15I21523-S2V1 HAC_App B System Check Plots

FCC ID: ZNFH790

Test Report

Download: PDF
FCCID_2766822

Test Laboratory: UL Verification Services Inc. SAR Lab C                               Date: 8/25/2015

HAC-RF Emission

Communication System: UID 0, CW; Frequency: 835 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2339; ConvF(1, 1, 1); Calibrated: 2/26/2015;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE3 Sn427; Calibrated: 1/14/2015
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/835 MHz/Hearing Aid Compatibility Test at 15mm distance (41x361x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 118.0 V/m; Power Drift = -0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 106.2 V/m
Near-field category: M4 (AWF 0 dB)
                                                  PMF scaled E-field
                                                  Grid 1 M4 Grid 2 M4 Grid 3 M4
                                                  104.4 V/m 106.2 V/m 104.5 V/m
                                                  Grid 4 M4 Grid 5 M4 Grid 6 M4
                                                  60.39 V/m 61.30 V/m 60.56 V/m
                                                  Grid 7 M4 Grid 8 M4 Grid 9 M4
                                                  103.9 V/m 105.1 V/m 102.7 V/m
Cursor:
Total = 106.2 V/m
E Category: M4
Location: 0, -78, 9.7 mm




              0 dB = 106.2 V/m = 40.52 dBV/m




                                                                                                                Plot No. 1


Test Laboratory: UL Verification Services Inc. SAR Lab C                               Date: 8/25/2015

HAC-RF Emission

Communication System: UID 0, CW; Frequency: 1880 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2339; ConvF(1, 1, 1); Calibrated: 2/26/2015;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE3 Sn427; Calibrated: 1/14/2015
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/1880 MHz/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 137.9 V/m; Power Drift = 0.02 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 92.32 V/m
Near-field category: M3 (AWF 0 dB)
                                                 PMF scaled E-field
                                                  Grid 1 M3 Grid 2 M3 Grid 3 M3
                                                  79.66 V/m 80.85 V/m 80.22 V/m
                                                  Grid 4 M3 Grid 5 M3 Grid 6 M3
                                                  72.46 V/m 73.08 V/m 71.91 V/m
                                                  Grid 7 M3 Grid 8 M3 Grid 9 M3
                                                  91.95 V/m 92.32 V/m 89.39 V/m
Cursor:
Total = 92.32 V/m
E Category: M3
Location: 1.5, 31, 9.7 mm




              0 dB = 92.32 V/m = 39.31 dBV/m




                                                                                                                Plot No. 2


Test Laboratory: UL Verification Services Inc. SAR Lab C                               Date: 9/1/2015

HAC-RF Emission

Communication System: UID 0, CW; Frequency: 1880 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2339; ConvF(1, 1, 1); Calibrated: 2/26/2015;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE3 Sn427; Calibrated: 1/14/2015
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/1880 MHz/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 133.7 V/m; Power Drift = 0.02 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 89.83 V/m
Near-field category: M3 (AWF 0 dB)
                                                 PMF scaled E-field
                                                  Grid 1 M3 Grid 2 M3 Grid 3 M3
                                                  84.09 V/m 84.96 V/m 83.66 V/m
                                                  Grid 4 M3 Grid 5 M3 Grid 6 M3
                                                  71.36 V/m 72.62 V/m 72.15 V/m
                                                  Grid 7 M3 Grid 8 M3 Grid 9 M3
                                                  88.36 V/m 89.83 V/m 88.43 V/m
Cursor:
Total = 89.83 V/m
E Category: M3
Location: 0, 30.5, 9.7 mm




              0 dB = 89.83 V/m = 39.07 dBV/m




                                                                                                                Plot No. 3


Test Laboratory: UL Verification Services Inc. SAR Lab C                               Date: 8/25/2015

HAC-RF Emission

Communication System: UID 0, CW (0); Frequency: 2600 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: ER3DV6 - SN2339; ConvF(1, 1, 1); Calibrated: 2/26/2015;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE3 Sn427; Calibrated: 1/14/2015
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement/2600 MHz/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 68.28 V/m; Power Drift = -0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 90.99 V/m
Near-field category: M3 (AWF 0 dB)
                                                  PMF scaled E-field
                                                  Grid 1 M3 Grid 2 M3 Grid 3 M3
                                                  84.08 V/m 84.62 V/m 83.17 V/m
                                                  Grid 4 M3 Grid 5 M3 Grid 6 M3
                                                  87.27 V/m 88.25 V/m 86.62 V/m
                                                  Grid 7 M3 Grid 8 M3 Grid 9 M3
                                                  90.18 V/m 90.99 V/m 88.79 V/m
Cursor:
Total = 90.99 V/m
E Category: M3
Location: 1, 20, 9.7 mm




              0 dB = 90.99 V/m = 39.18 dBV/m




                                                                                                                Plot No. 4



Document Created: 2015-09-14 21:50:30
Document Modified: 2015-09-14 21:50:30

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC