HAC RF Appendix_14.1

FCC ID: ZNFLS770

Test Report

Download: PDF
FCCID_2542831

Test Laboratory: UL Verification Services Inc., SAR Lab 2                           Date: 1/22/2015

HAC-RF Emission
Communication System: UID 0, CW; Frequency: 835 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2540; ConvF(1, 1, 1); Calibrated: 8/26/2014;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1359; Calibrated: 2/17/2014
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/835 MHz/Hearing Aid Compatibility Test at 15mm distance
(41x361x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 124.7 V/m; Power Drift = 0.16 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 110.7 V/m
Near-field category: M4 (AWF 0 dB)

                                          PMF scaled E-field
                                          Grid 1 M4 Grid 2 M4 Grid 3 M4
                                          109.0 V/m         110.6 V/m   109.1 V/m
                                          Grid 4 M4         Grid 5 M4   Grid 6 M4
                                          64.08 V/m         65.47 V/m   65.47 V/m
                                          Grid 7 M4         Grid 8 M4   Grid 9 M4
                                          107.7 V/m         110.7 V/m   109.1 V/m




                      0 dB = 110.7 V/m = 40.88 dBV/m


Test Laboratory: UL Verification Services Inc., SAR Lab 2                           Date: 1/22/2015

HAC-RF Emission
Communication System: UID 0, CW; Frequency: 1880 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2540; ConvF(1, 1, 1); Calibrated: 8/26/2014;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1359; Calibrated: 2/17/2014
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/1880 MHz/Hearing Aid Compatibility Test at 15mm
distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 133.8 V/m; Power Drift = 0.04 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 87.98 V/m
Near-field category: M3 (AWF 0 dB)

                                          PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          83.01 V/m         84.01 V/m   82.28 V/m
                                          Grid 4 M3         Grid 5 M3   Grid 6 M3
                                          67.29 V/m         69.51 V/m   69.15 V/m
                                          Grid 7 M3         Grid 8 M3   Grid 9 M3
                                          85.70 V/m         87.98 V/m   87.04 V/m




                      0 dB = 87.98 V/m = 38.89 dBV/m


Test Laboratory: UL Verification Services Inc., SAR Lab 2                           Date: 2/5/2015

HAC-RF Emission
Communication System: UID 0, CW; Frequency: 835 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2540; ConvF(1, 1, 1); Calibrated: 8/26/2014;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1359; Calibrated: 2/17/2014
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/835 MHz/Hearing Aid Compatibility Test at 15mm distance
(41x361x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 121.4 V/m; Power Drift = -0.03 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 109.2 V/m
Near-field category: M4 (AWF 0 dB)

                                          PMF scaled E-field
                                          Grid 1 M4 Grid 2 M4 Grid 3 M4
                                          98.61 V/m         102.8 V/m   99.39 V/m
                                          Grid 4 M4         Grid 5 M4   Grid 6 M4
                                          61.60 V/m         63.46 V/m   63.07 V/m
                                          Grid 7 M4         Grid 8 M4   Grid 9 M4
                                          107.6 V/m         109.2 V/m   109.1 V/m




                      0 dB = 109.2 V/m = 40.76 dBV/m


Test Laboratory: UL Verification Services Inc., SAR Lab 2                           Date: 2/5/2015

HAC-RF Emission
Communication System: UID 0, CW; Frequency: 1880 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2540; ConvF(1, 1, 1); Calibrated: 8/26/2014;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1359; Calibrated: 2/17/2014
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/1880 MHz/Hearing Aid Compatibility Test at 15mm
distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 127.2 V/m; Power Drift = 0.04 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 88.62 V/m
Near-field category: M3 (AWF 0 dB)

                                          PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          87.78 V/m         88.62 V/m   86.16 V/m
                                          Grid 4 M3         Grid 5 M3   Grid 6 M3
                                          66.25 V/m         67.80 V/m   67.46 V/m
                                          Grid 7 M3         Grid 8 M3   Grid 9 M3
                                          83.53 V/m         84.95 V/m   83.87 V/m




                      0 dB = 88.62 V/m = 38.95 dBV/m



Document Created: 2015-02-12 18:36:35
Document Modified: 2015-02-12 18:36:35