Test report 75941097-01 Issue 1

FCC ID: YOM-6961-MON

Test Report

Download: PDF
FCCID_3677536

Report on the FCC Testing of the
Monica Healthcare Ltd
Interface unit. Model: Novii System Interface Unit
POD. Model: Novii System Pod
In accordance with FCC 47 CFR Part 15B
Prepared for:                  Monica Healthcare Ltd
                               Interchange 25 business Park
                               Unit 8
                               Bostocks lane
                               Nottingham
                               NG10 5QG
                               United Kingdom

FCC ID:
Novii Interface-YOM-6961-MON, Novii Pod-YOM-6960-MON

COMMERCIAL-IN-CONFIDENCE
Date: December 2017
Document Number: 75941097-01 | Issue: 01

RESPONSIBLE FOR                       NAME                                                                   DATE                                SIGNATURE

Project Management                    Clare Wright                                                           13 December 2017

Authorised Signatory                  Matthew Russell                                                        13 December 2017
Signatures in this approval box have checked this document in line with the requirements of TÜV SÜD Product Service document control rules.

ENGINEERING STATEMENT
The measurements shown in this report were made in accordance with the procedures described on test pages. All reported testing
was carried out on a sample equipment to demonstrate limited compliance with FCC 47 CFR Part 15B. The sample tested was
found to comply with the requirements defined in the applied rules.
RESPONSIBLE FOR                       NAME                                                                   DATE                                SIGNATURE

Testing                               Paul Joynson                                                           13 December 2017

Testing                               Graeme Lawler                                                          13 December 2017
FCC Accreditation
90987 Octagon House, Fareham Test Laboratory
EXECUTIVE SUMMARY
A sample of this product was tested and found to be compliant with FCC 47 CFR Part 15B: 2016.


                                           DISCLAIMER AND COPYRIGHT
                                           This non-binding report has been prepared by TÜV SÜD Product Service with all reasonable skill and care. The document is confidential to the
                                           potential Client and TÜV SÜD Product Service. No part of this document may be reproduced without the prior written approval of TÜV SÜD
                                           Product Service. © 2017 TÜV SÜD Product Service.
                                           ACCREDITATION
                                           Our UKAS Accreditation does not cover opinions and interpretations and any expressed are outside the scope of our UKAS Accreditation.
                                           Results of tests not covered by our UKAS Accreditation Schedule are marked NUA (Not UKAS Accredited).




TÜV SÜD Product Service                          TUV SUD Ltd is a                           Phone: +44 (0) 1489 558100                    TÜV SÜD Product Service
is a trading name of TUV SUD Ltd                 TÜV SÜD Group Company                      Fax: +44 (0) 1489 558101                      Octagon House
Registered in Scotland at East Kilbride,                                                    www.tuv-sud.co.uk                             Concorde Way
Glasgow G75 0QF, United Kingdom                                                                                                           Fareham
Registered number: SC215164                                                                                                               Hampshire PO15 5RL
                                                                                                                                          United Kingdom



     TÜV SÜD Product Service


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




Contents
1           Report Summary .........................................................................................................................2
1.1         Report Modification Record...........................................................................................................2
1.2         Introduction....................................................................................................................................2
1.3         Brief Summary of Results .............................................................................................................3
1.4         Declaration of Build Status ............................................................................................................4
1.5         Product Information .......................................................................................................................6
1.6         Deviations from the Standard........................................................................................................6
1.7         EUT Modification Record ..............................................................................................................6
1.8         Test Location .................................................................................................................................7
2           Test Details ..................................................................................................................................8
2.1         Radiated Emissions.......................................................................................................................8
2.2         AC Power Line Conducted Emissions ........................................................................................13
3           Measurement Uncertainty ........................................................................................................17




      COMMERCIAL-IN-CONFIDENCE                                                                                                               Page 1 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




1          Report Summary

1.1        Report Modification Record

           Alterations and additions to this report will be issued to the holders of each copy in the form of a
           complete document.

            Issue           Description of Change                                               Date of Issue

            1               First Issue                                                         13 December 2017


                                                             Table 1

1.2        Introduction

           Applicant                                Monica Healthcare Ltd
           Manufacturer                             Monica Healthcare Ltd
           Model Number(s)                          Novii System Interface Unit (107-PT-001)
           Serial Number(s)                         Novii System Pod (107-PT-003)
                                                    Interface: TA1772
                                                    Pod: AA5233
           Hardware Version(s)                      Interface: Rev L
                                                    Pod: Rev H
           Software Version(s)                      Interface: V2.71
                                                    Pod: V2.54
           Number of Samples Tested                 1 System
           Test Specification/Issue/Date            FCC 47 CFR Part 15B: 2016
           Order Number                             Issue 2 501559
           Date                                     30-November-2017
           Date of Receipt of EUT                   03-December-2017 and 05-December-2017
           Start of Test                            03-December-2017
           Finish of Test                           05-December-2017
           Name of Engineer(s)                      Paul Joynson and Graeme Lawler
           Related Document(s)                      ANSI C63.4 (2014)




      COMMERCIAL-IN-CONFIDENCE                                                                       Page 2 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




1.3       Brief Summary of Results

          A brief summary of the tests carried out in accordance with FCC 47 CFR Part 15B is shown below.

           Section      Specification Clause    Test Description                                      Result   Comments/Base Standard

           Configuration and Mode: Interface + Pod – Standby
           2.1          15.109                  Radiated Emissions                                    Pass     ANSI C63.4
           2.2          15.107                  AC Power Line Conducted Emissions                     Pass     ANSI C63.4


                                                                                    Table 2




      COMMERCIAL-IN-CONFIDENCE                                                                                                     Page 3 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




1.4        Declaration of Build Status

           Novii System Interface

                                                             MAIN EUT
                                                           Novii System Interface Unit for the Novii Wireless Patch
            MANUFACTURING DESCRIPTION
                                                           System
            MANUFACTURER                                   Monica Healthcare Ltd
            MODEL NAME/NUMBER                              Novii System Interface Unit
            PART NUMBER                                    107-PT-001
            SERIAL NUMBER                                  TA1763 (TA1764, TA1762, TA1759, TA1752)
            HARDWARE VERSION                               Rev L
            SOFTWARE VERSION                               Rev 2.71
            PSU VOLTAGE/FREQUENCY/CURRENT                  100-240V / 50-60Hz / 2.5A
            HIGHEST INTERNALLY GENERATED /
                                                           26 MHz
            USED FREQUENCY
            FCC ID (if applicable)                         YOM-6961-MON
            INDUSTRY CANADA ID (if applicable)
                                                       The Novii System Interface Unit is part of the Novii Wireless
            TECHNICAL DESCRIPTION
                                                       Patch System: a Maternal/Fetal monitor that records Fetal
            (a brief description of the intended use and
                                                       heart rate, Maternal Heart Rate and Uterine Contractions from
            operation)
                                                       a pregnant subject.
            COUNTRY OF ORIGIN                          United Kingdom
                                               RF CHARACTERISTICS (if applicable)
            TRANSMITTER FREQUENCY
                                              2402 to 2480 (Bluetooth) and 0.110 to 0.205 (Qi charger)
            OPERATING RANGE (MHz)
            RECEIVER FREQUENCY OPERATING
                                              2402 to 2480 (Bluetooth) and 0.110 to 0.205 (Qi charger)
            RANGE (MHz)
            INTERMEDIATE FREQUENCIES          2440 MHz (Bluetooth) and 0.157 MHz (Qi charger)
            EMISSION DESIGNATOR(S):
                                              1M00F1D
            (i.e. G1D, GXW)
            MODULATION TYPES:                 V2.1 + EDR (GFSK. Bluetooth) and frequency modulation (Qi
            (i.e. GMSK, QPSK)                 charger)
            OUTPUT POWER (W or dBm)           10 dBm (Bluetooth), 37dBm (Qi charger)
                                SEPARATE BATTERY/POWER SUPPLY (if applicable)
            MANUFACTURING DESCRIPTION         Novii Interface Unit Power Supply
            MANUFACTURER                      XP Power
            TYPE                              AC/DC power supply
            PART NUMBER                       ACM18US05
            PSU VOLTAGE/FREQUENCY/CURRENT     100-240V / 50-60Hz / 2.5A
            COUNTRY OF ORIGIN                 China
                                           MODULES (if applicable)
            MANUFACTURING DESCRIPTION
            MANUFACTURER
            TYPE
            POWER
            FCC ID
            INDUSTRY CANADA ID
            EMISSION DESIGNATOR
            DHSS/FHSS/COMBINED OR OTHER
            COUNTRY OF ORIGIN
                                         ANCILLARIES (if applicable)
            MANUFACTURING DESCRIPTION         Novii System Pod
            MANUFACTURER                      Monica Healthcare Ltd
            TYPE                              Novii Pod
            PART NUMBER                       107-PT-003
            SERIAL NUMBER                     AA5426, AA5431
            COUNTRY OF ORIGIN                 United Kingdom




      COMMERCIAL-IN-CONFIDENCE                                                                                        Page 4 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




          Novii System Pod

                                                             MAIN EUT
           MANUFACTURING DESCRIPTION                      Novii System Pod for the Novii Wireless Patch System
           MANUFACTURER                                   Monica Healthcare Ltd
           MODEL NAME/NUMBER                              Novii System Pod
           PART NUMBER                                    107-PT-003
           SERIAL NUMBER                                  AA5426, AA5431 (AA5425, AA5235, AA5266, AA5233)
           HARDWARE VERSION                               Rev H
           SOFTWARE VERSION                               Rev 2.54
           PSU VOLTAGE/FREQUENCY/CURRENT                  Qi wireless charging (from Novii System Interface Unit)
           HIGHEST INTERNALLY GENERATED /
                                                          26MHz
           USED FREQUENCY
           FCC ID (if applicable)                         YOM-6960-MON
           INDUSTRY CANADA ID (if applicable)
                                                      The Novii System Pod is part of the Novii Wireless Patch
           TECHNICAL DESCRIPTION
                                                      System: a Maternal/Fetal monitor that records Fetal heart rate,
           (a brief description of the intended use and
                                                      Maternal Heart Rate and Uterine Contractions from a pregnant
           operation)
                                                      subject.
           COUNTRY OF ORIGIN                          United Kingdom
                                              RF CHARACTERISTICS (if applicable)
           TRANSMITTER FREQUENCY
                                             2402 to 2480 (Bluetooth) and 0.110 to 0.205 (Qi charger)
           OPERATING RANGE (MHz)
           RECEIVER FREQUENCY OPERATING
                                             2402 to 2480 (Bluetooth) and 0.110 to 0.205 (Qi charger)
           RANGE (MHz)
           INTERMEDIATE FREQUENCIES          2440 MHz (Bluetooth) and 0.157 MHz (Qi charger)
           EMISSION DESIGNATOR(S):
                                             1M00F1D
           (i.e. G1D, GXW)
           MODULATION TYPES:                 V2.1 + EDR (GFSK. Bluetooth) and frequency modulation (Qi
           (i.e. GMSK, QPSK)                 charger)
           OUTPUT POWER (W or dBm)           10 dBm (Bluetooth)
                               SEPARATE BATTERY/POWER SUPPLY (if applicable)
           MANUFACTURING DESCRIPTION         Novii System Interface Unit
           MANUFACTURER                      Monica Healthcare Ltd
           TYPE                              Qi wireless charger
           PART NUMBER                       107-PT-001
           PSU VOLTAGE/FREQUENCY/CURRENT     On receiver: 5V DC / 380mA
           COUNTRY OF ORIGIN                 United Kingdom
                                          MODULES (if applicable)
           MANUFACTURING DESCRIPTION
           MANUFACTURER
           TYPE
           POWER
           FCC ID
           INDUSTRY CANADA ID
           EMISSION DESIGNATOR
           DHSS/FHSS/COMBINED OR OTHER
           COUNTRY OF ORIGIN
                                        ANCILLARIES (if applicable)
                                             Novii System
           MANUFACTURING DESCRIPTION                                Novii System Patch
                                             Interface Unit
                                             Monica Healthcare      Monica Healthcare
           MANUFACTURER
                                             Ltd                    Ltd
           TYPE                              Novii Interface        Novii Patch
           PART NUMBER                       107-PT-001             107-PT-004
                                             TA1763 (TA1764,
           SERIAL NUMBER                     TA1762, TA1759,        N/A (disposable)
                                             TA1752)
           COUNTRY OF ORIGIN                 United Kingdom         USA




   COMMERCIAL-IN-CONFIDENCE                                                                                         Page 5 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




1.5        Product Information


1.5.1      Technical Description

           The Monica Novii POD is an intrapartum Maternal/Fetal Monitor that non-invasively measures and
           displays fetal heart rate (FHR), uterine activity (UA) and maternal heart rate (MHR).

           The Novii POD acquires and displays the FHR tracing from abdominal surface electrodes that pick
           up the fetal ECG (fECG) signal. Using the same surface electrodes, the POD also acquires and
           displays the UA tracing from the uterine electromyography (EMG) signal and the MHR tracing from
           the maternal ECG signal (mECG).

           The POD is indicated for use on women who are at >36 completed weeks, in labor, with singleton
           pregnancies, using surface electrodes on the maternal abdomen.

           The Novii Patch is an accessory to the Novii POD that connects directly to the Novii POD and
           contains the surface electrodes that attach to the abdomen. The Novii Interface is an accessory to
           the Novii POD which provides a means of interfacing the wireless output of the Novii POD to the
           transducer inputs of a Maternal/Fetal Monitor.

           The Novii Interface enables signals collected by the Novii POD to be printed and displayed on a
           Maternal/Fetal Monitor and sent on to a central network, if connected.

1.6        Deviations from the Standard

           No deviations from the applicable test standard were made during testing.

1.7        EUT Modification Record

           The table below details modifications made to the EUT during the test programme.
           The modifications incorporated during each test are recorded on the appropriate test pages.

                                                                                                            Date Modification
            Modification State   Description of Modification still fitted to EUT   Modification Fitted By
                                                                                                            Fitted
            Serial Number: TA1772

            0                    As supplied by the customer                       Not Applicable           Not Applicable

            Serial Number: AA5233

            0                    As supplied by the customer                       Not Applicable           Not Applicable


                                                                     Table 3




      COMMERCIAL-IN-CONFIDENCE                                                                                   Page 6 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




1.8        Test Location

           TÜV SÜD Product Service conducted the following tests at our Fareham Test Laboratory.

            Test Name                                Name of Engineer(s)              Accreditation

            Configuration and Mode: Interface + Pod – Idle

            Radiated Emissions                       Paul Joynson and Graeme Lawler   UKAS

            AC Power Line Conducted Emissions        Paul Joynson and Graeme Lawler   UKAS


                                                                 Table 4


           Office Address:

           Octagon House
           Concorde Way
           Segensworth North
           Fareham
           Hampshire
           PO15 5RL
           United Kingdom




      COMMERCIAL-IN-CONFIDENCE                                                                        Page 7 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




2          Test Details

2.1        Radiated Emissions

2.1.1      Specification Reference

           FCC 47 CFR Part 15B, Clause 15.109

2.1.2      Equipment Under Test and Modification State

           Interface, S/N: TA1772 - Modification State 0
           Pod, S/N: AA5233- Modification State 0

2.1.3      Date of Test

           04-December-2017 to 05-December-2017

2.1.4      Test Method

           The test was performed in accordance with ANSI C63.4, clause 8.

2.1.5      Environmental Conditions

           Ambient Temperature         18.1 °C
           Relative Humidity           37.0 %

2.1.6      Test Results

           Interface + Pod - Standby

           Highest frequency generated or used within the EUT: 2.5 GHz
           Upper frequency test limit: 13.5 GHz

            Frequency       QP Level         QP Limit          QP Margin
                                                                                Angle(Deg)   Height(m)   Polarity
            (MHz)           (dBuV/m)         (dBuV/m)          (dBuV/m)
             83.835         19.3             39.1              -19.8            43           1.00        Vertical
             84.761         24.2             39.1              -14.9            116          1.00        Vertical
             85.628         17.8             39.1              -21.3            244          1.00        Vertical

             493.318        34.1             46.4              -12.3            160          1.00        Vertical
             618.755        29.6             46.4              -16.8            151          1.16        Vertical

             690.703        38.7             46.4              -7.7             233          1.00        Horizontal
             887.805        34.5             46.4              -11.9            129          1.00        Vertical


                                                    Table 5 - 30 MHz to 1 GHz




      COMMERCIAL-IN-CONFIDENCE                                                                           Page 8 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




                                  Figure 1 - 30 MHz to 1 GHz - Horizontal and Vertical




   COMMERCIAL-IN-CONFIDENCE                                                              Page 9 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




           Frequency          Result (µV/m)                    Limit (µV/m)                    Margin (µV/m)       Angle (°)          Height       Polarisation
           (GHz)                                                                                                                      (m)
                          Peak             Average           Peak        Average              Peak     Average
           1.085332      45.16               41.61           69.54           49.54           24.45       7.93      173                350          Vertical
           1.282613      44.49               39.22           69.54           49.54            25.5      10.32      165                302          Vertical


                                                                Table 6 - 1 GHz to 13.5 GHz

          No other emissions were detected within 6 dB of the limit.


                                                                                     * RBW   1 MHz     Marker 1 [T1 ]
                                                                                     * VBW50 kHz               41.74 dBµV/m
                              Ref        80 dBµV/m            * Att   0 dB            SWT 160 ms            1.280448718 GHz

                               80       Offset    -10 dB
                                                                                                                                *
                               FC15BA_P                                                                                         A
                               70

                       1 PK
                       VIEW
                               60
                                                                                                                                LVL
                       2 PK
                       MAXH                                                                                                     TDS
                               FC15BA_A
                               50

                                    1                                                                                           PA
                                                                                                                                PS
                               40



                               30

                                                                                                                                6DB
                               20                                                                                               AC



                               10



                               0



                               -10


                               -20

                               Start      1 GHz                              700 MHz/                            Stop   8 GHz




                       Date: 4.DEC.2017           20:37:08




                                             Figure 2 - 1 GHz to 8 GHz - Horizontal and Vertical




   COMMERCIAL-IN-CONFIDENCE                                                                                                                    Page 10 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




                                                                               * RBW   1 MHz     Marker 1 [T1 ]
                                                                               * VBW50 kHz               35.55 dBµV/m
                               Ref     80 dBµV/m           * Att   0 dB         SWT 125 ms           12.971153846 GHz

                               80    Offset     -10 dB
                                                                                                                         *
                               FC15BA_P                                                                                  A
                               70

                        1 PK
                        VIEW
                               60
                                                                                                                         LVL
                        2 PK
                        MAXH                                                                                             TDS
                               FC15BA_A
                               50


                                                                                                                         PS
                               40
                                                                                                              1


                               30

                                                                                                                         6DB
                               20                                                                                        AC



                               10



                               0



                               -10


                               -20

                               Start    8 GHz                             550 MHz/                     Stop   13.5 GHz




                        Date: 4.DEC.2017        22:00:33




                                        Figure 3 - 8 GHz to 13.5 GHz - Horizontal and Vertical

          FCC 47 CFR Part 15, Limit Clause 15.109

           Frequency of Emission (MHz)                                               Field Strength (µV/m)
           30 to 88                                                                  90.0
           88 to 216                                                                 150.0

           216 to 960                                                                210.0
           Above 960                                                                 300.0


2.1.7




   COMMERCIAL-IN-CONFIDENCE                                                                                                    Page 11 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




2.1.8     Test Location and Test Equipment Used

          This test was carried out in EMC Chamber 5.

                                                                                  Calibration
           Instrument                Manufacturer        Type No          TE No   Period        Calibration Due
                                                                                  (months)

           Antenna (Bilog)           Schaffner           CBL6143          287     24            18-Apr-2018

           Pre-Amplifier             Phase One           PS04-0086        1533    12            31-Jul-2018

           Screened Room (5)         Rainford            Rainford         1545    36            20-Dec-2017

           Turntable Controller      Inn-Co GmbH         CO 1000          1606    -             TU

           Hygrometer                Rotronic            HYGROPALM 1      2338    12            24-Oct-2018

                                                         NPS-2302-8000-
           Cable (N-N, 8m)           Rhophase                             3248    12            02-May-2018
                                                         NPS

           EMI Test Receiver         Rohde & Schwarz     ESU40            3506    12            22-Nov-2018

           Tilt Antenna Mast         maturo Gmbh         TAM 4.0-P        3916    -             TU

           Mast Controller           maturo Gmbh         NCD              3917    -             TU

                                                         KPS-1501-2000-
           Cable (Rx, Km-Km 2m)      Scott Cables                         4526    6             22-May-2018
                                                         KPS
           Cable (Rx, SMAm-SMAm                          SLSLL18-SMSM-
                                     Scott Cables                         4528    6             03-Feb-2017
           0.5m)                                         00.50M
           Double Ridged
                                     ETS-Lindgren        3117             4722    12            17-Feb-2018
           Waveguide Horn Antenna
           Double Ridge Broadband
                                     Schwarzbeck         BBHA 9120 B      4848    12            17-Feb-2018
           Horn Antenna


                                                        Table 7

          TU - Traceability Unscheduled




   COMMERCIAL-IN-CONFIDENCE                                                                     Page 12 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




2.2        AC Power Line Conducted Emissions

2.2.1      Specification Reference

           FCC 47 CFR Part 15B, Clause 15.107

2.2.2      Equipment Under Test and Modification State

           Interface, S/N: TA1772 - Modification State 0

2.2.3      Date of Test

           03-December-2017 to 05-December-2017

2.2.4      Test Method

           The test was performed in accordance with ANSI C63.4, clause 7.

2.2.5      Environmental Conditions

           Ambient Temperature        18.1 °C
           Relative Humidity          37.0 %

2.2.6      Test Results

           Interface + Pod - Standby

           Applied supply Voltage: 60 Hz
           Applied supply frequency: 120 V AC

            Frequency      QP Level        QP Limit   QP Margin    AV Level    AV Limit   AV Margin
            (MHz)          (dBuV)          (dBuV)     (dBuV)       (dBuV)      (dBuV)     (dBuV)
             0.152         48.0            79.0       -31.0        35.9        66.0       -30.1

             0.288         45.6            79.0       -33.4        44.9        66.0       -21.1
             0.504         43.2            73.0       -29.8        42.9        60.0       -17.1

             0.790         42.8            73.0       -30.2        42.2        60.0       -17.8
             0.863         43.6            73.0       -29.4        42.8        60.0       -17.2

             1.152         43.5            73.0       -29.5        42.4        60.0       -17.6


                                           Table 8 - Live Line Emissions Results




      COMMERCIAL-IN-CONFIDENCE                                                              Page 13 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




                                           Figure 4 - Live Line - 150 kHz to 30 MHz




   COMMERCIAL-IN-CONFIDENCE                                                           Page 14 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




           Frequency       QP Level          QP Limit    QP Margin       AV Level        AV Limit        AV Margin
           (MHz)           (dBuV)            (dBuV)      (dBuV)          (dBuV)          (dBuV)          (dBuV)
            0.152          47.8              79.0        -31.2           36.7            66.0            -29.3

            0.288          45.5              79.0        -33.5           44.8            66.0            -21.2
            0.503          43.3              73.0        -29.7           42.9            60.0            -17.1

            0.791          43.4              73.0        -29.6           42.7            60.0            -17.3
            0.864          43.6              73.0        -29.4           42.9            60.0            -17.1
            1.151          43.8              73.0        -29.2           42.7            60.0            -17.3



                                           Table 9 - Neutral Line Emissions Results




                                           Figure 5 - Neutral Line - 150 kHz to 30 MHz

          FCC 47 CFR Part 15, Limit Clause 15.107

           Frequency of Emission (MHz)                                   Conducted Limit (dBµV)

                                                           Quasi-Peak                               Average

           0.15 to 0.5                                       66 to 56*                              56 to 46*
           0.5 to 5                                              56                                    46
           5 to 30                                               60                                    50


                                                            Table 10

          *Decreases with the logarithm of the frequency.




   COMMERCIAL-IN-CONFIDENCE                                                                                 Page 15 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




2.2.7     Test Location and Test Equipment Used

          This test was carried out in EMC Chamber 5.

                                                                                Calibration
           Instrument                Manufacturer         Type No       TE No   Period        Calibration Due
                                                                                (months)

           Transient Limiter         Hewlett Packard      11947A        15      12            30-May-2018

           LISN (1 Phase)            Chase                MN 2050       336     12            07-Apr-2018

           Screened Room (5)         Rainford             Rainford      1545    36            20-Dec-2017

           Hygrometer                Rotronic             HYGROPALM 1   2338    12            24-Oct-2018

           Digital Multimeter        Iso-tech             IDM-101       2895    12            20-Jul-2018

           EMI Test Receiver         Rohde & Schwarz      ESU40         3506    12            22-Nov-2018


                                                        Table 11




   COMMERCIAL-IN-CONFIDENCE                                                                   Page 16 of 17


Document Number: 75941097-01 | Issue: 01
COMMERCIAL-IN-CONFIDENCE




3         Measurement Uncertainty

          For a 95% confidence level, the measurement uncertainties for defined systems are:

           Test Name                                       Measurement Uncertainty

           Radiated Emissions                              30 MHz to 1 GHz: ±5.2 dB
                                                           1 GHz to 40 GHz: ±6.3 dB
           AC Power Line Conducted Emissions               150 kHz to 30 MHz, LISN, ±3.7 dB


                                                     Table 12




    COMMERCIAL-IN-CONFIDENCE                                                                   Page 17 of 17



Document Created: 2017-12-13 12:06:56