SAR_Appendix F

FCC ID: ZNFVC200

RF Exposure Info

Download: PDF
FCCID_2710862

Calibration Laboratory of                                          Puss;
                                                                  CRCE                                             Schweizerischer Kalibrierdienst
      K                                                           SAn"                                       S
Schmid & Partner                                                 i\laE\//mf                                  c     Service suisse d‘étalonnage
  Engineering AG                                                 s                                                 Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      a ToX                                      S     Swiss Calibration Service
                                                                    Il'llnll\\\‘\\




Accredited by the Swiss Accreditation Service (SAS)                                                          Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       UL CCS USA                                                                             Certificate No: D835V2—4d117_May15

[CALIBRATION CERTIFICATE                                                                                                                             I
 Object                               D835V2 — SN: 4d117


 Calibration procedure(s)             QA CAL—O5.v9
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                    May 18, 2015




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S!).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter EPM—442A                   GB37480704                07—Oct—14     (No.   217—02020)                       Oct—15
 Power sensor HP 8481A                  US37292783                07—Oct—14     (No.   217—02020)                       Oct—15
 Power sensor HP 8481A                  MY41092317                07—Oct—14     (No.   217—02021)                       Oct—15
 Reference 20 dB Attenuator             SN: 5058 (20k)            01—Apr—15     (No.   217—02131)                       Mar—16
 Type—N mismatch combination            SN: 5047.2 / 06327        01—Apr—15     (No.   217—02134)                       Mar—16
 Reference Probe ES3DV3                 SN: 3205                  30—Dec—14 (No. ES3—3205_Dec14)                        Dec—15
 DAE4                                   SN: 601                   18—Aug—14 (No. DAE4—601_Aug14)                        Aug—15


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 RF generator R&S SMT—O6                100005                    04—Aug—99 (in house check Oct—13)                     In house check: Oct—16
 Network Analyzer HP 8753E              US37390585 S4206          18—Oct—01 (in house check Oct—14)                     In house check: Oct—15




                                        Name                                         Function                            Signature


                                                                                                                   p en
 Calibrated by:                         Israe Elnaouq                                Laboratory Technician



 Approved by:                           Katja Pokovic                                Technical Manager

                                                                                                             metrigerom Issued: May 18, 2015
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D835V2—4d117_May15                                    Page 1 of 11


    &                                                            ol y
Calibration
      4
            Laboratory of                                   $
                                                             Sisl/‘n,            S
                                                                                     Schweizerischer Kalibrierdienst
Schmid & Partner                                            iE‘\\\—'-//E
                                                                    7 IEE//!‘_   c   Service suisse d‘étalonnage
  Engineering AG                                            ASe—HAU T                Servizio svizzero di taratura
                              &       ipSge                 z3 —wnWy             S      .      chrar;       .
Zeughausstrasse 43, 8004 Zurich, Switzerland                 *4/fi\\\\\               Swiss Calibration Service
                                                                 Arlnlas®



Accredited by the Swiss Accreditation Service (SAS)                              Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
Convr                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
    b) 1EC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
         devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
         February 2005
    c) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    d) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
     s   Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
     e   Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
     e   Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
     a   SAR measured: SAR measured at the stated antenna input power.
     e   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
     e   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.




  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D835V2—4d117_May15                               Page 2 of 11


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY5                               V52.8.8

     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                             dx, dy, dz =5.0 mm
     Frequency                                          835 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           41.5               0.90 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   42.1 +6 %          0.92 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C                                    ————


SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     2.30 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         9.08 W/kg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition

     SAR measured                                        250 mW input power                  1.50 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         5.93 Wikg + 16.5 % (k=2)

Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Body TSL parameters                                  22.0 °C            55.2              0.97 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   55.2 + 6 %         0.99 mho/m +6 %
     Body TSL temperature change during test                      <0.5 °C            ————                    ————


SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
     SAR measured                                        250 mW input power                     2.38 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         9.38 W/ikg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition
     SAR measured                                        250 mW input power                     1.57 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         6.20 W/kg + 16.5 %(k=2)



Certificate No: D835V2—4d117_May15                    Page 3 of 11


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           52.7 Q — 2.0 |Q

      Return Loss                                                                       — 29.7 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           48.2 ) — 4.9 jQ
      Return Loss                                                                      — 25.5 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.392 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                 June 29, 2010




Certificate No: D835V2—4d117_May15                       Page 4 of 11


Appgendix (Additional assessments outside the scope of SCS 0108)

Measurement Conditions
   DASY system configuration, as far as not g iven on page 1 and 3.
     Phantom                                           SAM Head Phantom


SAR result with SAM Head (Top)

     SAR averaged over 1 ecm* (1 g) of Head TSL                Condition

     SAR measured                                        250 mW input power           2.12 W/kg
     SAR for nominal Head TSL parameters                   normalized to 1W    8.61 Wikg + 17.5 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL               condition

     SAR measured                                        250 mW input power           1.40 W/kg
     SAR for nominal Head TSL parameters                   normalized to 1W    5.66 W/kg + 16.9 % (k=2)

SAR result with SAM Head (Mouth)

     SAR averaged over 1 cm* (1 g) of Head TSL                 Condition

     SAR measured                                        250 mW input power           2.34 Wikg
     SAR for nominal Head TSL parameters                   normalized to 1W    9.51 W/kg + 17.5 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL               condition

     SAR measured                                        250 mW input power           1.59 W/kg
     SAR for nominal Head TSL parameters                   normalized to 1W    6.43 Wi/kg * 16.9 % (k=2)

SAR result with SAM Head (Neck)

     SAR averaged over 1 ecm* (1 g) of Head TSL                Condition

     SAR measured                                        250 mW input power            2.22 Wikg
     SAR for nominal Head TSL parameters                   normalized to 1W    9.02 Wi/kg * 17.5 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL               condition

     SAR measured                                        250 mW input power           1.50 W/kg
     SAR for nominal Head TSL parameters                   normalized to 1W    6.09 W/kg + 16.9 % (k=2)

SAR result with SAM Head (Ear)

     SAR averaged over 1 cm* (1 g) of Head TSL                 Condition

     SAR measured                                        250 mW input power           1.94 W/kg
     SAR for nominal Head TSL parameters                   normalized to 1W    7.88 Wi/kg + 17.5 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL               condition

     SAR measured                                         250 mW input power          1.30 W/kg
     SAR for nominal Head TSL parameters                   normalized to 1W    5.26 W/ikg + 16.9 % (k=2)


Certificate No: D835V2—4d117_May15                    Page 5 of 11


BASY5 Validation Report for Head TSL.
                                                                               Date: 12.05.2015

Test Laboratory: SPEAG, Zurich, Switzerland

DUT : Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN; 4d117

Communication System: UID 0 — CW; Frequency: 835 MHz
Medium parameters used: f = 835 MHz; 0 = 0.92 S/m; & = 42.1; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   e   Probe: ES3DV3 — SN3205; ConvF(6.2, 6.2, 6.2); Calibrated: 30.12.2014;

       Sensor—Surface: 3mm (Mechanical Surface Detection)

       Electronics: DAFE4 Sn601; Calibrated: 18.08.2014

       Phantom: Flat Phantom 4.9L; Type: QDOOOP49AA; Serial: 1001

       DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
   *




BDipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=>mm
Reference Value = 55.92 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 3.42 W/kg
SAR(1 g) = 2.3 W/kg; SAR(10 g) = 1.5 W/kg
Maximum value of SAR (measured) = 2.69 W/kg


       dB
        0

       —2.40

       —4.80

       —£.20

       —9.60

       —12.00
                  0 dB = 2.69 W/kg = 4.30 dBW/kg




Certificate No: D835V2—4d117_May15             Page 6 of 11


impedance Measurement Plot for Head TSL




                                                               12 May 20150 11:42:25
         CHJ sii     i U ors         1:52.697 &   —2.0059 4 95.924 pF       835.000 BBA MH=z
                                              /ff—’_r—_\-‘_‘




         De 1

         CA




         Av q
         16


         H1 d

          CH2




         CA




         Ay
         169



         H1 d

                   START 635.000 000 MHz                              STOP 1 #35.0900 o8B0 MHz




Certificate No: D835V2—4d117_May15                  Page 7 of 11


DASY5 Validation Report for Body TSL

                                                                                   Date: 12.05.2015

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 4d117

Communication System: UID 0 — CW; Frequency: 835 MHz
Medium parameters used: f = 835 MHz; 0 = 0.99 S/m; s, = 55.2; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e    Probe: ES3DV3 — SN3205; ConvF(6.17, 6.17, 6.17); Calibrated: 30.12.2014;

        Sensor—Surface: 3mm (Mechanical Surface Detection)

        Elecironics: DAF4 Sn601; Calibrated: 18.08.2014

        Phantom: Flat Phantom 4.9L; Type: QDOO0OPA49AA; Serial: 1001

        DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
   La




Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 54.70 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 3.47 W/kg
SAR(1 g) = 2.38 W/kg; SAR(10 g) = 1.57 W/kg
Maximum value of SAR (measured) = 2.78 W/kg


        dB
         a

        —2.40

        —4.80

        —f.20

        —9.650

        —12.00
                  0 dB = 2.78 W/kg = 4.44 dBW/kg




Certificate No: D835V2—4d117_May15             Page 8 of 11


Impedance Measurement Plot for Body TSL




                                                                12 May 2015      10:37:06
         [CHI] S11    1 U FS          1: 49.184 a   —4.8984 m 38.911 pF          835.000 090 MHz
                                                     *



         Del

         C¥




         Avg
         16


         H1d

          CH2




         Del

         CA




         Aw
         fee

         H1 d

                     START 635.000 600 MHz                                STOP 1 035.000 00@ MHz




Certificate No: D835V2—4d117_May15                   Page 9 of 11


DASY5 Vailidation Report for SAM Head
                                                                                Date: 18.05.2015

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 4d117

Communication System: UID 0 — CW; Frequency: 835 MHz
Medium parameters used: £ = 835 MHz; 0 = 0.89 S/m; &, = 42.9; p = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e    Probe: ES3DV3 — SN3205; ConvF(6.2, 6.2, 6.2); Calibrated: 30.12.2014;

   e    Sensor—Surface: 3mim (Mechanical Surface Detection)

   e    Electronics: DA 34 Sn601; Calibrated: 18.08.2014

   s    Phantom Type: SAM Head

   »—   DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)


SAM Head/Top/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 54.15 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 3.09 W/kg
SAR(1 g) = 2.12 W/kg; SAR(10 g) = 1.4 W/kg
Maximum value of SAR (measured) = 2.47 W/kg

SAM Head/Mouth/Zoom Scan (8x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 57.10 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 3.37 W/kg
SAR(1 g) = 2.34 W/kg; SAR(10 g) = i1.59 W/kg
Maximum value of SAR (measured) = 2.67 W/kg

SAM Head/Neck/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 55.26 V/m; Power Drift = —0.03 dB
Peak SAR (extrapolated) = 3.13 W/kg
SAR(1 g) = 2.22 W/kg; SAR(10 g) = 1.5 W/kg
Maximum value of SAR (measured) = 2.57 W/kg

SAM Head/Ear/Zoom Scan (7x8x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 51.21 V/m; Power Drift = 0.09 dB
Peak SAR (extrapolated) = 2.74 W/kg
SAR(1 g) = 1.94 W/kg; SAR(10 g) = 1.3 W/kg
Maximum value of SAR (measured) = 2.23 W/kg




Certificate No: D835V2—4d117_May15             Page 10 of 11


       —1.00



       —2.00



       —3.00



       —4.00



       —5.00
                  0 dB = 2.47 W/kg =3.93 dBW/kg




Certificate No: D835V2—4d117_May15          Page 11 of 11


Calibration Laboratory of                                                                                          Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                                                                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      L(NX                                             Swiss Calibration Service
                                                                    t uls®



Accredited by the Swiss Accreditation Service (SAS)                                            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       UL CCS USA                                                                        Certificate No: D835V2—4d142_Sep14

CALIBRATION CERTIFICATE                                                                                                                   ommemun
 Object                               D835V2 — SN: 40142


 Calibration procedure(s)             QA CAL—O5.v9
                                      Calibration procedure for dipole validation kits above 700 MHz




 Calibration date:                    September 09, 2014




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter EPM—442A                   GB37480704                09—Oct—13 (No. 217—01827)                             Oct—14
 Power sensor HP 8481A                  US37292783                09—Oct—13 (No. 217—01827)                             Oct—14
 Power sensor HP 8481A                  MY41092317                09—Oct—13 (No. 217—01828)                             Oct—14
 Reference 20 dB Attenuator             SN: 5058 (20k)            03—Apr—14 (No. 217—01918)                             Apr—15
 Type—N mismatch combination            SN: 5047.2 / 06327        03—Apr—14 (No. 217—01921)                             Apr—15
 Reference Probe ES3DV3                 SN: 3205                  30—Dec—13 (No. ES3—3205_Dec13)                        Dec—14
 DAE4                                   SN: 601                   18—Aug—14 (No. DAE4—601_Aug14)                        Aug—15


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 RF generator R&S SMT—O6                100005                    04—Aug—99 (in house check Oct—13)                     In house check: Oct—16
 Network Analyzer HP 8753E              US37390585 $4206          18—Oct—01 (in house check Oct—13)                     In house check: Oct—14




                                                                                                                            422
                                        Name                                   Function                                  Signature
 Calibrated by:                        Jeton Kastrati                          Laboratory Technician



 Approved by:                           Katia Pokovic                          Technical Manager



                                                                                                                        Issued: September 9, 2014
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D835V2—4d142_Sep14                                    Page 1 of 8


    k       s                                                   9l fr,,
Calibration Laboratory of                                    \\\“\i\://’ i        Schweizerischer Kalibrierdienst
Schmid & Partner                                            i%                C   Service suisse d‘étalonnage
   Engineering AG                                           2y¢_—\\ >             Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 % ///\\\ ww      S   Swiss Calibration Service
                                                               CZululubs®



Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvrF                       sensitivity in TSL / NORM x,y,2
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
         Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
         Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
         devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
         February 2005
    c) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    d) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    «_   Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    *    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    s    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    e    SAR measured: SAR measured at the stated antenna input power.
    e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.




 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D835V2—4d142_Sep14                              Page 2 of 8


Measurement Conditions
   DASY system configuration, as far as not g iven on page 1.
     DASY Version                                               DASY5                               V52.8.8
     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom

     Distance Dipole Center — TSL                               15 mm                             with Spacer
     Zoom Scan Resolution                               dx, dy, dz =5 mm
     Frequency                                          835 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity

     Nominal Head TSL parameters                                   22.0 °C           41.5                0.90 mho/m
     Measured Head TSL parameters                               (22.0 + 0.2) °C   41.8 x 6 %          0.93 mho/m + 6 %
     Head TSL temperature change during test                       <0.5 °C


SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                   Condition
     SAR measured                                         250 mW input power                     2.28 Wi/kg
     SAR for nominal Head TSL parameters                     normalized to 1W          8.91 W/kg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                  condition

     SAR measured                                         250 mW input power                  1.47 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W          5.77 Wikg * 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
     Nominal Body TSL parameters                                   22.0 °C            55.2               0.97 mho/m
     Measured Body TSL parameters                               (22.0 + 0.2) °C   54.8 x 6 %          1.01 mho/m +6 %
     Body TSL temperature change during test                       <0.5 °C


SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                    Condition
     SAR measured                                         250 mW input power                     2.38 W/kg

     SAR for nominal Body TSL parameters                     normalized to 1W          9.22 Wikg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                  condition

     SAR measured                                         250 mW input power                     1.55 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W          6.05 W/kg * 16.5 % (k=2)



Certificate No: D835V2—4d142_Sep14                     Page 3 of 8


Appendix (Additional assessments outside the scope of SCS108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           50.8 Q — 4.7 j|Q
      Return Loss                                                                       — 26.5 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           46.5 Q — 6.5 jQ
      Return Loss                                                                      — 22.3 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1.392 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                March 27, 2012




Certificate No: D835V2—4d142_Sep14                        Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                  Date: 09.09.2014

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 4d142

Communication System: UID 0 — CW; Frequency: 835 MHz
Medium parameters used: £= 835 MHz; 0 = 0.93 S/m; & = 41.8; p = 1000 kg/m"
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

       Probe: ES3DV3 — SN3205; ConvF(6.22, 6.22, 6.22); Calibrated: 30.12.2013;

       Sensor—Surface: 3mm (Mechanical Surface Detection)

       Electronics: DAE4 S$n601; Calibrated: 18.08.2014

       Phantom: Flat Phantom 4.9L; Type: QDOOOP49AA; Serial: 1001

       DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=>mm, dy=5mm, dz=5mm
Reference Value = 55.45 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 3.42 W/kg
SAR(1 g) = 2.28 W/kg; SAR(10 g) = 1.47 W/kg
Maximum value of SAR (measured) = 2.67 W/kg




       —£.20

       —9.60

       —12.00
                  0 dB = 2.67 W/kg =4.27 dBW/kg




Certificate No: D835V2—4d142_Sep14             Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                9 Sep 2014 11:38:06
          [CHI s11     i uo rs         750830 a   —4.7090 2 40.477 pF      835.000 Add MHz



          De1


          C&




         Avg
         16


          HId

          CH2




          CA




          As
          16"

          H1d

                     START 635.000 000 MHz                          STOP 1 935.000 080 MHz




Certificate No: D835V2—4d142_Sep14                  Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                  Date: 09.09.2014

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 4d142

Communication System: UID 0 — CW; Frequency: 835 MHz
Medium parameters used: £= 835 MHz; 0 = 1.01 S/m; s, = 54.8; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e   Probe: ES3DV3 — SN3205; ConvF(6.09, 6.09, 6.09); Calibrated: 30.12.2013;

   e   Sensor—Surface: 3mm (Mechanical Surface Detection)

   e   Electronics: DAE4 Sn601; Calibrated: 18.08.2014

   «   Phantom: Flat Phantom 4.9L; Type: QDOOOPA9AA; Serial: 1001

   e   DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)


Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 54.65 V/m; Power Drift = —0.06 dB
Peak SAR (extrapolated) = 3.52 W/kg
SAR(1 g) = 2.38 W/kg; SAR(10 g) = 1.55 W/kg
Maximum value of SAR (measured) = 2.78 W/kg




       —2.00

       —4.00

       —6.00

       —8.00

       —10.00
                  0 dB = 2.78 W/kg = 4.44 dBW/kg




Certificate No: D835V2—4d142_Sep14            Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                      9 Sep 2014     10:45:45

                                                  sAF**
         [CHI) §s41     1 U FS          21 46.480 &   —6.5059 a   29.297 pF          835.000 080 MHz

                                                .

         De l

         C&




         Av
         :1.69


         H1d

          CH2




         CA




         Aw
         16 "

         H1d

                      START 6£35.000 900 MHz                                  STOP 1 935.000 000 MHz




Certificate No: D835V2—4d142_Sep14                      Page 8 of 8


Calibration Laboratory of                                                                                          Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                                                                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                               Accreditation No.: SCS 108
The Swiss Accreditation: Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       ULCCS USA                                                                            Certificate No: D1900V2—5d163_Sep14

CALIBRATION CERTIFEATE                                                                                                                               I
 Object                                D1900V2 — SN: 5d163


 Calibration procedure(s)              QA CAL—O5.v9
                                       Calibration procedure for dipole validation kits above 700 MHz




 Calibration date:                     September 11, 2014




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter EPM—442A                   GB37480704                09—Oct—13   (No.   217—01827)                         Oct—14
 Power sensor HP 8481A                  US37292783                09—Oct—13   (No.   217—01827)                         Oct—14
 Power sensor HP 8481A                  MY41092317                09—Oct—13   (No.   217—01828)                         Oct—14
 Reference 20 dB Attenuator             SN: 5058 (20k)            03—Apr—14   (No.   217—01918)                         Apr—15
 Type—N mismatch combination            SN: 5047.2 / 06327        03—Apr—14   (No.   217—01921)                         Apr—15
 Reference Probe ES3SDV3                SN: 3205                  30—Dec—13 (No. ES3—3205_Dec13)                        Dec—14
 DAE4                                   SN: 601                   18—Aug—14 (No. DAE4—601_Aug14)                        Aug—15

 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 RF generator R&S SMT—O6                100005                    04—Aug—99 (in house check Oct—13)                     In house check: Oct—16
 Network Analyzer HP 8753E              US37390585 $4206          18—Oct—01 (in house check Oct—13)                     In house check: Oct—14



                                        Name                                   Function
 Calibrated by:                         Claudio Leubler                        Laboratory Technician



 Approved by:                           Katjia Pokovic                         Technical Manager



                                                                                                                       Issued; September 11, 2014
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D1900V2—50163_Sep14                                   Page 1 of 8


Calibration Laboratory of                                                         Schweizerischer Kalibrierdienst
Schmid & Partner                                                                  Service suisse d‘étalonnage
  Engineering AG                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                      Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvrF                       sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
         Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
         Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
         devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
    c) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    d) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    *    Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    *«   Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    a    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    «    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    e    SAR measured: SAR measured at the stated antenna input power.
    *    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    »    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.




 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D1900V2—5d163_Sep14                              Page 2 of 8


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY5                               V52.8.8

     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom

     Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         1900 MHz +1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           40.0               1.40 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   39.5 £ 6 %         1.37 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C            wme                     mes:

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     10.1 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        40.8 W/kg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition

     SAR measured                                        250 mW input power                     5.27 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W        21.2 Wikg * 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C            53.3              1.52 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   52.6 +6 %          1.50 mho/m x 6 %
     Body TSL temperature change during test                      <0.5 °C


SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
     SAR measured                                        250 mW input power                     10.1 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         40.6 Wikg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition

     SAR measured                                        250 mW input power                  5.34 Wi/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         21.4 Wikg + 16.5 % (k=2)



Certificate No: D1900V2—5d163_Sep14                   Page 3 of 8


Appendix (Additional assessments outside the scope of SCS108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           52.3 Q + 4.9 jQ
      Return Loss                                                                       — 25.5 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           48.6 O + 5.4 jQ
      Return Loss                                                                      — 25.0 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.164 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                              December 20, 2011




Certificate No: D1900V2—5d163_Sep14                       Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                  Date: 11.09.2014

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 5d163

Communication System: UID 0 — CW; Frequency: 1900 MHz
Medium parameters used: £= 1900 MHz; o = 1.37 S/m; &, = 39.5; p = 1000 kg/m"
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e   Probe: ES3DV3 — SN3205; ConvF(5.06, 5.06, 5.06); Calibrated: 30.12.2013;

   e   Sensor—Surface: 3mm (Mechanical Surface Detection)

   e   Electronics: DAF4 Sn601; Calibrated: 18.08.2014

   e   Phantom: Flat Phantom 5.0 (front); Type: QDOOOP5OAA; Serial: 1001

   e   DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)


Dipole Calibration for Head Tissue/Pin=250 mW, d=1{mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 98.95 V/m; Power Drift = 0.06 dB
Peak SAR (extrapolated) = 18.4 W/kg
SAR(1 g) = 10.1 W/kg; SAR(10 g) = 5.27 W/kg
Maximum value of SAR (measured) = 12.7 W/kg




                            3.47




                            —6.94




                            ~10.40




                            13.87




                            ~17.34

                                     0 dB = 12.7 W/kg = 11.04 dBW/kg



Certificate No: D1900V2—5d163_Sep14                 Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                   11 Sep 28014    14:i17:47
         CHI S41     1 U FS           1:52,344 a   4.87F70A   408.52 pH           1 960.000 000A MHz
                                               onsse

         Del

         Cor




         Avg
         16


         H1d


          CH2




         Cor




         As
         16°

         H1d

                   START 1 790.900 900 MHz                                STOP 2 180,.000 900 MHz




Certificate No: D1900V2—5d163_Sep14                  Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                  Date: 11.09.2014

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 5d163

Communication System: UID 0 — CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; 0 = 1.5 S/m; s = 52.6; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e   Probe: ES3DV3 — SN3205; ConvF(4.76, 4.76, 4.76); Calibrated: 30.12.2013;

       Sensor—Surface: 3mm (Mechanical Surface Detection)

       Electronics: DAE4 S$Sn601; Calibrated: 18.08.2014

       Phantom: Flat Phantom 5.0 (back); Type: QDOOOP5OAA; Serial: 1002

       DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 96.06 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 17.5 W/kg
SAR(1 g) = 10.1 W/kg; SAR(10 g) = 5.34 W/kg
Maximum value of SAR (measured) = 12.7 W/kg




                            —3.32



                            —6.65




                            ~13.30



                            16.62

                                     0 dB = 12.7 W/kg = 11.04 dBW/kg




Certificate No: D1900V2—5d163_Sep14                 Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                  11 Sep 2014     14116132
         CHZl s11     1 U FS          1:48,534 a   5.3555 &   448.60 pH          1 9985.900 000 MHz




         Oe 1

         Cor




         Avg
         16


         H14
          CH2




         Cor




         AAw
         16°

         H1 d

                    START 1 708.000 900 MHz                               STOP 2 100.0080 000 MHz




Certificate No: D1900V2—5d163_Sep14                 Page 8 of 8



Document Created: 2015-06-30 16:45:33
Document Modified: 2015-06-30 16:45:33

© 2025 FCC.report
This site is not affiliated with or endorsed by the FCC