RF Exposure SAR Info

FCC ID: XNL-ORTHOSNSR6

RF Exposure Info

Download: PDF
FCCID_2164413

                                                             RF Exposure
                                                                   For the

                                                           OrthoSensor, Inc.

                                                VERASENSE Mako Restoris PKA



                                                           January 08, 2014

                                                              Prepared for:

                                                           OrthoSensor, Inc.

                                                 1855 Griffin Road, Suite A-310

                                                       Dania Beach, FL 33004



Prepared By:                                                                             Reviewed By:

H.B. Compliance Solutions

5005 S. Ash Avenue, Suite # A-10

Tempe, Arizona 85282                                                                     Hoosamuddin Bandukwala




                                                                                        Cert # ATL-0062-E




Certificates and reports shall not be reproduced except in full, without the written permission of H.B Compliance Solutions, LLC.


                             Standalone SAR Test Exclusion Consideration


According to KDB447498D01 General RF Exposure Guidance v05 4.2.4 (Transmitters implanted
in the body of a user) When the aggregate of the maximum power available at the antenna port
and radiating structures of an implanted transmitter, under all operating circumstances, is ≤
1.0mW, SAR test exclusion may be applied. The maximum available output power requirement
and worst case operating conditions must be supported by power measurements results and
fully justified in a SAR analysis report, in lieu of the SAR measurement or numerical simulation
according to design and implementation requirements of the device.

Maximum peak output power at antenna input terminal = 1.23 (dBm)
Maximum peak output power at antenna input terminal = 1.33 (mW)
Antenna gain (typical) = 1.9(dBi)
Prediction frequency = 404.34 (MHz) or 0.4 (GHZ)


To solve for the EIRP ;

[(max. power of channel, including tune-up tolerance, mW)/(min. test separation distance, mm)] · [√f(GHz)]

General RF Exposure = (3.23 / 5 mm ) x √0.4GHz = 0.408 ①

SAR requirement:

S = 3.0 ② ;

① < ②.

Therefore SAR report is not required.




END OF TEST REPORT




SAR Exclusion Test Report # EMC_13031                                                                    Page 2 of 2


                                                  )A
                                O RT HO©SENS OR
                                    DELIVERING INTELLIGENT ORTHOPEDICS

                            1855 Griffin Road, Ste A—310, Dania Beach, FL 33004—2200
                                   Telephone; 954—577—7770, Fax: 954.577.7343


                                          RF Exposure Statement



Date: 01/07/2014

American Certification Body, Inc.
6731 Whittier Avenue
Suite C110
McLean, VA 22101


Regarding: RF Exposure
FCC ID: XNL—ORTHOSNSR6



To Whom It May Concern:

Based upon output power, the implantable devise qualifies for SAR exclusion per Section 4.2.4 of
KDB publication 447498.


Thank you,



                 /                            j—g . zor4
By:          ;                                                         Juan Fernandez
                        (         e‘)                                   (Print name)

Title:           EVP of Operations and Corporate Compliance Officer


On behalf of: OrthoSensor, INC

Telephone:       954—577—7770



Document Created: 2014-01-13 11:15:23
Document Modified: 2014-01-13 11:15:23

© 2024 FCC.report