HAC RF Emission Appendix_B

FCC ID: ZNFUS550

Test Report

Download: PDF
FCCID_2562652

Test Laboratory: UL Verification Services Inc., SAR Lab 2                       Date: 2/23/2015

HAC-RF Emission
Communication System: UID 0, CW; Frequency: 835 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2540; ConvF(1, 1, 1); Calibrated: 8/26/2014;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1259; Calibrated: 1/14/2015
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/835 MHz/Hearing Aid Compatibility Test at 15mm distance
(41x361x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 128.5 V/m; Power Drift = -0.02 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 116.8 V/m
Near-field category: M4 (AWF 0 dB)

                                          PMF scaled E-field
                                          Grid 1 M4 Grid 2 M4 Grid 3 M4
                                          113.3 V/m     116.8 V/m   115.1 V/m
                                          Grid 4 M4     Grid 5 M4   Grid 6 M4
                                          65.05 V/m     66.88 V/m   66.38 V/m
                                          Grid 7 M4     Grid 8 M4   Grid 9 M4
                                          107.2 V/m     109.6 V/m   108.3 V/m




                      0 dB = 116.8 V/m = 41.35 dBV/m


Test Laboratory: UL Verification Services Inc., SAR Lab 2                       Date: 2/23/2015

HAC-RF Emission
Communication System: UID 0, CW; Frequency: 1880 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2540; ConvF(1, 1, 1); Calibrated: 8/26/2014;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1259; Calibrated: 1/14/2015
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/1880 MHz/Hearing Aid Compatibility Test at 15mm
distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 124.2 V/m; Power Drift = 0.07 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 92.76 V/m
Near-field category: M3 (AWF 0 dB)

                                          PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          91.21 V/m     92.76 V/m   90.62 V/m
                                          Grid 4 M3     Grid 5 M3   Grid 6 M3
                                          66.71 V/m     68.34 V/m   67.96 V/m
                                          Grid 7 M3     Grid 8 M3   Grid 9 M3
                                          82.98 V/m     84.66 V/m   83.50 V/m




                      0 dB = 92.76 V/m = 39.35 dBV/m