RF Exposure - SAR ReportI19Z60741-SEM01_SAR_Rev0_part3

FCC ID: ZNFX120EMW

RF Exposure Info

Download: PDF
FCCID_4337729

uy




CallbI"atIOH Laboratory of                                   ¢\“\\L\I//I'7’7¢   g   Schweizerischer Kalibrierdienst
Schmid & Partner                                            ;ia\%\fl_%é          c   Service suisse d‘étalonnage
  Engineering AG                                            uzol                    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 7'/,,//fi\“\?       S   swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                             Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvrF                       sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
     Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle! to the body axis.
     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
     SAR measured: SAR measured at the stated antenna input power.
     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D1900V2—5d101_Jul18                             Page 2 of 8


Measurement Conditions
    DASY system configuration, as far as not given on page 1.
     DASY Version                                             DASYS5                              V52.10.1
      Extrapolation                                  Advanced Extrapolation
      Phantom                                         Modular Flat Phantom
      Distance Dipole Center — TSL                             10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
      Frequency                                        1900 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
      Nominal Head TSL parameters                                 22.0 °C           40.0               1.40 mho/m
      Measured Head TSL parameters                             (22.0 + 0.2) °C   39.9 £ 6 %         1.34 mho/m +6 %
      Head TSL temperature change during test                     <0.5 °C                                    _——=


SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                 Condition
      SAR measured                                       250 mW input power                     9.84 Wikg
      SAR for nominal Head TSL parameters                    normalized to 1W        40.4 Wikg x 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                condition
      SAR measured                                       250 mW input power                 5.23 Wikg
      SAR for nominal Head TSL parameters                    normalized to 1W        21.3 Wikg # 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
      Nominal Body TSL parameters                                 22.0 °C           53.3               1.52 mho/m
      Measured Body TSL parameters                             (22.0 £ 0.2) °C   54.3 £6 %          1.46 mho/m + 6 %
      Body TSL temperature change during test                     <0.5 °C            ~———                    w


SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                 Condition
      SAR measured                                       250 mW input power                     9.83 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W        40.4 Wikg x 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                 5.26 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W        21.4 Wikg # 16.5 % (k=2)



Certificate No: D1900V2—5d101_Jul18                   Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           50.7 Q + 5.3 jQ
      Return Loss                                                                       —25.6 dB


Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           44.9 Q + 6.5 jQ
      Return Loss                                                                      —21.2 dB




General Antenna Parameters and Design

     I Electrical Delay (one direction)                                                 1.205 ns                        J

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                March 28, 2008




Certificate No: D1900V2—5d101_Jul18                      Page 4 of 8


a
N
(I
     DASY5 Validation Report for Head TSL

                                                                                         Date: 24.07.2018

     Test Laboratory: SPEAG, Zurich, Switzerland

     DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

     Communication System: UID 0 — CW; Frequency: 1900 MHz
     Medium parameters used: f = 1900 MHz; 0 = 1.34 S/m; & = 39.9; p = 1000 kg/m*
     Phantom section: Flat Section
     Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

     DASY52 Configuration:

         *   Probe: EX3DV4 — SN7349; ConvF(8.18, 8.18, 8.18) @ 1900 MHz; Calibrated: 30.12.2017

         *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

         *   Electronics: DAE4 Sn601; Calibrated: 26.10.2017

         e   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

         *   DASY32 52.10.1(1476); SEMCAD X 14.6.11(7439)


     Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
     Measurement grid: dx=5mm, dy=5mm, dz=5mm
     Reference Value = 110.6 V/m; Power Drift =—0.01 dB
     Peak SAR (extrapolated) = 17.9 W/kg
     SAR(1 g) = 9.84 W/kg; SAR(10 g) = 5.23 W/kg
     Maximum value of SAR (measured) = 15.0 W/kg




                                  12.00


                                  +16.00


                                  —20.00

                                           0 dB = 15.0 W/kg = 11.76 dBW/kg




     Certificate No: D1900V2—5d101_Jul18                  Page 5 of 8


®
Impedance Measurement Plot for Head TSL




          File    View    Channel    Sweep Calibration Trace Scale Marker System Window   Help

                                                                                        1.900000 GHz      50.660 O
                                                                                            441.05 pH     5.2653 O
                                                                                      , 1.900000 GHz    52.645 mU
                                                                                                           79.866 °




                          Ch1Aug= 20
                 Ch1: Start 1.70000 GHz   ——                                                            Stop 2.10000 GHz


                                                                                            00000 GHz   —2$.573 dB




                 00       Ch 1 Avg= |20
                 Ch1: Start 1.70000 GHa   ———                                                           Stop 2.10000 GHz




Certificate No: D1900V2—5d101_Jul18                        Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                    Date: 24.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

Communication System: UID 0 — CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; 0 = 1.46 S/m; & = 54.3; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e    Probe: EX3DV4 — SN7349; ConvF(8.15, 8.15, 8.15) @ 1900 MHz; Calibrated: 30.12.2017

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   &    DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 105.3 V/m; Power Drift = —0.04 dB
Peak SAR (extrapolated) = 17.2 W/kg
SAR(1 g) = 9.83 W/kg; SAR(10 g) = 5.26 W/kg
Maximum value of SAR (measured) = 14.8 W/kg




                            12.00


                            —16.00


                            —20.00
                                      0 dB = 14.8 W/kg = 11.70 dBW/kg




Certificate No: D1900V2—5d101_Jul18                  Page 7 of 8


©
Impedance Measurement Plot for Body TSL




          File View Channel       Sweep   Calibration   Trace Scale Marker System Window   Help

                                                                                           1.900000 GHz      44.864 0
                                                                                               547.21 pH     6.5327 O
                                                                                      y 1.900000 GHz       87.393 mU
                                                                                                              124.24 °




                       ChiAvg= 20
              Chi: Start 1.70000 GHz                                                                       Stop 2.10000 GHz


                                                                                                       2   __—2].170 dB




              00

              00
              00
              o0
              00          1     &
              Ch1: Start 1.70000 GHz                                                                       Stop 2.10000 GHz

                                                        C" 1—Port




Certificate No: D1900V2—5d101_Jul18                             Page 8 of 8


®
ne se en n on nge en        e ne n e+             100    >     22 %




       Calibration
             o     Laboratory of                                          :‘\‘wl‘m
                                                                               //z                                 5      Schweizerischer Kalibrierdienst
       Schmid & Partner                                                  iia\Eu/{fi                                 c Service suisse d‘étalonnage
         Engineering AG                                                  ;///—\\ >                                        Servizio svizzero di taratura
       Zeughausstrasse 43, 8004 Zurich, Switzerland                       *’7(\\\“‘\3                              S      Swiss Calibration Service
                                                                             Whiitals

        Accredited by the Swiss Accreditation Service (SAS)                                                        Accreditation No.: SCS 0108
        The Swiss Accreditation Service is one of the signatories to the EA
        Multilateral Agreement for the recognition of calibration certificates

        Client      CTTL (Auden)                                                                       Certificate No: D2450V2—853_Jul18

        CALIBRATION CERTIFICATE
        Object                                D2450V2 — SN:853



        Calibration procedure(s)              QA CAL—05.v10
                                              Calibration procedure for dipole validation kits above 700 MHz



        Calibration date:                     July 24, 2018


        This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
        The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


        All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


        Calibration Equipment used (M&TE critical for calibration)


        Primary Standards                      ID #                      Cal Date (Certificate No.)                           Scheduled Calibration
        Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                      Apr—19
        Power sensor NRP—Z91                   SN: 108244                04—Apr—18 (No. 217—02672)                            Apr—19
        Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02673)                            Apr—19
        Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                            Apr—19
        Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                            Apr—19
        Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                        Dec—18
        DAE4                                   SN: 601                   26—O0ct—17 (No. DAE4—601_Oct17)                      Oct—18


        Secondary Standards                    ID #                      Check Date (in house)                                Scheduled Check
        Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
        Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
        Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
        RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
        Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—17)                     In house check: Oct—18


                                               Name                                     Function
        Calibrated by:                         Claudio Leubler                          Laboratory Technician



        Approved by:                           KatJa Pokovic                            Technical Manager



                                                                                                                              Issued: July 24, 2018
        This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


       Certificate No: D2450V2—853_Jul18                                     Page 1 of 8


Calibration Laboratory of                                                     S   Schweizerischer Kalibrierdienst
Schmid & Partner                                                              C   Service suisse d‘étalonnage
  Engineering AG                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                  S swiss Calibration Service

Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle! to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D2450V2—853_Jul18                               Page 2 of 8


(flé"l’
         Measurement Conditions
             DASY system configuration, as far as not given on page 1.
               DASY Version                                              DASY5                              V52.10.1
               Extrapolation                                   Advanced Extrapolation
               Phantom                                         Modular Flat Phantom
               Distance Dipole Center— TSL                               10 mm                             with Spacer
               Zoom Scan Resolution                              dx, dy, dz =5 mm
               Frequency                                         2450 MHz + 1 MHz



         Head TSL parameters
            The following parameters and calculations were applied.
                                                                         Temperature       Permittivity         Conductivity
               Nominal Head TSL parameters                                  22.0 °C           39.2               1.80 mho/m
               Measured Head TSL parameters                              (22.0 + 0.2) °C   37.8 26 %          1.85 mho/m +6 %
               Head TSL temperature change during test                      <0.5 °C            ~——                     _

         SAR result with Head TSL

              SAR averaged over 1 cm* (1 g) of Head TSL                   Condition
               SAR measured                                       250 mW input power                      13.2 W/kg
               SAR for nominal Head TSL parameters                    normalized to 1W         51.7 Wikg + 17.0 % (k=2)


               SAR averaged over 10 cm* (10 g) of Head TSL                condition
               SAR measured                                       250 mW input power                  618 W/kg
               SAR for nominal Head TSL parameters                    normalized to 1W         24.2 Wikg £ 16.5 % (k=2)


         Body TSL parameters
            The following parameters and calculations were applied.
                                                                         Temperature       Permittivity         Conductivity
               Nominal Body TSL parameters                                  22.0 °C           §$2.7              1.95 mho/m
               Measured Body TSL parameters                              (22.0 + 0.2) °C   51.9 +6 %          2.02 mho/m x 6 %
               Body TSL temperature change during test                      <0.5 °C            ~———                    ~———


         SAR result with Body TSL

               SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
               SAR measured                                       250 mW input power                      18.1 W/kg
               SAR for nominal Body TSL parameters                    normalized to 1W         51.3 W/kg + 17.0 % (k=2)

               SAR averaged over 10 cm* (10 g) of Body TSL                 condition
               SAR measured                                       250 mW input power                      6.10 W/kg
               SAR for nominal Body TSL parameters                    normalized to 1W         24.1 Wikg # 16.5 % (k=2)



         Certificate No: D2450V2—853_Jul18                      Page 3 of 8


(EM,
       Appendix (Additional assessments outside the scope of SCS 0108)

       Antenna Parameters with Head TSL

             Impedance, transformed to feed point                                           54.1 Q + 3.9 jQ
             Return Loss                                                                       — 25.2 dB



       Antenna Parameters with Body TSL

             Impedance, transformed to feed point                                           48.5 Q + 6.0 jQ
             Return Loss                                                                      ~24.1 dB



       General Antenna Parameters and Design

             Electrical Delay (one direction)                                                   1.161 ns


       After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

       The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
       second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
       are added to the dipole arms in orderto improve matching when loaded according to the position as explained in the
       "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
       according to the Standard.
       No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
       feedpoint may be damaged.


       Additional EUT Data

             Manufactured by                                                                    SPEAG
             Manufactured on                                                              November 10, 2009




       Certificate No: D2450V2—853_Jul18                        Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                    Date: 24.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.85 $/m; s = 37.8; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    e   Probe: EX3DV4 — SN7349; ConvF(7.88, 7.88, 7.88) @ 2450 MHz; Calibrated: 30.12.2017

    e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    *   Electronics: DAE4 Sn601; Calibrated: 26.10.2017

    *   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

    *   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 115.3 V/m; Power Drift = —0.00 dB
Peak SAR (extrapolated) = 26.1 W/kg
SAR(I g) = 13.2 W/kg; SAR(10 g) = 6.13 W/kg
Maximum value of SAR (measured) = 21.6 W/kg




                             1320


                             —17.60


                             —22.00

                                      0 dB = 21.6 W/kg = 13.34 dBW/kg




Certificate No: D2450V2—853_Jul18                    Page 5 of 8


Impedance Measurement Plot for Head TSL




     Eile ViewChannel Sweep Calibration Trace Scale Marker System Window _

                                                                         2450000 GHz      54.148 Q
                                                                            254.98 pH     3.9251 0
                                                                         2450000 GHz    54.794 mU
                                                                                           41.259 °




                   Ch1Aug= 20
          Ch1: Start 2.25000 GHz   ——                                                   Stop 2.65000 GHz

        0.00
                                                                                        —2$.225 dB
       5.00
       0.00
        .00




          00

          00
          00
          00
          00       Ch 1     =
          Ch1: Start 2.25000 GHz   ——                                                   Stop 2.65000 GHz


       Status _   CH 1:                                                                      teos?




Certificate No: D2450V2—853_Jul18


n
S
(I
     DASY5 Validation Report for Body TSL

                                                                                          Date: 16.07.2018
     Test Laboratory: SPEAG, Zurich, Switzerland

     DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:853

     Communication System: UID 0 — CW; Frequency: 2450 MHz
     Medium parameters used: f = 2450 MHz; 0 = 2.02 S/m; &, = 51.9; p = 1000 kg/m‘
     Phantom section: Flat Section
     Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

     DASY52 Configuration:

         *    Probe: EX3DV4 — SN7349; ConvF(8.01, 8.01, 8.01) @ 2450 MHz; Calibrated: 30.12.2017

         e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

         e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

         *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

         «_   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


     Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
     Measurement grid: dx=5mm, dy=5mm, dz=5mm
     Reference Value = 108.0 V/m; Power Drift = —0.03 dB
     Peak SAR (extrapolated) = 25.6 W/kg
     SAR(I g) = 13.1 W/kg; SAR(10 g) = 6.1 W/kg
     Maximum value of SAR (measured) = 21.0 W/kg


              dB
              0


              ~4.00


              —8.00


              —12.00


              ~16.00


              —20.00
                        0 dB = 21.0 W/kg = 13.22 dBW/kg




     Certificate No: D2450V2—853_Jul18               Page 7 of 8


Impedance Measurement Plot for Body TSL




     Eile View Channel Sweep Calibration Trace Scale Marker System Window Help .
                                                                          2450000 GHz       48.464 O
                                                                             389.27 pH      5.9924 Q
                                                                          2450000 GHz     62.710 mU
                                                                                             100.89 °




                   Ch1Avg= 20
          Ch1: Start 2.25000 GHz   ——                                                     Stop 2.65000 GHz

         .00
        00
                                                                          2450000   GHz   —24.053 dB
        .00
        .00




         00

         00
         00
         00
         00       Ch 1     =
          Chi: Start 2.25000 GHz   ——

       Status    _ CH 1:




Certificate No: D2450V2—853_Jul18                 Page 8 of 8


                                                                                                                           ttgit      n         n     n   n +




    .       5                                                         w1 0j
Calibration Laboratory of                                         Q\\“\\\/\\\_/’//"/,;                     g        Schweizerischer Kalibrierdienst
Schmid & Partner                                                 i‘afi&%                                    c       Service suisse d‘étalonnage
  Engineering AG                                                  ud                                               Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      *g/I//\\\\\\\“                           S       swiss Calibration Service
                                                                      "dhilals


Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client       CTTL (Auden)                                                                        Certificate No: D2600V2—1012_Jul18

CALIBRATION CERTIFICATE
 Object                                D2600V2 — SN:1012


 Calibration procedure(s)              QA CAL—05.v10
                                       Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                     July 26, 2018


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                       Apr—19
 Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                             Apr—19
 Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02673)                             Apr—19
 Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                             Apr—19
 Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                             Apr—19
 Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                        Dec—18
 DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                        Oct—18


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
 Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—17)                     In house check: Oct—18


                                        Name                                       Function                              Signature
 Calibrated by:                         Michael Weber                              Laboratory Technician                         H-



 Approved by:                           Katia Pokovic                              Technical Manager                   ///%’C




                                                                                                                        Issued: July 26, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D2600V2—1012_Jul18                                     Page 1 of 8


(IEI)
            P      s                                                    w4 uj
        Callbl.'atlon Laboratory of                                  Q‘QM?Z           S   Schweizerischer Kalibrierdienst
        Schmid & Partner                                            m                 c   Service suisse d‘étalonnage
          Engineering AG                                            s o amls              Servizio svizzero di taratura
                                     ;      M
        Zeughausstrasse 43, 8004 Zurich, Switzerland
                                                                    x4rosng
                                                                       fan &
                                                                                      S      i      iFSH          M
                                                                                          Swiss Calibration Service
                                                                      Hlilals

        Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
        The Swiss Accreditation Service is one of the signatories to the EA
        Multilateral Agreement for the recognition of calibration certificates

        Glossary:
        TSL                          tissue simulating liquid
        ConvF                        sensitivity in TSL / NORM x,y,z
        N/A                          not applicable or not measured

        Calibration is Performed According to the Following Standards:
           a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
              Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
              Communications Devices: Measurement Techniques", June 2013
           b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
              (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
              300 MHz to 6 GHz)", July 2016
           c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
              communication devices used in close proximity to the human body (frequency range of 30
              MHz to 6 GHz)", March 2010
           d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

        Additional Documentation:
           e) DASY4/5 System Handbook

        Methods Applied and Interpretation of Parameters:
           e Measurement Conditions: Further details are available from the Validation Report at the end
             of the certificate. All figures stated in the certificate are valid at the frequency indicated.
           e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
             point exactly below the center marking of the flat phantom section, with the arms oriented
             paralle! to the body axis.
           e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
             positioned under the liquid filled phantom. The impedance stated is transformed from the
             measurement at the SMA connector to the feed point. The Return Loss ensures low
             reflected power. No uncertainty required.
           e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
             No uncertainty required.
           e SAR measured: SAR measured at the stated antenna input power.
           e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
                 connector.
            e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
                 nominal SAR result.
          The reported uncertainty of measurement is stated as the standard uncertainty of measurement
          multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
          probability of approximately 95%.




        Certificate No: D2600V2—1012_Jul18                              Page 2 of 8



Document Created: 2019-06-27 14:15:32
Document Modified: 2019-06-27 14:15:32

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC