HAC RF Test Report2

FCC ID: ZNFLS770

Test Report

Download: PDF
FCCID_2576244

Test Laboratory: UL Verification Services Inc., SAR Lab 2                           Date: 3/9/2015

HAC-RF Emission
Communication System: UID 0, CW; Frequency: 835 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2540; ConvF(1, 1, 1); Calibrated: 8/26/2014;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1259; Calibrated: 1/14/2015
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/835 MHz/Hearing Aid Compatibility Test at 15mm distance
(41x361x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 133.2 V/m; Power Drift = -0.05 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 118.1 V/m
Near-field category: M4 (AWF 0 dB)

                                          PMF scaled E-field
                                          Grid 1 M4 Grid 2 M4 Grid 3 M4
                                          114.2 V/m         118.1 V/m   117.3 V/m
                                          Grid 4 M4         Grid 5 M4   Grid 6 M4
                                          66.19 V/m         69.06 V/m   68.70 V/m
                                          Grid 7 M4         Grid 8 M4   Grid 9 M4
                                          110.1 V/m         114.0 V/m   112.6 V/m




                                                   0 dB = 118.1 V/m = 41.44 dBV/m



                                                                                                     Plot No. 1


Test Laboratory: UL Verification Services Inc., SAR Lab 2                           Date: 3/9/2015

HAC-RF Emission
Communication System: UID 0, CW; Frequency: 1880 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2540; ConvF(1, 1, 1); Calibrated: 8/26/2014;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1259; Calibrated: 1/14/2015
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)



Dipole E-Field measurement/1880 MHz/Hearing Aid Compatibility Test at 15mm
distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 140.6 V/m; Power Drift = 0.08 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 89.13 V/m
Near-field category: M3 (AWF 0 dB)

                                          PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          84.01 V/m         85.10 V/m   83.26 V/m
                                          Grid 4 M3         Grid 5 M3   Grid 6 M3
                                          68.46 V/m         70.94 V/m   70.83 V/m
                                          Grid 7 M3         Grid 8 M3   Grid 9 M3
                                          86.54 V/m         89.13 V/m   88.37 V/m




                      0 dB = 89.13 V/m = 39.00 dBV/m


                                                                                                     Plot No. 2


Test Laboratory: UL Verification Services Inc., SAR Lab 2                           Date: 3/10/2015

HAC-RF Emission
Communication System: UID 0, CW (0); Frequency: 2600 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2540; ConvF(1, 1, 1); Calibrated: 8/26/2014;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1259; Calibrated: 1/14/2015
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/2600 MHz/Hearing Aid Compatibility Test at 15mm
distance (41x361x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 71.96 V/m; Power Drift = -0.04 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 91.64 V/m
Near-field category: M3 (AWF 0 dB)

                                          PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          78.13 V/m         79.15 V/m   76.77 V/m
                                          Grid 4 M3         Grid 5 M3   Grid 6 M3
                                          89.43 V/m         91.64 V/m   90.57 V/m
                                          Grid 7 M3         Grid 8 M3   Grid 9 M3
                                          79.97 V/m         80.48 V/m   77.92 V/m




                       0 dB = 91.64 V/m = 39.24 dBV/m



                                                                                                      Plot No. 3