SAR report part 3

FCC ID: MSQX01BDA

RF Exposure Info

Download: PDF
FCCID_4077683

       Add:81 Xueyuan Road,Hsidan Distvics ReJig, 100191 China
       Ter ie nemoreians        hi iditamdisizastt
       Tmal: ttichinatlcom      hip/wnctinatien

 Appendix (Additional assessments outside the scope of CNAS L0570)

 Antenna Parameters with Head TSL

     Impedance,ranstormed to feed point                                so00+ saon
     Returm Loss                                                        «23608

 Antenna Parameters with Body TSL

     Impedance, transtormed to feed point                             «seove 110
     Retum Loss                                                        «22808                        _]

 General Antenna Parameters and Design

     Electical Delay(one drecton)                                        1087 ns


 After long term use with 10OW radiated power, only a slight warrming of the dipole near the feedpoint can
 be measured.
 The dipole is made of standard semiigid coaal cable. The center conductor of the feeding line is diectly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the posiion as explained in the *Measurement Conditons" paragraph. The SAR data are not
 affected by this change. The overalldiole length is stll according to the Standard.
 No excessive force must be applied to tdipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.

 Additional EUT Data

    Manufactured by                                                         sreis




Conificate No: 21707262                       Page s ofs


           r"                   in Colabotion h
           -= TTL               a
                                causramon taporatory
         Add No S1 Xueywan Rond sidan Distict bling 10019 China
         Tes einamiorenain)         ic asinezmoisnzastt
         Eul eilichinulcom          hiphnwwchinarton

 DASYS Validation Report for Head TSL                                   Dare: 12062017
 Test Laboratory: CITL, Beijing, China
       : Dipole 1900 MHz; Type: D1900V2; Serial: DI900V2 — SN: 5d1 18
     Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cyele:1:1
     Medium parameters used: F= 1900 MH; a = 1.409$/m; er =39.36; p = 1000 kg/m3
     Phantom section: Center Section
     Measurement Standard: DASYS (IEEE/TEC/ANSI C63.19—2007)
 DASYS Configuration:
             +   Probe: EX3DV4 — SN3617; Conv(8.26, 8.26, 8.26); Calibrated: 1/23/2017;
             +   Sensor—Surface: 1. Amm (Mechanical Surfice Detection)
             +   Electronics: DAE3 Sn536; Calibrated: 1092017
             +   Phantom:     Triple Flat Phantom 5.1 ype: QD 000 PS1 CA; Serial: 1161/1
             +   Mcasurement SW DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                 (A17)

         System Performance Check/Zoom Sean (7x7x7) (7x7x7)/Cube 0: Measurement grid:
         dxcSmm, dy=Smm,         mm
         Reference Value= 101.5 Vim; Power Drif = 0.01 dB
         Peak SAR (extrapolated) = 19.1 Wikg
         SARCL g) = 10 W/kze: SAR(IO g)         19 We
         Maximum value of SAR (measured)         156 Wike

            «8
            o

            —3.56

            432

            n.a7

            463

            a8.29
                         L.
                     0 4B = 15.6 Wikg = 11.93 dBW/kg




Conificate No: 217.974                         Page s ors


        Add: No31 Xvevuan oad, aidan Distict Beiing, 100191 China
        tat «te—Inazsons—2070       Pncoi1noz30t6332504
        —mil:ctlchinatlcom          hipnomchinatlen




        Impedance Measurement Plot for Head TSL

         Te an tepagooo war crovan TT
          29 ror mm ce caremas
           w00
           100
           10.00
           .00
           a.oooy                                                   «
          ce
          .0
          co.ee
          ca.co
          ce
         umm sis seten Cseate 1.c00s (rs on
          m rsooooe on a0.0000 6 iom a u}yfl\’_‘\




Conificate No: 21797262                          Page ors


                          *     in Calsboridonvith
        =_ TTL a
       Add: No 1 Xueyuan Road, aldan Disvics Deing, 100191 China
       Telsene—ioazsten3200)       recomiosmoiemastt
       E—nalt ctichinatlcom        hip hnewcchiatten


DASYS Validation Report for BodyTSL                                         Dare: 12062017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 1900 MHz; Type: DI900V2; Serial: D1900V2 — SN: Sd1 18
   Communication System: UID0, CW;      Frequency: 1900 MHz; Duty Cyele: 1:1
   Medium parameters used:f= 1900 MHz; & = 1.542 $/m; c, = 52.89; p = 1000 kg/m‘
   Phantom section: Left    Section
   Measurement Standard: DASYS (1 MEC/ANSI C63.19—2007)
DASYS Configuration:
          + Probe: EX3DV4 — SN3617; ConvF(Z.95, 7.95, 7.95);       Calibrated: 1/23/2017;
          +    Sensor—Surface: 1 Amm (Mechanical Surface Detection)
          *    Electronics: DAE3 Sn$36; Calibrated: 1092017
          + Phantom: Triple Flat Phantom 5.1C; Type: QD 000 PS1 CA; S
          + Measurement SW DASY52, Version 52.10 (0); SEMCADX Vers
            i)
      System Performance Check/Zoom Sean (7«7x7) (7x7x7)Cube 0: Measurement grid:
      dc=Smm, dy=Smm, dz=Sm
      Reference Value =96.27 Vm; Power Drif = 0.01 dB
      Peak SAR (extrapolated) 18.5 W/kg
      SAR(T ) =10.2 W/kg SAR(IO g) =5.3 Wikg
      Maximumvalue of SAR (measured) = 15.5 Wke




                   0 4BB = 155 W/kg= 11.90 dBWikg

Contfate No: 1797262                           Pageofs


        AddNo1 Xueyoan owd, aidan iss Hoiing, 100191 China
        Teb olocasoieisamh      h e insmiotetsastt
        Bonalt cttachinatlsom   MWip/Aianschinatlon

        Impedance Measurement Plot for Body TSL


         Ti stt toe was 15. o0007 ser t.oooss [
          190 e       m             en e
           m
           Sese
           se
           was
           s.cooy                                            x

          —ese
          —a.co
              s                                          a
         omm six seich (@ep) seute .co0s ts on
          a ce on cname taom o m.W\
                               /




Conificate No: 217—07262                    Parssors


                     TTL p_e s\ m BX
             AddNoSH
             N       tasom fondHaidon"erememtnian
                memtanen              Dtiic, Hoying 100091Chine               m
             Intsgaten Sifoidneds
      Client           Sporton                                      Certificate No:         217—97263
CALIBRATION CERTIFICATE

Object                                  24502 — SN; 840

Calibration Procedure(s)                mmmame
                                        Calibration Procedures for dipole validation kits
Caltbration date:                       December 7, 2017
This calibration Certficate documents the traceabilty to national standards, which realize the physical units of
measurements(S|). The measurements and the uncertaintes with confidence probabilty are given on the following
pages and are partofthe certficate.
All ealbtions have been conducted in the closed laboratory facilty: environment temperature@zs3)tc and
humidty<70%.
Callbration Equipment used (M&TE criical for callbration)

Primary Sendards              o#                  Gal Date(Callorated by. Gertiicate No )      ___Scheduled Gabration
 Power Meter NRVD             102196             O2—Mar—17 (CTTL NoJ17x01254)                           Mar18
 Power sensor       NRVZ5     100506             02—Mar—17 (CTTL, No J17X01254)                         Mard8
 Reference Probe EX3DV4       SN 3617            23—Jan—17(SPEAGNoEX3—3617_Jant?7)                      Jan—18
 oaes                         sh sse             00—0017(CTT—SPEAGNo.217—07108)                         octi8
 Secondary Standards          D#                 Gal Date(Calibrated by. Certficate No.)       Scheduled Callbration
 Signal Generator E4436G MYA9071430 13—Jan—17 (CTTL NoJ17X00280)                                        Jan—ts
 Natwork Analyzer ESO71G      MY4G110673         13—Jan—17 (CTTL, NoJ17X00285)                          Jan—t8



                               Name                      Function                                 Signature
Caltrated by:                 Zhao Jing                SAR Test Enginoer                          41
Reviewed by:                  Lin Hao                  SAR Test Engineer                         xfii}: y
Approved by:                  Oi Dianyuan              SAR Project Leader                    —fovg___
                                                                           Issued: December 10, 2017
This callbraton certficate shall not be reproduced exceptinfull wihout writen approval ofthe laboratory.

    Cenificate No: 21797263                           Page i ofs


         Adds51 Xueywn Rond.Haidan Davic, Heling, 100191, China
         reneoanounan            ns meloemoimsasnt
         Enal: aiiichiut.con        ipshchnatien
Glossary:
TSL                    tissue simulating liquid
Cone                    sensitvity in TSL / NORMxy.z
N/A                    not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
    Spatial—Averaged Specific Absorption Rate (SAR)in the Human Head from Wireless
  Communications Devices: Measurement Techniques®, June 2013
b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
    exposure to radio frequency fields from hand—held and body—mounted wireless
    communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
    6GHz)% July 2016
) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
    communication devices used in close proximity to the human body (frequency rangeof
   30MHz to 6GHz)\, March 2010
d) KDB865G64, SAR Measurement Requirements for 100 MHz to 6 GHz:


Additional Documentation:
e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
*     Measurement Condiions: Further details are available from the Validation Report at the end
      of the certificate. All figures stated in the certficate are valid at the frequency indicated.
*     Antenina Parameters with TSL: The dipole is mounted with the spacer to position its feed
  point exactly below the center marking of the flat phantom section, with the arms oriented
  parallel to the body axis.
+ Feed Point Impedance and Return Loss: These parameters are measured with the dipole
      positioned under the liquid fled phantom. The impedance stated is transformed from the
      measurement at the SMA connector to the feed point. The Return Loss ensures low
      reflected power. No uncertainty required.
+     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
      No uncertainty required.
+     SAR measured: SAR measured at the stated antenna input power.
+     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
*     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result


    The reported uncertainty of measurement is stated as the standard uncertainty: of
    Measurement mulipled by the coverage factor k=2, which for a normal distrbution
    Corresponds to a coverage probabiity of approximately 95%.



Centficate No: 21797263                       Page 2 ofs


                           "      in Cotaborsionwith
                                  a
                                 causmimon Lasoratory
        AddN1 XueyunnRoud. dn Diio. elin, 100191, China
        fet oninemsoreinan   rac meloamotemasit
        E—mal: citichnatcom         hipwwwchinatien

Measurement Conditions
   DASY system confiuration, as for as no gven onpage 1
    DASY version                                 nasyse                                    s2.100.145
    Extrapotation                          Advanced Extrapoiation
     Phantom                               "Tile Flat Phantom 5.1C

     Distance Dipole CenterTSL                    10 mm                                  win Spacer
     Z0om Scan Resolution                     o,0y, tz =5 mm
     Frequency                                2450 Mz 1 e
Head TSL paramoters
   The folowing parameters and calculatons were applied.
                                                   Temperature             Pormitivity          Conductiviy
     NominalHead TSL paramoters                     220°c                    EB                180 mhorm
     Measured Hoad TSL parameters                 gzo02)‘c                  s0810%            183moin£6%
     Head TSL temperature change during test           «10c                    a~                     am
SAR result with Head TSL
     SAR averaged over 1_cm)_(1 g)of Head TSL                   Condton
     SAR measured                                           250 mW noutpower            192mw/g
     SAR for nominal Head TSL parameters                     normalized t 1W        62.mW /g t 188 % (t)
     SAR averaged over 10 on‘ (10g)of Hoad TSL                  Condiion
     SAR measured                                           250 mW nput power            614 mw/g
     SAF fo nominalHead TSL parameters                       normalzed o 1W         24. mW ig 4187 % (ke2)
Body TSL parameters
   ‘he folowing paramaters and calcultons were appled
                                                  Temporature              Porminivity         Conductivity
     Nominal Body TSL paramoters                    z20‘c                     ser               195 mhoim
     Monsured Body TSL parameters                @zos02‘c                   s2526%            189 mhoma6%
     Body TSL temporature change during test           ks                      >«                     _
SAR result with Body TSL
    SAR averaged over 1_cm"_(1g) o Body TSL                     Congtion
     SAR measured                                           250 mW input power               120 mW1g
    SAR fornominal Body TSL. parameters                      nomalized to 1W        61.9 mW i# 18.8 % (ort)

    SAR averaged over 10 cm" (10.) of Body TSL                  Condtion
    SAR measured                                            250 mW nput power             ssomw»o
    SA fornominal Body TSL parametors                        normalized to 1W       24.0 mW 1g 21827 % (ke2)

Cenificae No: Z17—97263                        Page s ofs


                    TL a
        Add No S1 Xueywan RouktidanDivic, Heling, 100191, China
        ter nineanouian           rac sneloentoienastt
        E—mal: ctlchinant com     hopiAnwchinaten

 Appendix (Additional assessments outside the scope of CNAS LO570)
 Antenna Parameters with Head TSL


    TImpedance, ranstormed tofead point                               szon+«sto
     Retum Loss                                                         ~25 308


 Antenna Parameters with Body TSL

     Impedance, transtormed to feed point                             si.10—s000
     Retum Loss                                                         —25800

 General Antenna Parameters and Design

     Electical Delay(one diection)                                      1025 ns


 Afterlong term use with 100W radiated power,only a sight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semniigid coanial cable. The center conductor of the feeding line is directly
 connected to the second arm ofthe dipole. The antenna is therefore short—ircuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in ordertoimprove matching when loaded
 according to the posiion as explained in the "Measurement Conditions" paragraph, The SAR data are not
 affected by this change. The overall dipole length is sil according to the Standard.
 No excessive force must be applied tothe dipole arms, because they might bend orthe soldered
 connections near the feedpoint may be damaged.

 Additional EUT Data

     Manutactured by                                                       sreas




Corificate No: 21797263                       Page 4 ofs


        Add NsS1 XueyuanRont dn Divic, Heling,1091; China
        Tat «se inczsoueas                  oressasoe
        E—mal:ctiichinattcon   iphrewchiatten

DASVS Validation Report for Head TSL                                 Dae: 12062017
Test Laborstory: CTTL, Bcijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450OV2 — SN: 840
   Communication System: UID 0, CW; Fr         50 MHz; Duty Cyele:11
   Medium parameters use                                      ; p= 1000 ka/m3
   Phantom section: Center Section
  Measurement Standard: DASYS (IEEE/TEC/ANSI C63,19—2007)
DASYS Configuration:
           *   Probe: EX3DV4 — SN3617; ConvF(7.74, 7.74, 7.74); Calibrated: 1/23/2017;
           +   Sensor—Surface: 1 Ammm (Mechanical Surface Detection)
           +   Electronics: DAE3 Sn536; Calibrated: 1092017
           +   Phantom: Triple Flat Phantom 5.1    ype: QD 000 PS1 CA; Serial: 1161/1
           +   Measurement SW: DASY32, Version 52.10 (0);       SEMCAD X Version 14.6.10
               (7417)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dc=Smm,
        dy=Smm, dz=Smm
       Reference Value = 106.0 V/m; Power Drift = 0.00dB
       Peak SAR (extrapolated) = 27.3 Whkg
       SARCT g) = 13.2 Wikgs SAR(IO g) = 4 Wig
       Maximum value of SAR (measured) 22.0 Whe




           33

           a.s6

           1209

           arge

           anes |
                   0 dB =22.0 Wikg = 1342 dBWhkg




Cenifcate No: Z17—97263                   Page sors


                        L        in Cotaboration wik


        Add: NoS1 Xueyumn Rout, Maldan
                                   in; Davic,      100191 China
        Mdeibsshenam                 ho meivsonence
        Eonallcnlichinatlcon         hipiIwchinatlen




        Impedance Measurement Plot for Head TSL


         rran mnay mm ver mm Th
           100 anareees e
                        m en
           «.00
          100
          se
          sn.00
          a.ccey                                                  u
         cce                                                      k
         satce
         cce
         casreo                                                       |
          cce                                                 x
         im 5is sntth (130 seate2.co0s ts oel)
          ao rasooes o se n «sior n yfl’\




                                                                  al




Conificae No: 717—97263                       Page cof s


        Add No 1 XueyuaRoad Haidan Diatic, in 10019, China
        tel encinamiunaims       rac abinamoinasns
        Emal: cntchinanlcom      hipiAchiatlen

DASYS Validation Report for Body T                                         Date: 12072017
Test Laboratory: CTTL, Beijing,   China
D
    Communication System: um 0. Cw                             Duty Cyele: 1:1
    Mediumparameters      used   2450M          1.926 S/     =52.48; p = 1000 ka/m?
    Phantom section: Left Section
    Measurement Standard: DASYS (IEEE/IE       ANSI C63.19—2007)
DASY5 Configuration:

           *    Probe: EX3DV4 — SN3617; ConvB(7.8, 7.8, 7.8); Calibrated: 1/23/2017;
            +   Sensor—Surface: 1 Amm (Mechanical Surface Detection)
           * Electroni    AE3 Sn536; Calibrated: 10/9/2017
           + Phantom: Triple Flat Phantom5.1    ype: QD 000 PS1 CA; Serial: 1161/1
           + Measurement SW: DASY32, Version 52.10 (0) MCAD X Version 14.6.10
             (aim)
       Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dxc—Smm,
       dy—Smm, de=Smm
       Reference Value = 99.77 V/m; Power Drift = 0.02 dB
       Peak SAR (extrapolated) = 26.8 W/kz
       SAR(T g) = 12.9 Wikg: SAR(TO g) 5.99 We
       Maximum value of SAR (measured) = 21.7 Wke




           130

           ase

           12.09

           a7ag

           2148
                    0 dBt =21.7 Wikg = 13.36 dBWhkg




Conificate No: 21797263                    Page 7 ofs


                           J   inColibontonwith

                             causmmon tapomarory
        Ad No1 Xuenn Road,Haidan Disvict Neling, 10019China
        Tolon inamsostsann       Tc ditiroitzsaset
        Hmall aiiichinatlcon     hipwowchinatlon

        Impedance Measurement Plot for Body TSL


         r e ty wag to: oo wer o. oo Te
          ©9 mm              am:
          «.00
           100
           ro.00
           sn0o
           s.ooom                                             k
         ss.co                                                4
          m.so
          se
         —asioo
          n00                                             +
         um sis snich (heps seate ccows t o)
          m nssm oe mome com e ne.
                                           /




Conificate No: 217—97263                    Page ofs


            ssm
            ds No S1 Xueyuan
                                     r—se‘as _ T/
                             Rondsidan Disvct Being 10019 Chin
                                                                                                             dhg uoy
                                                                                                             CABRATION


      Client           Sporton                                        Certificate No:        217—97255
CALIBRATION CERTIFICAT

Object                                   ©2002 SN: 1061

Callbration Procedure(s)                 FzH—000.01
                                         Callbration Procedures for dipole validation kits

Callbration date:                        December 7, 2017

This caltbation Cortficate documents the traceabilty to national standards, which realize the physical units of
measurements(S|). The measurements and the uncertainties with confidence probabity are given on the following
pages and are part of the certficate.
All caltrations have been conducted in the closed laboratory facity: environment temperature@z2s3)c and
humidty<70%.
Calibration Equipment used (M&TE criical forcallbration)

Prmary Standards               1#                 Gal Date(Calibrated by. Gertfieate No.)       Scheduled Calibration
 Power Meter NRVD              102196             02Mar—17 (CTTL, No.J17X01254)                          Mar18
 Power sensor NRVZ5            100506             02:Mar—17 (CTTL, NoJ17X01254)                          Mart8
 Reference Probe EX3DV4        SN 3617            23—Jan—17(SPEAGNo EX3—3617_.Jant7)                     Jan—18
 DAE3                          Sh sas             08—Oct—17(CTTL—SPEAGNo217—97198)                        Oct18

 Secondary Standards           1#                 Cal Date(Calibrated by. Gertficate No.)        Scheduled Calbration
 Signal Generalor E438C        MY49071430         13—Jan—17 (CTTL, No.J17X00286)                        Jan—18
 Network Analyzer ESO71C MY46110673 13—Jan—17 (CTTL, No.J17X00285)                                       Jan—t8

                                Name                       Function                                  Signature
potbrated by:                  Zhao Jing                SAR Test Engineer                           bt
Reviewed by:                   Lin Hao                  SAR Test Engineer                         Tfi‘f. #>
Avenstiltby:                   Oi Dianyuan              SAR Project Leader                     _#A
                                                                             Issued: December 10, 2017
This calbration certficate shall not be reproduced except in full wihout writen approval ofthe laboratory.


    Cenificate No: Z17—07258                            Page i ofs


         ‘TTL Ll_.!:L!.__
        Addo1 Xueyuan Road,Haidan itc Reli100191 China
        Tek ote—iemsoreians     rax ie ioamoicizastt
        E—maltctachinanlcom      hipchiatlen
Glossary:
TSL                   tissue simulating liquid
ConvF                 sensitiity in TSL/ NORMxy,z
NA                   not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Sid 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
  Communications Devices: Measurement Techniques, June 2013
b) IEEC 62200—1, "Measurement procedure for assessment of specific absorption rate of human
   exposure to radio frequency fields from hand—held and body—mounted wireless
    communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
    6GHz)®, July 2016
c IEC 62200—2, ‘Procedure to measure the Specific Absorption Rate (SAR) For wireless
    communication devices used in close proximity to the human body (frequency range of
   3OMHz to GGHz)\, March 2010
d) KDBB5G4, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parametors:
*    MeasurementConditions: Further details are available from the Validation Report at the end
  of the certficate. All figures stated in the certiicate are valid at the frequency indicated.
+ Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle!to the body axis.
+ Feed Point Impedance and Retum Loss: These parameters are measured with the dipole
     positioned under the liquid filed phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
*    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
+ SAR measured: SAR measured at the stated antenna input power.
+    SAR normalized: SAR as measured, normalized to an input powerof 1 W at the antenna
     connector.
*    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result


    The reported uncertainty of measurement is stated as the standard uncertainty of
    Measurement multiplied by the coverage factor k=2, which for a normal distrbution
    Corresponds to a coverage probabilty of approximately 95%.

Cenificat No: 217.9725                      Page a ofs


        Add NoS1 Xveua Roud, Haidan Distic Reing, 100191, hina
        Toh aeanamsoiesum n w incroimasis
            ib ertchinantcom      hipotcchinatl en
Measurement Conditions
   DASY aystem contiquration,asfar s not iven on page 1
     DASY Version                                   nasvse                                   521001418
     Extrapolation                          Advanced Extrapolation
     Phantom                                "Tisle Fat Phantom 5.1C
     Distance Dipole Center— TSL                    10 mm                                  win Spacer
     Zoom Sean Resolution                      x,dy. cz =$ mm
     Frquency                                  2600 Mrics 1 it
Hoad TSL parametors
   ‘The folowing parameters and calculations were
                                              appled.
                                             Tomperature                     Pormitivity         Conductiviy
     Nominal Head TSL paramotors                zo‘                             se0               18 mhotm
     Measured Hoad TSL parametors            @zo02)c                          30426%            109 mhoin£6%
     Head TSL tomporature change during test    <1.0°C                           —                   e
SAR result with Head TSL
   SAR averaged over 1_cm_(1g) of Head TSL                        Coniion
     SAR measured                                           250 mW nput power               148 mW/q
     SAR for nominal Head TSL parameters                     normalized t 1W          $8.2 mW /q 188 % (ke2)
     SAR averaged over 10 cm" (10 ) of Hoad TSL                   Condtion
     SAR measured                                           250 mW nput power               s50mw19
     SA fornominal Head TSL paramoters                       nomalized t 1W           26.0 mW ig 187 % (ket)
Body TSL parameters
   Thefolowing parameters and calculatons were appled
                                               Temperature                   Porminivity         Conductivity
     Nominal Body TSL parametors                 rmo‘c                          sas              218 mhaim
     Measured Body TSL paramaters             @zoz02)°C                       s2626%            |213moma6%
     Body TSL temperature change during test|    <1.0 °C                         zs                  ==
SAR result with Body TSL
     SAR averaged over 1_cm"_(1g) o Body TSL                      Goniion
     SAR measured                                           250 mW input power                 140mw19
     SAR fornomnal Body TSL parameters                          normalized t 1W       50.4 mW q 2 18.8 % (cet)
     SAR averaged over 10 en (10.g)or Body TSL                  Condition
     SAR measured                                           250 mW inout power             623 mW/o
     SA# fornominal Body TSL parametars                      normatized to 1W         280 mW ig 187 % (e2)

Conificate No: 217.97255                            Pages ors


                    TL               in Coaboraion wit


       Add: No 1 Xueyua Road, Haidan Ditit Belig, 100191 China
       To iosmiemans               rec asiamoizasts
       Eemail: ctlchinatlcom       hapihwectinaten

 Appendix(Additional assessments outside the scope of CNAS LO570)
 Antenna Parameters with Hoad TSL

     Impedance, ransformed to fed point                              «240—6050
     Retum Loss                                                       «24308

 Antenna Parameters with Body TSL

     Impedance,ranstormed to foed point                              «s20—5.100
     Retum Loss                                                        «23500

 General Antenna Parameters and Design
     Electical Dotay(one diection)                                     1013ns


 Afterlong term use with 100W radiated power, only a sight warming ofthe dipole near the feedpoint can
 be measured.

 The dipole is made of standard semiigid coaxial cabl. The center conductor of the feeding line is drectly
 connected to the second arm ofthe dipole. The antenna is therefore short—cirouited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order t improve matching when loaded
 according to the posiion as explained in the *Measurement Conditons" paragraph. The SAR data are not
 affected by this change. The overal diole length is stl according to the Standard
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.

 Additional EUT Data

    Manufactured by                                                       seers




Cenificae No: 217—97255                           Page 4 ors


                               in Cotuboreion wth
               TTL a
                   cAusmanion taporarory
       Add NoS1 Xueyuan Rowd Haidan Distic, Nng 10019, China
       Tot oniGeasousamm)         mc s inamoiennastt
       E—mal: ctltchinantcom      hihvcchinatlon

DASYS Validation Report for Head TSL                                       Date: 12.07.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2600 MHz; Type: D2600V2; Serial; D260OV2 — SN: 1061
   Communication System: UID 0, CW                                       ds
   Medium parameters us         2600 MHz; a = 1.985 S/m; er=39.42; =     1000 ke/im3
   Phantomsection: Center Section
   Measurement Standard: DASYS (1 MEC/ANSI Cé3.19—2007)
DASY5 Configuration:
           *   Probe: EX3DV4 — SN3617; ConvB(Z.3, 7.3, 7.3); Calibrated: 1/23/2017;
           + Sensor—Surface:1.4mm (Mechanical Surface Detection)
                            AE3 Sn536; Calibrated: 101972017
                      : Triple Flat Phantom5.1C; Type: QD 000 PS1 CA; Serial: 1161/1
           + Measurement SW DASY52, Version 52.10 (0); SEMCADX Version 14.6.10
               (rai)

       Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)Cube 0: Measurement grid: dx=Smm,
       dy=Smm, dz=Smm
       Reference Value= 107.7 V/m; Power Dri= 0.00dB
       Peak SAR (extrapolated) = 31.5 Wikg
       SAR(L g) = 14.6 Wkgs SAR(IO g)       Why
       Maximum value of SAR (measured) 25.1 Wike




          164

          .29

          1303

          «10.50


          32e                                   =
                   0 dBB =25.1 Wikg = 14.00 dBW/kg




Cenifieate No: 2179725                       Pages ofs


                                  in Cotabormtion wih

                                 causmmon asoratory
        Add: No1 Xueyuan Road, aidan Distics Neling, 100191 Chna
        Teloonieaaoteisarnd rac eioamoienastt
        E—nall ctashinatcom        hipAmchinatlen



        Impedance Measurement Plot for Head TSL


         Te itt ty ray To: oo war ac ooo Triy
          100 nmeene
          «se
           100
           100
          aniee
           s.0009
          00
          c
          cce
          c
           .w                                                  x
            sin seith (90 seates.000 ts ort)
          m ns on esn e sn in




Cenifieate No: 217.97255                        Page 6ofs


                           *   inColibontonnith
                TTL a          causmnon taporarory




DASYS Validation Repor         r Body TSL                                  Date: 12072017
Test Laboratory: CTTL, B       s, China
DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D260OV2 —       1061
   Communication System: UID 0, CW; Frequency: 2600 MHz; Duty Cycle:1:1
   Medium parameters us          2600 M           127 S/m £, = 52.63; p= 1000 kg/e‘
  Phantomsection: Left Section
  Measurement Standard: DASYS (IE              ANSI C63,19—2007)
DASY5 Configuration:
            +   Probe: EX3DV4 — SN3617; ConvF(7.48, 7.48, 7.48); Calibrated: 1/23/2017;
            +   Sensor—Surface: 1.Amm(Mechanical Surface Detection)
            + Electronics: DAE3 $n§36; Calibrated: 1092017
            + Phantom:         lat Phantom 5.1C; Type: QD 000 PS1 CA; Serial: 1161/1
            + Measurement SW DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
              @ain
        Dipole Calibration/ZoomSean (7x7x7) (Zx7x7)Cube 0: Measurement grid: da=Smm,
        dy=Smm, da=Smm
        Reférence Value = 96.43 V/m; Power Drift =0.03 dB
        Peak SAR (extrapolated) = 30.0 Wz
        SARQ g) =14 W/kkg: SAR(IO g) 623 Wikg
        Maximum value ofSAR (measured) = 23.8 Wkz


           «8
           a

           ara

           448

           423

           10.48

           aare
                      &
                    0 dBB =23.8 Wikg = 13.77 dBW/ke




Cenificate No: 217—9725s                    Page ? ofs


                           *   inCofibontionnith
               TTL a
       Add: No1 Xueyian ood, aidan Diseic Neing 100191 Chia
       To ce—osnnienammy         hec usinemoizaset
       E—mal: ctitchinattcom     MipiAmctination
       Impedance Measurement Plot for Body TSL


         Te srcas wan 16reom7 wer acecoms T
          20 re—remme on     mm ce
          ss.00
          woe
          w0
          in.0o




          o.0e                                    x
         m 5i2 sntch Gups) seate 2.0000 (rs ont1
         » aciosme oc ssm n cnasis e airuie_    ¥
                                            /                 \   |




        n




Corificate No: 217.9725$                    Page t ofs


                                                                                    U Jn
                                        in Collaboration with                    sfi‘&//./"'», A t BBiAaJ

                                                                                    ons                           CALIBRATION
            Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China     "'{\‘
                                                                                    is         v CNAS LO570
            Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504               stt
            E—mail: cttl@chinattl.com        Http://www.chinattl.cn

      Client :        Auden                                                     Certificate No: 218—60107


           RationceRtiFIcATE _
Object                                    DAE4 — SN: 1305

Calibration Procedure(s)                  FF—Z211—002—01

                                          Calibration Procedure for the Data Acquisition Electronics
                                          (DAEx)
Calibration date:                         May 11, 2018

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(Sl). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)*c and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)


Primary Standards                ID #          Cal Date(Calibrated by, Certificate No.)       Scheduled Calibration



Process Calibrator 753        1971018              27—Jun—17 (CTTL, No.J17X05859)                     June—18




                                Name                       Function                               Signature
Calibrated by:                   Yu Zongying              SAR Test Engineer                      %

Reviewed by:                     Lin Hao                  SAR Test Engineer                      fii{fi%
Approved by:                     Qi Dianyuan              SAR Project Leader               %

                                                                                           Issued: May 12, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




     Certificate No: Z18—60107                                  Page 1 of 3


        (TTL spea g
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
       E—mail: cttl@chinattl.com      Http:/www.chinattl.on

Glossary:
DAE                           data acquisition electronics
Connector angle                information used in DASY system to align probe sensor X
                              to the robot coordinate system.


Methods Applied and Interpretation of Parameters:
e   DC Voltage Measurement: Calibration Factor assessed for use in DASY
    system by comparison with a calibrated instrument traceable to national
    standards. The figure given corresponds to the full scale range of the
    voltmeter in the respective range.

e   Connector angle: The angle of the connector is assessed measuring the
    angle mechanically by a tool inserted. Uncertainty is not required.

e   The report provide only calibration results for DAE, it does not contain other
    performance test results.




Certificate No: Z18—60107                               Page 2 of 3


                                   in Collaboration with

               T"TL s_p e a g
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504
       E—mail: cttl@chinattl.com        Http:/www.chinattl.on



DC Voltage Measurement
   A/D — Converter Resolution nominal
       High Range:       1LSB =       6.1uV ,  full range =   —100...+300 mV
       Low Range:        1LSB =        61inV , full range =   ~Arseerrs +3mV
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


    Calibration Factors                      X                             Y                 2Z
     High Range                    403.659 + 0.15% (k=2) 403.993 + 0.15% (k=2) 404.315 + 0.15% (k=2)
     Low Range                     3.98260 + 0.7% (k=2)         3.99157+0.7% (k=2)   3.99746 +0.7% (k=2)



Connector Angle

    Connector Angle to be used in DASY system                                             97°%1*°




Certificate No: Z18—60107                                  Page 3 of 3


                                                                                         asth®
                                                                                           |   14



                      TTL ssesns BemCNAS ::
                                    ~    in Collaboration with                      \‘\$               A aP BM iA hJ
                                                                                    &                                   RBHHAWV
                                                                                    '.
                                                                                     o {'
                                                                                        R _s‘ v CALIBRATI
                                                                                                          ON
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China          /,,,
                                                                                           | \‘.\
             Tel: +86—10—62304633—2512                                                                                  CNAS LO570
                                                Fax: +86—10—62304633—2504                  um
             E—mail: cttl@chinattl.com          Http://www.chinattl.cn
      Client               Sporton                                           Certificate No:        2Z18—60364
\CALIBRATION CERTIFICATE
Object                                     EX3DV4 — SN:3843

Calibration Procedure(s)
                                           FF—Z11—004—01
                                           Calibration Procedures for Dosimetric E—field Probes

Calibration date:                          September 27, 2018

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(Sl). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.

All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)°c and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)
Primary Standards                        ID # Cal Date(Calibrated by, Certificate No.)                   Scheduled Calibration
 Power Meter        NRP2          101919        20—Jun—18 (CTTL, No.J18X05032)                             Jun—19
 Power sensor NRP—Z91             101547        20—Jun—18 (CTTL, No.J18X05032)                             Jun—19
 Power sensor NRP—Z91             101548        20—Jun—18 (CTTL, No.J18X05032)                             Jun—19
 Reference10dBAttenuator          18N50W—10dB    09—Feb—18(CTTL, No.J18X01133)                             Feb—20
 Reference20dBAttenuator          18N50OW—20d4B          09—Feb—18(CTTL, No.J18X01132)                      Feb—20
 Reference Probe EX3DV4           SN 3846                25—Jan—18(SPEAG,No.EX3—3846_Jan18)                 Jan—19
 DAE4                             SN 777                 15—Dec—17(SPEAG, No.DAE4—777_Dec17)                Dec —18


 Secondary Standards              ID #                   Cal Date(Calibrated by, Certificate No.)        Scheduled Calibration
 SignalGeneratorMG3700A           6201052605             21—Jun—18 (CTTL, No.J18X05033)                   Jun—19
 Network Analyzer E5071C          MY46110673             14—Jan—18 (CTTL, No.J18X00561)                   Jan —19
                                Name                           Function                                    S/ignature
Calibrated by:                   Yu Zongying                   SAR Test Engineer                       "Ng       sR

Reviewed by:                     Lin Hao                      SAR Test Engineer

Approved by:                     Qi Dianyuan                   SAR Project Leader

                                                                                         Issued: September 29, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



     Certificate No: Z18—60364                                Page 1 of 11


                              "    in Collaboration with
          %TTL s
               §
                 _p_e _2 _g
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Http://www.chinattl.cn


Glossary:
TSL                   tissue simulating liquid
NORMx,y,z             sensitivity in free space
ConvF                 sensitivity in TSL / NORMx,y,z
DCP                   diode compression point
CF                    crest factor (1/duty_cycle) of the RF signal
A,B,C,D               modulation dependent linearization parameters
Polarization &        4 rotation around probe axis
Polarization 0        6 rotation around an axis that is in the plane normal to probe axis (at measurement center),
                                                                                                                   i
                      0=0 is normal to probe axis
Connector Angle       information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "EEE Recommended Practice for Determining the Peak Spatial—Averaged
   Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
   Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from
   lJalndégciléi and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)",
     uly
c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March
   2010
d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
e   NORMx,y,z: Assessed for E—field polarization 0=0 (fs900MHz in TEM—cell; f> 1800MHz: waveguide).
    NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the
     E —field uncertainty inside TSL (see below ConvF).
e   NORMMx,y,z = NORMx,y,z* frequency_response (see Frequency Response Chart). This
    linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of the
    frequency response is included in the stated uncertainty of ConvEF.
e   DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep
    (no uncertainty required). DCP does not depend on frequency nor media.
e   PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
    characteristics.
e   Ax,yz Bx,y,z; Cx,y,z;VRx,y,z:A,B,C are numerical linearization parameters assessed based on the
    data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
    media. VR is the maximum calibration range expressed in RMS voltage across the diode.
e   ConvrF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
    Transfer Standard for f<800MHz) and inside waveguide using analytical field distributions based on
    power measurements for f >800MHz. The same setups are used for assessment of the parameters
    applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
    These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
    The sensitivity in TSL corresponds to NORMx,y,z* ConvF whereby the uncertainty corresponds to
    that given for Convr. A frequency dependent ConvF is used in DASY version 4.4 and higher which
    allows extending the validity from+50MHz to+100MHz.
e   Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat
    phantom exposed by a patch antenna.
e   Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the
    probe tip (on probe axis). No tolerance required.
e   Connector Angle: The angle is assessed using the information gained by determining the NORMx
    (no uncertainty required).


Certificate No: Z18—60364                             Page 2 of 11



Document Created: 2019-05-08 05:10:15
Document Modified: 2019-05-08 05:10:15

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC