TCoil Test Report

FCC ID: ZNFAN160

Test Report

Download: PDF
FCCID_1853853

                                                                                                                            o
                                       HEARING AID COMPATIBILITY
Applicant Name:                                                               Date of Testing:
LG Electronics MobileComm U.S.A                                               11/15/2012
1000 Sylvan Avenue                                                            Test Site/Location:
Englewood Cliffs, NJ 07632                                                    PCTEST Lab, Columbia, MD, USA
United States                                                                 Test Report Serial No.:
                                                                              OY1211121642.Z2NF


[FCC ID:                                           ZNFAN160
APPLICANT:                                         LG ELECTRONICS MOBILECOMM U.S.A


Scope of Test:                                     Audio Band Magnetic Testing (T—Coil)
Application Type:                                  Certification
FCC Rule Part(s):                                  CFR §20.19(b)
HAC Standard:                                      ANSI ©63.19—2007 §6.3(v), §7.3(v)
EUT Type:                                          Portable Handset
Model(s):                                          LG—AN160, AN160, LGAN160, LG236C
Tx Frequencies Tested:                             824.70 — 848.31 MHz (Cellular CDMA)
                                                   1851.25 — 1908.75 MHz (PCS CDMA)
Test Device Serial No.:                            Pre—Production Sample [S/N: HAC #2]



C63.19—2007 HAC Category:                          T4 (SIGNAL TO NOISE CATEGORY)

This wireless portable device has been shown to be hearing—aid compatible under the above rated category, specified in ANSVIEEE
Std. ©63.19—2007 and had been tested in accordance with the specified measurement procedures. Hearing—Aid Compatibility is
based on the assumption thatall production units will be designed electrically identical to the device tested in this report. Test results
reported herein relate only to the item(s) tested. For North American bands only.

I attest to the accuracy of data. All measurements reported herein were performed by me or were made under my supervision and
are correct to the best of my knowledge and belief. | assume full responsibility for the completeness of these measurements and
vouch for the qualifications of all persons taking them.

PCTEST certifies that no party to this application has been subject to a denial of Federal benefits that includes FCC benefits
pursuant to Section 5301 of the Anti—Drug Abuse Act of 1988, 21 U.S.C. 862.




      Randy Ortaner
      President                                                                                                         ACCREDITED'
                                                                                                                       Testio Lasoratory
                                                                                                                       ceanricare azoa1.01




FCC ID: ZNFAN160                                            HAC (T—COIL) TEST REPORT                       @uc            Reviewed by:
                                                                                                                          Quality Manager
Filename:                     Test Dates:                      EUT Type:
OY1211121642.2NF              11952012                         Portable Handset                                           Page 1 of 37
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                        REV 6.4C
                                                                                                                                   04/05/10


                                   T A B L E                   O F          C O N T E N T S




1.          INTRODUCTION ............................................................................................................................. 3

2.          TEST SITE LOCATION ................................................................................................................... 4

3.          EUT DESCRIPTION........................................................................................................................ 5

4.          ANSI C63.19-2007 PERFORMANCE CATEGORIES .................................................................... 6

5.          METHOD OF MEASUREMENT ...................................................................................................... 8

6.          TEST SUMMARY .......................................................................................................................... 18

7.          FCC 3G MEASUREMENTS .......................................................................................................... 22

8.          MEASUREMENT UNCERTAINTY ................................................................................................ 23

9.          EQUIPMENT LIST......................................................................................................................... 24

10.         CALIBRATION CERTIFICATES.................................................................................................... 25

11.         CONCLUSION............................................................................................................................... 32

12.         REFERENCES .............................................................................................................................. 33

13.         TEST SETUP PHOTOGRAPHS ................................................................................................... 35




                                                                                                                                      Reviewed by:
FCC ID: ZNFAN160                                                   HAC (T-COIL) TEST REPORT
                                                                                                                                      Quality Manager
Filename:                        Test Dates:                         EUT Type:
                                                                                                                                      Page 2 of 37
0Y1211121642.ZNF            11/15/2012                               Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                                    REV 6.4C
                                                                                                                                               04/05/10


1.          INTRODUCTION

On July 10, 2003, the Federal Communications Commission (FCC) adopted new rules requiring wireless
manufacturers and service providers to provide digital wireless phones that are compatible with hearing
aids. The FCC has modified the exemption for wireless phones under the Hearing Aid Compatibility Act of
1998 (HAC Act) in WT Docket 01-309 RM-86581 to extend the benefits of wireless telecommunications to
individuals with hearing disabilities. These benefits encompass business, social and emergency
communications, which increase the value of the wireless network for everyone. Approximately 500
million people worldwide and 30 million people in the United States suffer from hearing loss.


Compatibility Tests Involved:

The standard calls for wireless communications devices to be measured for:
    RF Electric-field emissions
    RF Magnetic-field emissions
    T-coil mode, magnetic-signal strength in the audio band
    T-coil mode, magnetic-signal frequency response through the audio band
    T-coil mode, magnetic-signal and noise articulation index

The hearing aid must be measured for:
    RF immunity in microphone mode
    RF immunity in T-coil mode

In the following tests and results, this report includes the evaluation for a wireless communications
device.




                                             Figure 1-1 Hearing Aid in-vitu




1
    FCC Rule & Order, WT Docket 01-309 RM-8658

                                                                                                Reviewed by:
FCC ID: ZNFAN160                                      HAC (T-COIL) TEST REPORT
                                                                                                Quality Manager
Filename:                     Test Dates:              EUT Type:
                                                                                                Page 3 of 37
0Y1211121642.ZNF            11/15/2012                 Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                              REV 6.4C
                                                                                                         04/05/10


2.          TEST SITE LOCATION

I.      Introduction
The map at the right shows the location of the PCTEST
LABORATORY in Columbia, Maryland. It is in proximity to the FCC
Laboratory, the Baltimore-Washington International (BWI) airport,
the city of Baltimore and Washington, DC (See Figure 2-1).

These measurement tests were conducted at the PCTEST
Engineering Laboratory, Inc. facility in New Concept Business Park,
Guilford Industrial Park, Columbia, Maryland. The site address is
6660-B Dobbin Road, Columbia, MD 21045. The test site is one of
the highest points in the Columbia area with an elevation of 390
feet above mean sea level. The site coordinates are 39 11’15” N
latitude and 76 49’ 38” W longitude. The facility is 1.5 miles north
of the FCC laboratory, and the ambient signal and ambient signal
strength are approximately equal to those of the FCC laboratory.
There are no FM or TV transmitters within 15 miles of the site. The                       Figure 2-1
detailed description of the measurement facility was found to be in              Map of the Greater Baltimore
compliance with the requirements of § 2.948 according to ANSI                   and Metropolitan Washington,
C63.4-2003 on January 27, 2006 and Industry Canada.                                       D.C. Area

II.     Test Facility / Accreditations:

Measurements were performed at an independent accredited PCTEST Engineering Lab located in
Columbia, MD 21045, U.S.A.
                           PCTEST Lab is accredited to ISO 17025-2005 by the American Association for
                            Laboratory Accreditation (A2LA) in Specific Absorption Rate (SAR) testing, Hearing-
                            Aid Compatibility (HAC), CTIA Test Plans, and wireless testing for FCC and Industry
                            Canada Rules.
                           PCTEST Lab is accredited to ISO 17025 by U.S. National Institute of Standards and
                            Technology (NIST) under the National Voluntary Laboratory Accreditation Program
                            (NVLAP Lab code: 100431-0) in EMC, FCC and Telecommunications.
                           PCTEST facility is an FCC registered (PCTEST Reg. No. 90864) test facility with the
                            site description report on file and has met all the requirements specified in Section
                            2.948 of the FCC Rules and Industry Canada (IC-2451).
                           PCTEST Lab is a recognized U.S. Conformity Assessment Body (CAB) in EMC and
                            R&TTE (n.b. 0982) under the U.S.-EU Mutual Recognition Agreement (MRA).
                           PCTEST TCB is a Telecommunication Certification Body (TCB) accredited to
                            ISO/IEC Guide 65 by the American National Standards Institute (ANSI) in all scopes
                            of FCC Rules and all Industry Canada Standards (RSS).
                           PCTEST facility is an IC registered (IC-2451) test laboratory with the site description
                            on file at Industry Canada.
                           PCTEST is a CTIA Authorized Test Laboratory (CATL) for AMPS and CDMA, and
                            EvDO mobile phones.
                           PCTEST is a CTIA Authorized Test Laboratory (CATL) for Over-the-Air (OTA)
                            Antenna Performance testing for AMPS, CDMA, GSM, GPRS, EGPRS, UMTS (W-
                            CDMA), CDMA 1xEVDO Data, CDMA 1xRTT Data.



                                                                                                      Reviewed by:
FCC ID: ZNFAN160                                     HAC (T-COIL) TEST REPORT
                                                                                                      Quality Manager
Filename:                     Test Dates:              EUT Type:
                                                                                                      Page 4 of 37
0Y1211121642.ZNF            11/15/2012                 Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                  REV 6.4C
                                                                                                             04/05/10


3.          EUT DESCRIPTION




FCC ID:                                             ZNFAN160
Applicant:                                          LG Electronics MobileComm U.S.A
                                                    1000 Sylvan Avenue
                                                    Englewood Cliffs, NJ 07632
                                                    United States
Model(s):                                           LG-AN160, AN160, LGAN160, LG236C
Serial Number:                                      HAC #2
Tx Frequencies Tested:                              824.70 - 848.31 MHz (Cellular CDMA)
                                                    1851.25 - 1908.75 MHz (PCS CDMA)
HW Version:                                         Revision B
SW Version:                                         AN160C01
Maximum Tested Conducted
                                                    24.80 dBm (Cell. CDMA), 25.08 dBm (PCS CDMA)
Power (HAC):
Antenna:                                            Internal Antenna
HAC Test Configurations:                            Cell. CDMA, 1013, 384, 777, BT Off
                                                    PCS CDMA, 25, 600, 1175, BT Off
EUT Type:                                           Portable Handset

                                                                                                     Reduced       Voice Over Digital
                       Band                               C63.19      Simultaneous Transmissions
  Air‐Interface                           Type                                                        power            Transport
                      (MHz)                               tested          (Not to be tested)
                                                                                                    20.19 (c)(1)         (Data)
                       850
     CDMA                                Voice             Yes                  Yes: BT                N/A                N/A
                       1900
      BT               2450               Data             N/A                 Yes: CDMA               N/A                N/A
     NOTE: HAC Rating was not based on concurrent voice and data modes. Standalone mode was found to represent worst case rating.
                                             Table 3: ZNFAN160 Air Interfaces




                                                                                                                                 Reviewed by:
FCC ID: ZNFAN160                                                 HAC (T-COIL) TEST REPORT
                                                                                                                                 Quality Manager
Filename:                      Test Dates:                         EUT Type:
                                                                                                                                 Page 5 of 37
0Y1211121642.ZNF            11/15/2012                             Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                              REV 6.4C
                                                                                                                                         04/05/10


4.          ANSI C63.19-2007 PERFORMANCE CATEGORIES

I.      MAGNETIC COUPLING

Axial and Radial Field Intensity

     All orientations of the magnetic field, in the axial, horizontal and vertical position along the
     measurement plane shall be ≥ -18 dB(A/m) at 1 kHz in a 1/3 octave band filter per 7.3.1.
Frequency Response

     The frequency response of the axial component of the magnetic field shall follow the response curve
     specified in EIA RS-504-1983, over the frequency range 300 Hz – 3000 Hz per 7.3.2.

                                                                                  MAGNETIC FIELD STRENGTH RELATIVE TO 1000 Hz
                                                                        20
                          dB Relative to Value at 1




                                                                                                     -4 dB/Oct .

                                                                        10
                                                                                                                                     +2 dB

                                                                                        -2 dB
                                    kHz




                                                                            0
                                                                                                                                                    -6 dB/ Oct.


                                                                       -10
                                                                                       6 dB/ Oct.


                                                                       -20
                                                                          300                       500                 1000           2000                       3000
                                                                                                                   FREQUENCY (Hz)

                                                    Figure 4-1
                    Magnetic field frequency response for Wireless Devices with an axial field
                                             ≤ -15 dB (A/m) at 1 kHz

                                                                                       M AGNETIC FIELD S TRENGTH RELATIV E TO 1 0 0 0 Hz
                                                                      20

                                                                                                 -4 d B /O ct.
                                      dB Relative to Value at 1 kHz




                                                                      10
                                                                                                                                      +2 dB


                                                                                        -2 d B
                                                                       0
                                                                                                                                                -7 d B /O ct.


                                                                      -10
                                                                                     6 d B /O ct.
                                                                                                                                           -1 2
                                                                                                                                           d B /O ct.
                                                                      -20
                                                                            300                     500                  1000         2000                        3000
                                                                                                                   FREQUENCY (Hz)

                                                 Figure 4-2
            Magnetic Field frequency response for wireless devices with an axial field that exceeds
                                           -15 dB(A/m) at 1 kHz
                                                                                                                                                                         Reviewed by:
FCC ID: ZNFAN160                                                                                          HAC (T-COIL) TEST REPORT
                                                                                                                                                                         Quality Manager
Filename:                                          Test Dates:                                               EUT Type:
                                                                                                                                                                         Page 6 of 37
0Y1211121642.ZNF            11/15/2012                                                                       Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                                                                     REV 6.4C
                                                                                                                                                                                04/05/10


Signal Quality

     The table below provides the signal quality requirement for the intended audio magnetic signal from a
     wireless device. Only the RF immunity of the hearing aid is measured in T-coil mode. It is assumed
     that a hearing aid can have no immunity to an interference signal in the audio band, which is the
     intended reception band for this mode. The only criterion that can be measured is the RF immunity in
     T-coil mode. This is measured using the same procedure as the audio coupling mode at the same
     levels.

     The signal quality of the axial and radial components of the magnetic field was used to determine the
     T-coil mode category.


                                                          Telephone RF Parameters
                                     Category
                                                        Wireless Device Signal Quality
                                                     (Signal + Noise-to-noise ratio in dB)
                                            T1                     0 to 10 dB
                                            T2                   10 to 20 dB
                                            T3                    20 to 30 dB
                                            T4                      > 30 dB

                                                           Table 4-1
                                                 Magnetic Coupling Parameters




                                                                                              Reviewed by:
FCC ID: ZNFAN160                                          HAC (T-COIL) TEST REPORT
                                                                                              Quality Manager
Filename:                     Test Dates:                   EUT Type:
                                                                                              Page 7 of 37
0Y1211121642.ZNF            11/15/2012                      Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                          REV 6.4C
                                                                                                     04/05/10


           METHOD OF MEASUREMENT


1.       Test Setup
     The equipment was connected as shown in an acoustic/RF hemi—anechoic chamber:




                                                              __________      _______________________




                                                                                          7/1/1, i
                                                                                                        I


                                                                                                                 oDe   sensor

                                                                                                                                                   Y¥

                                                                                                                                        D   Probe Amplifier


                              Sine tone, P.50, WN




                                                   Figure 5—1 Validation Setup with Helmholtz Coil




                                                                                                                                                    +
                                                                                                                                        %       Amplifier



                                                         /RF
                                                                                                                :)       Probe Sensor

                                                                          "@e7...


                                5.            &        —I ns                  E
                                     l   lc            -l-_-ll-

                                                  ~|   ocm
                                                        7  *J
                                                           i lklle
                                                                                    M
                                                                                    «l
                                                                  -*.J.



                                                                                    Sine Tone, P.50




                                                                                    Figure 5—2 T—Coil Test Setup




                                                                          >                                                                                    Reviewed by:
FCC ID: ZNFAN160                     LCTEST                                                         HAC (T—COIL) TEST REPORT                            @ 1c       _       *
                                     ee t                                                                                                                      Quality Manager
FiliIename :                 Test
                               est Dates:
                                    Dates                                                                   EUT Type:
                                                                                                                  ype                                          Page 8 of 37
OY1211121642.ZNF             11/15/2012                                                                     Portable Handset
 © 2012 PCTEST Engineering Laboratory, Inc.                                                                                                                            REV 6.4C
                                                                                                                                                                        04/05/10


II.          Scanning Mechanism

Manufacturer:                               TEM
Accuracy:                                   +0.83 cm/meter
Minimum Step Size:                          0.1 mm
Maximum speed                               6.1 cmisec
Line Voltage:                               115 VAC
Line Frequency:                             60 Hz
Material Composite:                         Delrin (Acetal)
Data Control:                               Parallel Port                                                Figure 5—3
Dynamic Range (X—Y—Z2):                     45x 31.75 x 47 cm                                       RF Near—Field Scanner
Dimensions:                                 36" x 25" x 38"
Operating Area:                             36" x 49" x 55"
Reflections:                                < —20 dB (in anechoic chamber)



III.         ITU—T P.50 Artificial Voice
Manufacturer:                      ITU—T
Active Frequency
Range:                             100 Hz — 8 kHz
Stimulus Type:                     Male and Female, no spaces
Single Sample Duration:            20.96 seconds
Activity Level:                    100%


                                                                                                         Figure 5—4
                                                                                           Spectral Characteristic of full P.50
            se   on      on    n     en      en   na   aa     aa     en    on    on   e   en   en   en   are   rea   sn     men    on
        s
       »
       as




                                                          Figure 5—5
                                              Temporal Characteristic of full P.50




FCC ID: ZNFAN160                                              HAC (T—COIL) TEST REPORT                   @uc              Reviewed by:
                                                                                                                          Quality Manager
Filename:                     Test Dates:                          EUT Type:
OY1211121642.2NF              11952012                             Portable Handset                                       Page 9 of 37
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                        REV 6.4C
                                                                                                                                   04/05/10


 ABM1 Measurement Block Diagram:

                  Raw Probe
                 Measurement
                    (mV)
                                                                 T-Coil Probe                            1 kHz bin,
 1kHz through                   40                                                      40 dB
                                                              Sensitivity Factor                         1/3-octave                                ABM1
     WD                         dB                                                  pre-gain factor
                                                              (-59.09 dBV/A/m)                              filter




 ABM2 Measurement Block Diagram:
                  Raw Probe
                 Measurement
                    (mV)
                                                T-Coil Probe            40 dB
                               40                                                       Half-Band           IEC179          100Hz-10kHz
      noise                                  Sensitivity Factor        pre-gain                                                                    ABM2
                               dB                                                    Integrator curve   A-weighting curve    Power Sum
                                             (-59.09 dBV/A/m)           factor



                                   Figure 5-6 Magnetic Measurement Processing Steps

IV.           Test Procedure

       1. Ambient Noise Check per C63.19 §6.2.1
             a. Ambient interference was monitored using a Real-Time Analyzer between 100-10,000 Hz
                 with 1/3 octave filtering.
             b. “A-weighting” and Half-Band Integration was applied to the measurements.
             c. Since this measurement was measured in the same method as ABM2 measurements,
                 this level was verified to be less than 10 dB below the lowest measurement signal (which
                 is the highest ABM2 measurement for a T4 WD). Therefore the maximum noise level for
                 a T4 WD with an ABM1 = -18 dBA/m is:
                                            -18 - 30 - 10 = -58 dBA/m

       2. Measurement System Validation (See Figure 5-1)
             a. The measurement system including the probe, pre-amplifier and acquisition system were
                validated as an entire system to ensure the reliability of test measurements.

                  b. ABM1 Validation
                     The magnetic field at the center of the Helmholtz coil is given by the equation (per
                     C63.19 Annex D.9.1):
                                                                                         V
                                                                                       N( )
                                                                             NI           R
                                                                      Hc           
                                                                                  3
                                                                           r 1.25     r 1.25 3
                                Where Hc = magnetic field strength in amperes per meter
                                      N = number of turns per coil

                        For the Helmholtz Coil, N=20; r=0.08m; R=10.193Ω and using V=57mV:

                                                                             0.057
                                                                           20  (  )
                                                                  Hc       10.193  1.0003 A / m
                                                                       0.08  1.25 3
                        Therefore a pure tone of 1kHz was applied into the coils such that 57 mV was observed
                        across the 10 Ω resistor. The voltmeter used for measurement was verified to be capable
                        of measurements in the audio band range. This theoretically generates an expected field
                        of 1 A/m in the center of the Helmholtz coil which was used to validate the probe


                                                                                                                                          Reviewed by:
FCC ID: ZNFAN160                                                     HAC (T-COIL) TEST REPORT
                                                                                                                                          Quality Manager
Filename:                      Test Dates:                              EUT Type:
                                                                                                                                          Page 10 of 37
0Y1211121642.ZNF            11/15/2012                                  Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                                      REV 6.4C
                                                                                                                                                 04/05/10


                     measurement at 1 A/m. This was verified to be within ± 0.5 dB of the 1 A/m value (see
                     Page 19).

                c.   Frequency Response Validation

                     The frequency response through the Helmholtz Coil was verified to be within 0.5 dB
                     relative to 1 kHz, between 300 – 3000 Hz using the ITU-P.50 artificial speech signal as
                     shown below:




                                       Figure 5-7 Frequency Response Validation

                d. ABM2 Measurement Validation

                     WD noise measurements are filtered with A-weighting and Half-Band Integration over a
                     frequency range of 100Hz – 10kHz to process ABM2 measurements. Below is the
                     verification of the system processing A-weighting and Half-Band integration between
                     system input to output within 0.5 dB of the theoretical result:


                                                               Table 5-1
                                                 ABM2 Frequency Response Validation
                                                       HBI, A -        HBI, A -
                                             f (Hz)   Measured      Theoretical dB Var.
                                                     (dB re 1kHz) (dB re 1kHz)
                                              100      -16.180         -16.170   -0.010
                                              125      -13.257         -13.250   -0.007
                                              160      -10.347         -10.340   -0.007
                                              200       -8.017          -8.010   -0.007
                                              250       -5.925          -5.920   -0.005
                                              315       -4.045          -4.040   -0.005
                                              400       -2.405          -2.400   -0.005
                                              500       -1.212          -1.210   -0.002
                                              630       -0.349          -0.350   0.001
                                              800        0.071           0.070   0.001
                                             1000       0.000           0.000    0.000
                                             1250       -0.503          -0.500   -0.003
                                             1600       -1.513          -1.510   -0.003
                                             2000       -2.778          -2.780   0.002
                                             2500       -4.316          -4.320   0.004
                                             3150       -6.166          -6.170   0.004
                                             4000       -8.322          -8.330   0.008
                                             5000      -10.573         -10.590   0.017
                                             6300      -13.178         -13.200   0.022
                                             8000      -16.241         -16.270   0.029
                                             10000     -19.495         -19.520   0.025


                                                                                                  Reviewed by:
FCC ID: ZNFAN160                                     HAC (T-COIL) TEST REPORT
                                                                                                  Quality Manager
Filename:                     Test Dates:             EUT Type:
                                                                                                  Page 11 of 37
0Y1211121642.ZNF            11/15/2012                Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                              REV 6.4C
                                                                                                         04/05/10


                                                                             ABM2 Frequency Response Validation (LISTEN)

                                                         10




                                                          0
                                                           100                                  1000                                            10000



                                                         -10




                                  dBV, A weighted, HBI
                                                         -20




                                                         -30

                                                                                                              HBI, A - Measured
                                                                                                              HBI, A - Theoretical

                                                         -40




                                                         -50
                                                                                            Frequency (Hz)


                                                                                Figure 5-8
                                                                    ABM2 Frequency Response Validation

                     The ABM2 result is a power sum from 100 Hz to 10 kHz with half-band integration and A-
                     weighting. To verify the power sum measurement, a power sum over the full band was
                     measured and verified to track with the source level (See Figure 5-9). Therefore the
                     setup in this step was used to verify the power sum post-processing for ABM2
                     measurements. See below block diagram:




                                                                          Figure 5-9
                                                                 ABM2 Validation Block Diagram

                     The power summed output results for a known input were compared to the multi-meter
                     results to verify any deviation in the post-processing implemented with the power-sum.

                                                                                  Table 5-2
                                                                           ABM2 Power Sum Validation
                                                                  WN Input      Power Sum              Multimeter-Full
                                                                                                                                     Dev (dB)
                                                                   (dBV)          (dBV)                     (dBV)

                                                                     -60           -60.36                    -60.2                     0.16
                                                                     -50           -50.19                    -50.13                    0.06
                                                                     -40           -40.14                    -40.03                    0.11
                                                                     -30           -30.13                    -30.01                    0.12
                                                                     -20           -20.12                     -20                      0.12
                                                                     -10           -10.14                     -10                      0.14


                                                                                                                                                        Reviewed by:
FCC ID: ZNFAN160                                                             HAC (T-COIL) TEST REPORT
                                                                                                                                                        Quality Manager
Filename:                     Test Dates:                                     EUT Type:
                                                                                                                                                        Page 12 of 37
0Y1211121642.ZNF            11/15/2012                                        Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                                                    REV 6.4C
                                                                                                                                                               04/05/10


                                                                              ABM2 Power Sum Validation (LISTEN)

                                         10                                                                                                            28



                                         —20 —                                                                                  C¥




                                         —30 —                                                             (Q
                                   >
                                   Co
                                   T
                                   0
                                   &
                                   — —40 —                                         ¢
                                   G
                                   O
                                   :73                                                                                   O Power Sum (dBV)
                                   8                                                                                     A Multimeter—Full (dBV)
                                   >
                                         —50 —                   A



                                         —60 Py




                                         '70           I          I      I          I            I          I       I            I                 I    |


                                               —60   —55         —50    —45        —40         —35         —30     —25         —20             —15     —10
                                                                                          WN Input (dBV)


                                                                              Figure 5—10
                                                                       ABM2 Power Sum Validation

      3.   Measurement Test Setup
              ad. Fine scan above the WD (TEM)
                          A multitone signal was applied to the handset such that the phone acoustic
                          output was stable within 1dB over the probe settling time and with the acoustic
                          output level at the C63.19 specified levels (below). The measurement step size
                          was in 2 mm increments at a distance of 10 mm between the surface of the
                          wireless device as shown below:
                                                                                                                 T—Coill Probe




                                                                                                                 T—coil Element
                                                      —a—
                                                     10 mm




                                                     —<4
                                                             (                           Figure 5—11
                                                                              Measurement Distance

                                  After scanning, the planar field maximum point was determined. The position of
                                  the probe was moved to this location to setup the test using the sound check
                                  system.
                                  These steps were repeated for the other T—coil orientations (of axial, radial
                                  transverse, or radial longitudinal) per Figure 5—16 after a T—colil orientation was
                                  fully measured with the sound check system.



FCC ID: ZNFAN160                g ECTEST —                             HAC (T—COIL) TEST REPORT                                         @ 1c                 Reviewed by:
                                                                                                                                                                 _       *
                                Yonsmsacmmean, a                                                                                                             Quality Manager
FiliIename :                 Test
                               est Dates:
                                     Dates                                   EUT Type:
                                                                                   ype                                                                       Page 13 of 37
OY1211121642.ZNF              11/15/2012                                     Portable Handset
 © 2012 PCTEST Engineering Laboratory, Inc.                                                                                                                          REV 6.4C
                                                                                                                                                                      04/05/10


                b. Speech Signal Setup to Base Station Simulator
                       i. C63.19 Table 6-1 states audio reference input levels for various technologies:

                                                                                                        Input Level
                                                 Standard                    Technology
                                                                                                          (dBm0)
                                             TIA/EIA/IS-2000            CDMA                                -18
                                             J-STD-007                  GSM (217 Hz)                        -16
                                             T1/T1P1/3GPP               UMTS (WCDMA)                        -16
                                             iDENTM                     TDMA (22 and 11 Hz)                 -18

                                The CMU200 audio levels were determined using base station simulator
                                manufacturer calibration procedures resulting in the below corresponding
                                voltages relative to handset test point level (in dBm0):

                                                                         Table 5-3
                                                            CMU200 Voltage Input Levels for Audio
                                             dBm0 Ref.              Input Voltage            Notes

                                                                                             From CDMA2K "DECODER CAL".
                                             3.14 dBm0      1052.0 mV              0.4 dBV   (What is needed through Encoder
                                                                                             for FS)
                                              -18 dBm0      92.260 mV          -20.7 dBV     For 8k Enhanced (Low)


                c. Real-Time Analyzer (RTA)
                       i. The Real-Time Analyzer was configured to analyze measurements using 1/3
                          Octave band weighted filtering.
                d. WD Radio Configuration Selection
                       i. The device was chosen to be tested in the worst-case ABM2 condition under
                          RC1/SO3 (EVRC) (see below):




                                                        Figure 5-12
                                               Vocoder Analysis for ABM Noise

     4. Signal Quality Data Analysis
            a. Narrow-band Magnetic Intensity
                     i. The standard specifies a 1 kHz 1/3 octave band minimum field intensity for a sine
                        tone. The ABM1 measurements were evaluated at 1kHz with 1/3 octave band
                        filtering over an averaged period of 10 seconds.
            b. Frequency Response
                     i. The appropriate frequency response curve was measured to curves in Figure 4-1
                        or Figure 4-2 between 300 – 3000 Hz using digital linear averaging (limit lines

                                                                                                                      Reviewed by:
FCC ID: ZNFAN160                                               HAC (T-COIL) TEST REPORT
                                                                                                                      Quality Manager
Filename:                     Test Dates:                       EUT Type:
                                                                                                                      Page 14 of 37
0Y1211121642.ZNF            11/15/2012                          Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                     REV 6.4C
                                                                                                                                04/05/10


                               chosen according to measurement found in step 4a.) A linear average over 3x
                               the length of the artificial voice signal (3x sampling) was performed. A 10 second
                               delay was configured in the measurement process of the stimulus to ensure
                               handset vocoder latency effects and echo cancellation devices (if any) were
                               appropriately stabilized during measurements.
                           ii. The appropriate post-processing was applied according to the system processing
                               chain illustrated in Figure 5-13. All R10 frequencies were plotted with respect to
                               0dB at 1 kHz value and aligned with respect to the EIA-504 mask.




                                   Figure 5-13 Frequency Response Block Diagram

                          iii. The margin is represented by the closest measured data point on the curve to
                               the EIA-504 limit lines, in dB.

                c.   Signal Quality Index
                           i. Ensuring the WD was at maximum RF power, maximum volume, backlight on,
                              display on, maximum contrast setting, keypad lights on (when possible) with no
                              audio signal through the vocoder, the WD was measured over at least 100 Hz –
                              10,000 Hz, maximized over 5 seconds with a 50ms sample time for the ABM2
                              measurement (5 second time period is used in noise measurements under
                              standards such as IEEE 269, etc.)
                          ii. After applying half-band integration and A-weighting to the result, a power sum
                              was applied over each 1/3 octave bandwidth frequency for an ABM2 value
                         iii. This result was subtracted from the ABM1 result in step a, to obtain the Signal
                              Quality.

V.      Test Setup

                                                                                                           Integrator



                                                                                                       1/3 Octave Bandpass
                                                                                                             Filter
                                                                   Probe Coil
                                                                 (Shown in Axial Position)

                                                                                                       RMS Voltmeter



                                                          Wireless Device                                  Graphics
                                                                                                           Recorder



                                                                              Reference
                                                                                                            Signal
                                                                             Base Station
                                             Antenna replaced                                               Source
                                                                              Simulator
                                             with Coaxial Connection




                                                                                                         Controller

                                                         Figure 5-14
                                                Audio Magnetic Field Test Setup

VI.     Deviation from C63.19 Test Procedure

Non-conducted RF connection due to placement of RF port.

                                                                                                                             Reviewed by:
FCC ID: ZNFAN160                                                            HAC (T-COIL) TEST REPORT
                                                                                                                             Quality Manager
Filename:                     Test Dates:                                     EUT Type:
                                                                                                                             Page 15 of 37
0Y1211121642.ZNF            11/15/2012                                        Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                         REV 6.4C
                                                                                                                                    04/05/10


VII. Wireless Device Channels and Frequencies

The frequencies listed in the table below are those that lie in the center of the bands used for cellular
telephony. Low, middle and high channels were tested in each band for FCC compliance evaluation to
ensure the maximum emission is captured across the entire band.

To facilitate setting of a base station simulator for ABM measurements, specific band plan channel
numbers are listed that may be used in lieu of the band center frequencies.

                                                        Table 5-4
                                             Center Channels and Frequencies
                                         Test frequencies & associated channels
                                                                                    Frequency
                                               Channel
                                                                                      (MHz)
                                                      Cellular 850
                                             384 (CDMA)                              836.52
                                                         PCS 1900
                                             600 (CDMA)                               1880

VIII. RF Emission Effect on T-coil Measurements




                                          Figure 5-15
    High power RF Emissions Effect with HAC Dipole on the T-coil Probe System 10mm between
                             dipole maximum and magnetic probe




                                                                                                Reviewed by:
FCC ID: ZNFAN160                                         HAC (T-COIL) TEST REPORT
                                                                                                Quality Manager
Filename:                     Test Dates:                 EUT Type:
                                                                                                Page 16 of 37
0Y1211121642.ZNF            11/15/2012                    Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                            REV 6.4C
                                                                                                       04/05/10


IX.     Test Flow
The flow diagram below was followed (From C63.19):

                                      Confirm calibration of test
                                       equipment
                                      Configure and validate test
                                       setup
                                      Establish WD reference level
                                      Scan for measurement locations




                                     Position and orient probe
                                     Measure desired audio band
                                       Signal strength
                                     Measure undesired audio
                                       Band signal strength



                                      Calculate signal strength
                                      Calculate signal quality
                                      Measure frequency response




                                              All 3 locations
                                               measured?
                                                                          N

                                                         Y



                                                Intensity and
                                             Frequency response            N
                                                 compliant?




                                                         Y


                                       Determine and record signal                   Done
                                            Quality category




                                                       Figure 5-16
                                             C63.19 T-Coil Signal Test Process
                                                                                            Reviewed by:
FCC ID: ZNFAN160                                          HAC (T-COIL) TEST REPORT
                                                                                            Quality Manager
Filename:                     Test Dates:                    EUT Type:
                                                                                            Page 17 of 37
0Y1211121642.ZNF            11/15/2012                       Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                        REV 6.4C
                                                                                                   04/05/10


6.          TEST SUMMARY

I.      T-Coil Test Summary
                                                          Table 6-1
                                                       Table of Results
                                                                                            Minimum
     C63.19 Sec.        Mode         Band                 Test Description                             Measured       Verdict
                                                                                             Limit*
                                                                                            dBA/m       dBA/m       PASS/FAIL
       7.3.1.1                                 Intensity, Axial                               -18        -5.3          PASS
       7.3.1.2                                 Intensity, RadialH                             -18        -13.1         PASS
       7.3.1.2                                 Intensity, RadialV                             -18        -12.2         PASS
        7.3.3          CDMA         Cellular   Signal-to-Noise/Noise, Axial                   20         51.8          PASS
        7.3.3                                  Signal-to-Noise/Noise, RadialH                 20         44.9          PASS
        7.3.3                                  Signal-to-Noise/Noise, RadialV                 20         39.4          PASS
        7.3.2                                  Frequency Response, Axial                      0          1.7           PASS


       7.3.1.1                                 Intensity, Axial                               -18        -5.5          PASS
       7.3.1.2                                 Intensity, RadialH                             -18        -13.2         PASS
       7.3.1.2                                 Intensity, RadialV                             -18        -12.8         PASS
        7.3.3          CDMA           PCS      Signal-to-Noise/Noise, Axial                   20         51.0          PASS
        7.3.3                                  Signal-to-Noise/Noise, RadialH                 20         46.6          PASS
        7.3.3                                  Signal-to-Noise/Noise, RadialV                 20         38.4          PASS
        7.3.2                                  Frequency Response, Axial                      0          1.8           PASS

Note: The above summary table represents the worst-case numerical values according to configurations
in Table 6-3.

                                                        Table 6-2
                                               Consolidated Tabled Results

                                      Volume                         Cellular                             PCS
                                      Setting
                                                       Axial         RadialH      RadialV      Axial     RadialH       RadialV
        Freq. Response Margin                         PASS            PASS        PASS         PASS       PASS          PASS
      Magnetic Intensity Verdict    Maximum           PASS            PASS        PASS         PASS       PASS          PASS
                FCC SNR Verdict                       PASS            PASS        PASS         PASS       PASS          PASS

Note: The above table represents the pass/fail verdict according to data in Table 6-3.




                                                                                                                   Reviewed by:
FCC ID: ZNFAN160                                          HAC (T-COIL) TEST REPORT
                                                                                                                   Quality Manager
Filename:                     Test Dates:                      EUT Type:
                                                                                                                   Page 18 of 37
0Y1211121642.ZNF            11/15/2012                         Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                               REV 6.4C
                                                                                                                          04/05/10


II.      Raw Handset Data
                                                           Table 6-3
                                                        Raw Data Results
                                                                                      Cellular Band
                                 Volume
                                                        Axial                           RadialH                       RadialV
                                               1013      384         777       1013      384        777       1013     384       777
               ABM1, dBA/m                     -5.31    -5.53       -4.96     -12.77    -12.60     -13.07    -12.19   -11.84    -11.70
               ABM2, dBA/m                    -57.22   -57.40      -56.73     -57.77    -57.45     -59.09    -51.59   -52.50    -52.16
       Ambient Noise, dBA/m                   -60.36   -60.36      -60.36     -61.44    -61.44     -61.44    -60.80   -60.80    -60.80
      Freq. Response Margin
                                               1.93     1.95        1.73       1.94       1.89        2.00    2.00     1.98      1.98
                        (dB) Maximum
                 S+N/N (dB)                   51.91    51.87       51.77      45.00      44.85     46.02     39.40     40.66    40.46
       S+N/N per orientation
                                                       51.77                             44.85                         39.40
                       (dB)

                                                                                       PCS Band
                                 Volume
                                                        Axial                           RadialH                       RadialV
                                                 25      600        1175        25       600        1175       25      600       1175
               ABM1, dBA/m                     -5.13    -5.37       -5.46     -12.54    -12.69     -13.19    -12.79   -12.15    -12.07
               ABM2, dBA/m                    -56.10   -56.57      -56.92     -59.20    -59.60     -59.74    -51.14   -51.91    -51.76
       Ambient Noise, dBA/m                   -60.36   -60.36      -60.36     -61.44    -61.44     -61.44    -60.80   -60.80    -60.80
      Freq. Response Margin
                                               1.89     1.82        1.90       1.95       2.00        1.99    2.00     1.95      2.00
                        (dB) Maximum
                 S+N/N (dB)                   50.97    51.20       51.46      46.66      46.91     46.55     38.35     39.76    39.69
       S+N/N per orientation
                                                       50.97                             46.55                         38.35
                       (dB)
                                 [x,y] from
  T-coil Coordinates (cm)                              2.7,2.8                           2.8,3.3                      3.5,2.7
                                bottom left


Note: ABM1 >> Ambient noise

WD Configuration

       1. Radio Configuration: RC1/SO3 (EVRC)
       2. Power Configuration: Power Control Bits = "All Up"
       3. Phone Condition: Mute on; Backlight on; Max Volume, Max Contrast




                                                                                                                         Reviewed by:
FCC ID: ZNFAN160                                               HAC (T-COIL) TEST REPORT
                                                                                                                         Quality Manager
Filename:                      Test Dates:                       EUT Type:
                                                                                                                         Page 19 of 37
0Y1211121642.ZNF            11/15/2012                           Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                      REV 6.4C
                                                                                                                                 04/05/10


III.    Frequency Response Graph




                                                                                                        Figure 6-1
                                                                                               Axial Frequency Response

Note: This frequency response represents the worst-case ABM2 test configuration according to Table
6-3.

IV.     1 kHz Vocoder Application Check

                                                                                                CDMA2000 Pure Tone suppression characterization (example)

                                                                                 30


                                                                                 20


                                                                                 10
                                       Magnetic Intensity at 1kHz, R10 (dBA/m)




                                                                                  0


                                                                                 -10


                                                                                 -20


                                                                                 -30


                                                                                 -40


                                                                                 -50


                                                                                 -60


                                                                                 -70
                                                                                   -50   -45    -40       -35      -30         -25         -20   -15        -10   -5   0
                                                                                                                     Audio input level (dBm0)




This model was verified to be within the linear region for ABM1 measurements. This measurement was
taken in the axial configuration above the ABM1 maximum location/configuration derived from Table 6-3.

V.      Undesirable Audio Magnetic Band Plot (ABM2)




                                                                                                   Figure 6-2
                                                                                          Worst-case ABM2 Plot for WD


                                                                                                                                                                           Reviewed by:
FCC ID: ZNFAN160                                                                                                HAC (T-COIL) TEST REPORT
                                                                                                                                                                           Quality Manager
Filename:                   Test Dates:                                                                           EUT Type:
                                                                                                                                                                           Page 20 of 37
0Y1211121642.ZNF            11/15/2012                                                                            Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                                                                       REV 6.4C
                                                                                                                                                                                  04/05/10


Note: This plot represents the data from the location/configuration resulting in the highest ABM2 result
shown in Table 6-3.

VI.     T-Coil Validation Test Results




                                                     Figure 6-3
                                 Helmholtz Coil Validation for Frequency Response



                                                       Table 6-4
                                       Helmholtz Coil Validation Table of Results

                         Item                             Target                 Result      Verdict

 Signal Validation
 Frequency Response, from limits                        0 ± 0.5 dB                0.30       PASS
 Magnetic Intensity, -10 dBA/m                         -10 ± 0.5 dB              -9.503      PASS
 Noise Validation

 Axial Environmental Noise                            < - 58 dBA/m               -60.36      PASS

 RadialH Environmental Noise                          < - 58 dBA/m               -61.44      PASS

 RadialV Environmental Noise                          < - 58 dBA/m               -60.80      PASS




                                                                                               Reviewed by:
FCC ID: ZNFAN160                                      HAC (T-COIL) TEST REPORT
                                                                                               Quality Manager
Filename:                   Test Dates:                EUT Type:
                                                                                               Page 21 of 37
0Y1211121642.ZNF            11/15/2012                 Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                           REV 6.4C
                                                                                                      04/05/10


7.          FCC 3G MEASUREMENTS

Radio Configuration 1, Service Option 3 (thick, green data curve) was used for the testing as the worst-
case configuration for the handset due to vocoder gating from the EVRC logic. See below plot for ABM
noise comparison between operational field service options and radio configurations for a CDMA2000
handset:




                                                    Figure 7-1
                                         CDMA2000 Audio Band Magnetic Noise

I.       ABM Measurements
                     ABM2 Pre-Test (dBA/m), A, HBI
                        RC1/SO3              RC3/SO3    RC4/SO3            Orientation   Channel

                          -50.96              -59.68      -59.92            RadialV        25

                     ABM1 Pre-Test (dBA/m)


                        RC1/SO3              RC3/SO3    RC4/SO3            Orientation   Channel

                          -11.980            -12.000     -12.220            RadialV        25


           Mute on; Backlight on; Max Volume, Max Contrast
           Power Control Bits = "All Up"




                                                 Figure 7-2
                            Audio Band Magnetic Curve Measurement Block Diagram




                                                                                                   Reviewed by:
FCC ID: ZNFAN160                                       HAC (T-COIL) TEST REPORT
                                                                                                   Quality Manager
Filename:                   Test Dates:                 EUT Type:
                                                                                                   Page 22 of 37
0Y1211121642.ZNF            11/15/2012                  Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                               REV 6.4C
                                                                                                          04/05/10


8.          MEASUREMENT UNCERTAINTY

                                                         Table 8-1
                                                Uncertainty Estimation Table
                                                                                                                      Standard
                                     Data +/- Data +/-                          Probability               Standard
             Contribution                                    Data Type                         Divisor               Uncertainty
                                       %        dB                              distribution             uncertainty
                                                                                                                        (dB)
 ABM Noise                            7.0%       0.29        Std. Dev.          Normal k=1      1.00        7.0%
 RF Reflections                       4.7%       0.20       Specification       Rectangular     1.73        2.7%
 Reference Signal Level              12.2%       0.50       Specification       Rectangular     1.73        7.0%
 Positioning Accuracy                10.0%       0.41       Uncertainty         Rectangular     1.73        5.8%
 Probe Coil Sensitivity              12.2%       0.50       Specification       Rectangular     1.73        7.0%
 Probe Linearity                      2.4%       0.10        Std. Dev.          Normal k=1      1.00        2.4%
 Cable Loss                           2.8%       0.12       Specification       Rectangular     1.73        1.6%
 Frequency Analyzer                   5.0%       0.21       Specification       Rectangular     1.73        2.9%
 System Repeatability                 5.0%       0.21        Std. Dev.          Normal k=1      1.00        5.0%
 WD Repeatability                     9.0%       0.37        Std. Dev.          Normal k=1      1.00        9.0%
 Positioner Accuracy                 1.0%        0.04       Specification       Rectangular     1.73        0.6%

 Combined standard uncertainty, uc (k=1)                                                                   17.7%          0.71
 Expanded uncertainty (k=2), 95% confidence level                                                          35.3%          1.31
             Notes:
            1. Test equipments are calibrated according to techniques outlined in NIS81, NIS3003 and NIST Tech Note 1297.
            2. All equipments have traceability according to NIST. Measurement Uncertainties are defined in further detail in
                NIS 81 and NIST Tech Note 1297 and UKAS M3003.

Measurement uncertainty reflects the quality and accuracy of a measured result as compared to the true
value. Such statements are generally required when stating results of measurements so that it is clear to
the intended audience that the results may differ when reproduced by different facilities. Measurement
results vary due to the measurement uncertainty of the instrumentation, measurement technique, and test
engineer. Most uncertainties are calculated using the tolerances of the instrumentation used in the
measurement, the measurement setup variability, and the technique used in performing the test. While
not generally included, the variability of the equipment under test also figures into the overall
measurement uncertainty. Another component of the overall uncertainty is based on the variability of
repeated measurements (so-called Type A uncertainty). This may mean that the Hearing Aid
compatibility tests may have to be repeated by taking down the test setup and resetting it up so that there
are a statistically significant number of repeat measurements to identify the measurement uncertainty. By
combining the repeat measurement results with that of the instrumentation chain using the technique
contained in NIS 81 and NIS 3003, the overall measurement uncertainty was estimated.




                                                                                                                     Reviewed by:
FCC ID: ZNFAN160                                           HAC (T-COIL) TEST REPORT
                                                                                                                     Quality Manager
Filename:                   Test Dates:                      EUT Type:
                                                                                                                     Page 23 of 37
0Y1211121642.ZNF            11/15/2012                       Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                 REV 6.4C
                                                                                                                            04/05/10


9.          EQUIPMENT LIST

                                                            Table 9-1
                                                          Equipment List
   Manufacturer            Model                        Description                Cal Date     Cal Interval   Cal Due   Serial Number
      Agilent          E4407B                      ESA Spectrum Analyzer            4/3/2012      Annual         4/3/2013 US39210313
      Agilent         E5515C                  Wireless Communications Test Set      2/14/2012     Annual        2/14/2013 GB43304447
      Agilent         E5515C                  Wireless Communications Tester        4/4/2012      Annual         4/4/2013 US41140256
 Control Company     36934-158                   Wall-Mounted Thermometer           1/4/2012      Biennial       1/4/2014  122014497
   Gigatronics         80701A                    (0.05-18GHz) Power Sensor         10/10/2012     Annual       10/10/2013   1833460
   Gigatronics          8651A                       Universal Power Meter          10/10/2012     Annual       10/10/2013   8650319
      Listen        SoundCheck                    Acoustic Analyzer System          10/4/2012     Annual        10/4/2013    979921
      Listen       SoundConnect                   Microphone Power Supply           4/26/2012     Annual        4/26/2013 0899-PS150
         NI              4474                        Data Acquisition Card             N/A                         N/A         N/A
 Rohde & Schwarz      CMU200                        Base Station Simulator          5/22/2012    Annual        5/22/2013     109892
    Seekonk            NC-100                       Torque Wrench (8" lb)          11/29/2011    Triennial     11/29/2014     21053
       TEM       Axial T-Coil Probe                   Axial T-Coil Probe            1/20/2012     Annual       1/20/2013   TEM-1124
       TEM                                              HAC Positioner                 N/A                         N/A         N/A
       TEM                                   HAC System Controller with Software       N/A                         N/A         N/A
       TEM         Helmholtz Coil                       Helmholtz Coil              4/27/2012    Annual         4/27/2013   SBI 1052
       TEM             C63.19                           Helmholtz Coil             11/11/2011    Biennial      11/11/2013      925
       TEM       Radial T-Coil Probe                  Radial T-Coil Probe           1/20/2012    Annual         1/20/2013  TEM-1130




                                                                                                                         Reviewed by:
FCC ID: ZNFAN160                                             HAC (T-COIL) TEST REPORT
                                                                                                                         Quality Manager
Filename:                   Test Dates:                        EUT Type:
                                                                                                                         Page 24 of 37
0Y1211121642.ZNF            11/15/2012                         Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                                                     REV 6.4C
                                                                                                                                04/05/10




                      West Caldwell Calibration Laboratories Inc.




P
222
e
Loe
              Certificate of Calibration
L
                                                                        for

                                                         Axial T Coil Probe
                                              Manufactured by:            TEM CONSULTING
                                              Model No:                   Axial T Coil Probe
                                              Serial No:                  TEM—1124
                                              Calibration Recall No:      21559
                                                                     Submitted By:

                                              Customer:         ~_      STEVE LU
                                              Company:                  PCTEST ENGINEERING LAB
                                              Address:                  6660—B DOBBIN ROAD
                                                                        COLUMBIA              MD 21045


        The subject instrument was calibrated to the indicated specification using standards traceable to the
        National Institute of Standards and Technology or to accepted values of natural physical constants.
        This document certifies that the instrument met the following specification uponits returnto the
        submitter,

        West Caldwell Calibration Laboratories Procedure No.                         Axial T Coi TEM
        Upon receipt for Calibration, the instrument was found to be:

                              Within         CX )       see attached Report of Calibration.

        the tolerance of the indicated specification,

        West Caldwell Calibration Laboratories‘ calibration control system meets the requirements, ISO
          1001241 MIL—STD—45662A, ANSI/NCSL Z540—1, IEC Guide 25, ISO 9001:2008 and ISO 17025.



          Note: With this Certificate, Report of Calibration is included.                  Approved by:
                                                           a"
          Calibration Date:                 20—Jan—12                % 2.                 o              e          _
          Certificate No:                   21559 ~1                                       Felix Christopher
                                                           Certificate Page 1 of 1         Quality Manager
          OA Doc. #1051 Rev. 2.0 101101

                                          West& Caldwell                                isogoor:200s                se
                                                   a                                     egistered  Company
                                                                                                     tn
      w                                    Calibration                                  Aivamiiadiai
      uncompromised calibration             Laboratorles, Inc.                          To5.T                      petne
      1575 State Route 96, Victor, NY 14564, U.S.A.                                                            585) 586—432




                           eereer                                                                              @:ic


                                                                 HCATEMC_TEM—1124_Jan—20—2012

                                                  West Caldwell                                                                                                   ‘"oreemose
                                                  Calibration
        uncompromised calibration                    Laboratories, Inc.                                                                                              cenes
1575 State Route 96, Victor NY 14564                                                                                                                                     1533.01




                                                                                                 for
                      TEM Consulting LP Axial T Coil Probe                                  Model No.: Axial T Coil Probe                              Serial No.: TEM—1124

       Company : Pctest Engineering Lab.                                                                                                                   1. D. No: 80578
Calibration results:                                                                                            Before data: ...                        After data: .........
                   Probe Sensitivity measured with Helmholtz Coil
                                   Helmholtz Coll;                                                                                   Before & after data same: ... X......
                  the number of turns on eachcoil;   10       No.
                              the radius of eachcoil, in meters;        0.204           m                                        Laboratory Environment:
                            the current in the coils, in amperes.;       0.08           A                                   Ambient Temperature:           221           °C
                                       Helmholtz Coll Constant;          6.98           AlmIV                                  Ambient Humidity:           30.8          % RH
                                 Helmholtz Cofl magnetic field;          5.90           Alm                                    Ambient Pressure:          100.1          kPa
                                                                                                                                 Calibration Date:     20—Jan—12          1:01 PM
                                          Probe Sensitivity at           1000           Hz,                                    Re—calibration Due:     20—Jan—13
                                                         was            —60.16          dBV/A/m                                   Report Number:         21559           ~1
                                                                        0.981           mV/A/m                                    Control Number:        21559
                                             Probe resistance            904            Ohms
The above listed instrument meets or exceeds the tested manufacturer‘s specifications.
‘This Calibration is traceable through NIST test numbers:                               205342
The expanded uncertainty of calibration: 0.3048 at 95% confidence level with a coverage factor of k=2.
Graph represents Probes Frequency Response.

                                                                                      Axtal Probe Response
                                                                                                                                         —&—Measured Probe..
                 20
Magnitude (dB)
                 o




                      100                                                       Freq. (Hz)         1000                                                                               10000

The above listed instrument was checked using calibration procedure documented in West Caldwell
Calibration Laboratories Inc. procedure :                                                                                Rev. 5.0 Sept. 10, 2010 Doc. # 1038 HCATEMC
Calibration was performed by West Caldwell Calibration Laboratories Inc. under Operating Procedures
intended to implementthe requirements of 1SO10012—1, IEC Guide 25, ANSI/NCSL 2540—1, (MIL—STD—45662A) and ISO 9001:2008, |37025
                                                                                                                                                                        _
                                     Cal. Date:    20—Jan—2012         1:01 PM                                      Measurements prfOIMEd by: .. .....ce           t Alecscceee e
Calibrated on WCCL system type 9700                                                                                                                  Felix Christopher
This document shall not be reproduced, except in full, without the written approval from West Caldwell Cal, Labs, Inc.                  Rev. 6.0 Sept. 10, 2010 Doc. # 1038 HCATEMC




                                                                                        Page 1 of 2



                                                                                                                                                        @1                      ‘


                                                               HCATEMC_TEM—1124_Jan—20—2012

                                            West Caldwell Calibration Laboratories Inc.
                                                                      1575 State Route 96, Victor NY 14564
                                                                     Tel. (585) 586—3900 FAX (585) 586—4327
                                                         Calibration Data Record
                                                                                                for
         TEM Consulting LP Axial T Coil Probe                                             Model No.: Axial T Coil Probe                            Serial No.: TEM1124

                                                                                          Company : Petest Engineering Lab.

Test                 Function                                                                  Tolerance                                    Measured _values
                                                                                                                                  Before          Out        Remarks

1.0                  Probe Sensitivity at                                      1000 Hz.                      dBV/A/m              »60.16

                                                                                                                     dB
2.0                  Probe Level Linearity                                                                            6            6.03
                                                                                       Ref. (0 dB)                    0            0.00
                                                                                                                     —6            —6.02
                                                                                                                     12           ~12.03

                                                                                                                     Hz
3.0                  Probe Frequency Response                                                                        100          ~19.9
                                                                                                                     126          17.9
                                                                                                                     158          ~15.9
                                                                                                                     200          ~13.9
                                                                                                                     251          11.9
                                                                                                                     316          ~10.0
                                                                                                                     398           —8.0
                                                                                                                     601           5.0
                                                                                                                     631           4.0
                                                                                                                     794           —2.0
                                                                                        Ref. (0 dB)                 1000           0.0
                                                                                                                    1259           2.0
                                                                                                                    1585           4.0
                                                                                                                    1995           6.0
                                                                                                                    2512            7.9
                                                                                                                    3162            9.9
                                                                                                                    3981           11.9
                                                                                                                    5012           13.9
                                                                                                                    6310           15.9
                                                                                                                    7943           18.0
                                                                                                                    10000          20.2




Instruments used for calibration:                                                                                  Date of Cal.                 Traceablity No.              Due Date
        HP                                           34401A                      SIN    US360641                   17—0ct—2011                  ;205342                   16—0ct—2012
        HP                                           34401A                      S/N    US361024                   17—0ct—2011                  ;205342                   16—0ct—2012
        HP                                           33120A                      S/N    S3604371                   17—0ct—2011                  205342                    16—0ct—2012
       B&K                                               2133                    S/N    1492410                     4—Nov—2011                  681/280411—11              4—Nov—2012




                            Cal. Date:           20—Jan—2012           1:01 PM                                                    Tested by: Felix Christopher
Calibrated on WCCL system type 9700
This document shall not be reproduced, except in full, without the written approval from West Caldwell Cal. Labs. Inc.              Rev. 5.0 Sopt. 10, 2010 Doc. # 1038 HGATEMC




                                                                                        Page 2 of 2




                                     eeresr                                                                                                         @:c


                            West Caldwell Calibration Laboratories Inc.




                    Certificate of Calibration
                                                                                  for

                                                             Radial T Coil Probe
                                                    Manufactured by:          TEM CONSULTING                                            x
                                                    Model No:                                Radial T Coil Probe                        74 es
                                                    Serial No:                               TEM—1130                                   oz
                                                    Calibration Recall No:                   21559
                                                                             Submitted By:
                                                    Customer:                    STEVE LIU
                                                    Company:                     PCTEST ENGINEERING LAB
                                                    Address:                     6660—B DOBBIN ROAD
                                                                                 COLUMBIA             MD 21045


Coka           The subject instrument was calibrated to the indicated specification using standards traceable to the
               National Institute of Standards and Technology or to accepted values of natural physical constants,
               This document certifics that the instrument met the following specification uponits return to the
               submitter.

               West Caldwell Calibration Laboratories Procedure No.                               Radial T C TEM
4   5‘?3;:     Upon receipt for Calibration, the instrument was found to be:

                                    Within         (Xx )          see attached Report of Calibration,

               the tolerance of the indicated specification,

               West Caldwell Calibration Laboratories‘ ealibration control system meets the requirements, ISO
               10012—1 MIL—STD—45662A, ANSI/NCSL Z540—1, IEC Guide 25, ISO 9001:2008 and ISO 17025.




               Note: With this Centificate,                alibration     is included.                  Approved by:

               Calibration Date:                  20—Jan—12

               Certificate No:                    21559 ~2                                              Felix Christopher
               QA Dos. #1051 Rev. 2.0 10/1101                           Certificate Page 1 of 1         Quality Manager

                                                                                                     180 2001:2008          repeonpps
                                                West Caldwell
                                                Calibration
                                                   #    s
                                                            &                                        cutmtormaameo
                                                                                                     Registered C
                                                                                                           ‘on Traceablo
             uncompromised calibration            Laboratories, Inc.                                 Tonbsr                 gesseome
                                                                                                                              dlser


                                                                   HCRTEMC_TEM—1130_Jan—20—2012


                                                                                                                                                                ACCREDITATION
                                                   West Caldwell                                                                                                  Boien hss
                                                    Calibration                                                                                                   &
     Uncompromised calibration\                       Laboratories, Inc.                                                                                           rcouebm®
1575 State Route 96, Victor NY 14564                                                                                                                                  1533.01

                                                                                                  T                          7:
                                                      REPORT OF CALIBRATION
                                                                                                for
                       TEM Consulting LP Radial T Coil Probe                               Model No.: Radial T Coil Probe                            Serial No.: TEM—1130

     Company : Petest Engineering Lab.                                                                                                                  1. D. No: 80579

Calibration results:                                                                                           Before data: .........                 After datai .........
                   Probe Sensitivity measured with Helmholtz Coil
                                                   Helmholtz Coil;                                                                Before & after data same: ... X......
                               the number of turns on each coil;          10           No.
                               the radius of eachcoil, in meters;       0.204          m                                       Laboratory Environment:
                             the current in the coils, in amperes.;      0.08          A                                 Ambient Temperature:           22.1          °C
                                        Helmholtz Coil Constant;         6.98          AlmIV                                 Ambient Humidity:          30.8          % RH
                                  Helmholtz Coil magnetic field;         5.90          Alm                                   Ambient Pressure:          100.1         kPa
                                                                                                                               Calibration Date:    20—Jan—12           1:30 PM
                                           Probe Sensitivity at         1000           Hz.                                  Re—calibration Due:     20—Jan—13
                                                          was          —60.50          dBV/A/m                                 Report Number:         21559           —2
                                                                       0.944           mV/Aim                                  Control Number:        21559
                                              Probe resistance           903           Ohms
The above listed instrument meets or exceeds the tested manufacturer‘s specifications.
‘This Calibration is traceable through NIST test numbers:                              205342
The expanded uncertainty of calibration: 0.30dB at 95% confidence level with a coverage factor of k=2.
Graph represents Probes Frequency Response.
                                                                                     Radial Probe Response
                                                                                                                                     —a— Measured Probe Resp.
                  20
 Magnitude (dB)
                  0




                       100                                                      Freq. (Hz)         1000                                                                             10000

The above listed instrument was checked using calibration procedure documented in West Caldwell
Calibration Laboratories Inc. procedure:                                  Rev. 5.0 Sept. 10, 2010 Doc. # 1038 HCRTEMC
Calibration was performed by West Caldwell Calibration Laboratories Inc. under Operating Procedures
intended to implement the requirements of ISO10012—1, IEC Guide 25, ANSI/NCSL 2540—1, (MIL—STD—45662A) and ISO 9001:2008, l?                                      17025


                                      Cal. Date:     20—Jan—2012       1:30 PM                                      Measurements performed by: .... oA              Alcl....c..}
Calibrated on WCCL system type 9700                                                                                                                Felix Christopher
This document shall not be reproduced, except in full, without the written approval from West Caldwell Cal. Labs. Inc.                Rev. $.0 Sopt. 10, 2010 Doc. # 1038 HCRTEMC




                                                                                        Page 1 of 2



                                              eeresr                                                                                                  @:c


                                                               HCRTEMC_TEM—1130_Jan—20—2012

                                            West Caldwell Calibration Laboratories Inc.
                                                                      1575 State Route 96, Victor NY 14564
                                                                     Tel. (585) 586—3900 FAX (585) 586—4327

                                                         Calibration Data Record
                                                                                             for
         TEM Consulting LP Radial T Coil Probe                                            Model No.: Radial T Coil Probe                          Serial No.: TEM—1130

                                                                                          Company : Potest Engineering Lab.

Test                 Function                                                                  Tolerance                                    Measured values
                                                                                                                                  Before          Out       Remarks

1.0                  Probe Sensitivity at                                      1000 Hz.                      dBV/Aim              +60.50

                                                                                                                         dB
2.0                  Probe Level Linearity                                                                                6        6.03
                                                                                       Ref. (0 dB)                        0        0.00
                                                                                                                         —6        +6.02
                                                                                                                         12       ~12.02

                                                                                                                      Hz
3.0                  Probe Frequency Response                                                                        100          20.0
                                                                                                                     126          18.0
                                                                                                                     158          15.9
                                                                                                                     200          13.9
                                                                                                                     251          411.9
                                                                                                                     316          ~10.0
                                                                                                                     398           —8.0
                                                                                                                     501           —6.0
                                                                                                                     631           4.0
                                                                                                                     794           2.0
                                                                                        Ref. (0 dB)                 1000           0.0
                                                                                                                    1259            2.0
                                                                                                                    1585           4.0
                                                                                                                    1995            6.0
                                                                                                                    2512            7.9
                                                                                                                    3162            9.9
                                                                                                                     3981          11.9
                                                                                                                     5012          13.9
                                                                                                                     6310          15.9
                                                                                                                     7943          18.0
                                                                                                                    10000          20.1




Instruments used for calibration:                                                                                  Date of Cal.                 Traceability No.             Due Date
        HP                                           34401A                      SN     US360641                   17—0ct—2011                  ;205342                   16—0ct—2012
        HP                                           34401A                      S/N    US361024                   17—0ct—2011                  ;205342                   16—0ct—2012
        HP                                           33120A                      S/N    $3604371                   17—0ct—2011                  ;205342                   16—0ct—2012
       B&K                                               2133                    S/N    4492410                     4—Nov—2011                  681/280411—11              4—Nov—2012




                            Cal, Date:           20—Jan—2012           1:30 PM                                                    Tested by: Felix Christopher
Calibrated on WCCL system type 9700
This document shall not be reproduced, except In full, without the written approval from West Galdwell Cal, Labs, Inc.              Rov. 5.0 Sept, 10, 2010 Dos. # 1038 HCRTEMC




                                                                                        Page 2 of 2




                                      @\eeresr                                                                                                      @:«


11.         CONCLUSION

The measurements indicate that the wireless communications device complies with the HAC limits
specified in accordance with the ANSI C63.19 Standard and FCC WT Docket No. 01-309 RM-8658.
Precise laboratory measures were taken to assure repeatability of the tests. The tested device complies
with the requirements in respect to all parameters specific to the test. The test results and statements
relate only to the item(s) tested.

The measurement system and techniques presented in this evaluation are proposed in the ANSI standard
as a means of best approximating wireless device compatibility with a hearing-aid. The literature is under
continual re-construction.




                                                                                             Reviewed by:
FCC ID: ZNFAN160                              HAC (T-COIL) TEST REPORT
                                                                                             Quality Manager
Filename:                   Test Dates:         EUT Type:
                                                                                             Page 32 of 37
0Y1211121642.ZNF            11/15/2012          Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                         REV 6.4C
                                                                                                    04/05/10


12.         REFERENCES

     1. ANSI C63.19-2007, American National Standard for Methods of Measurement of Compatibility
        between Wireless communication devices and Hearing Aids.”, New York, NY, IEEE, June 2007

     2. FCC Public Notice DA 06-1215, Wireless Telecommunications Bureau and Office of Engineering
        and Technology Clarify Use of Revised Wireless Phone Hearing Aid Compatibility Standard, June
        6, 2006

     3. FCC 3G Review Guidance, Laboratory Division OET FCC, May/June 2006

     4. Berger, H. S., “Compatibility Between Hearing Aids and Wireless Devices,” Electronic Industries
        Forum, Boston, MA, May, 1997

     5. Berger, H. S., “Hearing Aid and Cellular Phone Compatibility: Working Toward Solutions,”
        Wireless Telephones and Hearing Aids: New Challenges for Audiology, Gallaudet University,
        Washington, D.C., May, 1997 (To be reprinted in the American Journal of Audiology).

     6. Berger, H. S., “Hearing Aid Compatibility with Wireless Communications Devices, “ IEEE
        International Symposium on Electromagnetic Compatibility, Austin, TX, August, 1997.

     7. Bronaugh, E. L., “Simplifying EMI Immunity (Susceptibility) Tests in TEM Cells,” in the 1990 IEEE
        International Symposium on Electromagnetic Compatibility Symposium Record, Washington,
        D.C., August 1990, pp. 488-491

     8. Byme, D. and Dillon, H., The National Acoustics Laboratory (NAL) New Procedure for Selecting
        the Gain and Frequency Response of a Hearing Aid, Ear and Hearing 7:257-265, 1986.

     9. Crawford, M. L., “Measurement of Electromagnetic Radiation from Electronic Equipment using
        TEM Transmission Cells, “ U.S. Department of Commerce, National Bureau of Standards, NBSIR
        73-306, Feb. 1973.

     10. Crawford, M. L., and Workman, J. L., “Using a TEM Cell for EMC Measurements of Electronic
         Equipment,” U.S. Department of Commerce, National Bureau of Standards. Technical Note 1013,
         July 1981.

     11. EHIMA GSM Project, Development phase, Project Report (1st part) Revision A. Technical-
         Audiological Laboratory and Telecom Denmark, October 1993.

     12. EHIMA GSM Project, Development phase, Part II Project Report. Technical-Audiological
         Laboratory and Telecom Denmark, June 1994.

     13. EHIMA GSM Project Final Report, Hearing Aids and GSM Mobile Telephones: Interference
         Problems, Methods of Measurement and Levels of Immunity. Technical-Audiological Laboratory
         and Telecom Denmark, 1995.

     14. HAMPIS Report, Comparison of Mobile phone electromagnetic near field with an upscaled
         electromagnetic far field, using hearing aid as reference, 21 October 1999.

     15. Hearing Aids/GSM, Report from OTWIDAM, Technical-Audiological Laboratory and Telecom
         Denmark, April 1993.

                                                                                             Reviewed by:
FCC ID: ZNFAN160                               HAC (T-COIL) TEST REPORT
                                                                                             Quality Manager
Filename:                   Test Dates:         EUT Type:
                                                                                             Page 33 of 37
0Y1211121642.ZNF            11/15/2012          Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                         REV 6.4C
                                                                                                    04/05/10


     16. IEEE 100, The Authoritative Dictionary of IEEE Standards Terms, Seventh Edition.

     17. Joyner, K. H, et. al., Interference to Hearing Aids by the New Digital Mobile Telephone System,
         Global System for Mobile (GSM) Communication Standard, National Acoustic Laboratory,
         Australian Hearing Series, Sydney 1993.

     18. Joyner, K. H., et. al., Interference to Hearing Aids by the Digital Mobile Telephone System, Global
         System for Mobile Communications (GSM), NAL Report #131, National Acoustic Laboratory,
         Australian Hearing Series, Sydney, 1995.

     19. Kecker, W. T., Crawford, M. L., and Wilson, W. A., “Contruction of a Transverse Electromagnetic
         Cell”, U.S. Department of Commerce, National Bureau of Standards, Technical Note 1011, Nov.
         1978.

     20. Konigstein, D., and Hansen, D., “A New Family of TEM Cells with enlarged bandwidth and
         Optimized working Volume,” in the Proceedings of the 7th International Symposium on EMC,
         Zurich, Switzerland, March 1987; 50:9, pp. 127-132.

     21. Kuk, F., and Hjorstgaard, N. K., “Factors affecting interference from digital cellular telephones,”
         Hearing Journal, 1997; 50:9, pp 32-34.

     22. Ma, M. A., and Kanda, M., ”Electromagnetic Compatibility and Interference Metrology,” U.S.
         Department of Commerce, National Bureau of Standards, Technical Note 1099, July 1986, pp.
         17-43.

     23. Ma, M. A., Sreenivashiah, I. , and Chang, D. C., “A Method of Determining the Emission and
         Susceptibility Levels of Electrically Small Objects Using a TEM Cell,” U.S. Department of
         Commerce, National Bureau of Standards, Technial Note 1040, July 1981.

     24. McCandless, G. A., and Lyregaard, P. E., Prescription of Gain/Output (POGO) for Hearing Aids,
         Hearing Instruments 1:16-21, 1983

     25. Skopec, M., “Hearing Aid Electromagnetic Interference from Digital Wireless Telephones, “IEEE
         Transactions on Rehabilitation Engineering, vol. 6, no. 2, pp. 235-239, June 1998.

     26. Technical Report, GSM 05.90, GSM EMC Considerations, European Telecommunications
         Standards Institute, January 1993.

     27. Victorian, T. A., “Digital Cellular Telephone Interference and Hearing Aid Compatibility—an
         Update,” Hearing Journal 1998; 51:10, pp. 53-60

     28. Wong, G. S. K., and Embleton, T. F. W., eds., AIP Handbook of Condenser Microphones: Theory,
         Calibration and Measurements, AIP Press.




                                                                                                  Reviewed by:
FCC ID: ZNFAN160                                 HAC (T-COIL) TEST REPORT
                                                                                                  Quality Manager
Filename:                   Test Dates:            EUT Type:
                                                                                                  Page 34 of 37
0Y1211121642.ZNF            11/15/2012             Portable Handset
© 2012 PCTEST Engineering Laboratory, Inc.                                                              REV 6.4C
                                                                                                         04/05/10



Document Created: 2019-10-28 21:34:07