                                                TEST REPORT
  Reference No. ....................        :   WTS19S02008374-4W

  FCC ID ................................   :   ZLZ-EPM

  Applicant ............................    :   Shenzhen Mindray BIO-Medical electronics Co.,LTD.

  Address ..............................    :   Mindray Building ,Keji 12th Road South,Hi-tech Ind , Shenzhen, China
  Manufacturer ....................         :   The same as above
  Address ..............................    :   The same as above

  Product...............................    :   Patient Monitor
  Model(s). ............................    :   ePM 15M, ePM 12M, ePM 10M, ePM 15, ePM 12, ePM 10
  Brand Name .......................        :   Mindray

  Standards...........................      :   FCC 2.1091

  Date of Receipt sample ....               :   2019-02-20

  Date of Test .......................      :   2019-02-21 to 2019-03-26

  Date of Issue ......................      :   2019-03-27
  Test Result .........................     :   Pass
  Remarks:
  The results shown in this test report refer only to the sample(s) tested, this test report cannot be
  reproduced, except in full, without prior written permission of the company. The report would be invalid
  without specific stamp of test institute and the signatures of compiler and approver.

                                                  Prepared By:
                                     Waltek Services (Shenzhen) Co., Ltd.
        Address: 1/F., Fukangtai Building, West Baima Road, Songgang Street, Baoan District, Shenzhen,
                                               Guangdong, China
                                             Tel :+86-755-83551033
                                             Fax:+86-755-83552400

          Compiled by:                                                     Approved by:




          Ford Wang / Project Engineer                                     Philo Zhong / Manager




Waltek Services (Shenzhen) Co.,Ltd.
http://www.waltek.com.cn                                  Page 1 of 9


Reference No.: WTS19S02008374-4W                  Page 2 of 9



2     Laboratories Introduction
Waltek Services (Shenzhen) Co., Ltd is a professional third-party testing and certification laboratory with
multi-year product testing and certification experience, established strictly in accordance with ISO/IEC 17025
requirements, and accredited by ILAC (International Laboratory Accreditation Cooperation) member. A2LA
(American Association for Laboratory Accreditation) of USA, Meanwhile, Waltek has got recognition as
registration and accreditation laboratory from EMSD (Electrical and Mechanical Services Department), and
American Energy star, FCC(The Federal Communications Commission), CEC(California energy efficiency),
IC(Industry Canada). It’s the strategic partner and data recognition laboratory of international authoritative
organizations, such as Intertek(ETL-SEMKO), TÜV Rheinland, TÜV SÜD, etc.




Waltek Services (Shenzhen) Co., Ltd is one of the largest and the most comprehensive third party testing
laboratory in China. Our test capability covered four large fields: safety test. Electro Magnetic Compatibility
(EMC), and energy performance, wireless radio. As a professional, comprehensive, justice international test
organization, we still keep the scientific and rigorous work attitude to help each client satisfy the international
standards and assist their product enter into globe market smoothly.




Waltek Services (Shenzhen) Co.,Ltd.
http://www.waltek.com.cn


Reference No.: WTS18S07118477-4W              Page 3 of 9



Test Facility:


      A. Accreditations for Conformity Assessment (International)
        Country/Region        Scope Covered By              Scope                Note
        USA                                                 FCC ID \ DOC \ VOC   1
        Canada                                              IC ID \ VOC          2
        Japan                                               MIC-T \ MIC-R        -
        Europe                                              EMCD \ RED           -
        Taiwan                                              NCC                  -
                              ISO/IEC 17025
        Hong Kong                                           OFCA                 -
        Australia                                           RCM                  -
        India                                               WPC                  -
        Thailand                                            NTC                  -
        Singapore                                           IDA                  -
        Note:
        1.   FCC Designation No.: CN1201. Test Firm Registration No.: 523476.
        2.   ISED CAB identifier: CN0013


      B. TCBs and Notify Bodies Recognized Testing Laboratory.
        Recognized Testing Laboratory of …                  Notify body number
        TUV Rheinland
        Intertek
                                                            Optional.
        TUV SUD
        SGS
        Phoenix Testlab GmbH                                0700
        Element Materials Technology Warwick Ltd            0891
        Timco Engineering, Inc.                             1177
        Eurofins Product Service GmbH                       0681




Waltek Services (Shenzhen) Co.,Ltd.
http://www.waltek.com.cn


Reference No.: WTS18S07118477-4W                                         Page 4 of 9


3     Contents
                                                                                                                                                                 Page
1    COVER PAGE ........................................................................................................................................... 1
2    LABORATORIES INTRODUCTION ................................................................................................................... 2
3    CONTENTS ............................................................................................................................................................ 4
4    REVISION HISTORY ............................................................................................................................................ 5
5    GENERAL INFORMATION .................................................................................................................................. 6
     5.1     GENERAL DESCRIPTION OF E.U.T. ................................................................................................................ 6
     5.2     DETAILS OF E.U.T. ......................................................................................................................................... 6
6    TEST SUMMARY .................................................................................................................................................. 7
7    RF EXPOSURE ...................................................................................................................................................... 8
     7.1     REQUIREMENTS............................................................................................................................................... 8
     7.2     THE PROCEDURES / LIMIT................................................................................................................................ 8
     7.3     MPE CALCULATION METHOD ......................................................................................................................... 9




Waltek Services (Shenzhen) Co.,Ltd.
http://www.waltek.com.cn


Reference No.: WTS18S07118477-4W            Page 5 of 9



4     Revision History

                    Date of
                                                  Date of
Test report No.     Receipt      Date of Test                Purpose    Comment   Approved
                                                   Issue
                    sample

                                  2019-02-21
WTS19S02008
                   2019-02-20     to 2019-03-   2019-03-27   original      -        Valid
    374-4W                             26




Waltek Services (Shenzhen) Co.,Ltd.
http://www.waltek.com.cn


Reference No.: WTS18S07118477-4W            Page 6 of 9



5     General Information
5.1   General Description of E.U.T.

      Product:                   Patient Monitor
      Model(s):                  ePM 15M, ePM 12M, ePM 10M, ePM 15, ePM 12, ePM 10
      Model Description:         Only the model names and antenna connect line length are different.
                                 2.4G-802.11b/g/n HT20
      Wi-Fi Specification:       5G-802.11a/ n HT20
      Hardware Version:          2.0

      Software Version:          01.06.00.01


5.2   Details of E.U.T.
      Ratings:                   Input:AC 100~240V~ 50/60Hz, 2.0~0.9A




Waltek Services (Shenzhen) Co.,Ltd.
http://www.waltek.com.cn


Reference No.: WTS18S07118477-4W           Page 7 of 9



6     Test Summary


                      Test Items                     Test Requirement   Result

            Maximum Permissible Exposure                  2.1091        PASS




Waltek Services (Shenzhen) Co.,Ltd.
http://www.waltek.com.cn


Reference No.: WTS18S07118477-4W                Page 8 of 9



7      RF Exposure
      Test Requirement:                 FCC Part 2.1091
      Test Mode:                        The EUT work in test mode(Tx).
7.1    Requirements

      Systems operating under the provisions of this section shall be operated in a manner that ensures that
      the public is not exposed to radio frequency energy levels in excess limit for maximum permissible
      exposure. In accordance with 47 CFR FCC Part 2 Subpart J, section 2.1091 this device has been
      defined as a mobile device whereby a distance of 0.2m normally can be maintained between the user
      and the device.
7.2    The procedures / limit

      (A) Limits for Occupational / Controlled Exposure
                                                Magnetic Field                       Averaging Time
      Frequency Range        Electric Field                      Power Density (S)
                                                 Strength (H)                          |E|2,|H|2or S
            (MHz)         Strength (E) (V/m)                        (mW/ cm2)
                                                    (A/m)                               (minutes)
           0.3-3.0               614                1.63                 (100)*              6
            3.0-30             1842 / f            4.89 / f          (900 / f)*              6
           30-300                61.4              0.163                  1.0                6
          300-1500                                                       F/300               6
        1500-100,000                                                       5                 6

       (B) Limits for General Population / Uncontrolled Exposure
                                               Magnetic Field                        Averaging Time
      Frequency Range        Electric Field                      Power Density (S)
                                                Strength (H)                          |E|2,|H|2 or S
            (MHz)         Strength (E) (V/m)                        (mW/ cm2)
                                                    (A/m)                               (minutes)
           0.3-1.34              614                1.63                 (100)*             30
           1.34-30              824/f              2.19/f             (180/f)*              30
           30-300                27.5              0.073                  0.2               30
          300-1500                                                    F/1500                30
        1500-100,000                                                      1.0               30
        Note: f = frequency in MHz ;
        *Plane-wave equivalent power density




Waltek Services (Shenzhen) Co.,Ltd.
http://www.waltek.com.cn


Reference No.: WTS18507118477—4w                 Page 9 of 9



7.3      MPE Calculation Method

Predication of MPE limit at a given distance
         m



                 4nR
8 = power densily (in appropriate units, e.g. mW/om®)
P = powerinput to the antenna(in appropriate units, e.9., mW).
G = power gain of the antennain the cirection of interest relative to an isotropic radiator, the power gain
factor, is normally numeric gain.
R = cistance to the canter of radiation of the antenna(appropriate units, e.4., or)

From the peak EUT RF output power, the minimum mobile separation distance, d=20om, as well as the
gain of the used antenna, the RF powerdensity can be obtained
Remark

Mode         Antenna     Antenna Gain Max.Peak Output| Peak Output| Power Density|             Limit of Power
cwir |       Gain (dB)    {numeric)        Power {dBm)      |Power {mW)]     (mWor®)        Density (mW/or®)

246            270          1.001              1808             so.er        0.030800                1
 se            338          2.i78              1878             as.os        oomiss                  1