RF Exposure Info 5

FCC ID: ZNFX320PM

RF Exposure Info

Download: PDF
FCCID_4428039









APPENDIX D: SAR TISSUE SPECIFICATIONS

Measurement Procedure for Tissue verification:
                  1) The network analyzer and probe system was configured and calibrated.
                  2) The probe was immersed in the tissue. The tissue was placed in a nonmetallic container.
                     Trapped air bubbles beneath the flange were minimized by placing the probe at a slight angle.
                  3) The complex admittance with respect to the probe aperture was measured
                  4) The complex relative permittivity ε’ can be calculated from the below equation (Pournaropoulos
                     and Misra):
                         j 2ωε r ε 0                 b     b   π               [                      ]
                                                                           exp − jωr ( µ 0ε r' ε 0 )1 / 2
                     Y=
                        [ln(b a )]2                 ∫ ∫ ∫
                                                    a      a   0
                                                                   cos φ ′
                                                                                     r
                                                                                                          dφ ′dρ ′dρ
where Y is the admittance of the probe in contact with the sample, the primed and unprimed coordinates refer to
source and observation points, respectively, r 2 = ρ 2 + ρ ′2 − 2 ρρ ′ cos φ ′ , ω is the angular frequency, and          j = −1 .




                                                         Figure D-1
                                          Composition of Tissue Equivalent Matter
      Note: Liquid recipes are proprietary SPEAG. Since the composition is approximate to the actual liquids utilized, the
                             manufacturer tissue-equivalent liquid data sheets are provided below.




                                                                                                                  Approved by:
      FCC ID: ZNFX320PM                                             SAR EVALUATION REPORT
                                                                                                                  Quality Manager
      Test Dates:                       DUT Type:                                                                 APPENDIX D:
      07/22/19-08/12/19                 Portable Handset                                                           Page 1 of 3
© 2019 PCTEST Engineering Laboratory, Inc.                                                                         REV 21.3 M
                                                                                                                    02/15/2019


                                                               Figure D-2
                                               750 – 5800 MHz Body Tissue Equivalent Matter


                                                                                              Approved by:
      FCC ID: ZNFX320PM                                      SAR EVALUATION REPORT
                                                                                              Quality Manager
      Test Dates:                       DUT Type:                                             APPENDIX D:
      07/22/19-08/12/19                 Portable Handset                                       Page 2 of 3
© 2019 PCTEST Engineering Laboratory, Inc.                                                     REV 21.3 M
                                                                                                02/15/2019


                                                               Figure D-3
                                               750 – 5800 MHz Head Tissue Equivalent Matter



                                                                                              Approved by:
      FCC ID: ZNFX320PM                                      SAR EVALUATION REPORT
                                                                                              Quality Manager
      Test Dates:                       DUT Type:                                             APPENDIX D:
      07/22/19-08/12/19                 Portable Handset                                       Page 3 of 3
© 2019 PCTEST Engineering Laboratory, Inc.                                                     REV 21.3 M
                                                                                                02/15/2019


APPENDIX E: SAR SYSTEM VALIDATION

Per FCC KDB Publication 865664 D02v01r02, SAR system validation status should be documented to confirm
measurement accuracy. The SAR systems (including SAR probes, system components and software versions)
used for this device were validated against its performance specifications prior to the SAR measurements.
Reference dipoles were used with the required tissue- equivalent media for system validation, according to the
procedures outlined in FCC KDB Publication 865664 D01v01r04 and IEEE 1528-2013. Since SAR probe
calibrations are frequency dependent, each probe calibration point was validated at a frequency within the valid
frequency range of the probe calibration point, using the system that normally operates with the probe for routine
SAR measurements and according to the required tissue-equivalent media.

A tabulated summary of the system validation status including the validation date(s), measurement frequencies,
SAR probes and tissue dielectric parameters has been included.

                                                                    Table E-1
                                                        SAR System Validation Summary – 1g
  SAR                                                              COND.    PERM.                  CW VALIDATION                       MOD. VALIDATION
SYSTEM FREQ. [MHz]       DATE        PROBE SN   PROBE CAL. POINT                                       PROBE       PROBE     MOD.
                                                                    (σ)      (εr)    SENSITIVITY                                        DUTY FACTOR      PAR
   #                                                                                                 LINEARITY   ISOTROPY    TYPE
    E          750       2/6/2019      3589       750      Head     0.891   43.677      PASS            PASS       PASS       N/A            N/A           N/A
    E          835       2/6/2019      3589       835      Head     0.922   43.470      PASS            PASS       PASS      GMSK           PASS           N/A
    O          835       8/7/2019      7538       835      Head     0.926   42.299      PASS            PASS       PASS      GMSK           PASS           N/A
    O          1750      8/7/2019      7538      1750      Body     1.370   40.428      PASS            PASS       PASS       N/A            N/A           N/A
    G          1900      7/3/2019      7409      1900      Head     1.460   40.732      PASS            PASS       PASS      GMSK           PASS           N/A
    H          1900      7/16/2019     7406      1900      Head     1.437   38.944      PASS            PASS       PASS      GMSK           PASS           N/A
    D          1900      5/20/2019     3914      1900      Head     1.454   40.608      PASS            PASS       PASS      GMSK           PASS           N/A
    E          2450      2/5/2019      3589      2450      Head     1.825   39.836      PASS            PASS       PASS     OFDM/TDD        PASS         PASS
    E          2450      8/8/2019      7547      2450      Head     1.802   37.994      PASS            PASS       PASS     OFDM/TDD        PASS         PASS
    E          2600      2/7/2019      3589      2600      Head     1.964   40.460      PASS            PASS       PASS       TDD           PASS           N/A
    G          750       7/23/2019     7409       750      Body     0.941   56.074      PASS            PASS       PASS       N/A            N/A           N/A
    O          835       7/3/2019      7538       835      Body     0.972   55.349      PASS            PASS       PASS      GMSK           PASS           N/A
    I          1750      5/21/2019     7357      1750      Body     1.442   55.384      PASS            PASS       PASS       N/A            N/A           N/A
    G          1750      7/11/2019     7409      1750      Body     1.445   53.920      PASS            PASS       PASS       N/A            N/A           N/A
    J          1900      2/8/2019      7488      1900      Body     1.571   52.538      PASS            PASS       PASS      GMSK           PASS           N/A
    I          2450      5/16/2019     7357      2450      Body     2.014   53.910      PASS            PASS       PASS     OFDM/TDD        PASS         PASS
    I          2600      5/15/2019     7357      2600      Body     2.162   53.620      PASS            PASS       PASS       TDD           PASS           N/A




NOTE: While the probes have been calibrated for both CW and modulated signals, all measurements were
performed using communication systems calibrated for CW signals only. Modulations in the table above represent
test configurations for which the measurement system has been validated per FCC KDB Publication 865664
D01v01r04 for scenarios when CW probe calibrations are used with other signal types. SAR systems were
validated for modulated signals with a periodic duty cycle, such as GMSK, or with a high peak to average ratio (>5
dB), such as OFDM according to FCC KDB Publication 865664 D01v01r04.




                                                                                                                                         Approved by:
        FCC ID: ZNFX320PM                                           SAR EVALUATION REPORT
                                                                                                                                         Quality Manager
        Test Dates:                      DUT Type:                                                                                        APPENDIX E:
        07/22/19 - 08/12/19             Portable Handset                                                                                   Page 1 of 1
© 2019 PCTEST Engineering Laboratory, Inc.                                                                                                REV 21.3 M
                                                                                                                                           02/15/2019



Document Created: 2019-09-04 11:04:43
Document Modified: 2019-09-04 11:04:43