I19Z61729-SEM01_SAR_Rev0_6

FCC ID: ZNFX120HM

RF Exposure Info

Download: PDF
FCCID_4476663

m
as777
s




 Measurement Conditions
    DASY system configuration, as far as not given on page 1.
       DASY Version                                             DASYS                                    V52.10.1
       Extrapolation                                  Advanced Extrapolation
       Phantom                                         Modular Flat Phantom
       Distance Dipole Center — TSL                             10 mm                                  with Spacer
      Zoom Scan Resolution                              dx, dy, dz =5 mm
       Frequency                                        1900 MHz + 1 MHz



 Head TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature        Permittivity                Conductivity
       Nominal Head TSL parameters                                 22.0 °C               40.0                    1.40 mho/m
       Measured Head TSL parameters                             (22.0 £ 0.2) °C     39.9 £ 6 %              1.34 mho/m +# 6 %
       Head TSL temperature change during test                     «0.5°C                                              ———


 SAR result with Head TSL

      SAR averaged over 1 cm‘ (1 g) of Head TSL                  Condition
      SAR measured                                        250 mW input power                         9.84 Wi/kg
      SAR for nominal Head TSL parameters                     normalized to 1W            40.4 Wikg + 17.0 % (k=2)

       SAR averaged over 10 cm* (10 g) of Head TSL                condition

       SAR measured                                       250 mW input power                         5.23 Wi/kg
       SAR for nominal Head TSL parameters                    normalized to 1W            21.3 Wikg # 16.5 % (k=2)


 Body TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature        Permittivity                Conductivity
       Nominal Body TSL parameters                                 22.0 °C               53.3                    1.52 mho/m
       Measured Body TSL parameters                             (22.0 + 0.2) °C     54.3 £ 6 %              1.46 mho/m + 6 %
       Body TSL temperature change during test                     <0.5 °C                ~——

 SAR result with Body TSL

       SAR averaged over 1 cm* (1 g) of Body TSL                 Condition

       SAR measured                                       250 mW input power                         9.83 W/kg
       SAR for nominal Body TSL parameters                    normalized to 1W            40.4 Wikg x 17.0 % (k=2)


       SAR averaged over 10 cm* (10 g) of Body TSL                condition
       SAR measured                                       250 mW input power                         5.26 Wikg
       SAR for nominal Body TSL parameters                    normalized to 1W            21.4 Wikg x 16.5 % (k=2)




 Certificate No: D1900V2—5d101_Jul18                   Page 3 of 8




                                                                                  won mm t ngr n en n me n nggn n n n n n n n n n n m i


mm>
@7TTL
uy




 Appendix (Additional assessments outside the scope of SCS 0108)

 Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                           50.7 Q + 5.3 jQ
       Return Loss                                                                       —25.6 dB



 Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                           44.9 Q + 6.5 JQ
       Return Loss                                                                      —21.2 dB



 General Antenna Parameters and Design

      rEIectrical Delay (one direction)                                                  1.205 ns


 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
 second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
 are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
 "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
 according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
 feedpoint may be damaged.



 Additional EUT Data

       Manufactured by                                                                    SPEAG
       Manufactured on                                                                March 28, 2008




 Certificate No: D1900V2—5d101_Jul18                      Page 4 of 8




                                                                                      hn nna c ngn c n n e


DASY5 Validation Report for Head TSL

                                                                                            Date: 24.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

Communication System: UID 0 — CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; 0 = 1.34 S/m; & = 39.9; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN7349; ConvF(8.18, 8.18, 8.18) @ 1900 MHz; Calibrated: 30.12.2017

   e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 Sn601; Calibrated: 26.10.2017

        Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

        DASY352 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 110.6 V/m; Power Drift =—0.01 dB
Peak SAR (extrapolated) = 17.9 W/kg
SAR(1 g) = 9.84 W/kg; SAR(10 g) = 5.23 W/kg
Maximum value of SAR (measured) = 15.0 W/kg




                            +12.00


                            +16.00


                            —20.00

                                      0 dB = 15.0 W/kg = 11.76 dBW/kg




Certificate No: D1900V2—5d101_Jul18                  Page 5 of 8




                                                                        mm nama n ng n n ce ce >


Impedance Measurement Plot for Head TSL




          File    View Channel        Sweep Calibration   Trace Scale Marker   System   Window    Help

                                                                                                 1.900000 GHz       50.660 0
                                                                                                     441.05 pH      5.2653 O
                                                                                            y 1900000 GHz        52.645 mU
                                                                                                                    79.066 °




                          ChiAug= 20
                 Ch1: Start 1.70000 GHz     ——                                                                   Stop 2.10000 GHz




                  o       Ch 1 Aug= 120
                      : Start 1.70000 GHz                                                                        Stop 2.10000 GHz




Certificate No: D1900V2—5d101_Jul18                            Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                               Date: 24.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

Communication System: UID 0 — CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; 0 = 1.46 S/m; & = 54.3; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN7349; ConvF(8.15, 8.15, 8.15) @ 1900 MHz; Calibrated: 30.12.2017

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 26.10.2017

        Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

        DASY532 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 105.3 V/m; Power Drift = —0.04 dB
Peak SAR (extrapolated) = 17.2 W/kg
SAR(1 g) = 9.83 W/kg; SAR(10 g) = 5.26 W/kg
Maximum value of SAR (measured) = 14.8 W/kg




                            12.00


                            —16.00


                            —20.00
                                      0 dB = 14.8 W/kg = 11.70 dBW/kg




Certificate No: D1900V2—5d101_Jul18                  Page 7 of 8




                                                                        ~on nana cce c >   &                 —


®
Impedance Measurement Plot for Body TSL




          File View Channel Sweep Calibration Trace Scale Marker System Window Help
                                                                                 1.900000 GHz        44.864 0
                                                                                     547.21 pH       6.5327 O
                                                                               y 1.900000 GHz      87.393 mU
                                                                                                      124.24°




                       Chi1Avg= 20
              Ch1: Start 1.70000 GHz   ——                                                          Stop 2.10000 GHz


                                                                                   00000     GHz   —2].170 dB




              00

              .00
              00
              00
              00       Ch 1     =
              Ch1: Start 1.70000 GHz   ——                                                          Stop 2.10000 GHz

                                                                Avg=20 Delay




Certificate No: D1900V2—5d101_Jul18               Page 8 of 8




                                                                           non nanac en c


)
a
S
(I
 e




          &       5                                                       w
     Cahb"‘aflon Laboratory of                                          $\\\\\-_Z”'9                             S      Schweizerischer Kalibrierdienst
     Schmid & Partner                                                 M                                         c Service suisse d‘étalonnage
       Engineering AG                                                 pe                                               Servizio svizzero di taratura
     Zeughausstrasse 43, 8004 Zurich, Switzerland                      {f/,,/fi\\>3                              S      gwiss Calibration Service
                                                                            oib abe

     Accredited by the Swiss Accreditation Service (SAS)                                                        Accreditation No.: SCS 0108
     The Swiss Accreditation Service is one of the signatories to the EA
     Multilateral Agreementfor the recognition of calibration certificates

     Client       CTTL (Auden)                                                                      Certificate No: D2450V2—853_Jul18

     CALIBRATION CERTIFICATE
      Object                               D2450V2 — SN:853


      Calibration procedure(s)             QA CAL—O05.v10
                                           Calibration procedure for dipole validation kits above 700 MHz



      Calibration date:                    July 24, 2018


      This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
      The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


      All calibrations have been conducted in the closed laboratory facility: environment temperature (22 3)°C and humidity < 70%.


      Calibration Equipment used (M&TE critical for calibration)


      Primary Standards                      ID #                      Cal Date (Certificate No.)                           Scheduled Calibration
      Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                      Apr—19
      Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                            Apr—19
      Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02673)                            Apr—19
      Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                            Apr—19
      Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                            Apr—19
      Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                       Dec—18
      DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                       Oct—18


      Secondary Standards                    ID #                      Check Date (in house)                                Scheduled Check
      Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                    In house check: Oct—18
      Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—16)                    In house check: Oct—18
      Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                    In house check: Oct—18
      RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                    In house check: Oct—18
      Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—17)                    In house check: Oct—18


                                             Name                                  Function                                 Signatuire
      Calibrated by:                        Claudio Leubler                        Laboratory Technician                             a’

                                                                                                                                     Ee
      Approved by:                          Katja Pokovic                          Technical Manager                   W



                                                                                                                           Issued: July 24, 2018
      This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.


     Certificate No: D2450V2—853_Jul18                                     Page 1 of 8


                                                               t
Calibration Laboratory of
     V
Schmid & Partner
                                                            ND
                                                           ila\\é'-{/m%
                                                                             S   Schweizerischer Kalibrierdienst
                                                                             C   Service suisse d‘étalonnage
  Engineering AG                                                 oag             Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                2 /N 5           S   Swiss Calibration Service
                                                             utds AC

Accredited by the Swiss Accreditation Service (SAS)                          Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                         tissue simulating liquid
ConvF                       sensitivity in TSL / NORM x,y,z
N/A                         not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) 1EC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle! to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D2450V2—853_Jul18                              Page 2 of 8


(Ilém’
         Measurement Conditions
            DASY system configuration, as far as not g iven on page1.
              DASY Version                                              DASY5                              V52.10.1
              Extrapolation                                   Advanced Extrapolation
              Phantom                                          Modular Flat Phantom
              Distance Dipole Center — TSL                              10 mm                             with Spacer
              Zoom Scan Resolution                              dx, dy, dz =5 mm
              Frequency                                         2450 MHz &1 MHz



         Head TSL parameters
            The following parameters and calculations were applied.
                                                                        Temperature       Permittivity         Conductivity
              Nominal Head TSL parameters                                  22.0 °C           39.2               1.80 mho/m
              Measured Head TSL parameters                              (22.0 £0.2) °C    37.8 26 %          1.85 mho/m + 6 %
              Head TSL temperature change during test                      <0.5 °C            R———

         SAR result with Head TSL

              SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
              SAR measured                                        250 mW input power                     13.2 W/kg
              SAR for nominal Head TSL parameters                     normalized to 1W        51.7 Wikg + 17.0 % (k=2)

              SAR averaged over 10 cm* (10 g) of Head TSL                condition
              SAR measured                                        250 mW input power                     613 Wikg
              SAR for nominal Head TSL parameters                     normalized to 1W        24.2 Wikg # 16.5 % (k=2)


         Body TSL parameters
            The following parameters and calculations were applied.
                                                                        Temperature       Permittivity         Conductivity
               Nominal Body TSL parameters                                 22.0 °C           527                1.95 mho/m
               Measured Body TSL parameters                             (22.0 + 0.2) °C   51.9 26 %          2.02 mho/m x 6 %
               Body TSL temperature change during test                     <0.5 °C            ~———                    ————


         SAR result with Body TSL

               SAR averaged over 1 cm* (1 g) of Body TSL                 Condition
              SAR measured                                        250 mW input power                     13.1 W/kg

               SAR for nominal Body TSL parameters                    normalized to 1W        51.3 W/kg + 17.0 % (k=2)

               SAR averaged over 10 cm* (10 g) of Body TSL                condition

               SAR measured                                       250 mW input power                 6.10 Wikg
               SAR for nominal Body TSL parameters                    normalized to 1W        24.1 Wikg # 16.5 % (k=2)



         Certificate No: D2450V2—853_Jul18                     Page 3 of 8


(llém’
         Appendix (Additional assessments outside the scope of SCS 0108)

         Antenna Parameters with Head TSL

               Impedance, transformed to feed point                                           54.1 Q + 8.9 jQ
               Return Loss                                                                       —25.2 dB



         Antenna Parameters with Body TSL

               Impedance, transformed to feed point                                           48.5 Q + 6.0 jQ
               Return Loss                                                                      —24.1 dB




         General Antenna Parameters and Design

              | Electrical Delay (one direction)                                                 1.161 ns


         After long term use with 100W radiated power, only a slight warming of the dipole nearthe feedpoint can be measured.

         The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
         second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
         are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
         "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
         according to the Standard.
         No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
         feedpoint may be damaged.



         Additional EUT Data

               Manufactured by                                                                    SPEAG
               Manufactured on                                                              November 10, 2009




         Certificate No: D2450V2—853_Jul18                        Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                    Date: 24.07.2018
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.85 S/m; &, = 37.8; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN7349; ConvF(7.88, 7.88, 7.88) @ 2450 MHz; Calibrated: 30.12.2017

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 26.10.2017

        Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

        DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 115.3 V/m; Power Drift = —0.00 dB
Peak SAR (extrapolated) = 26.1 W/kg
SAR(I g) = 13.2 W/kg; SAR(10 g) = 6.13 W/kg
Maximum value of SAR (measured) = 21.6 W/kg




                             13,20


                             —17.60


                             —22.00

                                      0 dB = 21.6 W/kg = 13.34 dBW/kg




Certificate No: D2450V2—853_Jul18                    Page 5 of 8


©
Impedance Measurement Plot for Head TSL




     File View Channel Sweep Calbration Trace Scale Marker System Window Help.
                                                                        2450000 GHz        54.148 O
                                                                           254.98 pH       3.9251 O
                                                                      y 2450000 GHz      54.794 mU |
                                                                                            41.259 °




                Ch1Avg= 20
         Ch1: Start 2.25000 GHz   ——                                                     Stop 2.65000 GHz


                                                                         2450000   GHz    —2$ 225 dB




         00

         00
         00
         00
         00       Ch 1     =
         Ch1: Start 2.25000 GHz                                                          Stop 2.65000 GHz


       Status    CH 1:




Certificate No: D2450V2—853_Jul18                Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                     Date: 16.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 2.02 $/m; & = 51.9; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY352 Configuration:

    +    Probe: EX3DV4 — SN7349; ConvF(8.01, 8.01, 8.01) @ 2450 MHz; Calibrated: 30.12.2017

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    *    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

    *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

    «_   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 108.0 V/m; Power Drift = —0.03 dB
Peak SAR (extrapolated) = 25.6 W/kg
SAR(I g) =13.1 W/kg; SAR(10 g) = 6.1 W/kg
Maximum value of SAR (measured) = 21.0 W/kg


         dB
          0


         ~4.00


         —8.00


         —12.00


         —16.00


         —20.00
                   0 dB = 21.0 W/kg = 13.22 dBW/kg




Certificate No: D2450V2—853_Jul18               Page 7 of 8


Impedance Measurement Plot for Body TSL




     File View Channel Sweep             Calibration Trace Scale MarkerSystem Window Help .

                                                                                     2450000 GHz      48.464 O
                                                                                        389.27 pH     5.9924 Q
                                                                                     2450000 GHz    62.710 mU
                                                                                                       100.89 °




                   Ch1Aug= 20
           Ch1: Start 2.25000 GHa   ——                                                              Stop 2.65000 GHz


      (Jro.00                                                                        2450000    2   —24.053 0B
         .00
         .00
          00




          .00

          00
          00
          00

           Thi: Star 2.25000 GHe    ——                                                              Stop 2.65000 GHz

       Status     CH 1:             j             Cipa *              ~Avg=20Delay _           se




Certificate No: D2450V2—853_Jul18                            Page 8 of 8


Calibration Laboratory of                                         so
                                                                   t\_z// %,                                S      Schweizerischer Kalibrierdienst
Schmid & Partner                                                 m                                          c      Service suisse d‘étalonnage
  Engineering AG                                                 uk                                                Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                       */fi\\‘v‘                                 S      swiss Calibration Service
                                                                      "lilalst




Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client       CTTL (Auden)                                                                       Certificate No: D2600V2—1012_Jul18

[CALIBRATION CERTIFICATE
 Object                                D2600V2 — SN:1012


 Calibration procedure(s)             QA CAL—05.v10
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                    July 26, 2018


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                       Apr—19
 Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                             Apr—19
 Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02673)                             Apr—19
 Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                             Apr—19
 Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                             Apr—19
 Reference Probe EXSDV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                        Dec—18
 DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                        Oct—18


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: US37292783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
 Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—17)                     In house check: Oct—18


                                        Name                                     Function                               Signature
 Calibrated by:                         Michael Weber                            Laboratory Technician                           H—



 Approved by:                           Katja Pokovic                            Technical Manager                     ////r/_




                                                                                                                       Issued: July 26, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D2600V2—1012_Jul18                                    Page 1 of 8