SAR Appendix B

FCC ID: ZNFW200V

RF Exposure Info

Download: PDF
FCCID_2770796

Test Laboratory: UL Verification Services Inc. SAR Lab A                    Date/Time: 9/18/2015 10:31:28 AM

SystemPerformanceCheck-D2450V2 SN 748

Frequency: 2450 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 2450 MHz; σ = 2.03 S/m; ε r = 51.82; ρ = 1000 kg/m
                                                                               3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1357; Calibrated: 2/20/2015
- Probe: EX3DV4 - SN3901; ConvF(7.26, 7.26, 7.26); Calibrated: 1/27/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: SAM v5.0 (B); Type: QD000P40CD; Serial: 1628

Body/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 6.29 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 57.98 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 10.7 W/kg
SAR(1 g) = 5.08 W/kg; SAR(10 g) = 2.32 W/kg
Maximum value of SAR (measured) = 7.27 W/kg




                      0 dB = 7.27 W/kg = 8.62 dBW/kg




                                                                                                               Plot No. 1


Test Laboratory: UL Verification Services Inc. SAR Lab A    Date/Time: 9/18/2015 10:50:29 AM

SystemPerformanceCheck-D2450V2 SN 748
Frequency: 2450 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 6.57 W/kg




                                                                                           Plot No. 2


Test Laboratory: UL Verification Services Inc. SAR Lab B                    Date/Time: 9/21/2015 10:36:08 AM

SystemPerformanceCheck-D750V3 SN 1024

Frequency: 750 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 750 MHz; σ = 0.898 S/m; ε r = 41.057; ρ = 1000 kg/m
                                                                                3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1360; Calibrated: 3/12/2015
- Probe: EX3DV4 - SN3751; ConvF(9.31, 9.31, 9.31); Calibrated: 11/14/2014;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 (A); Type: QDOVA001BB; Serial: 1117

Head /Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 0.950 W/kg

Head /Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 33.109 V/m; Power Drift = -0.04 dB
Peak SAR (extrapolated) = 1.17 W/kg
SAR(1 g) = 0.780 W/kg; SAR(10 g) = 0.513 W/kg




                      0 dB = 0.950 W/kg = -0.22 dBW/kg




                                                                                                               Plot No. 3


Test Laboratory: UL Verification Services Inc. SAR Lab B    Date/Time: 9/21/2015 10:55:51 AM

SystemPerformanceCheck-D750V3 SN 1024
Frequency: 750 MHz; Duty Cycle: 1:1

Head/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 0.954 W/kg




                                                                                           Plot No. 4



Document Created: 2015-09-28 19:56:43
Document Modified: 2015-09-28 19:56:43