1900 dipole calibration report

FCC ID: YR8ES620

RF Exposure Info

Download: PDF
FCCID_2947285

      SAR Reference Dipole Calibration Report



                                       Ref: ACR.269.4.13.SATU.A




 SHENZHEN MORLAB COMMUNICATIONS
       TECHNOLOGY CO., LTD
  FL3, BUILDING A, FEIYANG SCIENCE PARK, NO.8
               LONGCHANG ROAD,
BLOCK 67, BAOAN DISTRICT, SHENZHEN, GUANGDONG
              PROVINCE, P.R. CHINA
    SATIMO COMOSAR REFERENCE DIPOLE
                      FREQUENCY: 1900 MHZ
                      SERIAL NO.: SN 30/13 DIP1G900—261

                          Calibrated at SATIMO US
                  2105 Barrett Park Dr. — Kennesaw, GA 30144




                                      |ACCREDITED|
                                     Calibration CERT #2246.02



                                      09/22/2014

    Summary:

    This document presents the method and results from an accredited SAR reference dipole calibration
    performed in SATIMO USA using the COMOSAR test bench. All calibration results are traceable
    to national metrology institutions.


                         SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.269.4.13.SATU.A
   TIMD




                                  Name                               Function                          Date                  Signature
Prepared by :                Jérome LUC                        Product Manager                    9/16/2014                   fi

Checked by :                 Jérome LUC                        Product Manager                    9/16/2014                    fi/

Approved by :          Kim RUTKOWSKI                            Quality Manager                   9/16/2014                 #pm AuthasDs




                                                                        Customer Name
                                                                       Shenzhen Morlab
                                          per        o    e            Communications
                                      Distribution :                   Technology Co..
                                                                                 Ltd




   Issue                  Date                                                          Modifications
     A                  9/16/2014                 Initial release




                                                              Page: 2/10

                This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO.
                The information contained hereinis to be used onlyfor the purpose for which it is submitted and is not to
                be released in whole or part without written approval ofSATIMO.


 ({ (_« \                        SAR REFERENCE DIPOLE CALIBRATION REPORT                                                           Ref: ACR.200.4.13.8ATU.A
SATIMD



                                                           TABLE OF CONTENTS



oolot e 4
2     Devige UAAGF T@SL . :=.—. :: .. :30 n 1) en omeerrrrene nr oo ene en idns ndra n ied rrniep n on ar en nara d Paie i rapanet 4
3     Produict DeSsCHptOH .........::...1./2:. s oo e ferrerfereiieererensanianievienineene fran id dinagarr insepossase ied 4
    3.1       General Information                                                                                                                        4

4     Measurement MEetAQOQ :.;:,::..:s:::=s:mm=m—mm ies reev ns terevn taverar ies rrs resrraenaso nb eoviar d ianaevages 5
    4.1       Return Loss Requirements                                                                                                                   W
    4.2       Mechanical Requirements                                                                                                                    5
5     MECASUFEGMECNE UNCETEAIAEY ..................).ssssevirscrsorsrerisirerr esns rrsryss 26 rarrathversrasrererntassraes 5

    5.1       Return Loss                                                                                                                                5
    5.2       Dimension Measurement                                                                                                                      5
    5.3        Validation Measurement                                                                                                                    5
6     Calibrati0ON MEASUL@MENt ReSUItS ..................0.20.0000000 e en en en e rerrrrerrereererreree e ies 6
    6.1       Return Loss                                                                                                                                6
    6.2       Mechanical Dimensions                                                                                                                      6
7     ValidatIOH MEASUFEMGAL .................... ... (tivlsccrrsrernavervenermenterervenserernensensbetirtencecbeeeeten ies 7
    7.1       Measurement Condition                                                                                                                      7
    7.2       Head Liquid Measurement                                                                                                                    7
    7.3       Measurement Result                                                                                                                         8
    74        Body Measurement Result                                                                                                                    9
§     List OF BQULDMENME——:. : : .. .. o+ :. ..200. .0. oys en vrviens en oo en ies ar in rl er ns in dn aPeca on ons on eopamaan dn oasty 10




                                                                         Page: 3/10
                       This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO.
                       The information containedhereinis to be used onlyfor the purpose for which it is submitted and is not to
                       be releasedin whole or part without written approval ofSATIMO.


    (/ us \              SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACK, 2059 4.13.8ATH .A



1             INTRODUCTION


This document contains a summary of the requirements set forth by the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards for reference dipoles used for SAR measurement system validations
and the measurements that were performed to verify that the product complies with the fore
mentioned standards.




2             DEVICE UNDER TEST




                                                     Device Under Test

         Device Type                                        COMOSAR 1900 MHz REFERENCE DIPOLE
         Manufacturer                                       Satimo
         Model                                              $1D1900
         Serial Number                                      SN 30/13 DIP1G900—261
         Product Condition (new / used)                     Used

A yearly calibration interval is recommended.

3             PRODUCT DESCRIPTION

3.1      GENERAL INFORMATION

Satimo‘s COMOSAR Validation Dipoles are built in accordance to the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards. The product is designed for use with the COMOSAR test bench
only.




                                Figure 1 — Satimo COMOSAR Validation Dipole




                                                              Page: 410

                This documentshall not be reproduced, except in full or in part, without the written approval ofSATIMO.
                The information contained hereinis to be used onlyfor the purposefor which it is submitted and is not to
                be released in whole or part without written approval ofSATIMO.


    {(    \               SAR REFERENCE DIPOLE CALIBRATION REPORT                                                   Ref: ACR.269.4.13.SATU.A



         4     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C and CEI/IEC 62209 standards provide requirements for
reference dipoles used for system validation measurements. The following measurements were
performed to verify that the product complies with the fore mentioned standards.

         4.1   RETURN LOSS REQUIREMENTS
The dipole used for SAR system validation measurements and checks must have a return loss of —20
dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
with the phantom constucted as outlined in the fore mentioned standards.

         4.2    MECHANICAL REQUIREMENTS
The IEEE Std. 1528 and CEI/IEC 62209 standards specify the mechanical components and
dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
mm phantom shell thickness.


5              MEASUREMENT UNCERTAINTY

All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
Measurement Uncertainty.

         5.1   RETURNLOSS
The following uncertainties apply to the return loss measurement:
                      Frequency band                                    Expanded Uncertainty on Return Loss
                        400—6000MHz                                                               0.1 dB

         5.2    DIMENSION MEASUREMENT
The following uncertainties applyto the dimension measurements:
                         Length (mm)                                        Expanded Uncertainty on Length
                              3 — 300                                                            0.05 mm



         5.3   VALIDATIONMEASUREMENT
The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC ENS0361 and CEI/IEC
62209 standards were followed to generate the measurement uncertainty for validation
measurements.
                         Scan Volume                                       Expanded Uncertainty

                                  1 g                                                   20.3 %
                                 10 g                                                   20.1 %

                                                              Page: 5/10
                This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO.
                The information contained hereinis to be used onlyfor the purpose for which it is submitted and is not to
                be released in whole or part without written approval ofSATIMO.


                                  SAR REFERENCE DIPOLE CALIBRATION REPORT                                           Ref: ACR.269.4.13.SATU A




      6     CALIBRATION MEASUREMENT RESULTS

      6.1       RETURNLOSS


                       S11, 48




                                 50—                                                                                   |
     Frequency (MHz)                                       Return Loss (dB)                                Requirement (dB)
                1900                                              —26.71                                               —20

      6.2       MECHANICAL DIMENSIONS

Frequency MHz                                  L mm                                 h mm                                      d mm


                                  required            measured          required            measured             required            measured

    300                          420.0 +1 %.                          250.0 +1 %.                                6.35 +1 %.

    450                          290.0 +1 %.                          166.7 +1 %.                                6.35 £1 %.

    750                          176.0 #1 %.                          100.0 +1 %.                                6.35 +1 %.

    835                          161.0 +1 %.                           89.8 +1 %.                                3.6 +1 %.

    900                          149.0 +1 %.                           83.3 +1 %.                                3.6 +1 %.

    1450                         89.1 +1 %.                            51.7 #1 %.                                3.6 +1 %.

    1500                         80.5 +1 %.                            50.0 £1 %.                                3.6 +1 %.

    1640                         79.0 +1 %.                            45.7 +1 %.                                3.6 £1 %.

    1750                         75.2 £1%.                             42.9 +1 %.                                3.6 +1 %.

    1800                         72.0 +1 %.                            41.7 +1 %.                                3.6 +1 %.

    1900                         68.0 +1 %.             PASS           39.5 +1 %.              PASS              3.6 +1 %.             PASS
    1950                         66.3 +1 %.                            38.5 +1 %.                                3.6 +1 %.

    2000                         64.5 +1 %.                            37.5 +1 %.                                3.6 +1 %.

    2100                         61.0 +1 %.                            35.7 +1 %.                                3.6 1 %.

    2300                         55.5 £1 %.                            32.6 41 %.                                3.6 +1 %.

    2450                         51.5 £1 %.                            30.4 +1 %.                                3.6 41 %.

    2600                         48.5 +1 %.                            28.8 +1 %.                                3.6 +1 %.

    3000                         41.5 +1 %.                           25.0 £1 %.                                 3.6 +1 %.

    3500                         37.0%1 %.                             26.4 +1 %.                                3.6 £1 %.

    3700                         34.7+1 %.                            26.4 +1 %.                                 3.6 +1 %.


                                                                 Page: 6/10

                This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO
                The information contained hereinis to be used only for the purpose for which it is submitted andis not to
                be released in whole or part without written approval ofSATIMO.


                            SAR REFERENCE DIPOLE CALIBRATION REPORT                                                   Ref: ACR.269.4.13.SATU.A




           7     VALIDATION MEASUREMENT
The IEEE Std. 1528, OET 65 Bulletin C and CEI/IEC 62209 standards state that the system
validation measurements must be performed using a reference dipole meeting the fore mentioned
return loss and mechanical dimension requirements. The validation measurement must be performed
against a liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned
standards. Per the standards, the dipole shall be positioned below the bottom of the phantom, with
the dipole length centered and parallel to the longest dimension of the flat phantom, with the top
surface of the dipole at the described distance fro m the bottom surface of the phantom.

           7.1    MEASUREMENT CONDITION

Software                                                                 OPENSAR V4
Phantom                                                                  SN 20/09 SAMT7I
Probe                                                                    SN 18/11 EPG122
Liquid                                                                   Head Liquid Values: eps‘ : 39.8 sigma : 1.42
Distance between dipole center and liquid                                10.0 mm
Area scan resolution                                                     dx=8mm/dy=8mm
Zoon Scan Resolution                                                     dx=8mm/dy=8m/dz=5mm
Frequency                                                                1900 MHz
Input power                                                              20 dBm
Liquid Temperature                                                       21 °C
Lab Temperature                                                          21 °C
Lab Humidity                                                             45 %

           72    HEAD LIQUID MEASUREMENT

                           Fre,?al:-lincv         Relative permittivity (€,‘)             Conductivity (c) S/m

                                                  required           measured           required          measured

                               300               45.3 15 %                              0.87 +5 %

                               450               43.5 45 %                              0.87 +5 %

                               750               41.9 £5 %                              0.89 +5 %

                               835               41.5 +5 %                              0.90 +5 %

                               900               41.5 +5 %                              0.97 £5 %

                              1450               40.5 £5 %                              1.20 +5 %

                              1500               40.4 +5 %                              1.23 +5 %

                              1640               40.2 £5 %                              1.31 +5 %

                              1750               40.1 £5 %                              1.37 +5 %

                              1800               40.0 £5 %                              1.40 +5 %

                              1900               40.0 £5 %              PASS            1.40 +5 %            PASS

                              1950               40.0 £5 %                              1.40 +5 %

                              2000               40.0 £5 %                              1.40 +5 %

                              2100               39.8 15 %                              1.49 +5 %

                              2300               39.5 £5 %                              1.67 45 %

                              2450               39.2 +5 %                              1.80 +5 %

                              2600               39.0 15 %                              1.96 +5 %

                              3000               38.5 15 %                              2.40 +5 %

                              3500               37.9 15 %                              2.91 15 %


                                                                Page: 7/10

                  This documentshall not be reproduced, except in fiull or in part, without the written approval ofSATIMO.
                  The information contained hereinis to be used onlyfor the purpose for which it is submitted and is not to
                  be releasedin whole or part without written approval ofSATIMO.


                             SAR REFERENCE DIPOLE CALIBRATION REPORT                                              Ref: ACR.269.4.13.SATU.A




      7.3    MEASUREMENT RESULT
The IEEE Std. 1528 and CEIM/IEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation. All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power.


                             F'e;‘A“:z“cV         1 g SAR (W/ke/w)                    10 g SAR (W/kg/w)
                                             required          measured            required         measured

                                300            2.85                                  1.94

                                450            4.58                                  3.06

                                750            8.49                                  5.95

                                835            9.56                                  6.22

                                900            10.9                                  6.99

                                1450             29                                    16

                                1500           30.5                                  16.8

                                1640           34.2                                  18.4

                                1750           36.4                                  19.3

                                1800           38.4                                  20.1

                                1900           59.7           39.36 (3.94)           20.5          20.04 (2.00)

                                1950           40.5                                  20.9

                                2000           41.1                                  2141

                                2100           43.6                                  21.9

                                2300           48.7                                  23.3

                                2450           52.4                                   24

                                2600           55.3                                  24.6

                                3000           63.8                                  25.7

                                3500           67.1                                   25



             Sutce. Aodund                   Zomivta




              BVKim) BC¥ en




                                                           Page: 8/10

             This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO.
             The information contained hereinis to be used only for the purpose for whichit is submitted andis not to
             bereleasedin whole or part without written approval ofSATIMO.


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                 Ref: ACR.269.4.13.SATU .A




           74   BODY MEASUREMENT RESULT
Software                                                               OPENSAR V4
Phantom                                                                SN 20/09 SAMT7I
Probe                                                                  SN 18/11 EPG122
Liquid                                                                 Body Liquid Values: eps‘ : 52.5 sigma : 1.50
Distance between dipole center and liquid                              10.0 mm
Area scan resolution                                                   dx=8mm/dy=8mm
Zoon Scan Resolution                                                   dx=8mm/dy=8m/dz=5mm
Frequency                                                              1900 MHz
Input power                                                            20 dBm
Liquid Temperature                                                     21 °C
Lab Temperature                                                        21 °C
Lab Humidity                                                           45 %




                          F'e;‘:’z“‘y                1 g SAR (W/kg/w)                    10 g SAR (W/kg/w)
                                                         measured                             measured

                              1900                      42.36 (4.24)                         22.23(2.22)




                                          102   &

                  6    X on      ¥ ima)




                                                              Page: 9/10

                 This documentshall not be reproduced, except in full or in part, without the written approval ofSATIMO
                 The information contained hereinis to be used onlyfor the purpose for whichit is submitted andis not to
                 bereleasedin whole or part without written approval ofSATIMO.


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.269.4.13.SATU.A




       8       LIST OF EQUIPMENT



                                                 Equipment Summary Sheet

   Equipment                 Manufacturer / Identification No        Current                                             Next Calibration
   Description                   Model                        *| Calibration Date                                              Date
                                          a                                            Validated. No cal              Validated.        No   call
   SAM Phantom                       Satimo                SN—20/09—SAM71              required.                      Fequired.

                                       i                                               Validated. No cal              Validated.        No   call
COMOSAR Test Bench                 Version 3                          NA               required.                      Fequired.

  Network Analyzer           RhodeZ&ViChwarz                    SN100132                         0212014                       0212016
      Calipers                       Carrera                   CALIPER—01                        1212013                       1212016

                                       5                                               Characterized prior to |\Characterized prior to
  feeference Srobs                   satme                 Cholee SNNTSA               test. No cal required. |test. No cal required.

     Multimeter                 Keithley 2000                     1188656                        1172013                       1172016

  Signal Generator             Agilent E4438C                 MY49070581                         1212013                       12/2016

           ;                                                                           Characterized prior to |\Characterized prior to
      Amplifier                  Asthercomm                       SN 045               test. No cal required. |test. No cal required.

    Power Meter                   HP E4418A                    US38261498                        11/2013                       11/2016

   Power Sensor                 HP ECP—E26A                    US37181460                        1172013                       11/2016

   m                                                                                   Characterized prior to\Characterized prior to
 Girection=l Coupler           herde 42 10—20                      88SS5                test. No cal required. |test. No cal required.
  Temperature and
  Humidity Sensor             Control Company                     11—661—9                        372014                       372016




                                                                Page: 10/10

                  This documentshall not be reproduced, except in full or in part, without the written approval ofSATIMO.
                  The information contained herein is to be used onlyfor the purpose for which it is submitted and is not to
                  be releasedin whole or part without written approval ofSATIMO.