Test Report HAC RF Appendix A

FCC ID: IHDT56XN2

Test Report

Download: PDF
FCCID_4043797

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2018/10/6


HAC_E_Dipole_835_181006
DUT: HAC-Dipole 835 MHz

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.2 ℃

DASY5 Configuration:
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2018/6/14
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan - measurement distance from the probe sensor center to CD835 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x361x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 123.4 V/m; Power Drift = -0.07 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 109.6 V/m
Average value of Total=(109.6+107.6) / 2 = 108.6 V/m

                                         PMF scaled E-field
                                          Grid 1 M4 Grid 2 M4 Grid 3 M4
                                          106.1 V/m      109.6 V/m   108.3 V/m
                                          Grid 4 M4      Grid 5 M4   Grid 6 M4
                                          62.70 V/m      63.77 V/m   63.09 V/m
                                          Grid 7 M4      Grid 8 M4   Grid 9 M4
                                          106.0 V/m      107.6 V/m   106.0 V/m
                                                        Cursor:
                                                   Total = 109.6 V/m
                                                    E Category: M4
                                               Location: -1, -75.5, 9.7 mm




                    0 dB = 109.6 V/m = 40.80 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2018/10/6


HAC_E_Dipole_1880_181006
DUT: HAC Dipole 1880 MHz

Communication System: CW; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.2 ℃

DASY5 Configuration:
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2018/6/14
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan - measurement distance from the probe sensor center to CD1880 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 145.0 V/m; Power Drift = 0.10 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 93.41 V/m
Average value of Total=(91.53+93.41) / 2 = 92.47 V/m

                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          90.26 V/m      91.53 V/m   90.07 V/m
                                          Grid 4 M3      Grid 5 M3   Grid 6 M3
                                          70.92 V/m      72.30 V/m   71.53 V/m
                                          Grid 7 M3      Grid 8 M3   Grid 9 M3
                                          91.64 V/m      93.41 V/m   91.47 V/m
                                                        Cursor:
                                                   Total = 93.41 V/m
                                                    E Category: M3
                                                 Location: 0, 31, 9.7 mm




                    0 dB = 93.41 V/m = 39.41 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab      Date: 2018/10/6


HAC_E_Dipole_2450_181006
DUT: HAC Dipole 2450 MHz

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: 0 = 0 $/m, &, = 1; p= 0 kg/m"
Ambient Temperature : 23.2 °C

DASYS5 Configuration:
— Probe: EF3DV3 — SN4053; ConvF(1, 1, 1); Calibrated: 2018/3/19;
— Sensor—Surface: (Fix Surface)
— Electronics: DAE4 Sn8§54; Calibrated: 2018/6/14
— Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
— Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan — measurement distance from the probe sensor center to CD2450 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 76.20 V/m; Power Drift = 0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E—field emissions = 90.85 V/m
Average value of Total=(89.03+90.85) / 2 = 89.94 V/m

                                         PMF scaled E—field

                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          8$8.32 V/m 8$9.03 V/m 86.25 V/m
                                          Grid 4 M3 Grid 5 M3 Grid 6 M3
                                          78.07 V/m 78.28 V/m 76.43 V/m
                                          Grid 7 M3 Grid 8 M3 Grid 9 M3
                                          8$9.87 V/m 90.85 V/m 8$7.96 V/m
                                                          Cursor:
                                                    Total =90.85 V/m
                                                     E Category: M3
                                               Location: 0.5, 24.5, 9.7 mm




        —1.38


        —2.16


        —4.14


        —5.52


        —6.90
                    0 dB =90.85 V/m =39.17 dBV/m



Document Created: 2019-03-23 12:02:36
Document Modified: 2019-03-23 12:02:36

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC