12524730-S1V1 FCC Report RF Exposure

FCC ID: XPY2AGQN4NNN

RF Exposure Info

Download: PDF
FCCID_4071756

      FCC Part 1 Subpart I
      FCC Part 2 Subpart J

    RF EXPOSURE REPORT

              FOR

        Smart Pill Bottle

MODEL NUMBER: 900-00003 Rev A

   FCC ID: XPY2AGQN4NNN


 REPORT NUMBER: 12524730-S1V1

     ISSUE DATE: 10/24/2018

          Prepared for
          AdhereTech
     11 Broadway, Suite 457
      New York, NY 10004
             USA

           Prepared by
 UL VERIFICATION SERVICES INC.
     47173 BENICIA STREET
   FREMONT, CA 94538, U.S.A.
       TEL: (510) 771-1000




    NVLAP LAB CODE 200065-0


REPORT NO: 12524730-S1V1                                                                 DATE: 10/24/2018
FCC ID: S5EST10417
                                             Revision History


            Issue
 Rev.       Date            Revisions                                                       Revised By
  V1        10/24/2018      Original issue




                                               Page 2 of 6
UL Verification Services Inc.                                                                            Doc. No.: 1.0
     This report shall not be reproduced without the written approval of UL Verification Services Inc.
                                                     .


REPORT NO: 12524730-S1V1                                                                                             DATE: 10/24/2018
FCC ID: S5EST10417
                                               TABLE OF CONTENTS
1.     ATTESTATION OF TEST RESULTS ................................................................................. 4

2.     TEST METHODOLOGY ..................................................................................................... 5

3.     REFERENCES ................................................................................................................... 5

4.     FACILITIES AND ACCREDITATION ................................................................................. 5

5.     DEVICE UNDER TEST ....................................................................................................... 5
     5.1.    Description ................................................................................................................... 5
     5.2.    Wireless Technologies and Output Power ................................................................... 5

6.     DUTY CYCLE POWER CORRECTION .............................................................................. 6

7.     STANDALONE SAR TEST EXCLUSION CONSIDERATIONS .......................................... 6

8.     SIMULTANEOUS TRANSMISSION SAR ANALYSIS ........................................................ 6




                                               Page 3 of 6
UL Verification Services Inc.                                                                                                             Doc. No.: 1.0
     This report shall not be reproduced without the written approval of UL Verification Services Inc.
                                                                      .


REPORT NO: 12524730-S1V1                                                                 DATE: 10/24/2018
FCC ID: S5EST10417

1. ATTESTATION OF TEST RESULTS
COMPANY NAME:                    AdhereTech
                                 11 Broadway, Suite 457
                                 New York, NY 10004
                                 USA

DUT DESCRIPTION:                 Smart Pill Bottle

MODEL:                           900-00003 Rev A

SERIAL NUMBER:                   N/A


                                       APPLICABLE STANDARDS
                            STANDARD                                          TEST RESULTS
          FCC PART 1 SUBPART I & PART 2 SUBPART J                                   Pass


Note: The results documented in this report apply only to the tested sample, under the
conditions and modes of operation as described herein. This document may not be altered or
revised in any way unless done so by UL Verification Services Inc. and all revisions are duly
noted in the revisions section. Any alteration of this document not carried out by UL Verification
Services Inc. will constitute fraud and shall nullify the document.

Approved & Released For
UL Verification Services Inc. By:




Dave Weaver
Operations Leader
UL Verification Services Inc.




                                               Page 4 of 6
UL Verification Services Inc.                                                                            Doc. No.: 1.0
     This report shall not be reproduced without the written approval of UL Verification Services Inc.
                                                     .


REPORT NO: 12524730-S1V1                                                                 DATE: 10/24/2018
FCC ID: S5EST10417

2. TEST METHODOLOGY
All calculations were made in accordance with FCC KDB 447498 D01 v06.


3. REFERENCES
Output power, Duty cycle is excerpted from the applicable test reports or client declarations.


4. FACILITIES AND ACCREDITATION
The test sites and measurement facilities used to collect data are located at 47173 and 47266
Benicia Street, Fremont, California, USA

UL Verification Services Inc. is accredited by NVLAP, Laboratory Code 200065-0.


5. DEVICE UNDER TEST
5.1.     Description
The AdhereTech Smart Pill Bottle is a device intended to generate alerts and reminders for pre-
determined medication dosing schedules. The device in is a sleep state for 99% of its life and
only powers on and connects very infrequently to deliver small packets of data. The device
powers on, connects to the base-station, and powers down as a function of two data events:
    1. Auto check-in, and
    2. Dose event.
For any given medication dose schedule (1, 2, or 3 doses per day) the device will
perform one auto check-in event and one dose event. The auto check-in is scheduled by
the device using an internal real-time clock and the dose event is triggered by the user
opening and closing the bottle cap.

In a worst-case scenario there could be two data events in a 30-minute window. This would
be one auto check-in and one dose event (i.e. the bottle has an auto check-in and then the
user opens the bottle within 30 minutes). During any given data event the device transmits for
no more than 3 seconds (one second for cell tower registration and two seconds to deliver
and receive one packet of data equal to no more than 2 kB).

5.2.    Wireless Technologies and Output Power
 Wireless technologies       Frequency bands                    Maximum Output Power

 LTE                     2, 4 and 12           200mW for all bands




                                               Page 5 of 6
UL Verification Services Inc.                                                                            Doc. No.: 1.0
     This report shall not be reproduced without the written approval of UL Verification Services Inc.
                                                     .


REPORT NO: 12524730-S1V1                                                                 DATE: 10/24/2018
FCC ID: S5EST10417



6. DUTY CYCLE POWER CORRECTION
The device supports LTE bands 2, 4 and 12. The minimum LTE bitrate, assuming an allocation
of 1 RB, is around 300 kb/s. The amount of data transferred per transmission is less than 2 kB
so 2 seconds to transfer the data is conservative.

The maximum transmit power for any supported band is 200 mW.

The worst-case duty cycle over a thirty-minute period (1800 seconds) is calculated assuming 6
seconds of total transmission time as described in Section 5.

Duty Cycle = 6/1800 = 0.0033

Scaling the output power based upon duty cycle gives 200 x 0.0033 = 0.66 mW

7. STANDALONE SAR TEST EXCLUSION CONSIDERATIONS
From KDB 447498, for transmission frequencies 100 MHz to 6 GHz and test separation
distances ≤ 50 mm, the 1-g SAR test exclusion thresholds are determined by the following:

[(max. power of channel, including tune-up tolerance, mW) / (min. test separation distance,
mm)] · [√f(GHz)] ≤ 3.0 where:
    f(GHz) is the RF channel transmit frequency in GHz;
    Power and distance are rounded to the nearest mW and mm before calculation;
    For a separation distance of less than 5mm, 5mm is used.

The result is rounded to one decimal place for comparison with the 3.0 threshold.
The table below shows that at the maximum power for all bands and technologies, after
accounting for source-based and operational duty cycles, and for a separation distance of 5mm
or less, SAR test exclusion applies.

The device was assessed the 1g SAR limits.

                                                     Max. tune-up       Min. test     SAR test
                     RF Exposure      Frequency       tolerance
 RF Air interface                                                      separation     exclusion
                      Conditions        (GHz)
                                                          (mW)       distance (mm)     Result*
LTE Band 2            Body-w orn         1.910             1                5            0.3
LTE Band 4            Body-w orn         1.755             1                5            0.3
LTE Band 12           Body-w orn         0.716             1                5            0.2
 Conclusion:
 *: The computed value is ≤ 3; therefore, this qualifies for SAR test exclusion.

8. SIMULTANEOUS TRANSMISSION SAR ANALYSIS
Simultaneous transmission is not supported.

                                         END OF REPORT

                                               Page 6 of 6
UL Verification Services Inc.                                                                            Doc. No.: 1.0
     This report shall not be reproduced without the written approval of UL Verification Services Inc.
                                                      .



Document Created: 2019-03-24 12:25:37
Document Modified: 2019-03-24 12:25:37