conversion factor

FCC ID: O25PJ-400NW

RF Exposure Info

Download: PDF
FCCID_405601

Software Reference Guide                                                                       Page 16 of 69



 2.6.1 E-field probe (3D)

 E-field probes have additional tabs for sensitivity and conversion factors. For general probe parameters
 see chapter 2.6 (General tab, Mechanical tab, Surface detection tab or Offset tab).

                                                     For detailed evaluations of the sensitivity factors
                                                     (Sensitivity, Diode Compression Point), see AN:
                                                     Data Storage and Evaluation. Source Impedance is
                                                     for later use and UM: 4.5 Field measurements.

                                                     This data corresponds to the sensitivity factors on
                                                     your probe calibration sheet and cannot be modified
                                                     in user or administrator modes.




 The conversion factor tab is for special calibrations in media:

                                                     For each medium and frequency range (for example
                                                     brain simulating liquid around 900 MHz) a conversion
                                                     factor for the probe can be defined (see figure on the
                                                     left side). If you have checked the box select
                                                     parameter automatically, the first conversion factor
                                                     which fits the solution and media range of your actual
                                                     setup is used for the evaluations. If no conversion
                                                     factor for this solution range can be found, a warning
                                                     will appear. If you have not checked the box select
                                                     parameter automatically, the highlighted set of
                                                     conversion factors will be used if the frequency of the
                                                     device and the medium is within the solution range.




 You can add your own new conversion factors for different frequencies or solution parameters by
 pressing the New button or delete a conversion factor set by pressing the Delete button. Pressing the
 Options button will display the conversion factor set.




Schmid & Partner Engineering AG, August 99