RF Exposure Info 2

FCC ID: YY3-01120709267

RF Exposure Info

Download: PDF
FCCID_1464513

APPENDIX B: BV ADT SAR MEASUREMENT SYSTEM


APPENDIX C: PHOTOGRAPHS OF SYSTEM VALIDATION


APPENDIX D: SYSTEM CERTIFICATE & CALIBRATION


D1: PHANTOM


Schmid & Partner Engineering AG
                                                               s p                e a g
Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 1 245 9700, Fax +41 1 245 9779
info@speag.com, hittp://www.speag.com




Certificate of Conformity / First Article Inspection

 Item                             Oval Flat Phantom EL1 4.0
 Type No                          QD OVA 001B
 Series No                        1003 and higher
 Manufacturer                     SPEAG
                                  Zeughausstrasse 43
                                  CH—8004 Zurich
                                  Switzerland

Tests
Complete tests were made on the prototype units QD OVA 001 AA 1001, QD OVA 001 AB 1002,
pre—series units QD OVA 001 BA 1003—1005 as well as on the series units QD OVA 001 BB, 1006 ff.

 Test                    Requirement                               Details                       Units tested
 Dimensions              Compliant with the standard IEC           Dimensions of bottom          Prototypes,
                         62209 — 2 [1] requirements                for 300 MHz — 6 GHz:          Samples
                                                                   longitudinal = 600 mm
                                                                   (max. dimension)
                                                                   width= 400 mm (min
                                                                   dimension)
                                                                   depth= 190 mm
                                                                   Shape: ellipse
 Material thickness      Compliant with the standard IEC           Bottom plate:                 Prototypes,
                         62209 —— 2 [1] requirements               2.0mm +/ 0.2mm                All items
 Material                Dielectric parameters for required        300 MHz — 6 GHz               Material
 parameters              frequencies                               Rel. permittivity = 4 +/—1,   sample
                                                                   Loss tangent s 0.05
 Material resistivity    The material has been tested to be        DEGMBE based                  Equivalent
                         compatible with the liquids defined in    simulating liquids            phantoms,
                         the standards if handled and cleaned                                    Material
                         according to the instructions.                                          sample
                         Observe Technical Note for material
                         compatibility.
 Sagging                 Compliant with the requirements           <1% typical < 0.8% if         Prototypes,
                         according to the standard.                filled with 155mm of          Sample
                         Sagging of the flat section when filled   HSL900 and without            testing
                         with tissue simulating liquid             DUT below


Standards
[1] EC 62209 — 2, Draft Version 0.9, "Evaluation of Human Exposure to Radio Frequency Fields from
    Handheld and Body—Mounted Wireless Communication Devices in the Frequency Range of 30 MHz
    to 6 GHz: Human models, Instrumentation and Procedures
    Part 2: Procedure to determine the Specific Absorption Rate (SAR) for ... including accessories and
    multiple transmitters", December 2004

Conformity
Based on the sample tests above, we certify that this item is in compliance with the standard [1].

Date                                07.07.2005          $           e__a
                                                         5chm%qAG
Signature / Stamp                                        Zoughasfstrasse 4378004 Z\Wfld/
                                                              f
                                                         Phong 41 124                aArdis —
                                                         info@speag.com, httpiiiwww.speag.com




Doc No   881—QD OVA 001B—C                                                                        Page         1(1)


D2: DOSIMETRIC E-FIELD PROBE


Calibration Laboratory of                                            wyt9,                                                loar          Snrieedi
                                                                   ‘\\\/_/ 7                                       Schweizerischer Kalibrierdienst
Schmid & Partner                                                 i\ia\Em{/é                                        Service suisse d‘étalonnage
  Engineering AG                                                  Ls                                               Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                       ’«/,////-_\T\\‘v‘                               Swiss Calibration Service
                                                                       flls


Accredited by the Swiss Accreditation Service (SAS)                                               Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       BV—ADT (Auden)                                                                       Certificate No: EX3—3504_Jan10

|CALIBRATION CERTIFICATE                                                                                                                                I
 Object                                EX3DV3 — SN:3504


 Calibration procedure(s)              QA CAL—01.v6, QA CAL—14.v3, QA CAL—23.v3 and QA CAL—25.v2
                                       Calibration procedure for dosimetric E—field probes



 Calibration date:                     January 26, 2010




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards              _       ID #         _            Cal Date (Certificate No.) ___________                Scheduled Calibration        ___|
 Power meter E44198                     GB41293874                 1—Apr—09 (No. 217—01030)                             Apr—10
 Power sensor E4412A                    MY41495277                 1—Apr—09 (No. 217—01030)                             Apr—10
 Power sensor E4412A                    MY41498087                 1—Apr—09 (No. 217—01030)                             Apr—10                              |
 Reference 3 dB Aftenuator              SN: $5054 (3c)            31—Mar—09 (No. 217—01026)                             Mar—10                              |
 Reference 20 dB Attenuator             SN: $5086 (20b)           31—Mar—09 (No. 217—01028)                             Mar—10                              ‘
 Reference 30 dB Attenuator             SN: $5129 (30b)           31—Mar—09 (No. 217—01027)                             Mar—10
 Reference Probe ES3DV2                 SN: 3013                  30—Dec—09 (No. ES3—3013_Dec08)                        Dec—10
 DAE4                                   SN: 660                   29—Sep—09 (No. DAE4—660_SepO9)                        Sep—10

 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 RF generator HP 8648C                  US3642U01700              4—Aug—99 (in house check Oct—09)                      In house check: Oct—11
 Network Analyzer HP 8753E              US37390585                 18—Oct—01 (in house check Oct—09)                    In house check: Oct10


                                        Name                                      Function                               Signature
 Calibrated by:                         KatJa Pokovic                              Technical Manager                                   é;‘



 Approved by:                           Fin Bomholt                                R&D Director                    —+—~/Z            /%/
                                                                                                                   en
                                                                                                                    y
                                                                                                                      P—fa


                                                                                                                        Issued: January 26, 2010
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: EX3—3504_Jan10                                        Page 1 of 11


Calibration Laboratory of                                                                     Schweizerischer Kalibrierdienst
Schmid & Partner                                                                              Service suisse d‘étalonnage
  Engineering AG                                                                              Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                  Swiss Calibration Service


Accredited by the Swiss Accreditation Service ($AS)                                       Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates



Glossary:
TSL                          tissue simulating liquid
NORMx,y,z                    sensitivity in free space
ConvF                        sensitivity in TSL / NORMx,y,z
DCP                          diode compression point
CF                           crest factor (1/duty_cycle) of the RF signal
A, B, C                      modulation dependentlinearization parameters
Polarization                 @ rotation around probe axis
Polarization 9               8 rotation around an axis that is in the plane normal to probe axis (at measurement center),
                             Le., 9 = 0 is normal to probe axis

Calibration is Performed According to the Following Standards:
    a)   IEEE Std 1528—2003, "EEE Recommended Practice for Determining the Peak Spatial—Averaged Specific
         Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement
         Techniques", December 2003
    b)   1EC 62208—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held devices used in close
         proximity to the ear (frequency range of 300 MHz to 3 GHz)", February 2005

Methods Applied and Interpretation of Parameters:
    *    NORMx,y,z: Assessed for E—field polarization 9 = 0 (f s 900 MHz in TEM—cell; > 1800 MHz: R22 waveguide}.
         NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the E*—field
         uncertainty inside TSL (see below ConvF).
    *    NORM(Mx,y,2 = NORMx,y,2 * frequency_response (see Frequency Response Chart). This linearization is
         implemented in DASY4 software versions later than 4.2. The uncertainty of the frequency response is included
         in the stated uncertainty of ConvF.
    «_   DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep with CW
         signal (no uncertainty required). DCP does not depend on frequency nor media.
    +    Axy,2; Bxy,2; Cxy,2, VRxy,z: A, B, C are numerical linearization parameters assessed based on the data of
         power sweep for specific modulation signal. The parameters do not depend on frequency nor media. VR is the
         maximum calibration range expressed in RMS voltage across the diode.
    *    ConvF and Boundary Effect Parameters: Assessed in flat phantomn using E—field (or Temperature Transfer
         Standard for f< 800 MHz) and inside waveguide using analytical field distributions based on power
         measurements for f > 800 MHz. The same setups are used for assessment of the parameters applied for
         boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
         used in DASY4 software to improve probe accuracy close to the boundary. The sensitivity in TSL corresponds
         to NORMx,y,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
         ConvF is used in DASY version 4.4 and higher which allows extending the validity from + 50 MHz to + 100
         MHz.
    *    Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat phantom
         exposed by a patch antenna.
    *    Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the probe tip
         (on probe axis). No tolerance required.




Certificate No: EX3—3504_Jan10                                  Page 2 of 11


EX3DV3 SN:3504                                                                       January 26, 2010




                             Probe EX3DV3

                                            SN:3504
                              Manufactured:                      December 15, 2003
                              Last calibrated:                   January 21, 2009
                              Recalibrated:                      January 26, 2010

                                      Calibrated for DASY Systems
                                       (Note: non—compatible with DASY2 system!)




  Certificate No: EX3—3504_Jan10                 Page 3 of 11


EX3DV3 SN:3504                                                                                                                              January 26, 2010


DASY — Parameters of Probe: EX3DV3 SN:3504


Basic Calibration Parameters
                                                                                             Sensor X Sensor Y Sensor Z |Unc (k=2)
Norm (uV/(V/im)")®                                                                              0.59           0.62           0.62          |+101%
DCP (my)®                                                                                       97.9           95.0           98.0


Modulation Calibration Parameters
UID             Communication System Name                               PAR                       A              B              C             VR          Unc®
                                                                                                  dB           dBuV                           mV          (k=2)
10000           CW                                                      0.00        X                 0.00           0.00           1.00|     300        # 1.5%
                                                                                    Y                 0.00           0.00           1.001     300
                                                                                    Z                 0.00           0.00           1.00|     300




 The reported uncertainty of measurement is stated as the standard uncertainty of measurement multiplied
 by the coverage factor k=2, which for a normal distribution corresponds to a coverage probability of
 approximately 95%.


* The uncertainties of NormX,Y,Z do not affect the E—field uncertainty inside TSL (see Pages 5 and 6).

* Numerical lingarization parameter: uncertainty not required
& Uncertainty is determined using the maximum deviation from linear response applying recatangular distribution and is expressed for the square of the field value.




      Certificate No: EX3—3504_Jan10                                    Page 4 of 11


EX3DV3 SN:3504                                                                                                                         January 26, 2010




DASY — Parameters of Probe: EX3DV3 SN:3504

Calibration Parameter Determined in Head Tissue Simulating Media


f [MHz]         Validity [MHz]®       Permittivity         Conductivity      ConvF X       ConvF Y__      ConvF 2             Alpha        Depth Une (k=2)

900             * 50 /+ 100           41.5 4 5%            0.97 £ 5%                9.80           9.80          9.80           0.48         0.73 £11.0%

1750            £ 50 / £ 100          40.1 4 5%            1.37 4 5%                8.70           8.70          8.70           0.50         0.67 £11.0%

1950            * 50 / £ 100          40.0 4 5%            1.40 2 5%                8.20           8.20          8.20           0.38         0.75 £11.0%

2450            * 50 / £ 100          39.2 £ 5%            1.80 £ 5%                7.77           7.77          7.77           0.21         1.06 £11.0%

2600            * 50 / £ 100          39.0 £ 5%            1.96 £ 5%                7.79           7.19          7.79           0.22         116 £110%

5200            £ 50 /+ 100           36.0 + 5%            4.66 £ 5%                4.87           4.87          4.87           0.45         1.10 £13.1%

5300            £ 50 / £ 100          35.9 £ 5%            4.176 £ 5%               4.62           4.62          4.62           0.45         1.70 £13.1%

5500            * 50 / £ 100          35.6 £ 5%            4.96 # 5%                4.51           4.51          4.51           0.50         1.70 £13.1%

5600            £ 50 /# 100           35.5 £ 5%            5.07 £ 5%                4.25           4.25          4.25           0.55         1.70 £13.1%

5800            * 50 /    100         35.3 + 5%            5.27 £ 5%                4.53           4.53           4.53          0.50         1.70 £13.1%




© The validity of + 100 MHz only applies for DASY v.4 and higher (see Page 2). The uncertainty is the RSS of the ConvF uncertainty at calibration frequency

and the uncertainty for the indicated frequency band




      Certificate No: EX3—3504_Jan10                                   Page 5 of 11


EX3DV3 SN:3504                                                                                                                          January 26, 2010




DASY — Parameters of Probe: EX3DV3 SN:3504

Calibration Parameter Determined in Body Tissue Simulating Media


f (MHz]         Validity [MHz]C        Permittivity        Conductivity       ConvFX        ConvFY         ConvF 2             Alpha        Depth Unc (k=2)

900             # 50 /# 100            55.0 £ 5%           1.05 £ 5%                 9.83          9.83           9.83           0.44         0.76 #11.0%

1750            £ 50 / £ 100           §3.4 £ 5%            1.49 £ 5%                8.64           8.64          8.64           0.44         0.74 #11.0%

1950            * 50 / £ 100           53.3 £ 5%           1.52 4 5%                 8.52           8.52          8.52           0.39         0.79 #11.0%

2450            * 50 /« 100            §2.17 £ 5%          1.95 4 5%                 7.91           7.91          7.91           0.32         0.86 #11.0%

2600            * 50 /    100          52.5 £ 5%           216 £ 5%                  7.80           7.80          7.80           0.27         0.90 #11.0%

5200            * 50 / £ 100           49.0 4 5%           5.30 £ 5%                 4.45           4.45          4.45           0.50         1.80 £13.1%

5300            * 50 / + 100           48.5 4 5%           542 + 5%                  4.18           4.18          4.18           0.55         1.80 £13.1%

§500            * 50 / 100             48.6 4 5%           5.65 x 5%                 3.91           3.91          3.91           0.60         1.80 £13.1%

5600            * 50 / + 100           48.5 * 5%           5.77 # 5%                 3.70           3.70          3.70           0.65         1.80 £13.1%

5800            £ 50 / £ 100           48.2 £5%            6.00 £ 5%                 3.95           3.95          3.95           0.60         1.75 £181%




° The validity of £ 100 MHz only applies for DASY v4.4 and higher (see Page 2). The uncertainty is the RSS of the ConvF uncertainty at calibration frequency

and the uncertainty for the indicated frequency band.




      Certificate No: EX3—3504_Jan10                                   Page 6 of 11


EX3DV3 SN:3504                                                                                                                       January 26, 2010



                                                                    Frequency Response of E—Field
                                                                      (TEM—Cell:ifi110 EXX, Waveguide: R22)


                                                            15

                                                            1.4:;

                                                            13
                          Frequency response (normalized)




                                                            12

                                                            11 L

                                                            10 ;

                                                            o.9 .

                                                            0.8 :


                                                            0.7 ;


                                                            0.6 :

                                                            o.s |
                                                                o         500       1000        1500     2000     2500        3000
                                                                                             f [MHz]

                                                                                —@—TEM                 ~@—R22




                                                                 Uncertainty of Frequency Response of E—field: £ 6.3% (k=2)




  Certificate No: EX3—3504_Jan10                                                    Page 7 of 11


EX3DV3 SN:3504                                                                                       January 26, 2010




                                      Receiving Pattern (¢), 9 = 0°

                      f = 600 MHz, TEM ifi110EXX                          f= 1800 MHz, WG R22




                                                                                             —O—30 MHz
                                                                                             —BI— 100 MHz
                                                                                             —@—600 MHz
                                                                                             —IB— 1800 MHz
                                                                                             —&—2500 MHz



                  0            60         120                       240




                                    Uncertainty of Axial Isotropy Assessment: £ 0.5% (k=2)




  Certificate No: EX3—3504_Jan10                     Page 8 of 11


EX3DV3 SN:3504                                                                                                             January 26, 2010



                                                          Dynamic Range f(SARpesq)
                                                                  (Waveguide R22, f = 1800 MHz)

                                           1.E+06 ,


                                           1.2+05


                                           1.2+04
                       Input Signal [xV]




                                           1.E+08



                                           1.E+02




                                           1.2+01 ;


                                           1.E+00
                                                0.0001     0.001       0.01             0.1        1         10      100

                                                                                SAR [mW/cm‘]
                                                            ~—@®—not compensated                   —®— compensated




                                                           0.01               0.1              1            10
                                                                                    SAR [mW/cm‘]




                                                         Uncertainty of Linearity Assessment: £ 0.6% (k=2)




  Certificate No: EX3—3504_Jan10                                         Page 9 of 11


EX3DV3 SN:3504                                                                                                                                          January 26, 2010



                                     Conversion Factor Assessment

                                                                                                        f= 1750 MHz, WGLS R22 (head)




                                                                       SAR[mW/cm?] / W
                                                                                         3
                                                                                         o
                                                                                         a
                                                                                         o
                                                                                         s30
                                                                                         ao
                                                                                         2
                                                                                         o
                                                                                               0                        10                20       30            40
                                                                                                                             z[mm]

          | —@—Analytical      —O—Measurements        ‘                                            l    —@— Analytical                     —O—Measurements   |




                                      Deviation from Isotropy in HSL
                                                          Error (¢, 9), f = 900 MHz
                                                                                                                 00 s
                                                                                                                  g2:
                                                                                                               poese
                                                                                                          ornona
                                                                                                         200620
                                                                                                        beSrrE



                                                                                                                             Error [dB]
                                                                                                        266




                                   BI—1.00—0.80 IMl—0.80——0.60 I—0.60——0.40 1—0.40——0.20 RI—0.20—0.00

                                   ©10.00—0.20   m10.20—0.40      0.40—0.60                ©0.60—0.80          M0.80—1.00




                                   Uncertainty of Spherical Isotropy Assessment: £ 2.6% (k=2)




  Certificate No: EX3—3504_Jan10                               Page 10 of 11


EX3DV3 SN:3504                                              January 26, 2010




Other Probe Parameters

Sensor Arrangement                                                  Triangular
Connector Angle (°)                                             Not applicable
Mechanical Surface Detection Mode                                     enabled
Optical Surface Detection Mode                                       disabled
Probe Overall Length                                                  337 mm
Probe Body Diameter                                                    10 mm
Tip Length                                                              9 mm
Tip Diameter                                                          2.5 mm
Probe Tip to Sensor X Calibration Point                                 1 mm
Probe Tip to Sensor Y Calibration Point                                 1 mm
Probe Tip to Sensor Z Calibration Point                                 1 mm
Recommended Measurement Distance from Surface                           2 mm




   Certificate No: EX3—3504_Jan10           Page 11 of 11


D3: DAE


Schmid & Partner Engineering AG                               s                e       a

Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 44 245 9700, Fax +41 44 245 9779
info@speag.com, http://www.speag.com




                                        IMPORTANT NOTICE


USAGE OF THE DAE 4
The DAE unit is a delicate, high precision instrument and requires careful treatment by the user. There are no
serviceable parts inside the DAE. Special attention shall be given to the following points:

Battery Exchange: The battery cover of the DAE4 unit is closed using a screw, over tightening the screw may
cause the threads inside the DAE to wear out.

Shipping of the DAE: Before shipping the DAE to SPEAG for calibration, remove the batteries and pack the
DAE in an antistatic bag. This antistatic bag shall then be packed into a larger box or container which protects the
DAE from impacts during transportation. The package shall be marked to indicate that a fragile instrument is
inside.

E—Stop Failures: Touch detection may be malfunctioning due to broken magnets in the E—stop. Rough handling
of the E—stop may lead to damage of these magnets. Touch and collision errors are often caused by dust and dirt
accumulated in the E—stop. To prevent Estop failure, the customer shall always mount the probe to the DAE
carefully and keep the DAE unit in a non—dusty environmentif not used for measurements.

Repair: Minor repairs are performed at no extra cost during the annual calibration. However, SPEAG reserves
the right to charge for any repair especially if rough unprofessional handling caused the defect.

DASY Configuration Files: Since the exact values of the DAE input resistances, as measured during the
calibration procedure of a DAE unit, are not used by the DASY software, a nominal value of 200 MOhm is given
in the corresponding configurationfile.

Important Note:
Warranty and calibration is void if the DAE unit is disassembled partly or fully by the
Customer.


Important Note:
Never attempt to grease or oil the E—stop assembly. Cleaning and readjusting of the E—
stop assembly is allowed by certified SPEAG personnel only and is part of the annual
calibration procedure.


Important Note:
To prevent damage of the DAE probe connector pins, use great care when installing the
probe to the DAE. Carefully connect the probe with the connector notch oriented in the
mating position. Avoid any rotational movement of the probe body versus the DAE
while turning the locking nut of the connector. The same care shall be used when
disconnecting the probe from the DAE.



Schmid & Partner Engineering


TN_BRO40315AD DAE4.doc                                                                                 11.12.2009


     &       :                                                        o¥49p
Calibration Laboratory Of                                           \“\\t//,”’/                                    Schweizerischer Kalibrierdienst
Schmid & Partner                                                 ;i\\_//l/;:                                       Service suisse d‘étalonnage
  Engineering AG                                                  ://R >                                           Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                       'f/,l///—\\\ y                                  Swiss Calibration Service
                                                                      alalAs®

Accredited by the Swiss Accreditation Service (SAS)                                            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client                                                                                         Certificate No: DAE4—861_Jan10

CALIBRATIONCERTIFICATE_
 Object                                DAE4— SD 000 pod4 BJ — SN: 861


 Calibration procedure(s)             QA CAL-06 vi2                          1.                      10
                                        ca brat|on‘rocedure orthe dataacqussn on



 Calibration date:                    January 22. 2010




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratoryfacility: environment temperature (22 « 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Cerfificate No.)                            Scheduled Calibration
 Keithiey Multimeter Type 2001          SN: 0810278               1—0ct—09 (No: 9055)                                   Oct—10

 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Calibrator Box V1.1                    SE UMS 006 AB 1004        0§—Jun—09 (in house check}                            In house check: Jun—10




                                        Name                                    Function                                 Signature
 Calibrated by:                         Ahdréa Guntli .                      . ‘Techhii: n



 Approved by:                           FinBomholt        _\ [R&D Director _                             10          l               e


                                                                                                                         Issued: January 22, 2010
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: DAE4—861_Jan10                                        Page 1 of 5


Calibration Laboratory of                                    ul2                 Schweizerischer Kalibrierdienst
Schmid & Partner                                            ig:\%//;%            Service suisse d‘étalonnage
  Engineering AG                                            7 n c                Servizio svizzero di taratura
                             .      .                       4w y                    a     mean         —
Zeughausstrasse 43, 8004 Zurich, Switzerland                ’f/,,/fi\\\\\*        Swiss Calibration Service
                                                                alalids

Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates



Glossary
DAE                          data acquisition electronics
Connector angle              information used in DASY system to align probe sensor X to the robot
                             coordinate system.

Methods Applied and Interpretation of Parameters
   e DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
         comparison with a calibrated instrument traceable to national standards. The figure given
         corresponds to the full scale range of the voltmeter in the respective range.
    «_   Connector angle: The angle of the connector is assessed measuring the angle
         mechanically by a tool inserted. Uncertainty is not required.
    *    The following parameters as documented in the Appendix contain technical information as a
         result from the performance test and require no uncertainty.
               DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
               the nominal calibration voltage. Influence of offset voltage is included in this
               measurement.
               Common mode sensitivity: Influence of a positive or negative common mode voltage on
               the differential measurement.
               Channel separation: Influence of a voltage on the neighbor channels not subject to an
               input voltage.
               AD Converter Values with inputs shorted: Values on the internal AD converter
               corresponding to zero input voltage
               Input Offset Measurement. Output voltage and statistical results over a large number of
               zero voltage measurements.
               Input Offset Current: Typical value for information; Maximum channel input offset
               current, not considering the input resistance.
               Input resistance: DAE input resistance at the connector, during internal auto—zeroing
               and during measurement.
               Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
               alarm signal is generated.
               Power consumption: Typical value for information. Supply currents in various operating
               modes.




Certificate No: DAE4—861_Jan1O                                  Page 2 of 5


DC Voitage Measurement
    A/D — Converter Resolution nominal
        High Range:          1LSB =     6AuV ,        full range =  ~100...+300 mV
        Low Range:           1LSB =     61nV ,        full range =  hase +3mV
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


      Calibration Factors                 X                        Y                     Z
      High Range                 404.395 + 0.1% (k=2) 404.784 + 0.1% (k=2)   405.737 £ 0.1% (k=2)
      Low Range                  4.01182 + 0.7% (k=2) 3.98893 + 0.7% (k=2)   4.01269 + 0.7% (k=2)


Connector Angle

      Connector Angle to be used in DASY system                                      123.0°%%1°




Certificate No: DAE4—861_Jan10                       Page 3 of 5


Appendix

1. DC Voltage Linearity
     High Range                                Reading (uV)             Difference (uV)              Error (%)
     Channel X        + Input                    200003.4                       —2.33                 —0.00
     Channel X        + Input                    19997.73                       ~1.97                 —0.01
     Channel X        — Input                   ~19999.33                       1.07                  —0.01
     Channel Y        + input                    200002.5                       216                   —0.00
     Channel Y        + Input                    19995.17                       4.43                  —0.02
     Channel Y        — Input                   20000.88                        —0.58                 0.00
     Channel Z        + input                    199999.9                       —3.99                 —0.00
     Channel Z2       + Input                    19995.97                       ~3.48                 —0.02
     Channel Z        — Input                   20002.39                        0.01                  0.01


      Low Range                                Reading (uV)             Difference (V)                Error (%)
     Channel X        + Input                     1999.9                        —0.34                 0.02
     Channel X        + Input                     199.26                        —0.64                 —0.32
     Channel X        — input                    +200.72                        —0.82                 0.41
     Channel Y        + Input                     1999.2                        —0.72                 —0.04
     Channel Y        + Input                     198.82                        —1.18                 +0.59
     Channel Y        — input                    —201.63                        ~1.83                 0.92
     Channel 2        + Input                     2001.1                        1.22                  0.06
     Channel 2        + Input                     197.99                        —2.01                 ~1.01
     Channel Z        — Input                    —201,59                        —1.79                 0.89



2. Common mode sensitivity
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                        Common mode                       High Range                               Low Range
                        Input Voltage (mV)            Average Reading (1V)                    Average Reading (V)

     Channel X                          200                           5.69                          4.19
                                       +200                          —2.41                          —3.99
     Channel Y                          200                           1.24                          1.40
                                       ~200                          +2.48                         ~2.38

     Channel Z                          200                          9. 16                          —9.25
                                       — 200                          8.58                          8.07


3. Channel separation
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 see
                       Input Voltage (mV)         Channel X (uV)              Channel Y (V)       Channel Z (uV)
    Channel X                    200                        —                      2.32                1.39
    Channel Y                    200                       1.89                         —              4.20
    Channel Z                    200                       1.32                    0.19                     —


Certificate No: DAE4—861_Jan10                                  Page 4 of 5


4. AD—Converter Values with inputs shorted
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                                                    High Range (LSB)                      Low Range (LSB)

     Channel X                                              15980                              16853
     Channel Y                                              16068                              14547
     Channel Z                                              16088                              17866


5. Input Offset Measurement
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
    Input 10M

                            Average (uV)        min. Offset (uV)         max. Offset (uv)      5* [():\‘,’;a“m
     Channel X                   0.16                 —1.64                    1.60                    0.66
     Channel Y                   +0.58                ~2.29                   0.44                     0.49
     Channel Z                   —1.24                ~3.18                   0.50                     0.68



6. Input Offset Current
     Nominal Input circuitry offset current on all channels: <25fA


7. Input Resistance
                                                  Zeroing (MOhm)                         Measuring (MOhm)

     Channel X                                         0.1999                                  199.8
     Channel Y                                         0.2000                                 201.8
     Channel Z                                         0.1999                                  199.7



8. Low Battery Alarm Voltage (verified during pre test)
    Typical values                                Alarm Level (VDC)
     Supply (+ Vec)                                                               +7.9
     Supply (— Vec)                                                               —7.6



9. Power Consumption (verified during pre test)
    Typical values                          Switched off (mA)        Stand by (mA)          Transmitting (mA)
    Supply (+ Vec)                                  +0.0                     +6                    +14

    Supply (— Vec)                                 —0.01                     —8                        —9g




Certificate No: DAE4—861_Jan10                             Page 5 of 5


D4: SYSTEM VALIDATION DIPOLE


     4       4                                                       «ul Um
Calibration Laboratory of                                         Cs/s                                             Schweizerischer Kalibrierdienst
      .                                                          gst
Schmid & Partner                                                iia%il/(&}é                                        Service suisse d‘étalonnage
   Engineering AG                                               J ids                                              Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                     41/,////-3\\\\3                                   Swiss Calibration Service
                                                                     Ahilals


Accredited by the Swiss Accreditation Service (SAS)                                             Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       ADT (Auden)                                                                        Certificate No: D835V2—4d021_Apr10

[CALIBRATION CERTIFICATE
 Object                                D835V2 — SN: 40021


 Calibration procedure(s)              QA CAL—O5.v7
                                       Calibration procedure for dipole validation kits



 Calibration date:                     April 29, 2010




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE criticalfor calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter EPM—442A                   GB37480704                06—Oct—09 (No. 217—01086)                             Oct—10
 Power sensor HP 8481A                  US37202783                06—Oct—09 (No. 217—01086)                             Oct—10
 Reference 20 dB Attenuator             SN: 5086 (20g)            30—Mar—10 (No. 217—01158)                             Mar—11
 Type—N mismatch combination            SN: 5047.2 / 06327        30—Mar—10 (No. 217—01162)                             Mar—11
 Reference Probe ESSDV3                 SN: 3205                  26—Jun—09 (No. ES3—3205_Jun09)                        Jun—10
 DAE4                                   SN: 601                   02—Mar—10 (No. DAE4—601_Mar10)                        Mar—11


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power sensor HP 8481A                  MY41092317                18—Oct—02 (in house check Oct—09)                     In house check: Oct—11
 RF generator R&S SMT—06                100005                    4—Aug—99 (in house check Oct—09)                      In house check: Oct—11
 Network Analyzer HP 8753E              US37390585 $4206          18—Oct—01 (in house check Oct—09)                     In house check: Oct—10



                                        Name                                   Function                                  Signature


                                                                                                                     C
 Calibrated by:                         Dimcelliev                             Laboratory Technician




                                                                                                                     TeMss
 Approved by:                           KatJa Pokovic                          Technical Manager



                                                                                                                        Issued: April 29, 2010
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D835V2—4d021_Apr10                                    Page 1 of 9


Calibration Laboratory of                                                       Schweizerischer Kalibrierdienst
Schmid & Partner                                                                Service suisse d‘étalonnage
  Engineering AG                                                                Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                    Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                          Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                         tissue simulating liquid
ConvF                       sensitivity in TSL / NORM x.y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
         Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
         Communications Devices: Measurement Techniques", December 2003
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
       devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
    c) Federal Communications Commission Office of Engineering & Technology (FCC OET),
       "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency
       Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
       Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions",
       Supplement C (Edition 01—01) to Bulletin 65

Additional Documentation:
    d) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    *    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    *    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    e    SAR measured: SAR measured at the stated antenna input power.
    e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    &    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.



Certificate No: D835V2—4d021_Apr10                             Page 2 of 9


Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                              DASY5                               V5.2

      Extrapolation                                   Advanced Extrapolation

      Phantom                                       Modular Flat Phantom V4.9
      Distance Dipole Center — TSL                              15 mm                            with Spacer
      Zoom Scan Resolution                              dx, dy, dz =5 mm
      Frequency                                          835 MHz &1 MHz


Head TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity        Conductivity
      Nominal Head TSL parameters                                  22.0 °C           41.5              0.90 mho/m
      Measured Head TSL parameters                              (22.0 # 0.2) °C   41.8 x 6 %        0.89 mho/m + 6 %
      Head TSL temperature during test                          (22.0 + 0.2) °C



SAR result with Head TSL

      SAR averaged over 1 em* (1 g) of Head TSL                  Condition
      SAR measured                                        250 mW input power                   2.37 mW /g
      SAR normalized                                          normalized to 1W                 9.48 mW / g
      SAR for nominal Head TSL parameters                     normalized to 1W        9.58 mW /g « 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                 condition
      SAR measured                                        250 mW input power                   1.55 mWw /g
      SAR normalized                                          normalized to 1W                 6.20 mW /g
      SAR for nominal Head TSL parameters                     normalized to 1W        6.24 mW ig * 16.5 % (k=2)




Certificate No: D835V2—44021_Apr10                     Page 3 of 9


Body TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity        Conductivity
      Nominal Body TSL parameters                                  22.0 °C           55.2              0.97 mho/m
      Measured Body TSL parameters                              (22.0 + 0.2) °C   54.9 * 6 %        0.99 mho/m + 6 %
      Body TSL temperature during test                          (22.0 + 0.2) °C



SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
      SAR measured                                        250 mW input power                   2.52 mW / g
      SAR normalized                                          normalized to 1W                 10.1 mWw /g
      SAR for nominal Body TSL parameters                     normalized to 1W       9.91 mW / g # 17.0 % (k=2)


      SAR averaged over 10 cm* (10 g) of Body TSL                 condition
      SAR measured                                        250 mW input power                   1.65 mWw /g
      SAR normalized                                          normalized to 1W                 6.60 mWw /g
      SAR for nominal Body TSL parameters                     normalized to 1W        6.52 mW / g #16.5 % (k=2)




Certificate No: DB35V2—4d021_Apr10                     Page 4 of 9


Appendix

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           51.5 Q —2.5 |Q

      Return Loss                                                                       — 31.0 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           48.1 Q — 3.9 jQ

      Return Loss                                                                       —27.1 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.392 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                 April 22, 2004




Certificate No: D835V2—4d021_Apr10                       Page 5 of 9


DASY5 Validation Report for Head TSL

                                                                        Date/Time: 21.04.2010 10:38:05

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN:4d021

Communication System: CW; Frequency: 835 MHz; Duty Cycle: 1: 1
Medium: H$L900
Medium parameters used: f = 835 MHz; 0 = 0.89 mho/m; & = 41.8; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)

DASYS5 Configuration:

        Probe: ES3DV3 — SN3205; ConvF(6.04, 6.04, 6.04); Calibrated: 26.06.2009

        Sensor—Surface: 3mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 02.03.2010

        Phantom: Flat Phantom 4.9L; Type: QDOOOP49AA; Serial: 1001

        Measurement SW: DASY5, V5.2 Build 162; SEMCAD X Version 14.0 Build 57


Pin=250 mW /d=15mm, dist=3.0mm (ES—Probe)/Zoom Scan (7x7x7) /Cube 0: Measurement
grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 57.3 V/m; Power Drift = 0.018 dB
Peak SAR (extrapolated) = 3.55 W/kg
SAR(I g) = 2.37 mW/g; SAR(10 g) = 1.55 mW/g
Maximum value of SAR (measured) = 2.77 mW/g




                            0 dB = 2.77mW/g


Certificate No: D835V2—4d021_Apr10             Page 6 of 9


Impedance Measurement Plot for Head TSL



                                                               21 fipr 2010       @9:10:02
     (CHI) s1     1 U FS             4: 51.45%a   —24531 8   77.699 pF          835.000 000 MHz




     Del

     Cor




     fAvg
     16




     CH2




     Cor




     Av
     16 *




                START 635.000   000 MHz                                  STOP 1 109.000 000 MHz




Certificate No: D835V2—4d021_Apr10                       Page 7 of 9


DASY5 Validation Report for Body
                                                                        Date/Time: 29.04.2010 13:27:42

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN:4d021

Communication System: CW; Frequency: 835 MHz; Duty Cycle: 1: 1
Medium: MSL900
Medium parameters used: f = 835 MHz; 0 = 0.99 mho/m; &, = 54.9; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)

DASYS5 Configuration:

    e   Probe: ES3DV3 — SN3205; ConvF(5.97, 5.97, 5.97); Calibrated: 26.06.2009

    e   Sensor—Surface: 3mm (Mechanical Surface Detection)

    e   Electronics: DAE4 Sn601; Calibrated: 02.03.2010

    e   Phantom: Flat Phantom 4.9L; Type: QDOOOP49AA; Serial: 1001

    e   Measurement SW: DASY5, V5.2 Build 162; SEMCAD X Version 14.0 Build 57


Pin250 mW /d=15mm, dist=3.0mm (ES—Probe)/Zoom Scan (7x7x7) /Cube 0: Measurement
grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 56.1 V/m; Power Drift = 0.030 dB
Peak SAR (extrapolated) = 3.73 W/kg
SAR(1 g) = 2.52 mW/g; SAR(10 g) = 1.65 mW/g
Maximum value of SAR (measured) = 2.93 mW/g




                  12



                  18


                           0 dB = 2.93mW/g


Certificate No: D835V2—4d021_Apr10             Page 8 of 9


Impedance Measurement Plot for Body TSL



                                                                 29 fApr 2010      10:04:42
     (CHI) sii     1 U Fs            1: 48.107 a   —3.9023 6   48.844 pF          835.000 000 MHz




                                                                                                    CH1 Markers
     Del
                                                                                                     2: 58.350 a
                                                                                                       32.508 a
     Cor                                                                                            900.000 MHz




     fvg
     16




     CH2




                                                                                                    CH2 Markers

     Cor                                                                                            21—10,592 d8
                                                                                                    $00.000 MHz




     fv
     16°



                 START 635.000 000 MHz                                     STOP 1 100.000 000 MHz




Certificate No: D835V2—4d021_Apr10                        Page 9 of 9



Document Created: 2011-03-31 10:20:41
Document Modified: 2011-03-31 10:20:41

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC