B_15I20528v0 SAR System Check Plots

FCC ID: ZNF329G

RF Exposure Info

Download: PDF
FCCID_2819007

Test Laboratory: UL Verification Services Inc. SAR Lab F                    Date/Time: 5/12/2015 10:35:29 PM

SystemPerformanceCheck-D835V2 SN 4d002

Frequency: 835 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0C; Liquid Temperature: 23.0C
                                                                               3
Medium parameters used: f = 835 MHz; σ = 1.016 S/m; εr = 54.889; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1359; Calibrated: 2/18/2015
- Probe: EX3DV4 - SN3936; ConvF(9.33, 9.33, 9.33); Calibrated: 7/24/2014;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 (A); Type: QDOVA001BB; Serial: 1117

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 1.10 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 34.837 V/m; Power Drift = -0.04 dB
Peak SAR (extrapolated) = 1.33 W/kg
SAR(1 g) = 0.886 W/kg; SAR(10 g) = 0.582 W/kg
Maximum value of SAR (measured) = 1.08 W/kg




                      0 dB = 1.08 W/kg = 0.33 dBW/kg




                                                                                                               Plot No. 1


Test Laboratory: UL Verification Services Inc. SAR Lab F   Date/Time: 5/12/2015 11:40:23 PM

SystemPerformanceCheck-D835V2 SN 4d002
Frequency: 835 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.20 W/kg




                                                                                          Plot No. 2


Test Laboratory: UL Verification Services Inc. SAR Lab H                    Date/Time: 5/12/2015 12:12:19 AM

SystemPerformanceCheck-D1900V2 SN 5d140

Frequency: 1900 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0C; Liquid Temperature: 23.0C
                                                                                3
Medium parameters used: f = 1900 MHz; σ = 1.571 S/m; εr = 51.847; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1472; Calibrated: 3/5/2015
- Probe: EX3DV4 - SN7335; ConvF(7.9, 7.9, 7.9); Calibrated: 3/13/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 B; Type: QDOVA002AA; Serial: 1248

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 5.49 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 61.875 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 7.38 W/kg
SAR(1 g) = 4.05 W/kg; SAR(10 g) = 2.1 W/kg
Maximum value of SAR (measured) = 5.48 W/kg




                      0 dB = 5.48 W/kg = 7.39 dBW/kg




                                                                                                               Plot No. 3


Test Laboratory: UL Verification Services Inc. SAR Lab H    Date/Time: 5/12/2015 12:30:14 AM

SystemPerformanceCheck-D1900V2 SN 5d140
Frequency: 1900 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 5.49 W/kg




                                                                                           Plot No. 4



Document Created: 2015-05-15 18:51:43