Maximum SAR Test Results

FCC ID: ZNFX440IM

Test Report

Download: PDF
FCCID_4415302

   ORLAB,
   '                                                                            REPORT No. : $Z19060060S01


Annex D Plots of Maximum SAR Test Results




           SHENZHEN MORLAB COMMUNICATIONS TECHNOLOGY Co., Ltd.                  Tel: 86—755—36898555   Fax: 86—766—36008525
M o RLAB   FL1—3, Building A, FeiYang Science Park, No.8 LongChang Road,
           Block67, BacAn Distrct, ShenZhen , GuangDong Province, P. R. China   Hiteitwvwmortab.on     E—mail: service@morlab.on


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.16

GSM850_GPRS(4 TX slots)_Right Cheek_Ch251



Communication System: UID 0, GSM850(class 12) (0); Frequency: 848.8 MHz;Duty Cycle: 1:2.08
Medium: HSL_835 Medium parameters used: f = 849 MHz; σ = 0.935 S/m; εr = 42.132; ρ = 1000
kg/m3
Ambient Temperature:23.2 ℃; Liquid Temperature:22.5 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(9.32, 9.32, 9.32); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch251/Area Scan (81x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.152 W/kg

Ch251/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 3.919 V/m; Power Drift = 0.13 dB
Peak SAR (extrapolated) = 0.178 W/kg
SAR(1 g) = 0.146 W/kg; SAR(10 g) = 0.116 W/kg
Maximum value of SAR (measured) = 0.153 W/kg




               0 dB = 0.152 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.17

GSM1900_GPRS(4 TX slots)_Left Cheek_Ch661



Communication System: UID 0, PCS1900(class 12) (0); Frequency: 1880 MHz;Duty Cycle: 1:2.08
Medium: HSL_1900 Medium parameters used: f = 1880 MHz; σ = 1.321 S/m; εr = 41.384; ρ = 1000
kg/m3
Ambient Temperature:23.4 ℃; Liquid Temperature:22.3 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.68, 7.68, 7.68); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch661/Area Scan (81x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.155 W/kg

Ch661/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 5.937 V/m; Power Drift = 0.14 dB
Peak SAR (extrapolated) = 0.213 W/kg
SAR(1 g) = 0.141 W/kg; SAR(10 g) = 0.089 W/kg
Maximum value of SAR (measured) = 0.153 W/kg




               0 dB = 0.153 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.16

LTE Band 5_10MHz_QPSK_1RB_25Offset_Right Cheek_Ch20450



Communication System: UID 0, LTE (0); Frequency: 829 MHz;Duty Cycle: 1:1
Medium: HSL_835 Medium parameters used: f = 829 MHz; σ = 0.917 S/m; εr = 42.33; ρ = 1000
kg/m3
Ambient Temperature:23.2 ℃; Liquid Temperature:22.5 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(9.32, 9.32, 9.32); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch20450/Area Scan (81x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.0991 W/kg

Ch20450/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 2.306 V/m; Power Drift = 0.18 dB
Peak SAR (extrapolated) = 0.112 W/kg
SAR(1 g) = 0.095 W/kg; SAR(10 g) = 0.076 W/kg
Maximum value of SAR (measured) = 0.0986 W/kg




               0 dB = 0.0991 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.16

LTE Band 40_10MHz_QPSK_1RB_0Offset_Left Cheek_Ch38750



Communication System: UID 0, LTE (0); Frequency: 2310 MHz;Duty Cycle: 1:1
Medium: HSL_2300 Medium parameters used: f = 2310 MHz; σ = 1.707 S/m; εr = 41.263; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.4 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.55, 7.55, 7.55); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch38750/Area Scan (81x111x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.0788 W/kg

Ch38750/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 2.784 V/m; Power Drift = 0.09 dB
Peak SAR (extrapolated) = 0.137 W/kg
SAR(1 g) = 0.075 W/kg; SAR(10 g) = 0.045 W/kg
Maximum value of SAR (measured) = 0.0812 W/kg




               0 dB = 0.0788 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.16

LTE Band 40_10MHz_QPSK_1RB_49Offset_Left Cheek_Ch39200



Communication System: UID 0, LTE (0); Frequency: 2355 MHz;Duty Cycle: 1:1
Medium: HSL_2300 Medium parameters used: f = 2355 MHz; σ = 1.742 S/m; εr = 41.172; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.4 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.55, 7.55, 7.55); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch39200/Area Scan (81x111x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.0821 W/kg

Ch39200/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 2.958 V/m; Power Drift = -0.09 dB
Peak SAR (extrapolated) = 0.137 W/kg
SAR(1 g) = 0.074 W/kg; SAR(10 g) = 0.044 W/kg
Maximum value of SAR (measured) = 0.0789 W/kg




               0 dB = 0.0821 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.16

LTE Band 41_20MHz_QPSK_1RB_49Offset_Left Cheek_Ch40600



Communication System: UID 0, LTE (0); Frequency: 2591 MHz;Duty Cycle: 1:1
Medium: HSL_2600 Medium parameters used: f = 2591 MHz; σ = 1.997 S/m; εr = 40.27; ρ = 1000
kg/m3
Ambient Temperature:23.1 ℃; Liquid Temperature:22.3 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(6.98, 6.98, 6.98); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch40600/Area Scan (81x111x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.147 W/kg

Ch40600/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 2.888 V/m; Power Drift = 0.15 dB
Peak SAR (extrapolated) = 0.258 W/kg
SAR(1 g) = 0.135 W/kg; SAR(10 g) = 0.075 W/kg
Maximum value of SAR (measured) = 0.146 W/kg




               0 dB = 0.146 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.19

WLAN 2.4GHz_802.11b 1Mbps_Right Tilt_Ch6



Communication System: UID 0, WLAN 2.4GHz 802.11b (0); Frequency: 2437 MHz;Duty Cycle: 1:1
Medium: HSL_2450 Medium parameters used: f = 2437 MHz; σ = 1.828 S/m; εr = 40.968; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.7 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.34, 7.34, 7.34); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch6/Area Scan (81x131x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 1.06 W/kg

Ch6/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 10.43 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 2.25 W/kg
SAR(1 g) = 0.880 W/kg; SAR(10 g) = 0.393 W/kg
Maximum value of SAR (measured) = 0.943 W/kg




               0 dB = 1.06 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.15

GSM850_GPRS(4 TX slots)_Back Side_10mm_Ch251



Communication System: UID 0, GSM850(class 12) (0); Frequency: 848.8 MHz;Duty Cycle: 1:2.08
Medium: MSL_835 Medium parameters used: f = 849 MHz; σ = 0.993 S/m; εr = 55.699; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.6 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(9.24, 9.24, 9.24); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch251/Area Scan (81x161x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.519 W/kg

Ch251/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 15.79 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 0.812 W/kg
SAR(1 g) = 0.471 W/kg; SAR(10 g) = 0.264 W/kg
Maximum value of SAR (measured) = 0.524 W/kg




               0 dB = 0.519 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.16

GSM1900_GPRS(4 TX slots)_Back Side_10mm_Ch661



Communication System: UID 0, PCS1900(class 12) (0); Frequency: 1880 MHz;Duty Cycle: 1:2.08
Medium: MSL_1900 Medium parameters used: f = 1880 MHz; σ = 1.497 S/m; εr = 54.63; ρ = 1000
kg/m3
Ambient Temperature:23.2 ℃; Liquid Temperature:22.7 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.4, 7.4, 7.4); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch661/Area Scan (81x161x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.559 W/kg

Ch661/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 11.35 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 0.856 W/kg
SAR(1 g) = 0.482 W/kg; SAR(10 g) = 0.258 W/kg
Maximum value of SAR (measured) = 0.542 W/kg




               0 dB = 0.559 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.15

LTE Band 5_10MHz_QPSK_1RB_25Offset_Back Side_10mm_Ch20450



Communication System: UID 0, LTE (0); Frequency: 829 MHz;Duty Cycle: 1:1
Medium: MSL_835 Medium parameters used: f = 829 MHz; σ = 0.974 S/m; εr = 55.761; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.6 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(9.24, 9.24, 9.24); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch20450/Area Scan (81x161x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.318 W/kg

Ch20450/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 13.22 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 0.512 W/kg
SAR(1 g) = 0.292 W/kg; SAR(10 g) = 0.163 W/kg
Maximum value of SAR (measured) = 0.327 W/kg




               0 dB = 0.318 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.14

LTE Band 40_10MHz_QPSK_1RB_0Offset_Back Side_10mm_Ch38750



Communication System: UID 0, LTE (0); Frequency: 2310 MHz;Duty Cycle: 1:1
Medium: MSL_2300 Medium parameters used: f = 2310 MHz; σ = 1.691 S/m; εr = 53.432; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.5 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.38, 7.38, 7.38); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch38750/Area Scan (81x161x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.246 W/kg

Ch38750/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 5.692 V/m; Power Drift = -0.15 dB
Peak SAR (extrapolated) = 0.465 W/kg
SAR(1 g) = 0.216 W/kg; SAR(10 g) = 0.096 W/kg
Maximum value of SAR (measured) = 0.250 W/kg




               0 dB = 0.246 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.14

LTE Band 40_10MHz_QPSK_1RB_49Offset_Back Side_10mm_Ch39200



Communication System: UID 0, LTE (0); Frequency: 2355 MHz;Duty Cycle: 1:1
Medium: MSL_2300 Medium parameters used: f = 2355 MHz; σ = 1.749 S/m; εr = 53.361; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.5 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.38, 7.38, 7.38); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch39200/Area Scan (81x161x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.286 W/kg

Ch39200/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 5.111 V/m; Power Drift = -0.11 dB
Peak SAR (extrapolated) = 0.421 W/kg
SAR(1 g) = 0.227 W/kg; SAR(10 g) = 0.111 W/kg
Maximum value of SAR (measured) = 0.261 W/kg




               0 dB = 0.286 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.13

LTE Band 41_20MHz_QPSK_1RB_49Offset_Back Side_10mm_Ch40600



Communication System: UID 0, LTE (0); Frequency: 2591 MHz;Duty Cycle: 1:1
Medium: MSL_2600 Medium parameters used: f = 2591 MHz; σ = 2.06 S/m; εr = 52.58; ρ = 1000
kg/m3
Ambient Temperature:23.2 ℃; Liquid Temperature:22.2 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.01, 7.01, 7.01); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch40600/Area Scan (81x161x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.340 W/kg

Ch40600/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 4.014 V/m; Power Drift = 0.13 dB
Peak SAR (extrapolated) = 0.560 W/kg
SAR(1 g) = 0.301 W/kg; SAR(10 g) = 0.158 W/kg
Maximum value of SAR (measured) = 0.331 W/kg




               0 dB = 0.340 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.19

WLAN 2.4GHz_802.11b 1Mbps Back Side_10mm_Ch6



Communication System: UID 0, WLAN 2.4GHz 802.11b (0); Frequency: 2437 MHz;Duty Cycle: 1:1
Medium: MSL_2450 Medium parameters used: f = 2437 MHz; σ = 1.849 S/m; εr = 53.049; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.5 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.15, 7.15, 7.15); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch6/Area Scan (81x151x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.366 W/kg

Ch6/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 5.283 V/m; Power Drift = 0.07 dB
Peak SAR (extrapolated) = 0.657 W/kg
SAR(1 g) = 0.313 W/kg; SAR(10 g) = 0.148 W/kg
Maximum value of SAR (measured) = 0.351 W/kg




               0 dB = 0.366 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.19

WLAN 2.4GHz_802.11b 1Mbps_Right Tight_Ch6_Repeated



Communication System: UID 0, WLAN 2.4GHz 802.11b (0); Frequency: 2437 MHz;Duty Cycle: 1:1
Medium: HSL_2450 Medium parameters used: f = 2437 MHz; σ = 1.828 S/m; εr = 40.968; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.7 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.34, 7.34, 7.34); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch6/Area Scan (81x131x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 1.05 W/kg

Ch6/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 10.43 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 2.23 W/kg
SAR(1 g) = 0.879 W/kg; SAR(10 g) = 0.392 W/kg
Maximum value of SAR (measured) = 0.941 W/kg




               0 dB = 1.05 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.15

GSM850_GPRS(4 TX slots)_Back Side_10mm_Ch251



Communication System: UID 0, GSM850(class 12) (0); Frequency: 848.8 MHz;Duty Cycle: 1:2.08
Medium: MSL_835 Medium parameters used: f = 849 MHz; σ = 0.993 S/m; εr = 55.699; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.6 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(9.24, 9.24, 9.24); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch251/Area Scan (81x161x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.519 W/kg

Ch251/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 15.79 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 0.812 W/kg
SAR(1 g) = 0.471 W/kg; SAR(10 g) = 0.264 W/kg
Maximum value of SAR (measured) = 0.524 W/kg




               0 dB = 0.519 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.16

GSM1900_GPRS(4 TX slots)_Back Side_10mm_Ch661



Communication System: UID 0, PCS1900(class 12) (0); Frequency: 1880 MHz;Duty Cycle: 1:2.08
Medium: MSL_1900 Medium parameters used: f = 1880 MHz; σ = 1.497 S/m; εr = 54.63; ρ = 1000
kg/m3
Ambient Temperature:23.2 ℃; Liquid Temperature:22.7 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.4, 7.4, 7.4); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch661/Area Scan (81x161x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.559 W/kg

Ch661/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 11.35 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 0.856 W/kg
SAR(1 g) = 0.482 W/kg; SAR(10 g) = 0.258 W/kg
Maximum value of SAR (measured) = 0.542 W/kg




               0 dB = 0.559 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.15

LTE Band 5_10MHz_QPSK_1RB_25Offset_Back Side_10mm_Ch20450



Communication System: UID 0, LTE (0); Frequency: 829 MHz;Duty Cycle: 1:1
Medium: MSL_835 Medium parameters used: f = 829 MHz; σ = 0.974 S/m; εr = 55.761; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.6 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(9.24, 9.24, 9.24); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch20450/Area Scan (81x161x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.318 W/kg

Ch20450/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 13.22 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 0.512 W/kg
SAR(1 g) = 0.292 W/kg; SAR(10 g) = 0.163 W/kg
Maximum value of SAR (measured) = 0.327 W/kg




               0 dB = 0.318 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.14

LTE Band 40_10MHz_QPSK_1RB_0Offset_Back Side_10mm_Ch38750



Communication System: UID 0, LTE (0); Frequency: 2310 MHz;Duty Cycle: 1:1
Medium: MSL_2300 Medium parameters used: f = 2310 MHz; σ = 1.691 S/m; εr = 53.432; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.5 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.38, 7.38, 7.38); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch38750/Area Scan (81x161x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.246 W/kg

Ch38750/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 5.692 V/m; Power Drift = -0.15 dB
Peak SAR (extrapolated) = 0.465 W/kg
SAR(1 g) = 0.216 W/kg; SAR(10 g) = 0.096 W/kg
Maximum value of SAR (measured) = 0.250 W/kg




               0 dB = 0.246 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.14

LTE Band 40_10MHz_QPSK_1RB_49Offset_Back Side_10mm_Ch39200



Communication System: UID 0, LTE (0); Frequency: 2355 MHz;Duty Cycle: 1:1
Medium: MSL_2300 Medium parameters used: f = 2355 MHz; σ = 1.749 S/m; εr = 53.361; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.5 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.38, 7.38, 7.38); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch39200/Area Scan (81x161x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.286 W/kg

Ch39200/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 5.111 V/m; Power Drift = -0.11 dB
Peak SAR (extrapolated) = 0.421 W/kg
SAR(1 g) = 0.227 W/kg; SAR(10 g) = 0.111 W/kg
Maximum value of SAR (measured) = 0.261 W/kg




               0 dB = 0.286 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.13

LTE Band 41_20MHz_QPSK_1RB_49Offset_Back Side_10mm_Ch40600



Communication System: UID 0, LTE (0); Frequency: 2591 MHz;Duty Cycle: 1:1
Medium: MSL_2600 Medium parameters used: f = 2591 MHz; σ = 2.06 S/m; εr = 52.58; ρ = 1000
kg/m3
Ambient Temperature:23.2 ℃; Liquid Temperature:22.2 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.01, 7.01, 7.01); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch40600/Area Scan (81x161x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.340 W/kg

Ch40600/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 4.014 V/m; Power Drift = 0.13 dB
Peak SAR (extrapolated) = 0.560 W/kg
SAR(1 g) = 0.301 W/kg; SAR(10 g) = 0.158 W/kg
Maximum value of SAR (measured) = 0.331 W/kg




               0 dB = 0.340 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.19

WLAN 2.4GHz_802.11b 1Mbps Back Side_10mm_Ch6



Communication System: UID 0, WLAN 2.4GHz 802.11b (0); Frequency: 2437 MHz;Duty Cycle: 1:1
Medium: MSL_2450 Medium parameters used: f = 2437 MHz; σ = 1.849 S/m; εr = 53.049; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.5 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.15, 7.15, 7.15); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch6/Area Scan (81x151x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.366 W/kg

Ch6/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 5.283 V/m; Power Drift = 0.07 dB
Peak SAR (extrapolated) = 0.657 W/kg
SAR(1 g) = 0.313 W/kg; SAR(10 g) = 0.148 W/kg
Maximum value of SAR (measured) = 0.351 W/kg




               0 dB = 0.366 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.19

WLAN 2.4GHz_802.11b 1Mbps_Right Tight_Ch6_Repeated



Communication System: UID 0, WLAN 2.4GHz 802.11b (0); Frequency: 2437 MHz;Duty Cycle: 1:1
Medium: HSL_2450 Medium parameters used: f = 2437 MHz; σ = 1.828 S/m; εr = 40.968; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.7 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(7.34, 7.34, 7.34); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn480; Calibrated: 2019.04.11
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch6/Area Scan (81x131x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 1.05 W/kg

Ch6/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 10.43 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 2.23 W/kg
SAR(1 g) = 0.879 W/kg; SAR(10 g) = 0.392 W/kg
Maximum value of SAR (measured) = 0.941 W/kg




               0 dB = 1.05 W/kg