: 1000 mm; Width: 500 mm
                                                        Dimensions
                                                                                         Height: adjustable feet
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                        Filling Volume                   approx. 25 liters




                                                                                                              ELI PHANTOM V4.0
SAR_Report_FCC_ISED_Body_v2.1




                                                                                         Phantom for compliance testing of handheld and body-mounted wireless devices in the frequency
                                                                                         range of 30 MHz to 6 GHz.
                                                                                         The details and the Certificate of conformity can be found in Fig. 8

                                                        Shell Thickness                  2.0 ± 0.2 mm (bottom plate)
                                                                                         Major axis: 600 mm
                                                        Dimensions
                                                                                         Minor axis: 400 mm
                                              444/AR




                                                        Filling Volume                   approx. 30 liters




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                      Page 8 of 109


                                                       Issue Date: April 15, 2019                        Revision Date: July 09, 2019                     Revision No.: 1


                                                       3.2 E-Field-Probes
                                                       For the measurements the Dosimetric E-Field Probes ET3DV6R or EX3DV4 with following specifications are
                                                       used. They are manufactured and calibrated in accordance with FCC and IEEE 1528-2013
                                                       recommendations annually by Schmid & Partner Engineering AG.

                                                                                                                  ET3DV6R
                                                                                    Symmetrical design with triangular core
                                                                                    Built-in optical fiber for surface detection system (ET3DV6 only)
                                                        Construction
                                                                                    Built-in shielding against static charges
                                                                                    PEEK enclosure material (resistant to organic solvents, e.g., DGBE)

                                                                                    Overall length: 337 mm (Tip: 16 mm)
                                                        Dimensions                  Tip diameter: 6.8 mm (Body: 12 mm)
                                                                                    Distance from probe tip to dipole centers: 2.7 mm

                                                                                    10 MHz to 2.3 GHz
                                                        Frequency
                                                                                    Linearity: ± 0.2 dB (30 MHz to 2.3 GHz)

                                                                                    Axial isotropy: ± 0.2 dB in TSL (rotation around probe axis)
                                                        Directivity
                                                                                    Spherical isotropy: ± 0.4 dB in TSL (rotation normal to probe axis)

                                                        Dynamic Range               5 µW/g to > 100 mW/g; Linearity: ± 0.2 dB

                                                        Calibration Range           450 MHz / 750 MHz / 900 MHz / 1750 MHz / 1900 MHz for head and body simulating liquid



                                                                                                                   EX3DV4
                                                                                    Symmetrical design with triangular core
                                                        Construction                Built-in shielding against static charges
                                                                                    PEEK enclosure material (resistant to organic solvents, e.g., DGBE)

                                                                                    Overall length: 337 mm (Tip: 20 mm)
                                                        Dimensions                  Tip diameter: 2.5 mm (Body: 12 mm)
                                                                                    Typical distance from probe tip to dipole centers: 1 mm

                                                                                    10 MHz to > 6 GHz
                                                        Frequency
                                                                                    Linearity: ± 0.2 dB (30 MHz to 6 GHz)

                                                                                    Axial isotropy: ± 0.3 dB in TSL (rotation around probe axis)
                                                        Directivity
                                                                                    Spherical isotropy: ± 0.5 dB in TSL (rotation normal to probe axis)

                                                                                    10 µW/g to > 100 mW/g
                                                        Dynamic Range
                                                                                    Linearity: ± 0.2 dB (noise: typically < 1 µW/g)

                                                        Calibration Range           2450 MHz / 2600 MHz / 5250 MHz / 5600 MHz / 5800 MHz for head and body simulating liquid
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                     Page 9 of 109


                                                       Issue Date: April 15, 2019                            Revision Date: July 09, 2019                       Revision No.: 1


                                                       4 Measurement Procedure

                                                       4.1 General Requirement
                                                       The test shall be performed in a laboratory with an environment which avoids influence on SAR
                                                       measurements by ambient EM sources and any reflection from the environment itself. The ambient
                                                       temperature shall be in the range of 20°C to 26°C and 30-70% humidity. All tests have been conducted
                                                       according the latest version of all relevant KDBs.

                                                       4.2 Measurement Procedure
                                                       The following steps are used for each test position:

                                                                 Establish a call with the maximum output power with a base station simulator. The connection
                                                                  between the mobile phone and the base station simulator is established via air interface.
                                                                 Measurement of the local E-field value at a fixed location (P1). This value serves as a reference
                                                                  value for calculating a possible power drift.
                                                                 Measurement of the SAR distribution with resolution settings for area scan and zoom scan according
                                                                  KDB 865664 D01 as shown in Table 5.
                                                                 The used extrapolation and interpolation routines are all based on the modified Quadratic Shepard´s
                                                                  method [DASY4].
                                                                 Repetition of the E-field measurement at the fixed location (P1) and repetition of the whole
                                                                  procedure if the two results differ by more than  0.21dB.

                                                                                                                                            ≤ 3 GHz                          ≥ 3 GHz
                                                        Maximum      distance     from    closest    measurement       point
                                                                                                                                            5 ± 1 mm                   ½δln(2) ± 0.5 mm
                                                        (geometric center of probe sensors) to phantom surface
                                                        Maximum probe angle from probe axis to phantom surface normal
                                                                                                                                            30° ± 1°                         20° ± 1°
                                                        at the measurement location
                                                                                                                                     ≤ 2 GHz: ≤ 15 mm                  3 - 4 GHz: ≤ 12 mm
                                                                                                                                    2 - 3 GHz: ≤ 12 mm                 4 - 6 GHz: ≤ 10 mm
                                                                                                                               When the x or y dimension of the test device, in the
                                                        Maximum area scan spatial resolution: Δx Area, Δy Area                 measurement plane orientation, is smaller than the above,
                                                                                                                               the measurement resolution must be ≤ the corresponding x or
                                                                                                                               y dimension of the test device with at least one measurement
                                                                                                                               point on the test device.
                                                                                                                                     ≤ 2 GHz: ≤ 8 mm                   3 - 4 GHz: ≤ 5 mm*
                                                        Maximum zoom scan spatial resolution: ΔX Zoom, ΔY Zoom
                                                                                                                                    2 - 3 GHz: ≤ 5 mm*                 4 - 6 GHz: ≤ 4 mm*
                                                                                                                                                                       3 - 4 GHz: ≤ 4 mm
                                                                                                                                                                       4 - 5 GHz: ≤ 3 mm
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                        Maximum zoom       Uniform grid: ΔZ Zoom(n)                                         ≤ 5 mm
                                                                                                                                                                       5 - 6 GHz: ≤ 2 mm
                                                        scan spatial
                                                        resolution,                                                                                                    3 - 4 GHz: ≤ 3 mm
                                                        normal to                      ΔZ Zoom(1): between 1st two points
                                                                                                                                            ≤ 4 mm                     4 - 5 GHz: ≤ 2.5 mm
                                                        phantom            graded      closest to phantom surface
                                                                                                                                                                       5 - 6 GHz: ≤ 2 mm
                                                        surface            grid
SAR_Report_FCC_ISED_Body_v2.1




                                                                                       ΔZ Zoom(n>1): between subsequent
                                                                                                                                                       ≤ 1.5 ΔZ Zoom(n-1)
                                                                                       points
                                                                                                                                                                       3 - 4 GHz: ≥ 28 mm
                                                        Minimum zoom
                                                                           x, y, z                                                          ≥ 30 mm                    4 - 5 GHz: ≥ 25 mm
                                                        scan volume
                                                                                                                                                                       5 - 6 GHz: ≥ 22 mm
                                                        Note: δ is the penetration depth of a plane-wave at normal incidence to the tissue medium: see EEE 1528-2013 for details.
                                                        * When zoom scan is required and the reported SAR from the area scan based 1-g SAR estimation procedures of KDB 447498 is ≤
                                                        1.4 W/kg, ≤ 8 mm, ≤ 7 mm and ≤ 5 mm zoom scan resolution may be applied, respectively, for 2 GHz to 3 GHz, 3 GHz to 4 GHz and
                                                        4 GHz to 6 GHz
                                              444/AR




                                                       Table 5:      Parameters for SAR scan procedures.



                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                           Page 10 of 109


                                                       Issue Date: April 15, 2019                      Revision Date: July 09, 2019               Revision No.: 1


                                                       4.3 Additional Information for IEEE 802.11 (WiFi) Transmitters
                                                       For both DSSS and OFDM wireless modes an Initial Test Position must be established for each applicable
                                                       exposure configuration using either:

                                                                Design implementation defined by the manufacturer, or

                                                                Investigative results by the test lab based on:

                                                                      o    Exclusions based on the distance from the antenna to the surface, or

                                                                      o    Highest measured SAR from the area-scan-only measurements on all applicable test
                                                                           positions at the Initial Test Configuration, if found to require SAR tests.

                                                       Then, the initial test position procedure defines the required complete SAR scan measurements on each
                                                       exposure configuration as following:

                                                                When the reported SAR of the initial test position is ≤ 0.4 W/kg, further SAR measurements is not
                                                                 required for the remaining test positions in that configuration as well as 802.11 transmission mode
                                                                 combinations within the frequency or aggregated band.

                                                                When the reported SAR of the initial test position is > 0.4 W/kg, further SAR measurements is
                                                                 required in the initial test position or next closest/smallest test separation distance based on
                                                                 manufacturer justification, on the following highest maximum output power channel, until the
                                                                 reported SAR is ≤ 0.8 W/kg or all required test positions are tested.

                                                                When the reported SAR for all initial and subsequent test positions is > 0.8 W/kg, further SAR
                                                                 measurements is required on these positions on the subsequent next highest measured output
                                                                 power channels, until the reported SAR is ≤ 1.2 W/kg or all required channels have been tested.

                                                       For OFDM transmission configurations in 2.4 GHz and 5 GHz bands, it is important to determine SAR Initial
                                                       Test Configuration for each stand alone and aggregated frequency band according to the maximum output
                                                       power and tune-up tolerance specified for production units. The procedure is as following:

                                                                Highest output power channel is chosen; if there are channels with same maximum output power
                                                                 then the closest to the mid-band frequency is preferred. If there are more than one channel with
                                                                 same maximum output power and same distance to the mid-band frequency, then the channel with
                                                                 the higher frequency is preferred.

                                                                When SAR measurement is required for a subsequent test configuration and the channel bandwidth
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                                 is smaller than that in the initial test configuration, all channels in the subsequent test configuration
                                                                 that overlap with the larger bandwidth channel tested in the initial test configuration should be used
                                                                 to determine the highest maximum output power channel in the subsequent test configuration.
SAR_Report_FCC_ISED_Body_v2.1




                                                       Along with the initial test position reduction guidelines, the following procedures are also applied to SAR
                                                       measurement requirements when multiple OFDM configurations are supported:

                                                                When the reported SAR of the initial test configuration with the highest output power channel is > 0.8
                                                                 W/kg, further SAR measurements is required for next highest output power channel in the initial test
                                                                 configuration, until the reported SAR is ≤ 1.2 W/kg or all required channels have been tested.
                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                             Page 11 of 109


                                                       Issue Date: April 15, 2019                   Revision Date: July 09, 2019             Revision No.: 1


                                                                When the reported SAR of the subsequent test configuration with the highest output power channel
                                                                 is > 1.2 W/kg, further SAR measurements is required for next highest output power channel in this
                                                                 test configuration, until the reported SAR is ≤ 1.2 W/kg or all required channels have been tested.

                                                                When the reported SAR of the subsequent test configuration is > 1.2 W/kg, further SAR
                                                                 measurements for the following subsequent test configurations are required.

                                                       4.4 Measurement Variability
                                                       According KDB 865664 repeated measurements are required only when the measured SAR is ≥ 0.80 W/kg.
                                                       If the measured SAR value of the initial repeated measurement is < 1.45 W/kg with ≤ 20% variation, only one
                                                       repeated measurement is required to reaffirm that the results are not expected to have substantial variations,
                                                       which may introduce significant compliance concerns. A second repeated measurement is required only if
                                                       the measured result for the initial repeated measurement is within 10% of the SAR limit and vary by more
                                                       than 20%, which are often related to device and measurement setup difficulties. The following procedures
                                                       are applied to determine if repeated measurements are required.

                                                           1)   Repeated measurement is not required when the original highest measured SAR is < 0.80 W/kg;
                                                                steps 2) through 4) do not apply.

                                                           2)   When the original highest measured SAR is ≥ 0.80 W/kg, repeat that measurement once.

                                                           3)   Perform a second repeated measurement only if the ratio of largest to smallest SAR for the original
                                                                and first repeated measurements is > 1.20 or when the original or repeated measurement is ≥ 1.45
                                                                W/kg (~ 10% from the 1-g SAR limit).

                                                           4)   Perform a third repeated measurement only if the original, first or second repeated measurement is ≥
                                                                1.5 W/kg and the ratio of largest to smallest SAR for the original, first and second repeated
                                                                measurements is > 1.20.
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                        Page 12 of 109


                                                       Issue Date: April 15, 2019                                  Revision Date: July 09, 2019                       Revision No.: 1


                                                       5 System Verification and Test Conditions

                                                       5.1 Date of Testing
                                                                                                                       Date of Testing
                                                                      Band                         Frequency [MHz]                        Date of System Check        Date of SAR Measurement

                                                            GSM 850 / WCDMA B5                              850                               April 10, 2019                    April 10, 2019
                                                                    GSM 1900                                1900                              April 01, 2019                    April 01, 2019
                                                                   WCDMA B2                                 1900                              April 01, 2019                    April 02, 2019
                                                              IEEE 802.11 b/g/n                             2450                             March 06, 2019                  March 07, 2019

                                                       Table 6:       Date of testing.

                                                       5.2 Environment Conditions
                                                                                                              Environment Conditions
                                                                  Ambient Temperature[°C]                           Liquid Temperature [°C]                             Humidity [%]

                                                                           22.0  2                                            22.0  2                                   40.0  10

                                                          Notes: To comply with the required noise level (less than 12 mW/kg) periodically measurements without a DUT were conducted.

                                                       Table 7:       Environment Conditions.

                                                       5.3 Tissue Simulating Liquid Recipes
                                                                                                              Tissue Simulating Liquid
                                                        Frequency Range               Water       Tween 20            Tween 80               Salt         Preventol       DGME           Triton X/100
                                                              [MHz]                    [%]            [%]                 [%]                [%]               [%]         [%]                   [%]
                                                                                                                          Head Tissue
                                                                   300                50.4           47.3                  -                 2.2               0.1          -                    -
                                                                   450                50.8           47.5                  -                 1.6               0.1          -                    -
                                                              700 - 1000              52.8           46.0                  -                 1.1               0.1          -                    -
                                                             1600 - 1800              55.4           44.1                  -                 0.4               0.1          -                    -
                                                             1850 - 1980              55.2           44.5                  -                 0.2               0.1          -                    -
                                                             2000 - 2700              55.7           45.2                  -                  -                0.1          -                    -
                                                             5000 - 6000              65.5             -                   -                  -                 -         17.25              17.25
                                                                                                                          Body Tissue
                                                                   300                70.3           28.6                  -                 1.0               0.1          -                    -
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                                   450                71.0           28.0                  -                 0.9               0.1          -                    -
                                                              700 - 1000              71.2           28.0                  -                 0.7               0.1          -                    -
                                                             1600 - 1800              71.4           28.0                  -                 0.5               0.1          -                    -
                                                             1850 - 1980              71.5           28.0                  -                 0.4               0.1          -                    -
SAR_Report_FCC_ISED_Body_v2.1




                                                             2000 - 2700              71.6           28.0                  -                 0.3               0.1          -                    -
                                                             5000 - 6000              79.9             -                 20.0                 -                0.1          -                    -
                                                        Notes: Used liquid for measurement is checked above.

                                                       Table 8:       Recipes of the tissue simulating liquid.
                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                                 Page 13 of 109


                                                       Issue Date: April 15, 2019                           Revision Date: July 09, 2019                      Revision No.: 1


                                                       5.4 Tissue Simulating Liquid Parameters
                                                       For the measurement of the following parameters the Speag DAK-3.5 dielectric probe kit is used,
                                                       representing the open-ended coaxial probe measurement procedure.

                                                       Recommended values for the dielectric parameters of the tissue simulating liquids are given in IEEE 1528
                                                       and FCC published RF Exposure KDB Procedures. All tests were carried out using liquids with dielectric
                                                       parameters within +/- 5% of the recommended values. The dielectric properties of the tissue simulating liquid
                                                       have been measured within 24 h before SAR testing. The depth of the tissue simulant was at least 15.0 cm
                                                       for all system check and device tests, measured from the ear reference point in case of the SAM phantom
                                                       and from the inner surface of the flat phantom.
                                                                                                         Tissue Simulating Liquids
                                                                       Ambient / Liquid Temperature(C) : 22.0 ± 2                                            Humidity (%) : 40.0 ± 10
                                                                                                                             Permittivity                            Conductivity
                                                                          Frequency
                                                            Band                               Channel            Measured       Target       Delta      Measured        Target        Delta
                                                                            [MHz]                                    ε'             ε'      +/- 5 [%]      σ [S/m]       σ [S/m]     +/- 5 [%]
                                                                             835.0          System Check            55.1          55.2        -0.3          0.993         0.982         1.1
                                                                             824.2                128               55.1          55.2        -0.2          0.99          0.98          1.0
                                                          GSM 850
                                                                             836.6                190               55.1          55.2        -0.3          0.99          0.98          1.1
                                                                             848.8                251               55.0          55.2        -0.3          1.00          0.99          1.3
                                                                             835.0          System Check            55.1          55.2        -0.3          0.99          0.98          1.1
                                                                             826.4               4132               55.1          55.2        -0.2          0.99          0.98          1.0
                                                         WCDMA 5
                                                                             836.6               4183               55.1          55.2        -0.3          0.99          0.98          1.1
                                                                             846.6               4233               55.0          55.2        -0.3          1.00          0.99          1.2
                                                                            1900.0           System Check           53.5          53.3         0.3          1.59          1.52          4.3
                                                                            1850.2                512               53.6          53.3         0.6          1.54          1.52          1.2
                                                         PCS 1900
                                                                            1880.0                661               53.5          53.3         0.4          1.57          1.52          3.1
                                                                            1909.8                810               53.4          53.3         0.2          1.60          1.52          4.9

                                                                            1900.0           System Check           53.5          53.3         0.3          1.59          1.52          4.3
                                                                            1852.4               9262               53.6          53.3         0.6          1.54          1.52          1.2
                                                         WCDMA 2
                                                                            1880.0               9400               53.5          53.3         0.4          1.57          1.52          3.1
                                                                            1907.6               9538               53.4          53.3         0.2          1.60          1.52          4.9
                                                                             2450            System Check           51.5          52.7        -2.2          2.02          1.95          3.7
                                                                             2412                  1                51.6          52.8        -2.2          1.98          1.91          3.3
                                                           WLAN
                                                                             2437                  6                51.6          52.7        -2.2          2.01          1.94          3.6
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                                             2462                 11                51.5          52.7        -2.3          2.03          1.96          3.6
                                                       Notes: The dielectric properties of the tissue simulating liquid must be measured within 24 h before the SAR testing.

                                                       Table 9:     Parameters of the tissue simulating liquid.
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                          Page 14 of 109


                                                       Issue Date: April 15, 2019                            Revision Date: July 09, 2019                           Revision No.: 1


                                                       5.5 Simplified Performance Checking
                                                       The simplified performance check was realized using the dipole validation kit. The input power of the dipole
                                                       antenna was 250 mW (CW) and it was placed under the flat part of the SAM phantom. The target and
                                                       measured results are listed in the Table 10 and shown in Appendix C - System Verification Plots. The target
                                                       values were adopted from the calibration certificates found also in the appendix.

                                                                                                             System Check Results
                                                                                                                        SAR Values with Body TSL [W/kg]

                                                        Frequency                                             Measured                             Target                    Delta
                                                                            Dipole #SN
                                                          [MHz]                                   with 250 mW           scaled to 1 W     normalized to 1 W              +/- 10 [%]             Date
                                                                                                  1g         10g         1g       10g        1g             10g       1g             10g
                                                            835       D835V2 #470                2.50        1.67       10.00    6.68       9.56            6.24     4.60        7.05        10. Apr 19
                                                           1900       D1900V2 #535               9.46        5.11       37.84    20.44      38.20       20.36        -0.94       0.39        01. Apr 19
                                                           2450       D2450V2 #709               13.80       6.37       55.20    25.48      51.20       24.00        7.81        6.17        06. Mar 19

                                                       Table 10:     Dipole target and measured results.


                                                       6 Measurement Conditions

                                                       6.1 SAR Test Conditions
                                                                                                                    Test Conditions
                                                                                             TX Range                                                              Crest
                                                                   Band                                                          Used Channels                                         Phantom
                                                                                               [MHz]                                                               Factor

                                                            GSM / GPRS850                   824.2 - 848.8                        128 / 190 / 251                   8.3 / 2

                                                           GSM / GPRS1900                  1850.2 - 1909.8                       512 / 661 / 810                   8.3 / 2

                                                                                                                                                                                      SAM
                                                               WCDMA 5                      826.4 - 846.6                       4132 / 4183 / 4233                   1
                                                                                                                                                                                Twin Phantom V4.0
                                                               WCDMA 2                     1852.4 - 1907.6                      9262 / 9400 / 9538                   1

                                                              IEEE 802.11                    2412 - 2462                            1 / 6 / 11                       1

                                                          Notes: To establish a connection at a specific channel and with maximum output power, base station simulator has been used for
                                                                   2G and 3G testing. Engineering test software has been used for WLAN measurements.

                                                       Table 11:     Used channels and crest factors during the test.
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                                    Page 15 of 109


                                                       Issue Date: April 15, 2019                           Revision Date: July 09, 2019                         Revision No.: 1


                                                       6.2 Tune-Up Information

                                                       6.2.1 Maximum Transmitting Output Power Values

                                                                                                Output Power Tune-Up Information for 2G
                                                                                                    Frequency Range                        Averaged Output Power (RMS) [dBm]
                                                                     Band / Mode
                                                                                                         [MHz]                                        Tune-Up Limit

                                                                    2G                850              824.2 - 848.8                                        33.0
                                                                   GSM                1900            1850.2 - 1909.8                                       30.0
                                                                                        number of uplink time slots               1 TX             2 TX              3 TX                 4TX
                                                               2G
                                                           GPRS / EDGE                850              824.2 - 848.8              33.0             31.5              30.0              28.0
                                                             (GMSK)
                                                                                      1900            1850.2 - 1909.8             30.0             27.0              25.0              24.0
                                                                  2G                  850              824.2 - 848.8              28.0             28.0              26.0              25.0
                                                                EDGE
                                                               (8-PSK)                1900            1850.2 - 1909.8             26.0             26.0              24.0              23.0
                                                         Notes:

                                                                                                Output Power Tune-Up Information for 3G
                                                                                                    Frequency Range                        Averaged Output Power (RMS) [dBm]
                                                                     Band / Mode
                                                                                                         [MHz]                                        Tune-Up Limit

                                                               3G                    Band 2           1852.4 - 1907.6                                       23.0
                                                              WCDMA                  Band 5            826.4 - 846.6                                        23.0
                                                         Notes:

                                                                                             Output Power Tune-Up Information for Bluetooth
                                                                                                                                           Averaged Output Power (RMS) [dBm]
                                                                                                    Frequency Range
                                                                     Band / Mode                                             Max. Output Power       Internal Loss of
                                                                                                         [MHz]                                                                Tune-Up Limit
                                                                                                                              u-blox ELLA W1         K+K Trizeps VII

                                                             Bluetooth              v3.0+HS           2402.0 – 2480.0                10.0                   0.5                      9.5
                                                              Classic               v2.1+EDR          2402.0 – 2480.0                10.0                   0.5                      9.5
                                                         Notes:

                                                                                              Output Power Tune-Up Information for WLAN
                                                                                                                                           Averaged Output Power (RMS) [dBm]
                                                                                                    Frequency Range
                                                                     Band / Mode                                             Max. Output Power       Internal Loss of
                                                                                                         [MHz]                                                                Tune-Up Limit
                                                                                                                              u-blox ELLA W1         K+K Trizeps VII
                                                                                                      2412.0 – 2462.0                19.0                   0.5                     18.5
                                                                                    802.11b             CH1 / CH11                   19.0                   0.5                     18.5
                                                                                                      CH 12 / CH 13                    x                     x                        x
                                                                                                      2412.0 – 2462.0                16.0                   0.5                     15.5
                                                           WLAN 2.4 GHz
                                                                                    802.11g             CH1 / CH11                   16.0                   0.5                     15.5
                                                              DTS
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                                                                      CH 12 / CH 13                    x                     x                        x
                                                                                                      2412.0 – 2462.0                16.0                   0.5                     15.5
                                                                                    802.11n             CH1 / CH11                   16.0                   0.5                     15.5
                                                                                                      CH 12 / CH 13                    x                     x                        x
SAR_Report_FCC_ISED_Body_v2.1




                                                         Notes: Product hardware has the capability to operate on channel 12 and 13.
                                                                However, these channels will be disabled via software and will not be accessible to user.

                                                       Table 12:     Maximum transmitting output power declared by the manufacturer.
                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                            Page 16 of 109


                                                       Issue Date: April 15, 2019                          Revision Date: July 09, 2019                    Revision No.: 1


                                                       6.3 Measured Output Power

                                                       6.3.1 Output Power Values for 2G
                                                       This device supports GPRS/EDGE multislot class 12. According the following tables, GPRS 850/1900 with
                                                       4 TX represent the worst case, therefore measurements with four active time slots are conducted for
                                                       GPRS 850/1900.

                                                                                          Max. Burst-Averaged Output Power (RMS) [dBm]
                                                                   Frequency                                 GPRS (GMSK / CS1)                           EDGE (8PSK / MCS5)
                                                         Band                       CH    GSM
                                                                     [MHz]                          1 TX        2 TX       3 TX       4 TX       1 TX       2 TX          3 TX      4 TX
                                                                      824.2         128    32.3     32.3        29.0       27.6       26.4       26.7       26.4          24.2      23.0
                                                          850         836.6         190    32.0     32.0        29.0       27.3       26.5       26.9       26.5          24.7      23.3
                                                                      848.8         251    31.7     31.7        28.9       27.2       26.4       26.7       26.3          24.3      23.1
                                                                     1850.2         512    29.4     29.4        26.3       24.7       23.8       26.0       26.0          24.4      23.2
                                                         1900        1880.0         661    29.3     29.3        26.0       24.4       23.4       25.7       25.9          23.8      22.6
                                                                     1909.8         810    29.3     29.3        25.8       23.9       23.0       25.6       25.2          23.5      22.2

                                                                                          Max. Frame-Averaged Output Power (RMS) [dBm]
                                                                   Frequency                                 GPRS (GMSK / CS1)                           EDGE (8PSK / MCS5)
                                                         Band                       CH    GSM
                                                                     [MHz]                          1 TX        2 TX       3 TX       4 TX       1 TX       2 TX          3 TX      4 TX
                                                                      824.2         128    23.3     23.3        23.0       23.3       23.4       17.7       20.4          19.9      20.0
                                                          850         836.6         190    23.0     23.0        23.0       23.0       23.5       17.9       20.5          20.4      20.3
                                                                      848.8         251    22.7     22.7        22.9       22.9       23.4       17.7       20.3          20.0      20.1
                                                                     1850.2         512    20.4     20.4        20.3       20.4       20.8       17.0       20.0          20.1      20.2
                                                         1900        1880.0         661    20.3     20.3        20.0       20.1       20.4       16.7       19.9          19.5      19.6
                                                                     1909.8         810    20.3     20.3        19.8       19.6       20.0       16.6       19.2          19.2      19.2
                                                       Notes: The frame-averaged output power is linearly scaled burst averaged power over 8 time slots as shown below:
                                                              Frame-averaged power = maximum burst avg. power for 1 TX slot – 9.0 dB
                                                                Frame-averaged power = maximum burst avg. power for 2 TX slots – 6.0 dB
                                                                Frame-averaged power = maximum burst avg. power for 3 TX slots – 4.26 dB
                                                                Frame-averaged power = maximum burst avg. power for 4 TX slots – 3.0 dB

                                                       Table 13:     Measured output power for 2G bands.
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                      Page 17 of 109


                                                       Issue Date: April 15, 2019                          Revision Date: July 09, 2019                     Revision No.: 1


                                                       6.3.2 Output Power Values for 3G
                                                       For measurements in WCDMA, without HSDPA or HSUPA, the default test configuration is to establish a
                                                       radio link between the DUT and a communication test set using a 12.2 kbps RMC configured Test Loop
                                                       Mode 1 and TPC bits configured to all “1”. The SAR will be tested for all bands using a Rel99 call configured
                                                       to transmit at maximum output power per 3GPP 34.121 [3GPP 34.121]. The Rel99 parameters are
                                                       summarized in Table 14.

                                                       WCDMA was tested in RMC mode without HSPA. According KDB 941225 D01 HSPA SAR is not required
                                                       when the averaged output power of the HSPA subtests are not higher than 0.25 dB then measured in RMC
                                                       mode and the assessed SAR value in this mode is not higher than 1.2 W/kg.

                                                                                             Max. Averaged Output Power (RMS) [dBm]
                                                                     Freq.               WCDMA                     HSDPA                                      HSUPA
                                                          Band                      CH
                                                                     [MHz]                RMC    Subt. 1    Subt. 2   Subt. 3   Subt. 4   Subt. 1   Subt. 2   Subt. 3   Subt. 4    Subt. 5
                                                                      826.4     4132      21.0    20.9        21.0      20.7      20.6     21.0      20.9       20.6      20.9       20.8
                                                          850
                                                                      836.6     4183      21.3    21.2        21.2      20.7      20.8     20.9      21.0       20.5      20.9       20.7
                                                        (FDD 5)
                                                                      846.6     4233      21.0    21.1        21.1      20.7      20.6     20.6      21.0       20.6      21.0       20.8
                                                                     1852.4     9626      23.0    23.0        23.0      22.6      22.7     23.0      23.0       22.6      22.7       22.5
                                                          1900
                                                                     1880.0     9400      23.0    23.0        23.0      22.8      22.8     23.0      23.0       22.5      23.0       22.8
                                                        (FDD 2)
                                                                     1907.6     9538      22.5    22.5        22.5      22.1      22.0     22.5      22.2       22.1      22.3       22.1

                                                                     βc                           2/15       12/15     15/15     15/15    11/15      6/15      15/15      2/15      15/15

                                                                     βd                           15/15      15/15      8/15      4/15    15/15     15/15       9/15     15/15      15/15

                                                            ΔACK. ΔNACK. ΔCQI                       8          8           8       8        8         8          8            8       8

                                                       Table 14:     Measured output power for 3G bands.
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                       Page 18 of 109


                                                       Issue Date: April 15, 2019                         Revision Date: July 09, 2019                             Revision No.: 1


                                                       6.3.3 Output Power Values for WLAN 2.4 GHz
                                                                                            Max. Averaged Output Power (RMS) [dBm]
                                                                       Frequency                                                          Data Rate [Mbit/s]
                                                          Mode                            CH
                                                                         [MHz]                                 1                     2                        5.5                      11
                                                                          2412             1                  15.70                   -                        -                        -
                                                             b            2437             6                  17.50                 17.50                    17.50                    17.50
                                                                          2462            11                  17.50                   -                        -                        -

                                                                       Frequency                                                          Data Rate [Mbit/s]
                                                          Mode                            CH
                                                                         [MHz]                          6.0            9      12            18        24             36         48            54
                                                                          2412             1           13.6            -       -             -         -              -          -             -
                                                             g            2437             6           13.7            -       -             -         -              -          -             -
                                                                          2462            11           13.7            -       -             -         -              -          -             -

                                                                       Frequency                                                            MCS Index No.
                                                          Mode                            CH
                                                                         [MHz]                        MCS0         MCS1      MCS2         MCS3       MCS4           MCS5      MCS6          MCS7
                                                                          2412             1           13.6            -       -             -         -              -          -             -
                                                            n
                                                                          2437             6           13.7            -       -             -         -              -          -             -
                                                           HT20
                                                                          2462            11           13.9           13.9   13.9           13.9      13.9           13.9      13.9           13.9
                                                                          2422             3           13.9            -       -             -         -              -          -             -
                                                            n
                                                                          2437             6           13.9           13.8   13.8           13.8      13.8           13.8      13.8           13.8
                                                           HT40
                                                                          2452             9           13.8            -       -             -         -              -          -             -
                                                        Notes: CH12/13 are disabled by manufacturer and will not be accessible to user.

                                                       Table 15:     Conducted output power for WLAN 2.4 GHz.
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                              Page 19 of 109


                                                       Issue Date: April 15, 2019                           Revision Date: July 09, 2019                   Revision No.: 1


                                                       6.4 Standalone SAR Test Exclusion
                                                       SAR test exclusion is determined for the EUT according to KDB 447498 D01 with 1g SAR exclusion
                                                       thresholds for 100 MHz to 6GHz at test separation distances ≤ 50 mm determined by:
                                                            [(max power of channel. incl. tune-up tolerance. mW) / (min test separation distance. mm)] * [√f(GHz)]
                                                                                       ≤ 3.0 for 1g SAR and ≤ 7.5 for 10g extremity SAR, where

                                                       • f(GHz) is the RF channel transmit frequency in GHz
                                                       • Power and distance are rounded to the nearest mW and mm before calculation
                                                       • The result is rounded to one decimal place for comparison

                                                       When the minimum test separation distance is < 5 mm, a distance of 5 mm is applied to determine SAR test
                                                       exclusion.
                                                                                    Transmission Scenario for Test Exclusion Considerations
                                                                      Exposure Configuration        Body            Body           Body           Body           Body            Body
                                                                             Antenna             WLAN 2.4         Bluetooth      GPRS 850     GPRS 1900       WCDMA 2         WCDMA 5
                                                        Exposure    Frequency [GHz]                 2.412           2.402          0.824          1.850          1.852           0.826
                                                        Position
                                                                    Frame Avg. Power [dBm]          18.5             9.5            25.0          21.0           23.0            23.0
                                                                    Frame Avg. Power [mW]           70.8             8.9           316.2          125.9          199.5           199.5
                                                                    Antenna to user [mm]            58.0             58.0           58.0          58.0           58.0            58.0
                                                                    SAR exclusion threshold         176.6           176.8          209.2          190.3          190.2           209.1
                                                           Back
                                                                    SAR testing required?            no               no            yes            no             yes             no
                                                                    Estimated SAR [W/kg]          measured          0.400        measured      measured        measured        measured
                                                                    Antenna to user [mm]            187.0           187.0          187.0          187.0          187.0           187.0
                                                           Bottom   SAR exclusion threshold        1466.6           1466.8         917.8         1480.3         1480.2           919.5
                                                                    SAR testing required?            no               no             no            no             no              no
                                                       Notes: According to the customer’s requirements SAR measurements have been performed for all supported bands on the back side
                                                              of DUT (body supported / lap-held conditions). BT test has been excluded due to the output power below the given threshold.

                                                       Table 16:    SAR test exclusion for body-supported exposure configuration against different device edges according to KDB 447498.


                                                       When the standalone SAR test exclusion applies to an antenna that transmits simultaneously with other
                                                       antennas the standalone SAR must be estimated according to KDB 447498 in order to determine
                                                       simultaneous transmission SAR test exclusion:
                                                        (max. power of channel including tune-up tolerance. mW)/(min. test separation distance. mm)]·[√f(GHz)/x]
                                                         W/kg for test separation distances ≤ 50 mm;
                                                            where x = 7.5 for 1-g SAR and x = 18.75 for 10-g SAR
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                       When the minimum test separation distance is < 5 mm, a distance of 5 mm is applied to determine SAR test
                                                       exclusion.
                                                             0.4 W/kg for 1g SAR and 1.0 W/kg for 10g SAR. when the test separation distance is > 50 mm
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                      Page 20 of 109


                                                       Issue Date: April 15, 2019                             Revision Date: July 09, 2019                     Revision No.: 1


                                                       7 SAR Test Results

                                                       The tables below contain the measured SAR values averaged over a mass of 1g. SAR assessment was
                                                       conducted in the worst case configuration with output power values according to Table 13 – Table 15.
                                                       According to KDB 447498 D01, the reported SAR is the measured SAR value adjusted for maximum tune-
                                                       up tolerance limit shown in Table 12.

                                                       Reported SAR is calculated by the following formulas:
                                                                  Scaling factor tune up limit = tune-up limit power (mW) / RF power (mW)
                                                                  Scaling factor max. duty cycle = max. possible duty cycle / used duty cycle for SAR measurement
                                                                  Reported SAR = measured SAR * scaling factor tune up limit * scaling factor max. duty cycle

                                                       The plots with the highest measured SAR values are shown in Appendix B - SAR Distribution Plots.

                                                                                                   SAR Measurement Results for WWAN
                                                                                                                               Measured               Tune-Up    Reported        SAR1g
                                                                                 Freq.              Test        Gap     Pic.                 Drift                                        Plot
                                                            Band / Mode                     CH                                  SAR1g                 Scaling     SAR1g           Limit
                                                                                 [MHz]             Position    [mm]     No.                  [dB]                                         No.
                                                                                                                                [W/kg]                 Factor     [W/kg]         [W/kg]
                                                                       Voice
                                                                       1TX
                                                                                    836.6   190        back      0       3        0.017      -0.125    1.259        0.021

                                                          GSM                       836.6   190        back      0       3        0.017      -0.181    1.445        0.025         1.6
                                                           850         GPRS
                                                                        4TX
                                                                                    824.2   128        back      0       3        0.016      0.177     1.413        0.023

                                                                                    848.8   251        back      0       3        0.018      0.060     1.445        0.026                  1
                                                                       Voice
                                                                       1TX
                                                                                 1880.0     661        back      0       3        0.009      0.142     1.175        0.011

                                                          GSM                    1880.0     661        back      0       3        0.011      0.048     1.047        0.012         1.6
                                                           1900        GPRS
                                                                        4TX
                                                                                 1850.2     512        back      0       3        0.013      0.069     1.148        0.015                  2

                                                                                 1909.8     810        back      0       3        0.006      -0.073    1.259        0.008

                                                                                    836.6   4183       back      0       3        0.017      -0.106    1.585        0.027                  3
                                                         WCDMA
                                                                       RMC          826.4   4132       back      0       3        0.014      -0.014    1.479        0.021         1.6
                                                            B5
                                                                                    846.8   4233       back      0       3        0.012      0.081     1.585        0.019

                                                                                 1880.0     9400       back      0       3        0.019      0.134     1.000        0.019
                                                         WCDMA
                                                                       RMC       1852.0     9262       back      0       3        0.017      0.134     1.000        0.017         1.6
                                                            B2
                                                                                 1908.0     9538       back      0       3        0.022      0.115     1.122        0.025                  4

                                                        Notes:

                                                       Table 17:      SAR results for 2G and 3G bands.
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                                                                   SAR Measurement Results for WLAN
                                                                                                                               Measured               Tune-Up    Reported        SAR1g
                                                                                 Freq.              Test        Gap     Pic.                 Drift                                        Plot
                                                            Band / Mode                     CH                                  SAR1g                 Scaling     SAR1g           Limit
                                                                                 [MHz]             Position    [mm]     No.                  [dB]                                         No.
                                                                                                                                [W/kg]                 Factor     [W/kg]         [W/kg]
SAR_Report_FCC_ISED_Body_v2.1




                                                                                    2437     6         back      0       3        0.005      -0.110    1.259        0.006                  5
                                                          IEEE
                                                                      1 Mbps        2412     1         back      0       3        0.002      0.092     1.905        0.004         1.6
                                                         802.11 b
                                                                                    2462    11         back      0       3        0.005      0.119     1.259        0.006
                                                        Notes:

                                                       Table 18:      SAR results for IEEE 802.11 b.
                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                          Page 21 of 109


                                                       Issue Date: April 15, 2019                                Revision Date: July 09, 2019                 Revision No.: 1


                                                       General Note:

                                                       According to the declaration of the manufacturer the Tempus Pro device is intended to be used in body-
                                                       supported conditions (lap-held) with the back or bottom side towards the human body. Therefore, only two
                                                       sides of DUT have been considered for testing. The bottom side of DUT has been excluded from testing
                                                       according to Table 16.

                                                       Measured maximum SAR is < 0.8 W/kg, thus measurement variability assessment according to
                                                       KDB 865664 is not applicable.



                                                       8 Simultaneous Transmission Consideration

                                                                                            Simultaneous Transmission Capabilities of DUT
                                                                           WWAN                                            WLAN                                   Bluetooth
                                                                              X                                               X                                        X
                                                                              V                                               X                                        V
                                                                              X                                               V                                        V

                                                        Notes: Only WWAN + BT or WLAN + BT antennas are able to transmit simultaneously.
                                                       Table 19:       Simultaneous transmission capabilities.


                                                       For the following simultaneous transmission analysis the worst case SAR results shown in Table 17 –
                                                       Table 18 have been used.

                                                       According to KDB 447498, the following table gives an overview about the ∑SAR for simultaneous
                                                       transmitting modes. When ∑SAR > 1.6 W/kg. a SAR test exclusion is determined by the SAR to peak
                                                       location separation ratio.

                                                                                                    Simultaneous Transmission Analysis
                                                                                                                                                                         Σ
                                                         Exposure Position                 WWAN                           WLAN                    Bluetooth                             SPLSR
                                                                                                                                                                       SAR1g
                                                              of DUT                   max SAR [W/kg]                 max SAR [W/kg]            max SAR [W/kg]                         Analysis
                                                                                                                                                                       [W/kg]
                                                                   Back                     0.027                             X                     0.400*                 0.427         NO
                                                                                              X                             0.006                   0.400*                 0.406         NO

                                                        Notes: Analysis taken into consideration of the simultaneous transmission capabilities is shown in Table 19.
                                                                   *Estimated SAR values according to Table 17.
                                                       Table 20:       SAR for simultaneous transmission scenario.
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                            Page 22 of 109


                                                       Issue Date: April 15, 2019                        Revision Date: July 09, 2019                Revision No.: 1


                                                       9 Administrative Measurement Data
                                                       9.1 Calibration of Test Equipment

                                                                                                       Test Equipment Overview
                                                                                                                                          Serial        Last              Next
                                                                     Test Equipment                   Manufacturer         Model
                                                                                                                                         Number      Calibration       Calibration
                                                        DASY System Components
                                                            Software Versions DASY4                      SPEAG              V4.7           N/A           N/A               N/A
                                                            Software Versions SEMCAD                     SPEAG              V1.8           N/A           N/A               N/A
                                                            Dosimetric E-Field Probe                     SPEAG            ET3DV6R          1579        02/2018           02/2019
                                                            Dosimetric E-Field Probe                     SPEAG            ET3DV6R          1669        02/2019           02/2020
                                                            Dosimetric E-Field Probe                     SPEAG            EX3DV4           3536        09/2018           09/2020
                                                            Dosimetric E-Field Probe                     SPEAG            EX3DV4           3860        09/2017           09/2019
                                                            Data Acquisition Electronics                 SPEAG             DAE 3           335         02/2018           02/2019
                                                            Data Acquisition Electronics                 SPEAG             DAE 4           631         09/2018           09/2019
                                                            Phantom                                      SPEAG              SAM            1059          N/A               N/A
                                                            Phantom                                      SPEAG              SAM            1176          N/A               N/A
                                                            Phantom                                      SPEAG              SAM            1340          N/A               N/A
                                                            Phantom                                      SPEAG              SAM            1341          N/A               N/A
                                                            Phantom                                      SPEAG              ELI4           1004          N/A               N/A
                                                        Dipoles
                                                            System Validation Dipole                     SPEAG             D450V2          1014        03/2018           03/2021
                                                            System Validation Dipole                     SPEAG             D835V2          470         03/2018           03/2021
                                                            System Validation Dipole                     SPEAG            D1640V2          311         09/2018           09/2021
                                                            System Validation Dipole                     SPEAG            D1750V2          1005        03/2018           03/2021
                                                            System Validation Dipole                     SPEAG            D1900V2          535         03/2018           03/2021
                                                            System Validation Dipole                     SPEAG            D2450V2          709         11/2018           11/2021
                                                            System Validation Dipole                     SPEAG            D2600V2          1019        11/2018           11/2021
                                                            System Validation Dipole                     SPEAG            D5GHzV2          1028        05/2017           05/2020
                                                        Material Measurement
                                                            Network Analyzer                             Agilent           E5071C       MY46103220     08/2017           08/2019
                                                            Dielectric Probe Kit                         SPEAG            DAK-3.5          1234        02/2018           02/2020
                                                           Thermometer                                LKMelectronic       DTM3000          3511        02/2018           02/2020
                                                        Power Meters and Sensors
                                                            Power Meter                                  Anritsu          ML2487A       6K00002319     06/2018           06/2020
                                                            Power Sensor                                 Anritsu          MA2472A         990365       06/2018           06/2020
                                                            Power Meter                                  Anritsu          ML2488A       6K00002078     06/2018           06/2020
                                                            Power Sensor                                 Anritsu          MA2472A         002122       06/2018           06/2020
                                                            Spectrum Analyzer                        Rohde & Schwarz        FSP7          100433       04/2018           04/2020
                                                        RF Sources
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                            Network Analyzer                             Agilent           E5071C       MY46103220     08/2017           08/2019
                                                           RF Generator                              Rohde & Schwarz       SM300          100142         N/A               N/A
                                                        Amplifiers
                                                            Amplifier 10 MHz – 4200 MHz                Mini Circuits    ZHL-42-42W       D080504-1       N/A               N/A
                                                            Amplifier 2 GHz – 6 GHz                   Ciao Wireless       CA26-451         37452         N/A               N/A
SAR_Report_FCC_ISED_Body_v2.1




                                                        Radio Tester
                                                            Radio Communication Tester                   Anritsu          MT8815B       6200576536     04/2018           04/2020
                                                            Radio Communication Tester                   Anritsu          MT8820C       6200918336     04/2018           04/2020
                                                        Notes: Used test equipment for measurement is checked above.

                                                       Table 21:    Calibration of test equipment.
                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                Page 23 of 109


                                                       Issue Date: April 15, 2019                          Revision Date: July 09, 2019                       Revision No.: 1


                                                       9.2 Uncertainty Assessment

                                                                           Uncertainty Budget for SAR Measurements according to IEEE 1528-2013
                                                                                                                 (300 MHz - 6 GHz)

                                                                                                      Uncertainty                                                      Standard             vi²
                                                                                                                        Probability
                                                                        Error Sources                   Value                             Divisor      ci      ci     Uncertainty           or
                                                                                                                        Distribution
                                                                                                        [± %]                                                            [± %]             veff
                                                       Measurement System                                                                             1g      10g      1g           10g
                                                       Probe calibration                                   6.7             Normal            1         1       1       6.7          6.7     
                                                       Axial isotropy                                      0.3          Rectangular         3              0.1          0.1     
                                                       Hemispherical isotropy                              1.3          Rectangular         3              0.5          0.5     
                                                       Boundary effects                                    1.0          Rectangular         3        1       1       0.6          0.6     
                                                       Linearity                                           0.3          Rectangular         3        1       1       0.2          0.2     
                                                       System detection limit                              1.0          Rectangular         3        1       1       0.6          0.6     
                                                       Modulation response                                 4.0          Rectangular         3        1       1       2.3          2.3     
                                                       Readout electronics                                 0.3             Normal            1         1       1       0.3          0.3     
                                                       Response time                                       0.8          Rectangular         3        1       1       0.5          0.5     
                                                       Integration time                                    1.4          Rectangular         3        1       1       0.8          0.8     
                                                       RF ambient conditions - noise                       3.0          Rectangular         3        1       1       1.7          1.7     
                                                       RF ambient conditions - refl.                       3.0          Rectangular         3        1       1       1.7          1.7     
                                                       Probe positioner mech. tol.                         0.4          Rectangular         3        1       1       0.2          0.2     
                                                       Probe positioning                                   2.9          Rectangular         3        1       1       1.7          1.7     
                                                       Algorithms for max SAR eval.                        4.0          Rectangular         3        1       1       2.3          2.3     
                                                       Test Sample Related

                                                       Test sample positioning                             2.9             Normal            1         1       1       2.9          2.9    145
                                                       Device holder uncertainty                           3.6             Normal            1         1       1       3.6          3.6     5
                                                       SAR drift measurement (< 0.2 dB)                    4.7          Rectangular         3        1       1       2.7          2.7     
                                                       SAR scaling                                         2.0          Rectangular         3        1       1       1.2          1.2     
                                                       Phantom and Set-up
                                                       Phantom uncertainty                                 4.0          Rectangular         3        1       1       2.3          2.3     
                                                       SAR correction for perm./cond.                      1.9             Normal            1         1      0.84     1.9          1.6     
                                                       Liquid conductivity (meas.)                         5.0             Normal            1       0.78     0.71     3.9          3.6     
                                                       Liquid permittivity (meas.)                         5.0             Normal            1       0.23     0.26     1.2          1.3     
                                                       Liquid conductivity temp. unc.                      2.9          Rectangular         3      0.78     0.71     1.3          1.2     
                                                       Liquid permittivity temp. unc.                      1.8          Rectangular         3      0.23     0.26     0.2          0.3     
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                       Combined Standard Uncertainty                                                                                  11.1          11.0

                                                       Coverage Factor for 95%                                                                                               kp=2

                                                       Expanded Standard Uncertainty                                                                                  22.2      21.9
SAR_Report_FCC_ISED_Body_v2.1




                                                       Notes: Worst case probe calibration uncertainty has been applied for all available probes and frequencies.

                                                       Table 22:     Uncertainty budget for SAR measurements.
                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                         Page 24 of 109


                                                       Issue Date: April 15, 2019                          Revision Date: July 09, 2019                      Revision No.: 1


                                                                          Uncertainty Budget for SAR System Validation according to IEEE 1528-2013
                                                                                                                 (300 MHz - 6 GHz)

                                                                                                       Uncertainty                                                    Standard             vi²
                                                                                                                        Probability
                                                                        Error Sources                    Value                            Divisor      ci      ci    Uncertainty           or
                                                                                                                        Distribution
                                                                                                         [± %]                                                          [± %]             veff

                                                       Measurement System                                                                             1g      10g     1g           10g
                                                       Probe calibration                                   6.7             Normal            1         1       1     6.7           6.7     
                                                       Axial isotropy                                      0.3          Rectangular         3        1       1     0.1           0.1     
                                                       Hemispherical isotropy                              1.3          Rectangular         3        0       0     0.0           0.0     
                                                       Boundary effects                                    1.0          Rectangular         3        1       1     0.6           0.6     
                                                       Linearity                                           0.3          Rectangular         3        1       1     0.2           0.2     
                                                       System detection limit                              1.0          Rectangular         3        1       1     0.6           0.6     
                                                       Modulation response                                 0.0          Rectangular         3        0       0     0.0           0.0     
                                                       Readout electronics                                 0.3             Normal            1         1       1     0.3           0.3     
                                                       Response time                                       0.0          Rectangular         3        0       0     0.0           0.0     
                                                       Integration time                                    0.0          Rectangular         3        0       0     0.0           0.0     
                                                       RF ambient conditions - noise                       1.0          Rectangular         3        1       1     0.6           0.6     
                                                       RF ambient conditions - refl.                       1.0          Rectangular         3        1       1     0.6           0.6     
                                                       Probe positioner mech. tol.                         0.4          Rectangular         3        1       1     0.2           0.2     
                                                       Probe positioning                                   2.9          Rectangular         3        1       1     1.7           1.7     
                                                       Algorithms for max SAR eval.                        4.0          Rectangular         3        1       1     2.3           2.3     
                                                       Validation Dipole
                                                       Dev. of exp. dipole from num.                       5.0             Normal            1         1       1     5.0           5.0     
                                                       Input power and SAR drift (< 0.2 dB)                4.7          Rectangular         3        1       1     2.7           2.7     
                                                       Dipole axis to liquid distance (< 2deg)             2.0          Rectangular         3        1       1     1.2           1.2     
                                                       Phantom and Set-up
                                                       Phantom uncertainty                                 4.0          Rectangular         3        1       1     2.3           2.3     
                                                       SAR correction for perm./cond.                      1.9             Normal            1         1     0.84    1.9           1.6     
                                                       Liquid conductivity (meas.)                         5.0             Normal            1       0.78    0.71    3.9           3.6     
                                                       Liquid permittivity (meas.)                         5.0             Normal            1       0.23    0.26    1.2           1.3     
                                                       Liquid conductivity temp. unc.                      2.9          Rectangular         3      0.78    0.71    1.3           1.2     
                                                       Liquid permittivity temp. unc.                      1.8          Rectangular         3      0.23    0.26    0.2           0.3     

                                                       Combined Standard Uncertainty                                                                                 10.7          10.6

                                                       Coverage Factor for 95%                                                                                              kp=2
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06




                                                       Expanded Standard Uncertainty                                                                                 21.5          21.2

                                                       Notes: Worst case probe calibration uncertainty has been applied for all available probes and frequencies.

                                                       Table 23:     Uncertainty budget for SAR system validation.
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                        Page 25 of 109


                                                       Issue Date: April 15, 2019                            Revision Date: July 09, 2019                    Revision No.: 1


                                                       10 Report History

                                                                                                                Revision History
                                                        Revision                       Description of Revision                                  Date         Revised Page       Revised By

                                                            /       Initial Release                                                         April 15, 2019         -                 -

                                                                    IC No. corrected (typo), test date included on each SAR plot,
                                                            1                                                                               July 09, 2019     1, 14, 31-38          AR
                                                                    description of tissue simulating liquid depth included



                                                                                                         END OF THE SAR REPORT


                                                       Please refer to separated appendix file for the following data:

                                                                Appendix A - Pictures
                                                                Appendix B - SAR Distribution Plots
                                                                Appendix C - System Verification Plots
                                                                Appendix D – Certificates of Conformity
                                                                Appendix E – Calibration Certificates for DAEs
                                                                Appendix F – Calibration Certificates for E-Field Probes
                                                                Appendix G – Calibration Certificates for Dipoles
SAR_Report_EN50566_EN62209-2_1.0.docx/09.06
SAR_Report_FCC_ISED_Body_v2.1


                                              444/AR




                                                       SAR_Report_FCC_ISED_60320_6190299_TempusPro_rev1                                                                        Page 26 of 109



Document Created: 2019-07-10 10:27:18
Document Modified: 2019-07-10 10:27:18

© 2024 FCC.report