I19Z61729-SEM01_SAR_Rev0_7

FCC ID: ZNFX120HM

RF Exposure Info

Download: PDF
FCCID_4476664

Calibration Laboratory of                                   s\“\‘w’/"";            S    Schweizerischer Kalibrierdienst
Schmid & Partner                                           flafi&&                        Service suisse d‘étalonnage
  Engineering AG                                           oe                      C    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                */7\\\\\\@              S   Swiss Calibration Service
                                                               "thilulst


Accredited by the Swiss Accreditation Service (SAS)                                 Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                         tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
         MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle! to the body axis.
   * Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2600V2—1012_Jul18                              Page 2 of 8




                                                                              o opyrrgire im begries revier vex wy sb a ol


—CZLpS
<—                                                                                               J               e




 Measurement Conditions
     DASY system configuration, as far as not given on page 1.
       DASY Version                                              DASY5                                 V52.10.1
       Extrapolation                                   Advanced Extrapolation
       Phantom                                         Modular Flat Phantom
       Distance Dipole Center — TSL                              10 mm                                with Spacer
       Zoom Scan Resolution                              dx, dy, dz =5 mm
       Frequency                                         2600 MHz +1 MHz



 Head TSL parameters
     The following parameters and calculations were applied.
                                                                 Temperature        Permittivity            Conductivity
       Nominal Head TSL parameters                                  22.0 °C              39.0                1.96 mho/m
       Measured Head TSL parameters                              (22.0 £0.2) °C      37.2 26 %           2.02 mho/m + 6 %
       Head TSL temperature change during test                      <0.5 °C               w=                      ——

 SAR result with Head TSL

       SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
       SAR measured                                        250 mW input power                        14.2 W/kg
       SAR for nominal Head TSL parameters                     normalized to 1W           55.4 Wikg + 17.0 % (k=2)

       SAR averaged over 10 cm* (10 g) of Head TSL                condition
       SAR measured                                        250 mW input power                    6.33 W/kg
       SAR for nominal Head TSL parameters                     normalized to 1W           24.9 Wikg # 16.5 % (k=2)


 Body TSL parameters
     The following parameters and calculations were applied.
                                                                 Temperature         Permittivity           Conductivity
       Nominal Body TSL parameters                                  22.0 °C              52.5                2.16 mho/m
       Measured Body TSL parameters                              (22.0 £0.2) °C      51.5 +6 %            2,20 mho/m +# 6 %
       Body TSL temperature change during test                      <0.5 °C               ———


 SAR result with Body TSL

       SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
       SAR measured                                        250 mW input power                        13.7 Wikg
       SAR for nominal Body TSL parameters                     normalized to 1W           54.1 Wikg + 17.0 % (k=2)

       SAR averaged over 10 cm* (10 g) of Body TSL                 condition
       SAR measured                                        250 mW input power                    6.17 Wikg
       SAR for nominal Body TSL parameters                     normalized to 1W           24.5 Wikg £ 16.5 % (k=2)



 Certificate No: D2600V2—1012_Jul18                     Page 3 of 8




                                                                                  Revvpey eng ies e aee ergerew rnevieen e veur wy wor d k.


m©
@777                                                                                               l
_




 Appendix (Additional assessments outside the scope of SCS 01 08)

 Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                           474 Q —7 4 jQ
       Return Loss                                                                       —21.9 dB



 Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                           44.1 Q — 4.9 jQ
       Retum Loss                                                                       —21.8 dB




 General Antenna Parameters and Design

       Electrical Delay (one direction)                                                   1.154 ns

 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
 second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
 are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
 "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
 according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
 feedpoint may be damaged.


 Additional EUT Data

       Manufactured by                                                                    SPEAG
       Manufactured on                                                               October 30, 2007




 Certificate No: D2600V2—1012_Jul18                        Page 4 of 8




                                                                                     C 8 t prgnngenee n en reggree e n e e en e n e e y e es


mm>                                                                     Ds
@777
uo




 DASY5 Validation Report for Head TSL

                                                                                               Date: 26.07.2018

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1012

 Communication System: UID 0 — CW; Frequency: 2600 MHz
 Medium parameters used: £ = 2600 MHz; 0 = 2.02 $/m; & = 37.2; p = 1000 kg/m?*
 Phantom section: Flat Section
 Measurement Standard: DASY5 (IEEE/IEC/ANSI €63.19—2011)

 DASY52 Configuration:

    e   Probe: EX3DV4 — SN7349; ConvF(7.7, 7.7, 7.7) @ 2600 MHz; Calibrated: 30.12.2017

    e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e   Electronics: DAE4 Sn601; Calibrated: 26.10.2017

    e    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

    &    DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


 Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 118.3 V/m; Power Drift = —0.04 dB
 Peak SAR (extrapolated) = 28.3 W/kg
 SAR(I g) = 14.2 W/kg; SAR(10 g) = 6.33 W/kg
 Maximum value of SAR (measured) = 23.7 W/kg


        dB
         0

         —5.20

         —10.40

         —15.60

         —20.80

         —26.00
                    0 dB = 23.7 W/kg = 13.75 dBW/kg




 Certificate No: D2600V2—1012_Jul18              Page 5 of 8




                                                                      S 8 t prgnngeree n en reggree e n e e en n n e e y e e   e


©
Impedance Measurement Plot for Head TSL




 File View Channel Sweep        Calibration   Trace   Scale   Marker   System Window    Help

                                                                                        2.600000 GHz            47.371 0
                                                                                            82358 pF           —7.4326 0
                                                                                      y 2600000 GHz           80.734 mU
                                                                                                                ~105.12°




            Ch1Avg= 20
     Ch1: Start 2.40000 GHz                                                                                   Stop 2.90000 GHz


                                                                                       2.§00000         CHz   —2] 859 dB




     00      Ch 1     &
     Ch1: Start 2.40000 GHz                                                                                   Stop 2.80000 GHz

  Status     CH 1:        511                 C" 1—Port                Avg=20 Delay                                LCL




Certificate No: D2600V2—1012_Jul18                              Page 6 of 8




                                                                                           en peg en e ie e aer eragerew e nevner e neve wey n d a is


8
DASY5 Validation Report for Body TSL



Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1012
                                                                                               Date: 26.07.2018




Communication System: UID 0 — CW; Frequency: 2600 MHz
Medium parameters used: £ = 2600 MHz; 0 = 2.2 S/m; & = 51.5; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI €63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN7349; ConvF(7.81, 7.81, 7.81) @ 2600 MHz; Calibrated: 30.12.2017

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 Sn601; Calibrated: 26.10.2017

        Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

        DASY32 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 107.5 V/m; Power Drift = —0.07 dB
Peak SAR (extrapolated) = 27.7 W/kg
SAR(I g) = 13.7 W/kg; SAR(10 g) = 6.17 W/kg
Maximum value of SAR (measured) = 22.6 W/kg


       dB
        0

        —5.20

        —10.40

        —15.60

        —20.80

        —26.00
                   0 dB = 22.6 W/kg = 13.54 dBW/kg




Certificate No: D2600V2—1012_Jul18             Page 7 of 8




                                                                    h s vepeg eng en n e en cez n neviver ue veve wey w a ib all


©
Impedance Measurement Plot for Body TSL




 File    View Channel Sweep        Calibration Trace   Scale Marker   System   Window   Help

                                                                                       2600000 GHz           44.132 O
                                                                                           12.374 pF        —4.9469 C)
                                                                                     \ 2600000 GHz         81.418 mU
                                                                                                             —136.86 °




                 Ch1Avug= 20
        Ch1: Start 2.40000 GHz   ——                                                                       Stop 2.80000 GHz


                                                                                        2000000     QHz    —2].786 dB




        100

        00
        00
        00
        00
        Ch1: Start 2.40000 GHa   ———                                                                      Stop 2.80000 GHz

   Status       CH 1:    EH                  | C" 1—Port              Avg=20 Delay                             LCL




Certificate No: D2600V2—1012_Jul18                             Page 8 of 8




                                                                                          hypeg enge is e sns rg! rev e nevaer ns wey ns


Calibration Laboratory of                                                                                         Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                  Service suisse d‘étalonnage
  Engineering AG                                                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                      Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                                                        Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client      CTTL (Auden)                                                                      Certificate No: D750V3—1017_Jul19

CALIBRATION CERTIFICATE
 Object                               D750V3 — SN: 1017


 Calibration procedure(s)             QA CAL—O5.v11
                                      Calibration Procedure for SAR Validation Sources between 0.7—3 GHz



 Calibration date:                    July 18, 2019



 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                     ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                       SN: 104778                03—Apr—19 (No. 217—02892/02893)                       Apr—20
 Power sensor NRP—Z91                  SN: 103244                03—Apr—19 (No. 217—02892)                             Apr—20
 Power sensor NRP—Z91                  SN: 103245                03—Apr—19 (No. 217—02893)                             Apr—20
 Reference 20 dB Attenuator            SN: 5058 (20k)            04—Apr—19 (No. 217—02894)                             Apr—20
 Type—N mismatch combination           SN: 5047.2 / 06827        04—Apr—19 (No. 217—02895)                             Apr—20
 Reference Probe EX3DV4                SN: 7349                  29—May—19 (No. EX3—7349_May19)                        May—20
 DAE4                                  SN: 601                   30—Apr—19 (No. DAE4—601_Apr19)                        Apr—20

 Secondary Standards                    ID #                      Check Date (in house)                                Scheduled Check
 Power meter E44198                    SN:     GB39512475        30—Oct—14 (in house check Feb—19)                     In   house   check:   Oct—20
 Power sensor HP 8481A                 SN:     US37292783        07—Oct—15 (in house check Oct—18)                     In   house   check:   Oct—20
 Power sensor HP 8481A                 SN:     MY41092317        07—Oct—15 (in house check Oct—18)                     In   house   check:   Oct—20
 RF generator R&S SMT—06               SN:     100972            15—Jun—15 (in house check Oct—18)                     In   house   check:   Oct—20
 Network Analyzer Agilent E8358A       SN:     US41080477        31—Mar—14 (in house check Oct—18)                     In   house   check:   Oct—19


                                        Name                                  Function                                  Bighalure
 Calibrated by:                        Claudio Leubler                        Laboratory Technician
                                                                                                                                L7

 Approved by:                          Katja Pokovic                          Technical Manager                   /Wé


                                                                                                                      Issued: July 19, 2019
 This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.


Certificate No: D750V3—1017_Jul19                                     Page 1 of 8




                                                                                                       e opyrrgine r un regrre rever vev wy sb e


Calibration Laboratory of                                                      S   Schweizerischer Kalibrierdienst
Schmid & Partner                                                               e   Service suisse d‘étalonnage
   Engineering AG                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                   S   Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x.y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
     SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
     e   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D750V3—1017_Jul19                                Page 2 of 8


¢S)
                                                                                       on e         —           220   _




 Measurement Conditions
    DASY system configuration, as far as not given on page 1.
       DASY Version                                             DASY5                              V52.10.2
       Extrapolation                                  Advanced Extrapolation
       Phantom                                        Modular Flat Phantom
       Distance Dipole Center — TSL                             15 mm                             with Spacer
      Zoom Scan Resolution                              dx, dy, dz =5 mm
       Frequency                                         750 MHz +1 MHz



 Head TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
       Nominal Head TSL parameters                                 22.0 °C           41.9               0.89 mho/m
       Measured Head TSL parameters                             (22.0 + 0.2) °C   42.2 +6 %          0.89 mho/m + 6 %
       Head TSL temperature change during test                     <0.5 °C


 SAR result with Head TSL

       SAR averaged over 1 cm* (1 g) of Head TSL                 Condition
       SAR measured                                       250 mW input power                     2.14 W/kg
       SAR for nominal Head TSL parameters                    normalized to 1W         8.57 W/kg x 17.0 % (k=2)


       SAR averaged over 10 cm* (10 g) of Head TSL                condition
       SAR measured                                       250 mW input power                     1.39 W/kg
       SAR for nominal Head TSL parameters                    normalized to 1W         5.57 Wikg x 16.5 % (k=2)



 Body TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
       Nominal Body TSL parameters                                 22.0 °C           §5.5               0.96 mho/m
       Measured Body TSL parameters                             (22.0 + 0.2) °C   55.1 +6 %          0.96 mho/m + 6 %
       Body TSL temperature change during test                     <0.5 °C            sese                    wes—


 SAR result with Body TSL

       SAR averaged over 1 cm* (1 g) of Body TSL                 Condition
       SAR measured                                       250 mW input power                     2.14 W/kg
       SAR for nominal Body TSL parameters                    normalized to 1W         8.55 W/kg + 17.0 % (k=2)


       SAR averaged over 10 cm* (10 g) of Body TSL                condition
       SAR measured                                       250 mW input power                  141 Wikg
       SAR for nominal Body TSL parameters                    normalized to 1W         5.63 W/kg + 16.5 % (k=2)



 Certificate No: D750V3—1017_Jul19                     Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           53.1 Q — 1.3 jQ
      Return Loss                                                                       —29.6 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           48.9 Q — 4.3 jQ
      Return Loss                                                                       — 27.0 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                       L                           1.041 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG




Certificate No: D750V3—1017_Jul19                        Page 4 of 8


mm>                                                                    2e n n eee br e ecrrke .
777                                                                        e eee e e.




DASY5 Validation Report for Head TSL
                                                                                     Date: 18.07.2019

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 — SN:1017

Communication System: UID 0 — CW; Frequency: 750 MHz
Medium parameters used: f = 750 MHz; 0 = 0.89 S/m; &, = 42.2; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    e    Probe: EX3DV4 — SN7349; ConvF(10.07, 10.07, 10.07) @ 750 MHz; Calibrated: 29.05.2019

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e    Electronics: DAE4 Sn601; Calibrated: 30.04.2019

    e    Phantom: Flat Phantom 4.9 (front); Type: QD OOL P49 AA; Serial: 1001

    «_   DASY52 52.10.2(1504); SEMCAD X 14.6.12(7470)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 59.72 V/m; Power Drift = 0.00 dB
 Peak SAR (extrapolated) =3.21 W/kg
 SAR(I g) = 2.14 W/kg; SAR(10 g) = 1.39 W/kg
 Maximum value of SAR (measured) = 2.84 W/kg


         dB
         0


         —2.00


         —4.00


         —6.00


         —8.00


         —10.00
                    0 dB = 2.84 W/kg = 4.53 dBW/kg




 Certificate No: D750V3—1017_Jul19               Page 5 of 8


()
n
S
Impedance Measurement Plot for Head TSL




       File View Channel Sweep Calibration Trace Scale Marker System Window Help
                                                                                                        53.148 O
                                                                                                       ~1.3174 0
                                                                                                      33.079 mU
                                                                                                        ~21.979°




                     Ch1Avg= 20
           Chi1: Start 550.000 MHz   ——                                                              Stop 950000 MHz


                                                                                  750.000000   MHz    __—2$9.609 dB




           00

           .00
           00
           .00
           00
           Chi: Start

        Status      CH 1:                                         Auvg=20 Delay




 Certificate No: D750V3—1017_Jul19                  Page 6 of 8


&TTL                                                                  se n e oo r l e eckle.




)
                                                                           e



 DASY5 Validation Report for Body TSL
                                                                                    Date: 18.07.2019

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 — SN:1017

 Communication System: UID 0 — CW; Frequency: 750 MHz
 Medium parameters used: £ = 750 MHz; 0 = 0.96 S/m; s = 55.1; p = 1000 kg/m3
 Phantom section: Flat Section
 Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

 DASY352 Configuration:

    e   Probe: EX3DV4 — SN7349; ConvF(10.4, 10.4, 10.4) @ 750 MHz; Calibrated: 29.05.2019

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e    Electronics: DAE4 Sn601; Calibrated: 30.04.2019

    &    Phantom: Flat Phantom 4.9 (Back); Type: QD OOR P49 AA; Serial: 1005

    e    DASY52 52.10.2(1504); SEMCAD X 14.6.12(7470)


 Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 55.74 V/m; Power Drift = —0.02 dB
 Peak SAR (extrapolated) = 3.18 W/kg
 SAR(1 g) = 2.14 W/kg; SAR(10 g) = 1.41 W/kg
 Maximum value of SAR (measured) = 2.84 W/kg


        dB
         0


         —2.00


         ~4.00


         —6.00


         —8.00


         —10.00
                    0 dB = 2.84 W/kg = 4.53 dBW/kg




 Certificate No: D750V3—1017_Jul19              Page 7 of 8


6 S
Impedance Measurement Plot for Body TSL




     Eile View Channel Swesp Calibration Trace icdem!mdow»mh
      I                                                                  _     750.000000 MHz        48.897 O
                                                                                                    —4.2735 0
                                                                                                   44.586 mU
                                                                                                     —101.99 °




                    Ch1Avg= 20
            Chi: Start 550.000 MHz    ——    —                                                     Stop 950.000 MHz


                                                                               750.000000   MHz    —2f.016 dB




            .00

            .00
            .00
            00
            00       Ch 1      =
            Ch1: Start 5$50.000 MHa   ——                                                          Stop 950.000 MHz

          Status    CH 1: En'              ~TJCET—Pot          "Avg=20 Delay                            LCL




Certificate No: D750V3—1017_Jul19                       Page 8 of 8


                                       No. I19Z61729-SEM01
                                         Page 217 of 217


ANNEX I   Accreditation Certificate




                                      ©Copyright. All rights reserved by CTTL.



Document Created: 2019-10-11 13:31:02
Document Modified: 2019-10-11 13:31:02

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC