LG-X220dsh_TestRpt_RFExp Part 2

FCC ID: ZNFX220DSH

RF Exposure Info

Download: PDF
FCCID_2929607

Calibration Laboratory of                                                     Schweizerischer Kalibrierdienst
Schmid & Partner                                                         S
                                                                         ,.   Service suisse d'6talonnage
 Engineering AG                                                          r'   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                             S    Swiss Galibration Service


Accredited by the Swiss Accreditation Service (SAS)                      Accreditation Ho.:   SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

crienr      Spofton-KS (Auden)


                                                                                Schweizerischer Kalibrierdienst
                                                                                Service suisse d‘étalonnage
                                                                                Servizio svizzero di taratura
                                                                                Swiss Calibration Service

Accredited by the Swiss Accreditation Service (SAS)                          Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured
Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
       devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
         communication devices used in close proximity to the human body (frequency range of 30
         MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    *    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    *    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    *    SAR measured: SAR measured at the stated antenna input power.
    *    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurementis stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: DB35V2—40091_Nov15                             Page 2 of 8


   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY5                              V52.8.8
     Extrapolation                                   Advanced Extrapolation

     Phantom                                          Modular Flat Phantom
     Distance Dipole Center — TSL                              15 mm                            with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                          835 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C          41.5               0.90 mho/m
      Measured Head TSL parameters                             (22.0 £0.2) °C   42.6 £ 6 %         0.92 mho/m x 6 %
      Head TSL temperature change during test                     <0.5 °C                                   ——


SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                    2.91 Whkg
     SAR for nominal Head TSL parameters                     normalized to 1W       9.14 Wikg 2 17.0 % (k=2)


     SAR averaged over 10 cm(10 g) of Head TSL                  condition
      SAR measured                                       250 mW input power                    1.50 Wig
      SAR for nominal Head TSL parameters                    normalized to 1W       5.94 Wikg 2 16.5 % (k=2)



Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity
      Nominal Body TSL parameters                                 22.0 °C          55.2               0.97 mho/m
      Measured Body TSL parameters                             (22.0 £0.2) °C   55.6 16 %          0.99 mho/m + 6 %
      Body TSL temperature change during test                     <0.5°C            ——


SAR result with Body TSL

      SAR averaged over 1 cm° (1 g) of Body TSL                 Condition
      SAR measured                                       250 mW input power                    2.42 Wikg
      SAR for nominal Body TSL parameters                    normalized to 1W       9.55 Wikg x 17.0 % (k=2)


      SAR averaged over 10 cm° (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                    1.59 Wig
      SAR for nominal Body TSL parameters                    normalized to 1W        6.29 Wikg 2 16.5 % (k=2)




Certificate No: DB35V2—40091_Nov15                    Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

    lmpedance, transformed to feed point                  51.3O-4.3jC)
    Return Loss


DASY5 Validation Report for Head TSL

                                                                                   Date: 24.11.2015

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 40091

Communication System: UID 0 — CW; Frequency: 835 MHz
Medium parameters used: £ = 835 MHz; o = 0.92 S/m; & = 42.6; p = 1000 kg/m?
Phantom section: Flat Section
Measurement Standard: DASY3 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN7349; ConyF(9.77, 9.77, 9.77); Calibrated: 30.12.2014;

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 Sn601; Calibrated: 17.08.2015

       Phantom: Flat Phantom 4.9L; Type: QDOOOP49AA; Serial: 1001

       DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 60.87 V/m; Power Drift = —0.00 dB
Peak SAR (extrapolated) = 3.43 W/kg
SAR(I g) = 2.31 W/kg; SAR(10 g) = 1.5 W/kg
Maximum value of SAR (measured) =3.05 W/kg




                            1e
                                     0 dB =3.05 W/kg = 4.84 dB W/kg




Certificate No: DB35V2—40091_Nov15                  Page 5 of 8


                                                                 24 Nov 2015    15:46:17
          CHJ s1i     1 U FS          1: 51.328 &   —4.3398 n 43.920 pF        §35.000 00d MHz
                                                    po



          Del

          Ca




          A¥g
          16


          H1d


          CH2




          CaA




          Aw
          16

          H1d

                    START 635.000 000 MHz                             STOP 1 035.000 000 MHz




Certificate No: D835V2—4d091_Nov15                     Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                   Date: 24.11.2015

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 40091

Communication System: UID 0 — CW; Frequency: 835 MHz
Medium parameters used: £ = 835 MHz; 0 = 0.99 $/m; & = 55.6; p = 1000 kg/m?
Phantom section: Flat Section
Measurement Standard: DASYS (IEEEAEC/ANSI C63.19—2011)

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(9.72, 9.72, 9.72); Calibrated: 30.12.2014;

   *    Sensor—Surface; 1.4mm (Mechanical Surface Detection)

   +    Electronics: DAE4 $n601; Calibrated: 17.08.2015

   +    Phantom: Flat Phantom 4.9L; Type: QDOOOP49AA; Serial: 1001

   «_   DASY32 52.8.8(1222); SEMCAD X 14.6.10(7331)


Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 59.69 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 3.58 W/kg
SAR(I g) = 2.42 W/kg:; SAR(10 g) = 1.59 W/kg
Maximum value of SAR (measured) = 3.19 W/kg




                            uss
                                     0 dB =3.19 W/kg = 5.04 dBW/kg




Certificate No: DB35V2—40091_Nov15                 Page 7 of 8


                                                                        24 Now 2015     11144153
          CH #11      4 L FS          1: 47.342 a       —6.3320 m   30.102 pF          835.000 Agd MHz
                                                        eC_
                                                  Pag                *~,
                                             L.             *
          *                                 * *           _ h — _ &3 A_\"
          Del


          CA




          CH2




          CaA




          Aw
          159



          H1d

                    START 635.000 000 MHz                                       STOP 1 ©35.000 000 MHz




Certificate No: D835V2—4d091_Nov15                       Page 8 of 8


Calibration Laboratory of                                                     SchweizerischerKalibrierdienst
                                                                         S
Schmid & Partner                                                         ,.   Service suisse d'6talonnage
 Engineering AG                                                          \'   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                             S    Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                      Acqeditation   tlo.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certilicates

crient      Sporton-KS (Auden)


Calibration Laboratory of
Schmid & Paftner
 Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzerland


Accredited by the Swiss Accreditation Service (SAS)
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for lhe recognition of calibration ceriificates

Glossary:
TSL
ConvF
N/A


   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASYS                              V52.8.8
     Extrapolation                                   Advanced Extrapolation
     Phantom                                         Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                            with Spacer

     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         1900 MHz a 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity

     Nominal Head TSL parameters                                  22.0 °C          40.0               1.40 mho/m
     Measured Head TSL parameters                              (22.0 £0.2) °C   89.4 £ 6 %         1.39 mho/m 2 6 %
     Head TSL temperature change during test                      <0.5 °C


SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition

     SAR measured                                        250 mW input power                 9.85 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W        39.4 Wikg x 17.0 % (k=2)

     SAR averaged over 10 cm? (10 g) of Head TSL                condition
     SAR measured                                        250 mW input power                    5.15 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W        20.6 Wikg x 16.5 % (k=2)



Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature                            Conductivity
     Nominal Body TSL parameters                                  22.0 °C          53.3               1.52 mho/m
     Measured Body TSL parameters                              (22.0 £0.2) °C   52.2 26 %          1.52 mho/m 2 6 %
     Body TSL temperature change during test                      <0.5 °C

SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                  Condition

     SAR measured                                        250 mW input power                 10.2 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W        40.6 Wikg x 17.0 % (k=2)

     SAR averaged over 10 cm° (10 g) of Body TSL                 condition
     SAR measured                                        250 mW input power                    5.36 Wkg
      SAR for nominal Body TSL parameters                    normalized to 1W        21.4 Wikg 2 16.5 % (k=2)




Certificate No: D1900V2—5d118_Nov15                   Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

    lmpedance, transformed to feed point                  51.8O   +7.2ja
    Return Loss


DASYS Validation Report for Head TSL

                                                                                Date:23.11.2015

Test Laboratory: SPEAG, Zuich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 - SN: 5d118

Communication System: UID 0 - CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; o = 1.39 S/m; e. = 39.4;P = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63 .19 -201 l)

DASY52 Configuration:

   .   Probe: EX3DV4 - SN7349; ConvF(S.14,8.14,8.14); Calibrated: 30.12.2014;

   .   Sensor-Surface: 1.4mm (Mechanical Surface Detection)

   .   Electronics: DAE4 Sn601; Calibrated: I7.08.2015

   .   Phantom: Flat Phantom 5.0 (front); Type: QD000P50AA; Serial: 1001

   .   DASY52 52.8.8(1222); SEMCAD      x   14.6.t0(733r)


Dipole Calibration for Head Tissue/Pin=250 mW, d-l0mmlZoom Scan (7x7x7)lCube 0:
Measurement grid: dx=5mm, dy=Smm, dz=5mm
Reference Value = 107.5 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 13.5 W&g
SAR(I 8) = 9.85 Wkg; SAR(10 8) = 5.15 Wkg
Maximum value of SAR (measured) = 15.1 Wkg




                               0 dB = 15.1 Wkg = 11.79 dBWkg


                                                                      23 Nov 2015     16:02:12
          [CHI) Ss11     1 U OFS         1i 51.795 &   7.1895 4   602.23 pH          1 900.000 dao MHz
                                                       r—p——




          De l

          C




          Av q
          16


          H1 d


          CH2




          Del

          CaA




          Av
          169



          H1d

                       START 1 700.000 ado MHz                                STOP 2 100.000 000 MHz




Certificate No: D1900V2—5d118_Nov15                     Page 6 of 8


DASYS Validation Report for Body TSL

                                                                             Date:23.11.2015

Test Laboratory: SPEAG, Zunch, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 - SN: 5d118

Communication System: UID 0 - CW; Frequency: 1900 MHz
Mediumparametersused: f = 1900MHz; o=1.52 S/m; e. =52.2; p = 1000kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63 .T9 -201 1)

DASY52 Configuration:

   .   Probe: EX3DV4 - SN7349; ConvF(7.9,7.9,7.9); Calibrated: 30.12.2014;

   .   Sensor-Surface: 1.4mm (Mechanical Surface Detection)

   .   Electronics: DAE4 Sn601; Calibrated: 17.08.2015

   .   Phantom: Flat Phantom 5.0 (back); Type: QD000P50AA; Serial: 1002

   .   DASY5Z 52.8.8(1222); SEMCAD X 14.6.10(7331)


Dipole Calibration for Body Tissue/Pin=250 mW, d=l0mmlZoom Scan (7x7x7YCube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 105.1 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 18.3 Wkg
SAR(1 B) = 10.2 Wkg; SAR(10 8) = 5.36 Wkg
Maximum value of SAR (measured) = 15.5 Wkg


                                                                                                               23 Nov 2015               16:01:49
          CHJ s1i     £ UOFS          1: 47.791 &                               7.7031 a                   645.26 pH                   1 900.000 000 MHz

                                                                      C                                **_
                                                                A                            $         w         >
          %                                              5                              Fa       :F’        C ’\
                                                     /                              2                         /            % %
          Del                                 ',-f                            y §                  ¥    &   ’5 ~— _ ’}\\

                                          /£                                  $oCk               . 03            .             C3
                                                                                                                                A
          Ca                              {                               1              V                           k ~ .
                                      L                               |                                      ~ >
                                                                                                        *    '- ?"   _ \



                                      |                               ;,                     T             2 k s9
                                                                                             \               * teg #


          fAvg                                %                               «/                 * 6*~
                                                                                                   &     — /4
          16°                                  *%                             * .
                                                                                                       C
                                                                                                       \   /
                                                     %                                                                     &
                                                             '\\—‘{                                          u2" 4         7




          H1d

          CH2




          Del

          CA




          Aw
          159



          H1d


                    START 1 700.000 ddo MHz                                                                                      STOP 2 100.000 ooo MHz




Certificate No: D1900V2—5d118_Nov15                                                     Page 8 of 8


Calibration Laboratory of                                               S    SchweizerischerKalibrierdienst
Schmid & Partner                                                        A    Service suisse d'6talonnage
                                                                        ts   Servizio svizzero di taratura
 Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzerland                            S    Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                     Accreditation   tlo.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration cedificates

crient      Sporton-KS (Auden)


                                                                                  Schweizerischer Kalibrierdienst
                                                                              S
                                                                                  Service suisse d‘étalonnage
                                                                              C   Servizio svizzero di taratura
                                                                              S   Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured
Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
      devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
         communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    *    Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    *    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    *     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
    *     SAR measured: SAR measured at the stated antenna input power.
          SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    «_    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurementis stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2450V2—840_Nov15                               Page 2 of 8



Document Created: 2017-11-20 16:52:02
Document Modified: 2017-11-20 16:52:02

© 2025 FCC.report
This site is not affiliated with or endorsed by the FCC