I19Z60740-SEM01_SAR_Rev3_part4

FCC ID: ZNFX120HM

RF Exposure Info

Download: PDF
FCCID_4335886

©
Impedance Measurement Plot for Head TSL




          File         View Channel    Sweep Calibration   Trace Scale Marker System Window    Help

                                                                                            1.900000 GHz                       50.660 0
                                                                                                441.05 pH                      5.2653 O
                                                                                          & 1900000 GHz                      52.645 mU
                                                                                                                                79.866 °




                            Ch1Aug= 20
                   Ch1: Start 1.70000 GHz     ——                                                                         Stop 2.10000 GHz
                mas        m                man     mumumn:
                ooo        mm               mam
                10.00
                5.00
                0.00
                5.00
                F10.00
           L1s.00
                F20.00
           |
            2s.00
           (La0.00
           (Las.00
           L40.00           chtava= |2o                             _4
           (J      Chi: Start 1.70000 GHa     ——                                                                         Stop 2.10000 GHz




Certificate No: D1900V2—5d101_Jul18                             Page 6 of 8




                                                                                        Nes seperg n ragi en n en n nge i+           Roiny > >


DASY5 Validation Report for Body TSL

                                                                                                 Date: 24.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

Communication System: UID 0 — CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; 0 = 1.46 S/m; & = 54.3; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN7349; ConvF(8.15, 8.15, 8.15) @ 1900 MHz; Calibrated: 30.12.2017

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 Sn601; Calibrated: 26.10.2017

        Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

        DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 105.3 V/m; Power Drift = —0.04 dB
Peak SAR (extrapolated) = 17.2 W/kg
SAR(1 g) = 9.83 W/kg; SAR(10 g) = 5.26 W/kg
Maximum value of SAR (measured) = 14.8 W/kg




                            +12.00


                            —16.00


                            —20.00

                                      0 dB = 14.8 W/kg = 11.70 dBW/kg




Certificate No: D1900V2—5d101_Jul18                  Page 7 of 8




                                                                        Ne se vperg n rag en n mr n ngenene n en n n en e e c n   >


mm
TTL                                                                                            o2
y




Impedance Measurement Plot for Body TSL




          File View Channel Sweep Calibration   Trace Scale   Marker System Window   Help

                                                                                   1.900000 GHz               44.864 O
                                                                                       547.21 pH              6.5327 Q
                                                                                 y 1.900000 GHz             87.393 mU
                                                                                                               124.24°




                       Ch1Aug= 20
              Ch1: Start 1.70000 GHz   ——                                                                  Stop 2.10000 GHz

                                                                                                            —2].170 dB




              .00

              00
              00
              00
              on      Ch 1
              Thi: Start 1.                                                                                Stop 2.10000 GHz

                                $HOSs TECFror




Certificate No: D1900V2—5d101_Jul18                   Page 8 of 8




                                                                               Nes veperg n ragr en n n n mgnmene n mm en n en y e nc


©
ko en n en e en n n en n e e                     e e         s




           k       5                                                       y ty
      Calibration Laboratory of                                         lb3                                       g Schweizerischer Kalibrierdienst
      Schmid & Partner                                                 M                                          c Service suisse d‘étalonnage
         Engineering AG                                                pe                                               Servizio svizzero di taratura
      Zeughausstrasse 43, 8004 Zurich, Switzerland                       *g/I//fl\\\\\\\\“                        S     Swiss Calibration Service



       Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
       The Swiss Accreditation Service is one of the signatories to the EA
       Multilateral Agreement for the recognition of calibration certificates

       Client      CTTL (Auden)                                                                       Certificate No: D2450V2—853_Jul18

       CALIBRATION CERTIFICATE
       Object                                D2450V2 — SN:853


       Calibration procedure(s)             QA CAL—05.v10
                                            Calibration procedure for dipole validation kits above 700 MHz



       Calibration date:                    July 24, 2018


       This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
       The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


       All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


       Calibration Equipment used (M&TE critical for calibration)


       Primary Standards                      ID #                      Cal Date (Certificate No.)                           Scheduled Calibration
       Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02678)                      Apr—19
       Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                            Apr—19
       Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02673)                            Apr—19
       Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                            Apr—19
       Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                            Apr—19
       Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                       Dec—18
       DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                       Oct—18


       Secondary Standards                    ID #                      Check Date (in house)                                Scheduled Check
       Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                    In house check: Oct—18
       Power sensor HP 8481A                  SN: US37292783            07—Oct—15 (in house check Oct—16)                    In house check: Oct—18
       Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                    In house check: Oct—18
       RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                    In house check: Oct—18
       Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—17)                    In house check: Oct—18


                                              Name                                     Function                               Signatlire
       Calibrated by:                        Claudio Leubler                           Laboratory Technician                               ’

                                                                                                                                      C#
       Approved by:                          Katia Pokovic                             Technical Manager                 %


                                                                                                                             Issued: July 24, 2018
       This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.


      Certificate No: D2450V2—853_Jul18                                     Page 1 of 8




                                                                                                               wenvperg n rage en


    £      s                                                   al19
Calibration Laboratory of                                    TT                     g Schweizerischer Kalibrierdienst
Schmid & Partner                                            m                       c    Service suisse d‘étalonnage
  Engineering AG                                            5 //:\\ 7                    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 +4,//-\\\\\x‘          S    swiss Calibration Service
                                                                thilals®


Accredited by the Swiss Accreditation Service (SAS)                                  Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
     Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
         Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
         Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
         SAR measured: SAR measured at the stated antenna input power.
         SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
         SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D2450V2—853_Jul18                               Page 2 of 8




                                                                              Nes vperg n ragen e mr n nggenene n mssn n n y e 9c


)
1
S
uo




Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY5                              V52.10.1
     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom
     Distance Dipole Center— TSL                               10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         2450 MHz # 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C            39.2              1.80 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C    37.8 16 %         1.85 mho/m + 6 %
     Head TSL temperature change during test                      «08 °C              ~———                   ~—


SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     13.2 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         51.7 Wikg + 17.0 % (k=2)

     SAR averaged over 10 cm‘ (10 g) of Head TSL                condition
     SAR measured                                        250 mW input power                  613 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         24.2 Wikg # 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
      Nominal Body TSL parameters                                 22.0 °C            52.7              1.95 mho/m
      Measured Body TSL parameters                             (22.0 + 0.2) °C    51.9 £ 6 %        2.02 mho/m £ 6 %
      Body TSL temperature change during test                     <0.5 °C


SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                 Condition
      SAR measured                                       250 mW input power                     13.1 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         51.3 W/kg + 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                     6.10 W/kg

      SAR for nominal Body TSL parameters                    normalized to 1W         24.1 Wikg # 16.5 % (k=2)



Certificate No: D2450V2—853_Jul18                     Page 3 of 8




                                                                                 Nn epeg n n en n n n en c ++ >


(llélu)
          Appendix (Additional assessments outside the scope of SCS 0108)

          Antenna Parameters with Head TSL

                Impedance, transformed to feed point                                           54.1 Q + 8.9 jQ
                Return Loss                                                                       — 25.2 dB



          Antenna Parameters with Body TSL

                Impedance, transformed to feed point                                           48.5 Q + 6.0 jQ
                Return Loss                                                                      —24.1 dB




          General Antenna Parameters and Design

               l Electrical Delay (one direction)                                                 1.161 ns


          After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

          The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
          second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
          are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
          "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
          according to the Standard.
          No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
          feedpoint may be damaged.


          Additional EUT Data

                Manufactured by                                                                    SPEAG
                Manufactured on                                                              November 10, 2009




          Certificate No: D2450V2—853_Jul18                        Page 4 of 8




                                                                                               ~mvmun n n n e n m n ngn n~


DASY5 Validation Report for Head TSL

                                                                                                Date: 24.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.85 $/m; & = 37.8; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN7349; ConvF(7.88, 7.88, 7.88) @ 2450 MHz; Calibrated: 30.12.2017

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 26.10.2017

        Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

        DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 115.3 V/m; Power Drift = —0.00 dB
Peak SAR (extrapolated) = 26.1 W/kg
SAR(1 g) = 13.2 W/kg; SAR(10 g) = 6.13 W/kg
Maximum value of SAR (measured) = 21.6 W/kg




                             1320


                             —17.60


                             —22.00

                                      0 dB = 21.6 W/kg = 13.34 dBW/kg




Certificate No: D2450V2—853_Jul18                    Page 5 of 8




                                                                        Ns veperg n ragn en n en n en en n e n n n   y >


Impedance Measurement Plot for Head TSL




     Eile ViewChannel Sweep Calibration   Scale Marker System Window _

                                                                           2450000 GHz   54.148 O
                                                                              254.98 pH  3.9251 0
                                                                           2450000 GHz §4.794 mU
                                                                                          41.259 °




                   Ch1Avg= 20
          Ch1: Start 2.25000 GHz                                                            Stop 2.65000 GHz

        0.00                                                                            2    —2§5.   225 dB
       5.00
       0.00




          00                2
          Ch1: Start 2.25000 GHz   ——                                                       Stop 2.65000 GHz

       Status     CH 1:                               _ Avg=20   Delay   ________            Thcee"




Certificate No: D2450V2—853_Jul18            Page 6 of 8




                                                                           Nn ey n n en n n n ng n > n >


(ullu)
         DASY5 Validation Report for Body TSL

                                                                                                         Date: 16.07.2018
         Test Laboratory: SPEAG, Zurich, Switzerland

         DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:853

         Communication System: UID 0 — CW; Frequency: 2450 MHz
         Medium parameters used: f = 2450 MHz; 0 = 2.02 S/m; &, = 51.9; p = 1000 kg/m*
         Phantom section: Flat Section
         Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

         DASY52 Configuration:

             *   Probe: EX3DV4 — SN7349; ConvF(8.01, 8.01, 8.01) @ 2450 MHz; Calibrated: 30.12.2017

             e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

             e   Electronics: DAE4 Sn601; Calibrated: 26.10.2017

             *   Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

             +   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


         Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
         Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 108.0 V/m; Power Drift = —0.03 dB
         Peak SAR (extrapolated) = 25.6 W/kg
         SAR(I g) = 13.1 W/kg; SAR(10 g) = 6.1 W/kg
         Maximum value of SAR (measured) = 21.0 W/kg


                 dB
                 0


                 ~4.00


                 —8.00


                 —12.00


                 ~16.00


                 —20.00
                            0 dB = 21.0 W/kg = 13.22 dBW/kg




         Certificate No: D2450V2—853_Jul18              Page 7 of 8




                                                                              Nesn vperg n agi en n en n en ies n n e n n n n   y >


®
Impedance Measurement Plot for Body TSL




     File View Channel Sweep Calibration Trace Scale Marker System Window Help.
                                                                          2450000 GHz          48.464 O
                                                                             389.27 pH         5.9924 O
                                                                          2450000 GHz        62.710 mU
                                                                                                100.89 °




                  Ch1Avg= 20
         Chi: Start 2.25000 GHz   ——                                                        Stop 2.65000 GHz

         .00
        00
                                                                > 1       2450000     GHz    —24. 053 dB
        .00
        .00




         00

         00
         00
         00
         00
         Chi: Start 2                                                                       Stop 2.65000 GHz


       Status    _ CH 1:




Certificate No: D2450V2—853_Jul18                 Page 8 of 8




                                                                          ~nnm unn en n n c ngn en


§7T7Tr                                                                                                        keor n reeee e ce ecerkaes.




     M       3                                                         w9 0j
 Calibration Laboratory of                                         s‘\\\\\t’/l/fl”;                          g Schweizerischer Kalibrierdienst
 Schmid & Partner                                                 M                                         fe      Service suisse d‘étalonnage
   Engineering AG                                                 Lz re                                             Servizio svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                     *,/fi‘\\\“                                 S       swiss Calibration Service


 Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreementfor the recognition of calibration certificates

 Client       CTTL (Auden)                                                                       Certificate No: D2600V2—1012_Jul18

 CALIBRATION CERTIFICATE
  Object                                D2600V2 — SN:1012


  Calibration procedure(s)              QA CAL—05.v10
                                        Calibration procedure for dipole validation kits above 700 MHz



  Calibration date:                     July 26, 2018


  This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
  The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


  All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


  Calibration Equipment used (M&TE critical for calibration)


  Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
  Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                       Apr—19
  Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                             Apr—19
  Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—026783)                            Apr—19
  Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                             Apr—19
  Type—N mismatch combination            SN: 5047.2 / 06827        04—Apr—18 (No. 217—02683)                             Apr—19
  Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                        Dec—18
  DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                        Oct—18


  Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
  Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
  Power sensor HP 8481A                  SN: US37292783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
  Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
  RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
  Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—17)                     In house check: Oct—18


                                         Name                                   Function                                  Signature
  Calibrated by:                         Michael Weber                          Laboratory Technician                    %fi—



  Approved by:                           Katja Pokovic                          Technical Manager                       ///%r/—


                                                                                                                        Issued: July 26, 2018
  This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


 Certificate No: D2600V2—1012_Jul18                                    Page 1 of 8




                                                                                                       Coopyniyin. i ngine ieourveu uy ui iL.


    s      e                                                   w1 o
Calibration Laboratory of                                    i\‘\\\\_zz}           S   Schweizerischer Kalibrierdienst
Schmid & Partner                                            in ooe                 c   Service suisse d‘étalonnage
  Engineering AG                                            ;///—\\\?                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 */1//\\\\“\0          S   swiss Calibration Service
                                                                Wlitubs®

Accredited by the Swiss Accreditation Service (SAS)                                 Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvrF                       sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
         MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle! to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2600V2—1012_Jul18                              Page 2 of 8




                                                                              Coupynigin. um nigine reourveu uy ui iL.


8
                                                                                                  y       n 6 y     o




Measurement Conditions
   DASY system configuration, as far as not iven on page 1.
     DASY Version                                              DASY5                                V52.10.1
     Extrapolation                                   Advanced Extrapolation
      Phantom                                        Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                               with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
      Frequency                                        2600 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature         Permittivity         Conductivity
      Nominal Head TSL parameters                                 22.0 °C             39.0               1.96 mho/m
      Measured Head TSL parameters                             (22.0 + 0.2) °C     37.2 26 %          2.02 mho/m + 6 %
      Head TSL temperature change during test                     <85°C                ~———                    ——


SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                 Condition
      SAR measured                                       250 mW input power                       14.2 Wikg
      SAR for nominal Head TSL parameters                    normalized to 1W          55.4 Wikg + 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                condition
      SAR measured                                       250 mW input power                   6.33 Wikg
      SAR for nominal Head TSL parameters                    normalized to 1W          24.9 Wikg x 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.                                                              .
                                                               Temperature         Permittivity         Conductivity
      Nominal Body TSL parameters                                 22.0 °C             52.5               2.16 mho/m
      Measured Body TSL parameters                             (22.0 £0.2) °C      51.5 26 %          2.20 mho/m x 6 %
      Body TSL temperature change during test                     <0.5 °C              «~=


SAR result with Body TSL

      SAR averaged over 1 em* (1 g) of Body TSL                 Condition
      SAR measured                                       250 mW input power                       13.7 Wikg
      SAR for nominal Body TSL parameters                    normalized to 1W           54.1 Wikg £17.0 % (k=2)

      SAR averaged over 10 em* (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                    6.17 Wikg
      SAR for nominal Body TSL parameters                    normalized to 1W           24.5 Wikg # 16.5 % (k=2)



Certificate No: D2600V2—1012_Jul18                    Page 3 of 8




                                                                                 Coupyngin. rumngine 1Gouryou uy viaal.


                                                                                                     tw s      n n   n   >    n ——




Appendix (Additional assessments outside the scope of SCS 01 08)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           474 O — 7 4 |Q
      Return Loss                                                                       —21.9 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           44.1 Q — 4.9 jQ
      Return Loss                                                                      —21.8 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.154 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                               October 30, 2007




Certificate No: D2600V2—1012_Jul18                        Page 4 of 8




                                                                                    wervopg n ree a io rgn rewr a vewruer e tona wmz d e ae.


mm©                                                                        2e e e e e e e e e e e e .
‘/"/‘],
T                                                                                    t   e—       t n o    9     n ——




  DASY5 Validation Report for Head TSL

                                                                                                Date: 26.07.2018

  Test Laboratory: SPEAG, Zurich, Switzerland

  DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1012

  Communication System: UID 0 — CW; Frequency: 2600 MHz
  Medium parameters used: f = 2600 MHz; 0 = 2.02 S/m; & = 37.2; p = 1000 kg/m*
  Phantom section: Flat Section
  Measurement Standard: DASY5 (IEEE/IEC/ANSI €63.19—2011)

  DASY52 Configuration:

     e   Probe: EX3DV4 — SN7349; ConvF(7.7, 7.7, 7.7) @ 2600 MHz; Calibrated: 30.12.2017

     e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

     e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

     e    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

     &    DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


  Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
  Measurement grid: dx=5mm, dy=5mm, dz=5mm
  Reference Value = 118.3 V/m; Power Drift = —0.04 dB
  Peak SAR (extrapolated) = 28.3 W/kg
  SAR(I g) = 14.2 W/kg; SAR(10 g) = 6.33 W/kg
  Maximum value of SAR (measured) = 23.7 W/kg


         dB
         0

          —5.20

          —10.40

          —15.60

          —20.80

          —26.00
                     0 dB = 23.7 W/kg = 13.75 dBW/kg




  Certificate No: D2600V2—1012_Jul18              Page 5 of 8




                                                                       wervopg n ree a io rgn rewr a vewruer e tona wmz d e ae.


mm._.>
‘/"/"‘],
e




  Impedance Measurement Plot for Head TSL




   File    View    Channel Sweep   Calibration Trace      Scale   Marker   System   Window    Help

                                                                                           2600000 GHz          47.371 Q
                                                                                      5        8.2358 pF       —7.4326 Q
                                                                                          2600000 GHz         80.734 mU
                                                                                                                ~105.12 °




                  Ch1Avg= 20
          Ch1: Start 2.40000 GHz                                                                              Stop 2.90000 GHz


                                                                                             2,¢00000     2   —21. 859 dB




          00      Ch1        =
          Ch1: Start 2.40000 GHz                                                                              Stop 2.80000 GHz

     Status       ___CH 1:   &n               C* 1—Port                    Avg=20 Delay                            LCL




  Certificate No: D2600V2—1012_Jul18                                Page 6 of 8




                                                                                                 earopyirgire. muringrite ruousvuu y virdu.


DASY5 Validation Report for Body TSL

                                                                                      Date: 26.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1012

Communication System: UID 0 — CW; Frequency: 2600 MHz
Medium parameters used: f = 2600 MHz; 0 = 2.2 S/m; & = 51.5; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e    Probe: EX3DV4 — SN7349; ConvF(7.81, 7.81, 7.81) @ 2600 MHz; Calibrated: 30.12.2017

   e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   e    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   «_   DASY532 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 107.5 V/m; Power Drift = —0.07 dB
Peak SAR (extrapolated) = 27.7 W/kg
SAR(I g) = 13.7 W/kg; SAR(10 g) = 6.17 W/kg
Maximum value of SAR (measured) = 22.6 W/kg


        dB
         0

        —5.20

        —10.40

        ~15.60

        —20.80

        —26.00
                   0 dB = 22.6 W/kg = 13.54 dBW/kg




Certificate No: D2600V2—1012_Jul18             Page 7 of 8




                                                                    oropyrigin. munnigime ruourvou y virin.



Document Created: 2019-06-26 15:02:50
Document Modified: 2019-06-26 15:02:50

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC