DASY Calibration Certificate 2

FCC ID: ZNFX440IM

Test Report

Download: PDF
FCCID_4415304

Calibration Laboratory of                                                                                          Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                                                                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                             Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       Morlab (Auden)                                                                     Certificate No: DAE4—480_Apr19

CALIBRATION CERTIFICATE
 Object                                DAEA4 — SD 000 D04 BJ — SN: 480


 Calibration procedure(s)              QA CAL—O6.v29
                                       Calibration procedure for the data acquisition electronics (DAE)



 Calibration date:                     April 11, 2019




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Keithley Multimeter Type 2001          SN: 0810278               03—Sep—18 (No:23488)                                  Sep—19

 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Auto DAE Calibration Unit              SE UWS 053 AA 1001        O07—Jan—19 (in house check)                           In house check: Jan—20
 Calibrator Box V2.1                    SE UMS 006 AA 1002        O07—Jan—19 (in house check)                           In house check: Jan—20




                                        Name                                   Function                                 Signature
 Calibrated by:                         Adrian Gehring                         Laboratory Technician                     4 yif


 Approved by:                           Sven Kihn                              Deputy Manager                             \   V C    3 E    L‘ ” [   ‘
                                                                                                                                                     X




                                                                                                                        Issued: April 11, 2019
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: DAE4—480_Apr19                                        Page 1 of 5


Calibration Laboratory of                                    Pssn                Schweizerischer Kalibrierdienst
Schmid & Partner                                            i}h%//m              Service suisse d‘étalonnage
  Engineering AG
                                                            xA
                                                                                 Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                7 ///\\\\\           Swiss Calibration Service
                                                               ZAulululs®®

Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates


Glossary
DAE                          data acquisition electronics
Connector angle              information used in DASY system to align probe sensor X to the robot
                             coordinate system.

Methods Applied and Interpretation of Parameters
   e DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
     comparison with a calibrated instrument traceable to national standards. The figure given
         corresponds to the full scale range of the voltmeter in the respective range.
    e    Connector angle: The angle of the connector is assessed measuring the angle
         mechanically by a tool inserted. Uncertainty is not required.
    *    The following parameters as documented in the Appendix contain technical information as a
         result from the performance test and require no uncertainty.
         e     DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
               the nominal calibration voltage. Influence of offset voltage is included in this
               measurement.
         e     Common mode sensitivity: Influence of a positive or negative common mode voltage on
               the differential measurement.
         e     Channel separation: Influence of a voltage on the neighbor channels not subject to an
               input voltage.
         e     AD Converter Values with inputs shorted: Values on the internal AD converter
               corresponding to zero input voltage
         e     input Offset Measurement. Output voltage and statistical results over a large number of
               zero voltage measurements.
         e     Input Offset Current: Typical value for information; Maximum channel input offset
               current, not considering the input resistance.

         e     input resistance: Typical value for information: DAE input resistance at the connector,
               during internal auto—zeroing and during measurement.
         e     Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
               alarm signal is generated.
         e     Power consumption: Typical value for information. Supply currents in various operating
               modes.




Certificate No: DAE4—480_Apr19                                  Page 2 of 5


DC Voltage Measurement
   A/D — Converter Resolution nominal
       High Range:       1LSB =         6.1uV ,       full range =  —100...+300 mV
       Low Range:        1LSB =          61inv ,      full range =  —1....... +3mV
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


      Calibration Factors               X                          Y                  Z
      High Range             404.623 + 0.02% (k=2)   404.069 + 0.02% (k=2)   404.429 + 0.02% (k=2)
      Low Range              3.95542 + 1.50% (k=2)   3.94990 + 1.50% (k=2)   3.93793 + 1.50% (k=2)



Connector Angle

      Connector Angle to be used in DASY system                                  138.0 °+ 1 °




Certificate No: DAE4—480_Apr19                       Page 3 of 5


Appendix (Additional assessments outside the scope of SCS0108)

1. DC Voltage Linearity
      High Range                                Reading (uV)          Difference (uV)               Error (%)
     Channel X         + Input                   200025.15                   —7.75                  —0.00
     Channel X         + Input                    20010.70                   5.66                    0.03
     Channel X         — Input                   —20002.29                    3.03                  —0.02
     Channel Y         + Input                   200031.02                   —1.68                  —0.00
     Channel Y         + Input                    20009.03                   4.11                    0.02
     Channel Y         — Input                   —20003.75                    1.67                  —0.01
     Channel Z         + Input                   200037.99                   5.34                    0.00
     Channel Z         + Input                    20008.17                   3.28                    0.02
     Channel Z         — Input                   —20003.72                    1.83                  —0.01


     Low Range                                  Reading (uV)          Difference (uV)               Error (%)
     Channel X         + Input                    2000.99                    —0.02                  —0.00
     Channel X         + Input                     201.69                    0.59                    0.29
     Channel X         — Input                    —198.21                    0.80                   —0.40
     Channel Y         + Input                    2000.23                    —0.68                  —0.03
     Channel Y         + Input                     200.41                    —0.59                  —0.29
     Channel Y         — Input                    —199.51                    —0.44                   0.22
     Channel Z         + Input                    2001.10                    0.17                    0.01
     Channel Z         + Input                     200.14                    —0.85                  —0.42

     Channel Z         — Input                    —200.05                    —0.99                   0.50



2. Common mode sensitivity
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                         Common mode                       High Range                            Low Range
                         Input Voltage (mV)            Average Reading (uV)                 Average Reading (uV)

     Channel X                           200                       6.28                           4.13
                                        — 200                     —2.29                           —3.80
     Channel Y                           200                      —1.73                           —1.78
                                        — 200                      0.87                           0.61
     Channel Z                           200                      10.44                           10.41
                                        — 200                     —12.35                         —12.37



3. Channel separation
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                        Input Voltage (mV)         Channel X (uV)          Channel Y (uV)       Channel Z (uV)

     Channel X                    200                       —                   3.69                 —2.19

     Channel Y                    200                   8.09                        —                3.64

     Channel Z                    200                   7.06                    5.20                      —


Certificate No: DAE4—480_Apr1i9                              Page 4 of 5


4. AD—Converter Values with inputs shorted
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                                                   High Range (LSB)                        Low Range (LSB)

    Channel X                                               15805                               17908

    Channel Y                                               15802                               17649

    Channel Z                                               15727                               16748



5. Input Offset Measurement
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
    Input 10MQ

                           Average (uV)        min. Offset (uV)         max. Offset (uV)        Std. I?:‘\Il;atlon

    Channel X                    0.94                —0.16                   2.05                      0.39
    Channel Y                    0.33                —0.86                       1.28                   0.45
    Channel Z                    1.53                0.63                    2.34                       0.36



6. Input Offset Current
    Nominal Input circuitry offset current on all channels: <25fA


7. Input Resistance (Typical values for information)
                                                 Zeroing (kOhm)                           Measuring (MOhm)
    Channel X                                             200                                   200
    Channel Y                                             200                                   200
    Channel Z                                             200                                   200



8. Low Battery Alarm Voltage (Typical values for information)
    Typical values                                Alarm Level (VDC)
    Supply (+ Vec)                                                                +7.9
    Supply (— Vec)                                                                 —7.6



9. Power Consumption (Typical values for information)
    Typical values                         Switched off (mA)        Stand by (mA)            Transmitting (mA)
    Supply (+ Vec)                                +0.01                     +6                        +14
    Supply (— Vec)                                —0.01                     —8                        —9




Certificate No: DAE4—480_Apr19                            Page 5 of 5



Document Created: 2019-08-28 19:02:01