SAR Report

FCC ID: NM8BG

RF Exposure Info

Download: PDF
FCCID_579026

              Compliance Certification Services Inc.
              Report No: 50630102-SF            FCC ID: NM8BG                 Date of Issue: Jul 5, 2005




                              ANSI/IEEE Std. C95.1-1992
                               in accordance with the requirements of
            FCC Report and Order: ET Docket 93-62, and OET Bulletin 65 Supplement C
                                IC Report and Order: RSS-102



                                 FCC TEST REPORT

                                               For


                                          Pocket PC


                                       Trade Name: HP


                               Model: HSTNH-H09C-WL


                                             Issued to

                         HIGH TECH COMPUTER CORP.
                              23 Hsin Hua Rd, Taoyuan,R.O.C.


                                             Issued by

                       Compliance Certification Services Inc.
                      No. 11, Wu-Kung 6 Rd, Wu-Ku Hsiang,
                 Wu-Ku Industrial District, Taipei Hsien, (248) Taiwan.
                               TEL: 886-2-2299-9720
                              FAX: 886-2-2299-9721




Note: This report shall not be reproduced except in full, without the written approval of Compliance
 Certification Services Inc. This document may be altered or revised by Compliance Certification
      Services Inc. personnel only, and shall be noted in the revision section of the document.


                                                 Page 1                                   Total Page: 24
                                                                                                Rev. 00


                      Compliance Certification Services Inc.
                      Report No: 50630102-SF                                                                        Date of Issue: Jul 5, 2005




                                              TABLE OF CONTENTS
1.          EUT DESCRIPTION ....................................................................................................................4
2.          REQUIREMENTS FOR COMPLIANCE TESTING DEFINED BY THE FCC.........................5
3.          DOSIMETRIC ASSESSMENT SYSTEM ...................................................................................5
      3.1 MEASUREMENT SYSTEM DIAGRAM....................................................................................6
      3.2 SYSTEM COMPONENTS ...........................................................................................................7
4.          EVALUATION PROCEDURES ..................................................................................................9
5.          MEASUREMENT UNCERTAINTY .........................................................................................13
6.          EXPOSURE LIMIT ....................................................................................................................14
7.          TYPICAL COMPOSITION OF INGREDIENTS FOR LIQUID TISSUE PHANTOMS .........14
8.          MEASUREMENT RESULTS ....................................................................................................15
      8.1 SYSTEM PERFORMANCE CHECK ................................................................................................ 15
      8.2 TEST LIQUID CONFIRMATION...................................................................................................... 16
      8.3 EUT TUNE-UP PROCEDURES ......................................................................................................... 17
      8.4 SAR MEASUREMENTS RESULTS.................................................................................................. 17
9.          EUT PHOTOS ............................................................................................................................20
10.         EQUIPMENT LIST & CALIBRATION STATUS ....................................................................22
11.         FACILITIES ...............................................................................................................................23
12.         REFERENCES............................................................................................................................23
13.         ATTACHMENTS .......................................................................................................................24




                                                                         Page 2                                                             Rev. 00


                   Compliance Certification Services Inc.
                   Report No: 50630102—8F                                           Date ofIssue: Jul 5, 2005




      CERTIFICATE OF COMPLIANCE (SAR EVALUATION)

Applicant:                         HIGH TECH COMPUTER CORP.
                                   22 Chungshan N. Rd., 3 Sec.
Equipment Under Test:              Pocket PC
Trade Name:                        HP
Model Number:                      HSTNH—HO9C—WL
Date of Test:                      July 5, 2005
Device Category:                   PORTABLE DEVICES
Exposure Category:                 GENERAL POPULATION/UNCONTROLLED EXPOSURE


                                      APPLICABLE STANDARDS
                      STANDARD                                           TEST RESULT

                k6 OEIE ;Ssz'_lfg;emem &                             No non—compliancenoted

                                 Deviation from Applicable Standard
                                                  None

The device was tested by Compliance Certification Services Inc. in accordance with the measurement
methods and procedures specified in OET Bulletin 65 Supplement C (Edition 01—01). The test results in this
report apply only to thetested sample ofthe stated device/equipment. Other similar device/equipment will
not necessarily produce the sameresults due to production tolerance and measurement uncertainties.


Approved by:                                              Reviewed by:




Harris W. Lai                                             Chris Hsich
Executive Vice President                                  Section Manager
Compliance Certification Services Inc.                    Compliance Certification Services Inc.




                                                     Page3                                           Rev. 00


                   Compliance Certification Services Inc.
                   Report No: 50630102-SF                                   Date of Issue: Jul 5, 2005




1. EUT DESCRIPTION

 Product                        Pocket PC

 Trade Name                     HP

 Model Number                   HSTNH-H09C-WL

 Model Discrepancy              N/A

 Frequency Range                802.11b: 2412 ~ 2462 MHz

 Max. O/P Power:
 (Conducted/Average)
                                802.11b: 15.31dBm


 Max. SAR (1g):                 802.11b: 0.536 W/kg


 Modulation Technique           802.11b: Direct Sequence Spread Spectrum (DSSS)


 Antenna Specification          Antenna type: Main antenna: PIFA antenna


 Battery option:                Only one type / model with EUT

Remark: The sample selected for test was prototype that approximated to production product
and was provided by manufacturer.




                                                    Page 4                                    Rev. 00


               Compliance Certification Services Inc.
               Report No: 50630102-SF                                           Date of Issue: Jul 5, 2005



2. REQUIREMENTS FOR COMPLIANCE TESTING DEFINED BY THE FCC
  The US Federal Communications Commission has released the report and order “Guidelines for
  Evaluating the Environmental Effects of RF Radiation", ET Docket No. 93-62 in August 1996 [1].
  The order requires routine SAR evaluation prior to equipment authorization of portable transmitter
  devices, including portable telephones. For consumer products, the applicable limit is 1.6 mW/g for
  an uncontrolled environment and 8.0 mW/g for an occupational/controlled environment as
  recommended by the ANSI/IEEE standard C95.1-1992 [6]. According to the Supplement C of OET
  Bulletin 65 “Evaluating Compliance with FCC Guide-lines for Human Exposure to Radio frequency
  Electromagnetic Fields", released on Jun 29, 2001 by the FCC, the device should be evaluated at
  maximum output power (radiated from the antenna) under “worst-case” conditions for normal or
  intended use, incorporating normal antenna operating positions, device peak performance frequencies
  and positions for maximum RF energy coupling.
3. DOSIMETRIC ASSESSMENT SYSTEM
  These measurements were performed with the automated near-field scanning system DASY4 from
  Schmid & Partner Engineering AG (SPEAG). The system is based on a high precision robot
  (working range greater than 0.9 m) which positions the probes with a positional repeatability of better
  than ± 0.02 mm. Special E- and H-field probes have been developed for measurements close to
  material discontinuity, the sensors of which are directly loaded with a Schottky diode and connected
  via highly resistive lines to the data acquisition unit. The SAR measurements were conducted with
  the dosimetric probe EX3DV4-SN: 3554 (manufactured by SPEAG), designed in the classical
  triangular configuration and optimized for dosimetric evaluation. The probe has been calibrated
  according to the procedure with accuracy of better than ±10%. The spherical isotropy was evaluated
  with the procedure and found to be better than ±0.25 dB. The phantom used was the SAM Twin
  Phantom as described in FCC supplement C, IEEE P1528 and EN50361.




                                                  Page 5                                          Rev. 00


              Compliance Certification Services Inc.
              Report No: 50630102-SF                                                Date of Issue: Jul 5, 2005



3.1 MEASUREMENT SYSTEM DIAGRAM




     The DASY4 system for performing compliance tests consists of the following items:
     • A standard high precision 6-axis robot (St¨aubli RX family) with controller, teach pendant and
         software. An arm extension for accommodating the data acquisition electronics (DAE).
     • A dosimetric probe, i.e., an isotropic E-field probe optimized and calibrated for usage in tissue
         simulating liquid. The probe is equipped with an optical surface detector system.
     • A data acquisition electronics (DAE) which performs the signal amplification, signal multiplexing,
         AD-conversion, offset measurements, mechanical surface detection, collision detection, etc. The unit
         is battery powered with standard or rechargeable batteries. The signal is optically transmitted to the
         EOC.
     • The Electro-optical converter (EOC) performs the conversion between optical and electrical of the
         signals for the digital communication to the DAE and for the analog signal from the optical surface
         detection. The EOC is connected to the measurement server.
     • The function of the measurement server is to perform the time critical tasks such as signal filtering,
         control of the robot operation and fast movement interrupts.
     • A probe alignment unit which improves the (absolute) accuracy of the probe positioning.
     • A computer operating Windows 2000 or Windows XP.
     • DASY4 software.
     • Remote control with teach pendant and additional circuitry for robot safety such as warning lamps,
         etc.
     • The SAM twin phantom enabling testing left-hand and right-hand usage.
     • The device holder for handheld mobile phones.
     • Tissue simulating liquid mixed according to the given recipes.
     • Validation dipole kits allowing validating the proper functioning of the system.




                                                    Page 6                                            Rev. 00


                   Compliance Certification Services Inc.
                   Report No: 50630102-SF                                                    Date of Issue: Jul 5, 2005



3.2 SYSTEM COMPONENTS

DASY4 Measurement Server
                                    The DASY4 measurement server is based on a PC/104 CPU board with a
                                    166MHz low-power Pentium, 32MB chip disk and 64MB RAM. The
                                    necessary circuits for communication with either the DAE3 electronic box
                                    as well as the 16-bit AD-converter system for optical detection and digital
                                    I/O interface are contained on the DASY4 I/O-board, which is directly
                                    connected to the PC/104 bus of the CPU board.
                                    The measurement server performs all real-time data evaluation for field
                                    measurements and surface detection, controls robot movements and handles
                                    safety operation.
The PC-operating system cannot interfere with these time critical processes. All connections are supervised by a
watchdog, and disconnection of any of the cables to the measurement server will automatically disarm the robot
and disable all program-controlled robot movements. Furthermore, the measurement server is equipped with
two expansion slots which are reserved for future applications. Please note that the expansion slots do not have a
standardized pinout and therefore only the expansion cards provided by SPEAG can be inserted. Expansion
cards from any other supplier could seriously damage the measurement server. Calibration: No calibration
required.
Data Acquisition Electronics (DAE)
The data acquisition electronics (DAE3) consists of a highly sensitive
electrometer grade preamplifier with auto-zeroing, a channel and gain-
switching multiplexer, a fast 16 bit AD converter and a command
decoder and control logic unit. Transmission to the measurement
server is accomplished through an optical downlink for data and status
information as well as an optical uplink for commands and the clock.
The mechanical probe mounting device includes two different sensor
systems for frontal and sideways probe contacts. They are used for
mechanical surface detection and probe collision detection. The input
impedance of the DAE3 box is 200MOhm; the inputs are symmetrical
and floating. Common mode rejection is above 80 dB.
EX3DV4 Isotropic E-Field Probe for Dosimetric Measurements
Construction:  Symmetrical design with triangular core
               Built-in shielding against static charges
               PEEK enclosure material (resistant to organic solvents, e.g., DGBE)
Calibration:   Basic Broad Band Calibration in air: 10-3000 MHz.
               Conversion Factors (CF) for HSL 900 and HSL 1800
               CF-Calibration for other liquids and frequencies upon request.
Frequency:     10 MHz to > 6 GHz; Linearity: ± 0.2 dB (30 MHz to 3 GHz)
Directivity:   ± 0.3 dB in HSL (rotation around probe axis)
               ± 0.5 dB in HSL (rotation normal to probe axis)
Dynamic Range: 10 µW/g to > 100 mW/g; Linearity: ± 0.2 dB
                (noise: typically < 1 µW/g)

Dimensions:       Overall length: 330 mm (Tip: 20 mm)
                  Tip diameter: 2.5 mm (Body: 12 mm)
                  Distance from probe tip to dipole centers: 1 mm
Application:      High precision dosimetric measurements in any
                  exposure scenario (e.g., very strong gradient fields).
                  Only probe which enables compliance testing for
                  frequencies up to 6 GHz with precision of better
                  30%.                                                   Interior of probe
                                                                                             E-Field probe EX3DV4




                                                               Page 7                                          Rev. 00


                   Compliance Certification Services Inc.
                   Report No: 50630102-SF                                                    Date of Issue: Jul 5, 2005




SAM Phantom (V4.0)
Construction:    The shell corresponds to the specifications of the
                 Specific Anthropomorphic Mannequin (SAM)
                 phantom defined in IEEE 1528-200X, CENELEC
                 50361 and IEC 62209. It enables the dosimetric
                 evaluation of left and right hand phone usage as
                 well as body mounted usage at the flat phantom
                 region. A cover prevents evaporation of the liquid.
                 Reference markings on the phantom allow the
                 complete setup of all predefined phantom
                 positions and measurement grids by manually
                 teaching three points with the robot.
Shell Thickness: 2 ±0.2 mm
Filling Volume: Approx. 25 liters
Dimensions:      Height: 810mm; Length: 1000mm; Width:
                 500mm
Device Holder for SAM Twin Phantom
Construction:    In combination with the Twin SAM Phantom V4.0 or Twin SAM, the Mounting
                 Device (made from POM) enables the rotation of the mounted transmitter in
                 spherical coordinates, whereby the rotation point is the ear opening. The
                 devices can be easily and accurately positioned according to IEC, IEEE,
                 CENELEC, FCC or other specifications. The device holder can be locked at
                 different phantom locations (left head, right head, and flat phantom).




System Validation Kits
Construction:      Symmetrical dipole with l/4 balun Enables measurement of feedpoint
                   impedance with NWA Matched for use near flat phantoms filled with brain
                   simulating solutions Includes distance holder and tripod adaptor.
Frequency:         450, 900, 1800, 2450, 5800 MHz
Return loss:       > 20 dB at specified validation position
Power capability: > 100 W (f < 1GHz); > 40 W (f > 1GHz)
Dimensions:        450V2: dipole length: 270 mm; overall height: 330 mm
                   D900V2: dipole length: 149 mm; overall height: 330 mm
                   D1800V2: dipole length: 72 mm; overall height: 300 mm
                   D2450V2: dipole length: 51.5 mm; overall height: 300 mm
                   D5GHzV2: dipole length: 25.5 mm; overall height: 290 mm




                                                              Page 8                                           Rev. 00


                    Compliance Certification Services Inc.
                    Report No: 50630102-SF                                                                          Date of Issue: Jul 5, 2005




4. EVALUATION PROCEDURES

DATA EVALUATION
The DASY4 post processing software (SEMCAD) automatically executes the following procedures to calculate the
field units from the microvolt readings at the probe connector. The parameters used in the evaluation are stored in
the configuration modules of the software:
              Probe parameters:      - Sensitivity                  Normi, ai0, ai1, ai2
                                     - Conversion factor            ConvFi
                                     - Diode compression point      dcpi
              Device parameters:     - Frequency                    f
                                     - Crest factor                 cf
              Media parameters:      - Conductivity                 σ
                                     - Density                      ρ
These parameters must be set correctly in the software. They can be found in the component documents or be
imported into the software from the configuration files issued for the DASY components. In the direct measuring
mode of the multi-meter option, the parameters of the actual system setup are used. In the scan visualization and
export modes, the parameters stored in the corresponding document files are used.
The first step of the evaluation is a linearization of the filtered input signal to account for the compression
characteristics of the detector diode. The compensation depends on the input signal, the diode type and the DC-
transmission factor from the diode to the evaluation electronics. If the exciting field is pulsed, the crest factor of the
signal must be known to correctly compensate for peak power. The formula for each channel can be given as:
                                          2        cf
                    V =U +U
                       i      i           i
                                              ⋅
                                                  dcp i

            with    Vi        = Compensated signal of channel i                                  (i = x, y, z)
                    Ui        = Input signal of channel i                                        (i = x, y, z)
                    cf        = Crest factor of exciting field                                   (DASY parameter)
                    dcpi      = Diode compression point                                          (DASY parameter)

From the compensated input signals the primary field data for each channel can be evaluated:
            E-field probes:                        V
                              E       =                           i
                                  i
                                              Norm • ConvF    i


                                                                                                  2

            H-field probes:                                            a +a f +a f
                                                                      i10   i11            i12

                                                   H      i
                                                              =   Vi ⋅
                                                                            f
            with    Vi    = Compensated signal of channel i                                      (i = x, y, z)
                    Normi = Sensor sensitivity of channel i                                      (i = x, y, z)
                            µV/(V/m)2 for E0field Probes
                    ConvF = Sensitivity enhancement in solution
                    aij   = Sensor sensitivity factors for H-field probes
                    f     = Carrier frequency (GHz)
                    Ei    = Electric field strength of channel i in V/m
                    Hi    = Magnetic field strength of channel i in A/m




                                                                                  Page 9                                              Rev. 00


                    Compliance Certification Services Inc.
                    Report No: 50630102-SF                                                    Date of Issue: Jul 5, 2005




The RSS value of the field components gives the total field strength (Hermitian magnitude):
                                     2             2        2
                    E   tot
                              =    E +E +E
                                     x             y        z


The primary field data are used to calculate the derived field units.

                                         2             σ
                    SAR = E                    ⋅
                                         tot
                                                   ρ   ⋅1000

            with    SAR           = local specific absorption rate in mW/g
                    Etot          = total field strength in V/m
                    σ             = conductivity in [mho/m] or [Siemens/m]
                    ρ             = equivalent tissue density in g/cm3
Note that the density is normally set to 1 (or 1.06), to account for actual brain density rather than the density of the
simulation liquid.
The power flow density is calculated assuming the excitation field as a free space field.
                                     2
                                   E                   or       P
                                                                              2
                                                                          = H tot ⋅ 37.7
                    P         =      tot
                                                                    pwe
                        pwe
                                  3770
            with    Ppwe = Equivalent power density of a plane wave in mW/cm2
                    Etot    = total electric field strength in V/m
         Htot      = total magnetic field strength in A/m




                                                                                    Page 10                      Rev. 00


                   Compliance Certification Services Inc.
                   Report No: 50630102-SF                                                 Date of Issue: Jul 5, 2005



SAR MEASUREMENT PROCEDURES
The procedure for assessing the peak spatial-average SAR value consists of the following steps:
•   Power Reference Measurement
    The reference and drift jobs are useful jobs for monitoring the power drift of the device under test in the batch
    process. Both jobs measure the field at a specified reference position, at a selectable distance from the phantom
    surface. The reference position can be either the selected section’s grid reference point or a user point in this
    section. The reference job projects the selected point onto the phantom surface, orients the probe
    perpendicularly to the surface, and approaches the surface using the selected detection method.
•   Area Scan
    The area scan is used as a fast scan in two dimensions to find the area of high field values, before doing a finer
    measurement around the hot spot. The sophisticated interpolation routines implemented in DASY4 software can
    find the maximum locations even in relatively coarse grids. The scan area is defined by an editable grid. This
    grid is anchored at the grid reference point of the selected section in the phantom. When the area scan’s
    property sheet is brought-up, grid was at to 15 mm by 15 mm and can be edited by a user.
•   Zoom Scan
    Zoom scans are used to assess the peak spatial SAR values within a cubic averaging volume containing 1 g and
    10 g of simulated tissue. The default zoom scan measures 5x5x7 points within a cube whose base faces are
    centered around the maximum found in a preceding area scan job within the same procedure. If the preceding
    Area Scan job indicates more then one maximum, the number of Zoom Scans has to be enlarged accordingly
    (The default number inserted is 1).
•   Power Drift measurement
    The drift job measures the field at the same location as the most recent reference job within the same procedure,
    and with the same settings. The drift measurement gives the field difference in dB from the reading conducted
    within the last reference measurement. Several drift measurements are possible for one reference measurement.
    This allows a user to monitor the power drift of the device under test within a batch process. In the properties
    of the Drift job, the user can specify a limit for the drift and have DASY4 software stop the measurements if
    this limit is exceeded.
•   Z-Scan
    The Z Scan job measures points along a vertical straight line. The line runs along the Z-axis of a one-
    dimensional grid. A user can anchor the grid to the current probe location. As with any other grids, the local Z-
    axis of the anchor location establishes the Z-axis of the grid.




                                                         Page 11                                             Rev. 00


              Compliance Certification Services Inc.
              Report No: 50630102-SF                                                 Date of Issue: Jul 5, 2005



SPATIAL PEAK SAR EVALUATION
The procedure for spatial peak SAR evaluation has been implemented according to the IEEE1529 standard. It
can be conducted for 1 g and 10 g.
The DASY4 system allows evaluations that combine measured data and robot positions, such as:
   • maximum search
   • extrapolation
   • boundary correction
   • peak search for averaged SAR
During a maximum search, global and local maximum searches are automatically performed in 2-D after each
Area Scan measurement with at least 6 measurement points. It is based on the evaluation of the local SAR
gradient calculated by the Quadratic Shepard’s method. The algorithm will find the global maximum and all
local maxima within -2 dB of the global maxima for all SAR distributions.
Extrapolation
Extrapolation routines are used to obtain SAR values between the lowest measurement points and the inner
phantom surface. The extrapolation distance is determined by the surface detection distance and the probe
sensor offset. Several measurements at different distances are necessary for the extrapolation.
Extrapolation routines require at least 10 measurement points in 3-D space. They are used in the Cube Scan to
obtain SAR values between the lowest measurement points and the inner phantom surface. The routine uses the
modified Quadratic Shepard’s method for extrapolation. For a grid using 5x5x7 measurement points with 5mm
resolution amounting to 343 measurement points, the uncertainty of the extrapolation routines is less than 1%
for 1 g and 10 g cubes.
Boundary effect
For measurements in the immediate vicinity of a phantom surface, the field coupling effects between the probe
and the boundary influence the probe characteristics. Boundary effect errors of different dosimetric probe types
have been analyzed by measurements and using a numerical probe model. As expected, both methods showed
an enhanced sensitivity in the immediate vicinity of the boundary. The effect strongly depends on the probe
dimensions and disappears with increasing distance from the boundary. The sensitivity can be approximately
given as:




Since the decay of the boundary effect dominates for small probes (a<<λ), the cos-term can be omitted. Factors
Sb (parameter Alpha in the DASY4 software) and a (parameter Delta in the DASY4 software) are assessed
during probe calibration and used for numerical compensation of the boundary effect. Several simulations and
measurements have confirmed that the compensation is valid for different field and boundary configurations.
This simple compensation procedure can largely reduce the probe uncertainty near boundaries. It works well as
long as:
   • the boundary curvature is small
   • the probe axis is angled less than 30_ to the boundary normal
   • the distance between probe and boundary is larger than 25% of the probe diameter
   • the probe is symmetric (all sensors have the same offset from the probe tip)
Since all of these requirements are fulfilled in a DASY4 system, the correction of the probe boundary effect in
the vicinity of the phantom surface is performed in a fully automated manner via the measurement data
extraction during postprocessing.




                                                    Page 12                                            Rev. 00


                      Compliance Certification Services Inc.
                      Report No: 50630102-SF                                                 Date of Issue: Jul 5, 2005




5. MEASUREMENT UNCERTAINTY
                            UNCERTAINTY BUDGE ACCORDING TO IEEE P1528
                                                                                           Standard
                                  Uncertainty Probability
Error Description                 Value ±%
                                                                  Divisor   C1 1g          unc.(1g/10g)   V1 or Veff
                                              distribution                                 ±%
Measurement System
Probe calibration                 ±4.8         normal             1         1              ±4.8           ∞
                                                                                     1/2
Axial isotropy of probe           ±4.6         rectangular        √3        (1-Cp)         ±1.9           ∞

Sph. Isotropy of probe            ±9.7         rectangular        √3        (Cp)1/2        ±3.9           ∞

Probe linearity                   ±4.5         rectangular        √3        1              ±2.7           ∞

Detection Limit                   ±0.9         rectangular        √3        1              ±0.6           ∞

Boundary effects                  ±8.5         rectangular        √3        1              ±4.8           ∞

Readoutelectronics                ±1.0         normal             1         1              ±1.0           ∞
Response time                     ±0.9         rectangular        √3        1              ±0.5           ∞

Integration time                  ±1.2         rectangular        √3        1              ±0.8           ∞

Mech Constrains of robot          ±0.5         rectangular        √3        1              ±0.2           ∞

Probe positioning                 ±2.7         rectangular        √3        1              ±1.7           ∞

Extrap. And integration           ±4.0         rectangular        √3        1              ±2.3           ∞

RF ambient conditiona             ±0.54        rectangular        √3        1              ±0.43          ∞

Test Sample Related
Device positioning                ±2.2         normal             1         1              ±2.23          11
Device holder uncertainty         ±5           normal             1         1              ±5.0           7
Power drift                       ±5           rectangular        √3        1              ±2.9           ∞

Phantom and Set up
Phantom uncertainty               ±4           rectangular        √3        1              ±2.3           ∞

Liquid conductivity               ±5           rectangular        √3        0.6            ±1.7           ∞
Liquid conductivity               ±5           rectangular        √3        0.6            ±3.5/1.7       ∞

Liquid permittivity               ±5           rectangular        √3        0.6            ±1.7           ∞

Liquid permittivity               ±5           rectangular        √3        0.6            ±1.7           ∞


Combined Standard Uncertainty                                                              ±12.14/11.76

Coverage Factor for 95%                        kp=2
Expanded Standard Uncertainty                                                              ±24.29/23.51
    Table: Worst-case uncertainty for DASY4 assessed according to IEEE P1528.
    The budge is valid for the frequency range 300 MHz to 6G Hz and represents a worst-case analysis.




                                                        Page 13                                                Rev. 00


                   Compliance Certification Services Inc.
                   Report No: 50630102-SF                                                                              Date of Issue: Jul 5, 2005




6. EXPOSURE LIMIT
  (A). Limits for Occupational/Controlled Exposure (W/kg)
      Whole-Body            Partial-Body                Hands, Wrists, Feet and Ankles
      0.4                   8.0                         2.0


  (B). Limits for General Population/Uncontrolled Exposure (W/kg)
      Whole-Body            Partial-Body                Hands, Wrists, Feet and Ankles
      0.08                  1.6                         4.0
      NOTE:      Whole-Body SAR is averaged over the entire body, partial-body SAR is averaged over any
                 1 gram of tissue defined as a tissue volume in the shape of a cube. SAR for hands, wrists,
                 feet and ankles is averaged over any 10 grams of tissue defined as a tissue volume in the
                 shape of a cube.
      Population/Uncontrolled Environments:
      are defined as locations where there is the exposure of individuals who have no knowledge or
      control of their exposure.
      Occupational/Controlled Environments:
      are defined as locations where there is exposure that may be incurred by people who are aware of
      the potential for exposure, (i.e. as a result of employment or occupation).


                                          NOTE
                       GENERAL POPULATION/UNCONTROLLED EXPOSURE
                                   PARTIAL BODY LIMIT
                                         1.6 W/kg


7. TYPICAL COMPOSITION OF INGREDIENTS FOR LIQUID TISSUE PHANTOMS
  The following tissue formulations are provided for reference only as some of the parameters have not been
  thoroughly verified. The composition of ingredients may be modified accordingly to achieve the desired target
  tissue parameters required for routine SAR evaluation.
         Ingredients                                                               Frequency (MHz)
        (% by weight)                     450                         835                     915                   1900                 2450
         Tissue Type              Head          Body          Head          Body      Head          Body    Head           Body   Head          Body
             Water                38.56         51.16         41.45         52.4      41.05         56.0    54.9           40.4   62.7          73.2
         Salt (NaCl)              3.95          1.49          1.45          1.4       1.35          0.76    0.18           0.5    0.5           0.04
           Sugar                  56.32         46.78         56.0          45.0      56.5          41.76    0.0           58.0   0.0            0.0
             HEC                  0.98          0.52           1.0          1.0        1.0          1.21     0.0           1.0    0.0           0.0
         Bactericide              0.19          0.05           0.1          0.1        0.1          0.27     0.0           0.1    0.0           0.0
         Triton X-100              0.0           0.0           0.0          0.0        0.0           0.0     0.0           0.0    36.8          0.0
             DGBE                  0.0           0.0           0.0          0.0        0.0           0.0    44.92          0.0    0.0           26.7
      Dielectric Constant         43.42         58.0          42.54         56.1      42.0          56.8    39.9           54.0   39.8          52.5
      Conductivity (S/m)          0.85          0.83          0.91          0.95      1.0           1.07    1.42           1.45   1.88          1.78

  Salt: 99+% Pure Sodium Chloride            Sugar: 98+% Pure Sucrose
                             +
  Water: De-ionized, 16 MΩ resistivity       HEC: Hydroxyethyl Cellulose
  DGBE: 99+% Di(ethylene glycol) butyl ether, [2-(2-butoxyethoxy)ethanol]
  Triton X-100 (ultra pure): Polyethylene glycol mono [4-(1, 1, 3, 3-tetramethylbutyl)phenyl]ether




                                                                            Page 14                                                             Rev. 00


                      Compliance Certification Services Inc.
                      Report No: 50630102-SF                                                             Date of Issue: Jul 5, 2005




8. MEASUREMENT RESULTS
8.1 SYSTEM PERFORMANCE CHECK
  The system performance check is performed prior to any usage of the system in order to guarantee reproducible
  results. The system performance check verifies that the system operates within its specifications. The system
  performance check results are tabulated below. And also the corresponding SAR plot is attached as well in the
  SAR plots files.
  SYSTEM PERFORMANCE CHECK MEASUREMENT CONDITIONS
  •     The measurements were performed in the flat section of the SAM twin phantom filled with Body
        simulating liquid of the following parameters.
  •     The DASY4 system with an E-fileld probe EX3DV4 SN: 3554 was used for the measurements.
  •     The dipole was mounted on the small tripod so that the dipole feed point was positioned below the center
        marking of the flat phantom section and the dipole was oriented parallel to the body axis (the long side of
        the phantom). The standard measuring distance was 15 mm (below 1 GHz) and 10 mm (above 1 GHz)
        from dipole center to the simulating liquid surface.
  •     The coarse grid with a grid spacing of 10mm was aligned with the dipole.
  •     Special 5x5x7 fine cube was chosen for cube integration (dx=dy= 7.5 mm, dz= 3 mm).
  •     Distance between probe sensors and phantom surface was set to 3.0 mm.
  •     The dipole input power (forward power) was 250 mW±3%.
  •     The results are normalized to 1 W input power.
  Reference SAR values
  The reference SAR values were using measurement results indicated in the dipole calibration document (see
  table below)
       Frequency                                         Local SAR at Surface                    Local SAR at Surface
                 1g SAR                   10g SAR
         (MHz)                                            (Above Feed Point)                (y = 2cm offset from feed point)
             900              10.3              6.57                   16.4                                   5.4
            1800              38.2              20.3                   69.5                                   6.8
      2450(Body)              53.6              25.0                  107.2                                  N/A

  SYSTEM PERFORMANCE CHECK RESULTS
  Dipole: D2450V2 SN: 728
  Date: Jul 5 , 2005
  Ambient condition: Temperature 24.8°C; Relative humidity: 57%
               B od y S im u latin g Liq u id
                                                         P aram eters         T a rget   M easu red   D eviation [% ]   L im ited [% ]
      f(M H z)        T em p . [°C ]    D ep th [cm ]
                                                         P erm itivity:       5 2 .7 0    5 1 .1 0        -3 .0 4            ± 5
      2 4 5 0 .0 0        2 3 .7 0         1 5 .0 0     C on d u c tivity:    1 .9 5       1 .9 1         -2 .0 5            ± 5
                                                           1g SAR :           5 3 .6 0    5 3 .2 0        -0 .7 5            ± 5




                                                                    Page 15                                                   Rev. 00


                   Compliance Certification Services Inc.
                   Report No: 50630102-SF                                                Date of Issue: Jul 5, 2005



8.2 TEST LIQUID CONFIRMATION
   SIMULATING LIQUIDS PARAMETER CHECK
   The simulating liquids should be checked at the beginning of a series of SAR measurements to determine of the
   dielectric parameters are within the tolerances of the specified target values
   The relative permittivity and conductivity of the tissue material should be within ± 5% of the values given in the
   table below. 5% may not be easily achieved at certain frequencies. Under such circumstances, 10% tolerance
   may be used until more precise tissue recipes are available
   IEEE SCC-34/SC-2 P1528 RECOMMENDED TISSUE DIELECTRIC PARAMETERS
   The head tissue dielectric parameters recommended by the IEEE SCC-34/SC-2 in P1528 have been
   incorporated in the following table. These head parameters are derived from planar layer models simulating the
   highest expected SAR for the dielectric properties and tissue thickness variations in a human head. Other head
   and body tissue parameters that have not been specified in P1528 are derived from the tissue dielectric
   parameters computed from the 4-Cole-Cole equations and extrapolated according to the head parameters
   specified in P1528
       Target Frequency                        Head                                      Body
             (MHz)                     εr                σ (S/m)                 εr               σ (S/m)
                150                  52.3                   0.76                61.9               0.80
                300                  45.3                   0.87                58.2               0.92
                450                  43.5                   0.87                56.7               0.94
                835                  41.5                   0.90                55.2               0.97
                900                  41.5                   0.97                55.0               1.05
                915                  41.5                   0.98                55.0               1.06
                1450                 40.5                   1.20                54.0               1.30
                1610                 40.3                   1.29                53.8               1.40
           1800-2000                 40.0                   1.40                53.3               1.52
                2450                 39.2                   1.80                52.7               1.95
                3000                 38.5                   2.40                52.0               2.73
                5800                 45.3                   5.27                48.2               6.00


   SIMULATING LIQUIDS PARAMETER CHECK RESULTS
   Date: Jul 5 , 2005
   Ambient condition: Temperature 24.8°C; Relative humidity: 57%
            Body Simulating Liquid
                                               Parameters          Target   Measured   Deviation[%] Limited[%]
      f (MHz)      Temp. [°C]   Depth (cm)
                                               Permitivity:        52.70     51.10        -3.04             ±5
      2450.00          23.70      15.00
                                              Conductivity:        1.95      1.91         -2.05             ±5




                                                        Page 16                                              Rev. 00


                  Compliance Certification Services Inc.
                  Report No: 50630102-SF                                          Date of Issue: Jul 5, 2005



8.3 EUT TUNE-UP PROCEDURES
The following procedure had been used to prepare the EUT for the SAR test.
o   The client supplied a special driver to program the EUT, allowing it to continually transmit the
    specified maximum power and change the channel frequency.
o   The conducted power was measured at the high, middle and low channel frequency before and after
    the SAR measurement.
o   The output power(dBm) we measured before SAR test in different transition rate and channel


                               PK                     AV

           2412               15.16                  12.49

           2437               15.31                  12.77

           2462               15.20                  12.58




                                                    Page 17                                         Rev. 00


                   Compliance Certification Services Inc.
                   Report No: 50630102-SF                                               Date of Issue: Jul 5, 2005



8.4 SAR MEASUREMENTS RESULTS

EUT Setup Configuration 1




 802.11b (DSSS): Duty Cycle =100 %, Rate=1M, Crest Factor: 1                      Depth of liquid: 15.0 cm
                                                *Conducted                      Liquid
                                 Frequency                                                    SAR        Limit
 Sep. [mm] Antenna Channel                      Power dBm                       Temp
                                   [MHz]                                                     (W/kg)     (W/kg)
                                              Before      After                  [°C]

     0          Main             6             2437            15.31    15.30    23.7          0.329        1.6

 Notes:
      1.   *: Peak power.
      2.   Front panel in parallel with flat phantom.
      3.   See attachment for the result presentation in plot format.




                                                              Page 18                                     Rev. 00


                   Compliance Certification Services Inc.
                   Report No: 50630102-SF                                               Date of Issue: Jul 5, 2005




EUT Setup Configuration 2




 802.11b (DSSS): Duty Cycle =100 %, Rate=1M, Crest Factor: 1                      Depth of liquid: 15.0 cm
                                                *Conducted                      Liquid
                                 Frequency                                                    SAR        Limit
 Sep. [mm] Antenna Channel                      Power dBm                       Temp
                                   [MHz]                                                     (W/kg)     (W/kg)
                                              Before      After                  [°C]

     0           main            6             2437            15.31    15.30    23.7          0.536        1.6

 Notes:
      1. *: Peak power.
      2. Panel Bottom in parallel with flat phantom.
      3. See attachment for the result presentation in plot format.




                                                              Page 19                                     Rev. 00


                       Compliance Certification Services Inc.
                       Report No: 50630102-SF                                            Date of Issue: Jul 5, 2005



10.EQUIPMENT LIST & CALIBRATION STATUS
                                                                                         Calibration    Calibration
            Name of Equipment            Manufacturer      Type/Model    Serial Number
                                                                                         Cycle(days)       Due
S-Parameter Network Analyzer           Agilent           E8358A         US40260243       365           05/24/05

Electronic Probe kit                   Hewlett Packard   85070D         N/A              N/A           N/A

Power Meter                            Boonton           4531           13061            365           07/12/05

Power Sensor                           Boonton           56218          2240             365           07/12/05

Thermometer                            Amarell           4046           25060            3650          10/02/14

Thermometer                            Amarell           4046           25058            3650          10/02/14

Wireless Communication Test Set        Agilent           E5515C         GB44051665       365           07/23/05

Universal Radio Communication Tester   Rohde & Schwarz CMU 200          1100.0008.02     N/A           N/A

Signal Generator                       Agilent           83630B         3844A01022       365           01/14/05

Signal Generator                       Agilent           E8257C         US42340383       365           08/01/05

Amplifier                              Mini-Circuit      ZVE-8G         N/A              N/A           N/A

Amplifier                              Mini-Circuit      ZHL-1724HLN    N/A              N/A           N/A

DC Power generator                     ABM               8301HD         N/A              N/A           N/A

Data Acquisition Electronics (DAE)     SPEAG             DAE3           427              365           03/15/05

Data Acquisition Electronics (DAE)     SPEAG             DAE3           558              365           08/24/05

Dosimetric E-Field Probe               SPEAG             ET3DV6         1763             365           03/23/05

Dosimetric E-Field Probe               SPEAG             EX3DV4         3554             365           11/19/05

835 MHz System Validation Dipole       SPEAG             D835V2         4d015            730           12/09/06

900 MHz System Validation Dipole       SPEAG             D900V2         179              730           05/12/06

1800 MHz System Validation Dipole      SPEAG             D1800V2        2d062            730           03/17/06

1900 MHz System Validation Dipole      SPEAG             D1900V2        5d056            730           12/09/06

2450 MHz System Validation Dipole      SPEAG             D2450V2        728              730           03/22/06

5G MHz System Validation Dipole        SPEAG             D5GHz          1004             730           10/04/05

Probe Alignment Unit                   SPEAG             LB (V2)        348              N/A           N/A

Robot                                  Staubli           RX90B L        F02/5T69A1/A/01 N/A            N/A

SAM Twin Phantom V4.0                  SPEAG             N/A            N/A              N/A           N/A

Devices Holder                         SPEAG             N/A            N/A              N/A           N/A

Head/ Muscle 835 MHz                   CCS               H/M 835A       N/A              N/A           N/A

Head/ Muscle 900 MHz                   CCS               H/M 900A       N/A              N/A           N/A

Head/ Muscle 1800 MHz                  CCS               H/M 1800A      N/A              N/A           N/A

Head/ Muscle 1900 MHz                  CCS               H/M 1900A      N/A              N/A           N/A

Head/ Muscle 2450 MHz                  CCS               H/M 2450A      N/A              N/A           N/A

Head/ Muscle 5800 MHz                  CCS               H/M 5800A      N/A              N/A           N/A




                                                           Page 22                                           Rev. 00


                Compliance Certification Services Inc.
                Report No: 50630102-SF                                               Date of Issue: Jul 5, 2005




11.FACILITIES
  All measurement facilities used to collect the measurement data are located at
        No. 81-1, Lane 210, Bade Rd. 2, Luchu Hsiang, Taoyuan Hsien, Taiwan, R.O.C.
        No. 11, Wu-Kung 6 Rd, Wu-Ku Hsiang, Wu-Ku Industrial District, Taipei Hsien, (248)
        Taiwan.
        No. 199, Chunghsen Road, Hsintien City, Taipei Hsien, Taiwan, R.O.C.
12.REFERENCES
  [1]   Federal Communications Commission, \Report and order: Guidelines for evaluating the environ-mental
        effects of radiofrequency radiation", Tech. Rep. FCC 96-326, FCC, Washington, D.C. 20554, 1996.
  [2]   David L. Means Kwok Chan, Robert F. Cleveland, \Evaluating compliance with FCC guidelines for
        human exposure to radiofrequency electromagnetic fields", Tech. Rep., Federal Communication
        Commision, O_ce of Engineering & Technology, Washington, DC, 1997.
  [3]   Thomas Schmid, Oliver Egger, and Niels Kuster, \Automated E-_eld scanning system for dosimetric
        assessments", IEEE Transactions on Microwave Theory and Techniques, vol. 44, pp. 105{113, Jan. 1996.
  [4]   Niels Kuster, Ralph K.astle, and Thomas Schmid, \Dosimetric evaluation of mobile communications
        equipment with known precision", IEICE Transactions on Communications, vol. E80-B, no. 5, pp.
        645{652, May 1997.
  [5]   CENELEC, \Considerations for evaluating of human exposure to electromagnetic fields (EMFs) from
        mobile telecommunication equipment (MTE) in the frequency range 30MHz - 6GHz", Tech. Rep.,
        CENELEC, European Committee for Electrotechnical Standardization, Brussels, 1997.
  [6]   ANSI, ANSI/IEEE C95.1-1992: IEEE Standard for Safety Levels with Respect to Human Exposure to
        Radio Frequency Electromagnetic Fields, 3 kHz to 300 GHz, The Institute of Electrical and Electronics
        Engineers, Inc., New York, NY 10017, 1992.
  [7]   Katja Pokovic, Thomas Schmid, and Niels Kuster, \Robust setup for precise calibration of E-_eld probes
        in tissue simulating liquids at mobile communications frequencies", in ICECOM _ 97, Dubrovnik,
        October 15{17, 1997, pp. 120{124.
  [8]   Katja Pokovic, Thomas Schmid, and Niels Kuster, \E-_eld probe with improved isotropy in brain
        simulating liquids", in Proceedings of the ELMAR, Zadar, Croatia, 23{25 June, 1996, pp. 172{175.
  [9]   Volker Hombach, Klaus Meier, Michael Burkhardt, Eberhard K. uhn, and Niels Kuster, \The dependence
        of EM energy absorption upon human head modeling at 900 MHz", IEEE Transactions onMicrowave
        Theory and Techniques, vol. 44, no. 10, pp. 1865{1873, Oct. 1996.
  [10] Klaus Meier, Ralf Kastle, Volker Hombach, Roger Tay, and Niels Kuster, \The dependence of EM energy
       absorption upon human head modeling at 1800 MHz", IEEE Transactions on Microwave Theory and
       Techniques, Oct. 1997, in press.
  [11] W. Gander, Computermathematik, Birkhaeuser, Basel, 1992.
  [12] W. H. Press, S. A. Teukolsky,W. T. Vetterling, and B. P. Flannery, Numerical Recepies in C, The Art of
       Scientific Computing, Second Edition, Cambridge University Press, 1992..Dosimetric Evaluation of
       Sample device, month 1998 9
  [13] NIS81 NAMAS, \The treatment of uncertainity in EMC measurement", Tech. Rep., NAMAS Executive,
       National Physical Laboratory, Teddington, Middlesex, England, 1994.
  [14] Barry N. Taylor and Christ E. Kuyatt, \Guidelines for evaluating and expressing the uncertainty of NIST
       measurement results", Tech. Rep., National Institute of Standards and Technology, 1994. Dosimetric
       Evaluation of Sample device, month 1998 10




                                                     Page 23                                           Rev. 00


             Compliance Certification Services Inc.
             Report No: 50630102-SF                   Date of Issue: Jul 5, 2005



13.ATTACHMENTS
 Exhibit Content

   1     System Performance Check Plots

   2     SAR Test Plots




                                      END OF REPORT




                                           Page 24                      Rev. 00



Document Created: 2005-09-07 11:04:50
Document Modified: 2005-09-07 11:04:50

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC