SAR exemption analysis version E

FCC ID: YSGVR1240

RF Exposure Info

Download: PDF
FCCID_2782225

                 FCC SAR report analysis for SORIN Platinium
                        Implantable Medical Devices

        Reference:                      131328—664446—AR—SAR



        Version:                        E



        Status:                         Approved




        Date:                           06/0ct/2015



        Customer:                        SORIN CRM
                                        Parc d‘Affaires NOVEO
                                        4 avenue Réaumur
                                        92143 Clamart cedex


        Products:                       SORIN Platinium cardioverter defibrillator




        Author:                         Zorica COROLLEUR



              4
        Technical Manager:.             Marc CANCOUET                            P § C_'] ,         ‘T)s (')R &  50 D
                                                                                                              nrdelemrre

                                                                        HaF             INDUSTRIES              D
                                                                         *)             S.A.S au capital de 15.7
                                                                             4          RCS Nanterre B 408379
                                                                     [{—e—ice] 33 avenue du Cénéradt Leclere
                                                                               F — 92266 EONTENAY AUX ROSES


                               This document shall not be reproduced, except in full, without the written approval of the LCIE.
This document contains results related only to the items tested. It does not imply the conformity of the whole production to the items tested.




      LCW                                 §3 n du Genéni Leclere                  el—    +33 1 10 95 60 60    Ssociel¢ Anony me

      Labormoire Centrsd                  BP 8                                   Fux     +423 1 40 95 8h 36   au capriad de 15. 715 984 €
      des Iocusivies Eleciriques          922366 Ponteniv—aius Roses cedes       conmeriélcte 1               RCS Nanterre B 108 463 171
      tine sociéte de Bureau Yeritas      branc                                  www dore h


                         PUBLICATION HISTORY
VERSION      DATE           AUTHOR                MODIFICATION
A            25-Aug-15      Z.COROLLEUR           Creation of document
B            15-Sep-15      Z.COROLLEUR           Corrected FCC ID and Adding information on
                                                  RF conducted output power
C            23-Sep-15      Z.COROLLEUR           Adding information about conducted power
                                                  measurement
D            29-Sep-15      Z.COROLLEUR           Adding some internal photos
E            06-Oct-15      Z.COROLLEUR           Modification of document




              FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

    131328-664446-AR-SAR                     E                    06/10/2015            Page 2/18


                                                               CONTENTS

1.      INTRODUCTION ........................................................................................................................... 4

2.      RELATED DOCUMENTS ............................................................................................................. 5
     2.1.      APPLICABLE STANDARDS ................................................................................................ 5
     2.2.      REFERENCE DOCUMENTS ................................................................................................ 5

3.      SORIN PLATINIUM IMD ............................................................................................................... 6
     3.1.      GENERAL DESCRIPTION OF SYSTEM.............................................................................. 6
     3.2.      GENERAL DESCRIPTION OF IMPLANTABLE MEDICAL DEVICE .................................. 7
     3.3.      SCHEMATIC OF IMD ............................................................................................................ 9
     3.4.      ANTENNA CHARACTERISTIC OF IMD ............................................................................ 10
     3.5.      CHARACTERISTICS OF IMD............................................................................................. 11

4.      TESTS RESULTS ....................................................................................................................... 12
     4.1.      RESULTS OF RADIATED OUTPUT POWER .................................................................... 12
     4.2.      RF CONDUCTED OUTPUT POWER ................................................................................. 12
        4.2.1        RUNNING MODE ......................................................................................................... 12
        4.2.2        TEST CONDITIONS ..................................................................................................... 12
        4.2.3        TEST SETUP ............................................................................................................... 13
        4.2.4        The access point for these RF measurements ........................................................ 14
        4.2.5        TEST EQUIPMENT LIST ............................................................................................. 16
        4.2.6        GRAPHICS & RESULTS ............................................................................................. 16

5.      CONCLUSION ............................................................................................................................ 18




                                FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

                131328-664446-AR-SAR                                       E                             06/10/2015                            Page 3/18


1.       INTRODUCTION
          This document presents the results of tests performed on 8 SORIN PLATINIUM
          Implantable Medical Devices (IMD) to prove that all models are excluded of SAR
          evaluation.

          The SORIN PLATINIUM Implantable Medical Devices (IMD) is Compliant according to
          FCC part 95I standards.

  Hereafter is the list of models name and FCC ID:

         Models                                               FCC ID
         DR 1540                                        FCC ID YSGDR1540
         VR 1240                                        FCC ID YSGVR1240
       CRT-D 1741                                      FCC ID YSGCRTD1741
     SonR CRT-D 1841                                 FCC ID YSGCRTDSOR1841
         DR 1510                                        FCC ID YSGDR1510
         VR 1210                                        FCC ID YSGVR1210
       CRT-D 1711                                      FCC ID YSGCRTD1711
     SonR CRT-D 1811                                 FCC ID YSGCRTDSOR1811




            FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

 131328-664446-AR-SAR                      E                    06/10/2015            Page 4/18


        2.      RELATED DOCUMENTS



        2.1.    APPLICABLE STANDARDS
[A1]   KDB 865664 D02       RF Exposure Compliance Reporting and Documentation Considerations
[A2]   47 CFR Part95I       Medical Device Radiocommunication Service (MedRadio)
[A3]   47 CFR PART15        Radio frequency devices


        2.2.    REFERENCE DOCUMENTS


[R1]   131328-664446E                    Radio Test report of DR 1540
[R2]   131328-664446F                    Radio Test report of VR 1240
[R3]   131328-664446G                    Radio Test report of CRT-D 1741
[R4]   131328-664446H                    Radio Test report of SonR CRT-D 1841
[R5]   131328-664445E                    Radio Test report of DR 510
[R6]   131328-664445F                    Radio Test report of VR 1210
[R7]   131328-664445G                    Radio Test report of CRT-D 1711
[R8]   131328-664445H                    Radio Test report of SonR CRT-D 1811
[R9]   MISC2583B                         Tune up procedure




                    FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

          131328-664446-AR-SAR                     E                    06/10/2015            Page 5/18


3.     SORIN PLATINIUM IMD

3.1.   GENERAL DESCRIPTION OF SYSTEM


         The purpose of the paragraph is to provide a high-level description of the structure and
         operation of the following sub-systems of Sorin Remote Monitoring System: the Smartview
         Monitor and the Implantable Medical Device.

         The Smartview Monitor (SM) is intended to collect patient’s clinical data from an
         Implantable Medical Device (IMD) and transfer them to data management system (Back
         Office server).

         The IMD is implanted into the patient’s body. The Smartview Monitor is installed at patient
         home and is intended to collect data from the IMD remotely in absence of physician
         according to scheduled operation.




         The Implantable Medical Device uses two wireless RF bands:
             ISM band (2.45 GHz) for communication initialization (implant wake-up) receiver
                only
             MedRadio (402-405 MHz) band for data transfer Implantable Medical Device (IMD)
                and transfer them to data management system (Back Office server).




           FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

 131328-664446-AR-SAR                     E                    06/10/2015               Page 6/18


       3.2.    GENERAL DESCRIPTION OF IMPLANTABLE MEDICAL
               DEVICE


                 PLATINIUM IMD is declined in 8 models:

                                               Models
                                               DR 1540
                                               VR 1240
                                             CRT-D 1741
                                           SonR CRT-D 1841
                                               DR 1510
                                               VR 1210
                                             CRT-D 1711
                                           SonR CRT-D 1811



                 All models are an implantable cardioverter defibrillator for the recognition and treatment of
                 ventricular tachycardia and fibrillation, with ventricular resynchronization, in patients with
                 spontaneous or inducible tachyarrhythmias.
                 Each IMD is equipped with an accelerometer to allow adaptation of pacing to suit the
                 patient’s activity, and provide high energy shocks (42 J) for enhanced safety, as well as
                 automatic lead measurements to monitor system integrity.
                 PLATINIUM IMD is protected against high-frequency signals emitted by cellular
                 telephones.

                 Device and lead connections:

PLATINIUM CRT-D 1711                                   3*IS-1 bipolar (LV unipolar compatible), 2*DF-1

PLATINIUM CRT-D 1741                                   2*IS-1 bipolar (LV unipolar compatible), 1*DF4
PLATINIUM SonR CRT-D 1811                              2*IS-1 bipolar (LV unipolar compatible), 2*DF-1,
                                                       1*SonR (IS-1bipolar compatible)
PLATINIUM SonR CRT-D 1841                              1*IS-1 bipolar (unipolar compatible), 1*DF4, 1*SonR
                                                       (IS-1 bipolar compatible)
PLATINIUM DR 1510                                      2*IS-1 bipolar, 2*DF-1
PLATINIUM DR 1540                                      1*IS-1 bipolar, 1*DF4
PLATINIUM VR 1210                                      1*IS-1 bipolar, 2*DF-1
PLATINIUM VR 1240                                      1*DF4




                    FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

         131328-664446-AR-SAR                      E                    06/10/2015                 Page 7/18


                                                              Implantable
                                          Implantable         cardioverter
Type     Model        FCC Number          cardioverter         defibrillator       photos       Leads connection
                                           defibrillator      with cardiac
                                                           resynchronization



                        FCC ID
        DR 1510                                 X
                      YSGDR1510




                        FCC ID
       CRT-D1711     YSGCRTD1711                                  X




DF1

                         FCC ID
        VR 1210        YSGVR1210                X




                        FCC ID
       SonR CRT-D   YSGCRTDSOR181
                                                                  x
          1811            1




                        FCC ID
        DR 1540       YSGDR1540                 X




                        FCC ID
DF4    CRT-D1741                                                  X
                     YSGCRTD1741




                         FCC ID
        VR 1240                                 x
                       YSGVR1240




                      FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

            131328-664446-AR-SAR                       E                   06/10/2015           Page 8/18


                 FCC ID
SonR CRT-D   YSGCRTDSOR184
                                                           X
   1841            1




             DF-1 refers to the international standard for defibrillation lead connectors (ISO
             11318:2002).
             DF4 refers to the international standard for defibrillation lead connectors (ISO 27186:2010).




   3.3.      SCHEMATIC OF IMD

       All models are using the same architecture: Battery, Capacitor and RF hardware. The only
       difference is the dimension of antenna and the numbers of leads.




               FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

     131328-664446-AR-SAR                     E                    06/10/2015                 Page 9/18


3.4.   ANTENNA CHARACTERISTIC OF IMD
         The RF antenna consists in a gold wire loop of 0.4 mm diameter. It is the same for the
         SonR CRT-D, the CRT-D, the DR and the VR header of the DF1 family. The dimensions of
         antennas are illustrated in the following figures.




                     RF antenna of the DF1 SonR, Crtd, Dr and Vr headers




                     RF antenna of the DF4 SonR, Crtd, Dr and Vr headers




           FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

 131328-664446-AR-SAR                     E                    06/10/2015            Page 10/18


         3.5.      CHARACTERISTICS OF IMD

 Software identification:
-Software version: ROM V2

 Equipment information:
- Modulation: 2FSK
- Transmit operating mode:            Multiples antenna
                                      Single antenna

- Number of transmit chains:          1          2

- Number of receiver chains:          1          2

- Antenna type:                       Integral       External

- Type of the equipment:              Stand-alone equipment           Plug-in radio device     Combined
equipment

- Temperature range:      Tmin:       -20°C          0°C              25°C
                          Tnom:        37°C
                          Tmax:       +35°C          55°C             45°C

- Test source voltage:    Vmin:       207V/50Hz      2.5Vdc
                          Vnom:       230V/50Hz      2.62Vdc
                          Vmax:       253V/50Hz      3.24Vdc

- Type of power source:               Battery (Lithium-Ion)           Internal power supply
                                      External power supply           Car Charger

- Test sequence/test software used:
- Duty Cycle:                     Continuous duty        Intermittent duty       Continuous operation
- Equipment type:                 Representative production model         Pre-production model

Operating frequency range:
                         Frequency Band (MHz)
     2400MHz to 2483,5MHz
      5150MHz to 5350MHz
      5470MHz to 5725MHz
       402MHz to 405MHz

-Channel plan:
   Channel               Frequency (MHz)
  Cmin: 0                    402.15
    1                        402.45
    2                        402.75
    3                        403.05
    4                        403.35
  Cnom:5                     403.65
    6                        403.95
    7                        404.25
    8                        404.55
  Cmax: 9                    404.85


                         FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

             131328-664446-AR-SAR                         E                  06/10/2015              Page 11/18


     4.       TESTS RESULTS

     4.1.     RESULTS OF RADIATED OUTPUT POWER
    Below are the measured Radiated EIRP Powers, for details please refer to EMC test report.

                                                       Reference to test     Max measured    EIRP Limit
    Models                      FCC ID                                         EIRP(µW)         (µW)
                                                            Report
    DR 1540             FCC ID YSGDR1540               131328-664446E            0,047           25
    VR 1240             FCC ID YSGVR1240               131328-664446F            0,075           25
  CRT-D 1741           FCC ID YSGCRTD1741              131328-664446G            0,047           25
SonR CRT-D 1841      FCC ID YSGCRTDSOR1841             131328-664446H            0,059           25
    DR 1510             FCC ID YSGDR1510               131328-664445E            0,047           25
    VR 1210             FCC ID YSGVR1210               131328-664445F            0,059           25
  CRT-D 1711           FCC ID YSGCRTD1711              131328-664445G             0,076          25
SonR CRT-D 1811      FCC ID YSGCRTDSOR1811             131328-664445H             0,076          25



     4.2.     RF CONDUCTED OUTPUT POWER



     4.2.1 RUNNING MODE

        The EUT is set in the following mode during tests:

        - Continued transmission with modulation on an assigned channel at the highest power



     4.2.2 TEST CONDITIONS

        Test performed by                        : Mathieu CERISIER
        Date of test                             : 2015/07/31
        Ambient temperature                      : 26
        Relative humidity                        : 32%




                   FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

       131328-664446-AR-SAR                       E                    06/10/2015               Page 12/18


        4.2.3 TEST SETUP
           - The Equipment under Test is installed:

              In the climatic chamber
              On a table
              In an anechoic chamber


           -Measurement is performed with a spectrum analyzer

              On the EUT 50 ohms conducted access
              With a test fixture

           -Spectrum analyzer setup detail:
           RBW: 1 MHz
           VBW: 3 MHz
           Detector type: peak
           Sweep time: 1.9µs


           The spectrum analyzer marker peak function is used to find the maximum RF conducted output
           power.

           Since all 8 models use the same RF circuit and have identical RF characteristics (such as output
           power, Tx frequencies and modulations…etc) the tests were performed directly on one identical
           representative RF circuits mounted on SORIN workbench (note: SORIN workbench is a special
           test fixture provided by SORIN for controlling the testing mode and access to the EUT RF port for
           SORIN Implantable Medical Devices series and as shown below in the picture)




Photograph for RF conducted output power


                     FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

          131328-664446-AR-SAR                        E                  06/10/2015             Page 13/18


4.2.4 THE ACCESS POINT FOR THESE RF MEASUREMENTS

  All 8 models are using the same architecture: Battery, Capacitor and RF hardware. The only
  difference is the dimension of the antenna and the numbers of leads.
  The hardware components are soldered on flex PCB. The flex PCB is folded and integrated in the
  IMD. Therefore testing one representative RF circuit is enough to cover all 8 modules’ RF
  conducted characteristics.




         The flex PCB is folded and
         integrated in IMD.




         The access point for these RF measurements is illustrated above.




          FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

131328-664446-AR-SAR                     E                    06/10/2015            Page 14/18


                                                                        The flex PCB is folded
                                                                        and integrated in IMD.




                               Internal photo of EUT




                   BOTTOM VIEW of Flex PCB
                                                                TOP VIEW of Flex PCB




                                     Zoom on UFL connector




          FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

131328-664446-AR-SAR                     E                    06/10/2015               Page 15/18


              HYBRID RF PART (BOTTOM VIEW)                                     HYBRID RF PART (TOP VIEW)




               RF feedthrough connection                                   RF feedthrough connection



      Above figures show RF feedthrough pads on PCB for antenna connection and all RF traces on the
      PCB were designed for standard 50 ohm system. For reliable RF measurements, SORIN has
      installed a UFL connector directly on the antenna contact point for us to verify the actual RF output
      power to the antenna. The above setup pictures also show that the measurement system uses 50
      ohm coaxial cable for connecting the EUT to spectrum analyzer. Therefore, the load impedance
      seen by the EUT (Equipment Under Test) is at optimum 50 ohm for max power output and cable
      loss has been considered and added back to the reading.


  4.2.5 TEST EQUIPMENT LIST
 DESCRIPTION                       MANUFACTURER                    MODEL       N° LCIE      Cal_Date   Cal_Due
  EMI test receiver                        R&S                     ESR         A2642023      03/2015   03/2016
                                      Teledyne & MINI         920-0202-024 &
RF cable & Attenuator                                                          A5329674      10/2014   10/2015
                                         CIRCUITS                FW-20+




  4.2.6 GRAPHICS & RESULTS
      Since all 8 models share the same RF circuit and have identical RF characteristics, the following
      conducted power measurement result is applicable to all 8 models.
                                                            Tnom
                                                            Cmin




                      FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

    131328-664446-AR-SAR                                E                      06/10/2015              Page 16/18


                                                               Tnom
                                                               Cnom




                                                               Tnom
                                                               Cmax




                           Temperature                                                  Tnom
                              Voltage                                                   Vnom
                             Channel                                  Cmin              Cnom        Cmax
                RF conducted output power (dBm)                       -11,98            -11,59      -11,4
                 RF conducted output power (mW)                       0,063             0,069       0,072

All measurements are recorded in test reports listed in 4.1.


                          FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

             131328-664446-AR-SAR                              E               06/10/2015           Page 17/18


5.     CONCLUSION

         Since the test results, as reported in the previous section, demonstrate that both conducted
         and EIRP power plus 0.5dB tuneup tolerance are less than 1 mW, we would like to request
         FCC to grant SAR test exemption to all 8 of these applications according to FCC KDB
         447498 section 4.2.4, as reprinted below. Thank you.



         FCC:
         KDB 447498 section 4.2.4:
         4.2.4. Transmitters implanted in the body of a user
         When the aggregate of the maximum power available at the antenna port and radiating
         structures of an implanted transmitter, under all operating circumstances, is ≤ 1.0 mW,
         SAR test exclusion may be applied. The maximum available output power requirement and
         worst case operating conditions must be supported by power measurement results and
         fully justified in a SAR analysis report, in lieu of the SAR measurement or numerical
         simulation, according to design and implementation requirements of the device.




                           END OF DOCUMENT




           FCC SAR report analysis for SORIN Platinium Implantable Medical Devices

 131328-664446-AR-SAR                     E                    06/10/2015                Page 18/18



Document Created: 2015-10-07 09:06:24
Document Modified: 2015-10-07 09:06:24

© 2024 FCC.report