16I22652-S2V1 HAC_T_App C Probe Cal. Certificates

FCC ID: ZNFL61AL

Test Report

Download: PDF
FCCID_2902833

Calibration Laboratory of                                                                                           Schweizerischer Kalibrierdienst
                                                                                                              S
Schmid & Partner                                                                                                    Service suisse d‘étalonnage
  Engineering AG                                                                                              C     Servizio svizzero di taratura
                                                                                      y
Zeughausstrasse 43, 8004 Zurich, Switzerland                       L( NY                                      S     Swiss Calibration Service
                                                                    ’//’l/nh\\\‘\\\




Accredited by the Swiss Accreditation Service (SAS)                                                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       UL CCS USA                                                                             Certificate No: AM1DV3—3092_Jul15


CALIBRATION CERTIFICATE
 Object                                AM1DV3 — SN: 3092


 Calibration procedure(s)              QA CAL—24.v4
                                       Calibration procedure for AM1D magnetic field probes and TMFS in the
                                       audio range

 Calibration date:                     July 9, 2015




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (Sl).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                       ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Keithley Multimeter Type 2001           SN: 0810278               03—Oct—14 (No:15573)                                  Oct—15
 Reference Probe AM1DV2                  SN: 1008                  08—Jan—15 (No. AM1D—1008_Jan15)                       Jan—16
 DAE4                                    SN: 781                   12—Sep—14 (No. DAE4—781_Sep14)                        Sep—15



 Secondary Standards                     ID #                      Check Date (in house)                                 Scheduled Check
 AMCC                                    1050                      01—Oct—13 (in house check Oct—13)                     Oct—16
 AMMI Audio Measuring Instrument         1062                      26—Sep—12 (in house check Sep—12)                     Sep—15




                                         Name                                         Function                            Signature
  Calibrated by:



  Approved by:
                                         Leif Klysner



                                         Katia Pokovic
                                                                                      Laboratory Technician



                                                                                      Technical Manager       W
                                                                                                                     Re Bc
                                                                                                                          Issued: July 14, 2015
  This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



Certificate No: AM1DV3—3092_Jul15                                      Page 1 of 3


[References

[1]      ANSI—C63.19—2007
         American National Standard for Methods of Measurement of Compatibility between Wireless
         Communications Devices and Hearing Aids.
[2]       ANSI—C63.19—2011
         American National Standard, Methods of Measurement of Compatibility between Wireless
         Communications Devices and Hearing Aids.
[3]      DASY5 manual, Chapter: Hearing Aid Compatibility (HAC) T—Coil Extension

Description of the AM1D prokbe

      The AM1D Audio Magnetic Field Probe is a fully shielded magnetic field probe for the frequency
      range from 100 Hz to 20 kHz. The pickup coil is compliant with the dimensional requirements of
      [1+2]. The probe includes a symmetric low noise amplifier for the signal available at the shielded
      3 pin connector at the side. Power is supplied via the same connector (phantom power supply)
      and monitored via the LED near the connector. The 7 pin connector at the end of the probe does
      not carry any signals, but determines the angle of the sensor when mounted on the DAE. The
      probe supports mechanical detection of the surface.
      The single sensor in the probe is arranged in a tilt angle allowing measurement of 3 orthogonal
      field components when rotating the probe by 120° around its axis. It is aligned with the
      perpendicular component of the field, if the probe axis is tilted nominally 35.3° above the
      measurement plane, using the connector rotation and sensor angle stated below.
      The probe is fully RF shielded when operated with the matching signal cable (shielded) and
      allows measurement of audio magnetic fields in the close vicinity of RF emitting wireless devices
      according to [1+2] without additional shielding.

Handling of the item

      The probe is manufactured from stainless steel. In order to maintain the performance and
      calibration of the probe, it must not be opened. The probe is designed for operation in air and
      shall not be exposed to humidity or liquids. For proper operation of the surface detection and
      emergency stop functions in a DASY system, the probe must be operated with the special probe
      cup provided (larger diameter).

Methods Applied and Interpretation of Parameters

«     Coordinate System: The AM1D probe is mounted in the DASY system for operation with a HAC
      Test Arch phantom with AMCC Helmholtz calibration coil according to [3], with the tip pointing to
      "southwest" orientation.
*     Functional Test: The functional test preceding calibration includes test of
      Noise level
      RF immunity (1kHz AM modulated signal). The shield of the probe cable must be well connected.
      Frequency response verification from 100 Hz to 10 kHz.
=     Connector Rotation: The connector at the end of the probe does not carry any signals and is used
      for fixation to the DAE only. The probe is operated in the center of the AMCC Helmholtz coil using a
      1 kHz magnetic field signal. Its angle is determined from the two minima at nominally +120° and —
      120° rotation, so the sensor in the tip of the probe is aligned to the vertical plane in z—direction,
      corresponding to the field maximum in the AMCC Helmholtz calibration coil.
a     Sensor Angle: The sensor tilting in the vertical plane from the ideal vertical direction is determined
      from the two minima at nominally +120° and —120°. DASY system uses this angle to align the
      sensor for radial measurements to the x and y axis in the horizontal plane.
Sensitivity: With the probe sensor aligned to the z—field in the AMCC, the output of the probe is
         compared to the magnetic field in the AMCC at 1 kHz. The field in the AMCC Helmholtz coil is
         given by the geometry and the current through the coil, which is monitored on the precision
         shunt resistor of the coil.

Certificate No: AM1 DV3—3092_Jul15                   Page 2 of 3


AM1D probe identification and configuration data

    Item                        AM1DV3 Audio Magnetic 1D Field Probe
    Type No                     SP AM1 001 BA
    Serial No                   3092


    Overall length              296 mm
    Tip diameter                6.0 mm (at the tip)
    Sensor offset               3.0 mm (centre of sensor from tip)
    Internal Amplifier          20 dB


    Manufacturer / Origin       Schmid & Partner Engineering AG, ZUrich, Switzerland
    Manufacturing date          March 03, 2011
    Last calibration date       July 18, 2014




Calibration data



Connector rotation angle       (in DASY system)          276.8 °             +/— 3.6 ° (k=2)


Sensor angle                   (in DASY system)          0.82 °              +/— 0.5 ° (k=2)


Sensitivity at 1 kHz            (in DASY system)         0.00733 V / (A/m)   +/— 2.2 % (k=2)




 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.