sar system validations

FCC ID: KBCIX325-AC775BT

RF Exposure Info

Download: PDF
FCCID_599975

Celltech




           ———. ammonnc


                                                                            835 MHz System Validation Dipole (Brain)




                                835 MHz SYSTEM VALIDATION DIPOLE



                   Type:                                 835 MHz Validation Dipole




                                                                        411
                   Serial Number:




                                                               Celltech Labs Inc.
                   Place of Calibration:




                   Date of Calibration:                          March 30, 2005




Celltech Labs Inc. hereby certifies that this device has been calibrated on the date indicated above.




                   Calibrated by:




                   Approved by:




                             Celltech Labs Inc. 1955 Moss Court, Kelowna, B.C. Canada V1Y 9L3
                           Tel. 250-448-7047 x Fax. 250-448-7046 x e-mail: info@celltechlabs.com
                                                   www.celltechlabs.com
2005 Celltech Labs Inc.                                                                                       1 of 12


                                                                             835 MHz System Validation Dipole (Brain)




         1. Validation Dipole Construction & Electrical Characteristics

         The validation dipole was constructed in accordance with the IEEE Standard “Annex G (informative)
         Reference dipoles for use in system validation”. The electrical properties were measured using an HP
         8753ET Network Analyzer. The network analyzer was calibrated to the validation dipole N-type
         connector feed point using an HP85032E Type N calibration kit. The dipole was placed parallel to a
         planar phantom at a separation distance of 15.0mm from the simulating fluid using a loss-less dielectric
         spacer. The measured input impedance is:

                   Feed point impedance at 835MHz        Re{Z} = 47.627:
                                                         Im{Z} = -0.67188:

                   Return Loss at 835MHz                 -31.954dB




2005 Celltech Labs Inc.                                                                                        2 of 12


1stand Engneei Sevceas



                                                                     30 man 2ess       dssebiar
                  hE      4 ul fe          4 47.027 0|   ~67.00 ma 200.09 aF          #35.000 000 MHz




                       MPRKE
           tor           83




                         center se5.o00 one                                    seai. <aa.ona ceo itz
                                                                        on mar 2eas 14so5i es
                       nEX     tos     se dsrer o s8                   ai—o0.98e dn a95.000 600 kz



                          MARKER| 1
                 Cor        835




                             ceNTeR ©35.000 000 Miz                              seal a02.a00 000 Mz


                                                                          835 MHz System Validation Dipole (Brain)




         3. Validation Dipole Dimensions


                          Frequency (MHz)            L (mm)           h (mm)             d (mm)

                               300                    420.0           250.0               6.2
                               450                    288.0            167.0              6.2
                               835                    161.0             89.8              3.6
                               900                    149.0             83.3              3.6
                              1450                     89.1             51.7              3.6
                              1800                     72.0             41.7              3.6
                              1900                     68.0             39.5              3.6
                              2000                     64.5             37.5              3.6
                              2450                    51.8             30.6               3.6
                              3000                    41.5             25.0               3.6




         4. Validation Phantom

         The validation phantom is the SAM (Specific Anthropomorphic Mannequin) phantom manufactured by
         Schmid & Partner Engineering AG. The SAM phantom is a Fiberglass shell integrated in a wooden table.
         The shape of the shell corresponds to the phantom defined by SCC34-SC2. It enables the dosimetric
         evaluation of left and right hand phone usage as well as body mounted usage at the flat phantom region.
         A cover prevents evaporation of the liquid. Reference markings on the phantom allow the complete setup
         of all predefined phantom positions and measurement grids by manually teaching three points in the
         robot.

         Shell Thickness:            2.0 ± 0.1 mm
         Filling Volume:             Approx. 25 liters
         Dimensions:                 50 cm (W) x 100 cm (L)




2005 Celltech Labs Inc.                                                                                     4 of 12


Teste an Engneeren Sevices tas


Celltech
 Tasto and EnioerngSevcestas


                                                                            835 MHz System Validation Dipole (Brain)




         7. Measurement Conditions

         The SAM phantom was filled with 835 MHz simulated brain tissue mixture having the following
         parameters:

                   Relative Permittivity:      39.5
                   Conductivity:               0.90 mho/m
                   Fluid Temperature:          20.4 qC
                   Fluid Depth:                t 15.0 cm

                   Environmental Conditions:
                   Ambient Temperature: 20.5 qC
                   Barometric Pressure: 102.2 kPa
                   Humidity:            32 %

         Measurements were made at the planar section of the SAM phantom using a dosimetric E-field probe
         ET3DV5 (S/N: 1590, conversion factor 6.71).

         The 835 MHz simulated brain tissue mixture consisted of the following ingredients:



                                    Ingredient                   Percentage by weight

                                      Water                              40.71%

                                      Sugar                              56.63%

                                        Salt                             1.48%

                                       HEC                               0.99%

                                    Dowicil 75                           0.19%

                          Target Dielectric Parameters                Hr = 41.5
                                    at 22 qC                          V = 0.90 S/m




2005 Celltech Labs Inc.                                                                                       7 of 12


                                                                                  835 MHz System Validation Dipole (Brain)




         8. SAR Measurement

         The SAR measurement was performed with the E-field probe in mechanical detection mode only. The
         setup and determination of the forward power into the dipole was performed using the following
         procedures.




                                                                            3D Probe positioner

                                         d
                                                                    Field probe
                                                                                       Flat Phantom




                                                                                       Dipole




                                                Dir.Coupler
          Signal          Amp    Low    3dB                                        x
          Generator              Pass                             Cable                     Att1
                                        Att3
                                                                                                 PM1

                                                     Att2   PM3



                                               PM2




         First the power meter PM1 (including attenuator Att1) is connected to the cable to measure the forward
         power at the location of the dipole connector (X). The signal generator is adjusted for the desired
         forward power at the dipole connector (taking into account the attenuation of Att1) as read by power
         meter PM2. After connecting the cable to the dipole, the signal generator is readjusted for the same
         reading at power meter PM2. If the signal generator does not allow adjustment in 0.01dB steps, the
         remaining difference at PM2 must be taken into consideration. PM3 records the reflected power from the
         dipole to ensure that the value is not changed from the previous value. The reflected power should be
         20dB below the forward power.




2005 Celltech Labs Inc.                                                                                             8 of 12


                                                                                      835 MHz System Validation Dipole (Brain)




         9. Validation Dipole SAR Test Results

         Ten SAR measurements were performed in order to achieve repeatability and to establish an average
         target value.


                                 SAR @ 0.25W         SAR @ 1W         SAR @ 0.25W              SAR @ 1W
            Validation                                                                                          Peak SAR @
                                Input averaged     Input averaged    Input averaged          Input averaged
           Measurement                                                                                          0.25W Input
                                   over 1g            over 1g           over 10g                over 10g

                Test 1               2.45               9.80                1.60                  6.40             3.65
                Test 2               2.44               9.76                1.59                  6.36             3.66
                Test 3               2.45               9.80                1.60                  6.40             3.67
                Test 4               2.44               9.76                1.59                  6.36             3.64
                Test 5               2.44               9.76                1.59                  6.36             3.62
                Test 6               2.43               9.72                1.59                  6.36             3.61
                Test 7               2.45               9.80                1.60                  6.40             3.65
                Test 8               2.43               9.72                1.59                  6.36             3.62
                Test 9               2.43               9.72                1.59                  6.36             3.61
               Test10                2.45               9.80                1.60                  6.40             3.65

            Average SAR              2.44               9.76                1.59                  6.38             3.64



            IEEE Target SAR         Measured SAR       Deviation    IEEE Target SAR @            Measured SAR      Deviation
             @ 1 Watt Input         @ 1 Watt Input       from           1 Watt Input             @ 1 Watt Input      from
             averaged over          averaged over       Target         averaged over             averaged over      Target
              1 gram (W/kg)         1 gram (W/kg)         (%)         10 grams (W/kg)           10 grams (W/kg)       (%)

             9.5          +/- 10%           9.76         + 2.7        6.2          +/- 10%               6.38        + 2.9




2005 Celltech Labs Inc.                                                                                                   9 of 12


                                                                              835 MHz System Validation Dipole (Brain)



         835 MHz System Validation (Brain) - March 30, 2005
         DUT: Dipole 835 MHz; Type: D835V2; Serial: 411
         Ambient Temp: 20.5ºC; Fluid Temp: 20.4ºC; Barometric Pressure: 102.2 kPa; Humidity: 32%
         Communication System: CW
         Frequency: 835 MHz; Duty Cycle: 1:1
                                                                                                     3
         Medium: HSL835 Medium parameters used: f = 835 MHz; ı = 0.90 mho/m; İr = 39.5; ȡ = 1000 kg/m
         - Probe: ET3DV6 - SN1590; ConvF(6.71, 6.71, 6.71); Calibrated: 24/05/2004
         - Sensor-Surface: 4mm (Mechanical And Optical Surface Detection)
         - Electronics: DAE3 Sn370; Calibrated: 25/01/2005
         - Phantom: SAM 4.0; Type: Fiberglas; Serial: 1033
         - Measurement SW: DASY4, V4.5 Build 19; Postprocessing SW: SEMCAD, V1.8 Build 146

         835 MHz System Validation/Area Scan (6x10x1): Measurement grid: dx=10mm, dy=10mm
         Reference Value = 56.5 V/m; Power Drift = -0.031 dB

         835 MHz System Validation/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 56.5 V/m; Power Drift = -0.031 dB
         Peak SAR (extrapolated) = 3.65 W/kg
         SAR(1 g) = 2.45 mW/g; SAR(10 g) = 1.6 mW/g

         835 MHz System Validation/Zoom Scan 2 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 56.1 V/m; Power Drift = -0.00 dB
         Peak SAR (extrapolated) = 3.66 W/kg
         SAR(1 g) = 2.44 mW/g; SAR(10 g) = 1.59 mW/g

         835 MHz System Validation/Zoom Scan 3 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 55.9 V/m; Power Drift = -0.013 dB
         Peak SAR (extrapolated) = 3.67 W/kg
         SAR(1 g) = 2.45 mW/g; SAR(10 g) = 1.6 mW/g

         835 MHz System Validation/Zoom Scan 4 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 55.9 V/m; Power Drift = 0.00 dB
         Peak SAR (extrapolated) = 3.64 W/kg
         SAR(1 g) = 2.44 mW/g; SAR(10 g) = 1.59 mW/g

         835 MHz System Validation/Zoom Scan 5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 56.0 V/m; Power Drift = -0.017 dB
         Peak SAR (extrapolated) = 3.62 W/kg
         SAR(1 g) = 2.44 mW/g; SAR(10 g) = 1.59 mW/g

         835 MHz System Validation/Zoom Scan 6 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 55.8 V/m; Power Drift = -0.01 dB
         Peak SAR (extrapolated) = 3.61 W/kg
         SAR(1 g) = 2.43 mW/g; SAR(10 g) = 1.59 mW/g

         835 MHz System Validation/Zoom Scan 7 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 56.2 V/m; Power Drift = 0.00 dB
         Peak SAR (extrapolated) = 3.65 W/kg
         SAR(1 g) = 2.45 mW/g; SAR(10 g) = 1.6 mW/g

         835 MHz System Validation/Zoom Scan 8 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 56.2 V/m; Power Drift = 0.00 dB
         Peak SAR (extrapolated) = 3.62 W/kg
         SAR(1 g) = 2.43 mW/g; SAR(10 g) = 1.59 mW/g

         835 MHz System Validation/Zoom Scan 9 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 55.8 V/m; Power Drift = -0.00 dB
         Peak SAR (extrapolated) = 3.61 W/kg
         SAR(1 g) = 2.43 mW/g; SAR(10 g) = 1.59 mW/g

         835 MHz System Validation/Zoom Scan 10 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 56.2 V/m; Power Drift = -0.00 dB
         Peak SAR (extrapolated) = 3.65 W/kg
         SAR(1 g) = 2.45 mW/g; SAR(10 g) = 1.6 mW/g


2005 Celltech Labs Inc.                                                                                        10 of 12


                                                      835 MHz System Validation Dipole (Brain)




                          
         1 g average of 10 measurements: 2.44 mW/g
         10 g average of 10 measurements: 1.59 mW/g




2005 Celltech Labs Inc.                                                                11 of 12


nnnnnnnnnnnnnnnnnnnnnnnnnn




         System Validation — 835 MHz Dipole
         Measured Fluid Dielectric Parameters (Brain)
         Marh 2, 215


       Frequency               [         e"
       735000000 MHz         40.7992    19.7090
       745000000 MHz         40.6764     19.6562
       755000000 MHz          40.5150   19.6147
       765000000 MHz         40.3469     19.5936
       775000000 MHz         40.2286    19.5727
       785.000000 MHz        40.1120     19.5413
       795.000000 MHz        39.9862    19.4590
       $05.000000 MHz        39.8373     19.4821
       $15.000000 MHz        39.7113     19.4303
       $25.000000 MHz        39.5956     19.3828
       835000000 204525 19.3180
       $45.000000 MHz        39.3521     19.3009
       $55.000000 MHz        39.2084     19.3013
       $65.000000 MHz        39.0910    19.2701
       875000000 MHz         38.9606    19.2337
       885.000000 MHz        38.8205     19.2213
       895.000000 MHz        38.7043    191737
       905.000000 MHz        38.6586.   191569
       915.000000 MHz        38.4783    191542
       925.000000 MHz        38.3777    19.0771
       935.000000 MHz        38.2585     19.0264


                            1900 MHz SYSTEM VALIDATION DIPOLE


               Type:                                 1900 MHz Validation Dipole




                                                                     151
               Serial Number:




                                                            Celltech Labs Inc.
               Place of Calibration:




               Date of Calibration:                           June 18, 2004




Celltech Labs Inc. hereby certifies that this device has been calibrated on the date indicated above.




               Calibrated by:




               Approved by:




                          Celltech Labs Inc. 1955 Moss Court, Kelowna, B.C. Canada V1Y 9L3
                        Tel. 250-448-7047 x Fax. 250-448-7046 x e-mail: info@celltechlabs.com
                                                www.celltechlabs.com


1. Dipole Construction & Electrical Characteristics

The validation dipole was constructed in accordance with the IEEE Standard “Annex G (informative)
Reference dipoles for use in system validation”. The electrical properties were measured using an HP
8753E Network Analyzer. The network analyzer was calibrated to the validation dipole N-type connector
feed point using an HP85032E Type N calibration kit. The dipole was placed parallel to a planar phantom
at a separation distance of 10.0mm from the simulating fluid using a loss-less dielectric spacer. The
measured input impedance is:

       Feed point impedance at 1900MHz         Re{Z} = 50.115:
                                               Im{Z} = 6.2070:

       Return Loss at 1900MHz                  -24.205dB


                                                          18 Jun 2004 @9:26:48
[CHI) s11     1 U FS          1: 59.115 6   6.2070 4   519.94 pH      1 900.000 000 MHz
      t

PRm




Cor




            START 1 700.000 000 MHz                             STOP 2 100.900 000 MHz


                                       Wls

                                                       18 Jun 2004    @9:25:56
CHI]   Sii     LOG      10 dB/REF 0 dB                1:—24.205 dB   1 900.000 000 MHz
        y                        "


PRm                                          ts
                                         «        /




Ctor




 t



       f——_




             START 1 790.000 600 MHz                          STOP 2 100.000 000 MHz


Validation Dipole Dimensions


             Frequency (MHz)            L (mm)             h (mm)             d (mm)

                    300                  420.0              250.0               6.2
                    450                  288.0              167.0               6.2
                    835                  161.0              89.8                3.6
                    900                  149.0              83.3                3.6
                    1450                  89.1              51.7                3.6
                    1800                  72.0              41.7                3.6
                    1900                  68.0              39.5                3.6
                    2000                  64.5              37.5                3.6
                    2450                  51.8              30.6                3.6
                    3000                  41.5              25.0                3.6




2. Validation Phantom

The validation phantom is the SAM (Specific Anthropomorphic Mannequin) phantom manufactured by
Schmid & Partner Engineering AG. The SAM phantom is a Fiberglass shell integrated in a wooden table.
The shape of the shell corresponds to the phantom defined by SCC34-SC2. It enables the dosimetric
evaluation of left and right hand phone usage as well as body mounted usage at the flat phantom region.
A cover prevents evaporation of the liquid. Reference markings on the phantom allow the complete setup
of all predefined phantom positions and measurement grids by manually teaching three points in the
robot.

Shell Thickness:       2.0 ± 0.1 mm
Filling Volume:        Approx. 20 liters
Dimensions:            50 cm (W) x 100 cm (L)






3. Measurement Conditions

The SAM phantom was filled with 1900 MHz brain simulating tissue.

        Relative Permittivity:      38.3
        Conductivity:               1.43 mho/m
        Ambient Temperature: 24.0 qC
        Fluid Temperature:          22.6 qC
        Fluid Depth:                t 15.0 cm
        Barometric Pressure:        103.0 kPa
        Humidity:                   37%

The 1900 MHz tissue simulant consists of the following ingredients:


                         Ingredient                     Percentage by weight

                           Water                               55.85%

                           Glycol                              44.00%

                             Salt                               0.15%

               Target Dielectric Parameters                  Hr = 40.0
                         at 22 qC                            V = 1.40 S/m


4. SAR Measurement

The SAR measurement was performed with the E-field probe in mechanical detection mode only. The
setup and determination of the forward power into the dipole was performed using the following
procedures.




                                                                   3D Probe positioner

                                d
                                                           Field probe
                                                                             Flat Phantom




                                                                             Dipole




                                       Dir.Coupler
Signal         Amp      Low    3dB                                       x
Generator               Pass                             Cable                    Att1
                               Att3
                                                                                       PM1

                                            Att2   PM3



                                      PM2




First the power meter PM1 (including attenuator Att1) is connected to the cable to measure the forward
power at the location of the dipole connector (X). The signal generator is adjusted for the desired
forward power at the dipole connector (taking into account the attenuation of Att1) as read by power
meter PM2. After connecting the cable to the dipole, the signal generator is readjusted for the same
reading at power meter PM2. If the signal generator does not allow adjustment in 0.01dB steps, the
remaining difference at PM2 must be taken into consideration. PM3 records the reflected power from the
dipole to ensure that the value is not changed from the previous value. The reflected power should be
50dB below the forward power.


Ten SAR measurements were performed in order to achieve repeatability and to establish an average
target value.


Validation Dipole SAR Test Results


                    SAR @ 0.25W         SAR @ 1W        SAR @ 0.25W         SAR @ 1W
  Validation                                                                               Peak SAR @
                   Input averaged     Input averaged   Input averaged     Input averaged
 Measurement                                                                               0.25W Input
                      over 1g            over 1g          over 10g           over 10g
     Test 1             10.1              40.40             5.30              21.20             17.4
     Test 2             9.93              39.72             5.21              20.84             17.2
     Test 3             9.98              39.92             5.23              20.92             17.3
     Test 4             9.99              39.96             5.21              20.84             17.4
     Test 5             9.97              39.88             5.22              20.88             17.4
     Test 6             9.90              39.60             5.20              20.80             17.1
     Test 7             9.93              39.72             5.21              20.84             17.2
     Test 8             9.96              39.84             5.20              20.80             17.3
     Test 9             9.94              39.76             5.20              20.80             17.2
    Test 10             9.96              39.84             5.21              20.84             17.2

    Average            9.966              39.864           5.219             20.876            17.27


The results have been normalized to 1W (forward power) into the dipole.

                               Average Measured SAR       IEEE Target SAR
    1g/10g Averaged                                                                    Deviation (%)
                                    @ 1W Input              @ 1W Input
         1 gram                       39.864                       39.7                    + 0.413
         10 gram                      20.876                       20.5                    + 1.835


1900 MHz System Validation - June 18, 2004
DUT: Dipole 1900 MHz; Type: D1900V2; Serial: 151
Ambient Temp: 24.0°C; Fluid Temp: 22.6°C; Barometric Pressure: 103.0 kPa; Humidity: 37%
Communication System: CW
Frequency: 1900 MHz; Duty Cycle: 1:1
                                                         3
Medium: HSL1900 (ı = 1.43 mho/m; İr = 38.3; ȡ = 1000 kg/m )

- Probe: ET3DV6 - SN1387; ConvF(5.25, 5.25, 5.25); Calibrated: 18/03/2004
- Sensor-Surface: 4mm (Mechanical And Optical Surface Detection)
- Electronics: DAE3 Sn370; Calibrated: 14/05/2004
- Phantom: SAM 4.0; Type: Fiberglas; Serial: 1033
- Measurement SW: DASY4, V4.2 Build 44; Postprocessing SW: SEMCAD, V1.8 Build 112

1900 MHz System Validation/Area Scan (5x8x1): Measurement grid: dx=15mm, dy=15mm
Reference Value = 96.9 V/m; Power Drift = 0.1 dB
1900 MHz System Validation/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 96.9 V/m; Power Drift = 0.1 dB
Peak SAR (extrapolated) = 17.4 W/kg
SAR(1 g) = 10.1 mW/g; SAR(10 g) = 5.3 mW/g
1900 MHz System Validation/Zoom Scan 2 (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 94.8 V/m; Power Drift = 0.0 dB
Peak SAR (extrapolated) = 17.2 W/kg
SAR(1 g) = 9.93 mW/g; SAR(10 g) = 5.21 mW/g
1900 MHz System Validation/Zoom Scan 3 (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 94.2 V/m; Power Drift = 0.009 dB
Peak SAR (extrapolated) = 17.3 W/kg
SAR(1 g) = 9.98 mW/g; SAR(10 g) = 5.23 mW/g
1900 MHz System Validation/Zoom Scan 4 (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 94.9 V/m; Power Drift = 0.001 dB
Peak SAR (extrapolated) = 17.4 W/kg
SAR(1 g) = 9.99 mW/g; SAR(10 g) = 5.21 mW/g
1900 MHz System Validation/Zoom Scan 5 (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 95.2 V/m; Power Drift = -0.003 dB
Peak SAR (extrapolated) = 17.4 W/kg
SAR(1 g) = 9.97 mW/g; SAR(10 g) = 5.22 mW/g
1900 MHz System Validation/Zoom Scan 6 (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 94.8 V/m; Power Drift = -0.0 dB
Peak SAR (extrapolated) = 17.1 W/kg
SAR(1 g) = 9.9 mW/g; SAR(10 g) = 5.2 mW/g
1900 MHz System Validation/Zoom Scan 7 (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 94.8 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 17.2 W/kg
SAR(1 g) = 9.93 mW/g; SAR(10 g) = 5.21 mW/g
1900 MHz System Validation/Zoom Scan 8 (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 95.1 V/m; Power Drift = -0.007 dB
Peak SAR (extrapolated) = 17.3 W/kg
SAR(1 g) = 9.96 mW/g; SAR(10 g) = 5.2 mW/g
1900 MHz System Validation/Zoom Scan 9 (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 94.7 V/m; Power Drift = -0.0 dB
Peak SAR (extrapolated) = 17.2 W/kg
SAR(1 g) = 9.94 mW/g; SAR(10 g) = 5.2 mW/g
1900 MHz System Validation/Zoom Scan 10 (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 95.1 V/m; Power Drift = -0.0 dB
Peak SAR (extrapolated) = 17.2 W/kg
SAR(1 g) = 9.96 mW/g; SAR(10 g) = 5.21 mW/g


1 g average of 10 measurements: 9.966 mW/g
10 g average of 10 measurements: 5.219 mW/g


 1900 MHz System Validation
 Measured Fluid Dielectric Parameters (Brain)
 June 18, 2004



Frequency           e'         e''
1.800000000 GHz    38.7685    13.2945
1.810000000 GHz    38.7232    13.3253
1.820000000 GHz    38.6647    13.3519
1.830000000 GHz    38.6047    13.3737
1.840000000 GHz    38.5593    13.4078
1.850000000 GHz    38.5136    13.4244
1.860000000 GHz    38.4736    13.4289
1.870000000 GHz    38.4328    13.4399
1.880000000 GHz    38.3934    13.4856
1.890000000 GHz    38.3637    13.4872
1.900000000 GHz    38.3205    13.5178
1.910000000 GHz    38.2981    13.5327
1.920000000 GHz    38.2590    13.5755
1.930000000 GHz    38.2344    13.5976
1.940000000 GHz    38.2172    13.6297
1.950000000 GHz    38.1838    13.6574
1.960000000 GHz    38.1575    13.6807
1.970000000 GHz    38.1070    13.6962
1.980000000 GHz    38.0516    13.7296
1.990000000 GHz    38.0093    13.7634
2.000000000 GHz    37.9485    13.7978



Document Created: 2005-11-01 16:39:15
Document Modified: 2005-11-01 16:39:15

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC