Plots of System Performance Check - SAR

FCC ID: ZNFX440IM

Test Report

Download: PDF
FCCID_4415186

   ORLAB,
   '                                                                            REPORT No. : $Z19060060S01


Annex C Plots of System Performance Check




           SHENZHEN MORLAB COMMUNICATIONS TECHNOLOGY Co., Ltd.                  Tel: 86—755—36898555   Fax: 86—766—36008525
M o RLAB   FL1—3, Building A, FeiYang Science Park, No.8 LongChang Road,
           Block67, BacAn Distrct, ShenZhen , GuangDong Province, P. R. China   Hiteitwvwmortab.on     E—mail: service@morlab.on


Test Laboratory: SAR Lab, of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.16


System Check_835MHz_Head_190616



Communication System: UID 0, CW (0); Frequency: 835 MHz;Duty Cycle: 1:1
Medium: HSL_835 Medium parameters used: {= 835 MHz; 0 = 0.922 $/m; e, = 42.273; p = 1000
kg/m3
Ambient Temperature : 23.2 °C; Liquid Temperature : 22.5 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3823; ConvF(9.32, 9.32, 9.32); Calibrated: 2018.11.12;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn480; Calibrated: 2019.04.11
— Phantom: SAM 2; Type: QDO0OPA40CC; Serial: TP:1464
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW835/Area Scan (§1x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.45 W/kg

CW835/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=Smm
Reference Value= 53.72 V/m; Power Drift=—0.14 dB
Peak SAR (extrapolated) = 3.28 W/kg
SAR(I g) = 2.26 W/kg: SAR(1O g) = 1.5 W/kg
Maximum value of SAR (measured) = 2.43 W/kg

   dB
   0


    —2.00


    —3.99


    ~5.99


    ~7.98


    —9.98
               0 dB =2.45 Wikg


Test Laboratory: SAR Lab, of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.17


System Check_1900MHz_Head_190617



Communication System: UID 0, CW (0); Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: HSL_1900 Medium parameters used: £= 1900 MHz; a= 1.355 S/m; s = 41.316; p = 1000
kg/m3
Ambient Temperature : 23.4 °C; Liquid Temperature : 22.3 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3823; ConvF(7.68, 7.68, 7.68); Calibrated: 2018.11.12;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn480; Calibrated: 2019.04.11
— Phantom: SAM 2; Type: QDO0OPA40CC; Serial: TP:1464
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW 1900/Area Scan (81x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 10.6 W/kg

CW 1900/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=86.62 V/m; Power Drift= 0.17 dB
Peak SAR (extrapolated) = 17.5 W/kg
SAR(I g) = 9.59 W/kg: SAR(1O g) = 4.99 W/kg
Maximum value of SAR (measured) = 10.4 W/kg

   dB
   0


    ~3.65


    —7.29


    —10.94


    —14.58


    —18.23
               0 dB = 10.6 Wikg


Test Laboratory: SAR Lab, of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.16


System Check_Head_2300MHz_190616



Communication System: UID 0, CW (0); Frequency: 2300 MHz;Duty Cycle: 1:1
Medium: HSL_2300 Medium parameters used: £= 2300 MHz; 0 = 1.698 S/m; & = 41.26; p = 1000
kg/m3
Ambient Temperature : 23.3 °C; Liquid Temperature : 224 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3823; ConvF(7.55, 7.55, 7.55); Calibrated: 2018.11.12;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn480; Calibrated: 2019.04.11
— Phantom: SAM 2; Type: QDO0OPA40CC; Serial: TP:1464
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW2300/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 13.3 W/kg

CW2300/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=Smm, dy=5mm, dz=5mm
Reference Value=73.98 V/m; Power Drift= 0.19 dB
Peak SAR (extrapolated) = 17.9 W/kg
SAR(I g) = 12.4 W/kg: SAR(1O g) = 5.59 W/kg
Maximum value of SAR (measured) = 15.0 W/kg

   dB
   0


    ~3.49


    —6.99


    —10.48


    13.98


    —17.47
               0 dB = 13.3 Wikg


Test Laboratory: SAR Lab, of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.19


System Check_2450MHz_Head_190619



Communication System: UID 0, CW (0); Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: HSL_2450 Medium parameters used: £= 2450 MHz; 0 = 1.838 S/m; e = 40.92; p = 1000
kg/m3
Ambient Temperature : 23.3 °C; Liquid Temperature : 22.7 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3823; ConvF(7.34, 7.34, 7.34); Calibrated: 2018.11.12;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn480; Calibrated: 2019.04.11
— Phantom: SAM 2; Type: QDO0OPA40CC; Serial: TP:1464
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW 2450/Area Scan (101x101x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 15.4 W/kg

CW 2450/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=Smm, dz=Smm
Reference Value=90.11 V/m; Power Drift= 0.07 dB
Peak SAR (extrapolated) = 28.3 W/kg
SAR(I g) = 13.4 W/kg: SAR(1O g) = 6.14 W/kg
Maximum value of SAR (measured) = 15.3 W/kg

   dB
   0


    4.43

    —8.87

    —13.30

    —17.74


    —22.17
               0 dB =15.4 Wikg


Test Laboratory: SAR Lab, of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.16


System Check_2600MHz_Head_190616



Communication System: UID 0, CW (0); Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: HSL_2600 Medium parameters used: £= 2600 MHz; a =2.005 S/m; e = 40.198; p = 1000
kg/m3
Ambient Temperature : 23.1 °C; Liquid Temperature : 22.3 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3823; ConvF(6.98, 6.98, 6.98); Calibrated: 2018.11.12;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn480; Calibrated: 2019.04.11
— Phantom: SAM 2; Type: QDO0OPA40CC; Serial: TP:1464
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW 2600/Area Scan (101x101x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 16.7 W/kg

CW 2600/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=Smm, dz=Smm
Reference Value=87.06 V/m; Power Drift= 0.03 dB
Peak SAR (extrapolated) =31.9 W/kg
SAR(I g) = 14.3 W/kg: SAR(1O g) = 6.26 W/kg
Maximum value of SAR (measured) = 16.2 W/kg

   dB
   0


    ~4.91


    —9.82


    —14.72


    —19.63


    —24.54
               0 dB = 16.7 Wikg


Test Laboratory: SAR Lab, of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.15


System Check_835MHz_Body_190615



Communication System: UID 0, CW (0); Frequency: 835 MHz;Duty Cycle: 1:1
Medium: MSL_835 Medium parameters used: £= 835 MHz; 0 = 0.98 S/m; & = 55.749; p= 1000
kg/m3
Ambient Temperature : 23.3 °C; Liquid Temperature : 22.6 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3823; ConvF(9.24, 9.24, 9.24); Calibrated: 2018.11.12;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn480; Calibrated: 2019.04.11
— Phantom: SAM 2; Type: QDO0OPA40CC; Serial: TP:1464
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW835/Area Scan (§1x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.56 W/kg

CW835/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=Smm
Reference Value=51.37 V/m; Power Drift= 0.01 dB
Peak SAR (extrapolated) = 3.44 W/kg
SAR(I g) = 2.38 W/kg: SAR(1O g) = 1.59 W/kg
Maximum value of SAR (measured) = 2.56 W/kg

   dB
   0


    —1.98


    ~3.96


    ~5.94


    —7.92                                            /


    —9.90
               0 dB =2.56 Wikg = 4.08 dBWkg


Test Laboratory: SAR Lab, of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.16


System Check_1900MHz_Body_190616



Communication System: UID 0, CW (0); Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: MSL_1900 Medium parameters used: £=1900 MHz; 0 = 1.518 $/m; £,= 54.556; p = 1000
kg/m3
Ambient Temperature : 23.2 °C; Liquid Temperature : 22.7 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3823; ConvF(7.4,7.4, 7.4); Calibrated: 2018.11.12;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn480; Calibrated: 2019.04.11
— Phantom: SAM 2; Type: QDO0OPA40CC; Serial: TP:1464
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW 1900/Area Scan (81x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 11.0 W/kg

CW 1900/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value= 84.47 V/m; Power Drift= 0.02 dB
Peak SAR (extrapolated) = 17.2 W/kg
SAR(I g) = 9.58 W/kg: SAR(1O g) = 5.01 W/kg
Maximum value of SAR (measured) = 10.8 W/kg

   dB
   0


    ~3.40


    —6.80


    —10.20


    13.60


    —17.00
               0 dB =11.0 Wikg


Test Laboratory: SAR Lab, of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.14


System Check_2300MHz_Body_190614



Communication System: UID 0, CW (0); Frequency: 2300 MHz;Duty Cycle: 1:1
Medium: MSL_2300 Medium parameters used: £=2300 MHz; 0 = 1.769 $/m; & = 53.478; p = 1000
kg/m3
Ambient Temperature : 23.3 °C; Liquid Temperature : 22.5 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3823; ConvF(7.38, 7.38, 7.38); Calibrated: 2018.11.12;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn480; Calibrated: 2019.04.11
— Phantom: SAM 1; Type: QD00OP40CC; Serial: TP:1471
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW2300/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 13.2 W/kg

CW2300/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=74.36 V/m; Power Drift=0.11 dB
Peak SAR (extrapolated) = 17.7 W/kg
SAR(I g) = 12.51 W/kg: SAR(1O g) = 5.67 W/kg
Maximum value of SAR (measured) = 14.9 W/kg

   dB
   0


    ~3.49


    —6.98


    ~10.47


    13.96


    —17.45
               0 dB =13.2 W/kg=11.21 dBW/kg


Test Laboratory: SAR Lab, of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.19


System Check_2450MHz_Body_190619



Communication System: UID 0, CW (0); Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: MSL_2450 Medium parameters used: £=2450 MHz; 0 = 1.898 S/m; s = 53.046; p = 1000
kg/m3
Ambient Temperature : 23.3 °C; Liquid Temperature : 22.5 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3823; ConvF(7.15, 7.15, 7.15); Calibrated: 2018.11.12;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn480; Calibrated: 2019.04.11
— Phantom: SAM 1; Type: QD00OP40CC; Serial: TP:1471
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW 2450/Area Scan (101x101x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 14.3 W/kg

CW 2450/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=Smm, dz=Smm
Reference Value=87.03 V/m; Power Drift= 0.02 dB
Peak SAR (extrapolated) = 24.5 W/kg
SAR(I g) = 12.3 W/kg: SAR(1O g) = 5.73 Wkg
Maximum value of SAR (measured) = 14.2 W/kg

   dB
   0


    415


    —8.31


    ~12.46


    —16.62


    —20.77
               0 dB = 14.3 Wikg


Test Laboratory: SAR Lab, of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2019.06.13


System Check_2600MHz_Body_190613


Communication System: UID 0, CW (0); Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: MSL_2600 Medium parameters used: £=2600 MHz; 0 =2.17 S/m; & = 52.511; p= 1000
kg/m3
Ambient Temperature : 23.2 °C; Liquid Temperature : 22.2 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3823; ConvF(7.01, 7.01, 7.01); Calibrated: 2018.11.12;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn480; Calibrated: 2019.04.11
— Phantom: SAM 1; Type: QD00OP40CC; Serial: TP:1471
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW 2600/Area Scan (101x101x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 14.7 W/kg

CW 2600/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=Smm, dz=Smm
Reference Value=83.27 V/m; Power Drift= 0.03 dB
Peak SAR (extrapolated) = 26.2 W/kg
SAR(1 g) = 13.2 W/kg; SAR(10 g) = 5.96 W/kg
Maximum value of SAR (measured) = 14.6 W/kg

   dB
   0


    ~4.52


    —9.04


    ~13.57


    —18.09


    —22.61
               0 dB = 14.7 Wikg



Document Created: 2019-08-28 18:42:12
Document Modified: 2019-08-28 18:42:12

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC