SAR Efield Calibration Report 1

FCC ID: ZFT-K105

RF Exposure Info

Download: PDF
FCCID_1453225

    COMOSAR E—Field probe                                                               ix
          Calibration Report                                                         L2
                                                                 Ref: CR—131—1—09—SATB—B
                                                        Page: 1/17      Issue: B  Date: 2010/05/11




COMOSAR E—FIELD PROBE CALIBRATION REPORT



Prepared By:                    BUTET Romain, SATIMO

Project Description:            COMOSAR E—FIELD PROBE

Prepared For (End User):        CCS




This document is issued by SATIMO, in confidence and is not to be reproduced in whole or in part
without the prior written permission. The information contained herein is to be used only for the
purpose for which it is submitted and is not to be released in whole or in part without the prior written
permission of SATIMO.


    COMO'SAR.E —FiField probe                                             t(? 3 X
                                                                        S(QTEM
       Calibration Report                                                       2.
                                                                 Ref: CR-131-1-69-SATB—B
                                                       Page: 1/17        Issue: B        Date: 2010/05/1 1


              COMOSAR SEPT ISOTROPIC E—FIELD PROBE CALIBRATION REPORT


DATE: 5/11/2010

OFFER REFERENCE: PF.127.1.09.SATB.B

OBJECT: COMOSAR SEPT ISOTROPIC E—FIELD PROBE

MANUFACTURER: SATIMO

SERIAL NUMBER: SN 11/09 EP100

CUSTOMER: CCS

CONTRACT: B01351

DATE OF CALIBRATION: 5/5/2010


WARRANTY:
        This Calibration certificate may not be reproduced other than in full. Calibration certificates
without signature and seal are not valid. This documentation contains property information which is
protected by copyright. All right are reserved. No part of this document may be photocopied,
reproduced without the prior written agreement of SATIMO. SATIMO shall not be liable for errors
contained herein or for incidental or consequential in connection with the furnishing, performance or
use of this material. Warranty doesn‘t apply to Normal wear, Normal tear, Improper use, Improper
maintain, Improper installation.




Date




SAR TEAM MANAGER


 COMOSAR E—Field probe                                                  ((L. .,
       Calibration Report                                           SATE&@
                                                              Ref: CR—131—1—09—SATB—B
                                                     Page: 1/17       Issue: B        Date: 2010/05/1 1

                                  PRODUCT DESCRIPTION




 Frequency Range

 Probe length




: |Probe extremity diameter
                                                                                                 I



 Distance between dipoles/probe extremity                   <2.7 mm                              !

                                                            Dipole 1: R1=2.5307 MQ                   ‘
 Resistance of the three dipole (at the connector)          Dipole 2: R2=2.6353 MQ                   i
                                                            Dipole 3: R3=2.5471 MQ                   1



 Connector (HIROSE series SR30)                             6 wire male (Hirose SR3Oseries)


         The probe could be checked by measuring the resistance of the three dipoles.


                              CALIBRATION TEST EQUIPMENT



          TYPE                        IDENTIFICATION                DATE OF CALIBRATION
       ibrati                      CALISAR CALIBRATION                                                    |
    Calibration bench                  SYSTEM V2.0                                                        |
        Multimeter               Keithley (2000, SN: 1000572)     _|_Date of calibration: 01—04—2010


      COMOSAR E—Field probe                                              ((1.
                                                                     SATrIMC
          Calibration Report
                                                               Ref: CR—131—1—09—5ATB—B
                                                     Page: 1/17       Issue: B       Date: 2010/05/1 1

                                      MEASUREMENT PROCEDURE

Probe calibration is realized, in compliance with CENELEC EN 50361 and IEEE 1528 std, with
CALISAR, SATIMO proprietary calibration system. The calibration is performed with the EN 50361
annexe technique using reference guide at the five frequencies.




                  Waveguide




                      gan «. AlEr=—
                              (Pre — Pow) cos* (wg) e—(@=/85
                                          abd                  a
Where :
Piy        = Forward Power
Pbw        = Backward Power
a and b    =Waveguide dimensions
d          = Skin depth

Keithley configuration:

      Rate = Medium; Filter =ON; RDGS=10; FILTER TYPE =MOVING AVERAGE; RANGE AUTO

After each calibration, a SAR measurementis performed on a validation dl“fiole and compared with a
NPL calibrated probe, to verify it.


    COMOSAR E—Field probe                                                             ([( \
            Calibration Report                                                       S “TEE@@
                                                                         Ref: CR—131—1—09—SATB—B
                                                                Page: 1/17      Issue: B  Date: 2010/05/11

                                             PROBE UNCERTAINTIES




         ERROR sourcEs                     U:;:’:'(";";V   ;:‘t':::‘t':g‘   Divisor     ci   Uncs::::::;d(%X
Incident or forward power                     3,00%        Rectangular        Js_       1        1,732%

Reflected power                               3,00%        Rectangular        \/3_      1        1,732%

Liquid conductivity                           5,00%        Rectangular        1/5       4        2,887%

Liquid permittivity                           4,00%        Rectangular        J§        1        2,309%
Field homogeneity                             3,00%        Rectangular        \E        1        1,732%

Field probe positioning                       5,00%        Rectangular        \/5       1        2,887%
Field probe linearity                         3,00%        Rectangular        \/?T      1        1,732%
 Combined standard uncertainty                                         ~|                    C   4,761%
Expandedir::;:’rlt:flgtsyqli;:onfidenge,j                                                         9,331%


    COMOSAR E—Field probe                                                            ((c.,
          Calibration Report                                                    %fl@@                                     |
                                                                          Ref: CR—131—1—09—SATB—B
                                                         |    Page: 1/17           Issue: B          Date: 2010/05/1 1


                      1. Calibration at 835.00 MHz


A. Calibration parameters.
                     Label                                                              850
                    Epsilon                                                            41.81.
                     Sigma                                                           0.88 S/m
                  Temperature                                                          21°C
                   Cable loss                                                         0.12 dB
                  Coupler loss                                                       20.50 dB
                Waveguide $11                                                        —11.22 dB
               Low limit detection                                          0.824 V/im (0.604 mW/kg)


Calibration curves ei=f(V) (i=1,2,3) allow to obtain E—field value using the formula:
                                           E=(E‘+E,*+E,"



                                Calibration curves




          §                                                                              Dipole 1
          %                                                                               Dipole 2
          ui                                                                             Dipole 3




               _ o0   o1   o2    03   04    05 06        07    08   098    10 11
                                           Voltage (V)




The following tables represent the calibration curves linearization by curve segmentin CW signal.


      COMOSAR E—Field probe                                                                       ((/,X
                                                                                              SATIMo
        Calibration Report
                                                                                   Ref: CR-131-1-09-SATB—B
                                                                          Page: 1/17      Issue: B   Date: 2010/05/1 1

 Calibration coefficients for the three dipoles in CW:

                                                        Sensitivity in liquid:
      Liquid               Epsilon                 Sigma (S/m)         CF dipole 1           CF dipole 2   CF dipole 3
                                                                      (W.kg—1 (mV)—1) (W.kg—1 (mV)—1) (W.kg—1 (mV)—1)
       Head                    41.81                   0.90                  20.66               20.51       28.36
      Body                     55.51                   0.94                  20.00               19.88       27.77


  B. Isotropy.

 — Axial isotropy:                                    0.029 dB
 — Hemispherical isotropy:                            0.030 dB



                                 Isotropy curves


                                                                                     Hemispheric 30°
                                                                                     Hemizpheric "15°
                                                                                     Hemispheric 0°
                                                                                     Hemisphenc 15"
                                                                                     Hemicpheric 30°




           420.      08   06     04     02    00      0.20,30,40,50,60.70,8091.0




— C. Linearity.
                                                                                              y
                                       — Linearity:                                    0.04 dB


    COMOSAR E—Field probe                                                            ((f s K
                                                                              SATIMo
          Calibration Report
                                                                         Ref: CR—131—1—09—5ATB—B
                                                             Page: 1/17            Issue: B          Date: 2010/05/11



                      2. Calibration at 897.00 MHz


A. Calibration parameters.
                      Label                                                             900
                    Epsilon                                                          41.22
                     Sigma                                                         0.91 S/m
                 Temperature                                                         21°C
                  Cable loss                                                        0.11 dB
                 Coupler loss                                                      20.27 dB
                Waveguide $11                                                     —16.71 dB
               Low limit detection                                         0.795 V/im (0.59 mW/kg)


Calibration curves ei=f(V) (i=1,2,3) allow to obtain E—field value using the formula:
                                          E=4£#,*+£E,*+£E,*



                              Calibration curves




                                                                                          Dipole 1
                                                                                          Dipole 2
                                                                                          Dipole 3



             | 04—]      i     i           1     i      i       i    1
                 00     1     02     03   04    05      06     07   08       0.9
                                          Voltage (¥)




The following tables represent the calibration curves linearization by curve segment in CW signal.


     COMOSAR E—Field probe                                                                         ([( X
             Calibration Report                                                                 Tm@m
                                                                                      Ref: CR—131—1—09—SATB—B
                                                                           Page: 1/17           Issue: B    Date: 2010/05/11

Calibration coefficients for the three dipoles in CW:

                                                         Sensitivity in liquid:
     Liquid                 Epsilon                 Sigma (S/m)         CF dipole 1           CF dipole 2   CF dipole 3
                                                                      (W.kg—1 (mV)—1) (W.kg—1 (mV)—1) (W.kg—1 (mV)—1)
      Head        _             41.24                    0.94              22.06           22.01           30.16
      Body                      55.69                    1.00              21.56           21.36           29.10

B. Isotropy.

— Axial isotropy:                             0.029 dB
— Hemispherical isotropy:                     0.030 dB



                                  Isotropy curves
         1.07

         06                                                                           Hemispheric 30°
                                                                                      Hemispheric —15°
          0.6                                                                         Hemispheric 0‘
                                                                                      Hemisphenc 15"
                                                                                      Hericphenic 30°




         4.0                                                         ael
           C410       08   06     04      2    00     0.20.304050,60,70,00.91.0




C. Linearity.

                                       — Linearity:                                     0.04 dB"