DASY Calibration Certificate 1

FCC ID: ZNFX440IM

Test Report

Download: PDF
FCCID_4415303

  ORLAB,
  '                                                                              REPORT No. : $Z19060060S01

Annex E DASY Calibration Certificate




            SHENZHEN MORLAB COMMUNICATIONS TECHAOLOGY Co., Ltd.                    Tel: 86—755—36898555   Fax: 86—766—36008505
 M o RLAB   FL1—3, Building A, FeiYang Science Park, No.8 LongChang Road,
            Block87, BacAn Distrct, ShenZhen , GuangDong Province, P. R. China     Hiteitwvaw.mortab.on   E—mail: service@morlab.on


                                                                                           \\“\\ll" l’/,                              e   6

                        cmosmnan
                     es n                                                              S@'} BM P BWA




                                                                                       &
                     ———@                                                                 e  Ein4i
                     eaT TL sCALIBRATION
                     maaays>"     _p_e_a_q
                                         LABORATORY                                    %ޤ v CALIBRATION
                                                                                       *                     *   CNAS#*
               Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China           ""Ah, ,’,a\“\\‘                      CNAS LO570
               Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
               E—mail: cttl@chinattl.com           Http://www.chinattl.cn
        Client              Morlab                                               Certificate No:             2Z18—60438

CALIBRATION CERTIFICATE

Object                                        EX3DV4 — SN:3823

Calibration Procedure(s)                      FF—211—004—01
                                              Calibration Procedures for Dosimetric E—field Probes

Calibration date:                             November 12, 2018


This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(Sl). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)°c and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)
Primary Standards                           ID #          Cal Date(Calibrated by, Certificate No.)                 Scheduled Calibration
 Power Meter           NRP2          101919                 20—Jun—18 (CTTL, No.J18X05032)                           Jun—19
 Power sensor          NRP—Z91       101547                 20—Jun—18 (CTTL, No.J18X05032)                           Jun—19
 Power sensor          NRP—Z91       101548                 20—Jun—18 (CTTL, No.J18X05032)                           Jun—19
 Reference10dBAttenuator             18N50OW—10dB           09—Feb—18(CTTL, No.J18X01133)                            Feb—20
 Reference20dBAttenuator             18N50OW—20d4B          09—Feb—18(CTTL, No.J18X01132)                            Feb—20
 Reference Probe EX3DV4              SN 3846                25—Jan—18(SPEAG,No.EX3—3846_Jan18)                       Jan—19
 DAE4                                SN 777                  15—Dec—17(SPEAG, No.DAE4—777_Dec17)                     Dec —18


 Secondary Standards                 ID #                    Cal Date(Calibrated by, Certificate No.)             Scheduled Calibration
 SignalGeneratorMG3700A              6201052605              21—Jun—18 (CTTL, No.J18X05033)                        Jun—19
 Network Analyzer E5071C             MY46110673              14—Jan—18 (CTTL, No.J18X00561)                        Jan —19
                                  Name                             Function                                         Signature
   .             .                                                                                                    x
Calibrated by:                     Yu Zongying                    SAR Test Engineer                                 4/5>

Reviewed by:                       Lin Hao                        SAR Test Engineer                                 Tfifi%}
Approved by:                       Qi Dianyuan                    SAR Project Leader                             Ma\/

                                                                                              Issued: November 14, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



       Certificate No: Z18—60438                                  Page 1 of 11


          A                         In Collaboration with
         /// s p e a q
          *\
          uaagys"                   CALIBRATIONLABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com        Http:/www.chinattl.cn


Glossary:
TSL                   tissue simulating liquid
NORMx,y,z             sensitivity in free space
ConvrF                sensitivity in TSL / NORMx,y,z
DCP                   diode compression point
CF                    crest factor (1/duty_cycle) of the RF signal
A,B,C,D               modulation dependent linearization parameters
Polarization ®        ® rotation around probe axis
Polarization 0        0 rotation around an axis that is in the plane normal to probe axis (at measurement center), i
                      0=0 is normal to probe axis
Connector Angle       information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged
  Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
  Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from
    hand—held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)",
    July 2016
c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March
   2010
d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
e   NORMx,y,z: Assessed for E—field polarization 0=0 (fs900MHz in TEM—cell; > 1800MHz: waveguide).
     NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the
     E —field uncertainty inside TSL (see below ConvF).
e    NORM(P)x,y,z = NORMx,y,z* frequency_response (see Frequency Response Chart). This
     linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of the
     frequency response is included in the stated uncertainty of ConvEr.
e    DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep
     (no uncertainty required). DCP does not depend on frequency nor media.
e    PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
     characteristics.
e    Ax,yz; Bx,y,z; Cx,y,.z;VRx,y,z:A,B,C are numerical linearization parameters assessed based on the
     data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
     media. VR is the maximum calibration range expressed in RMS voltage across the diode.
e    ConvrF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
     Transfer Standard for fs800MH2z) and inside waveguide using analytical field distributions based on
     power measurements for f >800MHz. The same setups are used for assessment of the parameters
     applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
     These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
     The sensitivity in TSL corresponds to NORMx,y,z* ConvF whereby the uncertainty corresponds to
     that given for Convr. A frequency dependent ConvF is used in DASY version 4.4 and higher which
     allows extending the validity from+50MHz to+100MHz.
e    Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat
     phantom exposed by a patch antenna.
e    Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the
     probe tip (on probe axis). No tolerance required.
e    Connector Angle: The angle is assessed using the information gained by determining the NORMx
     (no uncertainty required).


Certificate No: Z18—60438                               Page 2 of 11


           A‘E                      In Collaboration with

         =7"7‘J,s a
         es CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com        Http:/www.chinattl.cn




                       Probe EX3DV4


                                              SN: 3823


                                     Calibrated: November 12, 2018

                                Calibrated for DASY/EASY Systems
                                        (Note: non—compatible with DASY2 system!)




Certificate No: Z18—60438                               Page 3 of 11


          ;A®                       In Collaboration with

          ®‘/"/‘/, s p e                                    a g
             razzar"                CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com        Http://www.chinattl.cn



     DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3823

Basic Calibration Parameters
                                     Sensor X                Sensor Y             Sensor Z         Unc (k=2)
 Norm(uV/(V/m)) A                    0.38                    0.38                 0.47            £10.0%
 DCP(mV)®                            101.7                   105.8                101.1


Modulation Calibration Parameters
 UID         Communication                         A              B         C             D      VR        Unc ®
             System Name                           dB             dB/uV                   dB     mV        (k=2)
 0           CW                           X        0.0            0.0       1.0           0.00   138.4     +2.7%
                                          Y        0.0            0.0       1.0                  137.8
                                          Z        0.0            0.0       1.0                  160.5



     The reported uncertainty of measurement is stated as the standard uncertainty of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probability of approximately 95%.


A The uncertainties of Norm X, Y, Z do not affect the E—field uncertainty inside TSL (see Page 5 and Page 6).
8 Numerical linearization parameter: uncertainty not required.
E Uncertainly is determined using the max. deviation from linear response applying rectangular distribution
and is expressed for the square of the field value.




Certificate No: Z18—60438                                Page 4 of 11


              ‘b‘@                        In Collaboration with

             =7"7"J,   a
             @imaayes" CALIBRATION LABORATORY
            Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
            Tel: +86—10—62304633—2512          Fax: +86—10—62304633—2504
            E—mail: cttl@chinattl.com          Http:/www.chinattl.cn


      DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3823


Calibration Parameter Determined in Head Tissue Simulating Media
                            i           Con      ivi                                                    Depth®   Unct.
 f [MHz]®      Pelrarili:tti'\\/’i:yF    o d(lsl;:‘):y    ConvF X       ConvF Y      ConvF Z   Alpha®    (r:m)   (k=2)

    900               41.5                  0.97             9.32            9.32     9.32      0.13     1.44    +£12.1%
   17850              40.1                  1.37             7.98            7.98     7.98      0.20     1.11    +12.1%
   1900               40.0                  1.40             7.68            7.68     7.68      0.27     0.92    £+12.1%
   2000               40.0                  1.40             7.75            7.75     7.75      0.23     0.99    £+12.1%
   2300               39.5                  1.67             1.66            1 .99    1.906     0.47     0.:75   +12.1%
   2450               39.2                  1.80             7.34            7.34     7.34      0.:52    0.75    +12.1%
   2600               39.0                  1.96             6.98            6.98     6.98      0.44     0.84    +12.1%
   5250               35.9                  4.71             5.28            5.28     5.28      0.40     1.30    +13.3%
   5600               35.5                  5.07             4.50            4.50     4.50      0.40     145     +13.3%
   5750               35.4                  5.22             4.60            4.60     4.60      0.45     1.50    +£13.3%



© Frequency validity above 300 MHz of £100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
+50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
F At frequency below 3 GHz, the validity of tissue parameters (s and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and 0) is
restricted to +£5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
S Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




   Certificate No: Z18—60438                                  Page 5 of 11


               A®                          In Collaboration with

             x7"7"J,   a
             ~iaages"" CALIBRATION LABORATORY
            Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
            Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
            E—mail: cttl@chinattl.com            Http://www.chinattl.cn


      DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3823


Calibration Parameter Determined in Body Tissue Simulating Media
                   Relati                Con          ivi                                                Depth®     nct.
 f [MHz]®      Perr(:liat‘ttil\\//i:yF    ° (:;;:;)Fty      ConvF X       ConvF Y     ConvF Z   Alpha®   :::1)    (L:(:;t)

    900               55.0                     1.05            9.24            9.24    9.24      0.19     1.22    #12.1%
   1750               53.4                     1.49            7.65            7.65    7.65      0.19     1.20    +£12.1%
   1900               53.9                     1.52            7.40            7.40    7.40      0.21     115     +£12.1%
   2000               53.3                     1.52            7.61            7.61    7.61      0.20     117     +£12.1%
   2300               52.9                     1.81            7.38            1.98    7.38      0.47     0.87    +12.1%
   2450               52.7                     1.95            7.15            715     716       0.56     0.77    +£12.1%
   2600               52.5                     2.16            7.01            7.01    7.01      0.67     0.68    +12.1%
   5250               48.9                     5.36            4.73            4.73    4.73      0.51     1.22    +13.3%
   5600               48.5                     5.77            3.96            3.96    3.96      0.52     1.39    +13.3%
   5750               48.3                     5.94            3.98            3.98    3.98      0.54     1.62    +13.3%



© Frequency validity above 300 MHz of +100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
+50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
" At frequency below 3 GHz, the validity of tissue parameters (s and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and 0) is
restricted to +5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
© Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




   Certificate No: Z18—60438                                    Page 6 of 11


                             !‘E                     In Collaboration with
                          a_‘/"/‘/ s p e a q
                                   ies               CALIBRATION LABORATORY
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2512                         Fax: +86—10—62304633—2504
         E—mail: cttl@chinattl.com                         Http://www.chinattl.cn



                                               Frequency Response of E—Field
                                               (TEM—Cell: ifi110 EXX, Waveguide: R22)
       Frequency response (normalized)




                                                      I               I              1        T            1      1
                                         0           500           1000            1500     2000         2500   3000

                                                           ce—                    f [MHz]          $«
                                                           TEM                                     R22

                                             Uncertainty of Frequency Response of E—field: £7.4% (k=2)




Certificate No: Z18—60438                                                 Page 7 of 11


                          m©                  In Collaboration with

                      =®‘7/"/‘J, a
                      ies CALIBRATION LABORATORY
                  Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
                  Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504
                  E—mail: cttl@chinattl.com        Http://www.chinattl.cn



                                         Receiving Pattern (®), 0=0°

                          f=600 MHz, TEM                                              f=1800 MHz, R22




                                                                                           PRAQDAYT—q
                                                                                           y hss eC
                                                                                             7 hi



            107 ;

            0.5 —
Error[dB]
            o
            o
                  l
            o
            on
            .




            1.0       —

                                                                           Rollf°]
                                     [=*< 100MHz           __—>——600MHz___—>— 1800MHz__—+— 2500MHz]
                                       Uncertainty of Axial Isotropy Assessment: £1.2% (k=2)




   Certificate No: Z18—60438                                      Page 8 of 11


                                  ®
                                       In Collaboration wi

         aw®‘/"/‘/ s _p_                              e            a               g
                                       CALIBRATION LABORATORY
       Add: No.51 Xueyuan Road, Ha         1an District, Beijing, 100191, China
       Tel +86—10—62304633—2512             Fax: +86—10—62304633—2504
       E      cttl@chinattl.com             Hitt I/Iwww.ch ina ttl.en
         &
         3


                                      Dynam ic Range f(SARnead)
                                                 (TEM cell                             f =   900 MHz)




                                                                                                                      auin sns felake t & ut
                                                             abusnsod ELE
                                                              1
               [ailjeu$is induy




                                                     4


                                                         4+
                                                         a
                                                         a
                                                         7
                                                         a
                                                         x
                                                         3
                                                         4
                                                          1




                                                          x
                                                          1

                                                          1
                                                          1




                                                          1
                                                          1
                                                          1
                                                          1
                                                          1




                                                                                   0                  1
                                                                              10                10              10

                                                                              SAR[m W/em         ‘I
                                                     —CM—)not compensated                    —®— compensated
          [ap JLoug3




                                                                        10°                                    10    19°
                                                                            SAR[mW/cm? 1
                                        [—=—| not compensated            —@— compensated
                                      Uncertainty of L inearl ty Assessment +0.9% (k=2)
Certificate No 7Z18—60438                                    Page 9 of 11


          fiin Collaboration with
         onl                                                      p__e€_                 _a         _0
          y                                         CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512                               Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com                               Http://www.chinattl.cn



                                Conversion Factor Assessment
      {=900 MHz, WGLS R9(H_convF)                                                                        f=1750 MHz, WGLS R22(H_convF)
              4.00                                                                                                30.00
              sso A
                    ‘L                                                                                            25.00
              3.00
              ,      \                                                                                            20.00
                .50
            ul
            iz‘oo                                                                                            &S 15.00
            h® 1.50                                                                                          e0
                                                                                                                  10.00
              1.00
                                                                                                                    5.00
              0.50
              0.00        iombrmale        is toeetsncteolorr      nc                                               009 L iLecc i P 5e
                      0               20               40               60          so          100                    00  40  200 30 40                        50   600   70
                                                                2{mm]                                                                                   2{mm]




                          Deviation from Isotropyin Liquid
                                           1.0
                                           0.8
                                           0.6
                                            0.4
                                            0.2
                                             0.0
                             7 Axis




                                            —0.2
                                             —0.4
                                             —0.6
                                              —0.8
                                              —1 8

                                                                 100                                                                     30
                                                                        150
                                                                 +            200                                                 20          o
                                                                    Aris            250
                                                                                              300
                                                                                                                            10           0@¥
                                                                                                      350

                                                          —1.0      —0.80 —0.60 —0.40 —0.20              0        0.20     0.40   0.60   080      1.0
                                           Uncertainty of Spherical Isotropy Assessment: £3.2% (K=2)



Certificate No: Z18—60438                                                           Page 10 of 11


         é.A@                     In Collaboration with
         /// s p e a g
          ~eaayys"                 CALIBRATION LABORATORY
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
       E—mail: cttl@chinattl.com      Http://www.chinattl.cn


   DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3823


    Other Probe Parameters
    Sensor Arrangement                                                     Triangular

    Connector Angle (°)                                                        169.7

    Mechanical Surface Detection Mode                                        enabled

    Optical Surface Detection Mode                                           disable

    Probe Overall Length                                                     337mm

    Probe Body Diameter                                                        10mm

    Tip Length                                                                  9mm

    Tip Diameter                                                              2.5mm

    Probe Tip to Sensor X Calibration Point                                     1mm

    Probe Tip to Sensor Y Calibration Point                                     1imm

    Probe Tip to Sensor Z Calibration Point                                     1imm

    Recommended Measurement Distance from Surface                             1.4mm




Certificate No: Z18—60438                             Page 11 of 11



Document Created: 2019-08-28 18:53:06
Document Modified: 2019-08-28 18:53:06

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC