04T2458 SAR Report

FCC ID: NM8VIVIDA

RF Exposure Info

Download: PDF
FCCID_463586

                in accordance with the requirements of
FCC Report and Order: ET Docket 93-62, and OET Bulletin 65 Supplement C

                                  for

            Dual Band (Cellular & PCS Band) Phone

                          MODEL: ST10B

                        FCC ID: NM8VIVIDA

                          January 29, 2004

                      REPORT NO: 04T2458-2

                             Prepared for

                    High Tech Computer Corp
                         23, Hsin-Hua Rd
                      Taoyuan, 330, Taiwan


                             Prepared by

            COMPLIANCE CERTIFICATION SERVICES
                  561F MONTEREY ROAD,
               MORGAN HILL, CA 95037, USA
                    TEL: (408) 463-0885


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


         CERTIFICATE OF COMPLIANCE (SAR EVALUATION)
                                        DATES OF TEST: January 29, 2004

 APPLICANT:                              High Tech Computer Corp
                                         23, Hsin-Hua Rd
                                         Taoyuan, 330, Taiwan
 MODEL:                                  ST10B
 FCC ID:                                 NM8VIVIDA
 DEVICE CATEGORY:                        PORTABLE DEVICES
 EXPOSURE CATEGORY:                      GENERAL POPULATION/UNCONTROLLED EXPOSURE

 Test Sample is a:                    Production unit

 Multiple Access Scheme: CDMA
 Tx Frequency:                        Cellular band: 824 MHz to 849 MHz
                                      PCS band: 1850 MHz to 1910 MHz

 Max. O/P Power:                      Cellular band: 23.90 dBm
 (Conducted/Average)                  PCS band: 24.75 dBm

 Max. SAR (1g):                       Cellular band:
                                      0.550 mW/g (Right head touched position)
                                      0.542 mW/g (Body worn)
                                      PCS band:
                                      1.35 mW/g (Right head tilted position)
                                      0.662 mW/g (Body worn)

 FCC Rule Part(s):                    Cellular band: 22H
                                      PCS band: 24E
 This wireless portable device has been shown to be capable of compliance for localized specific
 absorption rate (SAR) for uncontrolled environment/general population exposure limits specified in
 ANSI/IEEE Std. C95.1-1992 and had been tested in accordance with the measurement procedures
 specified in FCC OET 65 Supplement C (released on 6/29/2001 see Test Report).
 I attest to the accuracy of data. All measurements reported herein were performed by me or were
 made under my supervision and are correct to the best of my knowledge and belief. I assume full
 responsibility for the completeness of these measurements and vouch for the qualifications of all
 persons taking them.




   Steve Cheng
   EMC Engineering Manager




COMPLIANCE CERTIFICATION SERVICES                                                                                 Page: 2 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                                       DATE: January 29, 2004                                                FCC ID: NM8VIVIDA


                                                          TABLE OF CONTENTS


1.      EQUIPMENT UNDER TEST (EUT) DESCRIPTION ............................................................. 4
2.      REQUIREMENTS FOR COMPLIANCE TESTING DEFINED BY THE FCC............................... 4
3.      DOSIMETRIC ASSESSMENT SETUP ............................................................................... 4
     3.1.    MEASUREMENT SYSTEM DIAGRAM ..................................................................................................................5
     3.2.    SYSTEM COMPONENTS ...................................................................................................................................6
4.      EVALUATION PROCEDURES ......................................................................................... 8
5.      MEASUREMENT UNCERTAINTY................................................................................... 12
6.      EXPOSURE LIMIT ...................................................................................................... 13
7.      EUT ARRANGEMENT ................................................................................................ 14
8.      MEASUREMENT RESULTS .......................................................................................... 17
     8.1.    SIMULATING LIQUIDS PARAMETER CHECK .....................................................................................................17
     8.2.    SYSTEM PERFORMANCE CHECK ...................................................................................................................19
     8.3.    EUT TUNE-UP PROCEDURES ........................................................................................................................21
     8.4.    SAR MEASUREMENT RESULTS .....................................................................................................................22
9.      EUT PHOTOS ........................................................................................................... 27
10. EQUIPMENT LIST & CALIBRATION STATUS.................................................................. 30
11. REFERENCES ........................................................................................................... 31
12. ATTACHMENTS ......................................................................................................... 32




COMPLIANCE CERTIFICATION SERVICES                                                                                                             Page: 3 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


1. EQUIPMENT UNDER TEST (EUT) DESCRIPTION
     The following is the information provided by the applicant.
    Type of EUT:              Dual Band (Cellular & PCS Band) Phone
    Type/Model No.:           ST10B
    Modulation Type:          CDMA
     TX Frequency:            Cellular band: 824 MHz to 849 MHz
                              PCS band: 1850 MHz to 1910 MHz
    Duty Cycle:               100%
    Battery:                  Only one model with EUT
2. REQUIREMENTS FOR COMPLIANCE TESTING DEFINED BY THE FCC
     The US Federal Communications Commission has released the report and order “Guidelines for
     Evaluating the Environmental Effects of RF Radiation", ET Docket No. 93-62 in August 1996. The
     order requires routine SAR evaluation prior to equipment authorization of portable transmitter devices,
     including portable telephones. For consumer products, the applicable limit is 1.6 mW/g for an
     uncontrolled environment and 8.0 mW/g for an occupational/controlled environment as recommended
     by the ANSI/IEEE standard C95.1-1992. According to the Supplement C of OET Bulletin 65
     “Evaluating Compliance with FCC Guide-lines for Human Exposure to Radio frequency
     Electromagnetic Fields", released on Jun 29, 2001 by the FCC, the device should be evaluated at
     maximum output power (radiated from the antenna) under “worst-case” conditions for normal or
     intended use, incorporating normal antenna operating positions, device peak performance
     frequencies and positions for maximum RF energy coupling.
3. DOSIMETRIC ASSESSMENT SETUP
     These measurements were performed with the automated near-field scanning system DASY3 from
     Schmid & Partner Engineering AG (SPEAG). The system is based on a high precision robot (working
     range greater than 0.9 m), which positions the probes with a positional repeatability of better than ±
     0.02 mm. Special E- and H-field probes have been developed for measurements close to material
     discontinuity, the sensors of which are directly loaded with a Schottky diode and connected via highly
     resistive lines to the data acquisition unit. The SAR measurements were conducted with the
     dosimetric probe ET3DV6, SN: 1577 (manufactured by SPEAG), designed in the classical triangular
     configuration and optimized for dosimetric evaluation. The probe has been calibrated according to the
     procedure described in [7] with accuracy of better than ±10%. The spherical isotropy was evaluated
     with the procedure described in [8] and found to be better than ±0.25 dB. The phantom used was the
     SAM Twin Phantom as described in FCC supplement C, IEE P1528 and CENELEC EN50361.




COMPLIANCE CERTIFICATION SERVICES                                                                                 Page: 4 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


3.1. MEASUREMENT SYSTEM DIAGRAM




     The DASY4 system for performing compliance tests consists of the following items:
     •    A standard high precision 6-axis robot (St¨aubli RX family) with controller, teach pendant and
          software. An arm extension for accommodating the data acquisition electronics (DAE).
     •    A dosimetric probe, i.e., an isotropic E-field probe optimized and calibrated for usage in tissue
          simulating liquid. The probe is equipped with an optical surface detector system.
     •    A data acquisition electronics (DAE) which performs the signal amplification, signal multiplexing,
          AD-conversion, offset measurements, mechanical surface detection, collision detection, etc. The
          unit is battery powered with standard or rechargeable batteries. The signal is optically transmitted
          to the EOC.
     •    The Electro-optical converter (EOC) performs the conversion between optical and electrical of the
          signals for the digital communication to the DAE and for the analog signal from the optical
          surface detection. The EOC is connected to the measurement server.
     •    The function of the measurement server is to perform the time critical tasks such as signal
          filtering, control of the robot operation and fast movement interrupts.
     •    A probe alignment unit which improves the (absolute) accuracy of the probe positioning.
     •    A computer operating Windows 2000 or Windows XP.
     •    DASY4 software.
     •    Remote control with teach pendant and additional circuitry for robot safety such as warning lamps,
          etc.
     •    The SAM twin phantom enabling testing left-hand and right-hand usage.
     •    The device holder for handheld mobile phones.
     •    Tissue simulating liquid mixed according to the given recipes.
     •    Validation dipole kits allowing validating the proper functioning of the system.


COMPLIANCE CERTIFICATION SERVICES                                                                                 Page: 5 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


3.2. SYSTEM COMPONENTS
     DASY4 MEASUREMENT SERVER
                                          The DASY4 measurement server is based on a PC/104 CPU
                                          board with a 166MHz low-power Pentium, 32MB chip disk and
                                          64MB RAM. The necessary circuits for communication with
                                          either the DAE3 electronic box as well as the 16-bit AD-converter
                                          system for optical detection and digital I/O interface are contained
                                          on the DASY4 I/O-board, which is directly connected to the
                                          PC/104 bus of the CPU board.
                                          The measurement server performs all real-time data evaluation
     for field measurements and surface detection, controls robot movements and handles safety
     operation. The PC-operating system cannot interfere with these time critical processes. All
     connections are supervised by a watchdog, and disconnection of any of the cables to the
     measurement server will automatically disarm the robot and disable all program-controlled robot
     movements. Furthermore, the measurement server is equipped with two expansion slots which are
     reserved for future applications. Please note that the expansion slots do not have a standardized
     pinout and therefore only the expansion cards provided by SPEAG can be inserted. Expansion cards
     from any other supplier could seriously damage the measurement server. Calibration: No calibration
     required.
     DATA ACQUISITION ELECTRONICS (DAE)
     The data acquisition electronics (DAE3) consists of a highly sensitive
     electrometer grade preamplifier with auto-zeroing, a channel and gain-switching
     multiplexer, a fast 16 bit AD converter and a command decoder and control logic
     unit. Transmission to the measurement server is accomplished through an
     optical downlink for data and status information as well as an optical uplink for
     commands and the clock. The mechanical probe mounting device includes two
     different sensor systems for frontal and sideways probe contacts. They are used for mechanical
     surface detection and probe collision detection. The input impedance of the DAE3 box is 200MOhm;
     the inputs are symmetrical and floating. Common mode rejection is above 80 dB.
     ES3DV2 ISOTROPIC E-FIELD PROBE FOR DOSIMETRIC MEASUREMENTS
     Construction:         Symmetrical design with triangular core Interleaved sensors Built-in shielding against static
                           charges PEEK enclosure material (resistant to organic solvents, e.g., glycolether)
     Calibration:          Basic Broad Band Calibration in air: 10-2500 MHz. Conversion Factors (CF) for HSL 900
                           and HSL 1800 CF-Calibration for other liquids and frequencies upon request.
     Frequency:            10 MHz to > 6 GHz; Linearity: ± 0.2 dB
     Directivity:          ± 0.2 dB in HSL (rotation around probe axis);
                           ± 0.3 dB in tissue material (rotation normal to probe axis)
     Dynamic Range:        5 µW/g to > 100 mW/g; Linearity: ± 0.2 dB
     Dimensions:           Overall length: 330 mm (Tip: 20 mm)
                           Tip diameter: 3.9 mm (Body: 12 mm)
                           Distance from probe tip to dipole centers: 2.7 mm
     Application:          General dosimetry up to 6 GHz
                           Dosimetry in strong gradient fields
                           Compliance tests of mobile phones




                                                                           Interior of probe          Isotropic E-Field Probe




COMPLIANCE CERTIFICATION SERVICES                                                                                 Page: 6 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA



     SAM PHANTOM (V4.0)
     Construction:        The shell corresponds to the
                          specifications of the Specific
                          Anthropomorphic Mannequin (SAM)
                          phantom defined in IEEE 1528-200X,
                          CENELEC 50361 and IEC 62209. It
                          enables the dosimetric evaluation of left
                          and right hand phone usage as well as
                          body mounted usage at the flat phantom
                          region. A cover prevents evaporation of
                          the liquid. Reference markings on the
                          phantom allow the complete setup of all
                          predefined phantom positions and
                          measurement grids by manually teaching
                          three points with the robot.
     Shell Thickness:     2 ±0.2 mm
     Filling Volume:      Approx. 25 liters
     Dimensions:          Height: 810mm; Length: 1000mm; Width:
                          500mm

     DEVICE HOLDER FOR SAM TWIN PHANTOM
     Construction:        In combination with the Twin SAM Phantom V4.0 or Twin SAM, the
                          Mounting Device (made from POM) enables the rotation of the
                          mounted transmitter in spherical coordinates, whereby the rotation
                          point is the ear opening. The devices can be easily and accurately
                          positioned according to IEC, IEEE, CENELEC, FCC or other
                          specifications. The device holder can be locked at different
                          phantom locations (left head, right head, flat phantom).




     SYSTEM VALIDATION KITS
     Construction:     Symmetrical dipole with l/4 balun Enables measurement of
                       feedpoint impedance with NWA Matched for use near flat
                       phantoms filled with brain simulating solutions Includes distance
                       holder and tripod adaptor.
     Frequency:        450, 900, 1800, 2450, 5800 MHz
     Return loss:      > 20 dB at specified validation position
     Power capability: > 100 W (f < 1GHz); > 40 W (f > 1GHz)
     Dimensions:       450V2: dipole length: 270 mm; overall height: 330 mm
                       D900V2: dipole length: 149 mm; overall height: 330 mm
                       D1800V2: dipole length: 72 mm; overall height: 300 mm
                       D2450V2: dipole length: 51.5 mm; overall height: 300 mm
                       D5GHzV2: dipole length: 25.5 mm; overall height: 290 mm




COMPLIANCE CERTIFICATION SERVICES                                                                                 Page: 7 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                                         DATE: January 29, 2004                           FCC ID: NM8VIVIDA



4. EVALUATION PROCEDURES
     DATA EVALUATION
The DASY4 post processing software (SEMCAD) automatically executes the following procedures to
calculate the field units from the microvolt readings at the probe connector. The parameters used in the
evaluation are stored in the configuration modules of the software:
             Probe parameters:  - Sensitivity                                          Normi, ai0, ai1, ai2
                                - Conversion factor                                    ConvFi
                                - Diode compression point                              dcpi
             Device parameters: - Frequency                                            f
                                - Crest factor                                         cf
             Media parameters:               - Conductivity                            σ
                                             - Density                                 ρ
These parameters must be set correctly in the software. They can be found in the component documents
or be imported into the software from the configuration files issued for the DASY components. In the
direct measuring mode of the multi-meter option, the parameters of the actual system setup are used. In
the scan visualization and export modes, the parameters stored in the corresponding document files are
used.
The first step of the evaluation is a linearization of the filtered input signal to account for the compression
characteristics of the detector diode. The compensation depends on the input signal, the diode type and
the DC-transmission factor from the diode to the evaluation electronics. If the exciting field is pulsed, the
crest factor of the signal must be known to correctly compensate for peak power. The formula for each
channel can be given as:
                                             cf
                     V =U +U            ⋅
                                    2
                        i     i     i
                                            dcp i

             with    Vi       = Compensated signal of channel i                        (i = x, y, z)
                     Ui       = Input signal of channel i                              (i = x, y, z)
                     cf       = Crest factor of exciting field                         (DASY parameter)
                     dcpi     = Diode compression point                                (DASY parameter)

From the compensated input signals the primary field data for each channel can be evaluated:
             E-field probes:                                      V
                                             E           =             i
                                                 i
                                                             Norm • ConvF
                                                                  i




                                                                  a +a f +a f
                                                                                        2

             H-field probes:                                     i10       i11   i12
                                             H       i
                                                         =   Vi ⋅
                                                                       f
             with    Vi    = Compensated signal of channel i (i = x, y, z)
                     Normi = Sensor sensitivity of channel i (i = x, y, z)
                             µV/(V/m)2 for E0field Probes
                     ConvF = Sensitivity enhancement in solution
                     aij   = Sensor sensitivity factors for H-field probes
                     f     = Carrier frequency (GHz)
                     Ei    = Electric field strength of channel i in V/m
                     Hi    = Magnetic field strength of channel i in A/m

COMPLIANCE CERTIFICATION SERVICES                                                                                   Page: 8 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                                              DATE: January 29, 2004                FCC ID: NM8VIVIDA


The RSS value of the field components gives the total field strength (Hermitian magnitude):

                                =    E +E +E
                                      2              2        2
                      E   tot         x              y        z


The primary field data are used to calculate the derived field units.
                                                         σ
                      SAR = E                   ⋅
                                          2
                                          tot
                                                    ρ ⋅1000
             with    SAR            = local specific absorption rate in mW/g
                     Etot           = total field strength in V/m
                     σ              = conductivity in [mho/m] or [Siemens/m]
                     ρ              = equivalent tissue density in g/cm3
Note that the density is normally set to 1 (or 1.06), to account for actual brain density rather than the
density of the simulation liquid.
The power flow density is calculated assuming the excitation field as a free space field.
                                      2
                                     E                   or       P         = H tot ⋅ 37.7
                                                                                 2

                      P pwe =         tot
                                                                      pwe
                                    3770
             with    Ppwe = Equivalent power density of a plane wave in mW/cm2
                     Etot    = total electric field strength in V/m
          Htot      = total magnetic field strength in A/m




COMPLIANCE CERTIFICATION SERVICES                                                                                 Page: 9 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA



     SAR SYSTEM MEASUREMENT PROCEDURES
     The procedure for assessing the peak spatial-average SAR value consists of the following steps:
     • Power Reference Measurement
     The reference and drift jobs are useful jobs for monitoring the power drift of the device under test in
     the batch process. Both jobs measure the field at a specified reference position, at a selectable
     distance from the phantom surface. The reference position can be either the selected section’s grid
     reference point or a user point in this section. The reference job projects the selected point onto the
     phantom surface, orients the probe perpendicularly to the surface, and approaches the surface using
     the selected detection method.
     • Area Scan
     The area scan is used as a fast scan in two dimensions to find the area of high field values, before
     doing a finer measurement around the hot spot. The sophisticated interpolation routines implemented
     in DASY4 software can find the maximum locations even in relatively coarse grids. This grid is
     anchored at the grid reference point of the selected section in the phantom. When the area scan’s
     property sheet is brought-up, grid settings can be edited by a user. When an area scan has
     measured all reachable points, it computes the field maximum found in the scanned area, within a
     range of the global maximum. If only one Zoom Scan follows the Area Scan, then only the absolute
     maximum will be taken as reference. For cases where multiple maximums are detected, the number
     of Zoom Scans has to be increased accordingly.
     • Zoom Scan
     Zoom scans are used to assess the peak spatial SAR values within a cubic averaging volume
     containing 1 g and 10 g of simulated tissue. The default zoom scan measures 5 x 5 x 7 points within
     a cube whose base faces are centered around the maximum found in a preceding area scan job
     within the same procedure. If the preceding Area Scan job indicates more then one maximum, the
     number of Zoom Scans has to be enlarged accordingly. For dosimetric application, it is necessary to
     assess the peak spatial SAR value averaged over a volume. For this purpose, fine resolution volume
     scans need to be performed at the peak SAR location(s) determined during the Area Scan.
     • Power Drift measurement
     The drift job measures the field at the same location as the most recent reference job within the same
     procedure, and with the same settings. The drift measurement gives the field difference in dB from
     the reading conducted within the last reference measurement. Several drift measurements are
     possible for one reference measurement. This allows a user to monitor the power drift of the device
     under test within a batch process. In the properties of the Drift job, the user can specify a limit for the
     drift and have DASY4 software stop the measurements if this limit is exceeded.
     • Z-Scan
     The Z Scan job measures points along a vertical straight line. The line runs along the Z-axis of a one-
     dimensional grid. A user can anchor the grid to the current probe location. As with any other grids,
     the local Z-axis of the anchor location establishes the Z-axis of the grid.




COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 10 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


     SPATIAL PEAK SAR EVALUATION
     The procedure for spatial peak SAR evaluation has been implemented according to the IEEE1529
     standard. It can be conducted for 1 g and 10 g.
     The DASY4 system allows evaluations that combine measured data and robot positions, such as:
        • maximum search
        • extrapolation
        • boundary correction
        • peak search for averaged SAR
     During a maximum search, global and local maximum searches are automatically performed in 2-D
     after each Area Scan measurement with at least 6 measurement points. It is based on the evaluation
     of the local SAR gradient calculated by the Quadratic Shepard’s method. The algorithm will find the
     global maximum and all local maxima within -2 dB of the global maxima for all SAR distributions.
     Extrapolation
     Extrapolation routines are used to obtain SAR values between the lowest measurement points and
     the inner phantom surface. The extrapolation distance is determined by the surface detection
     distance and the probe sensor offset. Several measurements at different distances are necessary for
     the extrapolation.
     Extrapolation routines require at least 10 measurement points in 3-D space. They are used in the
     Cube Scan to obtain SAR values between the lowest measurement points and the inner phantom
     surface. The routine uses the modified Quadratic Shepard’s method for extrapolation. For a grid
     using 5x5x7 measurement points with 5mm resolution amounting to 343 measurement points, the
     uncertainty of the extrapolation routines is less than 1% for 1 g and 10 g cubes.
     Boundary effect
     For measurements in the immediate vicinity of a phantom surface, the field coupling effects between
     the probe and the boundary influence the probe characteristics. Boundary effect errors of different
     dosimetric probe types have been analyzed by measurements and using a numerical probe model.
     As expected, both methods showed an enhanced sensitivity in the immediate vicinity of the boundary.
     The effect strongly depends on the probe dimensions and disappears with increasing distance from
     the boundary. The sensitivity can be approximately given as:




     Since the decay of the boundary effect dominates for small probes (a<<λ), the cos-term can be
     omitted. Factors Sb (parameter Alpha in the DASY4 software) and a (parameter Delta in the DASY4
     software) are assessed during probe calibration and used for numerical compensation of the
     boundary effect. Several simulations and measurements have confirmed that the compensation is
     valid for different field and boundary configurations.
     This simple compensation procedure can largely reduce the probe uncertainty near boundaries. It
     works well as long as:
        • the boundary curvature is small
        • the probe axis is angled less than 30_ to the boundary normal
        • the distance between probe and boundary is larger than 25% of the probe diameter
        • the probe is symmetric (all sensors have the same offset from the probe tip)
     Since all of these requirements are fulfilled in a DASY4 system, the correction of the probe boundary
     effect in the vicinity of the phantom surface is performed in a fully automated manner via the
     measurement data extraction during postprocessing.




COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 11 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


5. MEASUREMENT UNCERTAINTY
                                   UNCERTAINTY BUDGE ACCORDING TO IEEE P1528

                                        Uncertainty     Prob.                 (ci)       (ci)       Std.         Std.         (vi)
            Error Description                                      Div.
                                         Value [%]      Dist.                 1g         10g      Unc.(1g)     Unc. (10g)     veff

      Measurement System

      Probe Calibration                     ±4.8          N          1         1          1        ±4.8%         ±4.8%         ∞
      Axial Isotropy                        ±4.7          R           3       0.7        0.7       ±1.9%         ±1.9%         ∞
      Hemispherical Isotropy                ±9.6          R           3       0.7        0.7       ±3.9%         ±3.9%         ∞
      Boundary Effects                      ±1.0          R           3        1          1        ±0.6%         ±0.6%         ∞
      Linearity                             ±4.7          R           3        1          1        ±2.7%         ±2.7%         ∞
      System Detection Limits               ±1.0          R           3        1          1        ±0.6%         ±0.6%         ∞
      Readout Electronics                   ±1.0          N           3        1          1        ±1.0%         ±1.0%         ∞
      Response Time                         ±0.8          R           3        1          1        ±0.5%         ±0.5%         ∞
      Integration Time                      ±2.6          R           3        1          1        ±1.5%         ±1.5%         ∞
      RF Ambient Condition                 ±1.59          R           3        1          1        ±0.9%         ±0.9%         ∞
      Probe Positioner                      ±1.6          R           3        1          1        ±0.2%         ±0.2%         ∞
      Probe Positioning                     ±2.9          R           3        1          1        ±1.7%         ±1.7%         ∞
      Max. SAR Eval.                        ±1.0          R           3        1          1        ±0.6%         ±0.6%         ∞
      Test sample Related

      Device Positioning                    ±1.1          N          1         1          1        ±1.1%         ±1.1%        145

      Device Holder                         ±3.6          N          1         1          1        ±3.6%         ±3.6%         5

      Power Drift                           ±5.0          R           3        1          1        ±2.9%         ±2.9%         ∞
      Phantom and Setup

      Phantom Uncertainty                   ±4.0          R           3        1          1        ±2.3%         ±2.3%         ∞
      Liquid Conductivity (target)          ±5.0          R           3      0.64       0.43       ±1.8%         ±1.2%         ∞
      Liquid Conductivity (meas.)           ±2.5          N          1       0.64       0.43       ±1.6%         ±1.1%         ∞
      Liquid Peermittivity (target)         ±5.0          R           3       0.6       0.49       ±1.7%         ±1.4%         ∞
      Liquid Permittivity (meas.)           ±2.5          N          1        0.6       0.49       ±1.5%         ±1.2%         ∞
      Combined Std. Uncertainty                                                                    ±9.8%         ±9.6%       330
      Expanded STD Uncertainty                                                                    ±19.6%       ±19.2%
     Table: Worst-case uncertainty for DASY4 assessed according to IEEE P1528.
     The budge is valid for the frequency range 300MHz – 3GHz and represents a worst-case analysis.




COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 12 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


6. EXPOSURE LIMIT
     (A) Limits for Occupational/Controlled Exposure (W/kg)


                  Whole-Body                        Partial-Body                      Hands, Wrists, Feet and Ankles

                       0.4                                8.0                                          20.0


     (B) Limits for General Population/Uncontrolled Exposure (W/kg)


                  Whole-Body                        Partial-Body                      Hands, Wrists, Feet and Ankles

                       0.08                               1.6                                           4.0


        NOTE 1: See Section 1 for discussion of exposure categories.


        NOTE 2: Whole-Body SAR is averaged over the entire body, partial-body SAR is averaged over
                any 1 gram of tissue defined as a tissue volume in the shape of a cube. SAR for hands,
                wrists, feet and ankles is averaged over any 10 grams of tissue defined as a tissue
                volume in the shape of a cube.

        NOTE 3: At frequencies above 6.0 GHz, SAR limits are not applicable and MPE limits for power
                density should be applied at 5 cm or more from the transmitting device.

        NOTE 4: The time averaging criteria for field strength and power density do not apply to general
                population SAR limit of 47 CFR §2.1093

                                                    NOTE
                                 GENERAL POPULATION/UNCONTROLLED EXPOSURE
                                             PARTIAL BODY LIMIT
                                                  1.6 mW/g




COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 13 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


7. EUT ARRANGEMENT
     Please refer to IEEE P1528 illustration below.
     7.1 ANTHROPOMORPHIC HEAD PHANTOM
     Figure 7-1a shows the front, back and side views of SAM. The point “M” is the reference point for the
     center of mouth, “LE” is the left ear reference point (ERP), and “RE” is the right ERP. The ERPs are
     15 mm posterior to the entrance to ear canal (EEC) along the B-M line (Back-Mouth), as shown in
     Figure 7-1b. The plane passing through the two ear reference points and M is defined as the
     Reference Plane. The line N-F (Neck-Front) perpendicular to the reference plane and passing
     through the RE (or LE) is called the Reference Pivoting Line (see Figure 7-1c). Line B-M is
     perpendicular to the N-F line. Both N-F and B-M lines should be marked on the external phantom
     shell to facilitate handset positioning. Posterior to the N-F line, the thickness of the phantom shell with
     the shape of an ear is a flat surface 6 mm thick at the ERPs. Anterior to the N-F line, the ear is
     truncated as illustrated in Figure 7-1b. The ear truncation is introduced to avoid the handset from
     touching the ear lobe, which can cause unstable handset positioning at the cheek.




                                                            Figure 7-1a
                                    Front, back and side view of SAM (model for the phantom shell)




                        Figure 7-1b                                                      Figure 7-1c
     Close up side view of phantom showing the ear region     Side view of the phantom showing relevant markings and the 7 cross
                                                              sectional plane locations




COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 14 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


     7.2 DEFINITION OF THE “CHEEK/TOUCH” POSITION
     The “cheek” or “touch” position is defined as follows:
     a. Ready the handset for talk operation, if necessary. For example, for handsets with a cover piece,
        open the cover. (If the handset can also be used with the cover closed both configurations must
        be tested.)
     b. Define two imaginary lines on the handset: the vertical centerline and the horizontal line. The
        vertical centerline passes through two points on the front side of the handset: the midpoint of the
        width wt of the handset at the level of the acoustic output (point A on Figures 7-2a and 7-2b), and
        the midpoint of the width wb of the bottom of the handset (point B). The horizontal line is
        perpendicular to the vertical centerline and passes through the center of the acoustic output (see
        Figure 7-2a). The two lines intersect at point A. Note that for many handsets, point A coincides
        with the center of the acoustic output. However, the acoustic output may be located elsewhere on
        the horizontal line. Also note that the vertical centerline is not necessarily parallel to the front face
        of the handset (see Figure 7-2b), especially for clamshell handsets, handsets with flip pieces, and
        other irregularly-shaped handsets.
     c. Position the handset close to the surface of the phantom such that point A is on the (virtual)
        extension of the line passing through points RE and LE on the phantom (see Figure 7-2c), such
        that the plane defined by the vertical center line and the horizontal line of the handset is
        approximately parallel to the sagittal plane of the phantom.
     d. Translate the handset towards the phantom along the line passing through RE and LE until the
        handset touches the pinna.
     e. e) While maintaining the handset in this plane, rotate it around the LE-RE line until the vertical
        centerline is in the plane normal to MB-NF including the line MB (called the reference plane).
     f. Rotate the handset around the vertical centerline until the handset (horizontal line) is symmetrical
        with respect to the line NF.
     g. While maintaining the vertical centerline in the reference plane, keeping point A on the line
        passing through RE and LE and maintaining the handset contact with the pinna, rotate the
        handset about the line NF until any point on the handset is in contact with a phantom point below
        the pinna (cheek). See Figure 7-2c. The physical angles of rotation should be noted.




                                                    Figure 7.2c
           Phone “cheek” or “touch” position. The reference points for the right ear (RE), left ear LE) and
           mouth (M), which define the reference plane for handset positioning, are indicated.




COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 15 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA




                     Figure 7.2a                                                                    Figure 7.2b


     7.3 DEFINITION OF THE “TILTED” POSITION
     The “tilted” position is defined as follows:
     a. Repeat steps (a) – (g) of 7.2 to place the device in the “cheek position.”
     b. While maintaining the orientation of the handset move the handset away from the pinna along the
        line passing through RE and LE in order to enable a rotation of the handset by 15 degrees.
     c. Rotate the handset around the horizontal line by 15 degrees.
     d. While maintaining the orientation of the handset, move the handset towards the phantom on a
        line passing through RE and LE until any part of the handset touches the ear. The tilted position
        is obtained when the contact is on the pinna. If the contact is at any location other than the pinna
        (e.g., the antenna with the back of the phantom head), the angle of the handset should be
        reduced. In this case, the tilted position is obtained if any part of the handset is in contact with the
        pinna as well as a second part of the handset is contact with the phantom (e.g., the antenna with
        the back of the head).




                                                       Figure 7-3
            Phone “tilted” position. The reference points for the right ear (RE), left ear (LE) and mouth (M),
            which define the reference plane for handset positioning, are indicated.


COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 16 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


8. MEASUREMENT RESULTS
     8.1. SIMULATING LIQUIDS PARAMETER CHECK
     The simulating liquids should be checked at the beginning of a series of SAR measurements to
     determine of the dielectric parameters are within the tolerances of the specified target values. The
     relative permittivity and conductivity of the tissue material should be within ± 5% of the values given in
     the table below. 5% may not be easily achieved at certain frequencies. Under such circumstances,
     10% tolerance may be used until more precise tissue recipes are available.
     TISSUE DIELECTRIC PARAMETERS FOR HEAD AND BODY PHANTOMS
     The head tissue dielectric parameters recommended by the IEEE SCC-34/SC-2 in P1528 have been
     incorporated in the following table. These head parameters are derived from planar layer models
     simulating the highest expected SAR for the dielectric properties and tissue thickness variations in a
     human head. Other head and body tissue parameters that have not been specified in P1528 are
     derived from the tissue dielectric parameters computed from the 4-Cole-Cole equations and
     extrapolated according to the head parameters specified in P1528.
                                                            Head                                     Body
         Target Frequency (MHz)
                                                   εr                σ (S/m)                 εr                σ (S/m)
                    150                           52.3                0.76                  61.9                0.80
                    300                           45.3                0.87                  58.2                0.92
                    450                           43.5                0.87                  56.7                0.94
                    835                           41.5                0.90                  55.2                0.97
                    900                           41.5                0.97                  55.0                1.05
                    915                           41.5                0.98                  55.0                1.06
                   1450                           40.5                1.20                  54.0                1.30
                   1610                           40.3                1.29                  53.8                1.40
                1800 – 2000                       40.0                1.40                  53.3                1.52
                   2450                           39.2                1.80                  52.7                1.95
                   3000                           38.5                2.40                  52.0                2.73
                   5800                           35.3                5.27                  48.2                6.00
                           (εr = relative permittivity, σ = conductivity and ρ = 1000 kg/m3)
     TYPICAL COMPOSITION OF INGREDIENTS FOR LIQUID TISSUE PHANTOMS
     The following tissue formulations are provided for reference only as some of the parameters have not
     been thoroughly verified. The composition of ingredients may be modified accordingly to achieve the
     desired target tissue parameters required for routine SAR evaluation.
          Ingredients                                               Frequency (MHz)
        (% by weight)                450                 835               915                 1900               2450
         Tissue Type           Head     Body       Head     Body     Head     Body        Head    Body       Head    Body
             Water             38.56 51.16         41.45     52.4    41.05     56.0        54.9    40.4      62.7     73.2
          Salt (NaCl)           3.95     1.49      1.45      1.4      1.35     0.76        0.18     0.5       0.5     0.04
             Sugar             56.32 46.78          56.0     45.0     56.5    41.76        0.0     58.0       0.0      0.0
              HEC               0.98     0.52        1.0     1.0       1.0     1.21        0.0      1.0       0.0      0.0
          Bactericide           0.19     0.05       0.1      0.1       0.1     0.27        0.0      0.1       0.0      0.0
         Triton X-100            0.0      0.0        0.0     0.0       0.0      0.0        0.0      0.0      36.8      0.0
             DGBE               0.0       0.0       0.0      0.0       0.0      0.0       44.92     0.0       0.0     26.7
      Dielectric Constant      43.42     58.0      42.54     56.1     42.0     56.8       39.9     54.0      39.8     52.5
      Conductivity (S/m)        0.85     0.83       0.91     0.95      1.0     1.07       1.42     1.45      1.88     1.78
     Salt: 99+% Pure Sodium Chloride          Sugar: 98+% Pure Sucrose
                                +
     Water: De-ionized, 16 MΩ resistivity HEC: Hydroxyethyl Cellulose
     DGBE: 99+% Di(ethylene glycol) butyl ether, [2-(2-butoxyethoxy)ethanol]
     Triton X-100 (ultra pure): Polyethylene glycol mono [4-(1,1, 3, 3-tetramethylbutyl)phenyl]ether




COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 17 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA



     SIMULATING LIQUIDS PARAMETER CHECK RESULTS
     Ambient condition: Temperature: 23.0°C; Relative humidity: 40%                         Date: January 29, 2004
               Head Simulating Liquid
                                                      Parameters        Target      Measured      Deviation[%]      Limited[%]
        f (MHz)      Temp. [°C]      Depth (cm)
                                                        Permitivity:     41.5       42.0951            1.43             ±5
          835            21.5             15
                                                      Conductivity:       0.9        0.9194            2.16             ±5


     Ambient condition: Temperature: 23.0°C; Relative humidity: 40%                         Date: January 29, 2004
               Body Simulating Liquid
                                                      Parameters        Target      Measured      Deviation[%]      Limited[%]
        f (MHz)      Temp. [°C]      Depth (cm)
                                                        Permitivity:     55.2         55.68            0.87             ±5
          835            21.5             15
                                                     Conductivity:       0.97        0.9646           -0.56             ±5

     Ambient condition: Temperature: 23.5°C; Relative humidity: 40%                         Date: January 29, 2004
               Head Simulating Liquid
                                                      Parameters        Target      Measured      Deviation[%]      Limited[%]
        f (MHz)      Temp. [°C]      Depth (cm)
                                                        Permitivity:      40        40.0674            0.17             ±5
         1900             22              15
                                                     Conductivity:        1.4        1.4568            4.06             ±5

     Ambient condition: Temperature: 23.0°C; Relative humidity: 40%                       Date: January 29, 2004
               Body Simulating Liquid
                                                      Parameters        Target      Measured      Deviation[%]      Limited[%]
        f (MHz)      Temp. [°C]      Depth (cm)
                                                        Permitivity:     53.3         52.83           -0.88             ±5
         1900             22              15
                                                     Conductivity:       1.52         1.594            4.87             ±5




COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 18 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


     8.2. SYSTEM PERFORMANCE CHECK
     The system performance check is performed prior to any usage of the system in order to guarantee
     reproducible results. The system performance check verifies that the system operates within its
     specifications. The system performance check results are tabulated below. And also the
     corresponding SAR plot is attached as well in the SAR plots files.
     SYSTEM PERFORMANCE CHECK MEASUREMENT CONDITIONS
     •    The measurements were performed in the flat section of the SAM twin phantom filled with Head
          simulating liquid of the following parameters.
     •    The DASY4 system with an Isotropic E-Field Probe ES3DV2-SN: 3021 was used for the
          measurements.
     •    The dipole was mounted on the small tripod so that the dipole feed point was positioned below
          the center marking of the flat phantom section and the dipole was oriented parallel to the body
          axis (the long side of the phantom). The standard measuring distance was 10 mm (above 1 GHz)
          and 15 mm (below 1 GHz) from dipole center to the simulating liquid surface.
     •    The coarse grid with a grid spacing of 15 mm was aligned with the dipole.
     •    Special 5 x 5 x 7 fine cube was chosen for cube integration(dx=dy=7.5mm; dz=5mm).
     •    Distance between probe sensors and phantom surface was set to 4 mm.
     •    The dipole input power (forward power) was 250 mW±3%.
     •    The results are normalized to 1 W input power.
     REFERENCE SAR VALUES
     The system performance check is performed prior to any usage of the system in order to guarantee
     reproducible results. The system performance check verifies that the system operates within its
     specifications of ±10%. The system performance check results are tabulated below. And also the
     corresponding SAR plot is attached as well in the SAR plots files.
     IEEE P1528 Recommended Reference Value
         Frequency                                             Local SAR at surface             Local SAR at surface
                           1 g SAR          10 g SAR
           (MHz)                                                (Above feed point)          (y=2cm offset from feed point)
           300                3.0               2.0                      4.4                                2.1
           450                4.9               3.3                      7.2                                3.2
           835                9.5               6.2                     14.1                                4.9
           900               10.8               6.9                     16.4                                5.4
           1450              29.0              16.0                     50.2                               6.5
           1800              38.1              19.8                     69.5                                6.8
           1900              39.7              20.5                     72.1                               6.6
           2450              52.4              24.0                    104.2                                7.7
           3000              63.8              25.7                    140.2                                9.5




COMPLIANCE CERTIFICATION SERVICES                                                                                 Page: 19 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                                    DATE: January 29, 2004                                      FCC ID: NM8VIVIDA


     SYSTEM PERFORMANCE CHECK RESULTS
     Dipole: D900V2 SN: 108
     Date: January 29, 2004
     Ambient condition: Temperature: 23.0°C; Relative humidity: 40%
                H e a d S im ula ting L iq uid
                                                            P a ra m e te rs      T a rg e t   M e a s ure d   D e v ia tio n[% ]   L im ite d [% ]
        f (M H z )     T e m p . [° C ]   D e p th [c m ]
                                                              P e rm itiv ity :    4 1 .5       4 1 .0 3 2          -1 .1 3              ± 5
          900             2 2 .0 0           1 5 .0 0       C o nd uc tiv ity :    0 .9 7       0 .9 7 0 2          0 .0 2               ± 5
                                                                  1g S A R :       1 0 .8        1 0 .8 4           0 .3 7              ± 10


     Dipole: D1800V2 SN: 294
     Date: January 29, 2004
     Ambient condition: Temperature: 23.5°C; Relative humidity: 40%
               H e a d S im ula ting L iq uid
                                                            P a ra m e te rs      T a rg e t   M e a s ure d   D e v ia tio n[% ]   L im ite d [% ]
        f (M H z )     T e m p . [°C ]    D e p th [c m ]
                                                              P e rm itiv ity :     40         4 0 .3 8 7 9         0 .9 7               ± 5
         1800             2 1 .5 0          1 5 .0 0        C o nd uc tiv ity :     1 .4         1 .4 6 8           4 .8 6               ± 5

                                                                  1g S A R :       3 8 .1        3 9 .4 0           3 .4 1              ± 10




COMPLIANCE CERTIFICATION SERVICES                                                                                              Page: 20 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


     8.3. EUT TUNE-UP PROCEDURES
           The following procedures had been used to prepare the EUT for the SAR test.

                To setup the desire channel frequency and the maximum output power. A Wireless
                Communication Test Set “Aglient, model 8960 Series” was used to program the EUT.

                 System config setting:
                 Operating mode - Active Cell
                 System Type - IS-2000
                 Cell Band - US Cellular (Rcvr Power Ctrl = 30 dBm)
                 Cell Band - US PCS (Rcvr Power Ctrl = 30 dBm)
                 Measurement Selection - Digital Average Power

           Maximum conducted power was measured by replacing the antenna with an RF cable for
           conductive measurements, before and after the SAR measurements was done.




COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 21 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


     8.4. SAR MEASUREMENT RESULTS
1. Left Touched Position




  CDMA Cellular Band: Duty Cycle: 100%, Crest Factor: 1                                 Depth of liquid: 15.0 cm

     EUT                               Frequency                  *Conducted              Liquid       SAR (1g)         Limit
                   Antenna
    Position                     Channel          MHz         Before         After      Temp_°C         (W/kg)        (W/kg)
                                    1013         824.70                                                    **
     Left
                    Fixed           384          836.52        24.02        23.85         21.50          0.355           1.6
   Touched
                                    777          848.31                                                    **

  CDMA PCS Band: Duty Cycle: 100%, Crest Factor: 1                                   Depth of liquid: 15.0 cm
     EUT                               Frequency                  *Conducted              Liquid       SAR (1g)         Limit
                   Antenna
    Position                     Channel          MHz          Before        After      Temp_°C         (W/kg)        (W/kg)
                                     25         1851.25                                                    **
     Left
                    Fixed           600         1880.00        24.75        24.60         21.50          0.748           1.6
   Touched
                                   1175         1908.75                                                    **

 Notes: 1. *: Average power
        2. **: The SAR measured at the middle channel for this configuration is at least 3 dB lower than SAR limit,
             testing at the high and low channels is option.
          3. Please refer to attachment for the result presentation in plot format.



COMPLIANCE CERTIFICATION SERVICES                                                                                Page: 22 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


2. Left Tilted Position




  CDMA Cellular Band: Duty Cycle: 100%, Crest Factor: 1                                 Depth of liquid: 15.0 cm
     EUT                             Frequency                  *Conducted               Liquid        SAR (1g)        Limit
                   Antenna
    Position                      Channel    MHz              Before   After            Temp_°C         (W/kg)        (W/kg)
                                    1013        824.70                                                     **
   Left Tilted      Fixed            384        836.52         24.02        23.85         21.50          0.331           1.6
                                     777        848.31                                                     **

  CDMA PCS Band: Duty Cycle: 100%, Crest Factor: 1                                  Depth of liquid: 15.0 cm
     EUT                             Frequency                  *Conducted               Liquid        SAR (1g)        Limit
                   Antenna
    Position                      Channel    MHz              Before   After            Temp_°C         (W/kg)        (W/kg)
                                     25         1851.25        24.30        24.15         22.00           1.04
   Left Tilted      Fixed            600          1880         24.75        24.60         22.00           1.18           1.6
                                    1175        1908.75        24.22        24.10         22.00          0.788

 Notes: 1. *: Average power
        2. **: The SAR measured at the middle channel for this configuration is at least 3 dB lower than SAR limit,
             testing at the high and low channels is option.
          3. Please refer to attachment for the result presentation in plot format.




COMPLIANCE CERTIFICATION SERVICES                                                                                Page: 23 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


3. Right Touched Position




  CDMA Cellular Band: Duty Cycle: 100%, Crest Factor: 1                                 Depth of liquid: 15.0 cm
     EUT                               Frequency                  *Conducted             Liquid        SAR (1g)        Limit
                   Antenna
    Position                      Channel         MHz         Before         After      Temp_°C         (W/kg)        (W/kg)
                                    1013         824.70        24.05        23.90         21.50          0.550
    Right
                    Fixed            384         836.52        24.02        23.85         21.50          0.511           1.6
   Touched
                                     777         848.31        24.00        23.80         21.50          0.508

  CDMA PCS Band: Duty Cycle: 100%, Crest Factor: 1                                   Depth of liquid: 15.0 cm
     EUT                               Frequency                  *Conducted             Liquid        SAR (1g)        Limit
                   Antenna
    Position                      Channel         MHz         Before         After      Temp_°C         (W/kg)        (W/kg)
                                     25         1851.25                                                    **
    Right
                    Fixed            600          1880         24.75        24.60          22.0          0.796           1.6
   Touched
                                    1175        1908.75                                                    **

 Notes: 1. *: Average power
        2. **: The SAR measured at the middle channel for this configuration is at least 3 dB lower than SAR limit,
             testing at the high and low channels is option.
          3. Please refer to attachment for the result presentation in plot format.




COMPLIANCE CERTIFICATION SERVICES                                                                                Page: 24 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


4. Right Tilted Position




  CDMA Cellular Band: Duty Cycle: 100%, Crest Factor: 1                                 Depth of liquid: 15.0 cm
     EUT                            Frequency                   *Conducted              Liquid        SAR (1g)         Limit
                   Antenna
    Position                     Channel   MHz                Before   After           Temp_°C         (W/kg)         (W/kg)
                                   1013         824.70         24.05        23.90                          **
     Right
                    Fixed           384         836.52         24.02        23.85        21.50           0.458           1.6
     Tilted
                                    777         848.31         24.00        23.80                          **

  CDMA PCS Band: Duty Cycle: 100%, Crest Factor: 1                                  Depth of liquid: 15.0 cm
     EUT                            Frequency                   *Conducted              Liquid        SAR (1g)         Limit
                   Antenna
    Position                     Channel   MHz                Before   After           Temp_°C         (W/kg)         (W/kg)
                                     25        1851.25        24.30        24.15          22.0            1.13
     Right
                    Fixed           600        1880.00        24.75        24.60          22.0            1.35           1.6
     Tilted
                                   1175        1908.75        24.22        24.10          22.0           0.825

 Notes: 1. *: Average power
        2. **: The SAR measured at the middle channel for this configuration is at least 3 dB lower than SAR limit,
              testing at the high and low channels is option.
          3. Please refer to attachment for the result presentation in plot format.




COMPLIANCE CERTIFICATION SERVICES                                                                                Page: 25 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


5. Body Worn Position




  CDMA Cellular Band: Duty Cycle: 100%, Crest Factor: 1                                 Depth of liquid: 15.0 cm
     Sep.                              Frequency                  *Conducted              Liquid       SAR (1g)         Limit
                   Antenna
   dist._cm                      Channel          MHz          Before        After      Temp_°C          (W/kg)        (W/kg)
                                   1013         824.70         24.05        23.90         21.50          0.542
     With
                    Fixed           384         836.52         24.02        23.85         21.50          0.396           1.6
    holster
                                    777         848.31         24.00        23.80         21.50          0.368

  CDMA PCS Band: Duty Cycle: 100%, Crest Factor: 1                                   Depth of liquid: 15.0 cm
                                                                  Conducted
     Sep.                              Frequency                                          Liquid       SAR (1g)         Limit
                   Antenna                                        Pwr dBm
   dist._cm                      Channel          MHz          Before   After           Temp_°C          (W/kg)        (W/kg)
                                     25         1851.25        24.30        24.15         22.00          0.466
     With
                    Fixed           600         1880.00        24.75        24.60         22.00          0.662           1.6
    holster
                                   1175         1908.75        24.22        24.10         22.00          0.417

 Notes:
 1. *: Average power
 2. A separation distance of 0 cm between the back of the EUT (holster) and flat phantom.
 3. The Ear-microphone wire connected to the phone to simulate hands-free operation in a body-worn
     configuration.
 4. Please refer to attachment for the result presentation in plot format.



COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 26 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.






fi   mones   O


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


10. EQUIPMENT LIST & CALIBRATION STATUS
     Name of Equipment                     Manufacturer         Type/Model     Serial Number         Cal. Due date
     S-Parameter Network Analyzer          Agilent              8753ES-6       US39173569            8/8/04
     Electronic Probe kit                  Hewlett Packard      85070C         N/A                   N/A
     Signal General                        HP                   83732B         US34490599            4/4/04
     Power Meter                           Giga-tronics         8651A          8651404               5/12/04
     Power Sensor                          Giga-tronics         80701A         1834588               2/18/04
     Amplifier                             Mini-Circuits        ZVE-8G         0360                  N/A
     Amplifier                             Mini-Circuits        ZHL-42W        D072701-5             N/A
     Wireless Communication Test Set Agilent                    8960           GP43344809            9/22/05
     Radio Communication Tester            Rohde & Schwarz CMU 200             838114/032            2/14/04
     Data Acquisition Electronics (DAE) SPEAG                   DAE3 V1        427                   2/4/04
     Dosimetric E-Field Probe              SPEAG                ES3DV2         3021                  7/29/04
     System Validation Dipole              SPEAG                D900V2         108                   4/10/04
     System Validation Dipole              SPEAG                D1800V2        294                   4/19/04
     Probe Alignment Unit                  SPEAG                LB (V2)        261                   N/A
     Robot                                 Staubli              RX90B L        F00/5H31A1/A/01 N/A
     SAM Twin Phantom                      SPEAG                TP-1785        QD 000 P40 CA         N/A
     SAM Twin Phantom                      SPEAG                TP-1015        N/A                   N/A
     Simulating Liquids                    SPEAG                H1800          N/A                   Within 24 hrs of first test
     Simulating Liquids                    SPEAG                H1900          N/A                   Within 24 hrs of first test
     Simulating Liquids                    SPEAG                M1900          N/A                   Within 24 hrs of first test
     Simulating Liquids                    SPEAG                H900           N/A                   Within 24 hrs of first test
     Simulating Liquids                    SPEAG                H835           N/A                   Within 24 hrs of first test
     Simulating Liquids                    SPEAG                M835           N/A                   Within 24 hrs of first test




COMPLIANCE CERTIFICATION SERVICES                                                                                Page: 30 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


11. REFERENCES
     [1]    Federal Communications Commission, \Report and order: Guidelines for evaluating the environ-
            mental effects of radiofrequency radiation", Tech. Rep. FCC 96-326, FCC, Washington, D.C.
            20554, 1996.
     [2]    David L. Means Kwok Chan, Robert F. Cleveland, \Evaluating compliance with FCC guidelines
            for human exposure to radiofrequency electromagnetic fields", Tech. Rep., Federal
            Communication Commision, O_ce of Engineering & Technology, Washington, DC, 1997.
     [3]    Thomas Schmid, Oliver Egger, and Niels Kuster, \Automated E-_eld scanning system for
            dosimetric assessments", IEEE Transactions on Microwave Theory and Techniques, vol. 44, pp.
            105{113, Jan. 1996.
     [4]    Niels Kuster, Ralph K.astle, and Thomas Schmid, \Dosimetric evaluation of mobile
            communications equipment with known precision", IEICE Transactions on Communications, vol.
            E80-B, no. 5, pp. 645{652, May 1997.
     [5]    CENELEC, \Considerations for evaluating of human exposure to electromagnetic fields (EMFs)
            from mobile telecommunication equipment (MTE) in the frequency range 30MHz - 6GHz", Tech.
            Rep., CENELEC, European Committee for Electrotechnical Standardization, Brussels, 1997.
     [6]    ANSI, ANSI/IEEE C95.1-1992: IEEE Standard for Safety Levels with Respect to Human
            Exposure to Radio Frequency Electromagnetic Fields, 3 kHz to 300 GHz, The Institute of
            Electrical and Electronics Engineers, Inc., New York, NY 10017, 1992.
     [7]    Katja Pokovic, Thomas Schmid, and Niels Kuster, \Robust setup for precise calibration of E-_eld
            probes in tissue simulating liquids at mobile communications frequencies", in ICECOM _ 97,
            Dubrovnik, October 15{17, 1997, pp. 120{124.
     [8]    Katja Pokovic, Thomas Schmid, and Niels Kuster, \E-_eld probe with improved isotropy in brain
            simulating liquids", in Proceedings of the ELMAR, Zadar, Croatia, 23{25 June, 1996, pp.
            172{175.
     [9]    Volker Hombach, Klaus Meier, Michael Burkhardt, Eberhard K. uhn, and Niels Kuster, \The
            dependence of EM energy absorption upon human head modeling at 900 MHz", IEEE
            Transactions onMicrowave Theory and Techniques, vol. 44, no. 10, pp. 1865{1873, Oct. 1996.
     [10]   Klaus Meier, Ralf Kastle, Volker Hombach, Roger Tay, and Niels Kuster, \The dependence of
            EM energy absorption upon human head modeling at 1800 MHz", IEEE Transactions on
            Microwave Theory and Techniques, Oct. 1997, in press.
     [11]   W. Gander, Computermathematik, Birkhaeuser, Basel, 1992.
     [12]   W. H. Press, S. A. Teukolsky,W. T. Vetterling, and B. P. Flannery, Numerical Recepies in C,
            The Art of Scientific Computing, Second Edition, Cambridge University Press, 1992..Dosimetric
            Evaluation of Sample device, month 1998 9
     [13]   NIS81 NAMAS, \The treatment of uncertainity in EMC measurement", Tech. Rep., NAMAS
            Executive, National Physical Laboratory, Teddington, Middlesex, England, 1994.
     [14]   Barry N. Taylor and Christ E. Kuyatt, \Guidelines for evaluating and expressing the uncertainty
            of NIST measurement results", Tech. Rep., National Institute of Standards and Technology,
            1994. Dosimetric Evaluation of Sample device, month 1998 10




COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 31 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.


REPORT NO: 04T2458-2                             DATE: January 29, 2004                                 FCC ID: NM8VIVIDA


12. ATTACHMENTS

         No.                                            Contents                                             No. of page (s)
          1        System Performance Check Plots                                                                     4
         2-1       SAR Test Plots - Cellular band                                                                    14
         2-2       SAR Test Plots - PCS band                                                                         16
          3        Probe_ES3DV2-SN: 3021                                                                             13
          4        System Validation Dipole - D900V2 SN: 108                                                          7
          5        System Validation Dipole - D1800V2 S/N: 294                                                        6




                                                      END OF REPORT




COMPLIANCE CERTIFICATION SERVICES                                                                               Page: 32 of 32
This report shall not be reproduced except in full, without the written approval of CCS. This document may be altered or revised by
Compliance Certification Services personnel only, and shall be noted in the revision section of the document.



Document Created: 2004-01-29 20:21:28
Document Modified: 2004-01-29 20:21:28

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC