Test report 75941097-05 Issue 1

FCC ID: YOM-6961-MON

Test Report

Download: PDF
FCCID_3677538

Report on the FCC Testing of the
Monica Healthcare Ltd
Interface unit. Model: Novii System Interface Unit
POD. Model: Novii System Pod
In accordance with FCC 47 CFR Parts 15 and 18
(Simultaneous Transmission)
Prepared for:                  Monica Healthcare Ltd
                               Interchange 25 business Park
                               Unit 8
                               Bostocks lane
                               Nottingham
                               NG10 5QG
                               United Kingdom

FCC ID:
YOM-6960-MON (Novii Pod) YOM -6961-MON (Novii Interface Unit)

COMMERCIAL-IN-CONFIDENCE
Date: December 2017
Document Number: 75941097-05 | Issue: 01

RESPONSIBLE FOR                       NAME                                                                   DATE                               SIGNATURE

Project Management                    Clare Wright                                                           13 December 2017

Authorised Signatory                  Matthew Russell                                                        13 December 2017
Signatures in this approval box have checked this document in line with the requirements of TÜV SÜD Product Service document control rules.

ENGINEERING STATEMENT
The measurements shown in this report were made in accordance with the procedures described on test pages. All reported testing
was carried out on a sample equipment to demonstrate limited compliance with FCC 47 CFR Parts 15 and 18 (Simultaneous
Transmission). The sample tested was found to comply with the requirements defined in the applied rules.
RESPONSIBLE FOR           NAME                                             DATE                       SIGNATURE

Testing                               Jack Tuckwell                                                          13 December 2017
FCC Accreditation
90987 Octagon House, Fareham Test Laboratory
EXECUTIVE SUMMARY
A sample of this product was tested and found to be compliant with FCC 47 CFR Part 15:2016 and FCC 47 CFR Part 18:2016.




                                           DISCLAIMER AND COPYRIGHT
                                           This non-binding report has been prepared by TÜV SÜD Product Service with all reasonable skill and care. The document is confidential to the
                                           potential Client and TÜV SÜD Product Service. No part of this document may be reproduced without the prior written approval of TÜV SÜD
                                           Product Service. © 2017 TÜV SÜD Product Service.
                                           ACCREDITATION
                                           Our UKAS Accreditation does not cover opinions and interpretations and any expressed are outside the scope of our UKAS Accreditation.
                                           Results of tests not covered by our UKAS Accreditation Schedule are marked NUA (Not UKAS Accredited).



TÜV SÜD Product Service                         TUV SUD Ltd is a                            Phone: +44 (0) 1489 558100                    TÜV SÜD Product Service
is a trading name of TUV SUD Ltd                TÜV SÜD Group Company                       Fax: +44 (0) 1489 558101                      Octagon House
Registered in Scotland at East Kilbride,                                                    www.tuv-sud.co.uk                             Concorde Way
Glasgow G75 0QF, United Kingdom                                                                                                           Fareham
Registered number: SC215164                                                                                                               Hampshire PO15 5RL
                                                                                                                                          United Kingdom



     TÜV SÜD Product Service


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




Contents
1           Report Summary .........................................................................................................................2
1.1         Report Modification Record...........................................................................................................2
1.2         Introduction....................................................................................................................................2
1.3         Brief Summary of Results .............................................................................................................3
1.4         Application Form ...........................................................................................................................4
1.5         Product Information .......................................................................................................................9
1.6         Deviations from the Standard........................................................................................................9
1.7         EUT Modification Record ..............................................................................................................9
1.8         Test Location ...............................................................................................................................10
2           Test Details ................................................................................................................................11
2.1         Radiated Spurious Emissions (Simultaneous Transmission) .....................................................11
3           Measurement Uncertainty ........................................................................................................19




      COMMERCIAL-IN-CONFIDENCE                                                                                                               Page 1 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




1          Report Summary

1.1        Report Modification Record

           Alterations and additions to this report will be issued to the holders of each copy in the form of a
           complete document.

            Issue           Description of Change                                               Date of Issue

            1               First Issue                                                         13 December 2017


                                                             Table 1

1.2        Introduction

           Applicant                                Monica Healthcare Ltd
           Manufacturer                             Monica Healthcare Ltd
           Model Number(s)                          Interface and Pod
           Serial Number(s)                         Interface: TA1772
                                                    Pod: AA5425, AA5431
           Hardware Version(s)                      Interface Rev L
                                                    Pod Rev H
           Software Version(s)                      Interface V2.71
                                                    Pod V2.54
           Number of Samples Tested                 1 interface and 1 pod
           Test Specification/Issue/Date            FCC 47 CFR Parts 15: 2016
                                                    FCC 47 CFR Parts 18: 2016
           Order Number                             Issue 2 501559
           Date                                     30-November-2017
           Date of Receipt of EUT                   04-December-2017
           Start of Test                            07-December-2017
           Finish of Test                           07-December-2017
           Name of Engineer(s)                      Jack Tuckwell
           Related Document(s)                      ANSI C63.10 (2013)




      COMMERCIAL-IN-CONFIDENCE                                                                       Page 2 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




1.3       Brief Summary of Results

          A brief summary of the tests carried out in accordance with FCC 47 CFR Parts 15 and 18 is shown below.

           Section                Specification Clause             Test Description                            Result   Comments/Base Standard

                           FCC Part 15           FCC Part 18

           Configuration and Mode: Wireless Charging + Bluetooth Transmit
           2.1            15.247 (d) and          18.305(b)        Radiated Spurious Emissions (Simultaneous   Pass     ANSI C63.10
                              15.205                               Transmission)


                                                                                       Table 2




      COMMERCIAL-IN-CONFIDENCE                                                                                                              Page 3 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




1.4        Application Form

                                                        EQUIPMENT DESCRIPTION
 Model Name/Number                             Novii Interface Unit

 Part Number                                   107-PT-001

 Hardware Version                              Rev L
 Software Version                              Rev 2.71

 FCC ID (if applicable)                                     YOM-6961_MON

 Industry Canada ID (if applicable)

 Technical Description (Please provide a brief              The Novii System Interface Unit is part of the Novii Wireless Patch System: a
 description of the intended use of the equipment)          Maternal/Fetal monitor that records Fetal heart rate, Maternal Heart Rate and
                                                            Uterine Contractions from a pregnant subject.


                                                        INTENTIONAL RADIATORS

                                Conducted
                                                                                                                      Test Channels (MHz)
                  Frequency      Declared        Antenna        Supported                               ITU
                                                                                      Modulation
  Technology         Band         Output          Gain         Bandwidth (s)                          Emission
                                                                                      Scheme(s)
                    (MHz)         Power           (dBi)           (MHz)                              Designator
                                                                                                                  Bottom    Middle    Top
                                  (dBm)

  Bluetooth      2042-2408            10             1.18         1 MHz/                V2.1+        1M00F1D       2402      2440     2480
  (Left Side)                                                     channel               EDR

  Bluetooth      2042-2408            10             3.24         1 MHz/                V2.1+        1M00F1D       2402      2440     2480
 (Right Side)                                                     channel               EDR

  QI Wireless       0.110 to          37                              0.095           Frequency
    Charger          0.205                                                            Modulation
  transmitter




                                                       UN-INTENTIONAL RADIATOR

 Highest frequency generated or used in the device or on which
                                                                              26MHz
 the device operates or tunes

 Lowest frequency generated or used in the device or on which
                                                                              0Hz
 the device operates or tunes


                                                               Power Source
                               Single Phase                             Three Phase                               Nominal Voltage
 AC
                                      x                                                                               100-240
                                          Nominal Voltage                                               Maximum Current
 External DC
                                                5V                                                             2.5A
                                          Nominal Voltage                                       Battery Operating End Point Voltage
 Battery
                                               N/A
 Can EUT transmit whilst being charged?                                             Yes    No




      COMMERCIAL-IN-CONFIDENCE                                                                                                   Page 4 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




                                                          EXTREME CONDITIONS

 Maximum temperature                 30          °C                      Minimum temperature     10   °C


                                                                   Ancillaries
 Please list all ancillaries which will be used with the device.
 Novii System Pod


                                                       ANTENNA CHARACTERISTICS
       Antenna connector                                                     State impedance   Ohm
       Temporary antenna connector                                           State impedance   Ohm
       Integral antenna                   Type        Ceramic (Antenova
                                                      SRCW004)
       External antenna                   Type


I hereby declare that the information supplied is correct and complete.
Name:                 Jean-Francois Pieri
Position held:        CTO                                  Date:       12th December 2017




   COMMERCIAL-IN-CONFIDENCE                                                                            Page 5 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




                                                       EQUIPMENT DESCRIPTION

 Model Name/Number                            Novii System Interface Unit / Novii System Pod
 Part Number                                  107-PT-001 / 107-PT-003

 Hardware Version                             Interface Rev_L / Pod Rev_H

 Software Version                             Interface v2.71 / Pod v2.54
 FCC ID (if applicable)                                     Interface: YOM-6961-MON
                                                            Pod: YOM-6960-MON

 Industry Canada ID (if applicable)                         N/A
 Technical Description (Please provide a brief              The Novii Wireless Patch System is a small, reliable, accurate intrapartum
 description of the intended use of the equipment)          Maternal/Fetal monitor. It monitors the Maternal and Fetal heart rate and
                                                            Uterine Contractins during labour and delivery.


                                                           EQUIPMENT SUPPLIED
 WPT Source

 WPT Client
 WPT System (Client and source
 designed to work exclusively together)


                                                               WPT SOURCE
               Type 1      No intelligent communication transmitted wirelessly
               Type 2      Transmission is modulated including load modulation techniques where:
                                 1.                         Fundamental is < 490 kHz and ;
                                 2.                         All emissions are > 40 dB below RSS-GEN field strength limits.

               Type 3      Neither type 1 or type 2, but uses some form of modulation to transmit intelligent communication.
 Is the device intended for us in any of the following?:
               High power WPT device (e.g charging electric vehicles)

               WPT over a distance of > 10 cm
               Medical Device
               WPT source operating at a frequency > 400 MHz

 Does the device support power management transfer?                                   Yes
 Can the source and client operate at different separation distances?                 No

 Minimum Distance:         5 mm                                     Maximum Distance         5 mm
 Does the EUT contain any other wireless modules (excluding WPT device)?              Yes

 Can the device transmit secondary frequencies?                                       Yes Bluetooth
 State Frequencies:         2402 to 2480MHz



                                                           WPT SOURCE DESIGN
               Single fixed power transfer zone – single client
               Multiple fixed power transfer zone – single client

               Multiple non-fixed power transfer zone – single client
               Multiple power transfer zone – multiples clients




   COMMERCIAL-IN-CONFIDENCE                                                                                                  Page 6 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




                                                                 POWER SOURCE
            AC mains                                                 State voltage
 AC supply frequency                   (Hz)
            VAC
            Max Current
            Hz
            Single phase                                                       Three phase
 And / Or
            External DC supply
            Nominal voltage                                          5V                Max Current        2.5      A
            Extreme upper voltage                                    5.125 V
            Extreme lower voltage                                    4.875 V
 Battery
            Nickel Cadmium                                                     Lead acid (Vehicle regulated)
            Alkaline                                                           Leclanche
            Lithium                                                            Other Details:
                           Volts nominal.
 End point voltage as quoted by equipment manufacturer                                           V


                                                        FREQUENCY INFORMATION
 Frequency Range                                  0.11 to 0.205       MHz
 Channel Spacing (where applicable)
 Receiver Frequency Range (if different)                to            MHz
 Channel Spacing (if different)
 Test Frequencies*                                Bottom                             MHz        Channel Number (if applicable)
                                                  Middle                             MHz        Channel Number (if applicable)
                                                  Top                                MHz        Channel Number (if applicable)
 Intermediate Frequencies                                                      MHz
 Highest Internally Generated Frequency:                                       MHz


                                                        POWER CHARACTERISTICS

 Maximum TX power                             5              W
 Minimum TX power                                            W (if variable)
 Is transmitter intended for:
 Continuous duty                                                                                                       Yes             No
 Intermittent duty                                                                                                     Yes             No
 If intermittent state DUTY CYCLE
 Transmitter ON                                     seconds
 Transmitter OFF                                    seconds




   COMMERCIAL-IN-CONFIDENCE                                                                                                      Page 7 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




                                                       ANTENNA CHARACTERISTICS

       Antenna connector                                                     State impedance                   Ohm
       Temporary antenna connector                                           State impedance                   Ohm
       Integral antenna                   Type                               State impedance                   dBi
       External antenna                   Type                               State impedance                   dBi


                                                      MODULATION CHARACTERISTICS

       Amplitude                                                                 Frequency
       Phase                                                                     Other (please provide details):
 Can the transmitter operate un-modulated?                                                                                 Yes           No


                                                        CLASS OF EMISSION USED
                                                 ITU designation or Class of Emission:

                                                                     1
                                                      (if applicable) 2
                                                      (if applicable) 3
 If more than three classes of emission, list separately:


                                                         BATTERY POWER SUPPLY

 Model name/number                                                        Identification/Part number

 Manufacturer                                                             Country of Origin


                                                        ANCILLARIES (If applicable)
 Model name/number                                                        Identification/Part number

 Manufacturer                                                             Country of Origin


                                                            EXTREME CONDITIONS
 Extreme test voltages (Max)         5.125 /     V                        Extreme test voltages (Mix)                            V
                                     4.2
 Nominal DC Voltage                  5/4.2       V                        DC Maximum Current                         2.5         A

 Maximum temperature                 43          °C                       Minimum temperature                        10          °C

I hereby declare that the information supplied is correct and complete.
Name: Simon Branson                                              Position held: Engineering Manager
Date: 08/12/17




   COMMERCIAL-IN-CONFIDENCE                                                                                                       Page 8 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




1.5        Product Information


1.5.1      Technical Description

           The Monica Novii POD is an intrapartum Maternal/Fetal Monitor that non-invasively measures and
           displays fetal heart rate (FHR), uterine activity (UA) and maternal heart rate (MHR).

           The Novii POD acquires and displays the FHR tracing from abdominal surface electrodes that pick
           up the fetal ECG (fECG) signal. Using the same surface electrodes, the POD also acquires and
           displays the UA tracing from the uterine electromyography (EMG) signal and the MHR tracing from
           the maternal ECG signal (mECG).

           The POD is indicated for use on women who are at >36 completed weeks, in labor, with singleton
           pregnancies, using surface electrodes on the maternal abdomen.

           The Novii Patch is an accessory to the Novii POD that connects directly to the Novii POD and
           contains the surface electrodes that attach to the abdomen. The Novii Interface is an accessory to
           the Novii POD which provides a means of interfacing the wireless output of the Novii POD to the
           transducer inputs of a Maternal/Fetal Monitor.

           The Novii Interface enables signals collected by the Novii POD to be printed and displayed on a
           Maternal/Fetal Monitor and sent on to a central network, if connected.

           The Novii Interface is the WPT transmitter and was tested with the Novii POD which is a WPT
           client only device.

1.6        Deviations from the Standard

           No deviations from the applicable test standard were made during testing.

1.7        EUT Modification Record

           The table below details modifications made to the EUT during the test programme.
           The modifications incorporated during each test are recorded on the appropriate test pages.

                                                                                                            Date Modification
            Modification State   Description of Modification still fitted to EUT   Modification Fitted By
                                                                                                            Fitted
            Serial Number: TA1772

            0                    As supplied by the customer                       Not Applicable           Not Applicable

            Serial Number: AA5425
            0                    As supplied by the customer                       Not Applicable           Not Applicable


                                                                     Table 3




      COMMERCIAL-IN-CONFIDENCE                                                                                   Page 9 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




1.8        Test Location

           TÜV SÜD Product Service conducted the following tests at our Fareham Test Laboratory.

            Test Name                              Name of Engineer(s)       Accreditation

            Configuration and Mode: Wireless Charging + Bluetooth Transmit

            Radiated Spurious Emissions            Jack Tuckwell             UKAS
            (Simultaneous Transmission)


                                                                Table 4

           Office Address:

           Octagon House
           Concorde Way
           Segensworth North
           Fareham
           Hampshire
           PO15 5RL
           United Kingdom




      COMMERCIAL-IN-CONFIDENCE                                                               Page 10 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




2          Test Details

2.1        Radiated Spurious Emissions (Simultaneous Transmission)

2.1.1      Specification Reference

           FCC 47 CFR Part 15, Clause 15.247 (d) and 15.205
           FCC 47 CFR Part 18, Clause 18.305(b)

2.1.2      Equipment Under Test and Modification State

           Interface, S/N: TA1772 - Modification State 0
           Pod, S/N: AA5425 - Modification State 0

2.1.3      Date of Test

           07-December-2017

2.1.4      Test Method

           The test was performed in accordance with ANSI C63.10, clauses 6.5 and 6.6.

2.1.5      Environmental Conditions

           Ambient Temperature        18.1 °C
           Relative Humidity          41.0 %

2.1.6      Test Results

           Wireless Charging + Bluetooth Transmit

           The EUT was configured for simultaneous transmission in the following mode of operation:

            Technology                          Frequency Band                 Channel Frequency
            Wireless Charging                   100 kHz to 300 kHz             172 kHz

            Bluetooth (GFSK/DH5)                2400 MHz to 2483.5 MHz         2441 MHz


                                                Table 5 - Modes of Operation

           The Interface and POD were both configured for Bluetooth transmissions at maximum power on
           2441 MHz. The Interface wireless charger was configured in a test mode to output at maximum
           amplitude. The POD was placed on the Interface unit to exercise the WPT.




      COMMERCIAL-IN-CONFIDENCE                                                                     Page 11 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




                                                                                 * RBW   1 kHz    Marker 1 [T1 ]
                                                                                 * VBW3 kHz               76.18 dBµV/m
                              Ref       110 dBµV/m          * Att   10 dB         SWT 145 ms           9.903846154 kHz

                                  110
                                  18_305
                                                                                                                           A
                                  100

                       1 PK
                       MAXH
                                  90


                                                                                                                           TDF
                              180


                                  70



                                  60

                                                                                                                           3DB
                                  50                                                                                       AC



                                  40



                                  30



                                  20


                                  10

                              Start      9 kHz                          14.1 kHz/                        Stop    150 kHz




                      Date: 7.DEC.2017           21:11:46




                                           Figure 1 – 9 kHz to 150 kHz - Horizontal and Vertical
                                                                                 * RBW   10 kHz   Marker 1 [T1 ]
                                                                                 * VBW30 kHz              60.31 dBµV/m
                              Ref       110 dBµV/m          * Att   10 dB         SWT 100 ms         150.000000000 kHz

                                  110
                                  18_305
                                                                                                                           A
                                  100

                       1 PK
                       VIEW
                                  90


                                                                                                                           TDF
                                  80



                                  70

                              1
                                  60

                                                                                                                           3DB
                                  50                                                                                       AC



                                  40



                                  30



                                  20


                                  10

                              Start      150 kHz                            985 kHz/                      Stop    10 MHz




                      Date: 7.DEC.2017           21:14:18




                                         Figure 2 – 150 kHz to 10 MHz - Horizontal and Vertical




   COMMERCIAL-IN-CONFIDENCE                                                                                                      Page 12 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




                                                                           * RBW   10 kHz   Marker 1 [T1 ]
                                                                           * VBW30 kHz              43.94 dBµV/m
                              Ref     110 dBµV/m       * Att   10 dB        SWT 200 ms          28.653846154 MHz

                              110
                              18_305
                                                                                                                    A
                              100

                       1 PK
                       VIEW
                              90


                                                                                                                    TDF
                              80



                              70



                              60

                                                                                                                    3DB
                              50                                                                                    AC
                                                                                                           1

                              40



                              30



                              20


                              10

                              Start    10 MHz                          2 MHz/                       Stop   30 MHz




                      Date: 7.DEC.2017      21:16:37




                                       Figure 3 – 10 MHz to 30 MHz - Horizontal and Vertical




   COMMERCIAL-IN-CONFIDENCE                                                                                               Page 13 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




           Frequency        QP Level         QP Limit      QP Margin
                                                                             Angle(Deg)   Height(m)    Polarity
           (MHz)            (dBuV/m)         (dBuV/m)      (dBuV/m)
            52.724          32.1             40.0          -7.9              360          1.00         Vertical
            163.096         40.5             43.5          -3.0              211          1.00         Vertical

            295.891         39.4             46.0          -6.6              245          1.18         Horizontal
            509.438         29.1             46.0          -16.9             28           1.00         Vertical

            690.434*        48.3                                             159          1.71         Vertical
            887.668*        49.0                                             253          1.73         Horizontal


                                       Table 6 - 30 MHz to 1 GHz Emissions Results

          *Emission is above -6 dB of the limit in the restricted band of 46 dBµV/m (Quasi-Peak) however the
          frequency does not fall in a restricted band and therefore the limit is -20 dBc of which there is more
          than 6 dB margin hence the emission was not further investigated.




                                   Figure 4 - 30 MHz to 1 GHz - Horizontal and Vertical




   COMMERCIAL-IN-CONFIDENCE                                                                           Page 14 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




           Frequency (GHz)                     Result (µV/m)                            Limit (µV/m)                                Margin (µV/m)

                                       Peak               Average                   Peak              Average                 Peak            Average
           1.085                       54.82                50.17                   74.00              54.00                  19.18             3.83

           1.282*

           4.88                        54.97                46.80                   74.00              54.00                  19.03             7.20
           7.3195                      56.19                47.02                   74.00              54.00                  17.81             6.98


                                                Table 7 - 1 GHz to 25 GHz Emissions Results

          *Emission is above -6 dB of the limit in the restricted band of 74 dBµV/m (Peak) or 54 dBµV/m
          (Average) however the frequency does not fall in a restricted band and therefore the limit is -20 dBc
          of which there is more than 6 dB margin hence the emission was not further investigated.
                                                                              * RBW   1 MHz       Marker 1 [T2 ]
                                                                              * VBW20 kHz                 49.19 dBµV/m
                              Ref     120 dBµV/m          * Att   0 dB         SWT 115 ms              1.085000000 GHz

                                                                                                  Marker 2 [T2 ]
                              120
                                                                                                          48.34 dBµV/m
                                                                                                                              *
                                                                                                       1.282285714 GHz  A
                              110
                                                                                                  Marker 3 [T2 ]
                       1 PK                                                                               44.55 dBµV/m
                       VIEW
                              100                                                                      2.374571429 GHz
                                                                                                  Marker 4 [T2 ]
                       2 PK
                                                                                                          44.20 dBµV/m TDS
                       VIEW
                              90                                                                       2.465428571 GHz
                                                                                                                              PA
                                                                                                                              PS
                              80
                              FC15C_P2

                              70

                                                                                                                              6DB
                              60                                                                                              AC

                              FC15C_A2
                                1          2
                              50
                                                                                              3   4


                              40



                              30


                              20

                              Start    1 GHz                             200 MHz/                              Stop   3 GHz




                      Date: 7.DEC.2017         19:20:04




                                         Figure 5 - 1 GHz to 3 GHz - Horizontal and Vertical




   COMMERCIAL-IN-CONFIDENCE                                                                                                              Page 15 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




                                                                               RBW 1 MHz          Marker 1 [T1 ]
                                                                              * VBW30 kHz                 50.17 dBµV/m
                              Ref     80 dBµV/m           * Att   0 dB         SWT 190 ms              4.880000000 GHz

                                                                                                  Marker 2 [T1 ]
                              80
                              FC15C_P2                                                                    54.50 dBµV/m
                                                                                                                           *
                                                                                                       7.319500000 GHz     A
                              70

                       1 PK
                       VIEW
                              60
                                                                                                           2
                              FC15C_A2
                       2 PK                                       1
                       VIEW                                                                                                TDS
                              50


                                                                                                                           PS
                              40



                              30

                                                                                                                           6DB
                              20                                                                                           AC



                              10



                              0



                              -10


                              -20

                              Center    5.5 GHz                          500 MHz/                          Span    5 GHz




                      Date: 7.DEC.2017         19:57:46




                                         Figure 6 - 3 GHz to 8 GHz - Horizontal and Vertical
                                                                              * RBW   1 MHz       Marker 1 [T1 ]
                                                                              * VBW30 kHz                 49.50 dBµV/m
                              Ref     80 dBµV/m           * Att   0 dB         SWT 380 ms              9.760500000 GHz

                                                                                                  Marker 2 [T1 ]
                              80
                              FC15C_P2                                                                    53.35 dBµV/m
                                                                                                                           *
                                                                                                      14.639500000 GHz     A
                              70

                       1 PK
                       VIEW
                              60
                              FC15C_A2                                                        2
                       2 PK
                                                1                                                                          TDS
                       MAXH
                              50


                                                                                                                           PS
                              40



                              30

                                                                                                                           6DB
                              20                                                                                           AC



                              10



                              0



                              -10


                              -20

                              Start    8 GHz                              1 GHz/                          Stop    18 GHz




                      Date: 7.DEC.2017         20:14:15




                                         Figure 7 - 8 GHz to 18 GHz - Horizontal and Vertical




   COMMERCIAL-IN-CONFIDENCE                                                                                                      Page 16 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




                                                                                 * RBW   1 MHz   Marker 1 [T1 ]
                                                                                 * VBW30 kHz             61.95 dBµV/m
                                Ref     110 dBµV/m           * Att   0 dB         SWT 270 ms         19.521000000 GHz

                                110
                                                                                                                         *
                                                                                                                         A
                                100

                         1 PK
                         VIEW
                                90
                                FCC_PEAK
                         2 PK
                         VIEW                                                                                            TDS
                                80



                                70
                                FCC_AVE              1

                                60

                                                                                                                         6DB
                                50                                                                                       AC



                                40



                                30



                                20


                                10

                                Start    18 GHz                             700 MHz/                     Stop   25 GHz




                         Date: 7.DEC.2017     21:40:16




                                          Figure 8 - 18 GHz to 25 GHz - Horizontal and Vertical


          FCC 47 CFR Parts 15.247(d), 15.205, 18.305(b)

          The least stringent limits from the applicable rule parts were used to determine compliance for
          Radiated Emissions testing of multiple transmission sources.

          The least stringent applicable limit was:

           Rule Part                                 Limit
           Part 15.247 (d)                           -20 dBc
           Part 15.205                               Peak: 74 dBμV/m at 3m, Average 54 dBμV/m at 3m
           Part 18.302(b)                            15 µV/m at 300m


                                                                     Table 8 - Limit Table




   COMMERCIAL-IN-CONFIDENCE                                                                                                    Page 17 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




2.1.7     Test Location and Test Equipment Used

          This test was carried out in EMC Chamber 5.

                                                                                     Calibration
           Instrument                Manufacturer           Type No          TE No   Period        Calibration Due
                                                                                     (months)

           Antenna (Bilog)           Schaffner              CBL6143          287     24            18-Apr-2018

           Antenna (Active Loop,
                                     Rohde & Schwarz        HFH2-Z2          333     24            09-Dec-2018
           9kHz-30MHz)
           Antenna 18-40GHz
                                     Q-Par Angus Ltd        QSH 180K         1511    24            07-Dec-2018
           (Double Ridge Guide)

           Pre-Amplifier             Phase One              PS04-0086        1533    12            31-Jul-2018

           Screened Room (5)         Rainford               Rainford         1545    36            20-Dec-2017

           Turntable Controller      Inn-Co GmbH            CO 1000          1606    -             TU

           Hygrometer                Rotronic               HYGROPALM 1      2338    12            24-Oct-2018

                                                            NPS-2302-8000-
           Cable (N-N, 8m)           Rhophase                                3248    12            02-May-2018
                                                            NPS

           Compliance 5 Emissions    Schaffner              C5e Software     3275    -             Software

           EMI Test Receiver         Rohde & Schwarz        ESU40            3506    12            22-Nov-2018

           Tilt Antenna Mast         maturo Gmbh            TAM 4.0-P        3916    -             TU

           Mast Controller           maturo Gmbh            NCD              3917    -             TU

           1GHz to 8GHz Low Noise
                                     Wright Technologies    APS04-0085       4365    12            18-Oct-2018
           Amplifier

                                                            KPS-1501-2000-
           Cable (Rx, Km-Km 2m)      Scott Cables                            4526    6             22-May-2018
                                                            KPS
           Double Ridged
                                     ETS-Lindgren           3117             4722    12            17-Feb-2018
           Waveguide Horn Antenna
           Double Ridge Broadband
                                     Schwarzbeck            BBHA 9120 B      4848    12            17-Feb-2018
           Horn Antenna


                                                           Table 9

          TU - Traceability Unscheduled




   COMMERCIAL-IN-CONFIDENCE                                                                        Page 18 of 19


Document Number: 75941097-05 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




3         Measurement Uncertainty

          For a 95% confidence level, the measurement uncertainties for defined systems are:

           Test Name                                        Measurement Uncertainty
           Radiated Spurious Emissions (Simultaneous        30 MHz to 1 GHz: ± 5.2 dB
           Transmission)                                    1 GHz to 40 GHz: ± 6.3 dB


                                                       Table 10




    COMMERCIAL-IN-CONFIDENCE                                                                   Page 19 of 19



Document Created: 2017-12-13 11:50:09
Document Modified: 2017-12-13 11:50:09

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC