I19Z61530-SEM01_SAR_Rev1_part4

FCC ID: ZNFX540HM

RF Exposure Info

Download: PDF
FCCID_4452305

                                                    No.I19Z61530-SEM01
                                                     Page 175 of 240


ANNEX H        Dipole Calibration Certificate
750 MHz Dipole Calibration Certificate




                                                ©Copyright. All rights reserved by CTTL.


Calibration Laboratory of                                                      S   Schweizerischer Kalibrierdienst
Schmid & Partner                                                               e   Service suisse d‘étalonnage
   Engineering AG                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                   S   Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x.y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
     SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
     e   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D750V3—1017_Jul19                                Page 2 of 8


¢S)
                                                                                                 o —                 2202   .e




 Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                              DASY5                              V52.10.2
      Extrapolation                                   Advanced Extrapolation
      Phantom                                         Modular Flat Phantom
      Distance Dipole Center — TSL                              15 mm                             with Spacer
      Zoom Scan Resolution                              dx, dy, dz =5 mm
      Frequency                                          750 MHz +1 MHz



 Head TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity           Conductivity
       Nominal Head TSL parameters                                 22.0 °C           41.9                0.89 mho/m
       Measured Head TSL parameters                             (22.0 + 0.2) °C   42.2 +6 %            0.89 mho/m + 6 %
       Head TSL temperature change during test                     <0.5 °C


 SAR result with Head TSL

       SAR averaged over 1 cm* (1 g) of Head TSL                 Condition
       SAR measured                                       250 mW input power                     2.14 W/kg
       SAR for nominal Head TSL parameters                    normalized to 1W         8.57 W/kg x 17.0 % (k=2)


       SAR averaged over 10 cm* (10 g) of Head TSL                condition
       SAR measured                                       250 mW input power                     1.39 W/kg
       SAR for nominal Head TSL parameters                    normalized to 1W         5.57 Wikg x 16.5 % (k=2)



 Body TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity           Conductivity
       Nominal Body TSL parameters                                 22.0 °C           §5.5                0.96 mho/m
       Measured Body TSL parameters                             (22.0 + 0.2) °C   55.1 +6 %            0.96 mho/m + 6 %
       Body TSL temperature change during test                     <0.5 °C            sese                    wes—


 SAR result with Body TSL

       SAR averaged over 1 cm* (1 g) of Body TSL                 Condition
       SAR measured                                       250 mW input power                     2.14 W/kg
       SAR for nominal Body TSL parameters                    normalized to 1W         8.55 W/kg + 17.0 % (k=2)


       SAR averaged over 10 cm* (10 g) of Body TSL                condition
       SAR measured                                       250 mW input power                  141 Wikg
       SAR for nominal Body TSL parameters                    normalized to 1W         5.63 W/kg + 16.5 % (k=2)



 Certificate No: D750V3—1017_Jul19                     Page 3 of 8


T77                                                                                       eee vve veme




Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           53.1 Q — 1.3 jQ
      Return Loss                                                                       —29.6 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           48.9 Q — 4.3 jQ
      Return Loss                                                                       — 27.0 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                       L                           1.041 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG




Certificate No: D750V3—1017_Jul19                        Page 4 of 8


mm>                                                                       l le l eene re ce eksrke.
777                                                                             8y e   e e e e




DASY5 Validation Report for Head TSL
                                                                                       Date: 18.07.2019

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 — SN:1017

Communication System: UID 0 — CW; Frequency: 750 MHz
Medium parameters used: f = 750 MHz; 0 = 0.89 S/m; &, = 42.2; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    e    Probe: EX3DV4 — SN7349; ConvF(10.07, 10.07, 10.07) @ 750 MHz; Calibrated: 29.05.2019

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e    Electronics: DAE4 Sn601; Calibrated: 30.04.2019

    e    Phantom: Flat Phantom 4.9 (front); Type: QD OOL P49 AA; Serial: 1001

    «_   DASY52 52.10.2(1504); SEMCAD X 14.6.12(7470)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 59.72 V/m; Power Drift = 0.00 dB
 Peak SAR (extrapolated) =3.21 W/kg
 SAR(I g) = 2.14 W/kg; SAR(10 g) = 1.39 W/kg
 Maximum value of SAR (measured) = 2.84 W/kg


         dB
         0


         —2.00


         —4.00


         —6.00


         —8.00


         —10.00
                    0 dB = 2.84 W/kg = 4.53 dBW/kg




 Certificate No: D750V3—1017_Jul19               Page 5 of 8


()
n
S
Impedance Measurement Plot for Head TSL




       File View Channel Sweep Calibration Trace Scale Marker System Window Help
                                                                                                        53.148 O
                                                                                                       ~1.3174 0
                                                                                                      33.079 mU
                                                                                                        ~21.979°




                     Ch1Avg= 20
           Chi1: Start 550.000 MHz   ——                                                              Stop 950000 MHz


                                                                                  750.000000   MHz    __—2$9.609 dB




           00

           .00
           00
           .00
           00
           Chi: Start

        Status      CH 1:                                         Auvg=20 Delay




 Certificate No: D750V3—1017_Jul19                  Page 6 of 8


&TTL                                                                    seo en oo s ece eckle.




)
                                                                             e



 DASY5 Validation Report for Body TSL
                                                                                    Date: 18.07.2019

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 — SN:1017

 Communication System: UID 0 — CW; Frequency: 750 MHz
 Medium parameters used: £ = 750 MHz; 0 = 0.96 S/m; s = 55.1; p = 1000 kg/m3
 Phantom section: Flat Section
 Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

 DASY352 Configuration:

    e   Probe: EX3DV4 — SN7349; ConvF(10.4, 10.4, 10.4) @ 750 MHz; Calibrated: 29.05.2019

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e    Electronics: DAE4 Sn601; Calibrated: 30.04.2019

    &    Phantom: Flat Phantom 4.9 (Back); Type: QD OOR P49 AA; Serial: 1005

    e    DASY52 52.10.2(1504); SEMCAD X 14.6.12(7470)


 Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 55.74 V/m; Power Drift = —0.02 dB
 Peak SAR (extrapolated) = 3.18 W/kg
 SAR(1 g) = 2.14 W/kg; SAR(10 g) = 1.41 W/kg
 Maximum value of SAR (measured) = 2.84 W/kg


        dB
         0


         —2.00


         ~4.00


         —6.00


         —8.00


         —10.00
                    0 dB = 2.84 W/kg = 4.53 dBW/kg




 Certificate No: D750V3—1017_Jul19              Page 7 of 8


(EI)
       Impedance Measurement Plot for Body TSL




            Eile View Channel Swesp Calibration Trace icdem!mdow»mh
             I                                                                  _     750.000000 MHz        48.897 O
                                                                                                           —4.2735 0
                                                                                                          44.586 mU
                                                                                                            —101.99 °




                           Ch1Avg= 20
                   Chi: Start 550.000 MHz    ——    —                                                     Stop 950.000 MHz


                                                                                      750.000000   MHz    —2f.016 dB




                   .00

                   .00
                   .00
                   00
                   00       Ch 1      =
                   Ch1: Start 5$50.000 MHa   ——                                                          Stop 950.000 MHz

                 Status    CH 1: En'              ~TJCET—Pot          "Avg=20 Delay                            LCL




       Certificate No: D750V3—1017_Jul19                       Page 8 of 8


T77                                                                                                           es vve veme .




Calibration Laboratory of                                           alll                                   S       Schweizerischer Kalibrierdienst




                                                             C
                                                                 m




                                                                                 oo
Schmid & Partner                                                     shes                                          Service suisse d‘étalonnage
  Engineering AG                                                 2 Pm p
                                                                                                           C       Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      4,,[[/7/\\\\\\\3                         S       Swiss Calibration Service
                                                                      "hiluls®


Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client      CTTL (Auden)                                                                        Certificate No: D835V2—4d069_Jul19

[CALIBRATION CERTIFICATE
 Object                                D835V2 — SN:4d069


 Calibration procedure(s)              QA CAL—O05.v11
                                       Calibration Procedure for SAR Validation Sources between 0.7—3 GHz



 Calibration date:                     July 18, 2019



 This calibration certificate documents the traceability to national standards, which realize the physicalunits of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN: 104778                03—Apr—19 (No. 217—02892/02893)                       Apr—20
 Power sensor NRP—Z91                   SN: 103244                03—Apr—19 (No. 217—02892)                             Apr—20
 Power sensor NRP—Z291                  SN: 103245                03—Apr—19 (No. 217—02893)                             Apr—20
 Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—19 (No. 217—02894)                             Apr—20
 Type—N mismatch combination            SN: 5047.2 / 06827        04—Apr—19 (No. 217—02895)                             Apr—20
 Reference Probe EX3DV4                 SN: 7349                  29—May—19 (No. EX3—7349_May19)                        May—20
 DAE4                                   SN: 601                   30—Apr—19 (No. DAE4—601_Apr19)                        Apr—20

 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter E4419B                     SN: GB39512475            30—Oct—14 (in house check Feb—19)                     In house check: Oct—20
 Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—18)                     In house check: Oct—20
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—18)                     In house check: Oct—20
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—18)                     In house check: Oct—20
 Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—18)                     In house check: Oct—19


                                        Name                                     Function


                                                                                                                                 6
 Calibrated by:                         Claudio Leubler                          Laboratory Technician



 Approved by:                           Katia Pokovie                            Technical Manager                     %



                                                                                                                        Issued: July 19, 2019
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D835V2—40069_Jul19                                    Page 1 of 8


(3 S
 Calibration Laboratory of
 Schmid & Partner
   Engineering AG
 Zeughausstrasse 43, 8004 Zurich, Switzerland


 Accredited by the Swiss Accreditation Service (SAS)
                                                             s\“\‘y/"";
                                                            ifififi
                                                            ooe
                                                             *z/,fi\\\\\x‘



 The Swiss Accreditation Service is one of the signatories to the EA
                                                                              S
                                                                              C
                                                                              S
                                                                                  Schweizerischer Kalibrierdienst
                                                                                  Service suisse d‘étalonnage
                                                                                  Servizio svizzero di taratura
                                                                                  Swiss Calibration Service



                                                                              Accreditation No.: SCS 0108

 Multilateral Agreementfor the recognition of calibration certificates

 Glossary:
 TSL                         tissue simulating liquid
 ConvF                       sensitivity in TSL / NORM x.y,z
 N/A                         not applicable or not measured

 Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
          Communications Devices: Measurement Techniques", June 2013
     b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
        (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
        300 MHz to 6 GHz)", July 2016
     c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
        communication devices used in close proximity to the human body (frequency range of 30
        MHz to 6 GHz)", March 2010
     d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

 Additional Documentation:
    e) DASY4/5 System Handbook

 Methods Applied and Interpretation of Parameters:
    e Measurement Conditions: Further details are available from the Validation Report at the end
      of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
      point exactly below the center marking of the flat phantom section, with the arms oriented
      paralle! to the body axis.
    e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
      positioned under the liquid filled phantom. The impedance stated is transformed from the
      measurement at the SMA connector to the feed point. The Return Loss ensures low
      reflected power. No uncertainty required.
    e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
      No uncertainty required.
    e SAR measured: SAR measured at the stated antenna input power.
    e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
     e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
   The reported uncertainty of measurement is stated as the standard uncertainty of measurement
   multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
   probability of approximately 95%.




 Certificate No: D835V2—4d069_Jul19                             Page 2 of 8


                                                                                    «o+ se e sea lz a Nesn               w crvin d

                                                                                            ~3g—       ks           is




Measurement Conditions
   DASY system configuration, as far as not given on page 1.
      DASY Version                                             DASY5                              V52.10.2
      Extrapolation                                  Advanced Extrapolation
      Phantom                                         Modular Flat Phantom
      Distance Dipole Center — TSL                             15 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
      Frequency                                         835 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
      Nominal Head TSL parameters                                 22.0 °C           41.5                0.90 mho/m
      Measured Head TSL parameters                             (22.0 + 0.2) °C   42.0 +6 %           0.91 mho/m + 6 %
      Head TSL temperature change during test                     «05 °C             ————                    ————


SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                 Condition
      SAR measured                                       250 mW input power                     2.44 W/kg
      SAR for nominal Head TSL parameters                    normalized to 1W        9.70 Wikg + 17.0 %(k=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                condition
      SAR measured                                       250 mW input power                  1.58 Wikg
      SAR for nominal Head TSL parameters                    normalized to 1W         6.29 W/kg + 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
      Nominal Body TSL parameters                                 22.0 °C           55.2                0.97 mho/m
      Measured Body TSL parameters                             (22.0 £ 0.2) °C   54.9 + 6 %          0.99 mho/m + 6 %
      Body TSL temperature change during test                     <0.5 °C

SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                 Condition
      SAR measured                                       250 mW input power                     2. 46 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         9.68 Wi/kg + 17.0 %(k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                     1.60 Wikg
      SAR for nominal Body TSL parameters                    normalized to 1W         6.32 W/kg + 16.5 % (k=2)




Certificate No: D835V2—40069_Jul19                    Page 3 of 8


4S
                                                                                                     o—      Li     2l   2l




 Appendix (Additional assessments outside the scope of SCS 0108)

 Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                           50.8 Q — 2.4 jQ
       Return Loss                                                                       — 32.1 dB



 Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                           47.1 Q — 3.9 jQ
       Return Loss                                                                       — 25.9 dB




 General Antenna Parameters and Design

       Electrical Delay (one direction)                     T                             1.393 ns


 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
 second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
 are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
 "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
 according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
 feedpoint may be damaged.


 Additional EUT Data

       Manufactured by                                                                    SPEAG




 Certificate No: D835V2—4d069_Jul19                       Page 4 of 8


)
@777
WGragzre‘"




 DASY5 Validation Report for Head TSL
                                                                                      Date: 15.07.2019

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN:4d069

 Communication System: UID 0 — CW; Frequency: 835 MHz
  Medium parameters used: £ = 835 MHz; 0 = 0.91 S/m; & = 42; p = 1000 kg/m3
  Phantom section: Flat Section
  Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

  DASY52 Configuration:

     e    Probe: EX3DV4 — SN7349; ConvF(9.89, 9.89, 9.89) @ 835 MHz; Calibrated: 29.05.2019

     e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

     e    Electronics: DAE4 Sn601; Calibrated: 30.04.2019

     *    Phantom: Flat Phantom 4.9 (front); Type: QD OOL P49 AA; Serial: 1001

     «_   DASY52 52.10.2(1504); SEMCAD X 14.6.12(7470)


  Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
  Measurement grid: dx=5mm, dy=5mm, dz=5mm
  Reference Value = 63.48 V/m; Power Drift = 0.01 dB
  Peak SAR (extrapolated) = 3.58 W/kg
  SAR(1 g) = 2.44 W/kg; SAR(10 g) = 1.58 W/kg
  Maximum value of SAR (measured) = 3.22 W/kg




          —2.20


          —4.40


          —6.60




          —11.00
                     0 dB = 3.22 W/kg = 5.08 dBW/kg




  Certificate No: D835V2—4d069_Jul19             Page 5 of 8


©
Impedance Measurement Plot for Head TSL




     Eile View Channel Swesp Calibration Trace Scale Marker System Window Help
                                                                       835.000000 MHz       50.811 0
                                                                                           ~2.3632 0
                                                                                          24.778 mU
                                                                                            —69.710 °




                   Ch1Avg= 20
          Ch1: Start 635000 MHz                                                           Stop 1.03500 GHz


                                                                       835.000000   MHz




         00       Ch 1     &
          Chi: Start 625.000                                                              Stop 1.02500 GHz

                                       C 1—Port                                                ECE




Certificate No: D835V2—4d069_Jul19                Page 6 of 8


771L                                                                      aeore n oo e e e enree s




)
                                                                             Ce ee e e e



 DASY5 Validation Report for Body TSL
                                                                                     Date: 18.07.2019

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN:4d069

 Communication System: UID 0 — CW; Frequency: 835 MHz
 Medium parameters used: f = 835 MHz; 0 = 0.99 S/m; &, = 54.9; p = 1000 kg/m3
 Phantom section: Flat Section
 Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

 DASY52 Configuration:

    e    Probe: EX3DV4 — SN7349; ConvF(10.16, 10.16, 10.16) @ 835 MHz; Calibrated: 29.05.2019

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e    Electronics: DAE4 Sn601; Calibrated: 30.04.2019

    e    Phantom: Flat Phantom 4.9 (Back); Type: QD OOR P49 AA; Serial: 1005

    «_   DASY52 52.10.2(1504); SEMCAD X 14.6.12(7470)


 Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 57.81 V/m; Power Drift = 0.01 dB
 Peak SAR (extrapolated) = 3.65 W/kg
 SAR(1 g) = 2.46 W/kg; SAR(10 g) = 1.6 W/ikg
 Maximum value of SAR (measured) = 3.28 W/kg




         —2.156


         —4.30


         —6.44


         —8.59


         —10.74
                    0 dB = 3.28 W/kg = 5.16 dBW/kg




 Certificate No: D835V2—40069_Jul19             Page 7 of 8



Document Created: 2019-09-19 14:35:07
Document Modified: 2019-09-19 14:35:07

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC