850 dipole calibration report

FCC ID: YR8ES620

RF Exposure Info

Download: PDF
FCCID_2947284

                                  S(('(l'iMD
                                  The microwave vision compan




      SAR Reference Dipole Calibration Report



                                        Ref: ACR.261.3.13.SATU.A




 SHENZHEN MORLAB COMMUNICATIONS
       TECHNOLOGY CO., LTD
  FL3, BUILDING A, FEIYANG SCIENCE PARK, NO.8
              LONGCHANG ROAD,
BLOCK 67, BAOAN DISTRICT, SHENZHEN, GUANGDONG
              PROVINCE, P.R. CHINA
   SATIMO COMOSAR REFERENCE DIPOLE
          FREQUENCY: 835 MHZ
          SERIAL NO.: SN 20/08 DIPCI9
                           Calibrated at SATIMO US
                  2105 Barrett Park Dr. — Kennesaw, GA 30144



                                      *a
                                      |ACCREDITED|
                                      Calibration CERT #2246.02



                                       09/22/2014

    Summary:

    This document presents the method and results from an accredited SAR reference dipole calibration
    performed in SATIMO USA using the COMOSAR test bench. All calibration results are traceable
    to national metrology institutions.


                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.261.3.13.SATU.A
SATIIV)D




                                  Name                                Function                          Date                   Signature
Prepared by :                Jérome LUC                         Product Manager                    9/18/2014                    fi’

Checked by :                  Jérome LUC                        Product Manager                    9/18/2014                    fi

Approved by:            Kim RUTKOWSKI                           Quality Manager                    9/18/2014                 / , purtmaush




                                                                         Customer Name
                                                                        Shenzhen Morlab
                                         oh.    Fane                    Communications
                                       Distribution :                   Fectnalogy Co..
                                                                                  Ltd




   Issue                  Date                                                           Modifications
     A                  9/18/2014                  Initial release




                                                              Page: 2/10

                This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO.
                The information contained herein is to be used onlyfor the purpose for which it is submitted and is not to
                be releasedin whole or part without written approval ofSATIMO.


      f     \                     SAR REFERENCE DIPOLE CALIBRATION REPORT                                                        Ref: ACR.261.3.13.SATU.A




                                                          TABLE OF CONTENTS



1      IMEOGUCEEON.:.—.———.>—>:+s:—:»sssssskssessritanrnessrynevayersatusyigne ru ce irvev esn ¢¥33s 6932¥ 83080 Pireveerrerescunn en neg 4
2      Device UnAEF T CSL .:. ssm en cear o in aen e ns dn anre d roreanvr ind en ons rone aca n n dn y 4
3      —PrOd#et DeSCrIPLON . ....;. .. :/.s .: mrvsm criainseaan prabane h on ioh d in wevg in is in InsivPrayeeyedian in behovanys 4
     3.1       General Information                                                                                                                     4

4      ~Measurenrent MetNOQ ... . ... ... »+.=.+>»>—2 moris«a32a) en vorn n enx crnin ns ns neonrshn rnnn dn do hob arenepnareging $
     4.1       Return Loss Requirements                                                                                                                5

     4.2       Mechanical Requirements                                                                                                                 $
5      Measurement UICEFEALNEY ....................c.scsvesen cce ons cersrrrrrervertensrrennerersarstrensursens in treracs $

     5.1       Return Loss                                                                                                                             5
     5.2       Dimension Measurement                                                                                                                   5
     5.3       Validation Measurement                                                                                                                  5
6      Calibratton MeaSUF@AENAE RESUIES,........................20.000seeevavrsrvererrscarrressrrerevtritritnerserass 6
     6.1       Return Loss                                                                                                                             6
     6.2       Mechanical Dimensions                                                                                                                   6
7.     ValldattoOn MEASUFEM@L ... ........ ... ...::).s y 0o renirviriae rrranirstriay ie Paviererenirensave oeseraares 7
     71        Measurement Condition                                                                                                                   7
     7.2       Head Liquid Measurement                                                                                                                 7
     7.3       Measurement Result                                                                                                                      8
     7.4       Body Measurement Result                                                                                                                 9
8      AASLOF EQUIPDMENL .. ... . :1..— :31 00000. 0 9e ie eia in en chafineay nhnh snn ranr en Perar in aieiinendre ns n aebaeg 10




                                                                        Page: 3/10

                       This documentshall
                       4      *           not be o
                                                 reproduced, except in full or in part, without the written approval ofSATIMO.
                        the information conamed       creinis to be used onlyforthe purposefor which it is submitted and is not to
                       be released in whote or part without written approval ofSATRAO.


    (({ . \                SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ret: ACB261.3 13.8ATH.A



1             INTRODUCTION


This document contains a summary of the requirements set forth by the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards for reference dipoles used for SAR measurement system validations
and the measurements that were performed to verify that the product complies with the fore
mentioned standards.




2             DEVICE UNDER TEST




                                                      Device Under Test
         Device Type                                         COMOSAR 835 MHz REFERENCE DIPOLE
         Manufacturer                                        Satimo
         Model                                               SID835
         Serial Number                                       SN 20/08 DIPCI9
         Product Condition (new / used)                      Used

A yearly calibration interval is recommended.

3             PRODUCT DESCRIPTION

3.1      GENERAL INFORMATION

Satimo‘s COMOSAR Validation Dipoles are built in accordance to the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards. The product is designed for use with the COMOSAR test bench
only.




                                 Figure 1 —Satimo COMOSAR Validation Dipole




                                                               Page: 4/10

                 This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO.
                 The information contained hereinis to be used onlyfor the purposefor which it is submitted and is not to
                 be releasedin whole or part without written approval ofSATIMO.


    (( s \                SAR REFERENCE DIPOLE CALIBRATION REPORT                                                   Ref: ACR.261.3.13.8ATU.A


       4     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C and CEI/IEC 62209 standards provide requirements for
reference dipoles used for system validation measurements. The following measurements were
performed to verify that the product complies with the fore mentioned standards.

       4.1   RETURN LOSS REQUIREMENTS
The dipole used for SAR system validation measurements and checks must have a return loss of —20
dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
with the phantom constucted as outlined in the fore mentioned standards.

       4.2      MECHANICAL REQUIREMENTS
The IEEE Std. 1528 and CEI/IEC 62209 standards specify the mechanical components and
dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
mm phantom shell thickness.


5            MEASUREMENT UNCERTAINTY

All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
Measurement Uncertainty.

       5.1      RETURNLOSS
The following uncertainties apply to the return loss measurement:
                      Frequency band                                    Expanded Uncertainty on Return Loss

                        400—6000MHz                                                               0.1 dB

       5.2      DIMENSION MEASUREMENT
The following uncertainties apply to the dimension measurements:
                         Length (mm)                                        Expanded Uncertainty on Length
                              3 — 300                                                            0.05 mm



       5.3      VALIDATIONMEASUREMENT
The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC ENS50361 and CEMIEC
62209 standards were followed to generate the measurement uncertainty for validation
measurements.
                         Scan Volume                                       Expanded Uncertainty

                                  1 g                                                   20.3 %
                                 10 g                                                   20.1 %

                                                              Page: 5/10
                This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO.
                The information contained hereinis to be used onlyfor the purpose for which it is submitted and is not to
                be releasedin whole or part without written approval ofSATIMO.


                         SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.261.3.13.8SATU .A




      6     CALIBRATION MEASUREMENT RESULTS

      6.1   RETURNLOSS




                         40


     Frequency (MHz)                                   Return Loss (dB)                                    Requirement (dB)
                835                                             =25.29                                                —20

      6.2       MECHANICAL DIMENSIONS

Frequency MHz                          L mm                                         h mm                                     d mm


                          required             measured                required             measured             required           measured

     300                420.0 +1 %.                                   250.0 £1 %.                               6.35 +1 %.

    450                 290.0 +1 %.                                   166.7 +1 %.                               6.35 +1 %.

     750                176.0 +1 %.                                   100.0 +1 %.                               6.35 +1 %.

    835                 161.0 +1 %.               PASS                89.8 +1 %.               PASS              3.6 +1 %.            PASS
     900                149.0 +1 %.                                   83.3 +1 %.                                 3.6 41 %.

    1450                 89.1 +1 %.                                   51.7 £1 %.                                 3.6 +1 %.

    1500                 80.5 +1 %.                                   50.0 £1 %.                                 3.6 £1 %.

    1640                 79.0 +1 %.                                   45.7 +1 %.                                 3.6 +1 %.

    1750                 75.2 1 %.                                    42.9 +1 %.                                 3.6 41 %.

    1800                 72.0 #1 %.                                   41.7 +1 %.                                 3.6 +1 %.

    1900                 68.0 +1 %.                                   39.5 +1 %.                                 3.6 +1 %.

    1950                 66.3 +1 %.                                   38.5 £1 %.                                 3.6 +1 %.

    2000                 64.5 +1 %.                                   37.5 +1 %.                                 3.6 +1 %.

    2100                 61.0 +1 %.                                   35.7 £1%.                                  3.6 +1 %.

    2300                 55.5 +1 %.                                   32.6 £1%.                                  3.6 +1 %.

    2450                 51.5 £1 %.                                   30.4 +1 %.                                 3.6 41 %.

    2600                 48.5 +1 %.                                   28.8 +1 %.                                 3.6 +1 %.

    3000                 41.5 +1 %.                                   25.0 +1 %.                                 3.6 +1 %.

    3500                 37.0+1 %.                                    26.4 +1 %.                                 3.6 +1 %.

    3700                 34.741%.                                     26.4 +1 %.                                 3.6 +1 %.


                                                              Page: 6/10

                This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO
                The information contained herein is to be used onlyfor the purpose for whichit is submitted and is not to
                be released in whole or part without written approval ofSATIMO.


                            SAR REFERENCE DIPOLE CALIBRATION REPORT                                                   Ref: ACR.261.3.13.SATU.A




           7     VALIDATION MEASUREMENT
The IEEE Std. 1528, OET 65 Bulletin C and CEI/IEC 62209 standards state that the system
validation measurements must be performed using a reference dipole meeting the fore mentioned
return loss and mechanical dimension requirements. The validation measurement must be performed
against a liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned
standards. Per the standards, the dipole shall be positioned below the bottom of the phantom, with
the dipole length centered and parallel to the longest dimension of the flat phantom, with the top
surface of the dipole at the described distance from the bottom surface of the phantom.

           7.1    MEASUREMENT CONDITION

Software                                                                 OPENSAR V4
Phantom                                                                  SN 20/09 SAMT7I
Probe                                                                    SN 18/11 EPG122
Liquid                                                                   Head Liquid Values: eps‘ : 42.6 sigma : 0.88
Distance between dipole center and liquid                                15.0 mm
Area scan resolution                                                     dx=8mm/dy=8mm
Zoon Scan Resolution                                                     dx=8mm/dy=8m/dz=5mm
Frequency                                                                835 MHz
Input power                                                              20 dBm
Liquid Temperature                                                       21 °C
Lab Temperature                                                          21 °C
Lab Humidity                                                             45 %

           7.2    HEAD LIQUID MEASUREMENT

                           Fre'?/‘t:zncy          Relative permittivity (s,‘)             Conductivity (6) S/m

                                                  required           measured           required          measured

                               300               45.3 45 %                              0.87 +5 %

                               450               43.5 +5 %                              0.87 +5 %

                               750               41.9 +5 %                              0.89 +5 %

                               835               41.5 45 %              PASS            0.90 +5 %            PASS

                               900               41.5 +5 %                              0.97 +5 %

                              1450               40.5 £5 %                              1.20 +5 %

                              1500               40.4 +5 %                              1.23 +5 %

                              1640               40.2 +5 %                              1.31 +5 %

                              1750               40.1 +5 %                              1.37 +5 %

                              1800               40.0 +5 %                              1.40 £5 %

                              1900               40.0 £5 %                              1.40 +5 %

                              1950               40.0 +5 %                              1.40 #5 %

                              2000               40.0 +5 %                              1.40 +5 %
                              2100               39.8 15 %                              1.49 +5 %

                              2300               39.5 45 %                              1.67 +5 %
                              2450               39.2 +5 %                              1.80 +5 %

                              2600               39.0 +5 %                              1.96 +5 %

                              3000               38.5 45 %                              2.40 +5 %

                              3500               37.9 +5 %                              2.91 +5 %


                                                                Page: 7/10

                  This document shall not be reproduced, except in fiull or in part, without the written approval ofSATIMO.
                  The information contained hereinis to be used onlyfor the purposefor which it is submitted and is not to
                  be releasedin whole or part without written approval ofSATIMO.


sarimo
 ((( s \               SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.261.3.13.SATU.A




      7.3   MEASUREMENT RESULT
The IEEE Std. 1528 and CEIIEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation. All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power.

                      F'E:A‘:Z"CV                1 g SAR (W/kg/w)                    10 g SAR (W/kg/w)
                                            required           measured            required         measured

                          300                  2.85                                  1.94

                          450                  4.58                                  3.06

                          750                  8.49                                  5.55

                          835                  9.56            9.68 (0.97)           6.22           6.29 (0.63)

                          900                  10.9                                  6.99

                         1450                   29                                    16

                         1500                  50.5                                  16.8

                         1640                  34.2                                  18.4

                         1750                  36.4                                  19.3

                         1800                  38.4                                  20.1

                         1900                  39.7                                  20.5

                         1950                  40.5                                  20.9

                         2000                  41.1                                  21.1

                         2100                  43.6                                  21.9

                         2300                  48.7                                  23.3

                         2450                  52.4                                   24

                         2600                  55.3                                  24.6

                         3000                  63.8                                  25.7

                         3500                  67.1                                   25




                °* Xioni CCC ¥iel




                                                           Page: 8/10

             This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO.
             The information contained hereinis to be used only for the purposefor whichit is submitted andis not to
             be releasedin whole or part without written approval ofSATIMO.


 ((L..X
SATIMGO
                            SAR REFERENCE DIPOLE CALIBRATION REPORT                                                 Ref: ACR.261.3.13.SATU.A




           74   BODY MEASUREMENT RESULT
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAMT7I
Probe                                                                   SN 18/11 EPG122
Liquid                                                                  Body Liquid Values: eps‘ : 55.3 sigma : 0.96
Distance between dipole center and liquid                               15.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=5mm
Frequency                                                               835 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %




                            F'e:fl‘;ez"“’             1 g SAR (W/kg/w)                     10 g SAR (W/ke/w)
                                                          measured                             measured

                                  835                    10.04 (1.00)                          6.52 (0.65)




                                                                                                               l _
       smus




                  f    X ind (o     ¥ ima)




                                                               Page: 910

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO
                 The information contained hereinis to be used onlyfor the purpose for whichit is submitted andis not to
                 be released in whole or part without written approval of SATIMO.


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.261.3.13.SATU.A




       8    LIST OF EQUIPMENT



                                                Equipment Summary Sheet

    Equipment                Manufacturer / Identification No                                  Current                 Next Calibration
   Description                      Model                                           ‘| Calibration Date                        Date
                                         §                                            Validated. No cal              Validated.        No   call
   SAM Phantom                       Satimo                SN—20/09—SAM71             required.                      required.

                                       ;                                              Validated. No cal              Validated.        No   cal
COMOSAR Test Bench                 Version 3                         NA               Fequired.                      Fequired.

  Network Analyzer           RhO4e8Schwarz                      SN100132                        0212014                       0212016
      Calipers                       Carrera                  CALIPER—01                        1212013                       1212016

                                       :                                               Characterized prior to\Characterized prior to
  Reference Fehs                     saume                Chetes SNNST                 test. No cal required. |test. No cal required.

     Multimeter                 Keithley 2000                     1188656                       1172013                       1172016

  Signal Generator             Agilent E4438C                 MY49070581                        12/2013                       1212016

           R                                                                           Characterized prior to\Characterized prior to
      Ampiifie!                  Asthercomin                      SN 0%S5              test,. No cal required. |test. No cal required.

    Power Meter                  HP E4418A                    US38261498                        1112013                       1172016

   Power Sensor                 HP ECP—E26A                    US37181460                        11/2013                      11/2016

   e                                                                                   Characterized prior to |\Characterized prior to
 @irectiohel Couplen           Nerde 4215—20                       O1SSs               test. No cal required. |test. No cal required.
  Temperature and
  Humidity Sensor             Control Company                    11—661—9                        372014                       372016




                                                               Page: 10/10

                  This document shall not be reproduced, exceptin full or in part, without the written approval ofSATIMO.
                  The information contained hereinis to be used onlyfor the purpose for which it is submitted and is not to
                  be released in whole or part without written approval ofSATIMO.