Novii_SAR_Considerations V1-0

FCC ID: YOM-6960-MON

RF Exposure Info

Download: PDF
FCCID_3677783

                                                                                     DOCUMENT                                              REVISION            PAGE


                                                                                                                                             1-0              1/7




       Product                                                                    Novii System




Document name                                                                     Novii System




 Document type                                                                         Technical

                                    This document specifies the technical details Novii System




       Abstract




                                                       MONICA HEALTHCARE LTD PROPRIETARY INFORMATION

Copyright © 2005 by MONICA HEALTCARE LTD. All rights reserved. This document and the information contained therein are MONICA HEALTCARE LTD’s exclusive property. Any copy
and/or disclosure thereof in any form whatsoever are subject to MONICA HEALTCARE LTD’s prior written consent.


                                                                                     DOCUMENT                                              REVISION            PAGE


                                                                                                                                             1-0              2/7



                                                            DOCUMENT HISTORY



                             SURNAME First name                              Responsibility                            Date                     Signature



   Issued by             Branson Simon                            Engineering                                  15 Nov. 2017




REVISION
                     WRITTEN BY                     DATE                                            DESCRIPTION/CHANGES
  INDEX


    1-0            Simon Branson                  30/11/17             First Issue




                                                       MONICA HEALTHCARE LTD PROPRIETARY INFORMATION

Copyright © 2005 by MONICA HEALTCARE LTD. All rights reserved. This document and the information contained therein are MONICA HEALTCARE LTD’s exclusive property. Any copy
and/or disclosure thereof in any form whatsoever are subject to MONICA HEALTCARE LTD’s prior written consent.


                                                                                     DOCUMENT                                              REVISION            PAGE


                                                                                                                                             1-0              3/7



                                                                        CONTENTS


                                                   SAR Considerations for Novii System

                                                                     POD Calculations

                                                              Interface Unit Calculation

                                                                          Calculations

                                                                          Conclusions




                                                       MONICA HEALTHCARE LTD PROPRIETARY INFORMATION

Copyright © 2005 by MONICA HEALTCARE LTD. All rights reserved. This document and the information contained therein are MONICA HEALTCARE LTD’s exclusive property. Any copy
and/or disclosure thereof in any form whatsoever are subject to MONICA HEALTCARE LTD’s prior written consent.


                                                                                     DOCUMENT                                              REVISION            PAGE


                                                                                                                                             1-0              4/7



This document details the SAR considerations for the Novii System.

For the FCC non ionising radiation limits are given;

see extract from FCC publication section 4.3 a)




Section 4.3

a) For 100 MHz to 6 GHz and test separation distances ≤ 50 mm, the 1-g and 10-g SAR test exclusion thresholds
are determined by the following:
[(max. power of channel, including tune-up tolerance, mW) / (min. test separation distance, mm)] · [√f(GHz)] ≤ 3.0
for 1-g SAR, and ≤ 7.5 for 10-g extremity SAR,30 where
  f(GHz) is the RF channel transmit frequency in GHz




                                                       MONICA HEALTHCARE LTD PROPRIETARY INFORMATION

Copyright © 2005 by MONICA HEALTCARE LTD. All rights reserved. This document and the information contained therein are MONICA HEALTCARE LTD’s exclusive property. Any copy
and/or disclosure thereof in any form whatsoever are subject to MONICA HEALTCARE LTD’s prior written consent.


CQA                                                                    Report No.: CQASZ170701391EW—02

1     SAR Evaluation

1.1    RF Exposure Compliance Requirement
1.1.1 Standard Requirement
      According to KDB447498D01 General RF Exposure Guidance vO6
      4.3.1. Standalone SAR test exclusion considerations
            Unless specifically required by the published RF exposure KDB procedures, standalone 1—g
      head or body and 10—g extremity SAR evaluation for general population exposure conditions, by
      measurement or numerical simulation, is not required when the corresponding SAR Exclusion
      Threshold condition, listed below, is satisfied.
1.1.2 Limits
      The 1—g and 10—g SAR test exclusion thresholds for 100 MHz to 6 GHz at test separation distances
      < 50 mm are determined by:

      [(max power of channel, including tune—up tolerance, mW)/(min. test separation distance, mm}] >
      [vf(GHz)] < 3.0 for 1—g SAR and = 7.5 for 10—g extremity SAR, where

        f(GHz) is the RF channel transmit frequency in GHz
        Power and distance are rounded to the nearest mW and mm before calculation‘"
        The result is rounded to one decimal place for comparison
      The test exclusions are applicable only when the minimum test separation distance is < 50 mm and
      for transmission frequencies between 100 MHz and 6 GHz. When the minimum test separation
      distance is < 5 mm, a distance of 5 mm is applied to determine SAR test exclusion
1.1.3EUT RF Exposure

      For BLE:

      Measurement Data
                                           GESK mode
                 Test channel                                  Peak Output Power (dBm)}
                    Lowest                                              —0.55

                     Middle                                              0.59
                    Highest                                              0.88

      The Max Conducted Peak Output Power is 0.88dBm in Highest channel(2.480GHz);
      The best case gain of the antenna is 1.0dBi.
      EIRP= 0.88dBm + 1.00Bi =1.88dBm
      1.88dBm logarithmic terms convert to numeric result is nearly 1.54mW
      According to the formula. calculate the EIRP test result:
      [{max. power of channel, including tune—up tolerance, mW)/{min. test separation distance, mm})]
      N(GHz)]
      General RF Exposure = (1.54mW / 5 mm ) x v2.480GHz = 0.485 I)
      SAR requirement:
      S 3.0                                                                 @;
      ® <@.
      So the SAR report is not required.
      Remark: The Max Conducted Peak Output Power data refer to report CQASZ170701391EW—01.



                                                 Page:1 of 1


                                                                                     DOCUMENT                                              REVISION            PAGE


                                                                                                                                             1-0              6/7



POD

For the POD, from the FCC extract the calculation is:


Power/Separation * sqrt (frequency) < 3 (SAR requirement) 1-g

The Panasonic module data sheet states a maximum power of 10.5 dBm (11.22 mW).

Frequency is 2.441 GHz, sqrt (2.441GHz) = 1.56


Separation is 7mm (measured using Vernier gauge from the antenna on the PCB to the part in
contact with the skin)


Calculation is 11.22mW / 7mm * sqrt(2.48GHz) = 2.52.


The POD is therefore SAR test excluded because it is less than 3.0.


It would also meet SAR test exclusion at 6mm separation.




                                                       MONICA HEALTHCARE LTD PROPRIETARY INFORMATION

Copyright © 2005 by MONICA HEALTCARE LTD. All rights reserved. This document and the information contained therein are MONICA HEALTCARE LTD’s exclusive property. Any copy
and/or disclosure thereof in any form whatsoever are subject to MONICA HEALTCARE LTD’s prior written consent.


                                                                                     DOCUMENT                                              REVISION            PAGE


                                                                                                                                             1-0              7/7


Interface Unit


For the Interface, from the FCC extract the calculation is:


Power/Separation * sqrt (frequency) < 7.5 (SAR requirement) 10-g


Assume the power is 2 x POD power = 22.5mW (excluding wireless charger)


Frequency is 2.441 GHz, sqrt (2.441GHz) = 1.56


Separation is 5mm


22.5/5*1.56 = 7.02 therefore SAR exempt < 7.5 (SAR 10-g requirement)


Also, for the Interface Unit, we state that the user is > 20cm away in the Instructions for Use and
is therefore SAR exempt.




                                                                 End of Document




                                                       MONICA HEALTHCARE LTD PROPRIETARY INFORMATION

Copyright © 2005 by MONICA HEALTCARE LTD. All rights reserved. This document and the information contained therein are MONICA HEALTCARE LTD’s exclusive property. Any copy
and/or disclosure thereof in any form whatsoever are subject to MONICA HEALTCARE LTD’s prior written consent.



Document Created: 2017-12-04 16:13:06
Document Modified: 2017-12-04 16:13:06

© 2024 FCC.report