FCC15. 247-68.930.14.004.01

FCC ID: OU9LS208-B06

Test Report

Download: PDF
FCCID_2507240

Report Number: 68.930.14.004.01




                                                               FCC - TEST REPORT



Report Number                                          : 68.930.14.004.01                               Date of Issue:                December 25, 2014



Model                                                  : LS208-B6

Product Type                                           : Body Fat Analyzer

Applicant                                              : Guangdong Transtek Medical Electronics Co., Ltd

Address                                                : Zone A, 5/F., Investment Building, No. 12 Huizhan East Rd.,

                                                            Torch Development District, Zhongshan, Guangdong, China

                                                            528437

Production Facility                                    : Guangdong Transtek Medical Electronics Co., Ltd

Address                                                : Zone A, 5/F., Investment Building, No. 12 Huizhan East Rd.,

                                                            Torch Development District, Zhongshan, Guangdong, China

                                                            528437




Test Result                                            :  Positive                     Negative



Total pages including
Appendices                                             : 23



TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch is a subcontractor to TÜV SÜD Product Service GmbH according to the principles outlined in
ISO 17025.

TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch reports apply only to the specific samples tested under stated test conditions. Construction of
the actual test samples has been documented. It is the manufacturer's responsibility to assure that additional production units of this model are manufactured with
identical electrical and mechanical components. The manufacturer/importer is responsible to the Competent Authorities in Europe for any modifications made to the
production units which result in non-compliance to the relevant regulations. TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch shall have no
liability for any deductions, inferences or generalizations drawn by the client or others from TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch
issued reports.

This report is the confidential property of the client. As a mutual protection to our clients, the public and ourselves, extracts from the test report shall not be
reproduced except in full without our written approval




EMC_SZ_FR_21.00 FCC                                     TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                                Page 1 of 23
                                           Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                                Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




1 Table of Contents
1     Table of Contents ................................................................................................................................................................... 2
2     Details about the Test Laboratory .......................................................................................................................................... 3
3     Description of the Equipment Under Test ............................................................................................................................. 4
4     Summary of Test Standards ................................................................................................................................................... 5
5     Summary of Test Results ....................................................................................................................................................... 6
6     General Remarks.................................................................................................................................................................... 7
7     Test Setups ............................................................................................................................................................................. 8
8     Systems test configuration ..................................................................................................................................................... 9
9     Technical Requirement ........................................................................................................................................................ 10
    9.1       Conducted peak output power ..................................................................................................................................... 10
    9.2       Power spectral density................................................................................................................................................. 11
    9.3       6 dB Bandwidth .......................................................................................................................................................... 12
    9.4       Spurious RF conducted emissions ............................................................................................................................... 14
    9.5       Band edge testing ........................................................................................................................................................ 18
    9.6       Spurious radiated emissions for transmitter ................................................................................................................ 20
10    Test Equipment List ............................................................................................................................................................. 22
11    System Measurement Uncertainty ....................................................................................................................................... 23




EMC_SZ_FR_21.00 FCC                                       TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                                              Page 2 of 23
                                             Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                                  Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




2 Details about the Test Laboratory
Details about the Test Laboratory
Test Site 1

Company name:         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch
                      Building 12&13, Zhiheng Wisdomland Business Park,
                      Nantou Checkpoint Road 2,
                      Nanshan District, 518052,
                      Shenzhen,P.R.C.

Telephone:            86 755 8828 6998
Fax:                  86 755 8828 5299


Test Site 2
Company name:         Audix Technology (shenzhen) Co.,Ltd
                      Block Shenzhen, Science & Industry Park,
                      Nantou, Shenzhen,
                      Guangdong,
                      China

Telephone:            86 755 2663 9496
Fax:                  86 755 2663 2877




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 3 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




3 Description of the Equipment Under Test

Product:                          Body Fat Analyzer

Model no.:                        LS208-B6

FCC ID:                           OU9LS208-B06

Options and accessories:          NIL

Rating:                           DC6.0V (Supplied by 4X1.5V AAA battery)

RF Transmission                   2402-2480MHz
Frequency:
No. of Operated Channel:          40

Modulation:                       GFSK

Antenna Type:                     PIFA

Antenna Gain:                     0dBi

Description of the EUT:           The Equipment Under Test (EUT) is a Bluetooth body fat analyzer




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 4 of 23
                          Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20               Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




4 Summary of Test Standards

                                         Test Standards
FCC Part 15 Subpart C       PART 15 - RADIO FREQUENCY DEVICES
10-1-2014 Edition           Subpart C - Intentional Radiators

All the test methods were according to KDB558074 D01 DTS Meas Guidance v03r02 issued by
FCC on July 05, 2014 and ANSI C63.10 (2009).




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 5 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


 Report Number: 68.930.14.004.01




 5 Summary of Test Results
                                         Technical Requirements
FCC Part 15 Subpart C
Test Condition                                                              Pages               Test Site                 Test Result
§15.207                     Conducted emission AC                            ---                   ---                       N/A
                            power port
§15.247(b)(1)               Conducted peak output                              10                  Site 2                    Pass
                            power
§15.247(e)                  Power spectral density                             11                  Site 2                    Pass

§15.247(a)(2)               6dB bandwidth                                      12                  Site 2                    Pass

§15.247(a)(1)               20dB bandwidth and                                 ---                    ---                    N/A
                            99% Occupied
                            Bandwidth
§15.247(a)(1)               Carrier frequency                                  ---                    ---                    N/A
                            separation
§15.247(a)(1)(iii)          Number of hopping                                  ---                    ---                    N/A
                            frequencies
§15.247(a)(1)(iii)          Dwell Time                                         ---                    ---                    N/A

§15.247(d)                  Spurious RF conducted                              14                  Site 2                    Pass
                            emissions
§15.247(d)                  Band edge                                          18                  Site 2                    Pass

§15.247(d) & §15.209        Spurious radiated                                  20                  Site 2                    Pass
&15.205                     emissions for transmitter

§15.203                     Antenna requirement                                      See note 2                              Pass

 Note 1: N/A=Not Applicable.
 Note 2: The EUT uses a permanently PIFA antenna, which gain is 0dBi. In accordance to §15.203, It is
 considered sufficiently to comply with the provisions of this section.




 EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                   Page 6 of 23
                          Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
 Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




6 General Remarks
Remarks

This submittal(s) (test report) is intended for FCC ID: OU9LS208-B06 complies with Section
15.207, 15.209, 15.247 of the FCC Part 15, Subpart C Rules.


SUMMARY:

All tests according to the regulations cited on page 5 were

 - Performed

 - Not Performed

The Equipment Under Test

 - Fulfills the general approval requirements.

 - Does not fulfill the general approval requirements.


Sample Received Date:                   November 27, 2014

Testing Start Date:                     November 28, 2014

Testing End Date:                       December 15, 2014

TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch

Reviewed by:                      Prepared by:                                                   Tested by:




_______________________                   _______________________                                 _______________________
      Phoebe Hu                                     Felix Li                                         Leo Li
  EMC Project Manager                         EMC Project Engineer                               EMC Test Engineer




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 7 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




7 Test Setups
7.1 Radiated test setups




7.2 Conducted RF test setups


                      Measuring
                      Receiver                                           EUT




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 8 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




8 Systems test configuration
Auxiliary Equipment Used during Test:

            DESCRIPTION            MANUFACTURER                    MODEL NO.(SHIELD)                     S/N(LENGTH)
              NoteBook                    Lenovo                              X220                               ---

Test software: BLE-CC254x-1.2.1 - Btool, which used to control the EUT in continues
transmitting mode

The system was configured to channel 0, 19, and 39 for the test.




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 9 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




9 Technical Requirement

9.1 Conducted peak output power

Test Method
   1. Use the following spectrum analyzer settings:
      RBW > the 6 dB bandwidth of the emission being measured, VBW≥3RBW, Span≥3RBW
      Sweep = auto, Detector function = peak, Trace = max hold.
   2. Add a correction factor to the display.
   3. Allow the trace to stabilize. Use the marker-to-peak function to set the marker to the
      peak of the emission. The indicated level is the peak output power.

Limits
According to §15.247 (b) (1), conducted peak output power limit as below:

                      Frequency Range                        Limit                          Limit
                            MHz                               W                             dBm
                         2400-2483.5                          ≤1                             ≤30



Test Result
                                                               Conducted Peak
                               Frequency                        Output Power                  Result
                                  MHz                               dBm
                         Low channel 2402MHz                        -2.54                      Pass
                        Middle channel 2440MHz                      -2.31                      Pass
                         High channel 2480MHz                       -2.77                      Pass




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 10 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




9.2 Power spectral density

Test Method
This procedure shall be used if maximum peak conducted output power was used to demonstrate
compliance:
       1. Set analyzer center frequency to DTS channel center frequency.
       RBW=3kHz,VBW≥3RBW,Span=1.5 times DTS bandwidth, Detector=Peak, Sweep=auto,
       Trace= max hold
       2. Allow trace to fully stabilize, use the peak marker function to determine the maximum
       amplitude level within the RBW .
       3. Repeat above procedures until other frequencies measured were completed
Limit

                                                       Limit [dBm]

                                                              ≤8

Test Result

                    Frequency          Power spectral                    Limit                Result
                       MHz                density                        dBm
                       2402                    -14.07                      8                       Pass
                       2440                    -13.69                      8                       Pass
                       2480                    -14.34                      8                       Pass




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 11 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




9.3 6 dB Bandwidth
Test Method
   1. Use the following spectrum analyzer settings:
      RBW=100K, VBW≥3RBW, Sweep = auto, Detector function = peak, Trace = max hold
   2. 2. Use the automatic bandwidth measurement capability of an instrument, may be
      employed using the X dB bandwidth mode with X set to 6 dB, care shall be taken so that
      the bandwidth measurement is not influenced by any intermediate power nulls in the
      fundamental emission that might be ≥ 6 dB.
   3. Allow the trace to stabilize, record the X dB Bandwidth value.

Limit

                                                        Limit [kHz]

                                                            ≥500


Test result
                   Frequency            6 dB Bandwidth                   Limit                 Result
                      MHz                     kHz                        kHz
                      2402                   664.3                       ≥500                      Pass
                      2440                   686.0                       ≥500                      Pass
                      2480                   591.0                       ≥500                      Pass




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 12 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 13 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




9.4 Spurious RF conducted emissions

Test Method
   1. Use the following spectrum analyzer settings:
      Span = wide enough to capture the peak level of the in-band emission and all spurious
      emissions (e.g., harmonics) from the lowest frequency generated in the EUT up through
      the 10th harmonic. Typically, several plots are required to cover this entire span.
      RBW = 100 kHz, VBW≥RBW, Sweep = auto, Detector function = peak, Trace = max hold
   2. Allow the trace to stabilize. Set the marker on the peak of any spurious emission
      recorded.
   3. The level displayed must comply with the limit specified in this Section. Submit these
      plots.
   4. Repeat above procedures until all frequencies measured were complete.

Limit

                                    Frequency Range                          Limit (dBc)
                                          MHz
                                        30-25000                                   -20




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 14 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




Spurious RF conducted emissions

2402MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 15 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




Spurious RF conducted emissions

2440MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 16 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




Spurious RF conducted emissions

2480MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 17 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




9.5 Band edge testing

Test Method
   1 Use the following spectrum analyzer settings:
      Span = wide enough to capture the peak level of the in-band emission and all spurious
      RBW = 100 kHz, VBW≥RBW, Sweep = auto, Detector function = peak, Trace = max hold
   2 Allow the trace to stabilize, use the peak and delta measurement to record the result.
   3 The level displayed must comply with the limit specified in this Section. .

Limit

                                    Frequency Range                          Limit (dBc)
                                          MHz
                                        30-25000                                   -20




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 18 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




Band edge testing




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 19 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




9.6 Spurious radiated emissions for transmitter

Test Method
   1. The EUT is placed on a turntable, which is 0.8m above ground plane.
   2. EUT is set 3m away from the receiving antenna, which is varied from 1m to 4m to find out
      the highest emissions.
   3. Use the following spectrum analyzer settings:
      Span = wide enough to fully capture the emission being measured, RBW = 1 MHz for f ≥
      1GHz, 100 kHz for f < 1 GHz, VBW ≥ RBW, Sweep = auto, Detector function = peak,
      Trace = max hold
   4. Follow the guidelines in ANSI C63.4-1992 with respect to maximizing the emission by
      rotating the EUT, adjusting the measurement antenna height and polarization, etc.
      The peak reading of the emission, after being corrected by the antenna factor, cable loss,
      pre-amp gain, etc., is the peak field strength,submit this data. Each emission was to be
      maximized by changing the polarization of receiving antenna both horizontal and vertical.
   5. Set the VBW to 10 Hz, while maintaining all of the other instrument settings. This peak
      level, once corrected, must comply with the limit specified in Section 15.209. If the duty
      cycle per channel of the hopping signal is less than 100 ms, then the reading obtained
      with the 10 Hz VBW may be further adjusted by a “duty cycle correction factor”, derived
      from 20log(duty cycle/100 ms), in an effort to demonstrate compliance with the 15.209
      limit. Submit this data.

Limit

According to part 15.247(d), the radio emission outside the operating frequency band shall be at
least 20 dB below that in the 100 kHz bandwidth within the band that contains the highest level
of the desired power. Radiated emissions which fall in the restricted bands, as defined in
section15.205, must comply with the radiated emission limits specified in section 15.209.

              Frequency                 Field Strength                 Field Strength                       Detector
                  MHz                        uV/m                         dBμV/m
                  30-88                       100                            40                                 QP
                 88-216                       150                           43.5                                QP
                216-960                       200                            46                                 QP
               960-1000                       500                            54                                 QP
              Above 1000                      500                            54                                 AV
              Above 1000                     5000                            74                                 PK




EMC_SZ_FR_21.00 FCC                     TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 20 of 23
                           Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


   Report Number: 68.930.14.004.01




   Spurious radiated emissions for transmitter

   According to ANSI C63.10, if the peak (or quasi-peak) measured value complies with the
   average limit, it is unnecessary to perform an average measurement, so AV emission value did
   not show in below table if the peak value complies with average limit.

   Transmitting spurious emission test result as below:

   2402MHz Test Result
             Antenna     Cable      Amp.                              Emission
Frequency                                          Reading                               Polarization           Limit          Detector   Result
              Factor     Loss       Factor                             Level
   MHz        dB/m         dB        dB              dBuV             dBuV/m                                  dBμV/m
 30-1000        --          --        --               --                --                Horizontal           --               QP        Pass
 30-1000        --          --        --               --                --                 Vertical            --               QP        Pass
  *4804       32.85       8.56      35.70            45.34             51.05               Horizontal           74               PK        Pass
  *4804       32.85       8.56      35.70              --                                  Horizontal           54               AV        Pass
  *4804       32.85       8.56      35.70            44.36               50.07              Vertical            74               PK        Pass
  *4804       32.85       8.56      35.70              --                  --               Vertical            54               AV        Pass

   2440MHz Test Result
             Antenna     Cable     Amp.                              Emission
Frequency                                         Reading                               Polarization            Limit          Detector   Result
              Factor     Loss      Factor                             Level
   MHz        dB/m         dB       dB              dBuV             dBuV/m                                   dBμV/m
  *4880       32.99       8.64     35.70            46.52             52.45               Horizontal            74               PK        Pass
  *4880       32.99       8.64     35.70              --                --                Horizontal            54               AV        Pass
  *4880       32.99       8.64     35.70            43.83             49.76                Vertical             74               PK        Pass
  *4880       32.99       8.64     35.70              --                --                 Vertical             74               AV        Pass

   2480MHz Test Result
             Antenna     Cable     Amp.                              Emission
Frequency                                         Reading                               Polarization            Limit          Detector   Result
              Factor     Loss      Factor                             Level
   MHz        dB/m         dB       dB              dBuV             dBuV/m                                   dBμV/m
  *4960       33.13       8.72     35.70            45.67             51.82               Horizontal            74               PK        Pass
  *4960       33.13       8.72     35.70              --                --                Horizontal            54               AV        Pass
  *4960       33.13       8.72     35.70            44.90             51.05                Vertical             74               PK        Pass
  *4960       33.13       8.72     35.70              --                --                 Vertical             54               AV        Pass

   Remark:
         (1) QP Emission Level= Antenna Factor +Cable Loss + Reading
             PK Emission Level= Antenna Factor +Cable Loss - Amp. factor + Reading
             AV Emission Level= PK Emission Level+20log(dutycycle)
         (2) Data of measurement within this frequency range shown “-” in the table above means the reading of
             emissions are attenuated more than 20db below the permissible limits or the field strength is too small
             to be measured.
         (3) “*” means the emission(s) appear within the restrict bands shall follow the requirement of section
             15.205.




   EMC_SZ_FR_21.00 FCC                      TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                  Page 21 of 23
                               Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
   Release 2014-03-20                 Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




10 Test Equipment List
                                              List of Test Instruments

                                                                                                                CAL. DUE
           DESCRIPTION            MANUFACTURER                 MODEL NO.              SERIAL NO.
                                                                                                                 DATE
           Test Receiver          Rohde & Schwarz                 ESHS10               838693/001               Nov.04, 15
             L.I.S.N.#1           Rohde & Schwarz                 ESH2-Z5              834066/011               Nov.04, 15
             L.I.S.N.#3                  Kyoritsu               KNW-242C                8-1920-1                May.07, 15
             RF Cable                     3D-2W                    Fujikura          LISN Cable 1#              May.07, 15
  CE
           Coaxial Switch                MP59B                     Anritsu               M55367                 May.07, 15
           Passive Probe                ESH2-Z3                   Rohde &
                                                                                      299.7810.52               May.07, 15
                                                                  Schwarz
            Pulse Limiter               ESH3-Z2                   Rohde &
                                                                                         100341                 May.07, 15
                                                                  Schwarz
   C         Spectrum                     Agilent                  E4446A             US44300459                May.08, 15
  RE       Test Receiver
                                  Rohde & Schwarz                 ESVS10               834468/011               May.07, 15
  <1          <1GHz
  GHz     Amplifier < 1 GHz                 HP                      8447D             2648A04738                May.07, 15
             HF Cable                 Hubersuhne                Sucoflex104              Room 2                 May.08, 15
           Bilog Antenna                Schaffner                CBL6111C                  2598                 Oct.25, 15
  RE     Spectrum > 1GHz                  Agilent                  E4446A             US44300459                May.08, 15
  >1
  GHz      Horn Antenna                   EMCO                       3115              9607-4877                Jun. 24, 15
            Amp > 1 Ghz                     HP                      8449B             3008A08495                May.08, 15
             HF Cable                 Hubersuhne                Sucoflex104               Room1                 May.08, 15

C - Conducted RF tests
   Conducted peak output power
   6dB bandwidth
   20dB bandwidth and 99% Occupied Bandwidth
   Carrier frequency separation
   Number of hopping frequencies
   Dwell Time
   Power spectral density
   Spurious RF conducted emissions
   Band edge




EMC_SZ_FR_21.00 FCC                      TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 22 of 23
                            Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                 Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.14.004.01




11 System Measurement Uncertainty
For a 95% confidence level, the measurement expanded uncertainties for defined systems, in
accordance with the recommendations of ISO 17025 were:

                                    System Measurement Uncertainty
                              Items                                         Extended Uncertainty
                                                                            4.32dB (30MHz-1GHz)
                   Radiated spurious emission
                                                                            2.27dB (1GHz -25GHz)
                  Conducted spurious emission                               2.10dB(30MHz-25GHz)
                        Bandwidth test                                              1*10-9
                     Conducted emission                                             2.4dB




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 23 of 23
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299



Document Created: 2015-01-20 03:33:14
Document Modified: 2015-01-20 03:33:14

© 2024 FCC.report