I17Z62297-SEM01_SAR_Rev3 Part3

FCC ID: ZNFX210FM

RF Exposure Info

Download: PDF
FCCID_3771649

TTL                                                                                    1

                 TTL in Collaboration with a


         Add: No.S1 Xueyuan Road, Haidian Distict, Bljng, 100191, China
         Tel:+86—10—62304633—2218       Faxc +86—10—0230463302200
         E—mail: ctl@chinatt.com            nifiwichinat en

Glossary:
TSL                   tissue simulating liquid
NORMxy.z              sensitivityin free space
ConvE                 sensitivity in TSL / NORMx.yz
DCP                   diode compression point
CF                    crest factor (1/duty_cycle) of the RF signal
AB,C,D                modulation dependent linearzation parameters
Polarization ©        © rotation around probe axis
Polarization 0        8 rotation around an axis that is in the plane normal to probe axis (at measurement center), i
                8=0 is normal to probe axis
Connector Angle information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged
   Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
   Measurement Techniques‘, June 2013
b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held devices used
   in close proximity to the ear (frequency range of 300MHz to 3GHz)®, February 2005
) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the humen body (frequency range of 30 MHz to 6 GHz)®, March
     2010
d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
*     NORMxy,z: Assessed for E—field polarization 6=0 (fs900MHz in TEM—cell 1:> 1800MHz: waveguide).
      NORMx,y,z are only intermediate values, ie., the uncertainties of NORMx,y,z does not effect the
      E field uncertainty inside TSL (see below ConvF).
*     NORM(Mxy,z = NORMxy,z* frequency_response (see Frequency Response Chart). This
      linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of the
      frequency responseis included in the stated uncertainty of ConvF.
*     DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep
      (no uncertainty required). DCP does not depend on frequency nor media.
*     PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
      characteristics.
*     Auy.zBuyz; Coy,2;VRuy.zAB.C are numerical lingarization parameters assessed based on the
      data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
      media. VR is the maximum calibration range expressed in RMS voltageacross the diode.
*     ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
      Transfer Standard for s800MHz) and inside waveguide using analytical feld dlistributions based on
      power measurements for f »800MHz. The same setups are used for assessment of the parameters
      applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
      These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
      The sensitivity in TSL corresponds to NORMx,y,z" ConvF whereby the uncertainty corresponds to
      that given for ConvF. A frequency dependent Conve is used in DASY version 4.4 and higher which
      allows extending the validity from:50MHztox100MHz.
*     Spherical isotropy (3D deviation from isotropy); in a field of low gradients realized using a flat
      phantom exposed by a patch antenna.
«_    Sensor Offset: The sensoroffset corresponds to the offset of virtual measurement center from the
      probe tip (on probe axis). No tolerance required.
*     Connector Angle: The angle is assessed using the information gained by determining the NORMx
      (no uncertainty required).


Certificate No: Z16—9725 1                            Page 2 of 11


                                   in Collaboration with
                 TrL CauBRATION
                     a          LazoRArORY
         Add: No.51 Xueywian Road, Haidian Distict, Beffng, 100191, China
         Tel: +86—10—62300633—2218       Fmc +86—10—62304633—02200
         E—mail: ctl@chinatl.com         Hipo/wwwrchinatlen




                       Probe EX3DV4


                                             SN: 3846


                                      Calibrated: January 13, 2017

                               Calibrated for DASY/EASY Systems
                                   (Note: non—compatible with DASY2 system!)




Certificate No: Z16—97251                             Page 3 of 11


        Add:No 51 Xueyuan Road, Haidian Distict, Bing, 100191, China
        Tels +86—10—62304633—2218       Pno+86—10—62304633—2200
        E—mail: ctl@chinatl.com        Hitpo/wwuchinat

     DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3846

Basic Calibration Parameters
                                    Sensor X                Sensor Y         Sensor 2       Une (k=2)
 Norm(gVi(Vim)®)                    0.39                    0.47             0.47           £10.8%
 DGP(mV)°                           99.4                    se.9             so.6

Modulation Calibration Parameters
 uin         Communication                     A               B       6            p      vR        Unc®
             System Name                        dB             dB—pV                dB     mV        (k=2)
 0           cw                        x        _|0.0          0.0     1.0          0.00   175.0     22.1%
                                       Y        0.0            0.0     1.0                 188.3
                                       2        0.0            0.0     1.0                 190.7

     The reported uncertainty of measurement is stated as the standard uncertainty of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probability of approximately 95%.


* The uncertainties of Norm X, Y, Z do not affect the E*—field uncertainty inside TSL (see Page 5 and Page 6)
* Numericallinearization parameter: uncertainty not required.
© Uncertainly is determined using the max. deviation from linear response applying rectangular distribution
and is expressed for the square of the field value.




Certificate No: Z16—97251                               Page 4 of 11


      TL                                                                                         I
‘T./


                                       in Colaboration with


            Add: No.S1 Xueysan Road, Haidian Distict, Beiing, 100191, China
            Tels +86—10—42304633—2218      Fac +86—104230403—2200
            E—mail: ctl@chinatt.com        Hitpshonwcchinathn

       DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3846

 Calibration Parameter Determined in Head Tissue Simulating Media
                                                                                                             s
  Auron Pe:::v";y . c°“"(;’°:')':'y ConvF X ConvF ¥ ConvF Z Alpha® D(:":', :‘::;
    Tso             419                 ose              wes            ass       es         030     o70_|       £12%
     200            a15                 o.97             ase            oss       ess        oi6     127         £12%
    1450            405                 120              s42            s42       s42        o2s     092         £12%
    T750            401                 137              ais            _|_a16_|_a16_|       022     109|        £12%
    1900            400                 140              Tee            18e       18e        023     114         +12%
    2100            so8                 149              Teo            1so       1so        020     118         +12%
    2300            se5                 167              Tas            145       14s        oss     on          £12%
    2450            302                 1.80             Taa            122       122        oas     087         +12%
    2600            se                  196              Tiz            112       112        os2     080         £12%
    s250            359                 an               aar            sar       _|_s3r_|   045     115|        £13%
    5600            355                 5.07             4.72           4.72         4.72    0.45    1.30        £13%
    s750            354                 522              a9s            4ss          495     o4s_|   140         £13%

 © Frequency validity above 300 Mz of +100MHz only applies for DASY v4.4 and higher (Page 2), else it is restrcted to
 #50MHz. The uncertainty is the RSS of Convr uncertainty at calibration frequency and the uncertainty forthe indicated
 frequency band. Frequency validty below 300 MHz is + 10, 25, 40, 50 and 70 MHz for Conv assessments at 30, 64, 128,
 150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to £ 110 MHz.
 "Atfrequency below 3 GHz, the validity of issue paramaters (e and 0) can be relaxed to £10% if liquid compensation
 formula is applied to measured SAR values. Atfrequencies above 3 GHz, the validty of lissue parameters (e and 0) is
 restricted to £5%. The uncertainty is the RSS of the Convuncertainty for indicated target tissue parameters.
 °AiphafDepth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
 effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
 between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




    Certificate No: Z16—9725 1                            Page s of11


           Add: No.51 Xueyuan Road, Haidian Distict, Belng, 100191, China
           Tel: «#6—10—62304633—2218     Fac +86—10—42304633—2200
           E—mail: ctl@chinatl.com       Hitpohewwcchinatt  on

      DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3846

Calibration Parameter Determined in Body Tissue Simulating Media
                                                                                                       s
 t pae® P;::"::“,';y . c°""(;,°:‘)'fy ConvF X Conv¥ Conve Z Alpha® r‘;‘:::; :::;‘
    To             s55                 0.%             ase             ase    s0s    040       oss         £12%
    200            s5.0                105             ase             asr    ose    or1       123         £12%
   1450            540                 1.30            sz              s22    s22    o12       136         |+12%
   1750            534                 149             7so             1so    7so    oz2e      100         £12%
   1900            sas                 152             Tsr             1sr    1st    ois       126         |+12%
   2100            sa2                 162             Tes             res    1es    oir       186         +12%
   2300            520                 181             Tss             Tss    155    os2       ore         £12%
   2450            527                 195             Ta              731    131    oss       oss         +12%
   2600            525                 216             72s             12s    125    oss       o|          £12%
   s250            4s                  536             ass             a0s    495    oso       155|        £13%
   5600            48.5                5.77            4.18            4.18   4.18   0.55       160|       +13%
   s750            483                 5.04            ass             ass    453    oss        198        |+13%

© Frequency validty above 300 MHz of +100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
#50MHz. The uncertainty is the RSS of ConvF uncertainty at calbration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to : 110 MHz.
"Atfrequency below 3 GHz, the validty of tissue paramaters (€ and 0) can be relaxed to £10% if iquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validty of tissue parameters (c and 0)is
restricted to £5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters
° AlphalDepth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




   Certificate No: Z16—97251                            Page 6 of 11


                                                         "   in Colsborationwith
                                           TTL a             CAuBRATIoN LaBorAroRy
                 Add: No.51 Xueyuan Road, Haidian Distict,Beiing, 100191, China
                 Tel: +86—10—62304638—2218                         Faxc +86—10—2304633—2200
                 E—mail: tl@chinat.com                                  4       in

                                                      Frequency Response of E—Field
                                                     (TEM—Cell: ifi110 EXX, Waveguide: R22)

                                         1.5

                                         14
       Frequency response (normalized)




                                         13

                                         12

                                         1.1

                                         1.0

                                         0.9

                                         0.8




                                                                                                                      T
                                               0             500          1000           1500    2000        2500   3000
                                                                                        f [MHz
                                                                   TEM                   Miiz}          &_

                                                   Uncertainty of Frequency Response of E—field: £7.5% (k=2)




Certificate No: Z16—97251                                                      Page 7 of 11


        Add: No.51 Xueyuan Road, Haidian Distrit, Beling, 100191, China
        Tel: +86—10—62304633—2218      Tc +86—100230463302200
        E—mail:cxl@chinatt.com             xt         inatl on

                              Receiving Pattern (®), 0=0°

              f=600 MHz, TEM                                                 f=1800 MHz, R22




       00 Aem ne enentiminines ind titcm—4mmo9cgp


      as            4               i                     i          i       j

                    t               t             t              t                    v         t
                  —1s0            «190           so              o                   so        100        1s0
                            (~+ — 100Mz           >   600MAz             +       1800MHz   *   2500MHz]
                            Uncertainty of Axial Isotropy Assessment: £0.9% (k=2)




Certificate No: 216—97251                             Page 8 of 11


        Add:No.S1 Xueyuan Road, Haidian Distict, Bjng, 100191, China
        Tels +86—10—62304633—2218     Fac +86—10—62304600—2200
        E—mail:ci@chinait.com          Utpshnverchinaen

                                    Dynamic Range f(SARneaq)
                                                (TEM cell, f = 900 MHz)



                        1054



                g 10° 4
                C
                    &
                ®
                i
                3 04
                    2
                £

                        10‘ 4



                        10‘                 i                                T                 y     1
                              10°          10°                10°          10‘             10°      10
                                                              SAR[mW/cm‘]
                                                  CHWJnot compensated   ~#~ compensated
        Error|dB]




                                       t                  t               +—*             r—
                                      10                 10‘          1 o‘                10       10‘
                                                            SAR[mWicm
                                            ot compensated                       * compensated
                                    Uncertainty of Linearity Assessment: £0.9% (k=2)
Certificate No: Z16—97251                              Page 9 of 11


                              *     in Colaboration with
                TTL CALBRATON
                    a         LABORATORY
        Add: No.S1 Xueyuan Road, Haidian Distict, Beiing,100191, China
        Tel: +86—10—62304633—2218        Fax: +86—10—62304633—2200
        E—mail:ctl@chinat.com            Htiponwchinat.en

                         Conversion Factor Assessment
       £=900 MHz, WGLS R9(H_convF)                                    =1750 MHz, WGLS R22(H_convF)
              is |—                           ————                       ja           —    o




                             10
                             os
                             as
                       2A




                                      10 —om om «o oz o                  om on ow om 10
                             Uncertainty of Spherical Isotropy Assessment: £2.8% (K=2)



Certificate No: Z16—9725 1                            Page 10 of 11


          @ I;Cohbonfianwim a

                                  CAuBRATION LAsoRATORY
        Add:No.51 Xueyuan Road, Haidian District, Beljng, 100191, China
        Tels +86—10—62304633—2218     Fac +86—1062304003—2200
        E—mail: tl@chinant.com         Mip:wowwchinatl.en

   DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3846


     Other Probe Parameters
    Sensor Arrangement                                                    Triangular

    Connector Angle (°)                                                        47.9

    Mechanical Surface Detection Mode                                       enabled

    Optical Surface Detection Mode                                          disable

    Probe Overall Length                                                    337mm

    Probe Body Diameter                                                       10mm

    Tip Length                                                                 9mm

    Tip Diameter                                                             2.6mm

    Probe Tip to Sensor X Calibration Point                                    1mm

    Probe Tip to Sensor Y Calibration Point                                    1mm

    Probe Tip to Sensor Z Calibration Point                                    1mm

    Recommended Measurement Distancefrom Surface                             1.4mm




Certificate No: Z16—9725 1                           Page 11 of 11


                                                   No. I17Z62297-SEM01
                                                       Page 126 of 166


ANNEX H        Dipole Calibration Certificate
835 MHz Dipole Calibration Certificate




                                                ©Copyright. All rights reserved by CTTL.


®
Calibration
     4      Laboratory of
Schmid & Partner
  Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzeriand

Accredted by the Swiss Accredtation Sorvice (SAS)
                                                                            s   Sehweizerischer Kallbrierdienst
                                                                            q Service suisse i‘étatonnage
                                                                              Servitio svizzoro di taratura
                                                                            5. swiss Caltbration Service

                                                                             Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the aignatories to the EA
Multlateral Agroomantfor the recogn‘tion ofcallration certfieatos
Glossary:
TSL                        tissue simulating liquid
ConvF                       sensitivity in TSL / NORM x.y,z
NA                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
         300 MHz to 6 GHz)®, July 2016
    c  IEC 62209—2, "Procedureto determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)®, March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   * Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   * Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
     *    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
     *   SAR measured: SAR measured at the stated antenna input power.
     *   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
     *   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95% .




Certficate No: DB3V2—40060_Jut7                               Page 2 of 8


Measurement Conditions
   DASY system confiquration, as far as not given on page 1.
    DASY Version                                             Dasys                              vse.100
    Extrapolation                                    Advanced Extrapolation
    Phantom                                           Medular Fiat Phantom
    Distance Dipole Gnter — TSL                              15 mm                             with Spacer
     Zoom Sean Resolution                              t dy, dz ~5 mm
     Frequency                                         805 MHza 1 MHz



Head TSL parameters
   The following paramaters and calculations were applied.
                                                               Temperature      Permittivity        Conduetivity
     Nominal Head TSL parameters                                 220°0             415               0.90 mhoim
     Measured Head TSL parameters                             @20 02)°C         40826 %           0.91 mhoim6%
     Head TSL temperature change during test                    <o5°c                                   m
SAR result with Head TSL

     SAR averaged over 1 om(1 g) of Head TSL                    Conditin
     SAR measured                                        250 mW input power                297 Who
     SAR for nominal Head TSL parameters                  normalized to 1W          9.37 Whg = 17.0 % (ke2)

     SAR averaged over 10 em(10 g) of Head TSL                  condition
     SAR measured                                        250 mW input power                158 Wha
     SAR for nominal Head TSL paramsters                  normalized to 1W          6.06 Wikg =16.5 % (k=2)

Body TSL parameters
   The following parameters and calculations were apolied.
                                                               Temperature      Permitivity         Conductivity
     Nominal Body TSL parameters                                  22.0°C            562              0.87 mhoim
     Measured Body TSL parameters                              gao02)°C         54626%            1.01 mhoim6%
     Body TSL temperature change during test                     <05°C            ~—
SAR result with Body TSL

     SAR averaged over 1 om? (1 g) of Body TSL                  Gondition
     SAR measured                                        250 mW input power                 243 Whg
     SAR for nominal Body TSL parameters                     normalized to 1W        8.41 Whkg 2 17.0 % (k=2)
     SAR averaged over 10 cm? (10 g) of Body TSL                condition
     SAR measured                                        250 mW input power                    157 whg
     SAR for nominal Body TSL parametars                     normalized to 1W        6.12 Wikg 2 16.5 % (k=2)



Certficate No: DB35V2—40060_Jut17                     Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                          sein—12j0
      Retum Loss                                                                      ~$2.4 dB


Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                          479n—3.0 n
      Retum Loss                                                                      —25.9 0B


General Antenna Parameters and Design

    [Electrical Detay (one direction)                       |                          1302 ns
Alfter long term use with 100W radiated power, only a sight warming of the dipole near the feedpoint can be measured

The dipole is made of standard sermiigid coaxial cable. The center conductor of the feeding line is direcy connected to the
second arm ofthe dipole. The antenna is therefore shor:—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching whon loaded according to the position as explained in the
‘Measurement Condiions* paragraph. The SAR data are not affected by this change. The overall dipole length is stll
according to the Standard.
No excessive force must be applled to the dipole arms, because they might bnd or the soldered connections noar the
feedpoint may be damaged.

Additional EUT Data

      Menutactured by                                                                  sraag
      Menufactured on                                                             November 09, 2007




Certficate No: DB3SV2—44069_Ju17                         Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                      Date: 19.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN:40069

Communication System: UTD 0 — CW; Frequency: 835 MHz
Medium parameters used: (=835 MHz: a=0.91 S/m; &= 40.8; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEEAIEC/ANST C63.19—201 1

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(10.07, 10.07, 10.07); Calibrated: 31.05.2017;

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 4.9 (front); Type: QD OOL P49 AA; Serial: 1001

   «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=3mm, dy=5mm, dz=5mm
Reference Value = 62.08 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 3.65 W/kg
SAR(L g) =2.37 W/kgs SAR(IO g) = 1.53 Wikg
Maximum value of SAR (measured) = 3.21 Wke


        dB
         0

        —2.20

        ~4.40

        —6.60

        —80.00


        +11.00
                  0 dB =3.21 Wikg = 5.07 dBWike




Certficate No: De3SV2—40060_Jut7               Page 5 of 8


Impedance Measurement Plot for Head TSL



                                                        19 w 2eir. coiseres
               san     i uis           u samse rirsee d6241pr      #35.000 ac0 nz
                                             *       *




                     Sthet cas.o00 one nine                     Sto 1 aa5.000 s00 mz




Cortficate No: DB3SV2—44069.Jult7                 Page 6 of 8


(
DASY5 Validation Report for Body TSL



Test Laboratory: SPEAG, Zurich, Switzerland
                                                                                   Date: 19.07,2017



DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN:44069

Communication System: UID 0 — CW: Frequency: 835 MHz
Medium parameters used: {= 835 MHz; a = 1.01 S/m; e; = 54.8; p = 1000 kg/m
Phantom section: Flat Section
Measurement Standard: DASY5 (TEEE/AIEC/ANST C63,19—2011)

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(10.2, 10.2, 10.2); Calibrated: 31.05.2017;

   +    Sensor—Surface: 1.4mm(Mechanical Surface Detection)

   *    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 4.9 (Back); Type: QD 0OR P49 AA; Scrial: 1005

   «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW , d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=3mm, dy=Smm, d=5mm
Reference Value = 59.35 V/m; Power Drift =—0.03 dB
Peak SAR (extrapolated) = 3.67 Wikg
SAR(L g) =2.43 W/kg; SAR(IO g) =      .57 Wike
Maximum value of SAR (measured)        .22 whke

        dB
         0

        2.20

        ~4.40




        —11.00
                   0 dB =3.22 Wike= 5.08 dBWike




Cortficate No: DB3SV2—44060_.Jult?                Page 7 of 8


           *                                                             t
= TTL                                                                           F
mm                                                                                      ~



 Impedance Measurement Plot for Body TSL



                                                        18 aar ceir osierse
           Exn sa      i u‘rs         mamosse —assz5n «0sts pF      a35.000 o00 nuz
                                                *

           ve1                                          +7~.3
           c


           PS
           is                           \                 s



           cue



           ca




           ma
                     Stakt saz.o00 ooo nez                      sroe 1 eas.c00 s00 mz




  Certficate No: DB35V2—44060_Jult7               Page 8 of 8


Calibration Laboratory of                                   «9
                                                            sn                                         Schwoizeriachor Kallorierdienst




                                                                                               © o ®
                                                     )€E
Schmid & Partner




                                                         »,
                                                                                                       Service suisse «‘étalonnage
  Engineering AG                                                                                       Serviclo sviezoro dl taratura




                                                        o
Zeughausstrasse 43, 8004 Zurich, Suitzeriand
                                                            ihien
                                                               haad®                                   Swiss Calibration Service

Accredtad by the Siss Accreditaion Sorvice (GAS)                                                 Acereditation N   : SCs 0108
The Swiss Accreditation Service is one ofthe signatories tothe EA
MultLateral Agreemontfor the recognion ot callration corificates
Client      CTTL—BJ (Auden)                                                          Cerificate No: D1750V2—1003_Jul17

|CALIBRATION CERTIFICATE
 Object                           D1750V2 — SN:1008

 Calbration procedure(e)          QA CAL—O5.19
                                  Calibration procedure for dipole validation kits above 700 MHz



 Calbration date:                 July 21, 2017

 Thiscaibvaton cerficate documents th traceabityto natlonalstandards, which reaize the physical ts of measurements(S1.
 The measremonts and the uncrtantes wih confdence probabily are gven on the folowing pages and are pa of he corifeate.

 Allcalbrations have beanconducted in the clased aboratar faity: anvronment temperature(223)°C and hunidity <70%.
 Caltration Equipment used (MATE crtealfr catraton)
 Pimary Standards                  CB                      Gal Date (Conticate No                           Soeduled Caltration
 Poner meter NBP                   sn: rorrre              ocAp—17 (No. 217—cesevrsen)                      Aocie
 Power sensor NRP—281              sht tosoes              oeAper? (No. atz—caset)                          Apers
 Power sersor NRP—201              sn: rooms               dcApe—7 (No. 217.00520)                          Apers
 Raference 20 0B Attrator          sh: sose (209           0r—apet7 (No.217—00820)                          Aprrs
 Type—N mismatch combination       SN:5047.2/00027         Or—Ap—t7 (No.217—02509)                          Apers
 Reterence Probe EX3DV4            sn: ros                 3t—May—17 (No. EX3—7340Mayt?)                    May—1s
 page                              Sz or                   26Mar—17 (No. DAE+—001_Mart?)                    Maris
 Secondary Standards                e                      Check Date (n house)                             Soheduled Check
 Poer meter EPM—442A                SN: GBsrd8o70          0rOct18 (nhousecheck Oct18)                      in house check Oct18
 Power sensor HP 8481A              Sh: Ussrzeeres         070ct18 (n housecheck Oct18)                     In house check: Oct18
 Power sensor HP 8481A              SN: ivatoneat?         070ct18 (nhousecheck Oct16)                      In house cheok: Oct18
 AF generater nSMT—0                sht tooore             15—jun—15 (n house check Oct16)                  In house checic Oct18
 Network Anatyzer HP 8758E          sn ussrasoses          1EOct0f (inhousecheckOct18)                      In house chedc Oot17
                                    Name                               Function                             Signature
 Calbrated by:                     Michaot Wobr                        Laboretoy Technician                 1&.


 Approved by:                      Katia Pokovie                       Techrical Manager
                                                                                                        Ee
                                                                                                           Issuedt Juty 24, 2017
 Thiscaibrationcorficale shallno be repreduced excopt in ful wthout witten approval of thelaboratoy.

Certficate No: D1750¥2—1008_Jult?                              Page 1 of 8



Document Created: 2018-03-06 16:45:26
Document Modified: 2018-03-06 16:45:26

© 2025 FCC.report
This site is not affiliated with or endorsed by the FCC