RF Exposure_Annex C

FCC ID: YHLBLUJOY3G

RF Exposure Info

Download: PDF
FCCID_4074925

                                                                               REPORT No.:SZ18100073S01



 Annex C Plots of System Performance Check




         SHENZHEN MORLAB COMMUNICATIONS TECHNOLOGY Co., Ltd.                     Tel: 86-755-36698555   Fax: 86-755-36698525
MORLAB   FL1-3, Building A, FeiYang Science Park, No.8 LongChang Road,
         Block67, BaoAn District, ShenZhen , GuangDong Province, P. R. China     Http://www.morlab.cn   E-mail: service@morlab.cn


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.10.28

System Check_835MHz_Head_181028



Communication System: UID 0, CW (0); Frequency: 835 MHz;Duty Cycle: 1:1
Medium: HSL_835_181028 Medium parameters used: f = 835 MHz; σ = 0.89 S/m; εr = 41.76 ρ = 1000
kg/m3
Ambient Temperature:23.2 ℃; Liquid Temperature:22.2 ℃

DASY5 Configuration:
- Probe: ES3DV3 - SN3154; ConvF(6.26, 6.26, 6.26); Calibrated: 2017.10.30;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW835/Area Scan (81x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.70 W/kg

CW835/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 55.53 V/m; Power Drift = -0.07 dB
Peak SAR (extrapolated) = 3.62 W/kg
SAR(1 g) = 2.45W/kg; SAR(10 g) = 1.57W/kg
Maximum value of SAR (measured) = 2.69 W/kg




               0 dB = 2.70 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.10.28

System Check_1900MHz_Head_181028



Communication System: UID 0, CW (0); Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: HSL_1900_181028 Medium parameters used: f = 1900 MHz; σ = 1.43 S/m; εr = 40.885; ρ
= 1000 kg/m3
Ambient Temperature:23.2 ℃; Liquid Temperature:22.2 ℃

DASY5 Configuration:
- Probe: ES3DV3 - SN3154; ConvF(4.98, 4.98, 4.98); Calibrated: 2017.10.30;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW 1900/Area Scan (81x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 12.3 W/kg

CW 1900/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 92.78 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 20.1 W/kg
SAR(1 g) = 10.23 W/kg; SAR(10 g) = 5.26 W/kg
Maximum value of SAR (measured) = 12.2 W/kg




               0 dB = 12.3 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.10.28

System Check_835MHz_Body_181028



Communication System: UID 0, CW (0); Frequency: 835 MHz;Duty Cycle: 1:1
Medium: MSL_835_181028 Medium parameters used: f = 835 MHz; σ = 0.956 S/m; εr = 54.343; ρ =
1000 kg/m3
Ambient Temperature:23.2 ℃; Liquid Temperature:22.2 ℃

DASY5 Configuration:
- Probe: ES3DV3 - SN3154; ConvF(6.22, 6.22, 6.22); Calibrated: 2017.10.30;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW 835/Area Scan (81x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.52 W/kg

CW 835/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 51.04 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 3.44 W/kg
SAR(1 g) = 2.49 W/kg; SAR(10 g) = 1.56 W/kg
Maximum value of SAR (measured) = 2.55 W/kg




               0 dB = 2.55 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.10.28

System Check_1900MHz_Body_181028



Communication System: UID 0, CW (0); Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: MSL_1900_181028 Medium parameters used: f = 1900 MHz; σ = 1.533 S/m; εr = 52.385; ρ
= 1000 kg/m3
Ambient Temperature:23.2 ℃; Liquid Temperature:22.2 ℃

DASY5 Configuration:
- Probe: ES3DV3 - SN3154; ConvF(4.71, 4.71, 4.71); Calibrated: 2017.10.30;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:147
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


CW 1900/Area Scan (81x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 11.2 W/kg

CW 1900/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 79.19 V/m; Power Drift = 0.19 dB
Peak SAR (extrapolated) = 18.0 W/kg
SAR(1 g) = 9.99 W/kg; SAR(10 g) = 5.53 W/kg
Maximum value of SAR (measured) = 11.0 W/kg




               0 dB = 11.0 W/kg