RF exposure 6 of 6

FCC ID: ZNFW150

RF Exposure Info

Download: PDF
FCCID_2594634

Calibration Laboratory of                                           WY                                               Schweizerischer Kalibrierdienst




                                                              [
                                                                 ”/’/




                                                                                 &8
Schmid & Partner




                                                                                     W
                                                                                                                     Service suisse d‘étalonnage
  Engineering AG                                                                                                     Servizio svizzero di taratura




                                                               wl




                                                                                  /,/I//
Zeughausstrasse 43, 8004 Zurich, Switzerland                        4             8                                  Swiss Calibration Service




                                                                 8
                                                                     l/"llllll\“\\


Accredited by the Swiss Accreditation Service (SAS)                                                      Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Murltilateral Agreement for the recognition of calibration certificates

Client       UL CCS USA                                                                                  Certificate No: D2450V2—706_May14

|CALIBRATION CERTIFICATE
 Object                               D2450V2 — SN: 706


 Calibration procedure(s)             QA CAL—O05.vg
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                     May 20, 2014




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                        Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter EPM—442A                   GB37480704                  09—Oct—13 (No. 217—01827)                             Oct—14
 Power sensor HP 8481A                  US37292783                  09—Oct—13 (No. 217—01827)                             Oct—14
 Power sensor HP 8481A                  MY41092317                  09—Oct—13 (No. 217—01828)                             Oct—14
 Reference 20 dB Attenuator             SN: 5058 (20k)              03—Apr—14 (No. 217—01918)                             Apr—15
 Type—N mismatch combination            SN: 5047.2 / 06327          03—Apr—14 (No. 217—01921)                             Apr—15
 Reference Probe ES3DV3                 SN: 3205                    30—Dec—13 (No. ES3—3205_Dec13)                        Dec—14
 DAE4                                   SN: 601                     30—Apr—14 (No. DAE4—601_Apr14)                        Apr—15


 Secondary Standards                    ID #                        Check Date (in house)                                 Scheduled Check
 RF generator R&S SMT—O6                100005                          04—Aug—99 (in house check Oct—13)                 In house check: Oct—16
 Network Analyzer HP 8753E              US37390585 $4206                18—Oct—01 (in house check Oct—13)                 In house check: Oct—14



                                        Name                                               Function                         ign tl re
 Calibrated by:                         Claudio Leubler                                    Laboratory Technician



 Approved by:                           Katja Pokovic                                      Technical Manager          /é;i//g__‘



                                                                                                                          Issued: May 21, 2014
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



Certificate No: D2450V2—706_May14                                          Page 1 of 8


     x      «                                                     pupy
Calibration Laboratory of                                    S\\\\\@/"/,}         Schweizerischer Kalibrierdienst
Schmid & Partner                                            ila\&\—//mm%          Service suisse d‘étalonnage
  Engineering AG                                            os                    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 *A/7/\\\\\\\0        Swiss Calibration Service
                                                                Arud ubst



Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
Tok                          tissue simulating liquid
Convr                        sensitivity in TSL / NORM x.y,z2
N/A                          not applicable or not measured

Calibration is Performed Accordir:g to the Following Standards:
    a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
          Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
          Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
       devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
    c) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    d) DASY4/5 System Handbook

Methods Applied and Interpretation of Paramieters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
          of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    s     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
          point exactly below the center marking of the flat phantom section, with the arms oriented
          parallel to the body axis.
     e    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
     «*   Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
     *    SAR measured: SAR measured at the stated antenna input power.
     *    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
      e   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.




 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D2450V2—706_May14                                Page 2 of 8


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY5                              V52.8.8

     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                             with Spacer

     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         2450 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Head TSL parameters                                  22.0 °C           39.2                1.80 mho/m

     Measured Head TSL parameters                              (22.0 + 0.2) °C   38.5 + 6 %         1.83 mho/m +6 %
     Head TSL temperature change during test                      <0.5 °C                                    ————


SAR resuit with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     13.4 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         53.0 W/kg + 17.0 % (k=2)


     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        250 mW input power                     6.17 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         24.5 Wikg + 16.5 % (k=2)



Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Body TSL parameters                                  22.0 °C            52.7               1.95 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   50.8 + 6 %          2.03 mho/m +6 %
     Body TSL temperature change during test                      <0.5 °C            ————                    ————

SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
     SAR measured                                        250 mW input power                     12.9 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         50.2 W/ikg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition

     SAR measured                                        250 mW input power                     5.94 Wi/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         23.4 Wikg + 16.5 % (k=2)



Certificate No: D2450V2—706_May14                     Page 3 of 8


Appendix

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           52.5 Q + 4.8 jQ
      Return Loss                                                                       — 25.6 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           48.4 Q + 5.6 jQ
      Return Loss                                                                      — 24.5 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1.143 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipote length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                 May 28, 2002




Certificate No: D2450V2—706_May14                         Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                   Date: 20.05.2014

Test Laboratory: SPEAG, Zurich, Switzerland

DUT : Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 706

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: £f = 2450 MHz; 0 = 1.83 S/m; & = 38.5; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   e    Probe: ES3DV3 — SN3205; ConvF(4.53, 4.53, 4.53); Calibrated: 30.12.2013;

   e    Sensor—Surface: 3mm (Mechanical Surface Detection)

   e    Electronics: DAFE4 Sn601; Calibrated: 30.04.2014

   e    Phantom: Flat Phantom 5.0 (front); Type: QDOOOP50OAA; Serial: 1001

   «_   DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)


Dipole Calibration for Head Tissue/Pin=250 mW, d=18mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 101.6 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 27.8 W/kg
SAR(1 g) = 13.4 W/kg; SAR(10 g) = 6.17 W/kg
Maximum value of SAR (measured) = 17.6 W/kg


        dB



        —4.80

        —9.60

        —14.40

        —19.20

        —24.00
                  0 dB = 17.6 W/kg = 12.46 dBW/kg




Certificate No: D2450V2—706_May14               Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                                              12 May 2014             15:41:29
         CHI) Ssii     1 U FS           1152533 a              4.7617 a              309,.33 pH                      2 450.000 06A MHz
                                                          ArzaTAj

                                                      . $      e                              [»       .\\‘\

         De1l                                /&               y
                                                              4\            Xo     EP .   3 ~
                                                          &        "~            * l\                   \

         Cor                            {/            h                 1          yrey’f

                                        lf        4                     f           3}
                                         |        »                     (KX,                       s




                                                                                          \
         ts                                                        |
         H1d

          CH2




         Del


         Cor




         Aw
         16 °

         H1d

                     START 2 250.000 0090 MHz                                                                  STOP 2 650.000 000 MHz




Certificate No: D2450V2—706_May14                                   Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                  Date: 20.05.2014

Test Laboratory: SPEAG, Zurich, Switzerland

DUT : Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 706

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: £= 2450 MHz; 0 = 2.03 S/m; & = 50.8; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   e   Probe: ES3DV3 — SN3205; ConvF(4.35, 4.35, 4.35); Calibrated: 30.12.2013;

       Sensor—Surface: 3mm (Mechanical Surface Detection)

       Electronics: DAFE4 Sn601; Calibrated: 30.04.2014

       Phantom: Flat Phantom 5.0 (back); Type: QDOOOP50AA; Serial: 1002

       DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 9:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 94.57 V/m; Power Drift = 0.01 dB
Peak SAKR(extrapolated) = 27.0 W/kg
SAR(1 g) = 12.9 W/kg; SAR(10 g) = 5.94 W/kg
Maximum value of SAR (measured) = 17.0 W/kg




       —5.00

       —10.00


       —15.00

       —20.00


       —25.00
                  0 dB = 17.0 W/kg = 12.30 dBW/kg




Certificate No: D2450V2—706_May14              Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                     12 May 2014      15:40:14
         CHZ) s1i1     1 U FS          1: 48,.400 a   5.62141 a   365.15 pH         2 450.000 000 MHz
                                                  eT.
         Del

         Cor




         Avg
         16


         H1d

          CH2




         Del

         Cor




         Av
         18°

         H1d

                     START 2 259.000 050 MHz                                  STOP 2 650.0050 D00 MHz




Certificate No: D2450V2—706_May14                       Page 8 of 8



Document Created: 2015-02-06 22:26:49
Document Modified: 2015-02-06 22:26:49