Test Report_20170227_v1 - ZNFX240H_FCC RFExp_Appenidx A_20170227

FCC ID: ZNFX240H

RF Exposure Info

Download: PDF
FCCID_3301908

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/21


System Check_Head_750MHz
DUT: D750V3­1012

Communication System: CW; Frequency: 750 MHz;Duty Cycle: 1:1
Medium: HSL_750_170121 Medium parameters used: f = 750 MHz; σ = 0.9 S/m; εr = 43.374; ρ = 1000
kg/m3
A i       T            :23.4 ℃ ; Li i T                   :22.4 ℃

DASY5 Configuration:
­ Probe: EX3DV4 ­ SN3955; ConvF(10.65, 10.65, 10.65); Calibrated: 2016/11/24;
­ Sensor­Surface: 1.4mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn1399; Calibrated: 2016/11/17
­ Phantom: SAM_Right; Type: QD000P40CD; Serial: TP:1383
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.75 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 56.31 V/m; Power Drift = ­0.01 dB
Peak SAR (extrapolated) = 3.16 W/kg
SAR(1 g) = 2.17 W/kg; SAR(10 g) = 1.42 W/kg
Maximum value of SAR (measured) = 2.73 W/kg




                   0 dB = 2.73 W/kg = 4.36 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/20


System Check_Body_750MHz
DUT: D750V3­1012

Communication System: CW; Frequency: 750 MHz;Duty Cycle: 1:1
Medium: MSL_750_170120 Medium parameters used: f = 750 MHz; σ = 0.973 S/m; εr = 53.826; ρ = 1000
kg/m3
A i       T            :23.8 ℃ ; Li i T                   :22.8 ℃

DASY5 Configuration:
­ Probe: EX3DV4 ­ SN3753; ConvF(9.42, 9.42, 9.42); Calibrated: 2016/5/11;
­ Sensor­Surface: 1.4mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn1388; Calibrated: 2016/10/10
­ Phantom: SAM_Left; Type: QD 000 P40 CB; Serial: TP­1478
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.03 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 58.67 V/m; Power Drift = ­0.11 dB
Peak SAR (extrapolated) = 3.40 W/kg
SAR(1 g) = 2.31 W/kg; SAR(10 g) = 1.56 W/kg
Maximum value of SAR (measured) = 3.02 W/kg




                   0 dB = 3.02 W/kg = 4.80 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/20


System Check_Head_835MHz
DUT: D835V2­499

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: HSL_850_170120 Medium parameters used: f = 835 MHz; σ = 0.895 S/m; εr = 42.956; ρ = 1000
kg/m3
A i       T            :23.2 ℃ ; Li i T                   :22.2 ℃

DASY5 Configuration:
­ Probe: ES3DV3 ­ SN3270; ConvF(6.03, 6.03, 6.03); Calibrated: 2016/8/26;
­ Sensor­Surface: 3mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn778; Calibrated: 2016/5/12
­ Phantom: SAM_Left; Type: QD 000 P40 CB; Serial: TP­1478
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.01 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 58.16 V/m; Power Drift = ­0.01 dB
Peak SAR (extrapolated) = 3.66 W/kg
SAR(1 g) = 2.48 W/kg; SAR(10 g) = 1.63 W/kg
Maximum value of SAR (measured) = 2.88 W/kg




                   0 dB = 2.88 W/kg = 4.59 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/20


System Check_Head_835MHz
DUT: D835V2­499

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: HSL_850_170120 Medium parameters used: f = 835 MHz; σ = 0.895 S/m; εr = 42.956; ρ = 1000
kg/m3
A i       T            :23.2 ℃ ; Li i T                   :22.2 ℃

DASY5 Configuration:
­ Probe: EX3DV4 ­ SN3955; ConvF(10.33, 10.33, 10.33); Calibrated: 2016/11/24;
­ Sensor­Surface: 1.4mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn1399; Calibrated: 2016/11/17
­ Phantom: SAM_Left; Type: QD 000 P40 CB; Serial: TP­1478
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.09 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 61.18 V/m; Power Drift = ­0.08 dB
Peak SAR (extrapolated) = 3.48 W/kg
SAR(1 g) = 2.32 W/kg; SAR(10 g) = 1.54 W/kg
Maximum value of SAR (measured) = 3.06 W/kg




                   0 dB = 3.06 W/kg = 4.86 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/19


System Check_Body_835MHz
DUT : D835V2­499

Communication System: CW ; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: MSL_850_170119 Medium parameters used: f = 835 MHz; σ = 0.938 S/m; εr = 55.617; ρ = 1000
kg/m3
A i       T            :23.8 ℃ ; Li i T                   :22.8 ℃

DASY5 Configuration:
­ Probe: EX3DV4 ­ SN3697; ConvF(8.79, 8.79, 8.79); Calibrated: 2016/10/25;
­ Sensor­Surface: 1.4mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn1388; Calibrated: 2016/10/10
­ Phantom: SAM_Left; Type: QD 000 P40 CB; Serial: TP­1478
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.98 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 58.85 V/m; Power Drift = ­0.14 dB
Peak SAR (extrapolated) = 3.42 W/kg
SAR(1 g) = 2.34 W/kg; SAR(10 g) = 1.55 W/kg
Maximum value of SAR (measured) = 2.92 W/kg




                   0 dB = 2.92 W/kg = 4.65 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/19


System Check_Body_835MHz
DUT: D835V2­499

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: MSL_850_170119 Medium parameters used: f = 835 MHz; σ = 0.938 S/m; εr = 55.617; ρ = 1000
kg/m3
A i       T            :23.8 ℃ ; Li i T                   :22.8 ℃

DASY5 Configuration:
­ Probe: EX3DV4 ­ SN3753; ConvF(9.27, 9.27, 9.27); Calibrated: 2016/5/11;
­ Sensor­Surface: 1.4mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn1388; Calibrated: 2016/10/10
­ Phantom: SAM_Left; Type: QD 000 P40 CB; Serial: TP­1478
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.22 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 60.27 V/m; Power Drift = 0.11 dB
Peak SAR (extrapolated) = 3.70 W/kg
SAR(1 g) = 2.41 W/kg; SAR(10 g) = 1.59 W/kg
Maximum value of SAR (measured) = 3.23 W/kg




                   0 dB = 3.23 W/kg = 5.09 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/21


System Check_Head_1750MHz
DUT: D1750V2­1068

Communication System: CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: HSL_1750_170121 Medium parameters used: f = 1750 MHz; σ = 1.359 S/m; εr = 39.828; ρ = 1000
kg/m3
A i       T            :23.2 ℃ ; Li i T                   :22.2 ℃

DASY5 Configuration:
­ Probe: ES3DV3 ­ SN3270; ConvF(5.21, 5.21, 5.21); Calibrated: 2016/8/26;
­ Sensor­Surface: 3mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn778; Calibrated: 2016/5/12
­ Phantom: SAM_Right; Type: QD000P40CD; Serial: TP:1383
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 11.4 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 92.67 V/m; Power Drift = ­0.07 dB
Peak SAR (extrapolated) = 15.1 W/kg
SAR(1 g) = 8.91 W/kg; SAR(10 g) = 4.88 W/kg
Maximum value of SAR (measured) = 10.9 W/kg




                   0 dB = 10.9 W/kg = 10.37 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2017/1/26

System Check_Head_1750MHz

DUT: D1750V2-1068

Communication System: CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: HSL_1750_170126 Medium parameters used: f = 1750 MHz; σ = 1.365 S/m; εr = 39.707; ρ =
1000 kg/m3
                    23.4                      22.4

DASY5 Configuration
- Probe: EX3DV4 - SN3955; ConvF(8.68, 8.68, 8.68); Calibrated: 2016/11/24;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1399; Calibrated: 2016/11/17
- Phantom: SAM_RIGHT; Type: QD000P40CD; Serial: 1719
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7373)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 14.0 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 102.2 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 16.1 W/kg
SAR(1 g) = 9.06 W/kg; SAR(10 g) = 4.9 W/kg
Maximum value of SAR (measured) = 13.7 W/kg




                0 dB = 13.7 W/kg = 11.37 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/22


System Check_Body_1750MHz
DUT: D1750V2­1068

Communication System: CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: MSL_1750_170122 Medium parameters used: f = 1750 MHz; σ = 1.508 S/m; εr = 54.719; ρ =
1000 kg/m3
A i T                  :23.4 ℃ ; Li i T                   :22.4 ℃

DASY5 Configuration:
­ Probe: ES3DV3 ­ SN3270; ConvF(4.95, 4.95, 4.95); Calibrated: 2016/8/26;
­ Sensor­Surface: 3mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn778; Calibrated: 2016/5/12
­ Phantom: SAM_Right; Type: QD000P40CD; Serial: TP:1383
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 12.7 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 99.87 V/m; Power Drift = ­0.10 dB
Peak SAR (extrapolated) = 16.1 W/kg
SAR(1 g) = 8.91 W/kg; SAR(10 g) = 4.77 W/kg
Maximum value of SAR (measured) = 12.6 W/kg




                   0 dB = 12.6 W/kg = 11.00 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab    Date: 2017/1/26


System Check_Body_1750MHz
DUT: D1750V2­1068

Communication System: CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: MSL_1750_170126 Medium parameters used: f = 1750 MHz; σ = 1.498 S/m; εr = 54.639; ρ
= 1000 kg/m3
A i T                  :23.5 ℃; Li i T                   :22.5 ℃

DASY5 Configuration:
­ Probe: EX3DV4 ­ SN3925; ConvF(8.3, 8.3, 8.3); Calibrated: 2016/5/26;
­ Sensor­Surface: 1.4mm (Mechanical Surface Detection)
­ Electronics: DAE3 Sn495; Calibrated: 2016/5/27
­ Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1644
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 13.7 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 95.58 V/m; Power Drift = 0.11 dB
Peak SAR (extrapolated) = 16.6 W/kg
SAR(1 g) = 9.49 W/kg; SAR(10 g) = 5.15 W/kg
Maximum value of SAR (measured) = 14.3 W/kg




               0 dB = 14.3 W/kg = 11.55 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/21


System Check_Head_1900MHz
DUT: D1900V2­5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: HSL_1900_170121 Medium parameters used: f = 1900 MHz; σ = 1.419 S/m; εr = 39.244; ρ = 1000
kg/m3
A i       T            :23.2 ℃ ; Li i T                   :22.2 ℃

DASY5 Configuration:
­ Probe: ES3DV3 ­ SN3270; ConvF(5.08, 5.08, 5.08); Calibrated: 2016/8/26;
­ Sensor­Surface: 3mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn778; Calibrated: 2016/5/12
­ Phantom: SAM_Right; Type: QD000P40CD; Serial: TP:1383
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 12.9 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 97.56 V/m; Power Drift = ­0.02 dB
Peak SAR (extrapolated) = 17.3 W/kg
SAR(1 g) = 9.93 W/kg; SAR(10 g) = 5.29 W/kg
Maximum value of SAR (measured) = 12.1 W/kg




                   0 dB = 12.1 W/kg = 10.83 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2017/2/13

System Check_Head_1900MHz

DUT: D1900V2-5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: HSL_1900_170213 Medium parameters used: f = 1900 MHz;              = 1.416 S/m;   r   = 40.239;
 = 1000 kg/m3
Ambient Temperature 23.3    Liquid Temperature 22.3

DASY5 Configuration:
- Probe: EX3DV4 - SN3925; ConvF(8.3, 8.3, 8.3); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE3 Sn495; Calibrated: 2016/5/27
- Phantom: SAM_Right; Type: SM000T01DA; Serial: TP:1303
- Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7373)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 15.6 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 108.2 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 18.6 W/kg
SAR(1 g) = 10 W/kg; SAR(10 g) = 5.23 W/kg
Maximum value of SAR (measured) = 15.2 W/kg




               0 dB = 15.2 W/kg = 11.82 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/22


System Check_Body_1900MHz
DUT: D1900V2­5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: MSL_1900_170122 Medium parameters used: f = 1900 MHz; σ = 1.552 S/m; εr = 54.953; ρ =
1000 kg/m3
A i T                  :23.4 ℃ ; Li i T                   :22.4 ℃

DASY5 Configuration:
­ Probe: ES3DV3 ­ SN3270; ConvF(4.7, 4.7, 4.7); Calibrated: 2016/8/26;
­ Sensor­Surface: 3mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn778; Calibrated: 2016/5/12
­ Phantom: SAM_Right; Type: QD000P40CD; Serial: TP:1383
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 13.2 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 93.85 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 17.4 W/kg
SAR(1 g) = 10.1 W/kg; SAR(10 g) = 5.36 W/kg
Maximum value of SAR (measured) = 12.6 W/kg




                   0 dB = 12.6 W/kg = 11.00 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2017/2/13

System Check_Body_1900MHz

DUT: D1900V2-5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: MSL_1900_170213 Medium parameters used: f = 1900 MHz;              = 1.578 S/m;   r   = 54.868;
 = 1000 kg/m3
Ambient Temperature 23.3   Liquid Temperature 22.3

DASY5 Configuration:
- Probe: EX3DV4 - SN3925; ConvF(8, 8, 8); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE3 Sn495; Calibrated: 2016/5/27
- Phantom: SAM_Right; Type: SM000T01DA; Serial: TP:1303
- Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7373)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 16.5 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 105.7 V/m; Power Drift = -0.13 dB
Peak SAR (extrapolated) = 18.4 W/kg
SAR(1 g) = 10.2 W/kg; SAR(10 g) = 5.35 W/kg
Maximum value of SAR (measured) = 15.5 W/kg




               0 dB = 15.5 W/kg = 11.90 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2017/2/24

System Check_Head_2450MHz

DUT: D2450V2-926

Communication System: CW ; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: HSL_2450_170224 Medium parameters used: f = 2450 MHz;              = 1.789 S/m;   r   = 39.045;
 = 1000 kg/m3
                    23.6                      22.6

DASY5 Configuration:
- Probe: EX3DV4 - SN3578; ConvF(7.28, 7.28, 7.28); Calibrated: 2016/5/11;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn914; Calibrated: 2017/1/6
- Phantom: SAM_Right; Type: SM000T01DA; Serial: TP:1303
- Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7373)


Pin=250mW/Area Scan (81x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 21.7 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 112.4 V/m; Power Drift = -0.16 dB
Peak SAR (extrapolated) = 26.6 W/kg
SAR(1 g) = 13.2 W/kg; SAR(10 g) = 6.15 W/kg
Maximum value of SAR (measured) = 21.9 W/kg




               0 dB = 21.9 W/kg = 13.40 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/2/24

System Check_Body_2450MHz

DUT: D2450V2-926

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: MSL_2450_170224 Medium parameters used: f = 2450 MHz; σ = 1.97 mho/m; εr = 54.3; ρ
= 1000 kg/m3
A i T                  :23.3 ℃; Li i T                    :22.3 ℃

DASY4 Configuration:
- Probe: EX3DV4 - SN3925; ConvF(7.64, 7.64, 7.64); Calibrated: 2016/5/26
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE3 Sn495; Calibrated: 2016/5/27
- Phantom: SAM_Right; Type: QD000P40CD; Serial: TP:1383
- ;Postprocessing SW: SEMCAD, V1.8 Build 159


Pin=250mW/Area Scan (71x71x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (interpolated) = 21.0 mW/g

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 106.0 V/m; Power Drift = 0.032 dB
Peak SAR (extrapolated) = 26.3 W/kg
SAR(1 g) = 12.8 mW/g; SAR(10 g) = 5.97 mW/g
Maximum value of SAR (measured) = 21.4 mW/g




               0 dB = 21.4mW/g


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/20


System Check_Head_2600MHz
DUT: D2600V2­1008

Communication System: CW; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: HSL_2600_170120 Medium parameters used: f = 2600 MHz; σ = 2.01 S/m; εr = 37.83; ρ = 1000
kg/m3
A i       T            :23.5 ℃ ; Li i T                   :22.5 ℃

DASY5 Configuration:
­ Probe: ES3DV3 ­ SN3270; ConvF(4.37, 4.37, 4.37); Calibrated: 2016/8/26;
­ Sensor­Surface: 3mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn778; Calibrated: 2016/5/12
­ Phantom: SAM_Right; Type: QD000P40CD; Serial: TP:1383
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (81x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 21.4 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 103.6 V/m; Power Drift = ­0.02 dB
Peak SAR (extrapolated) = 33.5 W/kg
SAR(1 g) = 15.5 W/kg; SAR(10 g) = 6.99 W/kg
Maximum value of SAR (measured) = 20.8 W/kg




                   0 dB = 20.8 W/kg = 13.18 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/19


System Check_Body_2600MHz
DUT: D2600V2­1008

Communication System: CW ; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: MSL_2600_170119 Medium parameters used: f = 2600 MHz; σ = 2.192 S/m; εr = 52.56; ρ = 1000
kg/m3
A i       T            :23.5 ℃ ; Li i T                   :22.5 ℃

DASY5 Configuration:
­ Probe: EX3DV4 ­ SN3931; ConvF(7.46, 7.46, 7.46); Calibrated: 2016/10/3;
­ Sensor­Surface: 1.4mm (Mechanical Surface Detection)
­ Electronics: DAE3 Sn577; Calibrated: 2016/9/28
­ Phantom: SAM_Left; Type: QD 000 P40 CB; Serial: TP­1478
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 22.1 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 99.92 V/m; Power Drift = 0.17 dB
Peak SAR (extrapolated) = 30.0 W/kg
SAR(1 g) = 14.1 W/kg; SAR(10 g) = 6.31 W/kg
Maximum value of SAR (measured) = 21.9 W/kg




                   0 dB = 21.9 W/kg = 13.40 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab     Date: 2017/1/19


System Check_Body_2600MHz
DUT: D2600V2­1008

Communication System: CW; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: MSL_2600_170119 Medium parameters used: f = 2600 MHz; σ = 2.192 S/m; εr = 52.56; ρ = 1000
kg/m3
A i       T            :23.5 ℃ ; Li i T                   :22.5 ℃

DASY5 Configuration:
­ Probe: ES3DV3 ­ SN3270; ConvF(4.12, 4.12, 4.12); Calibrated: 2016/8/26;
­ Sensor­Surface: 3mm (Mechanical Surface Detection)
­ Electronics: DAE4 Sn778; Calibrated: 2016/5/12
­ Phantom: SAM_Left; Type: QD 000 P40 CB; Serial: TP­1478
­ Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (71x71x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 20.2 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 95.89 V/m; Power Drift = ­0.02 dB
Peak SAR (extrapolated) = 30.8 W/kg
SAR(1 g) = 14.4 W/kg; SAR(10 g) = 6.46 W/kg
Maximum value of SAR (measured) = 19.4 W/kg




                   0 dB = 19.4 W/kg = 12.88 dBW/kg