Sar report II

FCC ID: P4MAGT100D

Test Report

Download: PDF
FCCID_698537

  ET3DVS SN:1604                                                            May 2, 2006



                        Deviation from Isotropy in HSL
                                 Error (6, 8),£=900 MHz




                                                                Error 91


                   wiss as m es om m om—o0 ow m mo on
                   messem momom sowomw momom mersic



                    Uncartainty o Spheriea Isctropy Assossmontx 26% (rez)




Centcai n 131008woyos                 raescto


          NCL CALIBRATION LABORATORIES

                     Calibration File No: DC—0267
                Project Number: BACL—CAL2450—3985


CERTIFICATE OF CALIBRATION

     Is ourthed thatthe eqvipment idaifed below has beon caltrate in the
  wct catimration LABORATORIES by quaied personne falowng recognized
         proceturesand using uanster tandantstracentle t NRCAIET

                           BACL vaidaton Dipsle
                      Manstactrer: APREL Laboratores
                          Partnumber 0—100—5—1
                            Frequeney: 1000 it
                             Stal No:Bct—0r9
                   Customer Bay Area Complance Laborttoy

                          Caltrated:4" March 2005
                         Rteased on. 4" March 2008




  Reteased Oy



                   seecrmmnmay             Ovennotrentttas
                  ferows outime             Tee en es
                  Emwon rae e               in ie ons


      NCL Calibration Laboratories
      Divison of APREL Laborators:


      Conditions

      Dipole BCL—049 received ‘rom customer, with the attibute of having a repair to one
      radial arm, in good condition for re—calibration. SMA connector cleaned prior to
      calibration

      Ambient Temperature of the Laboratory:                22°C +0.5°C
      Temperature of the Tissus:                            21 °C +0.5°C

      We the undersigned attest that to the best of our knowledge the calibration of
      this device has been accurately conducted and that all information contained
      within this report has been reviewed for accuracy.




      Stuart Nicol
      Director Product Development




                y
       s        Lex.
       D. Brooks
       Member of Enginering Staff
       (Calibration Engineor)


       This page has been reviewed (or content and attested o bysignature wthin this document




nopmemsmerewe                                reg oo ave                                 w er mmnopos


NCL     Calibration Laboratories
Divisonof APREL Laboratoncs

Calibration Results Summary

The following results relate the Calbrated Dipole and should be used as a quick
reference for the user.

Mechanical Dimensions

Length:                72.0 mm
Height:                 41.7 mm
Electrical Specification
swa:                    109 U
Return Loss:           28. 0B
impedance:             so90
System Validation Rosults
[Frequency 1 Gram 10 Gram Poak
[ 1800 mHz 3881         20 Tise

                                  Area Scan
                                      ¥ Anis (mnd
                                 o    i8

 agoa
 az2
 asie
                                                                              twsox




 aois
          X Axt m




 2412
 180
 1205
 nsos
 vom



 "This page has been reviewed for contant and attested to by signature wihin ts document


NCL Calibration Laboratories
Divisin of APREL Tabormtons,



Introduction

This Calibration Report has been produced in line with the SSI Dipole Calibration
Procedure S8H—TP—018—ALSAS. The results contained within this report are for
Validation Dipole BCL—049. The calibration routine consisted of a three—step process
Step 1 was a mechanical verification of the dipole to ensure that it meets the
mechanical specifications. Step 2 was an Electrical Calibration for the Validation
Dipole, where the SWR, Impedance, and the Return loss were assessed. Step 3
involved a System Validation using the ALGAG—10U, along with APREL E—020 130
MHz to 26 GHz E—Field Probe Serial Number 212.

References
SS—TP—018—ALSAS Dipole Calibration Procedure
SST—TP—016 Tissue Calibration Procedure
IEEE 1528 *Recommended Practice for Determining the Peak Spatial—Average
Specific Absorption Rate (SAR) in the Human Body Due to Wireiess
Communications Devices: Experimental Techniques®

Conditions
Dipole BCL—048 received from customerin good condiion, with the attibute of a
repair of one radial arm.
Ambient Temperature of the Laboratory:                  22°C +/—0.5°C
Temporature of the Tissue:                              20 °C +/0.5°C




"This page has been reviewad for content and altested to by signature within this document


NCL Calibration Laboratories
Divison of APREL Labortons.


Dipole Calibration Results

Mechanical Verification

    APREL              APREL             Measured           Measured
    Length              Hoight             Length             Hoight
   720 mm           _ 41.7 mm              7i6mm        |    417mm         _|

Tissue Validation


 Head Tissue 1800 MHz            Measured

Dleleclnc
    con
Gonductivity, o [Slm]                140




 This page has been reviewed for content and altestad to by signature within is document


Electrical Calibration
      Test_                  Result
    s11 RC                 ~26.008
   _ swr                      1090
      edance                  soon
The Following Graphs are the results as displayed on the Vector Network Analyzer
S11 Paramoter Return Loss


  se nevense reruecrIon                                                        im a o s
                                                                               REFERENCE PLale
  Los nowtruoe               »ner« —20.000 09             4.000 anrory            0.0000 i
                                                                               nencen 1
                                                                                 1.799500 Oz
   mget         4      f      1                            :                      —26.091 0

                                                                T              wrker To m
                                                           &"                  mnker To Nt




        .                     -                                                »nanken reroour
  1.senz5e                          ie                              2.iere00    FuKCTIONS




                                                                                           s
This page has been reviewed for content and attested to bysignature wiin ths document


NCL Catibration Laboratories
Divisn of APREL Laboratones


swR



  see mevense REFLECTION                                                        ou a — see
                                                                                REFERENCE PLaW
  sur                        »REF+ 1.200. 0             500.000 aU/OIV             0.0000 mn
                                                                                nnuen 1
                                                                                  1.739500 e
                                                                                     12095 U

                                                                                mnkem To hok
                                                                                nemcer To n




                                                                               »nenken Resoour
  1.sanasa                          oie                         2.ror000        runcrions




"Tis page has been reviewed for content and altested o by sgnature within ths document


NCL Calibration Laboratories
Divison of APREL Labortores

Smith Chart Dipole Impedance




  see Reverse ReFLECTION               i
  mnecvance                                                                        BA »Lane
                                                                                   reserence
                                                                                      00000 an
                                                                                   neaken 1
                                                                                     1.299500 Oe
                                                                                       se.953 2
                                                                                        alaey jo
                                                                                   meaken To nek
                                                                                   manker To nIN




                                    3                                             »nanken rervout
                           1.seaese — 2.rete0e Oe                                  famchints




"This page has been reviewed for content and attested to by signature wihin this document


NCL Calibration Laboratories
Divison of APRELLabortoris



System Validation Results Using the Electrically Calibrated Dipole

[Head Tissue               1 Gram Gram             FPeak|
|_Frequency                                       AbovsFees Pint
(~      1800 viHz | Ss1 2001                     _|__7i5e


                                      Area Scan
                                            Y Asis (mn)
                                            10


     as2

     az2

     asta

                                                                                  (w) soex
             X Axis (mm)




     a05

     aa2

     1m9

     128

     oom

     oo                            4o       4o        6
                                             ¥ Aus u)




     "This page has been reviewed for content and attested to bysignature wihin ths document


NCL Calibration Laboratories
Divisionof APREL Laboratorcs


Test Equipment
The test equipment used during Probe Calibration, manufacturer, model number
and, current calibration status are listed and located on the main APREL server
R:NCL\Calibration Equipment\instrument List May 2004




This page has beon reviewed for cantent and altested to bysignature wihin this document


                                            si0 rateg                    sometzi—zncosa on smineo
                         ‘Teaser wais masice masn peme 14 urdvas perpanar sa ue n aecyons ecepes ul
   seee oi sure smren
                                                                                           Aurmcaoy
           enade                        wens                         mm
                        (soren empemuu oveshionous         comssesnsuiesn
   evecresemmne         Gvmeme ncuuroysisisofnis                saoceiman
                        (somo eememet w ovsesizrrons            aicesoumin
                                     Teroiaeaong                       +0           semsuns Insiams
                        loomariamsava on oveasiseczas:             uns                   rave
                         ic0iosieis en ovaesisnco                 wsine|     (edmonmomemmemy
                              Geroriscon sviand in 11        tonus no
                              tesvorsst on svian sedness      Cot ow as
                              (nsovist ov svianisncore         cussiesn
                              (nsorise arsvianisnzor           recomcoo           verrnis mm mues
          lnna
                           To msvino %a pmesaes) meg wo              ra               waniee Lewe
                                                           ons m ow sixrd porn nmusdaty roseneeo
        ‘war» Rsung por oi # t arsemti xmunnim onGxmcoam pmon m paorpass wog sn voreeame iy
    «ommo uet on nated Buromy m o vend e fnnoone on.mngins ine sscamooen o oo summmirereu out
     (n sumonnston omunraniic aa ces o muepors muoms s Apgnmses ag mcwurcop es voreruns s
                                            EFFFFEFIZE .——
                                                snroyies voneses jo vonsitonn on 2 wunty mmmminn
                                                w suros swosatn sut avo u eovins vonmpary seas ou.
                                                  sommpaccy se ftoren i sou0 rmoes sons oa tt pnpsecy
  wones vorsan mins
 umm pcnmeys omues
mepmamerommenes


Ralibration Laboratory of                                                    g.   SomaizericterKaterirsionst
3chmid & Partner                                                             (    Sevce se atiiomane
   Engineering AG                                                                 Brrvico esmm citaaies
 tevarsucevzese i3 8016 Zinch Sutzmrind                                      .    smesCataton Sarvice
                                                                              Accrataton io: SCS 108
Acrudte bie SFadent flce of Marsog and Revediation
e SuiseAccredtation Sevice s one oth ignstoris tothe EA
insttmwa Agreonent i th recogntion f catrtion coritcnies
Slossary:
TSL                     tssue simutating liquid
ComyE                   sonsitiity in TSL / NORM xyz
NA                      not applicable or not measured

Calibration is Performed According to the Following Standards:
                                                                      the Peak Spatial—
    a) IEEE Sid 1520—2003, "IEEE Recommended Practice for Determining
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques‘, December 2003
                                                                          Absorption Rate
    b) CENELEC EN 50361, "Basic standard for the measurement of Specific
                                                                                    MHz — 3 GHz),
         related to human exposure to electromagnetic fields from mobie phones (300
         July 2001
    c) Federal Communications Commission Officees offorEngineer ing & Technology (FCC OET),
         "Evaluating Compliance with FCC Guidelin       Human Exposure to Radiofrequency
                                                                               nce of Mobile and
         Electromagnetic Fields; Additional Information for Evaluating Complia
         Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions‘,
         Supplement C (Edition 01—01) to Buletin 65

Additional Documentation:
   d) DASY4 Systom Handbook
Methos Applied and Interpretation of Parameters:                                          at the end
     +   Measurement Conditions: Further details are available from the Validation Report
         of the certficate. Allfigures stated in the certificato aro valid at the frequency indicated
     + Antenna Parameters with TSL: The dipole is mounted with the spacer to position its food
       point exactly below the center marking of the flat phantom section, with the arms oriented
       paralle! to the body axis.
     + Feed PointImpedance and Return Loss: These parameters are measured with the dipols
         postioned under the liquid filed phantom. The impedance stated is transformed from the
       measurement at the SMA connector to the feed point. The Retun Loss ensures low
       reflected power. No uncertainty required.
     + Electrical Delay: One—way delay between the SMA connector and the anterna feed point. No
       uncertainty required
     +   SAR measured: SAR measured at the stated antenna input power.
     +   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
       connector.
     + SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.


 Certfcats No: D9t0ve—122_u006                      Page2 ot6


Measurement Conditions
    DASY system contguration,as for as not      gven on page
     ASY version                                            onsve                            ver
     Exvapoiation                                   Agranced Extrapolaion
     Prantom                                       Wodor Fat Phanion ¥$ 0
     Distance Bipote Genter— T9L                            15 mm                         with Soncer
     Area Scanresotuton                                 ox dy= 18 mm
     Zoom Scan Resolstion                             dn, dy. oz =5 om
     Frequency                                        o0 M + 1 Mitz

"lead TSL parameters
    ‘The fotaning parametere and caeuiaions were anoted
                                                               Temporature   Permitivty        Conductiity
      Nominal Head TSL paramaters                                2o‘           a1s             097 mhoim
      Mensured Head TSL paramaters                             @oz02‘c       41.126%         096 mhainz6%
      Head TSL temoerature during test                         @orea‘c          —                  —

    R result with Head TSL

      SA averaged over 1 en" (1 g)of Head TSL               Conttion
      SAR measires                                      260 m nout pover               273 mwig
    | SHR nomalzes                                       romatznd i 1w                 109 mivig
    SA fer nomina Head TSt parameters*                   romatzed t 1w          t0.9 mwily 2 170 % (c=2)
     Sh averaged over 19 cn? (10) of Head T9L                 conction
     SAR messured                                       260 mW nout poner              175010
     SA¥ nomatzes                                         normabiznd io 1W             70omiio
     SA fornominal Head TGL parametore®                   romatiznd io 1W       699 mW io : 165 % (eet)




 Sormeon o nonina TS parametirsaccoring to ) chacler ‘SAR Genaitutios®
C arthcats Ne:Daoov2.122.un06                        Pagesor6


Appondix
Antonna Parameters with Head TSL
     [impedance,ranstormed to fees pont                                         so2u—c4j0
     [Rewrntoss                                                                   —moco


General Antenna Parameters and Design
    Electical Delar (one drecton)                        I                         141218                         ]
After iong tm use with 100W radates power, ony a signt warming otme dipole neat feedpait can be measired
‘The dpole is made of standard semirgid coaial cable. The center conductor ofth feeding in is drecty comnectad t the
secand arm of thecipole e anienna is tnerefor atortrcuted for Do—ignaia
No excasaiveforse mustbe appliad t th dipae arms. because they might bend or th scidered connactions nearthe
fredgoint may be damages

Additional EUT Data

     Manatactres oy                                                               seene
     wanatacured on                                                             Juy o«, zoo




Cericate No: D00V2—122.ncs                          Pagect


   DASY4 Validation Report for Head TSL
                                                                             Date/Time: 16.06.2006 11:35:14
   Test Laboratory: SPEAG, Zurich, Switzerland
   DUT: Dipole 900 MHz: Type: D9UOV2; Serial: DOGV2 — SN:122
   Communication System: CW; Frequency: 900 MH;DutyCycle: I:1
   Medium: HSL 900 MHz;
    Medium parameters used:         900 Mitzz a =0.961 mhoim; s = 41.35 o — 1000 k/‘
   Phantom section: Flat Section
   Measurement Standard: DASY4 (Hligh Precision Assessment)

    DASY4 Configuration:
      + Risbe ETIDV6— SN1SD7(P): ConnF(SR, 58 5%; Calitraed: 28 10z008
      +. SensorSurfice dn (Medanical Surfce Daectin)
      + Hecvonics DMA Sn60l; Callbaiet 1512.005
      + Praniom: is Phaton 0 lacty Type:QDIOOPSOAA: :
      +. Measiement SWDASV VA Buld 25;Pstpicesing SW:SEMCAD, V1.a Bail171

    Pin = 250 mW;      10 mm/Area Sean (71x81x1): Measwrement grid: dx=15mm, dy=1 Smm
    Maximum value of SAR (nterpolated) = 2.94 mWig
                                                                              dy—Smm,
    Pin = 250 mW; d = 10 mm/Zoom Sean (7x7x7)/Cube0: Measurement grd: dx=Smm,
    d=Smm
    Reference Value — 57.7 Vim;     Power Drift   =—0.103 dB
    Peak SAR (extrapolated) 12 Wike
    SAR(I g) =2.73 mWig SAR(IO g) = 1.75 mWig
    Maximum value ofSAR (measured) = 2.97 mW/g
             «s
              uose
              s


              aze
              ns
                 a

              ase
            riwe 1



    Certfcate No: ©300vz—122_un06                      Prgesors




wnopounnunreuse                                   kess                                 onicveimaue nepurs


impedance Measurement Plot for Head TSL
                                            wn zons soresces
                                                      mose se mz




             Sn reuses sen w                     sroe i feasen se wis




Certfcate No: D900v2—122unds          Pagesors


AnyData Corporation                                                                 FCC ID: P4MAGT100D
APPENDIX D - TEST SYSTEM VERIFICATIONS SCANS
Liquid Measurement Result

835 MHz

                                                Liquid          Target   Measured   Deviation     Limits
 Simulant     Freq [MHz]      Parameters
                                               Temp [ºC]        Value     Value       [%]          [%]
                                   ε              22             41.5     41.3       -0.48          ±5
  Head            835              σ              22             0.90     0.89       -1.12          ±5
                                1g SAR            22             9.5      9.42       -0.85         ±10
                                   ε              22             55.2     55.3       +0.18          ±5
  Body            835              σ              22             0.97     0.99        +2.0          ±5
                                1g SAR            22            7.873     7.91       +0.47         ±10

1900 MHz

                                                Liquid          Target   Measured   Deviation     Limits
 Simulant     Freq [MHz]      Parameters
                                               Temp [ºC]        Value     Value       [%]          [%]
                                   ε              22             40.0      39.9      -0.25          ±5
  Head           1900              σ              22             1.40      1.38      -1.45          ±5
                                1g SAR            22             39.7      39.5      -0.51         ±10
                                   ε              22             53.3      53.2      -0.19          ±5
  Body           1900              σ              22             1.52      1.51      -0.66          ±5
                                1g SAR            22            24.97     24.88      -0.36         ±10

εr = relative permittivity, σ = conductivity and ρ=1000kg/m3




Report#R0607263S                                Page 52 of 68                        SAR Evaluation Report


AnyData Corporation                                                              FCC ID: P4MAGT100D
Test Laboratory: Bay Area Compliance Lab Corp.(BACL)
System Validation Check D835 Head
DUT: Dipole 900 MHz; Type: D900V2; Serial Number: 122

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium parameters used: f = 835 MHz; σ = 0.89 mho/m; εr = 41.3; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:

    •   Probe: ET3DV6 - SN1604; ConvF(6.6, 6.6, 6.6); Calibrated: 5/2/2006

    •   Sensor-Surface: 4mm (Mechanical Surface Detection)

    •   Electronics: DAE3 Sn456; Calibrated: 10/18/2005

    •   Phantom: SAM with CRP; Type: Twin SAM; Serial: TP-1032

    •   Measurement SW: DASY4, V4.6 Build 23; Postprocessing SW: SEMCAD, V1.8 Build 161


d=15mm, Pin=1W/Area Scan (61x121x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (interpolated) = 9.82 mW/g

d=15mm, Pin=1W/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 104.1 V/m; Power Drift = -0.085 dB
Peak SAR (extrapolated) = 14.4 W/kg
SAR(1 g) = 9.42 mW/g; SAR(10 g) = 5.99 mW/g
Maximum value of SAR (measured) = 10.2 mW/g




        0 dB = 10.2mW/g




Report#R0607263S                              Page 53 of 68                       SAR Evaluation Report


AnyData Corporation                                                              FCC ID: P4MAGT100D


Test Laboratory: Bay Area Compliance Lab Corp.(BACL)
System Validation Check D835 Body
DUT: Dipole 900 MHz; Type: D900V2; Serial Number: 122

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium parameters used: f = 835 MHz; σ = 0.99 mho/m; εr = 55.3; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:

    •   Probe: ET3DV6 - SN1604; ConvF(6.27, 6.27, 6.27); Calibrated: 5/2/2006

    •   Sensor-Surface: 4mm (Mechanical Surface Detection)

    •   Electronics: DAE3 Sn456; Calibrated: 10/18/2005

    •   Phantom: SAM with CRP; Type: Twin SAM; Serial: TP-1032

    •   Measurement SW: DASY4, V4.6 Build 23; Postprocessing SW: SEMCAD, V1.8 Build 161


d=15mm, Pin=1W/Area Scan (61x121x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (interpolated) = 8.01 W/g

d=15mm, Pin=1W/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 97.8 V/m; Power Drift = -0.065 dB
Peak SAR (extrapolated) = 12.9 W/kg
SAR(1 g) = 7.91 mW/g; SAR(10 g) = 5.89 mW/g
Maximum value of SAR (measured) = 8.04 mW/g




            0 dB = 8.04mW/g



Report#R0607263S                              Page 54 of 68                       SAR Evaluation Report


AnyData Corporation                                                               FCC ID: P4MAGT100D


Test Laboratory: Bay Area Compliance Lab Corp.(BACL)
System Validation Check_D1900_Head
DUT: Dipole 1900 MHz; Type: D-1800-S-1; Serial Number: BCL-049

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium parameters used: f = 1900 MHz; σ = 1.38 mho/m; εr = 39.9; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:

    •   Probe: ET3DV6 - SN1604; ConvF(5.29, 5.29, 5.29); Calibrated: 5/2/2006

    •   Sensor-Surface: 4mm (Mechanical Surface Detection)

    •   Electronics: DAE3 Sn456; Calibrated: 10/18/2005

    •   Phantom: SAM with CRP; Type: Twin SAM; Serial: TP-1032

    •   Measurement SW: DASY4, V4.6 Build 23; Postprocessing SW: SEMCAD, V1.8 Build 161


d=10mm, Pin=1W/Area Scan (51x91x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (interpolated) = 40.5 mW/g

d=10mm, Pin=1W/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 111.3 V/m; Power Drift = -0.154 dB
Peak SAR (extrapolated) = 45.4 W/kg
SAR(1 g) = 39.5 mW/g; SAR(10 g) = 19.9 mW/g
Maximum value of SAR (measured) = 40.1 mW/g




            0 dB = 40.1mW/g

Report#R0607263S                              Page 55 of 68                        SAR Evaluation Report


AnyData Corporation                                                               FCC ID: P4MAGT100D


Test Laboratory: Bay Area Compliance Lab Corp.(BACL)
System Validation Check_D1900_Body
DUT: Dipole 1900 MHz; Type: D-1800-S-1; Serial Number: BCL-049

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium parameters used: f = 1900 MHz; σ = 1.51 mho/m; εr = 53.2; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:

    •   Probe: ET3DV6 - SN1604; ConvF(4.88, 4.88, 4.88); Calibrated: 5/2/2006

    •   Sensor-Surface: 4mm (Mechanical Surface Detection)

    •   Electronics: DAE3 Sn456; Calibrated: 10/18/2005

    •   Phantom: SAM with CRP; Type: Twin SAM; Serial: TP-1032

    •   Measurement SW: DASY4, V4.6 Build 23; Postprocessing SW: SEMCAD, V1.8 Build 161


d=10mm, Pin=1W/Area Scan (51x91x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (interpolated) = 24.91 mW/g

d=10mm, Pin=1W/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 110.2 V/m; Power Drift = 0.002 dB
Peak SAR (extrapolated) = 56.9 W/kg
SAR(1 g) = 24.88 mW/g; SAR(10 g) = 11.24 mW/g
Maximum value of SAR (measured) = 25.7 mW/g




            0 dB = 25.7mW/g




Report#R0607263S                              Page 56 of 68                        SAR Evaluation Report


AnyData Corporation                                                                FCC ID: P4MAGT100D
APPENDIX E - EUT SCANS

Test Laboratory: Bay Area Compliance Lab Corp.(BACL)
Body worn Low CH with belt clip
DUT: AnyData AGT-100D; Type: Wireless Data Module; Serial Number: 2019418

Communication System: CDMA; Frequency: 824.7 MHz;Duty Cycle: 1:1
Medium parameters used: f = 824.7 MHz; σ = 0.99 mho/m; εr = 55.3; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:

    •   Probe: ET3DV6 - SN1604; ConvF(6.27, 6.27, 6.27); Calibrated: 5/2/2006

    •   Sensor-Surface: 4mm (Mechanical Surface Detection)

    •   Electronics: DAE3 Sn456; Calibrated: 10/18/2005

    •   Phantom: SAM with CRP; Type: Twin SAM; Serial: TP-1032

    •   Measurement SW: DASY4, V4.6 Build 23; Postprocessing SW: SEMCAD, V1.8 Build 161


Back Touch the Phantom with Belt Clip Low CH/Area Scan (61x81x1): Measurement grid: dx=15mm,
dy=15mm
Maximum value of SAR (interpolated) = 0.400 mW/g

Back Touch the Phantom with Belt Clip Low CH/Zoom Scan (5x5x5)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 20.5 V/m; Power Drift = 0.03dB
Peak SAR (extrapolated) = 0.590 W/kg
SAR(1 g) = 0.380 mW/g; SAR(10 g) = 0.247 mW/g
Maximum value of SAR (measured) = 0.429 mW/g




                   0 dB = 0.429mW/g
                                                   Plot#1
Report#R0607263S                              Page 57 of 68                         SAR Evaluation Report


AnyData Corporation                                                                 FCC ID: P4MAGT100D


Test Laboratory: Bay Area Compliance Lab Corp.(BACL)
Body worn Mid CH with belt clip
DUT: AnyData AGT-100D; Type: Wireless Data Module; Serial Number: 2019418

Communication System: CDMA; Frequency: 836.52 MHz;Duty Cycle: 1:1
Medium parameters used: f = 836.52 MHz; σ = 0.99 mho/m; εr = 55.2; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:

    •   Probe: ET3DV6 - SN1604; ConvF(6.27, 6.27, 6.27); Calibrated: 5/2/2006

    •   Sensor-Surface: 4mm (Mechanical Surface Detection)

    •   Electronics: DAE3 Sn456; Calibrated: 10/18/2005

    •   Phantom: SAM with CRP; Type: Twin SAM; Serial: TP-1032

    •   Measurement SW: DASY4, V4.6 Build 23; Postprocessing SW: SEMCAD, V1.8 Build 161


Back Touch the Phantom with Belt Clip Mid CH/Area Scan (61x81x1): Measurement grid: dx=15mm,
dy=15mm
Maximum value of SAR (interpolated) = 0.907 mW/g

Back Touch the Phantom with Belt Clip Mid CH/Zoom Scan (5x5x5)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 29.6 V/m; Power Drift = 0.02dB
Peak SAR (extrapolated) = 1.21 W/kg
SAR(1 g) = 0.821 mW/g; SAR(10 g) = 0.538 mW/g
Maximum value of SAR (measured) = 0.878 mW/g




            0 dB = 0.878mW/g
                                                   Plot#2

Report#R0607263S                              Page 58 of 68                          SAR Evaluation Report


AnyData Corporation                                                                 FCC ID: P4MAGT100D


Test Laboratory: Bay Area Compliance Lab Corp.(BACL)
Body worn High CH with belt clip
DUT: AnyData AGT-100D; Type: Wireless Data Module; Serial Number: 2019418

Communication System: CDMA; Frequency: 848.31 MHz;Duty Cycle: 1:1
Medium parameters used: f = 848.31 MHz; σ = 0.99 mho/m; εr = 55.2; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:

    •   Probe: ET3DV6 - SN1604; ConvF(6.27, 6.27, 6.27); Calibrated: 5/2/2006

    •   Sensor-Surface: 4mm (Mechanical Surface Detection)

    •   Electronics: DAE3 Sn456; Calibrated: 10/18/2005

    •   Phantom: SAM with CRP; Type: Twin SAM; Serial: TP-1032

    •   Measurement SW: DASY4, V4.6 Build 23; Postprocessing SW: SEMCAD, V1.8 Build 161


Back Touch the Phantom with Belt Clip High CH 3/Area Scan (61x91x1): Measurement grid: dx=15mm,
dy=15mm
Maximum value of SAR (interpolated) = 0.331 mW/g

Back Touch the Phantom with Belt Clip High CH 3/Zoom Scan (5x5x5)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 17.5 V/m; Power Drift = -0.03dB
Peak SAR (extrapolated) = 0.447 W/kg
SAR(1 g) = 0.302 mW/g; SAR(10 g) = 0.196 mW/g
Maximum value of SAR (measured) = 0.327 mW/g




                   0 dB = 0.327mW/g

                                                   Plot#3

Report#R0607263S                              Page 59 of 68                          SAR Evaluation Report


AnyData Corporation                                     FCC ID: P4MAGT100D




                                             Z-Axis




Test Laboratory: Bay Area Compliance Lab Corp.(BACL)

Report#R0607263S                        Page 60 of 68    SAR Evaluation Report


AnyData Corporation                                                               FCC ID: P4MAGT100D
Body worn Low CH with belt clip
DUT: AnyData AGT-100D; Type: Wireless Data Module; Serial Number: 2019418

Communication System: CDMA; Frequency: 1851.25 MHz;Duty Cycle: 1:1
Medium parameters used: f = 1900 MHz; σ = 1.53 mho/m; εr = 53.4; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:

    •   Probe: ET3DV6 - SN1604; ConvF(4.88, 4.88, 4.88); Calibrated: 5/2/2006

    •   Sensor-Surface: 4mm (Mechanical Surface Detection)

    •   Electronics: DAE3 Sn456; Calibrated: 10/18/2005

    •   Phantom: SAM with CRP; Type: Twin SAM; Serial: TP-1032

    •   Measurement SW: DASY4, V4.6 Build 23; Postprocessing SW: SEMCAD, V1.8 Build 161


Back Touch the Phantom with Belt Clip Low CH /Area Scan (61x81x1): Measurement grid: dx=15mm,
dy=15mm
Maximum value of SAR (interpolated) = 1.12 mW/g

Back Touch the Phantom with Belt Clip Low CH /Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 17.9 V/m; Power Drift = 0.04dB
Peak SAR (extrapolated) = 1.88 W/kg
SAR(1 g) = 0.943 mW/g; SAR(10 g) = 0.562 mW/g

Maximum value of SAR (measured) = 1.03 mW/g




            0 dB = 1.03mW/g

                                                   Plot#4




Report#R0607263S                              Page 61 of 68                        SAR Evaluation Report


AnyData Corporation                                                               FCC ID: P4MAGT100D
Test Laboratory: Bay Area Compliance Lab Corp.(BACL)
Body worn Mid CH with belt clip
DUT: AnyData AGT-100D; Type: Wireless Data Module; Serial Number: 2019418

Communication System: CDMA; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium parameters used: f = 1900 MHz; σ = 1.51 mho/m; εr = 53.2; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:

    •     Probe: ET3DV6 - SN1604; ConvF(4.88, 4.88, 4.88); Calibrated: 5/2/2006

    •     Sensor-Surface: 4mm (Mechanical Surface Detection)

    •     Electronics: DAE3 Sn456; Calibrated: 10/18/2005

    •     Phantom: SAM with CRP; Type: Twin SAM; Serial: TP-1032

    •     Measurement SW: DASY4, V4.6 Build 23; Postprocessing SW: SEMCAD, V1.8 Build 161


Back Touch the Phantom with Belt Clip Mid CH/Area Scan (61x81x1): Measurement grid: dx=15mm,
dy=15mm
Maximum value of SAR (interpolated) = 1.62 mW/g

Back Touch the Phantom with Belt Clip Mid CH/Zoom Scan (5x5x5)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 28.2 V/m; Power Drift = 0.03dB
Peak SAR (extrapolated) = 2.37 W/kg
SAR(1 g) = 1.46 mW/g; SAR(10 g) = 0.863 mW/g

Maximum value of SAR (measured) = 1.59 mW/g




        Plot#5




Test Laboratory:                                                                      Bay Area
Compliance Lab                 0 dB = 1.59mW/g                                        Corp.(BACL)
Body worn High CH with belt clip
DUT: AnyData AGT-100D; Type: Wireless Data Module; Serial Number: 2019418

Communication System: CDMA; Frequency: 1908.75 MHz;Duty Cycle: 1:1
Medium parameters used: f = 1900 MHz; σ = 1.51 mho/m; εr = 53.2; ρ = 1000 kg/m3
Phantom section: Flat Section
Report#R0607263S                               Page 62 of 68                       SAR Evaluation Report


AnyData Corporation                                                            FCC ID: P4MAGT100D
DASY4 Configuration:

   •   Probe: ET3DV6 - SN1604; ConvF(4.88, 4.88, 4.88); Calibrated: 5/2/2006

   •   Sensor-Surface: 4mm (Mechanical Surface Detection)

   •   Electronics: DAE3 Sn456; Calibrated: 10/18/2005

   •   Phantom: SAM with CRP; Type: Twin SAM; Serial: TP-1032

   •   Measurement SW: DASY4, V4.6 Build 23; Postprocessing SW: SEMCAD, V1.8 Build 161


Back Touch the Phantom with Belt Clip High CH/Area Scan (61x81x1): Measurement grid: dx=15mm,
dy=15mm
Maximum value of SAR (interpolated) = 1.60 mW/g

Back Touch the Phantom with Belt Clip High CH/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 21.6 V/m; Power Drift = 0.04dB
Peak SAR (extrapolated) = 2.14 W/kg
SAR(1 g) = 1.31 mW/g; SAR(10 g) = 0.765 mW/g

Maximum value of SAR (measured) = 1.45 mW/g




                   0 dB = 1.45mW/
                                                  Plot#6




Report#R0607263S                            Page 63 of 68                       SAR Evaluation Report


AnyData Corporation                   FCC ID: P4MAGT100D




                          Z-Axis




Report#R0607263S      Page 64 of 68    SAR Evaluation Report


AnyData Corporation                                                                FCC ID: P4MAGT100D


APPENDIX F – CONDUCTED OUTPUT POWER MEASUREMENT
Provision Applicable
The measured peak output power should be greater and within 5% than EMI measurement.


Test Procedure
The RF output of the transmitter was connected to the input of the spectrum analyzer through sufficient
attenuation.


Test equipment
Agilent PSA E4446A Spectrum Analyzer, Calibration Due Date: 2007-03-06.


Test Results

      Frequency (MHz)        Output Power in dBm            Output Power in Watt
           824.7                    26.15                          0.412
           836.52                   26.03                          0.400
           848.3                    25.97                          0.395
          1851.25                   26.24                          0.420
            1880                    26.14                          0.411
          1908.75                   26.27                          0.424




Report#R0607263S                            Page 65 of 68                           SAR Evaluation Report


AnyData Corporation                                    FCC ID: P4MAGT100D

APPENDIX G – EUT TEST POSITION PHOTOS
Body worm setup photo with belt clip




Report#R0607263S                       Page 66 of 68    SAR Evaluation Report


AnyData Corporation                     FCC ID: P4MAGT100D


APPENDIX H – EUT & ACCESSORIES PHOTOS
EUT - Top View




EUT - Bottom View




Report#R0607263S       Page 67 of 68     SAR Evaluation Report


AnyData Corporation                                                                       FCC ID: P4MAGT100D



APPENDIX I - INFORMATIVE REFERENCES
[1] Federal Communications Commission, \Report and order: Guidelines for evaluating the environmental effects of
radiofrequency radiation", Tech. Rep. FCC 96-326, FCC, Washington, D.C. 20554, 1996.

[2] David L. Means Kwok Chan, Robert F. Cleveland, \Evaluating compliance with FCC guidelines for human
exposure to radiofrequency electromagnetic fields", Tech. Rep., Federal Communication Commission, O_ce of
Engineering & Technology, Washington, DC, 1997.

[3] Thomas Schmid, Oliver Egger, and Niels Kuster, \Automated E-_eld scanning system for dosimetric
assessments", IEEE Transactions on Microwave Theory and Techniques, vol. 44, pp. 105{113, Jan. 1996.

[4] Niels Kuster, Ralph K.astle, and Thomas Schmid, \Dosimetric evaluation of mobile communications
equipment with known precision", IEICE Transactions on Communications, vol. E80-B, no. 5, pp. 645{652, May
1997.

[5] CENELEC, \Considerations for evaluating of human exposure to electromagnetic fields (EMFs) from mobile
telecommunication equipment (MTE) in the frequency range 30MHz - 6GHz", Tech. Rep., CENELEC, European
Committee for Electrotechnical Standardization, Brussels, 1997.

[6] ANSI, ANSI/IEEE C95.1-1992: IEEE Standard for Safety Levels with Respect to Human Exposure to Radio
Frequency Electromagnetic Fields, 3 kHz to 300 GHz, The Institute of Electrical and Electronics Engineers, Inc.,
New York, NY 10017, 1992.

[7] Katja Pokovic, Thomas Schmid, and Niels Kuster, \Robust setup for precise calibration of E-field probes in
tissue simulating liquids at mobile communications frequencies", in ICECOM _ 97, Dubrovnik, October 15{17,
1997, pp. 120-24.

[8] Katja Pokovic, Thomas Schmid, and Niels Kuster, \E-field probe with improved isotropy in brain simulating
liquids", in Proceedings of the ELMAR, Zadar, Croatia, 23{25 June, 1996, pp. 172-175.

[9] Volker Hombach, Klaus Meier, Michael Burkhardt, Eberhard K. uhn, and Niels Kuster, \The depen-dence of EM
energy absorption upon human head modeling at 900 MHz", IEEE Transactions on Microwave Theory and
Techniques, vol. 44, no. 10, pp. 1865-1873, Oct. 1996.

[10] Klaus Meier, Ralf Kastle, Volker Hombach, Roger Tay, and Niels Kuster, \The dependence of EM energy
absorption upon human head modeling at 1800 MHz", IEEE Transactions on Microwave Theory and Techniques,
Oct. 1997, in press.

[11] W. Gander, Computermathematik, Birkhaeuser, Basel, 1992.

[12] W. H. Press, S. A. Teukolsky,W. T. Vetterling, and B. P. Flannery, Numerical Recepies in C, The Art of
Scientific Computing, Second Edition, Cambridge University Press, 1992.Dosimetric Evaluation of Sample device,
month 1998 9

[13] NIS81 NAMAS, \The treatment of uncertainity in EMC measurement", Tech. Rep., NAMAS Executive,
National Physical Laboratory, Teddington, Middlesex, England, 1994.

[14] Barry N. Taylor and Christ E. Kuyatt, \Guidelines for evaluating and expressing the uncertainty of
NIST measurement results", Tech. Rep., National Institute of Standards and Technology, 1994.
Dosimetric Evaluation of Sample device, month 1998 10.




Report#R0607263S                                 Page 68 of 68                              SAR Evaluation Report