              HAC T-COIL SIGNAL TEST REPORT

               CLASS II PERMISSIVE CHANGE

                   FCC CFR47 PART 20.19
                     ANSI C63.19-2011

                          For
GSM/WCDMA/LTE PHABLET + BLUETOOTH, DTS/UNII a/b/g/n and NFC

                     FCC ID: ZNFH740
             Model Name: LG-H740, LGH740, H740


                Report Number: 15I21442-S2V1
                    Issue Date: 9/10/2015

                        Prepared for
           LG ELECTRONICS MOBILECOMM USA, INC.
                   1000 SYLVAN AVENUE
         ENGLEWOOD CLIFFS, NEW JERSEY 07632, USA

                         Prepared by
               UL VERIFICATION SERVICES INC.
                   47173 BENICIA STREET
                 FREMONT, CA 94538, U.S.A.
                     TEL: (510) 771-1000
                     FAX: (510) 661-0888




                   NVLAP LAB CODE 200065-0


Report No.: 15I21442-S2V1                                                                        Issue Date: 9/10/2015
Revision History
 Rev.    Date             Revisions                                                        Revised By
 V1      9/10/2015        Initial Issue                                                    --




                                                    Page 2 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                                                     Issue Date: 9/10/2015

                                                           Table of Contents
1.     Attestation of Test Results ............................................................................................................. 4

2.     Test Methodology ........................................................................................................................... 5

3.     Facilities and Accreditation ........................................................................................................... 5

4.     Calibration and Uncertainty ........................................................................................................... 5
     4.1.    Measuring Instrument Calibration .............................................................................................. 5
     4.2.    Measurement Uncertainty .......................................................................................................... 6

5.     Test Procedures for all Technologies ........................................................................................... 7

6.     Audio Level and Gain Measurements............................................................................................ 9

7.     T-coil Measurement Criteria ......................................................................................................... 10
     7.1.    Frequency Response ............................................................................................................... 10
     7.2.    Signal to Noise......................................................................................................................... 11

8.     Device Under Test......................................................................................................................... 12
     8.1.    Air Interfaces and Operating Mode .......................................................................................... 12

9.     HAC (T-coil) Test Results ............................................................................................................. 13
     9.1.    Worst Case T-Coil Test Plot..................................................................................................... 14

Appendix .............................................................................................................................................. 15
     15I21442-S2V1 HAC_T_App A Setup Photo ...................................................................................... 15
     15I21442-S2V1 HAC_T_App B Frequency Response & SNR Test Plots ........................................... 15
     15I21442-S2V1 HAC_T_App C Probe Cal. Certificates ...................................................................... 15




                                                    Page 3 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                        Issue Date: 9/10/2015

1. Attestation of Test Results
Applicant Name                    LG ELECTRONICS MOBILECOMM USA, INC.
FCC ID                            ZNFH740
Model Name                        LG-H740, LGH740, H740
                                  FCC 47 CFR § 20.19
                                  ANSI C63.19-2011
Applicable Standards
                                  KDB 285076 D01 HAC Guidance
                                  TCB workshop updates
HAC Rating                        T4
Date Tested                       8/4/2015 to 8/5/2015
Test Results                      Pass
 UL Verification Services Inc. tested the above equipment in accordance with the requirements set forth in the
 above standards. All indications of Pass/Fail in this report are opinions expressed by UL Verification Services Inc.
 based on interpretations and/or observations of test results. Measurement Uncertainties were not taken into
 account and are published for informational purposes only. The test results show that the equipment tested is
 capable of demonstrating compliance with the requirements as documented in this report.
Note: The results documented in this report apply only to the tested sample, under the conditions and modes of
operation as described herein. This document may not be altered or revised in any way unless done so by UL
Verification Services Inc. and all revisions are duly noted in the revisions section. Any alteration of this document
not carried out by UL Verification Services Inc. will constitute fraud and shall nullify the document. This report must
not be used by the client to claim product certification, approval, or endorsement by NVLAP, NIST, any agency of
the Federal Government, or any agency of any government (NIST Handbook 150, Annex A). This report is written
to support regulatory compliance of the applicable standards stated above.
 Approved & Released By:                                      Prepared By:




Devin Chang                                                 AJ Newcomer
Senior Engineer                                             Laboratory Technician
UL Verification Services Inc.                               UL Verification Services Inc.




                                                    Page 4 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                         Issue Date: 9/10/2015

2. Test Methodology
The tests documented in this report were performed in accordance with ANSI C63.19-2011 Methods of
Measurement of Compatibility between Wireless Communications Devices and Hearing Aids and FCC published
procedure KDB 285076 D01 HAC Guidance v04 and TCB workshop updates



3. Facilities and Accreditation
The test sites and measurement facilities used to collect data are located at
                   47173 Benicia Street                        47266 Benicia Street
                       SAR Lab C                                     SAR Lab 2

UL Verification Services Inc. is accredited by NVLAP, Laboratory Code 200065-0.


4. Calibration and Uncertainty
4.1.    Measuring Instrument Calibration
The measuring equipment used to perform the tests documented in this report has been calibrated in
accordance with the manufacturers’ recommendations, and is traceable to recognized national
standards.
Name of Equipment                     Manufacturer             Type/Model               Serial No.     Cal. Due Date
Robot - Six Axes                          Stäubli               TX90 XL                    N/A             N/A
Robot Remote Control                      Stäubli                 CS8C                     N/A             N/A
DASY5 Measurement Server                  SPEAG               SEUMS001BA                  1041             N/A
Probe Alignment Unit                      SPEAG                  LB (V2)                   261             N/A
Audio Magnetic Measuring Ins.I            SPEAG                   AMMI                    1127             N/A
Coordinating SystemI                      SPEAG                  AMCC                      N/A             N/A
ABM Probe                                 SPEAG                 AM1DV3                    3083           1/15/2016
Data Acquisition Electronics              SPEAG                   DAE4                    1259           1/14/2016
Radio Communication Tester                R&S                   CMU 500                  125236          1/19/2016




                                                    Page 5 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                          Issue Date: 9/10/2015

4.2.      Measurement Uncertainty
Measurement Uncertainty for Audio Band Magnetic Measurement
                                                        Uncertainty   Probe            c      c          Std. Unc.
Error Description                                       values (%)    Dist.   Div.   AMB1   AMB2   AMB1 (%)    AMB2 (%)
Probe Sensitivity
Reference level                                             3.0        N       1     1.00   1.00     3.00       3.00
AMCC geometry                                               0.4        R      1.73   1.00   1.00     0.23       0.23
AMCC current                                                0.6        R      1.73   1.00   1.00     0.35       0.35
Probe positioning during calibration                        0.1        R      1.73   1.00   1.00     0.06       0.06
Noise contribution                                          0.7        R      1.73   0.01   1.00     0.01       0.40
Frequency slope                                             5.9        R      1.73   0.10   1.00     0.34       3.41
Probe System
Repeatability / drift                                       1.0        R      1.73   1.00   1.00     0.58       0.58
Linearity / Dynamic range                                   0.6        R      1.73   1.00   1.00     0.35       0.35
Acoustic noise                                              1.0        R      1.73   0.10   1.00     0.06       0.58
Probe angle                                                 2.3        R      1.73   1.00   1.00     1.33       1.33
Spectral processing                                         0.9        R      1.73   1.00   1.00     0.52       0.52
Integration time                                            0.6        N      1.00   1.00   5.00     0.60       3.00
Field disturbation                                          0.2        R      1.73   1.00   1.00     0.12       0.12
Test Signal
Reference signal spectral response                          0.6        R      1.73   0.00   1.00     0.00       0.35
Positioning
Probe positioning                                           1.9        R      1.73   1.00   1.00     1.10       1.10
Phantom positioning                                         0.9        R      1.73   1.00   1.00     0.52       0.52
EUT positioning                                             1.9        R      1.73   1.00   1.00     1.10       1.10
External Contributions
RF interference                                             0.0        R      1.73   1.00   1.00     0.00       0.00
Test signal variation                                       2.0        R      1.73   1.00   1.00     1.15       1.15
Combined Std. Uncertainty (ABM field)                                                                4.02       6.08
Expanded Std. Uncertainty (%)                                                                        8.04       12.15
Notes for table
1. N - Nomal
2. R - Rectangular
3. Div. - Divisor used to obtain standard uncertainty




                                                    Page 6 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                           Issue Date: 9/10/2015

5. Test Procedures for all Technologies
ANSI C63.19-2011, Section 7
This document describes the procedures used to measure the ABM (T-Coil) performance of the WD. In addition to
measuring the absolute signal levels, the A-weighted magnitude of the unintended signal shall also be determined.
In order to assure that the required signal quality is measured, the measurement of the intended signal and the
measurement of the unintended signal must be made at the same location for all measurement positions. In
addition, the RF field strength at each measurement location must be at or below that required for the assigned
category.
Measurements shall not include undesired properties from the WD’s RF field; therefore, use of a coaxial connection
to a base station simulator or non-radiating load may be necessary. However, even then with a coaxial connection
to a base station simulator or non-radiating load there may still be RF leakage from the WD, which may interfere
with the desired measurement. Pre-measurement checks should be made to avoid this possibility. All
measurements shall be done with the WD operating on battery power with an appropriate normal speech audio
signal input level given in Table 7.1. If the device display can be turned off during a phone call then that may be
done during the measurement as well.
Measurements shall be performed at two locations specified in A.3, with the correct probe orientation for a
particular location, in a multistage sequence by first measuring the field intensity of the desired T-Coil signal
(ABM1) that is useful to a hearing aid T-Coil. The undesired magnetic components (ABM2) must be measured at
the same location as the desired ABM or T-Coil signal (ABM1), and the ratio of desired to undesired ABM signals
must be calculated. For the perpendicular field location, only the ABM1 frequency response shall be determined in
a third measurement stage. The flow chart in Figure 7.3 illustrates this three-stage, two orientation process.
                                                                              1            2
The following steps summarize the basic test flow for determining ABM1 and ABM2 . These steps assume that a
sine wave or narrowband 1/3 octave signal can be used for the measurement of ABM1.
      a. A validation of the test setup and instrumentation may be performed using a TMFS or Helmholtz coil.
         Measure the emissions and confirm that they are within the specified tolerance.
      b. Position the WD in the test setup and connect the WD RF connector to a base station simulator or a non-
         radiating load as shown in Figure 7.1 or Figure 7.2. Confirm that equipment that requires calibration has
         been calibrated, and that the noise level meets the requirements given in 7.3.1.
      c.   The drive level to the WD is set such that the reference input level specified in Table 7.1 is input to the
           base station simulator (or manufacturer’s test mode equivalent) in the 1 kHz, 1/3 octave band. This drive
           level shall be used for the T-Coil signal test (ABM1) at f = 1 kHz. Either a sine wave at 1025 Hz or a voice-
           like signal, band-limited to the 1 kHz 1/3 octave, as defined in 7.4.2, shall be used for the reference audio
           signal. If interference is found at 1025 Hz an alternative nearby reference audio signal frequency may be
           used.46 The same drive level will be used for the ABM1 frequency response measurements at each 1/3
           octave band center frequency. The WD volume control may be set at any level up to maximum, provided
           that a signal at any frequency at maximum modulation would not result in clipping or signal overload.
      d. Determine the magnetic measurement locations for the WD device (A.3), if not already specified by the
         manufacturer, as described in 7.4.4.1.1 and 7.4.4.2.




1
    Audio Band Magnetic signal - desired (ABM1): Measured quantity of the desired magnetic signal
2
  Audio Band Magnetic signal - undesired (ABM2): Measured quantity of the undesired magnetic signal, such as interference
from battery current and similar non-signal elements.

                                                    Page 7 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                         Issue Date: 9/10/2015
    e. At each measurement location, measure and record the desired T-Coil magnetic signals (ABM1 at fi) as
       described in 7.4.4.2 in each individual ISO 266-1975 R10 standard 1/3 octave band. The desired audio
       band input frequency (fi) shall be centered in each 1/3 octave band maintaining the same drive level as
       determined in item c) and the reading taken for that band.
                 Equivalent methods of determining the frequency response may also be employed, such as fast
                 Fourier transform (FFT) analysis using noise excitation or input–output comparison using simulated
                 speech. The full-band integrated or half-band integrated probe output, as specified in D.9, may be
                 used, as long as the appropriate calibration curve is applied to the measured result, so as to yield
                 an accurate measurement of the field magnitude. (The resulting measurement shall be an accurate
                 measurement in dB A/m.)
                 All measurements of the desired signal shall be shown to be of the desired signal and not of an
                 undesired signal. This may be shown by turning the desired signal ON and OFF with the probe
                 measuring the same location. If the scanning method is used the scans shall show that all
                 measurement points selected for the ABM1 measurement meet the ambient and test system noise
                 criteria in 7.3.1.
    f.   At the measurement location for each orientation, measure and record the undesired broadband audio
         magnetic signal (ABM2) as specified in 7.4.4.4 with no audio signal applied (or digital zero applied, if
         appropriate) using A-weighting and the half-band integrator. Calculate the ratio of the desired to undesired
         signal strength (i.e., signal quality).
    g. Obtain the data from the postprocessor, SEMCAD, and determine the category that properly classifies the
       signal quality based on Table 8.5.

                                                    Test Setup Diagram

                                         USB
                    USB
                                                                                                  AMMI
                                            Audio   Coil    Coil                          Probe
                PC                           Out    Out      In                             In




                                                                                                  EUT




                                    Audio In/Out
                                                                         Coil Monitor
                                                                                        AMCC

                               Base Station
                                                                           Coil In



              ----- For Cal Gain




                                                    Page 8 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                        Issue Date: 9/10/2015

6. Audio Level and Gain Measurements
Establish a call and obtain the 1 kHz speech output signal at 3.14 dBm0 from the CMU200 using the Network
Bitstream “Decoder Cal.” Perform a measurement job to measure the RMS Coil Signal. Note the voltage level and
calculate the desired level for GSM/W-CDMA and CDMA, which are, according to ANSI C63.19-2011 Section
7.4.2.1, -16 dBm0 and -18 dBm0 respectively. To determine the appropriate audio level, use the following
equation, where x dBm0 is the measured RMS Coil Signal:

W-CDMA/GSM
No correction gain factors were measured for W-CDMA/GSM due to the Rohde & Schwarz CMW500, Serial No.
125236, hosting a calibrated audio board. The gains used to measure W-CDMA/GSM are set to 100.




                                                    Page 9 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                        Issue Date: 9/10/2015

7. T-coil Measurement Criteria
7.1. Frequency Response
The frequency response of the axial component of the magnetic field, measured in 1/3 octave bands, shall follow
the response curve, over the frequency range 300 Hz to 3000 Hz.
Figure 8.1 and Figure 8.2 provide the boundaries for the specified frequency. These response curves are for true
field strength measurements of the T-Coil signal. Thus the 6 dB/octave probe response has been corrected from
the raw readings.




                                                   Page 10 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                        Issue Date: 9/10/2015

7.2. Signal to Noise
This specifies the signal-to-noise quality requirement for the intended T-Coil signal from a WD. The worst signal to
noise of the two T-Coil signal measurements, as determined in Clause 7, shall be used to determine the T-Coil
mode category per Table 8.5.
Only the RF immunity of the hearing aid is measured in T-Coil mode. It is assumed that a hearing aid can have no
immunity to an interference signal in the audio band, which is the intended reception band for this mode. So, the
only criterion that can be measured is the RF immunity in T-Coil Mode. This is measured using the same procedure
as for the audio coupling mode and at the same levels as specified in 6.4.




                Measurement locations and reference plane to be used for the T-coil measurements




                                                   Page 11 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                                        Issue Date: 9/10/2015

8. Device Under Test
 Normal operation             Held to head
 Back Cover                   ☐ Normal Battery Cover
                              ☒ Normal Battery Cover with NFC
                              ☐ Wireless Charger Battery Cover
                              ☐ Wireless Charger Battery Cover with NFC
                              ☐ The rechargeable battery is not user accessible.


8.1.          Air Interfaces and Operating Mode
                                                                                                     Voice over digital
     Air-             Bands                Type          HAC                                                                Power
                                                                     Simultaneous but not Tested      Transport OTT
  Interface           (MHz)              Transport      Tested                                                             Reduction
                                                                                                        Capability

                       850
                                             VO          Yes               Yes – Wi-Fi/BT                   NA                NA
   GSM                1900

                   GPRS/EDGE                 DT           No               Yes – Wi-Fi/BT                  Yes                NA

                       850
                                             VO          Yes               Yes - Wi-Fi/BT                   NA                NA
 W-CDMA               1900

                      HSPA                   DT           No               Yes – Wi-Fi/BT                  Yes                NA

                       700

                       850

 LTE - FDD            1700                   VD1          No               Yes – Wi-Fi/BT                  Yes                NA

                      1900

                    2300 (B30)

                      2450

                      5200

   Wi-Fi              5300                   VD1          No              Yes – WWAN/BT                     NA                NA

                      5500

                      5800

    BT                2450                   DT           No             Yes – WWAN/Wi-Fi                   NA                NA
   Type Transport                                                    Note:
   VO=CMRS Voice Service                                             1. No Associated T-Coil measurement has been made in
   DT – Digital Transport                                                accordance with 285076 D02 T-Coil testing for CMRS IP.
   VD=CMRS IP Voice Service and Digital Transport




                                                   Page 12 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                             Issue Date: 9/10/2015

9. HAC (T-coil) Test Results
                                                                             ABM1
                              Channel and        Probe          Frequency              BWC Factor   ABM SNR               Plots
           Mode                                                             ≥ -16 dB                          T-Rating
                               Frequency       Orientation      Response                 (dB)         (dB)               page #
                                                                              (A/m)

          GSM850                                 z (axial)                   12.77        0.16       48.21      T4         1
                                  190
        Voice Coder                                               Pass
                               836.6 MHz
      Speechcodec Low                         y (Transversal)                -5.97        0.16       40.31      T4         2

         GSM1900                                 z (axial)                   13.15        0.16       55.63      T4         3
                                 661
        Voice Coder                                               Pass
                               1880 MHz
      Speechcodec Low                         y (Transversal)                -1.76        0.16       46.92      T4         4

       W-CDMA Band V                             z (axial)                   11.52        0.16       54.57      T4         5
                                 4183
        Voice Coder                                               Pass
                               836.6 MHz
      Speechcodec Low                         y (Transversal)                1.83         0.16       51.70      T4         6

       W-CDMA Band II                            z (axial)                   8.98         0.16       53.95      T4         7
                                 9400
        Voice Coder                                               Pass
                               1880 MHz
      Speechcodec Low                         y (Transversal)                1.70         0.16       51.48      T4         8



Note:
The radial longitudinal (x axis) measurements are no longer required per ANSI C63.19-2011




                                                   Page 13 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                        Issue Date: 9/10/2015

9.1.    Worst Case T-Coil Test Plot
Test Laboratory: UL Verification Services Inc., SAR Lab 2                                Date: 8/4/2015

GSM 850
Communication System: UID 0, GPRS-FDD (TDMA, GMSK, 1 slot) (0); Frequency: 836.6 MHz; Duty Cycle: 1:8.00018
Phantom section: TCoil Section
DASY5 Configuration:
- Probe: AM1DV3 - 3083; ; Calibrated: 1/15/2015
- Sensor-Surface: 0mm (Fix Surface)
- Electronics: DAE4 Sn1259; Calibrated: 1/14/2015
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

T-Coil scan (scan for ANSI C63.19-2007 & 2011 compliance)/General Scans ch 190/y
(transversal) 4.2mm 50 x 50/ABM Interpolated SNR(x,y,z) (121x121x1): Interpolated grid:
dx=1.000 mm, dy=1.000 mm
Signal Type: Audio File (.wav) 48k_voice_1kHz_1s.wav
Output Gain: 100
Measure Window Start: 300ms
Measure Window Length: 1000ms
BWC applied: 0.16 dB
Device Reference Point: 0, 0, -6.3 mm

Cursor:
ABM1/ABM2 = 40.31 dB
ABM1 comp = -5.97 dBA/m
BWC Factor = 0.16 dB
Location: -8.3, -8.3, 3.7 mm




                       0 dB = 1.000 = 0.00 dB




                                                   Page 14 of 15
UL Verification Services Inc.                                                                             Doc. No.: 1.0
           This report shall not be reproduced without the written approval of UL Verification Services Inc.


Report No.: 15I21442-S2V1                                                                        Issue Date: 9/10/2015

Appendix
Refer to separated files for the following appendixes

15I21442-S2V1 HAC_T_App A Setup Photo

15I21442-S2V1 HAC_T_App B Frequency Response & SNR Test Plots

15I21442-S2V1 HAC_T_App C Probe Cal. Certificates


                                                 END OF REPORT




                                                   Page 15 of 15