RFExposure FCC

FCC ID: ZLZTDBTSE

RF Exposure Info

Download: PDF
FCCID_2724678

                                     FCC RF Exposure
FCC ID: ZLZTDBTSE
Applicant: Shenzhen Mindray Bio-Medical Electronics Co.,Ltd.
Exposure category: General population/uncontrolled environment
EUT Type: Production Unit
Device Type: Portable Device
Refer Standard: KDB 447498 D01 General RF Exposure Guidance v05r02
FCC Part 2 §2.1093
Evaluation method
According to KDB447498 D01 General RF Exposure Guidance v05r01Section 4.3.1 Standalone SAR
test exclusion considerations:“ Unless specifically required by the published RF exposure KDB
procedures, standalone 1-g head or body and 10-g extremity SAR evaluation for general
population exposure conditions, by measurement or numerical simulation, is not required when
the corresponding SAR Test Exclusion Threshold condition, listed below, is satisfied. These test
exclusion conditions are based on source-based time-averaged maximum conducted output
power of the RF channel requiring evaluation, adjusted for tune-up tolerance, and the minimum
test separation distance required for the exposure conditions.22 The minimum test separation
distance is determined by the smallest distance from the antenna and radiating structures or
outer surface of the device, according to the host form factor, exposure conditions and platform
requirements, to any part of the body or extremity of a user or bystander (see 5) of section 4.1).
To qualify for SAR test exclusion, the test separation distances applied must be fully explained and
justified by the operating configurations and exposure conditions of the transmitter and
applicable host platform requirements, typically in the SAR measurement or SAR analysis report,
according to the required published RF exposure KDB procedures.         When no other RF exposure
testing or reporting is required, a statement of justification and compliance must be included in
the equipment approval, in lieu of the SAR report, to qualify for the SAR test exclusion. When
required, the device specific conditions described in the other published RF exposure KDB
procedures must be satisfied before applying these SAR test exclusion provisions; for example,
handheld PTT two-way radios, handsets, laptops & tablets etc.23 “[(max. power of channel,
including tune-up tolerance, mW)/ (min. test separation distance, mm)] ·[√f (GHz)]≤ 3.0 for 1-g
SAR and ≤ 7.5 for 10-g extremity SAR, where:
• f (GHz) is the RF channel transmit frequency in GHz
• Power and distance are rounded to the nearest mW and mm before calculation
• The result is rounded to one decimal place for comparison
• 3.0 and 7.5 are referred to as the numeric thresholds in the step 2 below
The test exclusions are applicable only when the minimum test separation distance is ≤ 50 mm
and for transmission frequencies between 100 MHz and 6 GHz. When the minimum test
separation distance is <5 mm, a distance of 5 mm according to 5) in section 4.1 is applied to
determine SAR test exclusion.


Conducted Power Results

                                      Bluetooth BLE 4.0
       Channel               Channel 0                Channel 20                 Channel 39
Output Power (dBm)             1.96                       1.24                      1.35


Manufacturing tolerance

                                      Bluetooth BLE 4.0
       Channel               Channel 0                Channel 20                 Channel 39
    Target (dBm)                1.0                       1.0                       1.0
   Tolerance ±(dB)              1.0                       1.0                       1.0

Evaluation Results


                                      Bluetooth BLE 4.0
                 Antenna    RF output power (including
                                                            SAR Test Exclusion       SAR Test
Frequency        Distance       tune-up tolerance)
                                                               Threshold             Exclusion
                 (mm)          dBm             mW
2402MHz              5          2              1.59              0.5<3.0                   Yes
2442MHz              5          2              1.59              0.5<3.0                   Yes
2480MHz              5          2              1.59              0.5<3.0                   Yes


Conclusion
The measurement results comply with the FCC Limit per 47 CFR 2.1093 for the uncontrolled RF
Exposure and SAR Exclusion Threshold per KDB 447498 v05r02.



Document Created: 2015-08-18 08:55:04
Document Modified: 2015-08-18 08:55:04

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC