Users Guide

FCC ID: XO8-3000TX1

Users Manual

Download: PDF
FCCID_1830439

IC 3000 Wireless Pendant
Part# IC3000TX1

This Pendant is designed to operate with the IC 3000 Emergency Response / Medical Triage Console. The
Pendant is designed with two unique buttons. The large RED button allows a user to wirelessly trigger a
telephone call for help in cases of medical emergencies, accidents, etc. that happen at home. The second smaller
WHITE button with the blue medical symbol is used to place a call to a medical Triage center to ask medical
questions and receive personalized professional assessment.

   IMPORTANT: The pendant is shipped with Emergency button disabled (shipping mode).
To enable the Emergency button see Section E.

            A. Description of Pendant
            1. Emergency Help Button
                •   Pressing the RED button will
                    activate the console to dial
                    the emergency line and send an
                    Emergency Alarm report.
                •   Red light flashing: Signal is
                    transmitted.




            2. Nurse Call Button
                •   Pressing the WHITE button will activate
                    the console to dial the Nurse line
                •   Blue light flashing: Signal is transmitted.

            3. Adjustable Lanyard.


            B. Learning In Pendants
        Follow the steps below for learning in pendant

                Step 1. Press and hold the NURSE & CANCEL buttons on the console (approx. 5 sec.) until the
                        console emits a beep and STATUS LED illuminates PURPLE. Now the console is in
                        LEARNING MODE.

                Step 2. Learn Pendant 1: press and hold the EMERGENCY button on the console.

                        Learn Pendant 2: press and hold the NURSE Button on the console.

                Step 3. The status LED will indicate either Green (Zone available) or Red (Zone Occupied).

                        a. If the Status LED is GREEN this means that the Zone is available, press RED
                           pendant button to learn-in. The Status LED on the console will illuminate RED and
                           an audible beep will be heard.

                        b. If the Status LED is RED this means that the Zone is occupied, you may clear the
                           Zone by pressing the CANCEL button on the console until system emits a short beep
                           and the Status LED turns green. Go to Step 3a to learn in pendant.




                                                                                                             9-­‐19-­‐2012	
  

                                                                                                             Ver.	
  1.2	
  


  Step 4. To exit LEARNING MODE, press the CANCEL button.
  •    If no signal is received within 2 minutes after entering QUICK LEARNING MODE, the console
       will exit and return to idle mode.
  •    To add more than 2 pendants please reference Command 63 through 70 in the IC 3000
       system manual.

C. Operation
  •    Press the RED emergency help button to activate the console to dial the emergency line and
       send an Emergency Alarm report.
  •    Press the WHITE nurse call button to activate to dial the Nurse Line.

D. Low Battery Detection
  The IC 3000 Pendant provides Low Battery Detection and Supervision features.

  •    The IC 3000 Pendant has a battery pre-installed. The Pendants housing has been factory
       sealed to protect against water intrusion.
  •    When the battery voltage becomes low, a Low Battery signal will be sent to the console. The
       “Status LED” on the main console will illuminate Solid YELLOW to notify the user of the
       condition. Additionally when a low battery condition is present a CID alert E380 will be sent.
  •    The console will transmit a new report every 24 hours as long as the condition exists.
  •    The low battery pendant must be replaced with new a pendant. Discard the low battery
       pendant.

E. Enable/Disable Emergency Button (shipping mode)

  •    To enable the Emergency button, press & hold the RED button on the pendant for 15
       seconds. The pendant will flash RED to indicate the feature is enabled.
  •    To disable the Emergency button (or put pendent into shipping mode), press & hold the
       WHITE button on the pendant for 15 seconds. The pendant will flash BLUE to indicate the
       feature is disabled.


F. Electrical Specifications

  •    Frequency: 433.92 Mhz
  •    Battery: CR2032, 3V 230mAh
  •    Battery life: Typical life expectancy of 5 yrs at an average use rate of one activation per day.
  •    Open Field Range: Approximately 600 Ft.



  FCC Statement
  This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device
  may not cause harmful interference, and (2) This device must accept any interference received, including interference that
  may cause undesired operation.

  FCC Caution
  Changes or modifications made to this IC 3000 Pendant that are not expressly approved by Instant Care could void the
  user’s authority to operate this equipment.



Document Created: 0480-07-05 00:00:00
Document Modified: 0480-07-05 00:00:00

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC